Regl. 2480, art. 8(c)(3)
of the Act (29 U.S.C.
Cite as Reglamento Núm. 2480, Art. 8(c)(3)
polymers made from AN except for
out regard to the use of respiratory
657) requires employers to notify
materials exempted under paragraph
protection. Section 6(b)(7) of the Act
promptly all employees exposed above
(a)(2).
(29 U.S.C. 655) mandates that any
the PEL. The final standard imple-
This proposed definition of "PAN"
standard promulgated under section
ments section 8(c)(3) by requiring an
has been deleted from the final stand-
6(b) shall, where appropriate, provide
employer to notify each employee in
ard for several reasons. The standard
for monitoring or measuring employee
writing of the employee's measurewas and is intended to protect employexposure at such locations and interment within 5 working days after the
FEDERAL REGISTER, VOL. 43, NO. 192-TUESDAY, OCTOBER 3, 1978
45796
RULES AND REGULATIONS
receipt of the results of any required
to the ETS, OSHA was most conand precision requirements set forth
measurements.
cerned about this situation, particulartherein. The testing that OSHA has
OSHA recognizes that the standard
ly in light of the industry recommenperformed, supplemented by the
is intended to set only minimum monidations at that time that the standard
NIOSH data, serves to demonstrate
toring requirements for employers. It
be set as low as 2 ppm. Therefore, testthat there is at least one method availis reasonable, then, that the standard
ing was performed at the OSHA Anaable which meets those requirements.
should only require the notification to
lytical Laboratory in Salt Lake City,
We are fully aware that there are
employees when monitoring required
Utah, to determine whether NIOSH
many other methods in use throughby the standard is carried out. Addimethod S-156, which was the method
out industry to sample for AN, as well
tional actions by the employer in monused under the existing 20 ppm standas methods which are in varying
itoring exposures, while certainly enard, could also be used in sampling
stages of development. Such methods
couraged by OSHA, are not required
lower concentrations (see appendix D).
include the use of porous polymer
by the standard, and direct notifica-
The test results clearly showed that
tubes instead of charcoal tubes for coltion need not be given to employees.
S-156 could be used for sampling and
lecting the sample, gas bags, badge-
The standard requires that measureanalysis below 1 ppm with sufficient
type dosimeters, and the use of portaments be made by monitoring which is
accuracy (Exs. 18; 18A). Testing was
ble and fixed-point sampling devices
representative of each employee's experformed in the presence of high huand systems, Considerable data has
posure to AN over and 8-hour period
midity (85 percent) and other monbeen submitted to the record by many
without regard to respiratory protecomers, such as styrene and butadiene,
companies concerning their experience
tion. Exposure measurements for each
to determine whether these elements
with these methods. By demonstrating
individual employee would, certainly,
had any detrimental effect on the
the availability and effectiveness of S-
be an indication of that employee's exsampling and analysis. The effects of
156 as modified, we do not intend to
posure. However, this may be unnecesthese factors was negligible. The Dishort-circuit any of the efforts by
sarily burdensome in some instances.
rector of the OSHA Laboratory, Floyd
these companies to develop their own
Monitoring which can be shown to be
Madsen, appeared at the rulemaking
monitoring procedures, some of which
representative of an employee's expohearing and was questioned at length
may prove to be even more accurate.
sure would provide the necessary inby industry representatives concerning
Industry representatives have
formation and in many instances
his testing procedures and results. In
argued that it is essential for a method
would require fewer samples. This is
addition, representatives of SPI were
to be "field validated" in order to be
particularly true in manufacturing of
permitted to visit the Laboratory after
used for compliance with the AN
AN and other 24-hour continuous
the hearing. in order to talk with Mr.
standard (Exs. 75; 83). Yet it should be
processes.
