216-RICR-40-20-1
216-RICR-40-20-1. General Provisions and Standards for Protection Against Radiation (version Adoption, 01/01/2019 to 01/01/2019)
1.1 Authority
A. This Part is promulgated
pursuant to the authority conferred under R.I. Gen. Laws § 23-1.3 -5 .
B. This Part establishes
generally applicable provisions, including standards for protection
against radiation hazards. Except as otherwise specifically provided,
this Part applies to persons licensed or registered by the Agency to
receive, possess, use, transfer, or dispose of any source of
radiation; provided, however, that nothing in this Part shall apply
to any person to the extent such person is subject to regulation by
the U.S. Nuclear Regulatory Commission. The limits in this Part do
not apply to doses due to background radiation, to exposure of
patients to radiation for the purpose of medical diagnosis or
therapy, to exposure from individuals administered radioactive
material and released under § 9.5.16 of this Subchapter, or to
voluntary participation in medical research programs.
C. The requirements of this
Part are designed to control the receipt, possession, use, transfer,
and disposal of sources of radiation by any licensee or registrant in
such a manner that the total dose to an individual, including doses
resulting from all sources of radiation other than background
radiation, does not exceed the standards for protection against
radiation prescribed in this Part. However, nothing in this Part
shall be construed as limiting actions that may be necessary to
protect health and safety.
1.2 Incorporated Material
A. Except as provided in this
Part, the requirements of 10
C.F.R. Part 20 (2018) are incorporated by
reference, not including any further editions or amendments thereof
and only to the extent that the provisions therein are not
inconsistent with this Part.
B. Notwithstanding the
provisions of § 1.2(A) of this Part, §§ 20.1001, 20.1002, 20.1006,
20.1007, 20.1008, 20.1009, 20.1205, 20.1401, 20.1406(b), 20.1905(g),
20.2109, 20.2202, 20.2203(c), 20.2206(a)(1), (3), (4) and (5),
20.2205, 20.2206, 20.2301, 20.2302, 20.2401, 20.2402, Appendix D to
Part 20 and Appendix F to 10 C.F.R. Part 20 are not incorporated by
reference.
C. Effect of incorporation of
10 C.F.R. Part 20. To reconcile differences between this Part and the
incorporated sections of 10 C.F.R. Part 20, the following words and
phrases shall be substituted for the language in 10 C.F.R. Part 20 as
follows:
1. Any reference to NRC or
Commission shall be deemed to be a reference to the Agency.
2. Any reference to NRC or
agreement state shall be deemed to be a reference to the Agency, NRC,
or agreement state.
3. Any reference to byproduct
material shall be deemed to be a reference to radioactive material.
4. Any notifications, reports
or correspondence referenced in the incorporated sections of 10
C.F.R. Part 20 shall be directed to the Agency using contact
information specified in § 1.4 of this Part.
5. Any reference to licensee
shall be deemed to include registrant.
6. Any reference to license
shall be deemed to include registration.
7. Any reference to licensed
shall be deemed to include registered.
8. Any requirement to utilize
NRC Form 4 may also be satisfied by use of Agency
Form RCA-2.
9. Any requirement to utilize
NRC Form 5 may also be satisfied by use of Agency
Form RCA-3.
10. 10 C.F.R. Part 20
notwithstanding, exposures involving the use of X-rays may be
weighted, in a manner specified by the Agency, so that, with Agency
approval, the effective dose equivalent may be substituted for the
deep dose equivalent in determining compliance with occupational
exposure limits for specified groups of individuals.
1.3 Definitions
A. In addition to the
definitions contained in 10
C.F.R. § 20.1003 , whenever used in this Part, the
following terms shall be construed as follows:
1. “Act” means R.I. Gen.
Laws Chapter 23-1.3 ,
entitled "Radiation Control".
2. “Agency” means Rhode
Island Radiation Control Agency (RCA), Center for Health Facilities
Regulation - Radiation Control Program, Rhode Island Department of
Health.
3. “Annual” means an
interval not to exceed twelve (12) months.
4. “NARM” means any
naturally occurring or accelerator-produced radioactive material. It
does not include byproduct, source, or special nuclear material.
5. “Radioactive material”
means any material (solid, liquid, or gas) which emits radiation
spontaneously.
6. “Registrant” means any
person who is registered with the Agency and is legally obligated to
register with the Agency pursuant to this Subchapter and the Act.
7. “Registration” means
registration with the Agency pursuant to this Subchapter and the Act.
