216-RICR-50-05-1
216-RICR-50-05-1. Public Drinking Water (version Amendment, 05/22/2008 to 05/14/2009)
RULES AND REGULATIONS PERTAINING TO
PUBLIC DRINKING WATER
[R46-13-DWQ]
STATE OF RHODE ISLAND AND PROVIDENCE PLANTATIONS
DEPARTMENT OF HEALTH
September 1977
AS AMENDED:
January 1983
August 1996
September 1989 (E)
September 1999
December 1990
April 2000 (T)
May 1991 (E)
June 2001
July 1991
August 1991 (E)
November 1991 (E)
February 1992
February 1992 (E)
January 2002 (re-filing in
accordance with the
provisions of section 42-
35-4.1 of the Rhode
Island General Laws, as
amended)
July 1992 (E)
January 2003
December 1992 (E)
January 2005
March 1993 (T)
April 1993 (E)
June 1993
September 1993
March 1994
January 2007 (re-filing in
accordance with the
provisions of section 42-
35-4.1 of the Rhode
Island General Laws, as
amended)
July 1994
May 2008
January 1995
February 1996 (E)
June 1996 (E)
i
INTRODUCTION
These amendments to Rules and Regulations Pertaining To Public Drinking Water (R46-13-
DWQ) are promulgated pursuant to the authority conferred under Section 46-13-18 of the
General Laws of Rhode Island, as amended, for the purpose of adopting revised PWS initial and
annual renewal licensure fees.
Pursuant to the provisions of Section 42-35-3(c) of the General Laws of Rhode Island, as
amended, the following were given consideration in arriving at the amended regulations: (1)
alternative approaches to the regulations; and (2) duplication or overlap with other state
regulations. No alternative approach, duplication, or overlap was identified based on available
information. Consequently, these rules are adequate in the best interest of the health and safety
of the public.
These amended regulations shall supersede all previous Rules and Regulations Pertaining to
Public Drinking Water promulgated by the Department of Health and filed with the Secretary of
State.
iii
TABLE OF CONTENTS
Definitions
1.0
Definitions
1
Coverage
2.0
Coverage
10
3.0
New Water Sources
13
4.0
Approval of Treatment Works, Storage and Pumping Facilities
15
Disinfection
5.0
Filtration and Disinfection
18
5.1
General Requirements
18
5.2
Criteria for avoiding filtration
19
5.3
Disinfection
22
5.4
Filtration
28
5.5
Analytical and monitoring requirements
30
5.6
Monitoring requirements for systems that do not provide filtration
31
5.7
Monitoring requirements for systems using filtration equipment
34
5.8
Reporting and record keeping requirements
36
CT Values
43
Lead & Copper
6.0
Control of Lead and Copper
47
7.0
Disinfectant Residuals, Disinfection Byproducts and Disinfection
103
Byproduct Precursors
General Requirements
8.0
Reserved
129
9.0
Assurance of Safety in Public Supply
130
10.0
Correction of Unsafe Conditions
131
11.0
Reports as to Public Supplies
132
12.0
Certified Laboratories
133
13.0
Ground Water Microbiology
134
14.0
Consecutive Water System Monitoring
135
15.0
Variances and Exemptions
136
Community Water
16.0
Community Water System Requirements
150
16.1
Inorganic Chemicals
150
16.2
Organic Chemicals
160
16.2(b)
Volatile Organic Chemicals
170
16.3
Turbidity
178
16.4
Microbiological
179
16.5
Radioactivity
186
iv
TABLE OF CONTENTS (Continued)
16.6
Unregulated Contaminants
197
16.7
Special Monitoring
198
Community Water
16.8
Public Notification
200
16.9
Records
235
16.10
Consumer Confidence Reports
235
Non-Community Water
17.0
NonCommunity Water System Requirements
276
17.1
Microbiological
276
17.2
Inorganic Chemicals
277
17.3
Organic Chemicals
277
17.4
Turbidity
278
17.5
Unregulated Contaminants and Special Monitoring
278
17.6
Public Notification
278
17.7
Records
278
Fees and Enforcement
18.0
Fee Schedule
280
19.0
Rules Governing Practices and Procedures
283
20.0
Violations, Noncompliance and Enforcement
284
21.0
Severability
290
Appendix 1
Analytical Methodology
291
Appendix 2
Reserved
324
Appendix 3
DWQ Penalty Matrix (1)
325
Appendix 4
Potential Sources of Groundwater Contamination
333
1
SECTION 1.0 DEFINITIONS
Wherever used in these rules and regulations the following terms shall be construed as follows:
1.1
“Act” - means Chapter 46-13 of the General Laws of Rhode Island.
