216-RICR-40-20-4
216-RICR-40-20-4. Medical Diagnostic and Interventional X-Ray and Imaging Systems (version Adoption, 01/01/2019 to 01/04/2022)
To view the official Final Regulation for accessibility and readability, click Download Regulation:
Download Regulation
4.1 Authority
A. This Part is promulgated
pursuant to the authority conferred under R.I. Gen. Laws § 23-1.3-5.
B. This Part establishes
requirements, for which a registrant is responsible, for use of
diagnostic X-ray equipment and associated imaging systems in the
healing arts or veterinary medicine. The provisions of this Part are
in addition to, and not in substitution for, other applicable
provisions of this Subchapter.
C. The use of diagnostic X-ray
equipment and associated imaging systems for the intentional exposure
of individuals for diagnosis shall be by or under the supervision of
a licensed practitioner of the healing arts.
D. The use of diagnostic X-ray
equipment and associated imaging systems in the practice of
veterinary medicine shall be by or under the supervision of an
individual authorized by and licensed in accordance with R.I. Gen.
Laws Chapter
5-25 to practice veterinary medicine.
E. Any notifications, reports
or correspondence required by this Part shall be directed to the
Agency using contact information specified in § 1.4 of this
Subchapter.
4.1.1 Incorporation by
Reference
Except as provided in this
Part, the requirements of 21
CFR Part 900 (2018) are incorporated by reference, not including
any further editions or amendments thereof and only to the extent
that the provisions therein are not inconsistent with this Part.
4.2 Definitions
A. Whenever used in this Part,
the following terms shall be construed as follows:
1. "Accessible surface"
means the external surface of the enclosure or housing of the
radiation producing machine as provided by the manufacturer.
2. “Act” means R.I. Gen.
Laws Chapter 23-1.3, entitled "Radiation Control".
3. “Agency” means Rhode
Island Radiation Control Agency (RCA), Center for Health Facilities
Regulation - Radiation Control Program, Rhode Island Department of
Health.
4. "Air kerma" means
kerma in air (see definition of Kerma).
5. "Air kerma rate (AKR)"
means the air kerma per unit time.
6. "Alert value"
means a dose index (e.g., of CTDI vol (mGy) or DLP(mGy-cm))
that is set by the registrant to trigger an alert to the CT operator
prior to scanning within an ongoing examination. The Alert value
represents a universal dose index value well above the registrant's
established range for the examination that warrants more stringent
review and consideration before proceeding.
7. "Aluminum equivalent"
means the thickness of type 1100 aluminum alloy affording the same
attenuation, under specified conditions, as the material in question.
[The nominal chemical composition of type 1100 aluminum is 99.00
percent minimum aluminum, 0.12 percent copper.]
8. "Articulated joint"
means a joint between two separate sections of a tabletop which joint
provides the capacity of one of the sections to pivot on the line
segment along which the sections join.
9. "Attenuation block"
means a block or stack of type 1100 aluminum alloy, or aluminum alloy
having equivalent attenuation, with dimensions 20 centimeters (cm) or
larger by 20 cm or larger by 3.8 cm, that is large enough to
intercept the entire x-ray beam.
10. "Automatic exposure
control (AEC)" means a device which automatically controls one
or more technique factors in order to obtain at a preselected
location(s) a required quantity of radiation.
11. "Automatic exposure
rate control (AERC)" means a device which automatically controls
one or more technique factors in order to obtain, at a preselected
location(s), a required quantity of radiation per unit time.
12. "Barrier" (See
"Protective barrier").
13. "Beam axis"
means a line from the source through the centers of the x-ray fields.
14. "Beam-limiting
device" means a device which provides a means to restrict the
dimensions of the x-ray field.
15. "Bone densitometry"
means a noninvasive measurement of certain physical characteristics
of bone that reflect bone strength. Test results are typically
reported as bone mineral content or density and are used for
diagnosing osteoporosis, estimating fracture risk, and monitoring
changes in bone mineral content.
16. "Bone densitometer"
means a device intended for medical purposes to measure bone density
and mineral content by x-ray or gamma ray transmission measurements
through the bone and adjacent tissues. This generic type of device
may include signal analysis and display equipment, patient and
equipment supports, component parts, and accessories.
17. "C-arm fluoroscope"
means a fluoroscopic x-ray system in which the image receptor and the
x-ray tube housing assembly are connected or coordinated to maintain
a spatial relationship. Such a system allows a change in the
direction of the beam axis with respect to the patient without moving
the patient.
18. "Cantilevered
tabletop" means a tabletop designed such that the unsupported
portion can be extended at least 100 cm beyond the support.
19. "Cassette holder"
means a device, other than a spot-film device, that supports and/or
fixes the position of the image receptor during a radiographic
exposure.
20. "Coefficient of
variation (C)" means the ratio of the standard deviation to the
mean value of a population of observations. It is estimated using the
following equation:
where:
s = Estimated standard deviation of the population
= Mean value of
observations in sample
xi
= ith observation in sample
and
n = Number of observations in sample.
21. "Computed radiography
(CR; also see DR)" means a digital x-ray imaging method in which
a photo-stimulable phosphor is used to capture and store a latent
image. The latent image is read out by stimulating the phosphor with
a laser. Computed radiography systems may use cassettes to house the
phosphor, or it may be integrated into a digital radiography system.
22. "Computed tomography
(CT)" means the production of a tomogram by the acquisition and
computer processing of x-ray transmission data.
23. "Computed tomography
dose index" (CTDI) means the average absorbed dose, along the
z-axis, from a series of contiguous irradiations. It is measured from
one axial CT scan (one rotation of the x-ray tube), and is calculated
by dividing the integrated absorbed dose by the nominal total beam
collimation. The scattering media for CTDI consist of two (16 and 32
cm in diameter) polymethylmethacrylate (PMMA, e.g., acrylic or
Lucite) cylinders of 14 cm length. The equation is:
where:
D(z)
= the radiation dose profile along the z-axis,
N
= the number of tomographic sections imaged in a single axial
scan. This is equal to the number of data channels used in a
particular scan. The value of N may be less than or equal to the
maximum number of data channels available on the system, and
T
= the width of the tomographic section along the z-axis imaged by
one data channel. In multiple-detector-row (multislice) CT
scanners, several detector elements may be grouped together to
form one data channel. In single-detector-row (single-slice) CT,
the z-axis collimation (T) is the nominal scan width.
24. "CTDI 100 "
means the accumulated multiple scan dose at the center of a 100-mm
scan and underestimates the accumulated dose for longer scan lengths.
It is thus smaller than the equilibrium dose. The CTDI 100 ,
requires integration of the radiation dose profile from a single
axial scan over specific integration limits. In the case of CTDI 100 ,
the integration limits are +50 mm, which corresponds to the 100-mm
length of the commercially available “pencil” ionization chamber.
CTDI 100 is acquired using a 100-mm long, 3-cc active
volume CT “pencil” ionization chamber and one of the two standard
CTDI acrylic phantoms (16 and 32 cm diameter) and a stationary
patient table. The equation is:
25. "Cone Beam Computed
Tomography (CBCT)" is a volumetric imaging modality. Volumetric
data are acquired using two dimensional digital detector arrays, and
a cone-shaped x-ray beam (instead of fan-shaped) that rotates around
the patient. Reconstruction algorithms can be used to generate
images of any desired plane.
26. "Control panel"
means that part of the x-ray control upon which are mounted the
switches, knobs, pushbuttons, keypads, touchscreens, and other
hardware necessary for manually setting the technique factors.
27. "Cradle" means:
a. A removable device which
supports and may restrain a patient above an x-ray table; or
b. A device:
(1) Whose patient support
structure is interposed between the patient and the image receptor
during normal use;
(2) Which is equipped with
means for patient restraint; and
(3) Which is capable of
rotation about its long (longitudinal) axis.
28. "CT conditions of
operation" means all selectable parameters governing the
operation of a CT x-ray system including nominal tomographic section
thickness, filtration, and the technique factors as defined in § 4.2
of this Part.
29. "CT gantry"
means tube housing assemblies, beam-limiting devices, detectors, and
the supporting structures, frames, and covers which hold and/or
enclose these components within a computed tomography system.
30. "CT number"
means the number used to represent the x-ray attenuation associated
with each elemental area of the CT image:
where:
k = A constant
a
normal value of 1
000
when the Houndsfield scale of CT number is used; ux = Linear
attenuation coefficient of the material of interest; and uw =
Linear attenuation coefficient of water.
31. "Cumulative air
kerma" means the total air kerma accrued from the beginning of
an examination or procedure and includes all contributions from
fluoroscopic and radiographic irradiation.
32. "Detector" (See
"Radiation detector")
33. "Diagnostic reference
level (DRL)” means an investigational level used to identify
unusually high radiation doses or dose rates for common medical X-ray
imaging procedures. DRLs are suggested action levels above which a
facility should review its methods and determine if acceptable image
quality can be achieved at lower doses. DRLs should not be applied to
an individual patient.
34. "Diagnostic source
assembly" means the tube housing assembly with a beam-limiting
device attached.
35. "Diagnostic x-ray
system" means an x-ray system designed for irradiation of any
part of the human [or animal] body for the purpose of diagnosis or
visualization.
36. "Digital radiography
(DR)" means an x-ray imaging method (or radiography) which
produces a digital rather than analog image. DR includes both
computed radiography and direct digital radiography.
37. "Direct digital
radiography (DDR; also see CR and DR)" means an x-ray imaging
method in which a digital sensor, usually incorporating a thin-film
transistor, is used to capture an x-ray image. Some DDR systems use a
scintillator to convert x-rays to light and a photodiode array to
convert light to charge, while others use a photoconductor to convert
x-rays directly to charge, which is stored on the thin-film
transistor.
38. "Direct scattered
radiation" means that scattered radiation which has been
deviated in direction only by materials irradiated by the useful beam
(See "Scattered radiation").
39. "Direct supervision"
means a qualified practitioner must exercise general supervision and
be present in the facility and immediately available to furnish
assistance and direction throughout the performance of the procedure.
It does not mean that the licensed practitioner must be present in
the room when the procedure is being performed.
40. "Dose" means the
absorbed dose as defined by the International Commission on Radiation
Units and Measurements. The absorbed dose, D, is the quotient of de
by dm, where de is the mean energy imparted to matter of mass dm;
thus D=de/dm, in units of J/kg, where the special name of the unit of
absorbed dose is gray (Gy).
41. "Dose area product"
(DAP), a/k/a kerma-area product" (KAP) means the product of the
air kerma and the area of the irradiated field and is typically
expressed in Gy-cm2, so it does not change with distance from the
x-ray tube.
42. "Dose length product"
(DLP) means the indicator of the integrated radiation dose from a
complete CT examination. It addresses the total scan length by the
formula: DLP (mGy-cm) = CTDI vol (mGy) x scan length (cm).
43. "Dose profile"
means the dose as a function of position along a line.
44. "Effective dose (E)"
means the sum of the tissue-weighted equivalent doses for the
radiosensitive tissues and organs of the body. It is given by the
expression E = ?T (w T H T ), in which H T
is the equivalent dose in tissue or organ T and w T is the
tissue weighting factor for tissue or organ T. The unit of E and H T
is joule per kilogram (J·kg-1), with the special name sievert (Sv).
45. "Equipment" (See
"X-ray equipment") means x-ray equipment.
46. "Exposure (X)"
means the quotient of dQ by dm where dQ is the absolute value of the
total charge of the ions of one sign produced in air when all the
electrons and positrons liberated or created by photons in air of
mass dm are completely stopped in air; thus X=dQ/dm, in units of
C/kg. A second meaning of exposure is the process or condition
during which the x-ray tube produces x-ray radiation.
47. "Field emission
equipment" means equipment which uses an x-ray tube in which
electron emission from the cathode is due solely to the action of an
electric field.
48. "Filter" means
material placed in the useful beam to preferentially absorb selected
radiations.
49. "Fluoroscopic imaging
assembly" means a subsystem in which x-ray photons produce a set
of fluoroscopic images or radiographic images recorded from the
fluoroscopic image receptor. It includes the image receptor(s),
electrical interlocks, if any, and structural material providing
linkage between the image receptor and diagnostic source assembly.
50. "Fluoroscopic
irradiation time" means the cumulative duration during an
examination or procedure of operator-applied continuous pressure to
the device, enabling x-ray tube activation in any fluoroscopic mode
of operation.
51. "Fluoroscopically-Guided
Interventional (FGI) Procedures" means an interventional
diagnostic or therapeutic procedure performed via percutaneous or
other access routes, usually with local anesthesia or intravenous
sedation, which uses external ionizing radiation in the form of
fluoroscopy to localize or characterize a lesion, diagnostic site, or
treatment site, to monitor the procedure, and to control and document
therapy.
52. "Fluoroscopy"
means a technique for generating x-ray images and presenting them
simultaneously and continuously as visible images. This term has the
same meaning as the term “radioscopy” in the standards of the
International Electrotechnical Commission.
53. "Focal spot (actual)"
means the area projected on the anode of the x-ray tube bombarded by
the electrons accelerated from the cathode and from which the useful
beam originates.
54. "General purpose
radiographic x-ray system" means any radiographic x-ray system
which, by design, is not limited to radiographic examination of
specific anatomical regions.
55. "General supervision"
means the procedure is performed under the overall direction and
control of the qualified practitioner but who is not required to be
physically present during the performance of the procedure.
56. "Half-value layer
(HVL)" means the thickness of specified material which
attenuates the beam of radiation to an extent such that the AKR is
reduced by one-half of its original value. In this definition, the
contribution of all scattered radiation, other than any which might
be present initially in the beam concerned, is deemed to be excluded.
57. "Healing arts
screening" means the testing of human beings using x-ray
machines for the detection or evaluation of health indications when
such tests are not specifically and individually ordered by a
licensed practitioner of the healing arts legally authorized to
prescribe such x-ray tests for the purpose of diagnosis or treatment.
58. "Heat unit"
means a unit of energy equal to the product of the peak kilovoltage,
milliamperes, and seconds, i.e., kVp x mA x second.
59. "Image intensifier"
means a device, installed in its housing, which instantaneously
converts an x-ray pattern into a corresponding light image of higher
intensity.
60. "Image receptor"
means any device, such as a fluorescent screen, radiographic film,
x-ray image intensifier tube, solid-state detector, or gaseous
detector which transforms incident x-ray photons either into a
visible image or into another form which can be made into a visible
image by further transformations. In those cases where means are
provided to preselect a portion of the image receptor, the term
“image receptor” shall mean the preselected portion of the
device.
61. "Irradiation"
means the exposure of matter to ionizing radiation.
62. "Isocenter"
means the center of the smallest sphere through which the beam axis
passes when the equipment moves through a full range of rotations
about its common center.
63. "Kerma" means
the quantity defined by the International Commission on Radiation
Units and Measurements. The kerma, K, is the quotient of dEtr by dm,
where dEtr is the sum of the initial kinetic energies of all the
charged particles liberated by uncharged particles in a mass dm of
material; thus K=dEtr/dm, in units of J/kg, where the special name
for the unit of kerma is gray (Gy). When the material is air, the
quantity is referred to as "air kerma."
64. "Kerma-area product
(KAP) " (See "dose area product")
65. "Kilovolts peak"
(See "Peak tube potential").
66. "kV" means
kilovolts.
67. "kWs" means
kilowatt second.
68. "Last-image hold
(LIH) radiograph" means an image obtained either by retaining
one or more fluoroscopic images, which may be temporarily integrated,
at the end of a fluoroscopic exposure or by initiating a separate and
distinct radiographic exposure automatically and immediately in
conjunction with termination of the fluoroscopic exposure.
