6.1 Authority
A. This Part is promulgated
pursuant to the authority conferred under R.I. Gen. Laws § 23-1.3 -5 .
B. This Part establishes
special requirements for the use of industrial radiation machines not
otherwise covered by this Subchapter. The requirements of this Part
are in addition to, and not in substitution for, other applicable
requirements of this Subchapter.
C. Any notifications, reports
or correspondence required by this Part shall be directed to the
Agency using contact information specified in § 1.4 of this
Subchapter.
6.2 Exemptions
Uses of portable/handheld
fluorescence x-ray (open beam) devices that are manufactured without
safety devices are exempt from the requirements of § 6.5(A) of this
Part.
6.3 Definitions
A. Whenever used in this Part,
the following terms shall be construed as follows:
1. “Act” means R.I. Gen.
Laws Chapter 23-1.3 entitled "Radiation Control".
2. “Agency” means Rhode
Island Radiation Control Agency (RCA), Center for Health Facilities
Regulation - Radiation Control Program, Rhode Island Department of
Health.
3. "Bomb detection
radiation machine" means x-ray generating equipment used solely
for the purpose of remotely detecting explosive devices. This
definition does not include hand-held x-ray bomb detection equipment
for the purposes of this Part.
4. "Cabinet X-ray system”
means an X-ray system with the X-ray tube installed in an enclosure
(hereinafter termed "cabinet") that is independent of
existing architectural structures except the floor. The cabinet X-ray
system is intended to contain at least that portion of a material
being irradiated, provide radiation attenuation, and exclude
personnel from its interior during generation of radiation. Included
are all X-ray systems designed primarily for the inspection of
carry-on baggage at airline, railroad, and bus terminals, and in
similar facilities. An X-ray tube used within a shielded part of a
building, or X-ray equipment which may temporarily or occasionally
incorporate portable shielding is not considered a cabinet X-ray
system.
5. “Category A industrial
radiation machine” means a device capable of generating or emitting
fields of radiation in an open beam configuration during normal
conditions of use. This includes, but is not limited to,
portable/handheld fluorescence x-ray, fluoroscopy hand held
intensified, fluoroscopy x-ray, flash x-ray, flash x-ray for bomb
detection, spectrography x-ray, diffraction x-ray and uncertified
cabinet x-ray.
6. “Category B industrial
radiation machine” means a device capable of generating or emitting
fields of radiation where the beam is contained during normal
conditions of use. This includes, but is not limited to, package
x-ray, certified and certifiable cabinet x-ray, x-ray fluorescence
units and similar devices.
7. "Certifiable cabinet
X-ray system" means an existing uncertified X-ray system that
has been modified to meet the certification requirements specified in
21 C.F.R. § 1020.40.
8. "Certified cabinet
X-ray system" means an X-ray system which has been certified in
accordance with 21 C.F.R. § 1010.2 as being manufactured and
assembled pursuant to the provisions of 21 C.F.R. § 1020.40.
9. "Interlock" means
a device arranged or connected such that the occurrence of an event
or condition is required before a second event or condition can occur
or continue to occur.
10. "Leakage radiation"
means all radiation coming from within the source housing, except the
useful beam.
11. "Open-beam
configuration" means an analytical X-ray system in which an
individual could accidentally place some part of his body in the
primary beam path during normal operation.
12. “Registrant” means any
person who is registered with the Agency and is legally obligated to
register with the Agency pursuant to this Subchapter and the Act.
13. “Registration” means
registration with the Agency pursuant to this Subchapter and the Act.
6.4 General Requirements - All
Industrial Radiation Machines
6.4.1 Radiation Levels
The local components of an
industrial radiation machine shall be located and arranged and shall
include sufficient shielding or access control such that no radiation
levels exist in any area surrounding the local component group which
could result in a dose to an individual present in the area in excess
of the dose limits given in § 1.8.1 of this Subchapter.
6.4.2 Warning Devices
A. The x-ray control shall
provide visual indication whenever x-rays are produced.
B. All ancillary warning
devices shall be labeled so that their purpose is easily identified
and shall have fail-safe characteristics.
C. Posting. Each area or room
containing industrial radiation machines shall be conspicuously
posted with a sign or signs bearing the radiation symbol and the
words "CAUTION - X-RAY EQUIPMENT", or words having a
similar intent.
D. Ports. Unused ports on
industrial radiation machine source housings shall be secured in the
closed position in a manner which will prevent inadvertent opening.
E. Labeling. Each registrant
shall ensure that each industrial radiation machine is labeled in a
conspicuous manner to caution individuals that radiation is produced
when it is energized. This label shall be affixed in a clearly
visible location on the face of the control unit. If the industrial
radiation machine is not visible from the control unit, the
industrial radiation machine shall have a visible indication that it
is energized.
F. Radiation Source Housing.
Each radiation source housing shall be equipped with an interlock
that shuts off the tube if it is removed from the radiation source
housing or if the housing is disassembled.
6.5 Additional Requirements -
Category A Industrial Radiation Machines
A. Safety Device. A safety
device shall be provided on all open-beam configurations which
prevents the entry of any portion of an individual's body into the
primary X-ray beam path or which causes the beam to be shut off upon
entry into its path shall be provided on all open-beam
configurations.
1. A registrant may apply to
the Agency for an exemption from the requirement of a safety device.
Such application shall include:
a. A description of the
various safety devices that have been evaluated;
b. The reason each of these
devices cannot be used; and
c. A description of the
alternative methods that will be employed to minimize the possibility
of an accidental exposure, including procedures to assure that
operators and others in the area will be informed of the absence of
safety devices.