“Action level” - is the concentration of lead or copper in water specified in Section
6.80(c) which determines, in some cases, the treatment requirements contained in Section
6 of these regulations that a water system is required to complete.
1.2
“Administrative penalty” - “Penalty” shall mean a monetary sum assessed by the Director
pursuant to these regulations in response to a violation of, or a failure to comply with, 46-
13 or any rule, regulation, license, permit or order adopted pursuant to the Director's
authority thereunder.
1.3
“Best available technology” - means the best technology, treatment techniques, or other
means which the EPA Administrator finds, after examination for efficacy under field
conditions and not solely under laboratory conditions, are available for a specific
contaminant or category of contaminants.
1.4
“Certified laboratory” - means an analytical laboratory licensed by the Rhode Island
Department of Health under Chapter 16.2 “Laboratories”, to perform biological,
microbiological, chemical or radiochemical examination of potable water or a laboratory
exempt from this law as provided for in 23-16.2-3 but which shall be certified by the
State Certification official in accordance with 40 CFR 1422.10b.
1.5
“Change of use” - means a different or expanded activity at an existing PWS which
significantly uses more or less water, or changes the duration of consumption between
transient and non-transient, than previously approved through application or documented
historical use.
1.6
“Coagulation” - means a process using coagulant chemicals and mixing by which
colloidal and suspended materials are destabilized and agglomerated into flocs.
1.7
“Community water system” - means the PWS which serves at least fifteen (15) service
connections used by year-round residents or regularly serves at least twenty-five (25)
year-round residents.
1.8
“Compliance cycle” - means the nine-year calendar year cycle during which PWSs must
monitor. Each compliance cycle consists of three-year compliance periods. The first
calendar year cycle begins January 1, 1993 and ends December 31, 2001; the second
begins January 1, 2002 and ends December 31, 2010; the third begins January 1, 2011
and ends December 31, 2019.
1.9
“Compliance period” - means a three-year calendar year period within a compliance
cycle. Each compliance cycle has three (3), three-year compliance periods. Within the
2
first compliance cycle, the first compliance period runs from January 1, 1993 to
December 31, 1995; the second from January 1, 1996 to December 31, 1998; and the
third from January 1, 1999 to December 31, 2001.
1.10
“Comprehensive performance evaluation (CPE)” - means a thorough review and analysis
of a treatment plant's performance-based capabilities and associated administrative,
operation and maintenance practices. It is conducted to identify factors that may be
adversely impacting a plant's capability to achieve compliance and emphasizes
approaches that can be implemented without significant capital improvements. For
purposes of compliance with Section 5.0 (f) of these regulations, the comprehensive
performance evaluation must consist of at least the following components: Assessment of
plant performance; evaluation of major unit processes; identification and prioritization of
performance limiting factors; assessment of the applicability of comprehensive technical
assistance; and preparation of a CPE report.
1.11
“Confluent growth” - means a continuous bacterial growth covering the entire filtration
area of a membrane filter, or a portion thereof, in which bacterial colonies are not
discrete.
1.12
“Connection” - means the water service line connecting a structure to the water
distribution line. In the absence of data on the number of service connections, the
population served divided by 2.5 shall be used as the default value.
a)
The following are excluded from the “connection” component of the PWS
definition:
A connection to a system that delivers water through constructed conveyances
other than pipes is excluded from consideration as a “connection” under three (3)
circumstances:
1)
Where the water is used exclusively for purposes other than residential
uses (consisting of drinking, bathing, and cooking, or other similar uses);
2)
Where the Director determines that alternative water to achieve the
equivalent level of public health protection provided by the applicable
national primary drinking water regulations is provided for drinking and
cooking;
3)
Where the Director determines that the water provided for drinking,
cooking, and bathing is treated (centrally or by point of entry) by the
provider, a pass-through entity, or the user to achieve the equivalent level
of protection provided by the applicable national primary drinking water
regulations.
If the application of one (1) or more of these exclusions reduces the
“connections” of a system providing water for human consumption
3
(through construction conveyances other than pipes) to fewer than fifteen
(15) service connections that serve fewer than twenty-five (25)
individuals, the supplier’s water system is not a PWS.
However, if the supplier’s remaining connections number fifteen (15) or
more, or if its remaining connections [even if they number fewer than
fifteen (15)] regularly serve at least twenty-five (25) individuals, then the
system is a PWS although the excluded connections are not considered
part of the PWS for as long as the exclusions apply and the system
complies with any conditions governing their applicability.
b)
An irrigation district in existence prior to May 18, 1994 that provides primarily
agricultural service through a piped water system with only incidental residential
or similar use shall not be considered to be a PWS if the system or the residential
or similar users of the system comply with subsections (a)(2) and (3) of this
definition.
