Regl. 6303, art. 7.6
Report results as ppm by volume, dry basis.
Length: 2,808 wordsOfficial source
Cite as Reglamento Núm. 6303, Art. 7.6
7.5 Reporting of Results. At the completion of the field analysis portion of
the study, ensure that the data sheets shown in Figure 18-11 have been completed.
Summarize this data on the data sheets shown in Figure 18-15.
7.6 Recovery Study. After conducting the presurvey and identifying all of the
pollutants of interest, conduct the appropriate recovery study during the test
based on the sampling system chosen for the compounds of interest.
7.6.1 Recovery Study for Direct Interface or Dilution Interface Sampling. If
the procedures in Section 7.2 or 7.3 are to be used to analyze the stack gas,
conduct the calibration procedure as stated in Section 7.2.2 or 7.3.2, as
appropriate. Upon successful completion of the appropriate calibration
procedure, attach the mid-level calibration gas for at least one target compound
to the inlet of the probe or as close as possible to the inlet of the probe, but
before the filter. Repeat the calibration procedure by sampling and analyzing
the mid-level calibration gas through the entire sampling and analytical system
until two consecutive samples are within 5 percent of their mean value. The mean
of the calibration gas response directly to the analyzer and the mean of the
calibration gas response sampled through the probe shall be within 10 percent of
each other. If the difference in the two means is greater than 10 percent, check
for leaks throughout the sampling system and repeat the analysis of the standard
through the sampling system until this criterion is met.
7.6.2 Recovery Study for Bag Sampling. Follow the procedures for the bag
sampling and analysis in Section 7.1. After analyzing all three bag samples,
choose one of the bag samples and analyze twice more (this bag will become the
spiked bag). Spike the chosen bag sample with a known mixture (gaseous or
liquid) of all of the target pollutants. Follow a procedure similar to the
calibration standard preparation procedure listed in Section 6.2, as appropriate.
The theoretical concentration, in ppm, of each spiked compound in the bag shall
be 40 to 60 percent of the average concentration measured in the three bag
samples. If a target compound was not detected in the bag samples, the
concentration of that compound to be spiked shall be 5 times the limit of
detection for that compound. Analyze the bag three times after spiking.
Calculate the average fraction recovered (R) of each spiked target compound with
the following equation:
R = t-u
S
EMTIC M-18
EMTIC NSPS Test Method
Page 18
7.4.1.7 Barometer. Accurate to 5 mm Hg, to measure atmospheric pressure during
sampling and pump calibration.
7.4.1.8 Rotameter. O to 100 cc/min, to detect changes in flow rate during
sampling.
7.4.2 Sampling and Analysis. It is suggested that the tester follow the
sampling and analysis portion of the respective NIOSH method section entitled
"Procedure." Calibrate the pump and limiting orifice flow rate through
adsorption tubes with the bubble tube flowmeter before sampling. The sample
system can be operated as a "recirculating loop" for this operation. Record the
ambient temperature and barometric pressure. Then, during sampling, use the
rotameter to verify that the pump and orifice sampling rate remains constant.
Use a sample probe, if required, to obtain the sample at the centroid of the
duct, or at a point no closer to the walls than 1 m. Minimize the length of
flexible tubing between the probe and adsorption tubes. Several adsorption tubes
can be connected in series, if the extra adsorptive capacity is needed. Provide
the gas sample to the sample system at a pressure sufficient for the limiting
orifice to function as a sonic orifice. Record the total time and sample flow
rate (or the number of pump strokes), the barometric pressure, and ambient
temperature. Obtain a total sample volume commensurate with the expected
concentration of the volatile organic(s) present, and recommended sample
loading factors (weight sample per weight adsorption media). Laboratory tests
prior to actual sampling may be necessary to predetermine this volume. When more
than one organic is present in the emissions, then develop relative adsorptive
capacity information. If water vapor is present in the sample at concentrations
above 2 to 3 percent, the adsorptive capacity may be severely reduced. Operate
the gas chromatograph according to the manufacturer's instructions. After
establishing optimum conditions, verify and document these conditions during all
operations. Analyze the audit samples (see Section 7.4.4.3), then the emission
samples. Repeat the analysis of each sample until the relative deviation of two
consecutive injections does not exceed 5 percent.