Madsen further concerning sampling
noted that such representatives have
The proposal required monitoring
and analysis, in response to a direct realso stated that the range of field conmonthly for employees exposed above
quest at the hearing by both SPI and
ditions that may be encountered in
the permissible exposure limit and
Vistron. Although such a visit did,
the AN industry is so wide and variquarterly for employees exposed beindeed, go beyond what is required of
able that it would be necessary to
tween the action level and the PEL. As
the Agency during informal rulemak-
"field validate" a method in each indinoted above, many employers have ining. OSHA determined that it was, in
vidual workplace in order to show that
dicated that they perform monitoring
this limited circumstance, an appropriit was, indeed, a workable method
on a more frequent basis. It is evident
ate gesture toward developing a comacross the board (Tr. 933). It is evident
that the higher the frequency of meaplete record on the sampling issue.
that no single analytical method will
surement, the higher the accuracy of
Laboratory testing has also been
work in every imaginable workplace
the employee's exposure profile. The
performed by NIOSH at their laborasituation, either for AN or for most
companies involved with the manufactories in Cincinnati, also involving
any other organic vapor. In addition,
turing and polymerizing of AN indicatmethod S-156. Data submitted to the
it is also clear that whatever method
ed little, if any, disagreement with the
record by Dr. Judd Posner of NIOSH
an employer may choose to use to
monitoring frequency in the proposal.
also demonstrated that AN could be
monitor for AN, he must validate it
OSHA has carried forward these prosampled and analysed below 1 ppm
for use in his particular plant. The
visions to the final standard. We bewith accuracy (Exs. 43; 66). Since this
need for such individual validation was
lieve that in the industries affected by
data was submitted during the course
espoused by industry representatives
this standard, monthly and quarterly
of the hearing, Dr. Posner invited in-
(Tr. 933).
monitoring will reflect employee expoterested persons at the hearing to
The method of "field validation"
sure with sufficient accuracy to assure
submit questions to him in writing, to
proposed by Dr. Ward of Monsanto
that suitable precautions will be taken
which he would reply during the postpresents an interesting approach
as needed to protect employees from
hearing comment period. It is imporwhich shows promise (Ex. 136). Howexposure to AN.
tant to note that no such questions
ever, there is nothing to indicate that
Industry representatives have conwere received. A complete write-up of
a method is or has been in use in the
centrated much of their criticism on
the analytical method and sampling
past to "field validate" sampling methwhat they have contended to be the
technique used by NIOSH is available
ods for AN or other organic vapors. In
lack of a method of sampling and analin the record (see appendix D). In esaddition, Dr. Posner of NIOSH has
ysis for AN in the range of 1 ppm and
sence, it sets forth the same method
submitted an analysis of Dr. Ward's
below which will meet the accuracy
utilized by the OSHA Laboratory,
validation method which points up
and precision requirements set forth
with the major exceptions being the
that it, too, is not without pitfalls in
in the proposed standard. This issue
use of a different desorbing solvent
its implementation (Ex. 140). There is
was brought to the surface by the
and the use of flame ionization withno single accepted method of "field
original NIOSH recommendations in
out the nitrogen phosphorus detector.
validating" a sampling method to
September 1977 (Ex. 65). In those rec-
OSHA does not require that any one
obtain data which will support its use
ommendations, NIOSH stated that
specific method of sampling and analunder all workplace conditions.
OSHA should issue a 4 ppm standard,
ysis be used for compliance with the
A detailed exposition of the range of
based upon what they characterized as
monitoring provisions of the standard.
available methods of sampling and
the limitations of existing sampling
This provision is set forth in performanalysis for AN was contained in the
and analytical methods for AN. As we
ance language, and permits the use of
written comments submitted by Monpointed out at length in the preamble
any method which meets the accuracy
santo (Ex. 11:(27)). Table A contains a
FEDERAL REGISTER, VOL. 43, NO. 192-TUESDAY, OCTOBER 3, 1978
RULES AND REGULATIONS
45797
listing of five methods which were
recovery (with methanol or carbon dieither validated by Monsanto (field
sulfide) falls below 75 percent by deand/or laboratory validation) or, in
veloping a "working curve," in lieu of
the case of the gasbags, showed a dethe NIOSH protocol. They emphatection limit below 1 ppm. The field
sized that appropriate validation techvalidation was, presumedly, carried
niques are necessary for the use of all
out under the protocol submitted by
monitoring procedures in the work-
Ward (Ex. 136). Although Monsanto
place setting, whether charcoal tubes
reported difficulty in using NIOSH
or other means of sampling are used.
method S-156, it is clear that they
They have chosen to rely upon alterhave been able to develop or apply
native systems, such as porous polyother methods of sampling and analymer adsorption/thermal desorption,
sis to their many and varied opersince they consider them to be less
ations.