1.13
“Conventional filtration treatment” - means a series of processes including coagulation,
flocculation, sedimentation, and filtration resulting in substantial particulate removal.
1.14
“Corrosion inhibitor” - means a substance capable of reducing the corrosivity of water
toward metal plumbing materials, especially lead and copper, by forming a protective
film on the interior surface of those materials.
1.15
“CT” or “CTcalc” - is the product of “residual disinfectant concentration” C in mg/L
determined before or at the first customer, and the corresponding disinfectant contact
time (T) in minutes, i.e., “C” x “T”. “CT99.9” is the CT value required for 99.9 percent
(3-log) inactivation of Giardia lamblia cysts. CT99.9 for a variety of disinfectants and
conditions appear in Tables 1.1-1.6, 2.1, and 3.1 of Section 5.6. CTcalc/CT99.9, is the
inactivation ratio. The sum of the inactivation ratios, or total inactivation ratio shown as
the sum of (CTcalc)/(CT99.9), is calculated by adding together the inactivation ratio for
each disinfection sequence. A total inactivation ratio equal to or greater than 1.0 is
assumed to provide a 3-log inactivation of Giardia lamblia cysts.
1.16
“Diatomaceous earth filtration” - means a process resulting in substantial particulate
removal in which (1) a precoat cake of diatomaceous earth filter media is deposited on a
support membrane (septum), and (2) while the water is filtered by passing through the
cake on the septum, additional filter media known as body feed is continuously added to
the feed water to maintain the permeability of the filter cake.
1.17
“Direct filtration” - means a series of processes including coagulation and filtration but
excluding sedimentation resulting in substantial particulate removal.
1.18
“Director” - means the Director of the Rhode Island Department of Health or his duly
authorized agent.
4
1.19
“Disinfectant contact time” (“T” in CT calculations) - means the time in minutes that it
takes for water to move from the point of disinfectant application or the previous point of
disinfectant residual measurement to a point before or at the point where residual
disinfectant concentration (“C”) is measured. Disinfectant contact time in pipelines must
be calculated based on “plug flow” by dividing the internal volume of the pipe by the
maximum hourly flow rate through that pipe. Disinfectant contact time within mixing
basins and storage reservoirs must be determined by tracer studies or an equivalent
demonstration.
1.20
“Disinfection” - means a process which inactivates pathogenic organisms in water by
chemical oxidants or equivalent agents.
1.21
“Disinfection profile” - is a summary of daily Giardia lamblia inactivation through the
treatment plant. The procedure for developing a disinfection profile is contained in
Section 5.
1.22
“Domestic or other non-distribution system plumbing problem” - means a coliform
contamination problem in a PWS with more than one (1) service connection that is
limited to the specific service connection from which the coliform-positive sample was
taken.
1.23
“Dose equivalent” – means the absorbed dose from ionizing radiation expressed in terms
of Rads multiplied by such a factor as account for differences in biological effectiveness
due to the type of radiation and its distribution in the body as specified by the
International Commission on Radiological Units and Measurements (ICRU).
1.24
“Effective corrosion inhibitor residual” - for the purpose of Section 6, means a
concentration sufficient to form a passivating film on the interior walls of a pipe.
1.25
“Filter profile” - is a graphical representation of individual filter performance, based on
continuous turbidity measurements or total particle counts versus time for an entire filter
run, from startup to backwash inclusively, that includes an assessment of filter
performance while another filter is being backwashed.
1.26
“Enhanced coagulation” - means the addition of sufficient coagulant for improved
removal of disinfection byproduct precursors by conventional filtration treatment.
1.27
“Enhanced softening” - means the improved removal of disinfection byproduct
precursors by precipitative softening.
1.28
“Filtration” - means a process for removing particulate matter from water by passage
through porous media.
1.29
“First draw sample” - means a one-liter sample of tap water, collected in accordance with
Section 6.86(b) (2), that has been standing in plumbing pipes at least 6 hours and is
collected without flushing the tap.
5
1.30
“Flocculation” - means a process to enhance agglomeration or collection of smaller floc
particles into larger, more easily settleable particles through gentle stirring by hydraulic
or mechanical means.
1.31
“GAC10” - means granular activated carbon filter beds with an empty-bed contact time
of 10 minutes based on average daily flow and a carbon reactivation frequency of every
180 days.
1.32
“Gross alpha particle activity” – means the total radioactivity due to alpha particle
emission as determined from measurements on a dry sample.
1.33
“Gross beta particle activity” – means the total radioactivity due to beta particle emission
as determined from measurements on a dry sample.