7.4.3 Standards and Calibration. The standards can be prepared according to the
respective NIOSH method. Use a minimum of three different standards; select the
concentrations to bracket the expected average sample concentration. Perform the
calibration before and after each day's sample analyses. Prepare the calibration
curve by using the least squares method.
7.4.4 Quality Assurance.
7.4.4.1 Determine the recovery efficiency of the pollutants of interest
according to Section 7.6.
7.4.4.2 Determination of Sample Collection Efficiency. For the source samples,
analyze the primary and backup portions of the adsorption tubes separately. If
the backup portion exceeds 10 percent of the total amount (primary and back-up),
repeat the sampling with a larger sampling portion.
7.4.4.3 Analysis Audit. Immediately before the sample analyses, analyze the two
audits in accordance with Section 7.4.2. The analysis audit shall agree with the
audit concentration within 10 percent.
7.4.4.4 Pump Leak Checks and Volume Flow Rate Checks. Perform both of these
EMTIC M-18
EMTIC NSPS Test Method
Page 17
system to provide the desired dilution factors. Make this correction by diluting
a high-concentration standard gas mixture to adjust the dilution ratio as
required.
Once the dilution system and GC operations are satisfactory, proceed with the
analysis of source gas, maintaining the same dilution settings as used for the
standards. Repeat the analyses until two consecutive values do not vary by more
than 5 percent from their mean value are obtained.
Repeat the analysis of the calibration gas mixtures to verify equipment
operation. Analyze the two field audit samples using either the dilution system,
or directly connect to the gas sampling valve as required. Record all data and
report the results to the audit supervisor.
7.3.3 Determination of Stack Gas Moisture Content. Same as Section 7.2.3.
7.3.4 Quality Assurance. Same as Section 7.2.4.
7.3.5 Emission Calculations. Same as section 7.2.5, with the dilution factor
applied.
7.4 Adsorption Tube Procedure (Alternative Procedure). It is suggested that the
tester refer to the National Institute for Occupational Safety and Health (NIOSH)
method for the particular organics to be sampled. The principal interferent will
be water vapor. If water vapor is present at concentrations above 3 percent,
silica gel should be used in front of the charcoal. Where more than one compound
is present in the emissions, then develop relative adsorptive capacity
information.
7.4.1 Additional Apparatus. In addition to the equipment listed in the NIOSH
method for the particular organic(s) to be sampled, the following items (or
equivalent) are suggested.
7.4.1.1 Probe (Optional). Borosilicate glass or stainless steel, approximately
6-mm ID, with a heating system if water condensation is a problem, and a filter
(either in-stack or out-stack heated to stack
temperature) to remove particulate matter. In most instances, a plug of glass
wool is a satisfactory filter.
7.4.1.2 Flexible Tubing. To connect probe to adsorption tubes. Use a material
that exhibits minimal sample adsorption.
7.4.1.3 Leakless Sample Pump. Flow controlled, constant rate pump, with a set
of limiting (sonic) orifices to provide pumping rates from approximately 10 to
100 cc/min.
7.4.1.4 Bubble-Tube Flowmeter. Volume accuracy within +1 percent, to calibrate
pump.
7.4.1.5 Stopwatch. To time sampling and pump rate calibration.
7.4.1.6 Adsorption Tubes. Similar to ones specified by NIOSH, except the
amounts of adsorbent per primary/backup sections are 800/200 mg for charcoal
tubes and 1040/260 mg for silica gel tubes. As an alternative, the tubes may
contain a porous polymer adsorbent such as Tenax GC or XAD-2.
EMTIC M-18
EMTIC NSPS Test Method
Page 16
7.3.1 Apparatus. The equipment required in addition to that specified for the
direct interface system is as follows:
7.3.1.1 Sample Pump. Leakless Teflon-coated diaphragm-type that can withstand
being heated to 120°C and deliver 1.5 liters/minute.