69. "Lead equivalent"
means the thickness of lead affording the same attenuation, under
specified conditions, as the material in question.
70. "Leakage radiation"
means radiation emanating from the diagnostic source assembly except
for:
a. The useful beam; and
b. Radiation produced when the
exposure switch or timer is not activated.
71. "Leakage technique
factors" means the technique factors associated with the
diagnostic source assembly which are used in measuring leakage
radiation. They are defined as follows:
a. For diagnostic source
assemblies intended for capacitor energy storage equipment, the
maximum-rated peak tube potential and the maximum-rated number of
exposures in an hour for operation at the maximum-rated peak tube
potential with the quantity of charge per exposure being 10
millicoulombs (or 10 mAs) or the minimum obtainable from the unit,
whichever is larger;
b. For diagnostic source
assemblies intended for field emission equipment rated for pulsed
operation, the maximum-rated peak tube potential and the
maximum-rated number of x-ray pulses in an hour for operation at the
maximum-rated peak tube potential; and
c. For all other diagnostic
source assemblies, the maximum-rated peak tube potential and the
maximum-rated continuous tube current for the maximum-rated peak tube
potential.
72. "Light field"
means that area of the intersection of the light beam from the
beam-limiting device and one of the set of planes parallel to and
including the plane of the image receptor, whose perimeter is the
locus of points at which the illumination is one-fourth of the
maximum in the intersection.
73. "Line-voltage
regulation" means the difference between the no-load and the
load line potentials expressed as a percent of the load line
potential; that is, Percent line-voltage regulation = 100 (Vn-Vl)/Vl,
where: Vn = No-load line potential; and Vl = Load line potential.
74. "mA" means
milliampere.
75. "mAs" means
milliampere second.
76. "Mode of operation"
means, for fluoroscopic systems, a distinct method of fluoroscopy or
radiography provided by the manufacturer and selected with a set of
several technique factors or other control settings uniquely
associated with the mode. The set of distinct technique factors and
control settings for the mode may be selected by the operation of a
single control. Examples of distinct modes of operation include
normal fluoroscopy (analog or digital), high-level control
fluoroscopy, cineradiography (analog and digital), digital
subtraction angiography, electronic radiography using the
fluoroscopic image receptor, and photospot recording. In a specific
mode of operation, certain system variables affecting kerma, AKR, or
image quality, such as image magnification, x-ray field size, pulse
rate, pulse duration, number of pulses, source-image receptor
distance (SID), or optical aperture, may be adjustable or may vary;
their variation per se does not comprise a mode of operation
different from the one that has been selected.
77. "Multiple tomogram
system" means a computed tomography x-ray system which obtains
x-ray transmission data simultaneously during a single scan to
produce more than one tomogram.
78. “Noise” means the
standard deviation of the fluctuations in CTN expressed as a
percentage of the attenuation coefficient of water. Its estimate
(Sn) is calculated using the following expression:
where:
µx = Linear attenuation coefficient of the material of interest,
µw = Linear attenuation coefficient of water, and s = Estimated
standard deviation of the CTN of picture elements in a specified
area of the CT image.
79. "Nominal tomographic
section thickness" means the full width at half-maximum of the
sensitivity profile taken at the center of the cross-sectional volume
over which x-ray transmission data are collected.
80. "Notification value"
means a protocol-specific dose index (e.g. CTDI vol (mGy) or
of DLP(mGy-cm)) that is set by the registrant to trigger a
notification to the CT operator prior to scanning when the dose index
exceeds the established range for the examination.
81. "Patient" means
an individual or animal subjected to healing arts examination,
diagnosis or treatment.
82. "Picture element"
means an elemental area of a tomogram.
83. "PBL" See
"Positive beam limitation."
84. "Peak tube potential"
means the maximum value of the potential difference across the x-ray
tube during an exposure.
85. "Personal
supervision" means a qualified practitioner must exercise
General Supervision and be present in the room or adjacent control
area during the performance of the procedure.
86. "Phantom" means
a volume of material behaving in a manner similar to tissue with
respect to the attenuation and scattering of radiation. This requires
that both the atomic number (Z) and the density of the material be
similar to that of tissue.
87. "Photostimulable
storage phosphor (PSP)" means a material used to capture and
store radiographic images in computed radiography systems.
88. "Pitch" means
the table incrementation, in CT, per x-ray tube rotation, divided by
the nominal x-ray beam width at isocenter.
89. "Position indicating
device (PID)" means a device on dental x-ray equipment used to
indicate the beam position and to establish a definite source-surface
(skin) distance. It may or may not incorporate or serve as a
beam-limiting device.
90. "Positive beam
limitation" means the automatic or semi-automatic adjustment of
an x-ray beam to the size of the selected image receptor, whereby
exposures cannot be made without such adjustment.
91. "Primary protective
barrier" means the material, excluding filters, placed in the
useful beam to reduce the radiation exposure (beyond the patient and
cassette holder) for protection purposes.
92. "Protective apron"
means an apron made of radiation absorbing materials used to reduce
radiation exposure.
93. "Protocol" means
a collection of settings and parameters that fully describe an
examination.
94. "Pulsed mode"
means operation of the x-ray system such that the x-ray tube current
is pulsed by the x-ray control to produce one or more exposure
intervals of duration less than one-half second.
95. "Quality Assurance"
means a program providing for verification by written procedures such
as testing, auditing, and inspection to ensure that deficiencies,
deviations, defective equipment, or unsafe practices, or a
combination thereof, relating to the use, disposal, management, or
manufacture of radiation devices are identified, promptly corrected,
and reported to the appropriate regulatory authorities as required.
96. “Qualified Medical
Physicist” (for activities authorized pursuant to this Part) means
an individual registered to provide Radiation Physics Services
(Diagnostic X-ray Physics Services) in accordance with § 3.6 of this
Subchapter.
97. "Radiation detector"
means a device which in the presence of radiation provides a signal
or other indication suitable for use in measuring one or more
quantities of incident radiation.
98. “Radiation medical
event” means an event that meets the criteria in § 4.4.14(A) of
this Part.
99. "Radiation Protocol
Committee (RPC)" means the representative group of qualified
individuals in a CT or FGI facility responsible for the ongoing
review and management of CT or FGI protocols to ensure that exams
being performed achieve the desired diagnostic image quality at the
lowest radiation dose possible while properly exploiting the
capabilities of the equipment being used.
100. "Radiation therapy
simulation system" means a radiographic or fluoroscopic x-ray
system intended for localizing the volume to be exposed during
radiation therapy and confirming the position and size of the
therapeutic irradiation field.
101. "Radiograph"
means an image receptor on which the image is created directly or
indirectly by an x-ray pattern and results in a permanent record.
102. "Radiography"
means a technique for generating and recording an x-ray pattern for
the purpose of providing the user with an image(s) after termination
of the exposure.
103. "Recording"
means producing a retrievable form of an image resulting from x-ray
photons.
104. "Reference plane"
means a plane which parallel to and which can be offset (as specified
in manufacturer information provided to users) from the location of
the tomographic plane(s).
105. “Registrant” means
any person who is registered with the Agency and is legally obligated
to register with the Agency pursuant to this Subchapter and the Act.
106. “Registration” means
registration with the Agency pursuant to this Subchapter and the Act.
107. "Scan" means
the complete process of collecting x-ray transmission data for the
production of a tomogram. Data may be collected simultaneously
during a single scan for the production of one or more tomograms.
108. "Scan increment"
means the amount of relative displacement of the patient with respect
to the CT x-ray system between successive scans measured along the
direction of such displacement.
109. "Scan sequence"
means a pre-selected set of two or more scans performed consecutively
under pre-selected CT conditions of operation.
110. "Scan time"
means the time elapsed during the accumulation of x-ray transmission
data for a single scan.
111. "Scattered
radiation" means radiation that, during passage through matter,
has been deviated in direction (See "Direct scattered
radiation").
112. "Sensitivity
profile" means the relative response of the CT x-ray system as a
function of position along a line perpendicular to the tomographic
plane.
113. "Single tomogram
system" means a CT x-ray system which obtains x-ray transmission
data during a scan to produce a single tomogram.
114. "Shutter" means
a device attached to the tube housing assembly which can intercept
the entire cross sectional area of the useful beam and which has a
lead equivalency not less than that of the tube housing assembly.
115. "Size-specific dose
estimate" (SSDE) means a patient dose estimate which takes into
consideration corrections based on the size of the patient, using
linear dimensions measured on the patient or patient images.
116. "Source" means
the focal spot of the x-ray tube.
117. "Source-image
receptor distance" (SID) means the distance from the source to
the center of the input surface of the image receptor.
118. "Source-skin
distance" (SSD) means the distance from the source to the center
of the entrant x-ray field in the plane tangent to the patient skin
surface.
119. "Spot-film"
means a radiograph which is made during a fluoroscopic examination to
permanently record conditions which exist during that fluoroscopic
procedure. Digital image receptors used in place of film with
spot-film devices should be considered "spot-film".
120. "Spot-film device"
means a device intended to transport and/or position a radiographic
image receptor between the x-ray source and fluoroscopic image
receptor. It includes a device intended to hold a cassette over the
input end of the fluoroscopic image receptor for the purpose of
producing a radiograph.
121. "Stray radiation"
means the sum of leakage and scattered radiation.
122. "Substantial
radiation dose level (SRDL)” means an appropriately-selected dose
used to trigger additional dose-management actions during a procedure
and medical follow-up for a radiation level that might produce a
clinically-relevant injury in an average patient.
123. “Technique factors”
means the following conditions of operation:
a. For capacitor energy
storage equipment, peak tube potential in kilovolts (kV) and quantity
of charge in milliampere-seconds (mAs);
b. For field emission
equipment rated for pulsed operation, peak tube potential in kV, and
number of X-ray pulses;
c. For CT X-ray systems
designed for pulsed operation, peak tube potential in kV, scan time
in seconds, and either tube current in mA, X-ray pulse width in
seconds, and the number of x-ray pulses per scan, or the product of
tube current, X-ray pulse width, and the number of X-ray pulses in
mAs;
d. For CT X-ray systems not
designed for pulsed operation, peak tube potential in kV, and either
tube current in mA and scan time in seconds, or the product of tube
current and exposure time in mAs and the scan time when the scan time
and exposure time are equivalent; and
e. For all other equipment,
peak tube potential in kV, and either tube current in mA and exposure
time in seconds, or the product of tube current and exposure time in
mAs.
124. "Tomogram"
means the depiction of the x-ray attenuation properties of a section
through the body.
125. "Tomographic plane"
means that geometric plane which the manufacturer identified as
corresponding to the output tomogram.
126. "Tomographic
section" means the volume of an object whose x-ray attenuation
properties are imaged in a tomogram.
127. "Tube" means an
x-ray tube, unless otherwise specified.
128. "Tube housing
assembly" means the tube housing with tube installed. It
includes high-voltage and/or filament transformers and other
appropriate elements when such are contained within the tube housing.
129. "Unintended"
radiation dose in diagnostic or interventional x-ray means a patient
radiation dose resulting from a human error or equipment malfunction
during the procedure.
130. "Useful beam"
means the radiation which passes through the tube housing port and
the aperture of the beam limiting device when the exposure switch or
timer is activated.
131. "Visible area"
means that portion of the input surface of the image receptor over
which incident x-ray photons are producing a visible image.
132. "Volume Computed
Tomography Dose Index (CTDI vol )" means a radiation
dose parameter derived from the CTDI w (weighted or average
CTDI given across the field of view). The formula is: CTDI vol
= (N)(T)(CTDI w )/I, where N = number of simultaneous axial
scans per x-ray source rotation, T = thickness of one axial scan
(mm), and I = table increment per axial scan (mm). Thus, CTDI vol
= CTDI w / pitch.
133. "Weighted Computed
Tomography Dose Index (CTDI w )” means the estimated
average CTDI 100 across the field of view (FOV). The
equation is: Where 1/3 and 2/3 approximate the relative areas
represented by the center and edge values derived using the 16 or 32
cm acrylic phantom. CTDI w uses CTDI 100 and an
f-factor for air (0.87 rad/R or 1.0 mGy/mGy).
134. "X-ray control"
means a device which controls input power to the x-ray high-voltage
generator and/or the x-ray tube. It includes equipment such as
timers, phototimers, automatic brightness stabilizers, and similar
devices, which control the technique factors of an x-ray exposure.
135. "X-ray exposure
control" means a device, switch, button or other similar means
by which an operator initiates and/or terminates the radiation
exposure. The x-ray exposure control may include such associated
equipment as timers and back-up timers.
136. "X-ray equipment"
means an x-ray system, subsystem, or component thereof. Types of
x-ray equipment are as follows:
a. "Mobile x-ray
equipment" means x-ray equipment mounted on a permanent base
with wheels and/or casters for moving while completely assembled;
b. "Portable x-ray
equipment" means x-ray equipment designed to be hand-carried;
and
c. "Stationary x-ray
equipment" means x-ray equipment which is installed in a fixed
location.
d. "Hand-held x-ray
equipment" means x-ray equipment that is designed to be
hand-held during operation.
137. "X-ray field"
means that area of the intersection of the useful beam and any one of
a set of planes parallel to and including the plane of the image
receptor, whose perimeter is the locus of points at which the AKR is
one-fourth of the maximum in the intersection.
138. "X-ray high-voltage
generator" means a device which transforms electrical energy
from the potential supplied by the x-ray control to the tube
operating potential. The device may also include means for
transforming alternating current to direct current, filament
transformers for the x-ray tube(s), high-voltage switches, electrical
protective devices, and other appropriate elements.
139. "X-ray system"
means an assemblage of components for the controlled production of
x-rays. It includes minimally an x-ray high-voltage generator, an
x-ray control, a tube housing assembly, a beam-limiting device, and
the necessary supporting structures. Additional components which
function with the system are considered integral parts of the system
140. "X-ray table"
means a patient support device with its patient support structure
(tabletop) interposed between the patient and the image receptor
during radiography and/or fluoroscopy. This includes, but is not
limited to, any stretcher equipped with a radiolucent panel and any
table equipped with a cassette tray (or bucky), cassette tunnel,
fluoroscopic image receptor, or spot-film device beneath the
tabletop.
141. "X-ray tube"
means any electron tube which is designed for the conversion of
electrical energy into x-ray energy.
4.3 General and Administrative
Requirements
4.3.1 Administrative
Controls
The registrant shall be
responsible for directing the operation of the X-ray system(s) under
their administrative control. The registrant or the registrant's
agent shall assure that the requirements of this Subchapter are met
in the operation of the X-ray system(s).
4.3.2 Operation Prohibited
An X-ray system which does
not meet the provisions of this Part shall not be operated for
diagnostic purposes.
4.3.3 Individuals Operating
X-ray Systems for Healing Arts Use
A. Individuals who will be
operating the X-ray systems for healing arts use shall possess a
current license in accordance with Licensure of Radiographers,
Nuclear Medicine Technologists, Radiation Therapists and Radiologist
Assistants [Subchapter 05 Part 34 of this Chapter], unless the
individual is specifically exempted from licensure by said
regulations. Individuals who will be operating the X-ray systems and
who are not subject to licensure under Subchapter 05 Part 34 of this
Chapter shall be adequately instructed in the safe operating
procedures and be competent in the safe use of the equipment. As a
minimum, such instruction shall consist of subjects outlined in §
4.12 of this Part.
B. The names and
qualifications of all personnel operating X-ray equipment for healing
arts use must be kept on file for Agency inspection at each facility
location.