1.34
“Ground water under the direct influence of surface water” - means any water beneath the
surface of the ground with (1) significant occurrence of insects or other macroorganisms,
algae, or large-diameter pathogens such as Giardia lamblia or Cryptosporidium or (2)
significant and relatively rapid shifts in water characteristics such as turbidity,
temperature, conductivity, or pH which closely correlate to climatological or surface
water conditions. Direct influence must be determined for individual sources in
accordance with criteria established by the Director. The Director's determination of
direct influence may be based on site-specific measurements of water quality and/or
documentation of well construction characteristics and geology with field evaluation.
1.35
“Haloacetic acids (five) (HAA5)” - means the sum of the concentrations in milligrams
per liter of the haloacetic acid compounds (monochloroacetic acid, dichloroacetic acid,
trichloroacetic acid, monobromoacetic acid and dibromoacetic acid), rounded to two (2)
significant figures after addition.
1.36
“Initial compliance period” - means the first full three-year compliance period which
begins at least 18 months after promulgation, except for dichloromethane, 1,2,4
trichlorobenzene,
1,1,2-trichloroethane,
benzo[a]pyrene,
dalapon,
di(2-
ethylhexyl)adipate, di(2-ethylhexyl)phthalate, dinoseb, diquat, endothall, endrin,
glyphosate, hexachlorbenzene, hexachlorocyclopentadiene, oxamyl(Vydate), picloram,
simazine, 2,3,7,8-TCDD (Dioxin), antimony, beryllium, cyanide, nickle, and thallium,
initial compliance period means January 1993-December 1995 for systems with 150 or
more service connections and January 1996-December 1998 for systems having fewer
than 150 service connections.
1.37
“Large water system” - for the purpose of Section 6, means a water system that serves
more than 50,000 persons.
1.38
“Lead service line” - means a service line made of lead which connects the water main to
the building inlet and any lead pigtail, gooseneck or other fitting which is connected to
such lead line.
6
1.39
“Legionella” - means a genus of bacteria, some species of which have caused a type of
pneumonia called Legionnaires Disease.
1.40
“License” - means approval as specified in Section 46-13-2.1 of the General Laws of
Rhode Island, 1956 as amended.
1.41
“Manmade beta particle and photon emitters” - means all radionuclides emitting beta
particles and/or photons listed in Maximum Permissible Body Burdens and Maximum
Permissible Concentrations of Radionuclides in Air or Water for Occupational Exposure,
NBS Handbook 69, except the daughter products of thorium - 232, uranium - 235 and
uranium - 238.
1.42
“Maximum contaminant level” - means the maximum permissible level of a contaminant
in water which is delivered to any user of a PWS (PWS).
1.43
“Maximum residual disinfectant level (MRDL)” - means a level of a disinfectant added
for water treatment that may not be exceeded at the consumer's tap without an
unacceptable possibility of adverse health effects. For chlorine and chloramines, a PWS
is in compliance with the MRDL when the running annual average of monthly averages
of samples taken in the distribution system, computed quarterly, is less than or equal to
the MRDL. For chlorine dioxide, a PWS is in compliance with the MRDL when daily
samples are taken at the entrance to the distribution system and no two (2) consecutive
daily samples exceed the MRDL. MRDLs are enforceable in the same manner as
maximum contaminant levels under Section 1412 of the Safe Drinking Water Act. There
is convincing evidence that the addition of a disinfectant is necessary for control of
waterborne microbial contaminants. Notwithstanding the MRDLs listed in Section 7.2(a)
herein, operators may increase residual disinfectant levels of chlorine or chloramines (but
not chlorine dioxide) in the distribution system to a level and for a time necessary to
protect public health to address specific microbiological contamination problems caused
by circumstances such as distribution line breaks, storm runoff events, source water
contamination, or cross-connections.
1.44
“Maximum residual disinfectant level goal (MRDLG)” - means the maximum level of a
disinfectant added for water treatment at which no known or anticipated adverse effect on
the health of persons would occur, and which allows an adequate margin of safety.
MRDLGs are non-enforceable health goals and do not reflect the benefit of the addition
of the chemical for control of waterborne microbial contaminants.
1.45
“Medium-size water system” - for the purpose of Section 6 only, means a water system
that serves greater than 3,300 and less than or equal to 50,000 persons.
1.46
“Near the first service connection” - means at one (1) of the 20 percent of all service
connections in the entire system that are nearest the water supply treatment facility, as
measured by water transport time within the distribution system.
7
1.47
“Non-community water system” - means a PWS that is not a community water system.
1.48
“Noncompliance” - “Nonconformance” - “Failure to comply” - “Violation” - each mean
any act or failure to act which constitutes or results in or from:
(i)
engaging in any activity prohibited by, or not in compliance with the Act or any
rule, regulation, permit, approval or order adopted pursuant to the Director's
authority thereunder;
(ii)
engaging in any business or other activity without a necessary permit, or approval
that is required by law or regulation;
(iii)
the failure to perform, or the failure to perform in a timely fashion, anything
required by the Act, by a rule, regulation, permit, approval or order adopted
pursuant to the Director's authority.