7.3.1.2 Dilution Pumps. Two Model A-150 Komhyr Teflon positive displacement
type delivering 150 cc/minute, or equivalent. As an option, calibrated
flowmeters can be used in conjunction with Teflon-coated diaphragm pumps.
7.3.1.3 Valves. Two Teflon three-way valves, suitable for connecting to 6.4-mm
OD Teflon tubing.
7.3.1.4 Flowmeters. Two, for measurement of diluent gas, expected delivery flow
rate to be 1,350 cc/min.
7.3.1.5 Diluent Gas with Cylinders and Regulators. Gas can be nitrogen or clean
dry air, depending on the nature of the source gases.
7.3.1.6 Heated Box. Suitable for being heated to 120°C, to contain the three
pumps, three-way valves, and associated connections. The box should be equipped
with quick connect fittings to facilitate connection of: (1) the heated sample
line from the probe, (2) the gas sampling valve, (3) the calibration gas
mixtures, and (4) diluent gas lines. A schematic diagram of the components and
connections is shown in Figure 18-13.
(Note: Care must be taken to leak-check the system prior to the dilutions so as
not to create a potentially explosive atmosphere.)
The heated box shown in Figure 18-13 is designed to receive a heated line from
the probe. An optional design is to build a probe unit that attaches directly
to the heated box. In this way, the heated box contains the controls for the
probe heaters, or, if the box is placed against the duct being sampled, it may
be possible to eliminate the probe heaters. In either case, a heated Teflon line
is used to connect the heated box to the gas sampling valve on the chromatograph.
7.3.2 Procedure. Assemble the apparatus by connecting the heated box, shown in
Figure 18-13, between the heated sample line from the probe and the gas sampling
valve on the chromatograph. Vent the source gas from the gas sampling valve
directly to the charcoal filter, eliminating the pump and rotameter. Heat the
sample probe, sample line, and heated box. Insert the probe and source
thermocouple at the centroid of the duct, or to a point no closer to the walls
than 1 m. Measure the source temperature, and adjust all heating units to a
temperature 0 to 3°C above this temperature. If this temperature is above the
safe operating temperature of the Teflon components, adjust the heating to
maintain a temperature high enough to prevent condensation of water and organic
compounds. Verify the operation of the dilution system by analyzing a high
concentration gas of known composition through either the 10:1 or 100:1 dilution
stages, as appropriate. (If necessary, vary the flow of the diluent gas to obtain
other dilution ratios.) Determine the concentration of the diluted calibration
gas using the dilution factor and the calibration curves prepared in the
laboratory. Record the pertinent data on the data sheet shown in Figure 18-11.
If the data on the diluted calibration gas are not within 10 percent of the
expected values, determine whether the chromatograph or the dilution system is
in error, and correct it. Verify the GC operation using a low concentration
standard by diverting the gas into the sample loop, bypassing the dilution
system. If these analyses are not within acceptable limits, correct the dilution
EMTIC M-18
EMTIC NSPS Test Method
Page 15
7.2.1.9 Charcoal Adsorber. To adsorb organic vapor collected from the source
to prevent exposure of personnel to source gas.
7.2.1.10 Gas Cylinders. Carrier gas (helium or nitrogen), and oxygen and
hydrogen for a flame ionization detector (FID) if one is used.
7.2.1.11 Gas Chromatograph. Capable of being moved into the field, with
detector, heated gas sampling valve, column required to complete separation of
desired components, and option for temperature programming.
7.2.1.12 Recorder/Integrator. To record results.
7.2.2 Procedure. To obtain a sample, assemble the sampling system as shown in
Figure 18-12. Make sure all connections are tight. Turn on the probe and sample
line heaters. As the temperature of the probe and heated line approaches the
source temperature as indicated on the thermocouple readout device, control the
heating to maintain a temperature of O to 3°C above the source temperature.
While the probe and heated line are being heated, disconnect the sample line from
the gas sampling valve, and attach the line from the calibration gas mixture.
Flush the sample loop with calibration gas and analyze a portion of that gas.