C. All individuals operating
fluoroscopic X-ray systems shall have completed at least the
following training before using fluoroscopy independently:
1. Biological effects of
X-ray;
2. Principles of radiation
protection;
3. Factors affecting
fluoroscopic outputs;
4. Dose reduction techniques
for fluoroscopic X-ray systems;
5. Principles and operation of
the specific fluoroscopic X-ray system(s) to be used;
6. Fluoroscopic and
fluorographic outputs of each mode of operation on the system(s) to
be used clinically; and
7. Applicable requirements of
this Subchapter.
D. The registrant shall either
provide in-service training for all operators of fluoroscopic x-ray
systems used for high dose, high risk procedures, as defined in §
4.5.13 of this Part, at intervals not to exceed twenty-four (24)
months or require evidence of continuing medical education, in
fluoroscopic radiation safety and patient dose management at
intervals not to exceed twenty-four (24) months.
E. Documentation pertaining to
the requirements of §§ 4.3.3(c) and (d) of this Part shall be
maintained for review for three (3) years.
4.3.4 Written Technique
Information
A. Written technique
information shall be provided in the vicinity of the diagnostic X-ray
system's control panel, which specifies, for all examinations
performed with that system, the following information:
1. Patient's body part and
anatomical size, or body part thickness, or age (for pediatrics),
versus technique factors to be utilized;
2. Equivalent manual technique
information if AEC is not available;
3. Type and size of the image
receptor combination to be used, if any;
4. Source to image receptor
distance to be used (except for dental intraoral radiography, which
shall list cone length to be used);
5. Type and location of
placement of patient shielding (e.g., gonad, thyroid, lap apron,
etc.); and
6. For mammography, indication
of kVp/target/filter combination and, if phototimed setting is used,
the density setting.
4.3.5 Written Safety
Procedures
The registrant of a facility
shall create and make available to X-ray operators written safety
procedures, including patient holding and any restrictions of the
operating technique required for the safe operation of the particular
X-ray system. The operator shall be able to demonstrate familiarity
with these procedures.
4.3.6 Room Occupancy During
Radiographic Exposure
A. Except for patients who
cannot be moved out of the room, only the staff, ancillary personnel
or other persons required for the medical procedure or training shall
be in the room during the radiographic exposure. Other than the
patient being examined:
1. All individuals shall be
positioned such that no part of the body will be struck by the useful
beam unless protected by not less than 0.5 millimeter lead equivalent
material.
2. The X-ray operator, other
staff, ancillary personnel and other persons required for the medical
procedure shall be protected from the direct scatter radiation by
protective aprons or whole body protective barriers of not less than
0.25 millimeter lead equivalent material.
3. Human patients who cannot
be removed from the room shall be protected from the direct scatter
radiation by whole body protective barriers of not less than 0.25
millimeter lead equivalent or shall be so positioned that the nearest
portion of the body is at least two (2) meters from both the tube
head and the nearest edge of the image receptor.
4. Written safety procedures,
as required by § 4.3.5 of this Part, shall describe how the
requirements of this section will be met when using mobile or
portable X-ray systems.
4.3.7 Gonadal Shielding
Gonadal shielding of not less
than 0.5 millimeter lead equivalent material shall be used for
patients who have not passed the reproductive age during radiographic
procedures in which the gonads are in the useful beam, except for
cases in which this would interfere with the diagnostic procedure.
4.3.8 Non-Healing Arts
Exposure Prohibited
Individuals shall not be
exposed to the useful beam except for healing arts purposes and
unless such exposure has been ordered in writing by a licensed
practitioner of the healing arts. This provision specifically
prohibits deliberate exposure of an individual for training,
demonstration or other non-healing-arts purposes, and exposure of an
individual for the purpose of healing arts screening except as
authorized by § 4.3.12 of this Part.
4.3.9 When a Patient or
Image Receptor Must be Provided with Auxiliary Support During a
Radiation Exposure
A. Mechanical holding devices
shall be used when the technique permits. The written safety
procedures, required by § 4.3.5 of this Part, shall list individual
projections where holding devices cannot be utilized;
B. Written safety procedures,
as required by § 4.3.5 of this Part, shall indicate the requirements
for selecting a holder and the procedure the holder shall follow;
C. The human holder shall be
instructed in personal radiation safety and protected as required by
§ 4.3.6 of this Part;
D. No individual shall be used
routinely to hold image receptor or patients;
E. In those cases where the
patient must hold the image receptor, except during dental
examinations covered by this Part, any portion of the body other than
the area of clinical interest struck by the useful beam shall be
protected by not less than 0.5 millimeter lead equivalent material;
and
F. Each facility shall have
protective aprons and gloves available in sufficient numbers to
provide protection for all personnel who are involved with X-ray
operations and who are otherwise not shielded.
G. A record shall be made of
the examination and shall include the name of the human holder; date
of the examination, number of exposures and technique factors
utilized for the exposure(s).
4.3.10 Procedures and
Auxiliary Equipment Designed to Minimize Patient and Personnel
Exposure Commensurate with The Needed Diagnostic Information Shall Be
Utilized
A. The fastest imaging system
consistent with the diagnostic objective of the examinations shall be
used. Film cassettes without intensifying screens shall not be used
for any diagnostic radiological imaging, with the exception of
veterinary radiography and standard film packets for intraoral use in
dental radiography.
B. The radiation exposure to
the patient shall be the minimum exposure required to produce images
of good diagnostic quality.
C. Portable or mobile
equipment shall be used only for examinations where it is impractical
to transfer the patient(s) to a stationary X-ray installation.
D. Facilities shall establish
and implement a quality assurance program for X-ray film processing,
whether processing is manual or automatic.
E. X-ray Film Processing
Facilities and Practices. Each installation using a radiographic
X-ray system and using analog image receptors (e.g., radiographic
film) shall have available suitable equipment for handling and
processing radiographic film in accordance with the following
provisions.
1. Manual Processing of Films
a. Processing of film: The
temperature of solutions in the tanks shall be maintained within the
range of 60 degrees F to 80 degrees F (16 degrees C to 27 degrees C).
Film shall be developed in accordance with the time-temperature
relationships recommended by the film manufacturer.
b. Devices shall be utilized
which will:
(1) Indicate the actual
temperature of the developer; and
(2) Give an audible or visible
signal indicating the termination of a preset time.
c. Processing tanks shall be
constructed of mechanically rigid, corrosion resistant material.
2. Automatic Processors and
Other Closed Processing Systems
a. Films shall be processed in
accordance with the time temperature relationships recommended by the
film manufacturer; and
b. Processing deviations from
the requirements of § 4.3.10(E)(2)(a) of this Part shall be
documented by the registrant in such manner that the requirements are
shown to be met or exceeded (e.g., extended processing, and special
rapid chemistry).
F. If grids are used between
the patient and the image receptor to decrease scatter to the film
and improve contrast, the grid shall:
1. Be positioned properly
(i.e., tube side facing the proper direction) and grid centered to
the central ray.
2. If of the focused type, be
of the proper focal distance for the SID being used.
G. Other Requirements
1. Pass boxes, if provided,
shall be so constructed as to exclude light from the darkroom when
cassettes are placed in or removed from the boxes, and shall
incorporate adequate shielding from stray radiation to prevent
exposure of undeveloped film.
2. The darkroom shall be
light-tight and use proper safelighting such that any film type in
use exposed in a cassette to x-radiation sufficient to produce an
optical density from 1 to 2 when processed shall not suffer an
increase in density greater than 0.05 when exposed in the darkroom
for two (2) minutes with all safelights on. If used, daylight film
handling systems shall preclude fogging of the film.
3. Darkrooms typically used by
more than one individual shall be provided a method to prevent
accidental entry while undeveloped films are being handled or
processed.
4. Film shall be stored in a
cool, dry place and shall be protected from exposure to stray
radiation. Film in open packages shall be stored in a light-tight
container.
5. Film cassettes and
intensifying screens shall be inspected periodically and shall be
cleaned and replaced as necessary to assure radiographs of good
diagnostic quality.
6. Outdated x-ray film shall
not be used for diagnostic radiographs, unless the film has been
stored in accordance with the manufacturer's recommendations and a
sample of the film passes a sensitometric test for normal ranges of
base plus fog and speed.
7. Film developing solutions
shall be prepared in accordance with the directions given by the
manufacturer, and shall be maintained in strength by replenishment or
renewal so that full development is accomplished within the time
specified by the manufacturer.
H. The tube housing and the
position indicating device (PID) for a permanently mounted intraoral
dental system shall not be hand-held during an exposure. § 4.13 of
this Part specifies requirements for the use of intraoral dental
radiographic units designed to be hand-held during patient
examination.
I. Dental fluoroscopy without
image intensification shall not be used.
4.3.11 Additional
Compliance Required
All individuals who are
associated with the operation of an X-ray system are subject to the
applicable requirements of Parts 1 and 2 of this Subchapter.
4.3.12 Healing Arts
Screening
Any person proposing to
conduct a healing arts screening program shall not initiate such a
program without prior approval of the Agency. When requesting such
approval, that person shall submit the information outlined in §
4.11 of this Part. If any information submitted to the Agency
becomes invalid or outdated, the Agency shall be immediately
notified.
4.3.13 Information and
Maintenance Record and Associated Information.
A. The registrant shall
maintain the following information in a separate file or package in
chronological order for each X-ray system, for inspection by the
Agency:
1. Maximum rating of technique
factors;
2. Model and serial numbers of
all major components, and user's manuals for those components;
3. Aluminum equivalent
filtration in the useful beam, including any routine variation;
4. Tube rating charts and
cooling curves;
5. Records of surveys,
calibrations, maintenance, and modifications performed on the X-ray
system(s) with the names of persons who performed such services;
6. A scale drawing of the room
in which a stationary X-ray system is located with such drawing
indicating the current use of areas adjacent to the room and an
estimate of the extent of occupancy by an individual in such areas.
In addition, the drawing shall include the results of a survey for
radiation levels present at the operator's position and at pertinent
points outside the room at specified test conditions; or the type and
thickness of materials, or lead equivalency, of each protective
barrier.
7. A copy of all
correspondence with this Agency regarding that X-ray system.
4.3.14 X-Ray Utilization
Log
A. Except for veterinary
facilities, each facility shall maintain a record containing the
patient's name, the type of examinations, and the dates the
examinations were performed. The record shall also include the
following information:
1. Name of the licensed
practitioner of the healing arts ordering the examination.
2. Name(s) of individuals who
performed the examination.
3. Any deviation from the
standard procedure as specified on the technique chart, including all
repeat exposures.
4. When applicable, the fluoro
recordkeeping requirements of § 4.5.3(E) of this Part.
5. When applicable, the X-ray
system used.
6. When the patient or image
receptor must be provided with human auxiliary support, the name of
the human holder.
B. X-ray utilization logs
shall be maintained for a minimum of five (5) years following the
examination or treatment of adult patients. Records of examination
or treatment of minors shall be maintained for a minimum of five (5)
years beyond the age of majority.
C. If X-ray utilization logs
are stored electronically, records shall be maintained in a manner
that will allow retrieval of records for any specified time period.
4.3.15 Report and
Notification of a Dose to an Embryo/Fetus
A. A registrant shall report
any dose to an embryo/fetus that is greater than 50 mSv (5 rem) dose
equivalent that is a result of an administration of radiation to a
pregnant individual unless the dose to the embryo/fetus was
specifically approved, in advance, by the referring physician.
B. The registrant shall notify
the Agency by telephone no later than the next business day after
discovery of a dose to the embryo/fetus that requires a report in §
4.3.15(A) of this Part.
C. The registrant shall submit
a written report, prepared by a Qualified Medical Physicist, to the
Agency within fifteen (15) business days after discovery of a dose to
the embryo/fetus that requires a report in § 4.3.15(A) of this Part.
1. The written report shall
include:
a. The registrant's name and
registration number;
b. The name of the referring
physician;
c. A brief description of the
event;
d. Why the event occurred;
e. The effect, if any, on the
embryo/fetus;
f. What actions, if any, have
been taken or are planned to prevent recurrence; and
g. Certification that the
registrant notified the pregnant individual (or the pregnant
individual's responsible relative or guardian), and if not, why not.
2. The report must not contain
the individual's name or any other information that could lead to
identification of the individual.
D. The registrant shall
provide notification of the event to the referring physician and also
notify the pregnant individual, no later than twenty-four (24) hours
after discovery of an event that would require reporting under §
4.3.15(A) of this Part, unless the referring physician personally
informs the registrant either that he or she will inform the pregnant
individual or that, based on medical judgment, telling the pregnant
individual would be harmful. The registrant is not required to notify
the pregnant individual without first consulting with the referring
physician. If the referring physician or pregnant individual cannot
be reached within twenty-four (24) hours, the registrant shall make
the appropriate notifications as soon as possible thereafter. The
registrant may not delay any appropriate medical care for the
embryo/fetus, including any necessary remedial care as a result of
the event, because of any delay in notification. To meet the
requirements of this paragraph, the notification may be made to the
pregnant individual's responsible relative or guardian instead of the
pregnant individual. If a verbal notification is made, the registrant
shall inform the pregnant individual, or the pregnant individual's
responsible relative or guardian, that a written description of the
event can be obtained from the registrant upon request. The
registrant shall provide such a written description if requested.
E. A registrant shall:
1. Annotate a copy of the
report provided to the Agency with the:
a. Name of the pregnant
individual who is the subject of the event; and
b. Social security number or
other identification number, if one has been assigned, of the
pregnant individual who is the subject of the event; and
2. Provide a copy of the
annotated report to the referring physician, if other than the
registrant, no later than fifteen (15) days after the discovery of
the event.
4.4 General Requirements for All
Diagnostic X-Ray Systems
4.4.1 Applicability
In addition to other
requirements of this Part, all diagnostic X-ray systems shall meet
the requirements of § 4.4 of this Part.
4.4.2 Maintaining
Compliance
Diagnostic X-ray systems and
their associated components used on humans and certified pursuant to
the Federal X-ray Equipment Performance Standard (21 C.F.R. Part
1020) shall be maintained in compliance with applicable requirements
of that standard.
4.4.3 Warning Label
The control panel containing
the main power switch shall bear the warning statement, legible and
accessible to view: "WARNING: This X-ray unit may be dangerous
to patient and operator unless safe exposure factors and operating
instructions and maintenance schedules are observed".
4.4.4 Battery Charge
Indicator
On battery-powered X-ray
generators, visual means shall be provided on the control panel to
indicate whether the battery is in a state of charge adequate for
proper operation.
4.4.5 Leakage Radiation
from the Diagnostic Source Assembly
The leakage radiation from
the diagnostic source assembly measured at a distance of one (1)
meter in any direction from the source shall not exceed 0.88
milligray (mGy) air kerma [100 milliroentgen (mR) exposure] in one
(1) hour when the X-ray tube is operated at its leakage technique
factors. If the maximum rated peak tube potential of the tube
housing assembly is greater than the maximum rated peak tube
potential for the diagnostic source assembly, positive means shall be
provided to limit the maximum X-ray tube potential to that of the
diagnostic source assembly. Compliance shall be determined by
measurements averaged over an area of one-hundred square centimeters
(100 cm2) with no linear dimension greater than twenty (20)
centimeters.