1.4 Communications
A. All
communications and reports concerning this Subchapter, and
applications filed thereunder, should be addressed to the Agency at
its office located at:
Rhode
Island Department of Health
Center
for Health Facilities Regulation
Radiation
Control Program
Three
Capitol Hill - Room 305
Providence,
RI 02908-5097
B. During
normal business hours, the Agency may be contacted at (401) 222-2566.
At other times, this number will allow you to leave a message on the
answering machine. In case of an emergency when it is necessary to
immediately contact the Agency, utilize the RI Department of Health's
24-hour number (401) 272-5952 and indicate the nature of your
emergency. FAX communications may be sent 24 hours a day to (401)
222-3999. For non-emergency situations, any required report or other
routine correspondence may also be submitted via e-mail to
[email protected] .
1.5 General Provisions
1.5.1 Implementation
A. Any existing license or
registration condition that is more restrictive than this Part
remains in force until there is an amendment or renewal of the
license or registration.
B. If a license or
registration condition exempts a licensee or registrant from a
provision of this Part in effect on or before January 1, 1994, it
also exempts the licensee or registrant from the corresponding
provision of this Part.
C. If a license or
registration condition cites provisions of this Part in effect prior
to January 1, 1994, which do not correspond to any provisions of this
Part, the license or registration condition remains in force until
there is an amendment or renewal of the license or registration that
modifies or removes this condition.
1.5.2 Exemptions and
Additional Requirements
A. The Agency may, upon
application by a licensee or registrant or upon its own initiative,
grant an exemption from the requirements of this Subchapter if it
determines the exemption is authorized by law and would not result in
undue hazard to life or property.
B. The Agency may, by rule,
regulation, or order, impose requirements on a licensee or
registrant, in addition to those established in this Subchapter, as
it deems appropriate or necessary to protect health or to minimize
danger to life or property.
1.5.3 Inspections
A. Each licensee and
registrant shall afford the Agency at all reasonable times the
opportunity to inspect sources of radiation and the premises and
facilities wherein such sources of radiation are used or stored, and
the cooperation and assistance of the registrant or licensee, or his
staff, if needed.
B. Each licensee and
registrant shall make available to the Agency for inspection, upon
reasonable notice, records maintained pursuant to this Subchapter.
1.5.4 Tests
A. Each licensee and
registrant shall perform upon instructions from the Agency, or shall
permit the Agency to perform such reasonable tests as the Agency
deems appropriate or necessary including, but not limited to, tests
of:
1. Sources of radiation;
2. Facilities wherein sources
of radiation are used or stored;
3. Radiation detection and
monitoring instruments; and
4. Other equipment and devices
used in connection with utilization or storage of licensed or
registered sources of radiation.
1.5.5 Violations
An injunction or other court
order may be obtained prohibiting any violation of any provision of
the Act or any regulation or order issued thereunder. Any person who
willfully violates any provision of the Act or any regulation or
order issued thereunder may be guilty of a misdemeanor and upon
conviction, may be punished by fine or imprisonment or both, as
provided by law.
1.5.6 Units of Radiation
Dose
For the purpose of this Part,
the units of radiation dose are defined by 10 C.F.R. § 20.1004.
1.5.7 Units of
Radioactivity
For the purpose of this Part,
the units of radioactivity are defined by 10 C.F.R. § 20.1005.
1.5.8 Deliberate Misconduct
A. Any licensee, certificate
of registration holder, applicant for a license or certificate of
registration, employee of a licensee, certificate of registration
holder or applicant; or any contractor (including a supplier or
consultant), subcontractor, employee of a contractor or subcontractor
of any licensee or certificate of registration holder or applicant
for a license or certificate of registration, who knowingly provides
to any licensee, applicant, certificate holder, contractor, or
subcontractor, any components, equipment, materials, or other goods
or services that relate to a licensee’s, certificate holder’s or
applicant’s activities in this part, may not:
1. Engage in deliberate
misconduct that causes or would have caused, if not detected, a
licensee, certificate of registration holder, or applicant to be in
violation of any rule, regulation, or order; or any term, condition,
or limitation of any license issued by the Agency; or
2. Deliberately submit to the
Agency, a licensee, certificate of registration holder, an applicant,
or a licensee’s, certificate holder’s or applicant’s,
contractor or subcontractor, information that the person submitting
the information knows to be incomplete or inaccurate in some respect
material to the Agency.
B. A person who violates §§
1.5.8(A)(1) or (2) of this Part may be subject to enforcement action
in accordance with the procedures in § 2.11 of this Subchapter.