1.49
“Non-transient non-community water system” - means a non-community water system
that regularly services at least twenty-five (25) of the same persons over six (6) months
per year.
1.50
“Optimal corrosion control treatment” - for the purpose of Section 6, means the corrosion
control treatment that minimizes the lead and copper concentrations at users' taps while
insuring that the treatment does not cause the water system to violate any other
regulations herein (Rules and Regulations Pertaining to Public Drinking Water).
1.51
“Order” - means the whole or a part of a final disposition by the Department, whether
affirmative, negative, injunctive, consent or declaratory in form, other than rulemaking
but including notices of violation, compliance orders, permits, and approvals issued
pursuant to the Director's authority.
1.52
“Permit” - means an authorization, or equivalent control document issued by the
Department to implement the requirements of 46-13.
1.53
“Person” - shall include an individual, partnership, association, or corporation, or any
town or city or any agency thereof, or the state or any agency thereof, or any other legal
entity.
1.54
“Picocurie (pCi)” - means a unit of radioactivity equal to 2.22 nuclear transformations per
minute.
1.55
“Point of disinfectant application” - means the point where the disinfectant is applied and
water downstream of that point is not subject to recontamination by surface water runoff.
1.56
“Point-of-entry treatment device (POE)” - means a treatment device applied to the
drinking water entering a house or building for the purpose of reducing contaminants in
the drinking water distributed throughout the house or building.
8
1.57
“Point-of-use treatment device (POU)” - means a treatment device applied to a single tap
used for the purpose of reducing contaminants in drinking water.
1.58
“PWS” - means a system for the provision to the public of water for human consumption
through pipes or other constructed conveyances, if such system has at least fifteen (15)
service connections or regularly serves at least twenty-five (25) individuals daily at least
sixty (60) days out of the year. Such term includes:
(i)
any collection, treatment, storage and distribution facilities under control of the
operator of such system and used primarily in connection with such system, and
(ii)
any collection or pretreatment storage facilities not under such control which are
used primarily in connection with such system.
1.59
“Rad” - means a unit of absorbed dose equal to 100 ergs per gram in any medium. (100
rad = 1 gray)
1.60
“Rem” - means the unit of dose equivalent from ionizing radiation to the total body or
any internal organ or organ system. (100 rem = 1 sievert)
1.61
“Repeat compliance period” - means any subsequent compliance period after the initial
compliance period.
1.62
“Requirement” - means any provision of the Act, or any rule, regulation, permit, approval
or order adopted pursuant to the Director's authority.
1.63
“Residual disinfectant concentration” (“C” in CT calculations) - means the concentration
of disinfectant measured in mg/1 in a representative sample of water.
1.64
“Sanitary survey” – means an on-site review of the water source, facilities, equipment,
operation and maintenance of a PWS for the purpose of evaluating the adequacy of such
source, facilities, equipment, operation and maintenance for producing and distributing
safe drinking water.
1.65
“Sedimentation” - means a process for removal of solids before filtration by gravity or
separation.
1.66
“Service line sample” - means a one-liter sample of water, collected in accordance with
Section 6.86(b)(3), that has been standing for at least 6 hours in a service line.
1.67
“Single family structure” - for the purpose of Section 6 only, means a building
constructed as a single-family residence that is currently used as either a residence or a
place of business.
9
1.68
“Slow sand filtration” - means a process involving passage of raw water through a bed of
sand at low velocity (generally less than 0.4 m/h or 1 gal./ft2/h resulting in substantial
particulate removal by physical and biological mechanisms.
1.69
“Small water system” - for the purpose of Section 6 only, means a water system that
serves 3,300 persons or fewer.
1.70
“Subpart H systems” - means PWSs using surface water or ground water under the direct
influence of surface water as a source that are subject to the requirements of Section 5 of
these regulations.
1.71
“Surface water” - means all water which is open to the atmosphere and subject to surface
runoff.
1.72
“SUVA” - means Specific Ultraviolet Absorption at 254 nanometers (nm), an indicator of
the humic content of water. It is a calculated parameter obtained by dividing a sample's
ultraviolet absorption at a wavelength of 254 nm (UV 254) (in m-1) by its concentration
of dissolved organic carbon (DOC) (in mg/L).
1.73
“System with a single service connection” - means a system which supplies drinking
water to consumers via a single service line.
1.74
“Too numerous to count” - means that the total number of bacterial colonies exceeds 200
on a 47-mm diameter membrane filter used for coliform detection.