subject to the types of variables associ-
Monsanto, in their written comated with charcoal-based systems.
ments, discussed the various problems
As was noted by Madsen in his testithat they had experienced with
mony at the hearing, Monsanto con-
NIOSH method S-156 (Ex. 11:(27)).
curs that the use of a nitrogen-specific
They noted that the only charcoal lot
gas chromatographic (GC) detector
which had shown satisfactory results
has the potential of enhancing the dewas lot 107. This point was noted by
tection limit of all of the methods
Madsen in his testimony (Tr. 132). In
listed in the table by a factor of 10
dealing with the problems of sample
(Ex. 11:(27)). It is also important to
"breakthrough" reported by various
note that other methods which incompanies, including Borg-Warner
volve gas chromatographic analysis,
(Ex. 83), Monsanto noted that some
have this same potential.
companies are using two larger char-
Monsanto has also reported using
coal tubes, containing up to 1,000 mg.,
both the Century OVA 128 and Cenin series, to replace the 150-mg. tubes
tury flare tubes at levels well below 1
generally used. They have concluded,
ppm, noting only that "the methods
however, that there is no evidence of
are claimed to be selective." Century
any effect of humidity or temperature
Systems testified independently that
at low AN burden.
their instrumentation was capable of
Monsanto noted that charcoal tube
measuring AN at concentrations well
methods can be used for AN at concenbelow 1 ppm as an 8-hour TWA (Tr.
trations below 2 ppm even where AN
873-874).
FEDERAL REGISTER, VOL. 43, NO. 192-TUESDAY, OCTOBER 3, 1978
45798
RULES AND REGULATIONS
TABLE B: ADDITIONAL METHODS OF MONITORING AND ANALYSIS
FOR AN BELOW 1 PPM
(Adapted from Monsanto comments (Exhibit (27)))
Method
Adsorption/Desorption/ Validation:
Analysis
concentration, detection
limits
1. Monsanto
Poropak-N
field validated down to
PoropaK-N
(porous Thermal GC
0.1 ppm; detection limit
polymer)
approaching 0.02 ppm for
a 3-liter air sample.
2. Modified
1 to 2%
field validated at
NIOSH
Charcoal Acetone in/GC
levels above 0.08 ppm
P&CAM 127
CSâ‚‚
within ± 2 std. devs.
for a TWA6.
3. NIOSH
laboratory and field
P&CAM 127
Charcoal/CS2/GC
validated at and above
0.2 ppm for a TWAS;
Reported loss of AN when
desorbed sample is
stored over time.
4. Passive Dosimetry Abcor R Acetone / GC laboratory validated 0.94
Gasbadge/in CSâ‚‚/
ppm; field tested at
0.8-3.8 ppm by reference
to method no. 2 above.
5. Gas sample bags
Gas bag/GC
Detection limit of
ca a 2.5 CC. gas sample.
Reported memory effects
at low AN levels.
FEDERAL REGISTER, VOL. 43, NO. 192-TUESDAY, OCTOBER 3, 1978
RULES AND REGULATIONS
45799
Monsanto was confident that "the
The monitoring defined by the
is certainly one approach toward decharcoal tube or passive dosimeter
OSHA Laboratory involves the use of
termining representativeness of expomethods could be practiced by downcharcoal tubes and personal sampling
sure measurements. There are other
stream processors with some guidance
pumps, with subsequent analysis by
approaches which would be suitable,
from trained professional" (Ex.
gas chromatograph with nitrogen
as well. The employer has the burden
11:(27)). The record reflects that the
phosphorus detector. Laboratory valiof showing, by whatever means at his
employers covered by this standard
dation data indicates that this method
disposal, that his measurements are
possess considerably more technical
can be used with sufficient accuracy
representative of each employee's excapability in this area than downand precision down to the 0.5-ppm
posure to AN.
stream employers would be expected
level and below. Floyd Madsen was
The rulemaking record, along with
to have. The broadening of the exempquestioned extensively at the hearing
experience under the ETS, indicate
tion provisions will limit the scope of
as to his findings by industry reprethat most affected employers have
the standard to those companies best
sentatives (Tr. 117-159). In addition,
successfully implemented a monitorable to provide an effective monitoring
several industry personnel visited
ing method which gives satisfactory
program which will produce meaning-
Madsen at the Laboratory during the
ful results.