C. For the purposes of §
1.5.8(A)(1) of this Part, deliberate misconduct by a person means an
intentional act or omission that the person knows:
1. Would cause a licensee,
certificate of registration holder or applicant to be in violation of
any rule, regulation, or order; or any term, condition, or
limitation, of any license issued by the Agency; or
2. Constitutes a violation of
a requirement, procedure, instruction, contract, purchase order, or
policy of a licensee, certificate of registration holder, applicant,
contractor, or subcontractor.
1.6 Radiation Protection Programs
For the purpose of this Part,
the required radiation protection program is defined by 10 C.F.R. §
20.1101.
1.7 Occupational Dose Limits
1.7.1 Occupational Dose
Limits for Adults
A. For the purpose of this
Part, the occupational dose limits for adults are defined by 10
C.F.R. § 20.1201.
B. For sources of radiation
other than radioactive material, when a protective apron is worn and
monitoring is conducted as specified in § 1.10.3(B) of this Part,
the effective dose equivalent for external radiation shall be
determined as follows:
1. When only one individual
monitoring device is used and it is located at the neck outside the
protective apron, and the reported dose exceeds twenty-five percent
(25%) of the limit specified in § 1.7.1(A) of this Part, the
reported deep dose equivalent value multiplied by 0.3 shall be the
effective dose equivalent for external radiation; or
2. When individual monitoring
devices are worn, both under the protective apron at the waist and
outside the protective apron at the neck, the effective dose
equivalent for external radiation shall be assigned the value of the
sum of the deep dose equivalent reported for the individual
monitoring device located at the waist under the protective apron
multiplied by 1.5 and the deep dose equivalent reported for the
individual monitoring device located at the neck outside the
protective apron multiplied by 0.04.
1.7.2 Compliance with
Requirements for Summation of External and Internal Doses
For the purpose of this Part,
compliance with requirements for summation of external and internal
doses is defined by 10 C.F.R. § 20.1202.
1.7.3 Determination of
External Dose from Airborne Radioactive Material
For the purpose of this Part,
determination of external dose from airborne radioactive material is
defined by 10 C.F.R. § 20.1203.
1.7.4 Determination of
Internal Exposure
For the purpose of this Part,
determination of internal exposure is defined by 10 C.F.R. §
20.1204.
1.7.5 Determination of
Prior Occupational Dose
A. For the purpose of this
Part, determination of prior occupational dose is defined by 10
C.F.R. § 20.2104.
B. The licensee or registrant
shall retain the records of prior occupational dose and exposure
history as specified in § 1.7.5(A) of this Part on Agency Form RCA-2
or equivalent until the Agency terminates each pertinent license or
registration requiring this record. The licensee or registrant shall
retain records used in preparing Agency Form RCA-2 or equivalent for
three (3) years after the record is made.
C. Upon termination of the
license or registration, the licensee or registrant shall make
arrangements, satisfactory to the Agency, for permanent storage of
records contained on Agency Form RCA-2 or equivalent.
1.7.6 Planned Special
Exposures
For the purpose of this Part,
planned special exposures are defined by 10 C.F.R. § 20.1206.
1.7.7 Occupational Dose
Limits for Minors
For the purpose of this Part,
occupational dose limits for minors are defined by 10 C.F.R. §
20.1207.
1.7.8 Dose Equivalent to an
Embryo/Fetus
For the purpose of this Part,
dose equivalent to an embryo/fetus is defined by 10 C.F.R. §
20.1208.
1.8 Radiation Dose Limits for
Individual Members of the Public
1.8.1 Dose Limits for
Individual Members of the Public
A. For the purpose of this
Part, dose limits for individual members of the public are defined by
10 C.F.R. § 20.1301.
B. Each registrant shall
conduct operations so that the total effective dose equivalent to
individual members of the public does not exceed the original design
criteria of 5 mSv (0.5 rem) in a year at locations within registered
facilities where only radiation machines were installed prior to
January 1, 1994 and which continue to meet the original design
criteria (e.g. workload, type and use of radiation machine, room
configuration, etc.) on or after January 1, 1994.
1.8.2 Compliance with Dose
Limits for Individual Members of the Public
For the purpose of this Part,
compliance with dose limits for individual members of the public is
defined by 10 C.F.R. § 20.1302.
1.9 Radiological Criteria for
License Termination
1.9.1 General Provisions
and Scope
A. Applicability. The criteria
in §§ 1.9.1 through 1.9.6 of this Part apply to the decommissioning
of facilities licensed under Parts 7, 9, 10 and 11 of this
Subchapter, as well as other facilities subject to the Agency’s
jurisdiction.