1.75
“Total Organic Carbon (TOC)” - means total organic carbon in mg/L measured using
heat, oxygen, ultraviolet irradiation, chemical oxidants, or combinations of these oxidants
that convert organic carbon to carbon dioxide, rounded to two (2) significant figures.
1.76
“Transient non-community water system or TWS” - means a non-community water
system that does not regularly serve at least twenty-five (25) of the same persons over six
(6) months per year.
1.77
“Uncovered finished water storage facility” - means a tank, reservoir, or other facility
used to store water that will undergo no further treatment except residual disinfection and
is open to the atmosphere.
1.78
“Water purveyor” – means any person who owns or operates a PWS.
1.79
“Waterborne disease outbreak” - means the significant occurrence of acute infectious
illness, epidemiologically associated with the ingestion of water from a PWS which is
deficient in treatment, as determined by the appropriate local or State agency.
1.80
“Virus” - means a virus of fecal origin which is infectious to humans by waterborne
transmission.
10
SECTION 2.0 COVERAGE
2.1
These regulations apply to any PWS unless a PWS meets all of the following conditions:
a)
The system consists only of distribution or storage facilities (and does not have
any collection or treatment facilities);
b)
The system obtains all of its water from a PWS to which these regulations apply;
and
c)
The system does not sell water to any person.
2.2
General Requirements
a)
No person shall develop, maintain, or operate a public water supply system unless
said system is approved by the Director. Further, all public water supply systems
must be developed, operated and maintained in accordance with the requirements
and provisions of these regulations in order for a public water supply system to
maintain approval by the Director.
b)
Should the Director find that a public water supply system is not developed,
maintained, or operated in compliance with regulatory provisions, s/he may
revoke, suspend or otherwise limit the approval previously granted.
c)
The Director is authorized to enter at all reasonable times in or upon any private
or public property for the purpose of carrying out the provisions of these
regulations or making an inspection or investigation of a condition which the
Director believes may be hazardous to the health of the consumers serviced by
any public water supply system or in violation of the regulations or orders
promulgated under Chapter 46-13.
2.3
Licensing Requirement
a)
Applicability
Pursuant to the provisions of Section 46-13-2.1 of the General Laws of Rhode
Island, as amended, no person shall operate or maintain a public water supply
system unless the system is licensed by the Director under the provisions of this
subsection.
Persons subject to licensure shall be assessed initial and annual renewal licensure
fees in accordance with the fee schedule listed for each category of PWS in
Paragraph 2.3 c)2) of this subsection.
b)
License Application
1)
To apply for a license, a PWS shall submit a completed application to the
Director on forms provided for this purpose. The application shall include
all information required by these regulations, as well as by the form and
the accompanying instructions. Applications for a new community or
nontransient non-community PWS shall include a water system
management plan that demonstrates the financial, managerial, and
technical capacity to comply with statutory and regulatory requirements.
11
2)
The Director may at any time after filing of the original application require
further information in order to determine whether the application should
be approved or denied.
3)
Each application for a PWS license shall be signed by the applicant or a
person duly authorized to act on behalf of the applicant.
4)
No new PWS shall be licensed until: the application has been approved,
the PWS has been constructed in accordance with the approved plans and
the water has been sampled and found to be in compliance with the
requirements of these regulations.
c)
License Fees
1)
Pursuant to the provisions of Section 46-13-2.1 of the General Laws of
Rhode Island, as amended, the Director shall grant a license to a PWS that
meets the licensure requirements set forth in these regulations and upon
submission of the license fee as listed in Paragraph 2.3 c)2) of these
regulations made payable by check to the General Treasurer, State of
Rhode Island. Said license, unless sooner suspended or revoked, shall
expire on the 30th day of June following its issuance and must be renewed
from year-to-year.
2)
Effective 1 July 2009, the annual fee for licensure shall be as follows:
Transient non-community water system: two hundred dollars ($200.00).
Nontransient non-community water system: three hundred and thirty
dollars ($330.00).
Community water system: one dollar and fifty cents ($1.50) per
connection:
minimum fee = three hundred and thirty dollars ($330.00).
maximum fee = thirty two thousand-five hundred dollars ($32,500.00).
d)
Denial of License
1)
The Director may deny an application for a license if s/he determines that
the applicant has not demonstrated the ability to comply fully with the
applicable requirements established by the Act and/or by these regulations.
2)
An applicant whose application is denied may request a hearing in
accordance with the Administrative Procedures of the Rhode Island
Department of Health.
e)
Suspension or Revocation of a License
The Director may, for cause or for violation of these regulations, suspend or
revoke any license issued under this subsection. The Director may also review the
current status of any license with regard to current use of the water supply and
any change of use of the PWS.
12
f)
Renewal of License
1)
All licenses shall expire on the 30th day of June following its issuance
except as provided in 2.3 (f)(5).
