Regl. 3215, art. 4.4
be used to validate the analytical data.
Length: 616 wordsOfficial source
Cite as Reglamento Núm. 3215, Art. 4.4
[4.1.5 Note amended by 48 FR 49459,
October 25. 1983]
4.2 Single-Point. Integrated Sampling
and Analytical Procedure.
4.2.1 The sampling point in the duct
shall be located as specified in Section 4.1.1.
4.2.2 Leak-check (mandatory) the nexible bag as in Section 2.2.6. Set up the equipment as shown in Figure 3-2. Just prior to
sampling. leak-check (mandatory) the train
by placing a vacuum gauge at the condenser
inlet, pulling a vacuum of a least 250 mm Hg
(10 in. Hg). plugging the outlet at the quick
disconnect, and then turning off the pump.
The vacuum shall remain stable for at least
0.5 minute. Evacuate the flexible bag. Connect the probe and place It in the stack,
with the tip of the probe positioned at the
sampling point; purge the sampling line.
Next, connect the bag and make sure that
all connections are tight and leak free.
4.2.3 Sample at a constant rate. or as
specified by the Administrator. The sampling run must be simultaneous with, and
for the same total lengh of time as, the pollutant emission rate determination. Collect
at least 30 liters (1.00 ft') of sample gas.
Smaller volumes may be collected, subject
to approval of the Administrator.
4.2.4 Obtain one integrated flue gas
sample during each pollutant emission rate
determination. For emission rate correction
factor determination, analyze the sample
within 4 hours after it is taken for percent
CO, or percent O, (as outlined in Sections
4.2.5 through 4.2.7). The Orsat analyzer
must be leak-check (see Section 5) before
the analysis. If excess air is desired. proceed
as follows: (1) within 4 hours after the
sample is taken, analyze it (as in Sections
4.2.5 through 4.2.7) for percent CO2. O,, and
CO: (2) determine the percentage of the gas
that is N, by subtracting the sum of the percent CO., percent O., and percent CO from
100 percent: (3) calculate percent excess air,
as outlined in Section 6.2.
4.2.5 To insure complete absorption of
the CO2. O,, or if applicable, CO. make repeated passes through each absorbing solution until two consecutive readings are the
same. Several passes (three of four) should
be make between readings. (If constant
readings cannot be obtained after three consecutive readings. replace the absorbing SO.
lution.)
4.2.6 Repeat the analysis until the following criteria are met:
4.2.6.1 For percent CO2. repeat the analytical procedure until the results of any
three analyses differ by no more that (a) 0.3
percent by volume when CO2 is greater than
4.0 percent or (b) 0.2 percent by volume
when CO. is less than or equal to 4.0 percent. Average the three acceptable values of
percent CO, and report the results to the
nearest 0.1 percent.
121:1574.1
4.2.6.2 For percent O,, repeat the analytical procedure until the results of any three
analyses differ by no more than (a) 0.3 percent by volume when O, is less than 15.0
percent or (b) 0.2 percent by volume when
O2 is greater than or equal to 15.0 percent.
Average the three acceptable values of percent O, and report the results to the nearest
0.1 percent.
4.2.6.3 For percent CO. repeat the analytical procedure until the results of any
three analyses differ by no more than 0.3
percent. Average the three acceptable
values of percent CO and report the results
to the nearest 0.1 percent.
4.2.7 After the analysis is completed.
leak-check (mandatory) the Orsat analyzer
once again. as described in Section 5. For
the results of the analysis to be valid, the
Orsat analyzer must pass this leak test
before an after the analysis.
NOTE: Although in most instances only
CO, or O2 is required, it is recommended
that both CO2 and O, be measured, and that