B. After a site has been
decommissioned and the license terminated in accordance with the
criteria in §§ 1.9.1 through 1.9.6 of this Part, the Agency will
require additional cleanup only if, based on new information, it
determines that the criteria in §§ 1.9.1 through 1.9.6 of this Part
were not met and residual radioactivity remaining at the site could
result in significant threat to public health and safety.
C. When calculating TEDE to
the average member of the critical group the licensee shall determine
the peak annual TEDE dose expected within the first one thousand
(1000) years after decommissioning.
1.9.2 Radiological Criteria
for Unrestricted Use
For the purpose of this Part,
compliance with radiological criteria for unrestricted use is defined
by 10 C.F.R. § 20.1402.
1.9.3 Criteria for License
Termination Under Restricted Conditions
For the purpose of this Part,
criteria for license termination under restricted conditions is
defined by 10 C.F.R. § 20.1403.
1.9.4 Alternate Criteria
for License Termination
For the purpose of this Part,
alternate criteria for license termination is defined by 10 C.F.R. §
20.1404.
1.9.5 Public Notification
and Public Participation
For the purpose of this Part,
requirements for public notification and public participation are
defined by 10 C.F.R. § 20.1405.
1.9.6 Minimization of
Contamination
For the purpose of this Part,
requirements for minimization of contamination are defined by 10
C.F.R. § 20.1406, excluding 10 C.F.R. § 20.1406(b).
1.10 Surveys and Monitoring
1.10.1 Testing for Leakage
or Contamination of Sealed Sources
A. The licensee in possession
of any sealed source shall assure that:
1. Each sealed source, except
as specified in § 1.10.1(B) of this Part, is tested for leakage or
contamination and the test results are received before the sealed
source is put into use unless the licensee has a certificate from the
transferor indicating that the sealed source was tested within six
(6) months before transfer to the licensee.
2. Each sealed source that is
not designed to emit alpha particles is tested for leakage or
contamination at intervals not to exceed six (6) months or at
alternative intervals approved by the Agency, after evaluation of
information specified by § 7.6.17 of this Subchapter, another
Agreement State or the U.S. Nuclear Regulatory Commission.
3. Each sealed source that is
designed to emit alpha particles is tested for leakage or
contamination at intervals not to exceed three (3) months or at
alternative intervals approved by the Agency, after evaluation of
information specified by § 7.6.17 of this Subchapter, another
Agreement State or the U.S. Nuclear Regulatory Commission.
4. For each sealed source that
is required to be tested for leakage or contamination, at any other
time there is reason to suspect that the sealed source might have
been damaged or might be leaking, the licensee shall assure that the
sealed source is tested for leakage or contamination before further
use.
5. Tests for leakage for all
sealed sources, except brachytherapy sources manufactured to contain
radium, shall be capable of detecting the presence of 185 Bq (0.005
µCi) of radioactive material on a test sample. Test samples shall
be taken from the sealed source or from the surfaces of the container
in which the sealed source is stored or mounted on which one might
expect contamination to accumulate. For a sealed source contained in
a device, test samples are obtained when the source is in the "off"
position.
6. The test for leakage for
brachytherapy sources manufactured to contain radium shall be capable
of detecting an absolute leakage rate of 37 Bq (0.001 µCi) of
radon-222 in a twenty-four (24) hour period when the collection
efficiency for radon-222 and its daughters has been determined with
respect to collection method, volume and time.
7. Tests for contamination
from radium daughters shall be taken on the interior surface of
brachytherapy source storage containers and shall be capable of
detecting the presence of 185 Bq (0.005 µCi) of a radium daughter
which has a half-life greater than four (4) days.
B. A licensee need not perform
test for leakage or contamination on the following sealed sources:
1. Sealed sources containing
only radioactive material with a half-life of less than thirty (30)
days;
2. Sealed sources containing
only radioactive material as a gas;
3. Sealed sources containing
3.7 MBq (100 µCi) or less of beta or photon-emitting material or 370
kBq (10 µCi) or less of alpha-emitting material;
4. Sealed sources containing
only hydrogen-3;
5. Seeds of iridium-192
encased in nylon ribbon; and
6. Sealed sources, except
teletherapy and brachytherapy sources, which are stored, not being
used and identified as in storage. The licensee shall, however, test
each such sealed source for leakage or contamination and receive the
test results before any use or transfer unless it has been tested for
leakage or contamination within six (6) months before the date of use
or transfer. No sealed source shall be stored for a period of more
than ten (10) years without being tested for leakage and/or
contamination.