results in their operations, at concenposthearing comment period (Ex. 142).
trations in the 1- to 2-ppm range. How-
Companies such as Goodyear, Dow
Neither the questioning at the hearing
Badische, and Dow Chemical have
ever, OSHA cannot ignore the difficulnor the laboratory visit have raised
submitted details of the sampling and
ties that various companies have reany significant questions as to Madanalytical methods that they have
ported in demonstrating the required
sen's findings.
found valid for use in their respective
accuracy and precision specified in
The primary differences between
operation (Exs. 138; 11:(17); 11:(25)).
both the ETS and proposal at a (95
the NIOSH and OSHA variations on
Still others have installed continuous
percent confidence level, ±25 percent
S-156 are the method of desorption
monitoring systems which provide
at or above 2 ppm, and ±35 percent
(NIOSH does not use ultrasonic agitaarea measurements of airborne AN
below 2 ppm) (Exs. 83; 159). We have
tion of the sample), the desorbing solconcentrations (Exs. 11:(10); 11:(8)).
determined that these criteria should
vent (NIOSH use 1 percent acetone in
As the record makes clear, there are
carbon disulfide, whereas OSHA uses
be made more flexible, in order to
a number of adequate methods availamethanol), and the analytical equipassist employers who have encounble for sampling and analysis of AN
ment (NIOSH uses flame ionization,
tered unique sampling problems.
concentrations at and below 1 ppm.
whereas OSHA adds a nitrogen-phos-
Therefore, the final standard permits
OSHA recognizes that one method
phorus detector).
the use of methods with accuracy and
may not work under all workplace con-
Method S-156 and variations thereprecision at a 95-percent confidence
ditions due to such variables as temon have been adopted and accepted for
level, of ±35 percent at or above 2
perature, humidity, and other unique
general use in field monitoring at the
ppm, and ±50 percent below 1 and 2
factors. Regardless of the method se-
1- to 2-ppm level by several major comppm. The difference (35 percent VS. 50
lected, an employer must validate the
panies that manufacture and/or use
percent) reflects a recognition that acmethod that he intends to use under
AN in their facilities. Workplace concuracy of monitoring decreases below
the conditions present in the particuditions and other factors vary widely
2 ppm.
lar workplace to assure himself that it
among these companies (cf. Ex. 2:(22);
The accuracy and precision requirewill be sufficiently accurate. It is for
11:(14); 138; Tr. 1196-1199).
ments are intended to provide the emthis reason that the monitoring accu-
The standard requires that the meaployer with a degree of confidence in
racy and precision required by the
surements be made by monitoring
his sampling results. As noted earlier,
standard are stated in performance
which is representative of each emmonitoring is not carried out for its
terms. The OSHA and NIOSH methployee's exposure to AN over an 8-
own sake, but for the purposes of deods of monitoring employee exposures
hour period without regard to the use
termining what is required for compliare included as an appendix to the
of respiratory protection. Exposure
ance in a given operation. OSHA anfinal standard as a means of assisting
measurements for each individual emticipates that the wide range of availathe employer in the selection of an apployee would, of course, be an indicable methods, combined with more
propriate method for his workplace.
tion of that employee's exposure.
flexible criteria for the method em-
It was also contended that because
However, this may be unnecessarily
ployed, will enable employers to perneither the OSHA nor the NIOSH
burdensome in some instances. Moniform their required monitoring withvariations of method S-156 had been
toring which is truly representative of
out serious difficulties.
"field validated" under all workplace
an employee's exposure would provide
The employee exposure measureinsults and exposure conditions, they
the necessary information and in
ments are to be made without regard
could not serve as reference methods.
many instances would involve fewer
to the use of respiratory protection.
This assertion totally misconstrues the
samples.
One commentator took issue with this
true purpose of a so-called "reference
In their testimony, the NIOSH panel
provision, contending that the sammethod." The applicability of any
expressed concern that OSHA did not
pling does not reflect the actual expo-
"reference method" was described
specify a means of determining if an
sure of the employee who is being promost accurately at the hearing by Dr.
exposure measurement was "repretected by a respirator (Ex. 149).
Ward:
sentative" of an employee's exposure.
