Regl. 3497, art. 5.2.3
Then calculate the F. factor as
Length: 2,529 wordsOfficial source
Cite as Reglamento Núm. 3497, Art. 5.2.3
follows:
0.209 F.
F.-
Eq. 3-4
F.
Fuel type
F. range
Cost:
Anthracite and lignite
1.016-1.130
Bituminous
1.063-1.220
OIL
Distillate
1.260-1.413
Residual
1.210-1.370
Gas:
Natural
1.600-1.836
Propane
1.434-1.588
Butane
1.406-1.553
Wood
1.000-1.120
Wood bark
1.003-1.130
Calculated F. values beyond the
acceptable ranges shown in this table should
be investigated before accepting the test
results. For example. the strength of the
solutions in the gas analyzer and the
analyzing technique should be checked by
sampling and analyzing a known
concentration. such as air: the fuel factor
should be reviewed and verified. An
acceptability range of ±12 percent is
appropriate for the F. factor of mixed fuels
with variable fuel ratios. The level of the
emission rate relative to the compliance level
should be considered in determining if a
retest is appropriate, i.e.. if the measured
emissions are much lower or much greater
than the compliance limit, repetition of the
test would not significantly change the
compliance status of the source and would be
unnecessarily time-consuming and costly.
5. Leak-Check Procedure for Orsat Analyzers
Moving an Orsat analyzer frequently
causes it to leak. Therefore, an Orsat analyzer should be throughly leak-checked on
site before the flue gas sample is introduced
into it. The procedure for leak-checking an
Orsat analyzer is:
5.1.1 Bring the liquid level in each pipette up to the reference mark on the capillary tubing and then close the pipette stopcock.
5.1.2 Raise the leveling bulb sufficiently
to bring the confining liquid meniscus onto
the graduated portion of the burette and
then close the manifold stopcock.
5.1.3 Record the meniscus position.
5.1.4 Observe the menicus in the burette
and the liquid level in the pipette for movement over the next 4 minutes.
5.1.5 For the Orsat analyzer to pass the
leak-check, two conditions must be met.
5.1.5.1 The liquid level in each pipette
must not fall below the bottom of the capillary tubing during this 4-minute interval.
5.1.5.2 The meniscus in the burette must
not change by more than 0.2 ml during this
4-minute interval.
5.1.6 If the analyzer fails the leak-check
procedure. all rubber connections and stopcocks should be checked until the cause of
the leak is identified. Leaking stopcocks
must be disassembled. cleaned, and re.
greased. Leaking rubber connections must
be replaced. After the analyzer is reassembled, the leak-check procedure must be repeated.
STATIONARY SOURCES
S-700
121:1576.3
6. Calculations
6.1 Nomenclature.
Mc=Dry molecular weight, g/g-mole (lb/lbmole).
Percent excess air.
%CO,=Percent CO2 by volume (dry basis).
%O,=Percent O₂ by volume (dry basis).
%CO Percent CO by volume (dry basis).
Percent N, by volume (dry basis).
0.264 = Ratio of O2 to N, in air, V/V.
0.280 = Molecular weight of N2 or CO, divided by 100.
0.320 = Molecular weight of O, divided by
100.
Molecular weight of CO, divided by
100.
6.2 Percent Excess Air. Calculate the percent excess air (if applicable). by substituting the appropriate values of percent O2.
CO, and N2 (obtained from Section 4.1.3 or
4.2.4) into Equation 3-1.
100
Equation 3-1
NOTE: The equation above assumes that
ambient air is used as the source of O, and
that the fuel does not contain appreciable
amounts of N, (as do coke oven or blast furnace gases). For those cases when appreciable amounts of N, are present (coal. oil. and
natural gas do not contain appreciable
amounts of N,) or when oxygen enrichment
is used, alternate methods, subject to approval of the Administrator. are required.