C. Tests for leakage or
contamination from sealed sources shall be performed by persons
specifically authorized by the Agency, another Agreement State or the
U.S. Nuclear Regulatory Commission to perform such services.
D. Records of tests for
leakage or contamination of sealed sources required by § 1.10.1 of
this Part shall be kept in units of becquerel or microcurie and
maintained for inspection by the Agency for five (5) years after the
records are made.
E. The following shall be
considered evidence that a sealed source is leaking:
1. The presence of 185 Bq
(0.005 µCi) or more of removable contamination on any test sample;
or
2. Leakage of 37 Bq (0.001
µCi) of radon-222 per twenty-four (24) hours for brachytherapy
sources manufactured to contain radium.
F. The licensee shall
immediately withdraw a leaking sealed source from use and shall take
action to prevent the spread of contamination. The leaking sealed
source shall be repaired or disposed of in accordance with this Part.
G. The licensee shall file a
report within five (5) working days with the Agency if the test for
leakage or contamination indicates a sealed source is leaking or
contaminated. The report shall include the equipment involved, the
test results and the corrective action taken.
1.10.2 General Survey and
Monitoring Requirements
A. For the purpose of this
Part, general survey and monitoring requirements are defined by 10
C.F.R. § 20.1501.
B. Exposure of a personnel
monitoring device to deceptively indicate a dose delivered to an
individual is prohibited.
1.10.3 Conditions Requiring
Individual Monitoring of External and Internal Occupational Dose
A. For the purpose of this
Part, conditions requiring individual monitoring of external and
internal occupational dose are defined by 10 C.F.R. § 20.1502.
B. Individuals wearing a
protective apron, when personnel monitoring is otherwise required by
this Subchapter, shall position their individual monitoring devices
as follows:
1. An individual monitoring
device used for the dose to an embryo/fetus of a declared pregnant
woman, pursuant to § 1.7.8 of this Part, shall be located under the
protective apron at the waist.
a. It is recognized that, in
the specific work environment of medical fluoroscopic equipment, the
dose to the embryo/fetus is overestimated by the individual
monitoring device because of the overlying tissue of the pregnant
individual. A medical physicist who is registered with the Agency
pursuant to § 3.6 of this Subchapter as a Provider of Diagnostic
X-Ray Physics Services should be consulted to determine the dose to
the embryo/fetus for the rare occasion in which this individual
monitoring device has a monthly reported dose equivalent value in
excess of 0.5 mSv (50 mrem). Therefore, for purposes of this Part,
the value to be used for determining the dose to an embryo/fetus
pursuant to § 1.7.8 of this Part for occupational exposure to
radiation from medical fluoroscopic equipment may be the value
reported by the individual monitoring device worn at the waist
underneath the protective apron which has been corrected for the
particular individual and her work environment by the above
referenced medical physicist.
2. An individual monitoring
device used for eye dose equivalent shall be located at the neck, or
an unshielded location closer to the eye, outside the protective
apron.
3. When only one individual
monitoring device is used to determine the effective dose equivalent
for external radiation pursuant to § 1.7.1(B) of this Part, it shall
be located at the neck outside the protective apron. When a second
individual monitoring device is used, for the same purpose, it shall
be located under the protective apron at the waist. The second
individual monitoring device is required for a declared pregnant
woman.
C. An individual monitoring
device used for monitoring the dose to the extremities, to
demonstrate compliance with § 1.7.1(A) of this Part, shall be worn
on the extremity likely to receive the highest exposure. Each
individual monitoring device shall be oriented to measure the highest
dose to the extremity being monitored.
1.11 Control of Exposure from
External Sources in Restricted Areas
1.11.1 Control of Access to
High Radiation Areas
A. For the purpose of this
Part, control of access to high radiation areas is defined by 10
C.F.R. § 20.1601.
B. The registrant is not
required to control entrance or access to rooms or other areas
containing sources of radiation capable of producing a high radiation
area as described in § 1.11.1(A) of this Part if the registrant has
met all the specific requirements for access and control specified in
other applicable Parts of this Subchapter.
1.11.2 Control of Access to
Very High Radiation Areas
A. For the purpose of this
Part, control of access to very high radiation areas is defined by 10
C.F.R. § 20.1602.
B. The registrant is not
required to control entrance or access to rooms or other areas
containing sources of radiation capable of producing a very high
radiation area as described in § 1.11.2(A) of this Part if the
registrant has met all the specific requirements for access and
control specified in other applicable Parts of this Subchapter.