Whereas this statement may be true,
There is always a need for a reference
They believed that this presented a
it overlooks the fact that exposure
method. However, the reference method
potential for actually reducing the
monitoring is not a single-purpose accannot serve under all applications. It can
protection afforded to workers extivity. It is necessary to know employserve under laboratory conditions and it has
posed to AN. OSHA believes that the
ee exposure levels without the use of
merit in that respect (Tr. 791).
standard should require the desired
respiratory protection in order to
OSHA and NIOSH have both perperformance, i.e., a showing of repreevaluate the effectiveness of the reformed laboratory testing of their
sentativeness, rather than specifying
quired engineering and work practice
variations of NIOSH method S-156
one method for determining it. We
controls, and to determine whether
and have found the method to be a
recognize, however, that NIOSH has
additional controls must be instituted.
suitable reference method for use in
published a document outlining a sam-
In addition, monitoring is necessary to
the range of concentrations covered by
pling strategy which is directed at this
determine which, if any, respirator
the standard (Exs. 18; 18A; 43; 66).
problem (Ex. 14:(157)). That strategy
must be used by the employee.
PEDERAL REGISTER, VOL. 43, NO. 192-TUESDAY, OCTOBER 3, 1978
45800
RULES AND REGULATIONS
As noted by numerous participants
those operations where exposures
The alarm system required by parain this rulemaking proceeding, expowould be well below the permissible
graph (i) of the standard will detect
sure conditions vary throughout the
exposure limit.
AN concentrations which exceed the
day in many operations. At least one
The standard requires that whenceiling limit. When an alarm is soundsample is to be taken during each shift
ever there has been a production,
ed, the employer will have to take
in order to assure that exposure meaprocess, or control change which may
action to locate the source of the resurements represent exposures of emresult in new or additional exposures
lease which is causing the high levels
ployees on all shifts. Employees workto AN, or whenever the employer has
of AN, and will then have to deconing in the same area doing different
reason to suspect an increase in emtaminate the leak or spill area. In this
jobs may have different exposures.
ployee exposure, the employer shall
regard, the weekly survey, as pro-
Therefore, the standard requires samrepeat the required monitoring and
posed, would be redundant.
pling for each job classification, as
measurements for those employees af-
OSHA has determined, however,
well. The proposed standard did not
fected by such change or increase.
that some form of leak detection procontain detailed procedures in this
The redetermination procedure is
gram should be required by the standregard. However, in order to assure acnecessary to assure that the most
ard. The details of such a program,
curate representative monitoring of
recent monitoring results accurately
however, should be left to the employemployee exposures, OSHA has deterrepresent the existing exposure condier, who is already under the obligation
mined that the variability of expotions. This is necessary so that the emto control employee exposures to
sures warrants such procedures.
ployer may take appropriate actions
within the permissible exposure limits.
The standard requires that the inisuch as instituting additional engi-
The leak detection program in the
tial monitoring be conducted, and the
neering controls and providing approfinal standard has been placed in the
results thereof obtained, as of the efpriate respiratory protection.
housekeeping section of the standard,
fective date of the standard. OSHA be-
The final standard also provides
paragraph (k).
lieves that this provides more than
that employers may discontinue peri-
Regulated areas and notification:
enough time for employers to comodic monitoring for those employees
Paragraphs (d) and (f). The final
plete the sampling and obtain the refor whom two consecutive measurestandard requires the employer to essults. Moreover, the standard permits
ments, taken at least 7 days apart,
tablish regulated areas wherever AN
employers who have monitored under
show exposures to be below the action
exposures exceed the PEL, and to
the ETS, or since January 17, 1978, to
level. Where employee exposure mealimit access to these areas to authorutilize these measurements for pursurements fall below the permissible
ized persons. This requirement will
poses of complying with the initial
exposure limits but are at or above the
help to limit exposure to AN by premonitoring requirements of the final
action level, the employer may alter
venting persons who are not authorstandard. Employers who have already
the monitoring schedule for those emized from entering areas with AN exmonitored their employees' exposures
ployees from monthly to quarterly
posures above 2 ppm. It will also assist
within this period will, therefore, not
after two consecutive measurements
in designating areas where precautionhave to conduct initial monitoring
taken at least 7 days apart indicate
ary signs are to be posted, while also
unless, because of a process, control,
the reduction in levels.
designating areas where employees
or personnel change, they are required