4.4.6 Radiation from
Components Other Than the Diagnostic Source Assembly
The radiation emitted by a
component other than the diagnostic source assembly shall not exceed
an air kerma of eighteen (18) µgray [two (2) milliroentgens
exposure] in one (1) hour at five (5) centimeters from any accessible
surface of the component when it is operated in an assembled X-ray
system under any conditions for which it was designed. Compliance
shall be determined by measurements averaged over an area of
one-hundred square centimeters (100 cm2) with no linear dimension
greater than twenty centimeters (20 cm).
4.4.7 Beam Quality
A. Half-Value Layer (HVL)
1. The HVL of the useful beam
for a given X-ray tube potential shall not be less than the values
shown in § 4.4.7(B) of this Part [Table 1] under the heading
“Specified Dental Systems,” for any dental X-ray system designed
for use with intraoral image receptors and manufactured after
December 1, 1980; under the heading,” Other X-Ray Systems2” for
any dental X-ray system designed for use with intraoral image
receptors and manufactured before or on December 1,1980, and all
other X-ray systems subject to this section and manufactured before
June 10, 2006; and under the heading, “Other X-Ray Systems” for
all X-ray systems, except dental X-ray systems designed for use with
intraoral image receptors, subject to this section and manufactured
on or after June 10, 2006. If it is necessary to determine such
half-value layer at an X-ray tube potential which is not listed in §
4.4.7(B) of this Part [Table 1], linear interpolation or
extrapolation may be made. Positive means shall be provided to ensure
that at least the minimum filtration needed to achieve beam quality
requirements is in the useful beam during each exposure. In the case
of a system, which is to be operated with more than one thickness of
filtration, this requirement can be met by a filter interlocked with
the kilovoltage selector which will prevent X-ray emissions if the
minimum required filtration is not in place.
2. Optional Filtration.
Fluoroscopic systems manufactured on or after June 10, 2006,
incorporating an X-ray tube(s) with a continuous output of one (1)
kilowatt or more and an anode heat storage capacity of one-million
(1,000,000) heat units or more shall provide the option of adding
X-ray filtration to the diagnostic source assembly in addition to the
amount needed to meet the half-value layer provisions of §
4.4.7(A)(1) of this Part. The selection of this additional X-ray
filtration shall be either at the option of the user or automatic as
part of the selected mode of operation. A means of indicating which
combination of additional filtration is in the X-ray beam shall be
provided.
B. Table 1 - X-Ray Tube
Voltage (kilovolt peak)
Design
Operating Range
Measured
Operating Potential
Minimum
HVL (mm in Aluminum)
Specified
Dental Systems (Dental X-ray systems designed for use with
intraoral image receptors and manufactured after December 1,
1980)
Other
X-Ray Systems (Dental X-ray systems designed for use with
intraoral image receptors and manufactured before or on December
1, 1980, and all other X-ray systems subject to this section and
manufactured before June 10, 2006)
Other
X-Ray Systems (All X-ray systems, except dental X-ray systems
designed for use with intraoral image receptors, subject to this
section and manufactured on or after June 10, 2006)
Below
51
30
1.5
0.3
0.3
40
1.5
0.4
0.4
50
1.5
0.5
0.5
51
to 70
51
1.5
1.2
1.3
60
1.5
1.3
1.5
70
1.5
1.5
1.8
71
2.1
2.1
2.5
Above
70
80
2.3
2.3
2.9
90
2.5
2.5
3.2
100
2.7
2.7
3.6
110
3.0
3.0
3.9
120
3.2
3.2
4.3
130
3.5
3.5
4.7
140
3.8
3.8
5.0
150
4.1
4.1
5.4
C. Beryllium window tubes
shall have a minimum of 0.5 millimeter aluminum equivalent filtration
permanently installed in the useful beam.
D. Measuring Compliance. For
capacitor energy storage equipment, compliance shall be determined
with the maximum selectable quantity of charge per exposure.
E. Aluminum Equivalent of
Material Between Patient and Image Receptor. Except when used in a
CT X-ray system, the aluminum equivalent of each of the items listed
in § 4.3.7(F) of this Part [Table 2], which are used between the
patient and the image receptor, shall not exceed the indicated
limits. Compliance shall be determined by X-ray measurements made at
a potential of one-hundred (100) kilovolts peak and with an X-ray
beam that has an HVL specified in § 4.3.7(B) of this Part [Table 1]
for the potential. This requirement applies to front panel(s) of
cassette holders and film changers provided by the manufacturer for
patient support or for prevention of foreign object intrusions. It
does not apply to screens and their associated mechanical support
panels or grids.
F. Table 2 - Maximum Aluminum
Equivalent (millimeters)
ITEM
Maximum
Aluminum Equivalent (millimeters)
1.
Front panel(s) of cassette holders (total of all)
1.2
2.
Film panel(s) of film changer (total of all)
1.2
3.
Cradle
2.3
4.
Tabletop, stationary, without articulated joints
1.2
5.
Tabletop, movable, without articulated joint(s) (including
stationary subtop)
1.7
6.
Tabletop, with radiolucent panel having one articulated joint
1.7
7.
Tabletop, with radiolucent panel having two or more articulated
joints
2.3
8.
Tabletop, cantilevered
2.3
9.
Tabletop, radiation therapy simulator
5.0
G. Modification of Certified
Diagnostic X-ray Components and Systems
1. Diagnostic X-ray components
and systems certified in accordance with 21 C.F.R. Part 1020 shall
not be modified such that the component or system fails to comply
with any applicable provision of this Part unless a variance in
accordance with 21 C.F.R. 1010.4 or an exemption under § 534(a)(5)
or § 538(b) of the Federal Food, Drug, and Cosmetic Act has been
granted.
2. The owner of a diagnostic
X-ray system who uses the system in a professional or commercial
capacity may modify the system provided the modification does not
result in the failure of the system or component to comply with the
applicable requirements of his Part. The owner who causes such
modification need not submit the reports required by this Subchapter,
provided the owner records the date and the details of the
modification in the system records and maintains this information,
and provided the modification of the X-ray system does not result in
a failure to comply with this Subchapter.
H. kVp Limitations. Dental
X-ray machines with a nominal fixed kVp of less than fifty (50) kVp
shall not be used to make diagnostic dental radiographs of humans.
4.4.8 Multiple Tubes
Where two (2) or more
radiographic tubes are controlled by one (1) exposure switch, the
tube or tubes which have been selected shall be clearly indicated
prior to initiation of the exposure. This indication shall be both
on the X-ray control panel and at or near the tube housing assembly
which has been selected.
4.4.9 Mechanical Support of
Tube Head
The tube housing assembly
supports shall be adjusted such that the tube housing assembly will
remain stable during an exposure unless tube housing movement is a
designed function of the X-ray system.
4.4.10 Technique Indicators
A. The technique factors to be
used during an exposure shall be indicated before the exposure
begins. If automatic exposure controls are used, the technique
factors which are set prior to the exposure shall be indicated.
B. The requirement of §
4.4.10(A) of this Part may be met by permanent markings on equipment
having fixed technique factors. Indication of technique factors
shall be visible from the operator's position except in the case of
spot films.
4.4.11 Structural Shielding
Structural shielding shall be
provided whenever necessary to meet the requirements of §§ 1.7.1
and 1.8.1 of this Subchapter, in addition to specific requirements
contained in other parts of this Subchapter.
4.4.12 Locks
All position locking,
holding, and centering devices on x-ray system components and systems
shall function as intended.
4.4.13 Use of Calibrated
Dosimetry System
The measurement of the
radiation output of an X-ray system shall be performed with a
calibrated dosimetry system. The calibration of such a system shall
be traceable to a national standard. The dosimetry system shall have
been calibrated within the preceding two (2) years.
4.4.14 Reports and
Notifications of Radiation Medical Events
A. Other than events that
result from intervention by a patient or human research subject, a
registrant shall report any event in which the administration of
ionizing radiation from a diagnostic radiation machine meets one or
more of the following criteria:
1. A patient or human research
subject receives an unintended skin dose to the same area in a single
procedure greater than two (2) Gy [two-hundred (200) rads].
2. A patient or human research
subject receives an unintended dose other than skin dose in a single
procedure greater than:
a. Five (5) times the
facility’s established protocol, and five-hundred (500) mGy [fifty
(50) rads] to any organ; or
b. Five (5) times the
facility’s established protocol, and fifty (50) mSv [five (5) rem]
total effective dose.
3. Wrong patient or wrong site
for the entire procedure when the resultant dose:
a. Exceeds five-hundred (500)
mGy [fifty (50) rads] to any organ; or
b. Total effective dose is
greater than or equal to (>) fifty (50) mSv [five (5) rem].
4. Equipment failure,
personnel error, accident, mishap or other unusual occurrence with
the administration of ionizing radiation that exceeds fifty (50) mGy
[five (5) rads] total effective dose.
B. Any wrong patient or wrong
site imaged regardless of dose received should be reported,
documented and addressed internally within the facility.
C. The registrant shall notify
the Agency by telephone no later than the next business day after
discovery of the radiation medical event.
1. All required notifications
shall use Agency contact information specified in § 1.4 of this
Subchapter.
D. The registrant shall submit
a written report, prepared by a Qualified Medical Physicist, to the
Agency within fifteen (15) business days after discovery of the
radiation medical event. The written report shall include:
1. The registrant’s name;
2. Date of event and date
discovered;
3. The total estimated dose
received;
4. The imaging procedure(s)
performed;
5. The type of equipment in
use (e.g., CT, fluoroscopy, radiographic, other);
6. The manufacturer and model
of the unit used:
7. Why the event occurred;
8. How the event was
discovered;
9. The effect, if any, on the
individuals(s) who is the subject of the radiation medical event;
10. Actions, if any, that have
been taken, or are planned, to prevent recurrence;
11. Certification that the
registrant notified the individual (or the individual’s responsible
relative or guardian), and if not, why not; and
12. If there was notification,
what information was provided to the individual.
E. The registrant shall
provide a clinical summary of the radiation medical event to the
prescribing physician and patient within fifteen (15) business days.
4.4.15 Records of Radiation
Medical Events
A registrant shall retain a
record of a radiation medical event reported in accordance with §
4.4.14 of this Part as part of the patient's permanent medical
record.
4.5 FLUOROSCOPIC EQUIPMENT
4.5.1 Applicability
The provisions of § 4.5 of
this Part apply to equipment for fluoroscopic imaging or for
recording images from the fluoroscopic image receptor, except
computed tomography X-ray systems manufactured on or after November
29, 1984.
4.5.2 Primary Protective
Barrier
A. Limitation of Useful Beam.
The fluoroscopic imaging assembly shall be provided with a primary
protective barrier which intercepts the entire cross section of the
useful beam at any SID. The X-ray tube used for fluoroscopy shall not
produce X-rays unless the barrier is in position to intercept the
entire useful beam. The air kerma (exposure) rate due to transmission
through the barrier with the attenuation block in the useful beam
combined with radiation from the fluoroscopic imaging receptor shall
not exceed 3.34x10-3 percent of the entrance air kerma (exposure)
rate, at a distance of ten (10) cm from any accessible surface of the
fluoroscopic imaging assembly beyond the plane of the image receptor.
B. Measuring Compliance. The
air kerma (exposure) rate shall be measured in accordance with § 4.6
of this Part. The air kerma (exposure) rate due to transmission
through the primary barrier combined with radiation from the
fluoroscopic image receptor shall be determined by measurements
averaged over an area of one-hundred square centimeters (100 cm2)
with no linear dimension greater than twenty (20) cm. If the source
is below the tabletop, the measurement shall be made with the input
surface of the fluoroscopic imaging assembly positioned thirty (30)
cm above the tabletop. If the source is above the tabletop and the
SID is variable, the measurement shall be made with the end of the
beam-limiting device or spacer as close to the tabletop as it can be
placed, provided that it shall not be closer than thirty (30) cm.
Movable grids and compression devices shall be removed from the
useful beam during the measurement. For all measurements, the
attenuation block shall be positioned in the useful beam ten (10) cm
from the point of measurement of entrance air kerma (exposure) rate
and between this point and the input surface of the fluoroscopic
imaging assembly.
4.5.3 Equipment Operation
A. All imaging formed by the
use of fluoroscopic x-ray systems shall be viewed, directly or
indirectly, and interpreted by a licensed practitioner of the healing
arts.
B. The operation of mobile or
portable fluoroscopic x-ray systems, for positioning purposes only,
by radiologic technologists shall be performed under the direct
supervision of a licensed practitioner of the healing arts who meets
the requirements of § 4.3.3(C) of this Part.
C. Radiologic technology
students shall not be allowed to operate fluoroscopic x-ray systems
unless in the physical presence of a licensed practitioner of the
healing arts and a radiologic technologist, as specified in §
4.3.3(C) of this Part.
D. Overhead fluoroscopy shall
not be used as a positioning tool for general purpose radiographic
examinations.
E. Each registrant that uses
fluoroscopic x-ray systems shall maintain a record of the cumulative
fluoroscopic exposure time used and the number of images recorded
from the fluoroscopic image receptor for each examination. This
record shall include patient identification, type and date of
examination, the fluoroscopic system used, and operator’s name. The
record shall be maintained for five (5) years.
4.5.4 Field Limitation
A. Angulation. For
fluoroscopic equipment manufactured after February 25, 1978, when the
angle between the image receptor and the beam axis of the X-ray beam
is variable, means shall be provided to indicate when the axis of the
X-ray beam is perpendicular to the plane of the image receptor.
Compliance with §§ 4.5.4(D) and (E) of this Part shall be
determined with the beam axis indicated to be perpendicular to the
plane of the image receptor.
B. Further Means for
Limitation. Means shall be provided to permit further limitation of
the X-ray field to sizes smaller than the limits of §§ 4.5.4(D) and
(E) of this Part. Beam-limiting devices manufactured after May
22,1979, and incorporated in equipment with a variable SID and/or
capability of a visible area of greater than three-hundred square cm
(300 cm2), shall be provided with means for stepless adjustment of
the X-ray field. Equipment with a fixed SID and the capability of a
visible area of no greater than three-hundred square cm (300 cm2)
shall be provided with either stepless adjustment of the X-ray field
or with a means to further limit the X-ray field size at the plane of
the image receptor to one-hundred twenty five square cm (125 cm2) or
less. Stepless adjustment shall, at the greatest SID, provide
continuous field sizes from the maximum obtainable to a field size
containable in a square of five (5) cm by five (5) cm. This
paragraph does not apply to non-image-intensified fluoroscopy.
C. Non-Image-Intensified
Fluoroscopy. The X-ray field produced by non-image-intensified
fluoroscopic equipment shall not extend beyond the entire visible
area of the image receptor. Means shall be provided for stepless
adjustment of field size. The minimum field size, at the greatest
SID, shall be containable in a square of five (5) cm by five (5) cm.
D. Fluoroscopy and Radiography
Using the Fluoroscopic Imaging Assembly with Inherently Circular
Image Receptors
1. For fluoroscopic equipment
manufactured before June 10, 2006, other than radiation therapy
simulation systems, the following applies:
a. Neither the length nor
width of the X-ray field in the plane of the image receptor shall
exceed that of the visible area of the image receptor by more than
three percent (3%) of the SID. The sum of the excess length and the
excess width shall be no greater than four percent (4%) of the SID.
b. For rectangular X-ray
fields used with circular image receptors, the error in alignment
shall be determined along the length and width dimensions of the
X-ray field which pass through the center of the visible area of the
image receptor.