6.3 Dry Molecular Weight. Use Equation
3-2 to calculate the dry molecular weight of
the stack gas
0.280(%N,+%CO)
Equation 3-2
NOTE: The above equation does not consider argon in air (about 0.9 percent, moleculars weight of 37.7). A negative error of
about 0.4 percent is introduced. The tester
may opt to include argon in the analysis
using procedures subject to approval of the
Administrator.
7. Bibliography
1. Altshuller, A. P. Storage of Gases and
Vapors in Plastic Bags. International Journal of Air and Water Pollution. 6.75-81.
1963.
2. Conner. William D. and J. S. Nader. Air
Sampling with Plastice Bags. Journal of the
American Industrial Hygiene Association.
25:291-297. 1964.
3. Burrell Manual for Gas Analysts, Seventh edition. Burrell Corporation. 2223
Fifth Avenue, Pittsburgh, Pa. 15219. 1951.
4. Mitchell, W. J. and M. R. Midgett. Field
Reliability of the Orsat Analyzer. Journal
of Air Pollution Control Association 26:491-
495. May 1976.
5. Shigehara, R. T., R. M. Neulicht, and
W.S. Smith. Validating Orsat Analysis Data
from Fossil Fuel-Fired Units. Stack Sampling News. 4(2):21-26. August, 1976.
Method 3A-Determination of Oxygen and
Carbon Dioxide Concentrations in Emissions
From Stationary Sources (Instrumental
Analyzer Procedure)
1. Applicability and Principle.
1.1 Applicability. This method is
applicable to the determination of oxvgen
(0,) and carbon dioxide (CO2) concentrations
in emissions from stationary sources only
when specified within the regulations.
1.2 Principle. A sample is continuously
extracted from the effluent stream: a portion
of the sample stream is-conveyed to an
instrumental analyzer(s) for determination of
O2 and CO2 concentration(s). Performance
specifications and test procedures are
provided to ensure reliable data.
2. Range and Sensitivity.
Same as Method 6C. Sections 2.1 and 2.2.
except that the span of the monitoring system
shall be selected such that the average O2 or
CO2 concentration is not less than 20 percent
of the span.
3. Definitions.
3.1 Measurement System. The total
equipment required for the determination of
the O2 or CO2 concentration. The
measurement system consists of the same
major subsystems as defined in Method 6C.
Sections 3.1.1. 3.1.2, and 3.1.3.
3.2 Span, Calibration Gas. Analyzer
Calibration Error. Sampling System Bias,
Zero Drift. Calibration Drift, Response Time.
and Calibration Curve. Same as Method 6C.
Sections 3.2 through 3.8. and 3.10.
3.3 Interference Response. The output
response of the measurement system to a
component in the sample gas. other than the
gas component being measured.
4. Measurement System Performance
Specifications.
Same as Method 6C. Sections 4.1 through
4.4.
5. Apparatus and Reagents.
5.1 Measurement System. Any
measurement system for O₂ or CO2 that
meets the specifications of this method. A
schematic of an acceptable measurement
system is shown in Figure 6C-1 of Method 6C.
The essential components of the
measurement system are described below:
5.1.1 Sample Probe. A leak-free probe, of
sufficient length to traverse the sample
points.
5.1.2 Sample Line. Tubing. to transport the
sample gas from the probe to the moisture
removal system. A heated sample line is not
required for systems that measure the O₂ or
CO2 concentration on a dry basis. or
transport dry gases.
5.1.3 Sample Transport Line. Calibration
Value Assembly. Moisture Removal System.
Particulate Filter, Sample Pump. Sample Flow
Rate Control. Sample Gas Manifold. and Data
Recorder. Same as Method 6C. Sections 5.1.3
through 5.1.9, and 5.1.11. except that the
requirements to use stainless steel. Teflon.
and nonreactive glass filters do not apply.