1.12 Respiratory Protection and
Controls to Restrict Internal Exposure in Restricted Areas
1.12.1 Use of Process or
Other Engineering Controls
For the purpose of this Part,
use of process or other engineering controls is defined by 10 C.F.R.
§ 20.1701.
1.12.2 Use of Other
Controls
For the purpose of this Part,
use of other controls is defined by 10 C.F.R. § 20.1702.
1.12.3 Use of Individual
Respiratory Protection Equipment
A. For the purpose of this
Part, use of individual respiratory protection equipment is defined
by 10 C.F.R. § 20.1703.
B. For the purpose of this
Part, further restrictions on the use of respiratory protection
equipment are defined by 10 C.F.R. § 20.1704.
C. For the purpose of this
Part, authorization for use of higher assigned protection factors is
required by 10 C.F.R. § 20.1705.
1.13 Storage and Control of
Licensed Material
1.13.1 Security of Stored
Material
For the purpose of this Part,
security of stored material is defined by 10 C.F.R. § 20.1801.
1.13.2 Control of Material
Not in Storage
A. For the purpose of this
Part, control of material not in storage is defined by 10 C.F.R. §
20.1802.
B. The registrant shall
maintain control of radiation machines that are in an unrestricted
area and that are not in storage.
1.14 Precautionary Procedures
1.14.1 Caution Signs
For the purpose of this Part,
caution signs are defined by 10 C.F.R. § 20.1901.
1.14.2 Posting Requirements
For the purpose of this Part,
posting requirements are defined by 10 C.F.R. § 20.1902.
1.14.3 Exceptions to
Posting Requirements
A. For the purpose of this
Part, exceptions to posting requirements are defined by 10 C.F.R. §
20.1903.
B. A room or area is not
required to be posted with a caution sign because of the presence of
radiation machines used solely for diagnosis in the healing arts.
1.14.4 Labeling Containers
and Radiation Machines
A. For the purpose of this
Part, labeling of containers is defined by 10 C.F.R. § 20.1904.
B. Each registrant shall
ensure that each radiation machine is labeled in a conspicuous manner
which cautions individuals that radiation is produced when it is
energized.
1.14.5 Exemptions to
Labeling Requirements
For the purpose of this Part,
exemptions to labeling requirements are defined by 10 C.F.R. §
20.1905, excluding 10 C.F.R. § 20.1905(g).
1.14.6 Procedures for
Receiving and Opening Packages
For the purpose of this Part,
procedures for receiving and opening packages are defined by 10
C.F.R. § 20.1906.
1.15 Waste Disposal
1.15.1 General Requirements
for Waste Disposal
For the purpose of this Part,
general requirements for waste disposal are defined by 10 C.F.R. §
20.2001.
1.15.2 Method for Obtaining
Approval of Proposed Disposal Procedures
For the purpose of this Part,
the method for obtaining approval of proposed disposal procedures is
defined by 10 C.F.R. § 20.2002.
1.15.3 Disposal by Release
into Sanitary Sewerage
For the purpose of this Part,
disposal by release into sanitary sewerage is defined by 10 C.F.R. §
20.2003.
1.15.4 Treatment or
Disposal by Incineration
For the purpose of this Part,
treatment or disposal by incineration is defined by 10 C.F.R. §
20.2004.
1.15.5 Disposal of Specific
Wastes
For the purpose of this Part,
disposal of specific wastes is defined by 10 C.F.R. § 20.2005.
1.15.6 Transfer for
Disposal and Manifests
For the purpose of this Part,
transfer for disposal and manifests are defined by 10 C.F.R. §
20.2006.
1.15.7 Compliance with
Environmental and Health Protection Regulations
For the purpose of this Part,
compliance with environmental and health protection regulations is
defined by 10 C.F.R. § 20.2007.
1.15.8 Disposal of 11e(3)
and 11e(4) Byproduct Material
For the purpose of this Part,
disposal of 11e(3) and 11e(4) Byproduct Material is defined by 10
C.F.R. § 20.2008.
1.16 Records
1.16.1 General Provisions
A. For the purpose of this
Part, general recordkeeping provisions are defined by 10 C.F.R. §
20.2101.
B. Each licensee and
registrant shall maintain records showing the receipt, transfer, and
disposal of all sources of radiation. All records required by this
Subchapter shall be maintained indefinitely unless otherwise
specified in this Subchapter.