2)
A renewal application must be filed with the Director by the 31st day of
May of each year on forms provided for this purpose.
3)
The appropriate licensing fee must accompany the renewal application.
4)
Renewal of a license shall be based upon: satisfactory compliance with the
regulations and timely submission of a renewal application and fee.
5)
In any case in which a PWS not less than 30 days prior to expiration of an
existing license, has filed a renewal application and fee in proper form for
renewal, such existing license shall not expire until final action on the
application has been taken by the Director.
g)
Licenses shall be issued only for the public water supply system and persons
named on the application and shall not be transferable or assignable. Existing
PWSs which have significant change of use of the water supply shall be reviewed
and modified as deemed appropriate by the Director.
13
SECTION 3.0 NEW WATER SOURCES
3.1
No source of water shall be developed for a PWS until a site plan prepared by a
professional engineer or land surveyor registered in accordance with Chapter 5-8 of
General Laws of Rhode Island, 1956, as amended, has been approved by the Director.
a)
Approval of plans and specifications granted an applicant shall expire within two
(2) years if construction of the approved source has not begun within that period.
b)
Expired approvals may be renewed if the data provided in the application is
unchanged and attested to by the applicant; and the plans conform with all
construction standards and testing requirements in effect at the time of application
for renewal.
3.2
In the case of a proposed gravel packed or gravel developed well, the site plan shall
contain pertinent information within at least 1750 feet of the proposed well including, but
not limited to, the location of existing and proposed sewage disposal systems and any
other existing or proposed potential sources of pollution including, but not limited to,
those listed in Appendix 4. Generally, the land within 400 feet of such wells shall be
reserved for protection of the water quality of the well, and shall be delineated on the site
plan by a topographic mapping of the 400 foot area to an appropriate scale. This distance
may be modified at the discretion of the Director taking into consideration such factors as
the volume and type of waste material to be disposed or stored in close proximity to the
land area reserved for protection of the well, the projected yield of the well, the depth
below grade to impervious formation, the depth below grade to the water table, the type
of soil in the area, or any other factors the Director deems pertinent.
3.3
In the case of a proposed drilled (rock), driven, or dug well, the site plan shall show
pertinent information within at least 1750 feet of the proposed well including, but not
limited to, the location of existing and proposed sewage disposal systems and any other
existing or proposed potential sources of pollution including but not limited to those
listed in Appendix 4. Generally, the land within 200 feet of such wells shall be reserved
for protection of the water quality of the well, and shall be delineated on the site plan by a
topographic mapping of the 200 foot area to an appropriate scale. This distance may be
modified at the discretion of the Director taking into consideration such factors as the
volume and type of waste material to be disposed or stored in close proximity to the land
area reserved for protection of the well, the depth below grade to impervious formation,
the depth below grade to the water table, the type of soil in the area, or any other factors
the Director deems pertinent.
3.4
In the case of a proposed surface water source, the site plan shall show pertinent
information within the entire watershed of the proposed surface water supply, but not
limited to the location of existing and proposed sewage disposal systems and any other
existing or proposed potential sources of pollution including, but not limited to, those
listed in Appendix 4. The portion of the watershed owned or controlled by the water
purveyor shall be clearly indicated. All surface water sources shall be provided with
14
water treatment consisting, as a minimum, of coagulation, sedimentation, filtration and
disinfection.
3.5
All revisions to approved plans must be submitted to the Director for approval. The
Director may require a new application and/or site plan if the revisions are deemed
significant.
3.6
Land reserved for the protection of the well as (indicated on the plan) approved by the
Director must remain under the direct control of the water supplier by either continued
ownership or recorded easement unless written permission to modify this area is granted
by the Director.
3.7
It is the responsibility of the water supplier to maintain the protective well area free from
potential sources of contamination as listed in Appendix 4.
3.8
Connection to another public water supply - A new public water supply shall not be
approved for use at any facility if another community public water supply is reasonably
accessible to such facility as determined by the Director, and permission to connect can
be obtained from the authority having jurisdiction.
3.9
Applications for approval of new water sources must be accompanied by an assessment
of the financial viability for said water system to maintain compliance with the
requirements of these regulations. The assessment shall include a discussion of operation
costs including: operation, maintenance, monitoring, anticipated future improvements,
debt repayment, and unforeseen emergencies or system breakdowns, and a discussion of
how the necessary revenues to pay for these costs will be raised.