2. For fluoroscopic equipment
manufactured on or after June 10, 2006, other than radiation therapy
simulation systems, the maximum area of the X-ray field in the plane
of the image receptor shall conform with one of the following
requirements:
a. When any linear dimension
of the visible area of the image receptor measured through the center
of the visible area is less than or equal to thirty-four (34) cm in
any direction, at least eighty percent (80%) of the area of the X-ray
field overlaps the visible area of the image receptor, or
b. When any linear dimension
of the visible area of the image receptor measured through the center
of the visible area is greater than thirty-four (34) cm in any
direction, the X-ray field measured along the direction of greatest
misalignment with the visible area of the image receptor does not
extend beyond the edge of the visible area of the image receptor by
more than two (2) cm.
E. Fluoroscopy and Radiography
Using Fluoroscopic Imaging Assembly With Inherently Rectangular Image
Receptors. For X-ray systems manufactured on or after June 10, 2006,
the following applies:
1. Neither the length nor
width of the X-ray field in the plane of the image receptor shall
exceed that of the visible area of the image receptor by more than
three percent (3%) of the SID. The sum of the excess length and the
excess width shall be no greater than four percent (4%) of the SID.
2. The error in alignment
shall be determined along the length and width dimensions of the
X-ray field which pass through the center of the visible area of the
image receptor.
F. Override Capability. If the
fluoroscopic X-ray field size is adjusted automatically as the SID or
image receptor size is changed, a capability may be provided for
overriding the automatic adjustment in case of system failure. If it
is so provided, a signal visible at the operator’s position shall
indicate whenever the automatic field adjustment is overridden. Each
such system failure override switch shall be clearly labeled as
follows: FOR X-RAY FIELD LIMITATION SYSTEM FAILURE
4.5.5 Activation of the
Tube
X-ray production in the
fluoroscopic mode shall be controlled by a device which requires
continuous pressure by the operator for the entire time of any
exposure. When recording serial fluoroscopic images from the
fluoroscopic image receptor, the operator shall be able to terminate
the X-ray exposure(s) at any time, but means may be provided to
permit completion of any single exposure of the series in process.
4.5.6 Air Kerma (Exposure)
Rates.
A. For fluoroscopic equipment,
the following requirements apply:
1. Fluoroscopic equipment
manufactured before May 19, 1995.
a. Equipment provided with
automatic exposure rate control (AERC) shall not be operable at any
combination of tube potential and current that will result in an air
kerma (exposure) rate in excess of 88 mGy per minute (10 R/min
exposure rate) at the measurement point specified in 21 C.F.R.
1020.32(d)(3), except as specified in § 4.5.6(A)(1)(e) of this Part.
b. Equipment provided without
AERC shall not be operable at any combination of tube potential and
current that will result in an air kerma (exposure) rate in excess of
44 mGy per minute (5 R/min exposure rate) at the measurement point
specified in 21 C.F.R. 1020.32(d)(3), except as specified in §
4.5.6(A)(1)(e) of this Part.
c. Equipment provided with
both an AERC mode and a manual mode shall not be operable at any
combination of tube potential and current that will result in an air
kerma (exposure) rate in excess of 88 mGy per minute (10 R/min
exposure rate) in either mode at the measurement point specified in
21 C.F.R. 1020.32(d)(3), except as specified in § 4.5.6(A)(1)(e) of
this Part.
d. Equipment may be modified
in accordance with § 4.4.7(E)(1) of this Part to comply with §
4.5.6(A)(2) of this Part. When the equipment is modified, it shall
bear a label indicating the date of the modification and the
statement: MODIFIED TO COMPLY WITH 21 CFR 1020.32(H)(2)
e. Exceptions:
(1) During recording of
fluoroscopic images, or
(2) When a mode of operation
has an optional high-level control, in which case that mode shall not
be operable at any combination of tube potential and current that
will result in an air kerma (exposure) rate in excess of the rates
specified in §§ 4.5.6(A)(1)(a) through (c) of this Part at the
measurement point specified in 21 C.F.R. 1020.32(d)(3), unless the
high-level control is activated. Special means of activation of
high-level controls shall be required. The high-level control shall
be operable only when continuous manual activation is provided by the
operator. A continuous signal audible to the operator shall indicate
that the high-level control is being employed.
2. Fluoroscopic equipment
manufactured on or after May 19, 1995.
a. Shall be equipped with AERC
if operable at any combination of tube potential and current that
results in an air kerma (exposure) rate greater than 44 mGy per
minute (5 R/min exposure rate) at the measurement point specified in
21 C.F.R. 1020.32(d)(3). Provision for manual selection of technique
factors may be provided.
b. Shall not be operable at
any combination of tube potential and current that will result in an
air kerma (exposure) rate in excess of 88 mGy per minute (10 R/min
exposure rate) at the measurement point specified in 21 C.F.R.
1020.32(d)(3), except as specified in § 4.5.6(A)(2)(c) of this Part.
c. Exceptions
(1) For equipment manufactured
prior to June 10, 2006, during the recording of images from a
fluoroscopic image receptor using photographic film or a video camera
when the X-ray source is operated in a pulsed mode.
(2) For equipment manufactured
on or after June 10, 2006, during the recording of images from the
fluoroscopic image receptor for the purpose of providing the user
with a recorded image(s) after termination of the exposure. Such
recording does not include images resulting from a last-image-hold
feature that are not recorded.
(3) When a mode of operation
has an optional high-level control and the control is activated, in
which case the equipment shall not be operable at any combination of
tube potential and current that will result in an air kerma
(exposure) rate in excess of 176 mGy per minute (20 R/min exposure
rate) at the measurement point specified in 21 C.F.R. 1020.32(d)(3).
Special means of activation of high-level controls shall be required.
The high-level control shall be operable only when continuous manual
activation is provided by the operator. A continuous signal audible
to the operator shall indicate that the high-level control is
employed.
4.5.7 Measurement of
Entrance Air Kerma (Exposure) Rate.
A. Measurement of entrance air
kerma (exposure) rate shall be performed for both maximum and typical
values and shall be made at intervals not to exceed twelve (12)
months or after any maintenance of the system which might affect the
air kerma (exposure) rate. Results of these measurements shall be
posted where any fluoroscopist may have ready access to such results
during the fluoroscopic procedure and in the record required in §
4.3.13(E) of this Part. Results of the measurements shall include the
mGy per minute (R/min exposure rate), as well as the technique
factors used to determine such results. The name of the Qualified
Medical Physicist performing the measurements and the date the
measurements were performed shall be included in the results.
B. Conditions of measurement
of maximum entrance air kerma (exposure) rate are as follows:
1. The measurements shall be
made under conditions that satisfy the requirements of § 4.5.6(A) of
this Part;
2. The kVp, mA, and/or other
selectable parameters shall be adjusted to those settings which give
the maximum air kerma (exposure) rate; and
3. An X-ray system that
incorporates automatic exposure rate control (AERC) shall have
sufficient material placed in the useful beam to produce the maximum
output of that system.
C. Conditions of measurement
of typical air kerma (exposure) rate are as follows:
1. The measurements shall be
made under conditions that satisfy the requirements of § 4.5.7(D) of
this Part and are typical of clinical use of the X-ray system;
2. The kVp shall be that
typical of clinical use of the X-ray system;
3. An X-ray system(s) that
incorporates AERC shall have sufficient material placed in the useful
beam to produce operating parameters typical of the use of the X-ray
system; and
4. An X-ray system(s) that
does not incorporate an AERC shall utilize a milliamperage typical of
the clinical use of the X-ray system.
a. Material should be placed
in the useful beam when conducting these periodic measurements to
protect the imaging system.
D. Measuring Compliance.
Compliance with this subsection shall be determined as follows:
1. If the source is below the
X-ray table, the air kerma (exposure) rate shall be measured at one
(1) cm above the tabletop or cradle.
2. If the source is above the
X-ray table, the air kerma (exposure) rate shall be measured at
thirty (30) cm above the tabletop with the end of the beam-limiting
device or spacer positioned as closely as possible to the point of
measurement.
3. In a C-arm type of
fluoroscope, the air kerma (exposure) rate shall be measured at
thirty (30) cm from the input surface of the fluoroscopic imaging
assembly, with the source positioned at any available SID, provided
that the end of the beam-limiting device or spacer is no closer than
thirty (30) cm from the input surface of the fluoroscopic imaging
assembly.
4. In a C-arm type of
fluoroscope having an SID less than forty-five (45) cm, the air kerma
(exposure) rate shall be measured at the minimum SSD.
5. In a lateral type of
fluoroscope, the air kerma (exposure) rate shall be measured at a
point fifteen (15) cm from the centerline of the X-ray table and in
the direction of the X-ray source with the end of the beam-limiting
device or spacer positioned as closely as possible to the point of
measurement. If the tabletop is movable, it shall be positioned as
closely as possible to the lateral X-ray source, with the end of the
beam-limiting device or spacer no closer than fifteen (15) cm to the
centerline of the X-ray table.
4.5.8 Indication of
Potential and Current
During fluoroscopy and
cinefluorography, the X-ray tube potential and current shall be
continuously indicated. Deviation of X-ray tube potential and
current from the indicated value shall not exceed the maximum
deviation as stated by the manufacturer.
4.5.9 Source-Skin Distance
A. Means shall be provided to
limit the source-skin distance to not less than thirty-eight (38) cm
on stationary fluoroscopes and to not less than thirty (30) cm on
mobile and portable fluoroscopes. In addition, for fluoroscopes
intended for specific surgical application that would be prohibited
at the source-skin distances specified in this paragraph, provisions
may be made for operating at shorter source-skin distances but in no
case less than twenty (20) cm.
B. For stationary, mobile, or
portable C-arm fluoroscopic systems manufactured on or after June 10,
2006, having a maximum source-image receptor distance of less than
forty-five (45) cm, means shall be provided to limit the source-skin
distance to not less than nineteen (19) cm. Such systems shall be
labeled for extremity use only. In addition, for those systems
intended for specific surgical application that would be prohibited
at the source-skin distance specified in this paragraph, provisions
may be made for operation at shorter source-skin distances but in no
case less than ten (10) cm.
4.5.10 Fluoroscopic
Irradiation Time, Display and Signal
A. Fluoroscopic equipment
manufactured before June 10, 2006:
1. Shall be provided with
means to preset the cumulative irradiation time of the fluoroscopic
tube. The maximum cumulative time of the timing device shall not
exceed five (5) minutes without resetting. A signal audible to the
operator shall indicate the completion of any preset cumulative
irradiation time. Such signal shall continue to sound while X-rays
are produced until the timing device is reset. Fluoroscopic equipment
may be modified in accordance with 21 C.F.R. § 1020.30(q) to comply
with the requirements of § 4.5.10 of this Part. When the equipment
is modified, it shall bear a label indicating the statement: MODIFIED
TO COMPLY WITH 21 C.F.R. § 1020.32(H)(2)
B. For X-ray controls
manufactured on or after June 10, 2006, there shall be provided for
each fluoroscopic tube:
1. A display of the
fluoroscopic irradiation time at the operator’s working position.
This display shall function independently of the audible signal
described § 4.5.10(B)(2) of this Part. The following requirements
apply:
a. When the X-ray tube is
activated, the fluoroscopic irradiation time in minutes and tenths of
minutes shall be continuously displayed and updated at least once
every six (6) seconds.
b. The fluoroscopic
irradiation time shall also be displayed within six (6) seconds of
termination of an exposure and remain displayed until reset.
c. Means shall be provided to
reset the display to zero prior to the beginning of a new examination
or procedure.
2. A signal audible to the
operator shall sound for each passage of five (5) minutes of
fluoroscopic irradiation time during an examination or procedure.
The signal shall sound until manually reset or, if automatically
reset, for at least two (2) seconds.
4.5.11 Mobile and Portable
Fluoroscopes.
In addition to the other
requirements of § 4.5 of this Part, mobile and portable fluoroscopes
shall provide an image receptor incorporating more than a simple
fluorescent screen.
4.5.12 Control of Scattered
Radiation
A. Fluoroscopic table designs
when combined with procedures utilized shall be such that no
unprotected part of any staff or ancillary individual's body shall be
exposed to unattenuated scattered radiation which originates from
under the table. The attenuation required shall be not less than
0.25 millimeter lead equivalent.
B. Equipment configuration
when combined with procedures shall be such that no portion of any
staff or ancillary individual's body, except the extremities, shall
be exposed to the unattenuated scattered radiation emanating from
above the tabletop unless that individual:
1. Is at least one-hundred
twenty (120) centimeters from the center of the useful beam, or
2. The radiation has passed
through not less than 0.25 millimeter lead equivalent material (e.g.,
drapes, Bucky-slot cover-sliding or folding panel, or self supporting
curtains) in addition to any lead equivalency provided by the
protective apron referred to in § 4.3.6 of this Part.
C. The Agency may grant
exemptions to § 4.5.12(B) of this Part where a sterile field will
not permit the use of the normal protective barriers. Where the use
of prefitted sterilized covers for the barriers is practical, the
Agency shall not permit such exception.
4.5.13 Patient Dose
Evaluation
A. Each registrant performing
fluoroscopically-guided interventional procedures shall develop
written policies and procedures to:
1. Identify those procedures
which have a potential to result in patient doses exceeding the
threshold for injury;
2. Reduce the probability of
such exposures; and
3. Ensure that appropriate
action occurs for patients receiving doses that warrant follow-up.
B. The registrant shall have a
patient dose monitoring procedures in place and shall document (in
the patient’s medical record) an estimate of the absorbed dose to
the skin. When the fluoroscopy unit is equipped with an air kerma
dose readout, the recording of this value shall suffice as a patient
dose record.
C. The registrant shall
conduct patient dose evaluation for any procedure that has a
reasonable probability of resulting in a deterministic injury (i.e.,
a cumulative absorbed dose to the skin equal to or greater than 1 Gy
(100 rads)). This evaluation shall be noted in the patient’s
medical record and reviewed by the Radiation Safety Committee. If the
registrant does not have a Radiation Safety Committee, the review
shall be conducted by the Radiation Safety Officer and the
registrant’s medical physicist.
4.5.14 Radiation Therapy
Simulation Systems
A. Radiation therapy
simulation systems shall be exempt from the requirements of §
4.5.2(A), provided such systems are intended only for remote control
operation.
B. Radiation therapy
simulation systems shall be exempt from all the requirements of §§
4.5.4(D), 4.5.6, and 4.6.12(B)(2) of this Part when used for therapy
simulation purposes.
C. As an alternative to the
requirements of § 4.5.10 of this Part, radiation therapy simulation
systems may be provided with a means to indicate the total cumulative
exposure time during which X-rays were produced, and which is capable
of being reset between X-ray examinations.
4.5.15 Display of
Last-Image-Hold (LIH)
A. Fluoroscopic equipment
manufactured on or after June 10, 2006, shall be equipped with means
to display LIH image following termination of the fluoroscopic
exposure.
B. For an LIH image obtained
by retaining pretermination fluoroscopic images, if the number of
images and method of combining images are selectable by the user, the
selection shall be indicated prior to initiation of the fluoroscopic
exposure.
C. For an LIH image obtained
by initiating a separate radiographic-like exposure at the
termination of fluoroscopic imaging, the technique factors for the
LIH image shall be selectable prior to the fluoroscopic exposure, and
the combination selected shall be indicated prior to initiation of
the fluoroscopic exposure.
D. Means shall be provided to
clearly indicate to the user whether a displayed image is the LIH
radiograph or fluoroscopy. Display of the LIH radiograph shall be
replaced by the fluoroscopic image concurrently with re-initiation of
fluoroscopic exposure, unless separate displays are provided for the
LIH radiograph and fluoroscopic images.
4.5.16 Displays of Values
of Air Kerma (Exposure) Rate and Cumulative Air Kerma.