1.16.2 Records of Radiation
Protection Programs
For the purpose of this Part,
requirements for maintenance of records of radiation protection
programs are defined by 10 C.F.R. § 20.2102.
1.16.3 Records of Surveys
For the purpose of this Part,
requirements for maintenance of records of surveys are defined by 10
C.F.R. § 20.2103.
1.16.4 Records of Tests for
Leakage or Contamination of Sealed Sources
Records of tests for leakage
or contamination of sealed sources required by § 1.10.1(A) of this
Part shall be kept in units of becquerel or microcurie and maintained
for inspection by the Agency for five (5) years after the records are
made.
1.16.5 Records of Planned
Special Exposures
For the purpose of this Part,
requirements for maintenance of records of planned special exposures
are defined by 10 C.F.R. § 20.2105.
1.16.6 Records of
Individual Monitoring Results
For the purpose of this Part,
requirements for maintenance of records of individual monitoring
results are defined by 10 C.F.R. § 20.2106.
1.16.7 Records of Dose to
Individual Members of the Public
For the purpose of this Part,
requirements for maintenance of records of dose to individual members
of the public are defined by 10 C.F.R. § 20.2107.
1.16.8 Records of Waste
Disposal
For the purpose of this Part,
requirements for maintenance of records of waste disposal are defined
by 10 C.F.R. § 20.2108.
1.16.9 Form of Records
For the purpose of this Part,
requirements regarding the form of records are defined by 10 C.F.R. §
20.2110.
1.17 Reports
1.17.1 Reports of Theft or
Loss of Licensed Material
For the purpose of this Part,
requirements regarding reports of theft or loss of licensed material
are defined by 10 C.F.R. § 20.2201.
1.17.2 Notification of
Incidents
A. Immediate Notification.
Notwithstanding other requirements for notification, each licensee or
registrant shall immediately report each event involving a source of
radiation possessed by the licensee or registrant that may have
caused or threatens to cause any of the following conditions:
1. Immediately notify the
Agency of each event involving a source of radiation possessed by the
licensee or registrant that may have caused or threatens to cause any
of the following conditions:
a. An individual to receive:
(1) A total effective dose
equivalent of 0.25 Sv (25 rem) or more; or
(2) A lens dose equivalent of
0.75 Sv (75 rem) or more; or
(3) A shallow dose equivalent
to the skin or extremities or a total organ dose equivalent of 2.5 Gy
(250 rad) or more; or
b. The release of radioactive
material, inside or outside of a restricted area, so that, had an
individual been present for twenty-four (24) hours, the individual
could have received an intake five (5) times the occupational ALI.
This provision does not apply to locations where personnel are not
normally stationed during routine operations, such as hot-cells or
process enclosures.
2. Immediately notify the
Agency as soon as possible, but not later than four (4) hours after
the discovery, of an event (e.g., fire, explosion, toxic gas release,
etc.) that prevents immediate protective actions necessary to avoid
exposures to radiation or radioactive materials that could exceed
regulatory limits or releases of licensed material that could exceed
regulatory limits.
B. Twenty-Four Hour
Notification. Each licensee or registrant shall, within twenty-four
(24) hours of discovery of the event, report to the Agency each event
involving loss of control of a licensed or registered source of
radiation possessed by the licensee or registrant that may have
caused, or threatens to cause, any of the following conditions:
1. An individual to receive,
in a period of twenty-four (24) hours:
a. A total effective dose
equivalent exceeding 0.05 Sv (5 rem); or
b. A lens dose equivalent
exceeding 0.15 Sv (15 rem); or
c. A shallow dose equivalent
to the skin or extremities or a total organ dose equivalent exceeding
0.5 Sv (50 rem); or
2. The release of radioactive
material, inside or outside of a restricted area, so that, had an
individual been present for twenty-four (24) hours, the individual
could have received an intake in excess of one occupational ALI. This
provision does not apply to locations where personnel are not
normally stationed during routine operations, such as hot-cells or
process enclosures.
3. An unplanned contamination
event that:
a. Requires access to the
contaminated area, by workers or the public, to be restricted for
more than twenty-four (24) hours by imposing additional radiological
controls or by prohibiting entry into the area; and
b. Involves a quantity of
material greater than five times the lowest annual limit on intake
specified for the material in § 1.19 of this Part; and
c. Has access to the area
restricted for a reason other than to allow isotopes with a half-life
of less than twenty-four (24) hours to decay prior to
decontamination.