5.1.4 Gas Analyzer. An analyzer to
determine continuously the O₂ or CO2
concentration in the sample gas stream. The
analyzer shall meet the applicable
performance specifications of Section 4. A
means of controlling the analyzer flow rate
and a device for determining proper sample
flow rate (e.g., precision rotameter. pressure
gauge downstream of all flow controls. etc.)
shall be provided at the analyzer. The
5.2 Calibration Gases. The calibration
gases for CO2 analyzers shall be CO2 in N2 or
CO2 in air. Alternatively. CO2/SO2. O2/SO₂
or O2/CO₂/SO₂ gas mixtures in N₂ may be
used. Three calibration gases. as specified
Section 5.3.1 through 5.3.3 of Method 6C. shall
be used. For O₂ monitors that cannot analyze
zero gas. a calibration gas concentration
equivalent to less than 10 percent of the span
may be used in place of zero gas.
6. Measurement System Performance Test
Procedures.
Perform the following procedures before
measurement of emissions (Section 7).
6.1 Calibration Concentration
Verification. Follow Section 6.1 of Method
6C. except if calibration gas analysis is
required, use Method 3 and change the
acceptance criteria for agreement among
Method 3 results to 5 percent (or 0.2 percent
by volume. whichever is greater).
6.2 Interference Response. Conduct an
interference response test of the analyzer
prior to its initial use in the field. Thereafter.
recheck the measurement system if changes
are made in the instrumentation that could
alter the interference response (e.g., changes
in the type of gas detector). Conduct the
interference response in accordance with
Section 5.4 of Method 20.
6.3 Measurement System Preparation,
Analy: Calibration Error. and Sampling
System Bias Check. Follow Sections 6.2
through 6.4 of Method 6C.
7. Emission Test Procedure.
7.1 Selection of Sampling Site and
Sampling Points. Select 8 measurement site
and sampling points using the same criteria
that are applicable to tests performed using
Method 3.
7.2 Sample Collection. Position the
sampling probe at the first measurement
point. and begin sampling at the same rate as
used during the sampling system bias check.
Maintain constant rate sampling (i.e., +10
percent) during the entire run. The sampling
time per run shall be the same as for tests
conducted using Method 3 plus twice the
system response time. For each run, use only
those measurements obtained after twice the
response time of the measurement system has
elapsed to determine the average effluent
concentration.
7.3 Zero and Calibration Drift Test.
Follow Section 7.4 of Method 6C.
8. Quality Control Procedures.
The following quality control procedures
are recommended when the results of this
method are used for an emission rate
correction factor. or excess air determination.
The tester should select one of the following
options for validating measurement results:
8.1 If both O₂ and CO₂ are measured
using Method 3A. the procedures described in
Section 4.4 of Method 3 should be followed to
validate the O₂ and CO2 measurement
results.
8.2 If only O2 is measured using Method
3A. measurements of the sample stream CO2
concentration should be obtained at the
sample by-pass vent discharge using an Orsat
requirements for measuring and controlling
the analyzer flow rate are not applicable if
data are presented that demonstrate the
analyzer is insensitive to flow variations over
the range encountered during the test.
[Method 3A added by 51 FR 21165, June
11, 1986]
[Appendix A, Method 3A]
121:1576.4
or Fyrite analyzer. or equivalent. Duplicate
samples should be obtained concurrent with
at least one run. Average the duplicate Orsat
or Fyrite analysis results for each run. Use
the average CO2 values for comparison with
the O₂ measurements in accordance with the
procedures described in Section 4.4 of
Method 3.
8.3 If only CO2 is measured using Method
3A. concurrent measurements of the sample
stream CO2 concentration should be obtained
using an Orsat or Fyrite analyzer as
described in Section 8.2. For each run.
differences greater than 0.5 percent between
the Method 3A results and the average of the
duplicate Fyrite analysis should be
investigated.
9. Emission Calculation.
For all CO2 enalyzers, and for O₂ analyzers
that can be calibrated with zero gas, follow
Section 8 of Method 6C. except express all
concentrations as percent. rather than ppm.
For O2 analyzers that use a low-level
calibration gas in place of a zero gas,
calculate the effluent gas concentration using
Equation 3A-1.
Eq. 3A-1
where:
Cₘ=Effluent gas concentration, dry basis.
percent.