A. Fluoroscopic equipment
manufactured on or after June 10, 2006, shall display at the
operator’s working position the air kerma (exposure) rate and
cumulative air kerma. The following requirements apply for each X-ray
tube used during an examination or procedure:
B. When the X-ray tube is
activated and the number of images produced per unit time is greater
than six (6) images per second, the air kerma (exposure) rate in
mGy/min shall be continuously displayed and updated at least once
every second.
C. The cumulative air kerma in
units of mGy shall be displayed either within five (5) seconds of
termination of an exposure or displayed continuously and updated at
least once every five (5) seconds.
D. The display of the air
kerma (exposure) rate shall be clearly distinguishable from the
display of the cumulative air kerma.
E. The air kerma (exposure)
rate and cumulative air kerma shall represent the value for
conditions of free-in-air irradiation at one of the following
reference locations specified according to the type of fluoroscope.
1. For fluoroscopes with X-ray
source below the X-ray table, X-ray source above the table, or of
lateral type, the reference location shall be the respective
locations specified in §§ 4.5.7(D)(1), (2) or (5) of this Part.
2. For C-arm fluoroscopes, the
reference location shall be fifteen (15) cm from the isocenter toward
the X-ray source along the beam axis. Alternatively, the reference
location shall be at a point specified by the manufacturer to
represent the location of the intersection of the X-ray beam with the
patient’s skin.
F. Means shall be provided to
reset to zero the display of cumulative air kerma prior to the
commencement of a new examination or procedure.
G. The displayed air kerma
(exposure) rate and cumulative air kerma shall not deviate from the
actual values by more than ± thirty-five percent (± 35%) over the
range of 6 mGy/min and 100 mGy to the maximum indication of air kerma
(exposure) rate and cumulative air kerma, respectively. Compliance
shall be determined with an irradiation time greater than three (3)
seconds.
4.6 RADIOGRAPHIC EQUIPMENT
4.6.1 Beam Limitation,
Except Mammographic Systems.
The useful beam shall be
limited to the area of clinical interest. This shall be deemed to
have been met if a positive beam limiting device meeting
manufacturer's specifications and the requirements of § 4.4.2 of
this Part has been properly used or if evidence of collimation is
shown on at least three sides or three corners of the film (for
example, projections from the shutters of the collimator, cone
cutting at the corners, or borders at the film's edge).
4.6.2 Radiation Exposure
Control
A. Exposure Initiation. Means
shall be provided to initiate the radiation exposure by a deliberate
action on the part of the operator, such as the depression of a
switch. Radiation exposure shall not be initiated without such an
action. In addition, it shall not be possible to initiate an exposure
when the timer is set to a "zero" or "off"
position if either position is provided.
B. Exposure Indication. Means
shall be provided for visual indication observable at or from the
operator's protected position whenever x-rays are produced. In
addition, a signal audible to the operator shall indicate that the
exposure has terminated.
C. Operator Protection, Except
Veterinary Systems.
1. Stationary Systems.
Stationary X-ray systems shall be required to have the X-ray control
permanently mounted in a protected area so that the operator is
required to remain in that protected area during the entire exposure.
2. Mobile and Portable
Systems. Mobile and portable X-ray systems which are:
a. Used continuously for
greater than one (1) week in the same location (i.e., a room or
suite) shall meet the requirements of § 4.6.2(C)(1) of this Part;
b. Used for less than one (1)
week at the same location shall be provided with either a protective
barrier at least two (2) meters (6.5 feet) high for operator
protection during exposures, or means shall be provided to allow the
operator to be at least 2.7 meters (9 feet) from the tube housing
assembly during the exposure.
D. Operator Protection for
Veterinary Systems.
1. All stationary, mobile or
portable X-ray systems used for veterinary work shall be provided
with either a two (2) meter (6.5 feet) high protective barrier for
operator protection during exposures, or shall be provided with means
to allow the operator to be at least 2.7 meters (9 feet) from the
tube housing assembly during exposures. No individual other than the
operator shall be in the X-ray room while exposures are being made
unless such individual's assistance is required. Refer to § 4.13 of
this Part for hand-held intraoral dental radiographic units used in
veterinary practice.
2. When an animal must be held
in position during radiography, mechanical supporting or restraining
devices should be used. If necessary, general anesthesia, sedation or
tranquilization should be used. If the animal must be held by an
individual, that individual shall be protected with appropriate
shielding devices, such as protective gloves and apron, and shall be
so positioned that no part of their body will be struck by the useful
beam. No individual shall be used routinely to hold animals or film
during radiation exposures. The exposure of any individual used for
this purpose shall be monitored, and a record shall be made of the
examination, including the name of the human holder, date of the
examination, number of exposures and technique factors utilized for
the exposure(s).
4.6.3 Control and
Indication of Technique Factors
A. Visual Indication. The
technique factors to be used during an exposure shall be indicated
before the exposure begins, except when automatic exposure controls
are used, in which case the technique factors which are set prior to
the exposure shall be indicated. On equipment having fixed technique
factors, this requirement may be met by permanent markings.
Indication of technique factors shall be visible from the operator’s
position except in the case of spot films made by the operator.
B. Timers. Means shall be
provided to terminate the exposure at a preset time interval, a
preset product of current and time, a preset number of pulses, or a
preset radiation exposure to the image receptor.
1. Except during serial
radiography, the operator shall be able to terminate the exposure at
any time during an exposure of greater than one-half (0.5) second.
Except during panoramic dental radiography, termination of exposure
shall cause automatic resetting of the timer to its initial setting
or to zero. It shall not be possible to make an exposure when the
timer is set to a zero or off position if either position is
provided.
2. During serial radiography,
the operator shall be able to terminate the X-ray exposure(s) at any
time, but means may be provided to permit completion of any single
exposure of the series in process.
C. Automatic Exposure
Controls. When an automatic exposure control is provided:
1. Indication shall be made on
the control panel when this mode of operation is selected;
2. When the X-ray tube
potential is equal to or greater than fifty-one (51) kilovolts peak
(kVp), the minimum exposure time for field emission equipment rated
for pulse operation shall be equal to or less than a time interval
equivalent to two pulses and the minimum exposure time for all other
equipment shall be equal to or less than 1/60 second or a time
interval required to deliver five (5) milliampere-seconds (mAs),
whichever is greater;
3. Either the product of peak
X-ray tube potential, current, and exposure time shall be limited to
not more than sixty (60) kilowatt-seconds (kWs) per exposure or the
product of X-ray tube current and exposure time shall be limited to
not more than six-hundred (600) mAs per exposure, except when the
X-ray tube potential is less than fifty-one (51) kVp, in which case
the product of X-ray tube current and exposure time shall be limited
to not more than two-thousand (2,000) mAs per exposure; and
4. A visible signal shall
indicate when an exposure has been terminated at the limits described
in § 4.6.3(A)(3) of this Part, and manual resetting shall be
required before further automatically timed exposures can be made.
D. Accuracy. Deviation of
technique factors from indicated values shall not exceed the limits
given by the manufacturer.
4.6.4 Positive Beam
Limitation (PBL)
A. The requirements of §
4.6.4 of this Part shall apply to radiographic systems which contain
PBL.
B. Field Size. When a PBL
system is provided, it shall prevent X-ray production when:
1. Either the length or width
of the X-ray field in the plane of the image receptor differs from
the corresponding image receptor dimension by more than three percent
(3%) of the SID; or
2. The sum of the length and
width differences stated in § 4.6.4(B)(1) of this Part without
regard to sign exceeds four percent (4%) of the SID.
3. The beam-limiting device is
at an SID for which PBL is not designed for sizing.
C. Conditions For PBL. When
provided, the PBL system shall function as described in § 4.6.4(B)
of this Part whenever all the following conditions are met:
1. The image receptor is
inserted into a permanently mounted cassette holder;
2. The image receptor length
and width are less than fifty (50) cm;
3. The X-ray beam axis is
within ± three degrees (±3°) of vertical and the SID is ninety
(90) cm to one-hundred thirty (130) cm inclusive; or the X-ray beam
axis is within ± three degrees (±3°) of horizontal and the SID is
ninety (90) cm to two-hundred five (205) cm inclusive;
4. The X-ray beam axis is
perpendicular to the plane of the image receptor to within ± three
degrees (±3°); and
5. Neither tomographic nor
stereoscopic radiography is being performed.
D. Measuring Compliance.
Compliance with the requirements of § 4.6.4(B) of this Part shall be
determined when the equipment indicates that the beam axis is
perpendicular to the plane of the image receptor and the provisions
of § 4.6.4(B) of this Part are met. Compliance shall be determined
no sooner than five (5) seconds after insertion of the image
receptor.
E. Operator Initiated
Undersizing. The PBL system shall be capable of operating such that,
at the discretion of the operator, the size of the field may be made
smaller than the size of the image receptor through stepless
adjustment of the field size. Each dimension of the minimum field
size at an SID of one-hundred (100) cm shall be equal to or less than
five (5) cm. Return to PBL function as described in § 4.6.4(B) of
this Part shall occur automatically upon any change of image receptor
size or SID.
F. Override of PBL. A
capability may be provided for overriding PBL in case of system
failure and for servicing the system. This override may be for all
SIDs and image receptor sizes. A key shall be required for any
override capability that is accessible to the operator. It shall not
be possible to remove the key while PBL is overridden. Each such key
switch or key shall be clearly and durably labeled as follows: FOR
X-RAY FIELD LIMITATION SYSTEM FAILURE
1. The override capability is
considered accessible to the operator if it is referenced in the
operator’s manual or in other material intended for the operator or
if its location is such that the operator would consider it part of
the operational controls.
4.6.5 Source-to-Skin
Distance
A. X-ray systems designed for
use with an intraoral image receptor shall be provided with means to
limit the source-skin distance to not less than:
1. Eighteen (18) cm if
operable above fifty (50) kVp; or
2. Ten (10) cm if not operable
above fifty (50) kVp.
B. Mobile and portable X-ray
systems other than dental shall be provided with means to limit the
source-skin distance to not less than thirty (30) cm.
4.6.6 Air Kerma (Exposure)
Reproducibility.
A. The following requirements
shall apply when the equipment is operated on an adequate power
supply as specified by the manufacturer:
1. For any specific
combination of selected technique factors, the coefficient of
variation of the air kerma (exposure) shall not exceed 0.10 when all
technique factors are held constant. This requirement shall be
deemed to have been met if, when four exposures are made at identical
technique factors, the value of the average exposure (E) is greater
than or equal to 5 times the maximum exposure (E max ) minus
the minimum exposure (E min ); i.e., E > 5 (E max
- E min ).
2. For equipment having
automatic exposure controls, compliance shall be determined with a
sufficient thickness of attenuating material in the useful beam such
that the technique factors can be adjusted to provide individual
exposures of a minimum of twelve (12) pulses on field emission
equipment rated for pulsed operation or no less than one-tenth (0.1)
second per exposure on all other equipment.
4.6.7 Radiation from
Capacitor Energy Storage Equipment.
A. Radiation emitted from the
X-ray tube shall not exceed:
1. An air kerma of 0.26 µGy
(0.03 mR exposure) in one (1) minute at five (5) cm from any
accessible surface of the diagnostic source assembly, with the
beam-limiting device fully open, the system fully charged, and the
exposure switch, timer, or any discharge mechanism not activated.
Compliance shall be determined by measurements averaged over an area
of one-hundred square cm (100 cm2), with no linear dimensions greater
than twenty (20) cm: and
2. An air kerma of 0.88 mGy
(100 mR exposure) in one (1) hour at one-hundred (100) cm from the
X-ray source, with beam-limiting device fully open, when the system
is discharged through the X-ray tube either manually or automatically
by use of a discharge switch or deactivation of the input power.
Compliance shall be determined by measurements of the maximum air
kerma per discharge multiplied by the total projected number of
discharges in one (1) hour. The measurements shall be averaged over
an area of one-hundred square cm (100 cm2) with no linear dimension
greater than twenty (20) cm.
4.6.8 Tube Stands for
Portable X-Ray Systems.
A tube stand or other
mechanical support shall be used for portable X-ray systems, so that
the X-ray tube housing assembly need not be hand-held during
exposures.
4.6.9 Measurement of
Radiation Output.
A. Measurement of the
radiation output shall be performed at a specified distance and over
a range of clinical kVp values, and shall be made at intervals not to
exceed twelve (12) months or after any maintenance of the system
which might affect the radiation output. These measurements shall be
performed in-air with minimum scatter conditions. Results of the
measurements shall include the µGy/mAs (mR/mAs), as well as the
technique factors used to determine such results.
B. The name and signature of
the Qualified Medical Physicist performing the measurements, and the
date the measurements were performed, shall be included in the
results.
C. These measurements may be
used to estimate entrance skin exposure (ESE) for the average adult
patient for selected routine radiographic procedures. These values
should be compared with available national reference values.
4.6.10 Beam-on Indicators.
The X-ray control shall
provide visual indication whenever X-rays are produced. In addition,
a signal audible to the operator shall indicate that the exposure has
terminated.
4.6.11 Primary Protective
Barrier for Mammography X-ray Systems
A. For X-ray systems
manufactured after September 5,1978, and before September 30,1999,
which are designed only for mammography, the transmission of the
primary beam through any image receptor support provided with the
system shall be limited such that the air kerma five (5) cm from any
accessible surface beyond the plane of the image receptor supporting
device does not exceed 0.88 µGy (0.1 mR exposure) for each
activation of the tube.
B. For mammographic X-ray
systems manufactured on or after September 30, 1999:
1. At any SID where exposures
can be made, the image receptor support device shall provide a
primary protective barrier that intercepts the cross section of the
useful beam along every direction except at the chest wall edge.
2. The X-ray system shall not
permit exposure unless the appropriate barrier is in place to
intercept the useful beam as required in § 4.6.11(B)(1) of this
Part.
3. The transmission of the
useful beam through the primary protective barrier shall be limited
such that the air kerma five (5) cm from any accessible surface
beyond the plane of the primary protective barrier does not exceed
0.88 µGy (0.1 mR exposure) for each activation of the tube.
C. Compliance with the
requirements of §§ 4.6.11(A) and (B)(3) of this Part for
transmission shall be determined with the X-ray system operated at
the minimum SID for which it is designed, at maximum rated peak tube
potential, at the maximum rated product of X-ray tube current and
exposure time (mAs) for the maximum rated peak tube potential, and by
measurements averaged over an area of one-hundred square cm (100 cm2)
with no linear dimension greater than twenty (20) cm. The sensitive
volume of the radiation measuring instrument shall not be positioned
beyond the edge of the primary protective barrier along the chest
wall side.
4.6.12 Field Limitation and
Alignment for Mobile, Portable and Stationary General Purpose X-ray
Systems.
A. Except when spot-film
devices are in service, mobile, portable and stationary general
purpose radiographic X-ray systems shall meet the following
requirements:
1. Variable X-ray Field
Limitation. A means for stepless adjustment of the size of the X-ray
field shall be provided. Each dimension of the minimum field size at
an SID of one-hundred (100) cm shall be equal to or less than five
(5) cm.