4. An event in which equipment
is disabled or fails to function as designed when:
a. The equipment is required
by regulation or license/registration condition to prevent releases
exceeding regulatory limits, to prevent exposures to radiation and/or
radioactive materials exceeding regulatory limits, or to mitigate the
consequences of an accident; and
b. The equipment is required
to be available and operable when it is disabled or fails to
function; and
c. No redundant equipment is
available and operable to perform the required safety function.
5. An event that requires
unplanned medical treatment at a medical facility of an individual
with spreadable radioactive contamination on the individual's
clothing or body.
6. An unplanned fire or
explosion damaging any licensed material or any device, container, or
equipment containing licensed material when:
a. The quantity of material
involved is greater than five (5) times the lowest annual limit on
intake specified for the material in § 1.19 of this Part; and
b. The damage affects the
integrity of the licensed material or its container.
C. The licensee or registrant
shall prepare each report filed with the Agency pursuant to § 1.17.2
of this Part so that names of individuals who have received exposure
to sources of radiation are stated in a separate and detachable
portion of the report.
D. Licensees or registrants
shall make the reports required by §§ 1.17.2(A) and (B) of this
Part to the Agency by telephone, telegram, mailgram, or facsimile to
the Agency. To the extent that the information is available at the
time of notification, the information provided in these reports shall
include:
1. The name of the person
making the report and their call-back telephone number;
2. A description of the event,
including time and date;
3. The exact location of the
event;
4. The levels of radiation and
the isotopes, quantities, and chemical and physical form of the
licensed material involved; and
5. Any personnel radiation
exposure data available.
E. The provisions of § 1.17.2
of this Part do not apply to doses that result from planned special
exposures, provided such doses are within the limits for planned
special exposures and are reported pursuant to § 1.17.4 of this
Part.
1.17.3 Reports of
Exposures, Radiation Levels, and Concentrations of Radioactive
Material Exceeding the Constraints or Limits
For the purpose of this Part,
requirements regarding reports of exposures, radiation levels, and
concentrations of radioactive material exceeding the constraints or
limits are defined by 10 C.F.R. § 20.2203, excluding 10 C.F.R. §
20.2203(c).
1.17.4 Reports of Planned
Special Exposures
For the purpose of this Part,
requirements regarding reports of planned special exposures are
defined by 10 C.F.R. § 20.2204.
1.17.5 Notifications and
Reports to Individuals
When a licensee or registrant
is required pursuant to §§ 1.17.3 or 1.17.4 of this Part to report
to the Agency any exposure of an individual to radiation or
radioactive material, the licensee or registrant shall also notify
the individual. Such notice shall be transmitted at a time not later
than the transmittal to the Agency.
1.17.6 Reports of
Transactions Involving Nationally Tracked Sources
For the purpose of this Part,
requirements regarding reports of transactions involving nationally
tracked sources are defined by 10 C.F.R. § 20.2207.
1.17.7 Vacating Premises
Each specific licensee shall,
no less than thirty (30) days before vacating or relinquishing
possession or control of premises which may have been contaminated
with radioactive material as a result of his or her activities,
notify the Agency in writing of intent to vacate. When deemed
necessary by the Agency, the licensee shall decontaminate the
premises in such a manner as the Agency may specify.
1.18 Assigned Protection Factors
for Respirators
For the purpose of this Part,
assigned protection factors for respirators are defined in Appendix A
to 10 C.F.R. Part 20.
1.19 Annual
Limits on Intake (ALIs) and Derived Air Concentrations (DACs) of
Radionuclides for Occupational Exposure; Effluent Concentrations;
Concentrations for Release to Sewerage
For the purpose of this Part,
Annual
Limits on Intake (ALIs) and Derived Air Concentrations (DACs) of
radionuclides for occupational exposure; effluent concentrations; and
concentrations for release to sewerage are defined in Appendix B
to 10 C.F.R. Part 20.
1.20 Quantities of Licensed
Material Requiring Labeling
For the purpose of this Part,
quantities of licensed material requiring labeling are defined in
Appendix C to 10 C.F.R. Part 20.
1.21 Requirements
for Transfers of Low-Level Radioactive Waste Intended for Disposal at
Licensed Land Disposal Facilities and Manifests
For the purpose of this Part,
requirements for transfers of low-level radioactive waste intended
for disposal at licensed land disposal facilities and manifests are
defined in Appendix G to 10 C.F.R. Part 20.
1.22 Nationally Tracked Source
Thresholds
For the purpose of this Part,
requirements for nationally tracked source thresholds are defined in
Appendix E to 10 C.F.R. Part 20.