Actual concentration of the upscale
calibration gas, percent.
Actual concentration of the low-level
calibration gas, percent.
=Average of initial and final system
calibration bias check responses for the
upscale calibration gas. percent.
Average of initial and final system
calibration bias check responses for the
low-level gas, percent.
C=Average gas concentration indicated by
the gas analyzer, dry basis, percent.
10. Bibliography.
Same as bibliography of Method 6C.
METHOD 4-DETERMINATION OF MOISTURE
CONTENT IN STACK GASES
1. Principle and Applicability
1.1 Principle. A gas sample is extracted
at a constant rate from the source: moisture
is removed from the sample stream and determined either volumetrically or gravimetrically.
1.2 Applicability. This method is applicable for determining the moisture content of
stack gas.
Two procedures are given. The first is a
reference method. for accurate determinations of moisture content (such as are
needed to calculate emission data). The
second is an approximation method, which
provides estimates of percent moisture to
aid in setting isokinetic sampling rates prior
to a pollutant emission measurement run.
The approximation method described
herein is only a suggested approach; alternative means for approximating the moisture content. e.g., drying tubes, wet bulb-dry
bulb techniques, condensation techniques.
stoichiometric calculations, previous experience, etc., are also acceptable.
The reference method is often conducted
simultaneously with a pollutant emission
measurement run; when it is. calculation of
percent isokinetic, pollutant emission rate,
etc., for the run shall be based upon the results of the reference method or its equivalent; these calculations shall not be based
upon the results of the approximation
method, unless the approximation method
is shown. to the satisfaction of the Administrator, U.S. Environmental Protection
Agency, to be capable of yielding results
within 1 percent H2O of the reference
method.
NOTE: The reference method may yield
questionable results when applied to saturated gas streams or to streams that contain
water droplets. Therefore, when these conditions exist or are suspected, a second de.
termination of the moisture content shall
be made simultaneously with the reference
method, as follows: Assume that the gas
stream is saturated. Attach a temperature
sensor [capable of measuring to ±1' c (2'
F)] to the reference method probe. Measure
the stack gas temperature at each traverse
point (see Section 2.2.1) during the reference method traverse: calculate the average
stack gas temperature. Next, determine the
moisture percentage, either by: (1) using a
psychrometric chart and making appropriate corrections if stack pressure is different
from that of the chart, or (2) using saturation vapor pressure tables. In cases where
the pyschrometric chart or the saturation
vapor pressure tables are not applicable
(based on evaluation of the process). alternate methods. subject to the approval of the
Administrator, shall be used.
2. Reference Method
The procedure described in Method 5 for
determining moisture content is acceptable
as a reference method.
2.1 Apparatus. A schematic of the sampling train used in this reference method is
shown in Figure 4-1. All components shall
be maintained and calibrated according to
the procedure outlined in Method 5.
STACK
FILTER
WALL
CONDENSER-ICE BATH SYSTEM INCLUDING
(EITHER IN STACK
SILICA GEL TUBE
OR OUT OF STACK)
PROBE
VACUUM
THERMOMETERS
GAUGE
BY-PASS VALVE
ORIFICE
MAIN VALVE
DRY GAS
METER
AIR-TIGHT
PUMP
1
Figure 4-1. Moisture sampling train-reference method.
Environment Reporter
[Appendix A, Method 4]
- -1- -
3
6
7
8
15
2
3
5
7
25
49
45
3
3
2
7
3
7
9
11
12
22
5
2
31
30
34
73
2
3
5
7
11
2
3
X - 1
X - 2
X - 3
X -4
X - 5
X - 6
X 7
- 8
X - - 9
X - 10
X - 11
)
179
180
63
64
65
66
193
194
195
196
i
71
72
73
74
75
76
77
78
79
80
81
82
83
64
85
86
181
182
89
91
7
8
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
11
12
33
34
35
36
37
38
53
61
62
3
64
19
20
28
29
30
31
32
107
108