2. Visual Definition.
a. Means for visually defining
the perimeter of the X-ray field shall be provided. The total
misalignment of the edges of the visually defined field with the
respective edges of the X-ray field along either the length or width
of the visually defined field shall not exceed two percent (2%) of
the distance from the source to the center of the visually defined
field when the surface upon which it appears is perpendicular to the
axis of the X-ray beam.
b. When a light localizer is
used to define the X-ray field, it shall provide an average
illuminance of not less than 160 lux (15 footcandles) at one-hundred
(100) cm or at the maximum SID, whichever is less. The average
illuminance shall be based on measurements made in the approximate
center of each quadrant of the light field.
c. The edge of the light field
at one-hundred (100) cm or at the maximum SID, whichever is less,
shall have a contrast ratio, corrected for ambient lighting, of not
less than four (4) in the case of beam-limiting devices designed for
use on stationary equipment, and a contrast ratio of not less than
three (3) in the case of beam-limiting devices designed for use on
mobile and portable equipment. The contrast ratio is defined as
I 1 /I 2 , where I 1 is the illuminance
three (3) mm from the edge of the light field toward the center of
the field; and I 2 is the illuminance three (3) mm from the
edge of the light field away from the center of the field. Compliance
shall be determined with a measuring aperture of one (1) mm.
3. Portable X-Ray Systems
a. Portable X-ray systems
shall have an evaluation of light field vs. X-ray field alignment
performed at least every six (6) months to determine compliance with
both § 4.6.12(A)(2)(c) and § 4.6.13(A)(3) of this Part.
b. Portable X-ray systems
shall have an evaluation of centering alignment performed at least
every six (6) months to determine compliance with § 4.6.13(A) of
this Part.
4.6.13 Field Indication and
Alignment on Stationary General Purpose X-ray Equipment.
A. Except when spot-film
devices are in service, stationary general purpose X-ray systems
shall meet the following requirements in addition to those prescribed
in § 4.6.12 of this Part:
1. Means shall be provided to
indicate when the axis of the X-ray beam is perpendicular to the
plane of the image receptor, to align the center of the X-ray field
with respect to the center of the image receptor to within two
percent (2%) of the SID, and to indicate the SID to within two
percent (2%);
2. The beam-limiting device
shall numerically indicate the field size in the plane of the image
receptor to which it is adjusted;
3. Indication of field size
dimensions and SIDs shall be specified in centimeters and/or inches
and shall be such that aperture adjustments result in X-ray field
dimensions in the plane of the image receptor which correspond to
those indicated by the beam-limiting device to within two percent
(2%) of the SID when the beam axis is indicated to be perpendicular
to the plane of the image receptor; and
4. Compliance measurements
will be made at discrete SIDs and image receptor dimensions in common
clinical use (such as SIDs of 100, 150, and 200 cm and/or 36, 40, 48,
72 inches and nominal image receptor dimensions of 13, 18, 24, 30,
35, 40, and 43 cm and/or 5, 7, 8, 9, 10, 11, 12, 14, and 17 inches)
or at any other specific dimensions at which the beam-limiting device
or its associated diagnostic X-ray system is uniquely designed to
operate.
4.6.14 Linearity.
A. The following requirements
apply when the equipment is operated on a power supply as specified
by the manufacturer in accordance with 21 CFR Part 1020 for any fixed
X-ray tube potential within the range of forty percent (40%) to
one-hundred percent (100%) of the maximum rated:
1. Equipment Having
Independent Selection of X-Ray Tube Current (mA). The average ratios
(Xi) of air kerma (exposure) to the indicated milliampere-seconds
product (mGy/mAs or mR/mAs) obtained at any two (2) consecutive tube
current settings shall not differ by more than 0.10 times their sum:
X1-X2 ? 0.10 (X1+X2), where X1
and X2 are the average mGy/mAs (mR/mAs) values obtained at each of
two (2) consecutive tube current settings, or at two (2) settings
differing by no more than a factor of two (2) where the mA selector
provides continuous selection.
2. Equipment Having Selection
of X-Ray Tube Current-Exposure Time Product (mAs). For equipment
manufactured after 3 May 1994, the average ratios of air kerma
(exposure) to the indicated milliampere-seconds product (mGy/mAs or
mR/mAs ) obtained at any two (2) consecutive mAs selector settings
shall not differ by more than 0.10 times their sum: X1-X2 ?
0.10 (X1 +X2), where X1 and X2 are the average mGy/mAs values
obtained at any two (2) consecutive mAs selector settings, or at two
(2) settings differing by no more than a factor of two (2) where the
mAs selector provides continuous selection.
3. Measuring Compliance.
Determination of compliance will be based on consecutive exposures,
made within one (1) hour. These settings may include any two (2)
focal spot sizes except where one is equal to or less than 0.45 mm
and the other is greater than 0.45 mm. For purposes of this
requirement, focal spot size is the focal spot size specified by the
X-ray tube manufacturer.
4.6.15 Field Limitation on
Radiographic X-ray Equipment Other Than General Purpose Radiographic
Systems
A. Equipment for Use With
Intraoral Image Receptors. Radiographic equipment designed for use
with an intraoral image receptor shall be provided with means to
limit the X-ray beam such that:
1. If the minimum
source-to-skin distance (SSD) is eighteen (18) cm or more, the X-ray
field at the minimum SSD shall be containable in a circle having a
diameter of no more than seven (7) cm; and
2. If the minimum SSD is less
than eighteen (18) cm, the X-ray field at the minimum SSD shall be
containable in a circle having a diameter of no more than six (6) cm.
B. X-ray Systems Designed for
One Image Receptor Size. Radiographic equipment designed for only one
image receptor size at a fixed SID shall be provided with means to
limit the field at the plane of the image receptor to dimensions no
greater than those of the image receptor, and to align the center of
the X-ray field with the center of image receptor to within two
percent (2%) of the SID, or shall be provided with means to both size
and align the X-ray field such that the X-ray field at the plane of
the image receptor does not extend beyond the edge of the image
receptor.
C. Systems Designed for
Mammography.
1. Radiographic systems
designed only for mammography and general purpose radiography
systems, when special attachments for mammography are in service,
manufactured on or after November 1, 1977, and before September 30,
1999, shall be provided with means to limit the useful beam such that
the X-ray field at the plane of the image receptor does not extend
beyond any edge of the image receptor at any designated SID except
the edge of the image receptor designed to be adjacent to the chest
wall where the X-ray field may not extend beyond this edge by more
than two percent (2%) of the SID. This requirement can be met with a
system that performs as prescribed in §§ 4.6.15(D)(1), (2), and (3)
of this Part. When the beam-limiting device and image receptor
support device are designed to be used to immobilize the breast
during a mammographic procedure and the SID may vary, the SID
indication specified in §§ 4.6.15(D)(2) and (3) of this Part shall
be the maximum SID for which the beam-limiting device or aperture is
designed.
2. Mammographic beam-limiting
devices manufactured on or after September 30,1999, shall be provided
with a means to limit the useful beam such that the X-ray field at
the plane of the image receptor does not extend beyond any edge of
the image receptor by more than two percent (2%) of the SID. This
requirement can be met with a system that performs as prescribed in
§§ 4.6.15(D)(1), (2), and (3) of this Part. For systems that allow
changes in SID, the SID indication specified in §§ 4.6.15(D)(2) and
(3) of this Part shall be the maximum SID for which the beam-limiting
device or aperture is designed.
3. Each image receptor support
device manufactured on or after November 1, 1977, intended for
installation on a system designed for mammography shall have clear
and permanent markings to indicate the maximum image receptor size
for which it is designed.
D. Other X-ray Systems.
Radiographic systems not specifically covered in §§ 4.6.12, 4.6.13,
4.6.15(B), 4.6.15(C) of this Part, and systems covered in §
4.6.15(A) of this Part, which are also designed for use with
extraoral image receptors and when used with an extraoral image
receptor, shall be provided with means to limit the X-ray field in
the plane of the image receptor so that such field does not exceed
each dimension of the image receptor by more than two percent (2%) of
the SID, when the axis of the X-ray beam is perpendicular to the
plane of the image receptor. In addition, means shall be provided to
align the center of the X-ray field with the center of the image
receptor to within two percent (2%) of the SID, or means shall be
provided to both size and alignment the X-ray field such that the
X-ray field at the plane of the image receptor does not extend beyond
any edge of the image receptor. These requirements may be met with:
1. A system which performs in
accordance with §§ 4.6.12 and 4.6.13 of this Part; or when
alignment means are also provided, may be met with either;
2. An assortment of removable,
fixed-aperture, beam-limiting devices sufficient to meet the
requirement for each combination of image receptor size and SID for
which the unit is designed. Each such device shall have clear and
permanent markings to indicate the image receptor size and SID for
which it is designed; or
3. A beam-limiting device
having multiple fixed apertures sufficient to meet the requirement
for each combination of image receptor size and SID for which the
unit is designed. Permanent, clearly legible markings shall indicate
the image receptor size and SID for which each aperture is designed
and shall indicate which aperture is in position for use.
4.6.16 Field Limitation and
Alignment for Spot-Film Devices.
A. The following requirements
shall apply to spot-film devices, except when the spot-film device is
provided for use with a radiation therapy simulation system:
1. Means shall be provided
between the source and the patient for adjustment of the X-ray field
size in the plane of the image receptor to the size of that portion
of the image receptor which has been selected on the spot-film
selector. Such adjustment shall be accomplished automatically when
the X-ray field size in the plane of the image receptor is greater
than the selected portion of the image receptor. If the X-ray field
size is less than the size of the selected portion of the image
receptor, the field size shall not open automatically to the size of
the selected portion of the image receptor unless the operator has
selected that mode of operation.
2. Neither the length nor
width of the X-ray field in the plane of the image receptor shall
differ from the corresponding dimensions of the selected portion of
the image receptor by more than three percent (3%) of the SID when
adjusted for full coverage of the selected portion of the image
receptor. The sum, without regard to sign, of the length and width
differences shall not exceed four percent (4%) of the SID. On spot
film devices manufactured after February 25, 1978, if the angle
between the plane of the image receptor and beam axis is variable,
means shall be provided to indicate when the axis of the X-ray beam
is perpendicular to the plane of the image receptor, and compliance
shall be determined with the beam axis indicated to be perpendicular
to the plane of the image receptor.
3. The center of the X-ray
field in the plane of the image receptor shall be aligned with the
center of the selected portion of the image receptor to within two
percent (2%) of the SID.
4. Means shall be provided to
reduce the X-ray field size in the plane of the image receptor to a
size smaller than the selected portion of the image receptor such
that:
a. For spot-film devices used
on fixed-SID fluoroscopic systems which are not required to, and do
not provide stepless adjustment of the X-ray field, the minimum field
size, at the greatest SID, does not exceed one-hundred twenty-five
square cm (125 cm 2 ); or
b. For spot-film devices used
on fluoroscopic systems that have a variable SID and/or stepless
adjustment of the field size, the minimum field size, at the greatest
SID, shall be containable in a square of five (5) cm by five (5) cm.
5. A capability may be
provided for overriding the automatic X-ray field size adjustment in
case of system failure. If it is so provided, a signal visible at the
operator’s position shall indicate whenever the automatic X-ray
field size adjustment override is engaged. Each such system failure
override switch shall be clearly labeled as follows: FOR X-RAY FIELD
LIMITATION SYSTEM FAILURE
4.7 COMPUTED TOMOGRAPHY SYSTEMS
4.7.1 Requirements for
Equipment
A. Applicability. Unless
otherwise specified, the requirements for equipment contained in §
4.7.1 of this Part are applicable to CT X-ray systems manufactured or
remanufactured on or after September 3, 1985.
B. Termination of Exposure.
1. Means shall be provided to
terminate the X-ray exposure automatically by either de-energizing
the X-ray source or shuttering the X-ray beam in the event of
equipment failure affecting data collection. Such termination shall
occur within an interval that limits the total scan time to no more
than one hundred ten percent (110%) of its preset value through the
use of either a backup timer or devices which monitor equipment
function.
2. A visible signal shall
indicate when the X-ray exposure has been terminated through the
means required by § 4.7.1(B) of this Part.
3. The operator shall be able
to terminate the X-ray exposure at any time during a scan, or series
of scans under CT system control, of greater than one-half (0.5)
second duration.
C. Tomographic Plane
Indication and Alignment.
1. For any single tomogram
system, means shall be provided to permit visual determination of the
tomographic plane or a reference plane offset from the tomographic
plane.
2. For any multiple tomogram
system, means shall be provided to permit visual determination of the
location of a reference plane. This reference plane can be offset
from the location of the tomographic planes.
3. If a device using a light
source is used to satisfy §§ 4.7.1(C)(1) or (2) of this Part, the
light source shall provide illumination levels sufficient to permit
visual determination of the location of the tomographic plane or
reference plane under ambient light conditions of up to five hundred
(500) lux.
D. Beam-On and Shutter Status
Indicators and Control Switches.
1. The CT X-ray control and
gantry shall provide visual indication whenever X-rays are produced
and, if applicable, whether the shutter is open or closed.
2. Each emergency button or
switch shall be clearly labeled as to its function.
E. Indication of CT Conditions
of Operation. The CT system shall be designed such that the CT
conditions of operation to be used during a scan or a scan sequence
shall be indicated prior to the initiation of a scan or scan
sequence. On equipment having all or some of these conditions of
operation at fixed values, this requirement may be met by permanent
markings. Indication of CT conditions of operation shall be visible
from any position from which scan initiation is possible.
F. Extraneous Radiation. When
data are being collected for image production, the radiation adjacent
to the tube port shall not exceed that permitted by § 4.4.5 of this
Part.
G. Maximum Surface CTDI
Identification. The angular position where the maximum surface CTDI
occurs shall be identified to allow for reproducible positioning of a
CT dosimetry phantom.
H. Additional Requirements
Applicable to CT X-Ray Systems Containing a Gantry.
1. The total error in the
indicated location of the tomographic plane or reference plane shall
not exceed five (5) millimeters.
2. If the X-ray production
period is less than one-half (0.5) second, the indication of X-ray
production shall be actuated for at least one-half (0.5) second.
Indicators at or near the gantry shall be discernible from any point
external to the patient opening where insertion of any part of the
human body into the primary beam is possible.
3. The deviation of indicated
scan increment versus actual increment shall not exceed plus or minus
1 millimeter with any mass from zero (0) to one hundred (100)
kilograms resting on the support device. The patient support device
shall be incremented from a typical starting position to the maximum
incremented distance or 30 centimeters, whichever is less, and then
returned to the starting position. Measurement of actual versus
indicated scan increment may be taken anywhere along this travel.
4. Premature termination of
the X-ray exposure by the operator shall necessitate resetting of the
CT conditions of operation prior to the initiation of another scan.
4.7.2 Facility Design
Requirements.
A. Aural Communication.
Provision shall be made for two-way aural communication between the
patient and the operator at the control panel.
B. Viewing Systems.
1. Windows, mirrors,
closed-circuit television, or an equivalent shall be provided to
permit continuous observation of the patient during irradiation and
shall be so located that the operator can observe the patient from
the control panel.
2. When the primary viewing
system is by electronic means, an alternate viewing system (which may
be electronic) shall be available for use in the event of failure of
the primary viewing system.
4.7.3 Radiation Output
Measurements, Spot Checks, and Operating Procedures
A. Output Measurements.
1. The measurement of the
radiation output of the CT X-ray system shall be performed by, or
under the direction of, a Qualified Medical Physicist.
2. The measurement of the
radiation output of a CT X-ray system shall be performed:
a. Before the first medical
use following installation or reinstallation of the CT X-ray system;
and
b. At intervals not to exceed
twelve (12) months; and
c. After any change or
replacement of components which, in the opinion of the Qualified
Medical Physicist, could cause a change in the radiation output.
3. CT dosimetry phantom(s)
shall be used in determining the radiation output of a CT X-ray
system. Such phantom(s) shall meet the following specifications and
conditions of use:
a. CT dosimetry phantoms shall
be right circular cylinders of polymethyl methacrylate of density
1.19 plus or minus 0.01 grams per cubic centimeter (g/cm 3 ).