B. Warning Devices. Open-beam
configurations shall be provided with a visible indication of:
1. X-ray tube status (ON-OFF)
located near the radiation source housing, if the primary beam is
controlled in this manner; and/or
2. Shutter status
(OPEN-CLOSED) located near each port on the radiation source housing,
if the primary beam is controlled in this manner.
C. Shutters. On open-beam
configurations each port on the radiation source housing shall be
equipped with a shutter that cannot be opened unless a collimator or
a coupling has been connected to the port.
D. Surveys.
1. Radiation surveys, as
required by § 1.10.2 of this Subchapter, of industrial radiation
machines sufficient to show compliance with § 6.4(A) of this Part
shall be performed:
a. Upon installation of the
equipment, and at least once every twelve (12) months thereafter;
b. Following any change in the
initial arrangement, number, or type of local components in the
system;
c. Following any maintenance
requiring the disassembly or removal of a local component in the
system;
d. During the performance of
maintenance and alignment procedures if the procedures require the
presence of a primary X-ray beam when any local component in the
system is disassembled or removed; and
e. Any time a visual
inspection of the local components in the system reveals an abnormal
condition.
f. Whenever personnel
monitoring devices show a significant increase over the previous
monitoring period or the readings are approaching the radiation dose
limits.
2. Radiation survey
measurements shall not be required if a registrant can demonstrate,
to the satisfaction of the Agency, compliance with § 6.4(A) of this
Part in some other manner.
E. Generator Cabinet. Each
X-ray generator shall be supplied with a protective cabinet which
limits leakage radiation measured at a distance of five (5) cm from
its surface such that it is not capable of producing a dose in excess
of 0.5 mrem (5 µSv) in any one (1) hour.
6.6 Additional Requirements -
Category B Industrial Radiation Machines
A. All Category B industrial
radiation machines shall be evaluated in accordance with the
following requirements:
1. The registrant shall
perform an evaluation of the radiation dose limits to determine
compliance with §§ 1.8.1(A), (B) and (C) of this Subchapter at
intervals not to exceed twelve (12) months. The registrant shall
ensure that radiation emitted five (5) centimeters from the external
surface of the cabinet x-ray system does not exceed 0.5 millirem (5.0
µSv) in any one (1) hour;
2. Tests for proper operation
of interlocks shall be conducted and recorded at intervals not to
exceed twelve (12) months;
3. Records that demonstrate
compliance with § 6.6(A) of this Part shall be maintained by the
registrant for ten (10) years for inspection by the Agency.
B. Certified and Certifiable
Cabinet X-ray Systems. Certified and certifiable cabinet x-ray
systems, including those designed to allow admittance of individuals
shall also be maintained in compliance with 21 C.F.R. § 1020.40, and
no modification shall be made to the system unless prior Agency
approval has been granted.
6.7 Operating Requirements
A. Procedures. Operating and
safety procedures shall be written and made available to all
industrial radiation machine operators. No individual shall be
permitted to operate an industrial radiation machine in any manner
other than that specified in the procedures unless such individual
has obtained written approval of the radiation safety officer.
B. Bypassing. No individual
shall bypass a safety device or interlock unless such individual has
obtained the written approval of the radiation safety officer. Such
approval shall be for a specified period of time. When a safety
device or interlock has been bypassed, a readily discernible sign
bearing the words "SAFETY DEVICE NOT WORKING", or words
having a similar intent, shall be placed on the radiation source
housing.
C. Repair or Modification of
Industrial Radiation Machines. Except as specified in § 6.7(B) of
this Part, no operation involving removal of covers, shielding
materials or tube housing or modifications to shutters, collimators,
or beam stops shall be performed without ascertaining that the tube
is off and will remain off until safe conditions have been restored.
The main switch, rather than interlocks, shall be used for routine
shutdown in preparation for repairs.
6.8 Personnel Requirements
A. Instruction. No individual
shall be permitted to operate or maintain an industrial radiation
machine unless the individual has received instruction in and
demonstrated competence in the following:
1. Identification of radiation
hazards associated with the use of the industrial radiation machine;
2. Radiation warning and
safety devices incorporated into the industrial radiation machine, or
the reasons they have not been installed on certain pieces of
equipment and the extra precautions required in such cases;
3. Operating and safety
procedures for the industrial radiation machine; and
4. Proper procedures for
reporting an actual or suspected exposure in excess of the limits
specified in § 1.8.1 of this Subchapter.
B. Instructions for Bomb
Detection Radiation Machines. All personnel operating bomb detection
radiation machines shall be trained in the set-up and operation of
the radiation machine and in establishing a restricted area.
C. Individual Monitoring. In
addition to the requirements of § 1.10.3(A)(1) of this Subchapter,
finger dosimetric devices shall be provided to and shall be used by:
1. Industrial radiation
machine workers using systems having an open-beam configuration and
not equipped with a safety device; and
2. Personnel maintaining
industrial radiation machines if the maintenance procedures require
the presence of a primary X-ray beam when any local component in the
X-ray system is disassembled or removed.
D. Reported dose values shall
not be used for the purpose of determining compliance with § 1.7.1
of this Subchapter unless evaluated by an individual registered with
the Agency to provide General Radiation Physics Services.
E. Records and Documentation.
Records that demonstrate compliance with §§ 6.8(A) through (C) of
this Part shall be maintained by the registrant for ten (10) years
for inspection by the Agency. In addition to complying with the
requirements of §§ 6.8(A) through (C) of this Part, records of
individual monitoring results shall be maintained by the registrant
in accordance with § 1.16.6 of this Subchapter.