Record the results. After the calibration gas sample has been flushed into the
GC instrument, turn the gas sampling valve to flush position, then reconnect the
probe sample line to the valve. Place the inlet of the probe at the centroid of
the duct, or.at a point no closer to the walls than 1 m, and draw source gas into
the probe, heated line, and sample loop. After thorough flushing, analyze the
sample using the same conditions as for the calibration gas mixture. Repeat the
analysis on an additional sample. Measure the peak areas for the two samples,
and if they do not agree to within 5 percent of their mean value, analyze
additional samples until two consecutive analyses meet this criteria. Record the
data. After consistent results are obtained, remove the probe from the source
and analyze a second calibration gas mixture. Record this calibration data and
the other required data on the data sheet shown in Figure 18-11, deleting the
dilution gas information.
(NOTE: Take care to draw all samples, calibration mixtures, and audits through
the sample loop at the same pressure.)
7.2.3 Determination of Stack Gas Moisture Content. Use Method 4 to measure the
stack gas moisture content.
7.2.4 Quality Assurance. Same as Section 7.1.7. Introduce the audit gases in
the sample line immediately following the probe.
7.2.5 Emission Caloulations. Same as Section 7.1.8.
7.3 Dilution Interface Sampling and Analysis Procedure. Source samples that
contain a high concentration of organic materials may require dilution prior to
analysis to prevent saturating the GC detector. The apparatus required for this
direct interface procedure is basically the same as that described in the Section
7.2, except a dilution system is added between the heated sample line and the gas
sampling valve. The apparatus is arranged so that either a 10:1 or 100:1
dilution of the source gas can be directed to the chromatograph. A pump of larger
capacity is also required, and this pump must be heated and placed in the system
between the sample line and the dilution apparatus.
EMTIC M-18
EMTIC NSPS Test Method
Page 14
Eq. 18-5
where:
Cₛ
= Concentration of the organic from the calibration curve, ppm.
Pr
= Reference pressure, the barometric pressure or absolute sample
loop pressure recorded during calibration, mm Hg.
T1
= Sample loop temperature at the time of sample analysis, °K.
Fr
= Relative response factor (if applicable, see Section 6.4).
Pa
= Barometric or absolute sample loop pressure at time of sample
analysis, mm Hg.
Ti
= Reference temperature, the temperature of the sample loop
recorded during calibration, °K.
Bws
= Water vapor content of the bag sample or stack gas, proportion by
volume.
7.2 Direct Interface Sampling and Analysis Procedure. The direct interface
procedure can be used provided that the moisture content of the gas does not
interfere with the analysis procedure, the physical requirements of the equipment
can be met at the site, and the source gas concentration is low enough that
detector saturation is not a problem. Adhere to all safety requirements with
this method.
7.2.1 Apparatus.
7.2.1.1 Probe. Constructed of stainless steel, Pyrex glass, or Teflon tubing
as required by duct temperature, 6.4-mm OD. enlarged at duct end to contain glass
wool plug. If necessary, heat the probe with heating tape or a special heating
unit capable of maintaining duct temperature.
7.2.1.2 Sample Lines. 6.4-mm OD Teflon lines, heat-traced to prevent
condensation of material.
7.2.1.3 Quick Connects. To connect sample line to gas sampling valve on GC
instrument and to pump unit used to withdraw source gas. Use a quick connect or
equivalent on the cylinder or bag containing calibration gas to allow connection
of the calibration gas to the gas sampling valve.
7.2.1.4 Thermocouple Readout Device. Potentiometer or digital thermometer, to
measure source temperature and probe temperature.
7.2.1.5 Heated Gas Sampling Valve. Of two-position, six-port design, to allow
sample loop to be purged with source gas or to direct source gas into the GC
instrument.
7.2.1.6 Needle Valve. To control gas sampling rate from the source.
7.2.1.7 Pump. Leakless Teflon-coated diaphragm-type pump or equivalent, capable
of at least 1 liter/minute sampling rate.
7.2.1.8 Flowmeter. Of suitable range to measure sampling rate.
EMTIC M-18
EMTIC NSPS Test Method
Page 13
concentration.
7.1.5 Analysis of Bag Samples.
7.1.5.1 Apparatus. Same as Section 5. A minimum of three gas standards are
required.
7.1.5.2 Procedure. Establish proper GC operating conditions as described in