The phantoms shall be at least fourteen (14) centimeters in length
and shall have diameters of thirty-two (32.0) centimeters for testing
CT X-ray systems designed to image any section of the body and
sixteen (16.0) centimeters for systems designed to image the head or
for whole body scanners operated in the head scanning mode.
b. CT dosimetry phantom(s)
shall provide means for the placement of a dosimeter(s) along the
axis of rotation and along a line parallel to the axis of rotation
1.0 centimeter from the outer surface and within the phantom. Means
for the placement of dosimeters or alignment devices at other
locations may be provided.
c. Any effects on the doses
measured due to the removal of phantom material to accommodate
dosimeters shall be accounted for through appropriate corrections to
the reported data or included in the statement of maximum deviation
for the values obtained using the phantom.
d. All dose measurements shall
be performed with the CT dosimetry phantom placed on the patient
couch or support device without additional attenuation materials
present.
4. These radiation output
measurements shall be required for a representative type of head and
body scans performed at the facility.
5. The CTDI along the two (2)
axes specified in § 4.7.3(B)(4)(b) of this Part shall be measured.
The CT dosimetry phantom shall be oriented so that the measurement
point 1.0 centimeter from the outer surface and within the phantom is
in the same angular position within the gantry as the point of
maximum surface CTDI identified. The CT conditions of operation shall
correspond to typical values used by the registrant.
a. For the purpose of
determining the CTDI, the manufacturer's statement as to the nominal
tomographic section thickness for that particular system may be
utilized.
6. Procedures for measurement
of radiation output shall be in writing. Records of radiation
measurements performed shall be maintained for inspection by the
Agency.
7. The dose profile along the
center axis of the CT dosimetry phantom for the minimum, maximum, and
midrange values of the nominal tomographic section thickness used by
the registrant shall be readily available.
B. Spot-checks
1. The spot-check procedures
shall be in writing and shall have been developed by a Qualified
Medical Physicist.
2. The spot-check procedures
shall incorporate the use of a CT imaging phantom which has the
capability of providing an indication of contrast scale, noise,
nominal tomographic section thickness, the resolution capability of
the system for low and high contrast objects, and measuring the mean
CTN for water or other reference material.
3. Spot-checks shall be
evaluated for compliance with tolerance limits specified pursuant to
§ 4.7.3(C)(1) of this Part at the time the radiation measurements
required by § 4.7.3(B) of this Part are performed.
4. Spot-checks shall include
acquisition of images obtained with the CT imaging phantoms. The
images shall be retained, until a new set of radiation measurements
is performed as follows:
a. If applicable, photographic
copies of the images obtained from the image display device;
b. Images stored in digital
form on a storage medium compatible with the CT X-ray system; and
c. Acceptance criteria for
image validation shall be documented.
5. The registrant shall
maintain a record of each spot check required by § 4.7.3(C) of this
Part for three (3) years.
C. Operating Procedures
1. The CT X-ray system shall
not be operated except by an individual who has been specifically
trained in its operation.
2. Information shall be
readily available regarding the operation of the system. Such
information shall include the following:
a. The latest set of radiation
measurements and spot-checks;
b. Instructions on the use of
the CT imaging phantom, including a schedule of spot-checks
appropriate for the system, and allowable variations for the
indicated parameters;
c. The distance in millimeters
between the tomographic plane and the reference plane if a reference
plane is utilized; and
d. Current imaging protocols
shall be available at the control panel which specify the CT
conditions of operation and the number of scans for each routine
examination.
3. If the measurement of
radiation output or spot-check of the CT X-ray system identifies that
a system operating parameter has exceeded a tolerance established by
a Qualified Medical Physicist, report the problem to the service
engineer and notify the Qualified Medical Physicist. The registrant
shall maintain a record of all such notifications for three (3)
years.
4.7.4 CT X-ray System Used
for Radiation Therapy Simulation
A. A CT X-ray system used
solely for radiation therapy simulation is exempt from the specific
requirements of §§ 4.7.1, 4.7.2, and 4.7.3 of this Part, and is
only subject to the requirements of § 5.10 of this Subchapter.
B. A CT X-ray system used for
both diagnostic X-ray and radiation therapy simulation is subject to
the requirements of both § 4.7 of this Part and § 5.10 of this
Subchapter.
4.8 MAMMOGRAPHY
4.8.1 Applicability
The provisions of this
section are in addition to, and not in substitution for, other
applicable provisions of this Subchapter.
4.8.2 Certification
Requirements
A. Only X-ray systems in
compliance with the requirements of the Mammography Quality Standards
Reauthorization Act of 1998, Public Law 105-248, and 21 C.F.R. Part
900 shall be used for screening and diagnostic mammography.
B. A facility performing
mammography shall have a valid certificate issued by the U.S.
Department of Health and Human Services, pursuant to the Mammography
Quality Standards Reauthorization Act of 1998, Public Law 105-248,
and 21 C.F.R. Part 900.
C. A facility performing
mammography shall ensure that the additional mammography activities
of processing the x-ray film, interpreting the image, and maintaining
viewing conditions, wherever performed, meet all quality standards
pursuant to the Mammography Quality Standards Reauthorization Act of
1998, Public Law 105-248, and 21 C.F.R. Part 900.
4.8.3 Retention of
Mammography X-rays.
Pursuant to RI Gen. Laws §
23-4.9-1, each mammographic imaging facility that takes a mammography
x-ray of any individual within Rhode Island shall keep and maintain
that mammography x-ray for the life of the individual. However, any
mammography x-ray may be destroyed if the individual has had no
contact with the mammographic imaging facility for a period exceeding
fifteen (15) years.
4.9 BONE DENSITOMETRY
4.9.1 Bone Densitometry
Systems
A. Bone densitometry systems
shall be:
1. Certified by the
manufacturer pursuant to the Medical Device Act and Subchapter C -
Electronic Product Radiation Control (EPRC) of Chapter V of the
Federal Food, Drug and Cosmetic Act.;
2. Registered in accordance
with Part 3 of this Subchapter; and
3. Maintained and operated in
accordance with the manufacturer’s specification and
recommendations.
4.9.2 Equipment
Requirements
Systems with stepless
collimators shall be provided with means to both size and align the
X-ray field such that the X-ray field at the plane of the image
receptor does not extend beyond two percent (2%) of the SID.
4.9.3 Bone Densitometry
System Operators
A. Operators of bone
densitometry systems shall be:
1. Licensed as a practitioner
of the healing arts; or
2. Individuals who possess a
current license in accordance with Licensure of Radiographers,
Nuclear Medicine Technologists, Radiation Therapists and Radiologist
Assistants [Subchapter 05 Part 34 of this Chapter], unless the
individual is specifically exempted from licensure by said
regulations; or
3. Individuals who are not
subject to licensure under Subchapter 05 Part 34 of this Chapter, and
have been instructed in the proper use of the bone densitometry
system. As a minimum, such instruction shall include:
a. Basic radiation protection;
b. Operating procedures for
bone densitometry systems, to include use of various system
functions, safety, and maintenance; and
c. Patient positioning for the
types of examinations performed.
4.9.4 Bone Densitometry
System Operation
A. During the operation of any
bone densitometry system:
1. The operator, ancillary
personnel, and members of the general public shall be positioned at
least one meter from the patient and bone densitometry system during
the examination.
2. The operator shall advise
the patient that the bone densitometry examination is a type of X-ray
procedure.
4.9.5 Maintenance of
Records
The registrant shall keep
maintenance records for bone densitometry systems as prescribed by §
4.9.1(A)(3) of this Part. These records shall be maintained for
inspection by the Agency for five (5) years from the date the
maintenance action was completed.
4.9.6 Bone Densitometry
Examination Requirements
A. Bone densitometry on human
patients shall be conducted only:
1. Under a prescription of a
licensed practitioner of the healing arts; or
2. Under a screening program
approved by the Agency.
4.9.7 Submission of
Information
Any person proposing to
conduct a bone densitometry screening program shall submit the
information outlined in § 4.11 of this Part, and include the name
and address of the licensed practitioner of the healing arts who will
interpret the screening results.
4.10 QUALITY ASSURANCE PROGRAM.
4.10.1 Quality Assurance
A. Except where otherwise
specified by the provisions of § 4.10.1(A)(7) of this Part, all
registrants of diagnostic X-ray imaging equipment shall establish and
maintain a quality assurance program consisting of quality control
assessments addressing at least the following items:
1. Administration:
a. Written standard operating
procedures on radiation protection are reviewed and updated by
management at intervals not to exceed twelve (12) months;
b. Employee review and written
acknowledgement of standard operating procedures and policies on
radiation protection;
c. Credentialing of
practitioners, medical physicists, and X-ray equipment operators; and
d. Record retention in
accordance with applicable Rhode Island statutes and regulations, but
in no case less than three (3) years.
2. Image Processing Equipment:
Compliance with § 4.3.10 of this Part;
3. Radiographic Equipment:
a. Compliance with performance
standards in §§ 4.4 and 4.6 of this Part, as specified by a
Qualified Medical Physicist;
b. Estimated entrance skin
exposures for selected patient examinations;
c. Image printing and viewing
equipment;
d. Evaluation of image
quality; and
e. Radiation protection.
4. Fluoroscopic Equipment:
a. Compliance with performance
standards in §§ 4.4 and 4.5 of this Part, as specified by a
Qualified Medical Physicist;
b. Low and high contrast
resolution; and
c. Radiation protection.
5. Computerized Tomography
Equipment:
a. Compliance with performance
standards in § 4.7 of this Part, as specified by a Qualified Medical
Physicist;
b. CT number;
c. Low and high contrast
resolution;
d. Dosimetry of selected
patient examinations to include pediatric patients if applicable;
e. Image printing and viewing
equipment; and
f. Radiation protection.
6. Bone Densitometry
Equipment:
a. Compliance with
requirements in § 4.9 of this Part.
7. Clarification of required
quality assurance program elements for certain mammography and dental
X-ray facilities.
a. The requirements in § 4.10
of this Part do not pertain to diagnostic X-ray imaging equipment
subject to the Mammography Quality Standards Reauthorization Act of
1998, Public Law 105-248, and 21 C.F.R. Part 900.
b. Registrants performing
diagnostic radiography limited to intra-oral dental procedures and/or
panoramic procedures and cephalometric procedures which do not
utilize an open beam configuration are only required to comply with
§§ 4.10.1(A)(1)(a), (b), (d), and 4.10.1(A)(2) of this Part.
4.10.2 Availability of
Quality Assurance Program
The quality assurance program
shall be in written form and available for review by the Agency.
4.10.3 Implementation of
Quality Assurance Program
A. The registrant shall assign
qualified personnel to fully implement the quality assurance program.
Quality control assessments for §§ 4.10.1(A)(2), (3), (4) and (5)
of this Part shall be conducted by, or under the direction of, a
Qualified Medical Physicist.
B. A Qualified Medical
Physicist shall determine the frequency and nature of quality control
tests, except when the frequency for a specific quality control test
is defined by this Subchapter.
C. A Qualified Medical
Physicist shall perform a review of the Quality Assurance Program at
an interval not to exceed twelve (12) months, and shall provide a
written report which documents the results of this review.
4.11 INFORMATION TO BE SUBMITTED
BY PERSONS PROPOSING TO CONDUCT HEALING ARTS SCREENING
A. Persons requesting that the
Agency approve a healing arts screening program shall submit the
following information and evaluation:
1. Name and address of the
applicant and, where applicable, the names and addresses of agents
within Rhode Island.
2. Diseases or conditions for
which the X-ray examinations are to be used in diagnoses.
3. A description of the X-ray
examinations proposed in the screening program (i.e., type and number
of views).
4. Description of the
population to be examined in the screening program, i.e., age range,
gender, physical condition, and other appropriate information.
5. An evaluation of any known
alternate methods not involving ionizing radiation that could achieve
the goals of the screening program and why these methods are not used
in preference to the X-ray examinations.
6. An evaluation conducted by
a Qualified Medical Physicist, of the X-ray system(s) to be used in
the screening program. The evaluation shall include the following:
a. Documentation that such
system(s) satisfy all requirements of this Subchapter; and
b. Estimation of patient
entrance skin exposures from the X-ray examinations to be performed.
7. A description of the
diagnostic X-ray quality control program.
8. Documentation of the
techniques for the X-ray examination procedures to be used.
9. The name and RI license
number of each radiologic technologist who will be operating the
X-ray system(s).
10. The name and RI license
number of each health care provider who will be supervising the
operators of the X-ray system(s). The extent of supervision and the
method of work performance evaluation shall be specified.
11. The name and address of
the Rhode Island-licensed practitioner of the healing arts who will
interpret the images.
12. Procedures to be used in
advising the individuals screened and their health care provider(s)
of the results of the screening procedure and any further medical
needs indicated.
13. Procedures for the
retention or disposition of the images and other records pertaining
to the X-ray examinations.
14. Frequency of screening of
individuals.
15. The duration of the
screening program.
4.12 INSTRUCTION OF USERS OF X-RAY
EQUIPMENT IN THE HEALING ARTS
A. Fundamentals of Radiation
Safety
1. Characteristics of
x-radiation.
2. Units of radiation dose.
3. Hazards of excessive
exposure to radiation.
4. Levels of radiation from
sources of radiation.
5. Methods of controlling
radiation dose.
a. Working time.
b. Working distances.
c. Shielding.
B. Radiation Detection
Instrumentation to be Used
1. Radiation survey
instruments.
a. Operation.
b. Calibration.
c. Limitations.
2. Survey, monitoring and
spot-check techniques.
3. Personnel monitoring
devices.
4. Interpretation of personnel
monitoring reports.
C. Operation and Control of
X-ray Equipment
1. Collimation and filtration.
2. Exposure techniques for the
equipment used.
3. Image processing
techniques.
D. Anatomy and positioning
1. Relevant human anatomy.
2. Relevant human physiology.
3. Radiographic positioning.
E. The requirements of
pertinent federal and state regulations.
F. The licensee's or
registrant's written operating and emergency procedures.
4.13 REQUIREMENTS FOR USE OF
HAND-HELD INTRAORAL DENTAL RADIOGRAPHIC UNIT
A. The following requirements
are applicable to intraoral dental radiographic units designed to be
operated as a hand-held unit:
1. For All Uses:
a. Operators of hand-held
intraoral dental radiographic units shall be specifically trained to
operate such equipment.
b. When operating a hand-held
intraoral dental radiographic unit, operators shall wear a protective
apron and thyroid collar, unless otherwise authorized by the Agency
or recommended by a Qualified Medical Physicist.
c. A hand-held intraoral
dental radiographic unit shall be held with minimal motion during a
patient examination. A tube stand may be utilized to immobilize a
hand-held intraoral dental radiographic unit during patient
examination.
d. Unless otherwise authorized
by the Agency, a hand-held intraoral dental radiographic unit shall
be used with a secondary radiation block to shield the operator.
e. The operator shall ensure
there are no bystanders within a radius of six (6) feet from the
patient being examined with a hand-held intraoral radiographic unit.
f. Hand-held intraoral dental
radiographic units shall not be used for patient examinations in
hallways and waiting rooms.
g. The registrant shall comply
with any facility-specific requirements established by the Agency.
2. Additional Requirements for
Operatories in Permanent Facilities:
a. When hand-held intraoral
dental radiographic units are used for patient examinations in dental
operatories, that facility shall meet the structural shielding
requirements specified by the Agency or by a health physicist or
Qualified Medical Physicist.