15
SECTION 4.0 APPROVAL OF TREATMENT WORKS, STORAGE AND
PUMPING FACILITIES
4.1
No new water treatment works or water storage or pumping facilities shall be constructed
or such existing works or facilities substantially altered until design plans and
specifications prepared by a professional engineer registered in accordance with Chapter
5-8 of the General Laws of Rhode Island, as amended, and a plan for operation and
maintenance have been approved by the Director.
a)
Any chemical or substance added to a public water supply, any materials used in
the manufacture of public water supply components or appurtenances, or any
pipe, storage tank, valve, fixture or other materials which come in contact with
water intended for use in a public water supply shall meet American National
Standards Institute/NSF International standards, specifically ANSI/NSF Standard
60-1988 and ANSI/NSF Standard 61-1991 which are hereby adopted by
reference.
Only products which meet the standards adopted in or pursuant to this Section
shall be used by a supplier of water in a public water supply. Certification that a
product meets the standards adopted pursuant to this Section by an organization
having a third-party certification program accredited by American National
Standards Institute to test and certify products shall be prima facie evidence that a
product meets the standards.
Product Type
Standard
Drinking Water Treatment Chemicals
60
Pipes and Related Products
61
Protective (Barrier) Materials
61
Joining and Sealing Materials
61
Process Media
61
Mechanical Devices
61
Plumbing Devices
61
4.2
All newly constructed PWSs or additions to existing systems shall be flushed, adequately
disinfected, and the water examined for the presence of coliform organisms in accordance
with Appendix 1. No system shall be placed in use until such examination discloses the
absence of coliform organisms. Any waste water resulting from disinfection must be
disposed of properly, and with proper permits.
4.3
All revisions to approved plans must be submitted to the Director for approval. The
Director may require a new application and/or site plan if the revisions are deemed
significant.
4.4
Use of Non-Centralized Treatment Devices
16
a)
Criteria and procedures for PWSs using point-of-entry devices.
1)
PWSs may use point-of-entry devices to comply with maximum
contaminant levels only if they meet the requirements of this Section and
are approved by the Director.
2)
It is the responsibility of the PWS to operate and maintain the point-of-
entry treatment system.
3)
The PWS must develop and obtain the Director's approval for a
monitoring plan before point-of-entry devices are installed for compliance.
Under the plan approved by the Director, point-of-entry devices must
provide health protection equivalent to central water treatment.
“Equivalent” means that the water would meet all MCLS and would be of
acceptable quality similar to water distributed by a well-operated central
treatment plant. In addition to the VOCs, monitoring must include
physical measurements and observations such as total flow treated and
mechanical condition of the treatment equipment.
4)
Effective technology must be properly applied under a plan approved by
the Director and the microbiological safety of the water must be
maintained.
i)
Adequate certification of performance and field testing must be
provided as required by the Director.
ii)
The design and application of the point-of-entry devices must
consider the tendency for increase in heterotrophic bacteria
concentrations in water treated with activated carbon. It may be
necessary
to
use
frequent
backwashing,
post-contractor
disinfection, and Heterotrophic Plate Count monitoring to ensure
that the microbiological safety of the water is not compromised.
5)
All consumers shall be protected. Every building connected to the system
must have a point-of-entry device installed, maintained, and adequately
monitored. The Director must be assured that every building is subject to
treatment and monitoring, and that the rights and responsibilities of the
PWS customer convey with title upon sale of property.
4.5
Use of bottled water or point of use treatment devices
PWSs shall not use bottled water or point-of-use devices to achieve compliance with an
MCL. Bottled water or point-of-use devices may be used on a temporary basis to avoid
an unreasonable risk to health, and only with prior approval of the Director.
17
a)
Where bottled water is used, the PWS is fully responsible for the provision of
sufficient quantities of bottled water to every person supplied by the PWS. The
water system must use an approved bottled water supply.
b)
Where a point of use device is used, it must comply with the requirements of
Paragraph 4.4.
18
SECTION 5.0 FILTRATION AND DISINFECTION
5.1
General Requirements:
The requirements of this Section constitute Rhode Island’s primary drinking water
regulations. These regulations establish criteria under which filtration is required as a
treatment technique for PWSs supplied by a surface water source, or a ground water
source under the direct influence of surface water also referred to as Section 5.0 systems.
These regulations establish treatment technique requirements in lieu of maximum
contaminant levels for the following contaminants: Giardia lamblia, viruses,
heterotrophic plate count bacteria, Legionella, Cryptosporidium and turbidity.
Each Section 5.0 system must provide treatment of that source water that complies with
these treatment technique requirements.
5.1.1 The treatment technique requirements consist of installing and properly operating water
treatment processes which reliably achieve:
1)
At least 99.9 percent (3-log) removal and/or inactivation of Giardia lamblia cysts
between a point where the raw water is not subject to recontamination by surface
water runoff and a point downstream before or at the first customer, and
2)
At least 99.99 percent (4-log) removal and or inactivation of viruses between a
point where the raw water is not subject to recontamination by surface water
runoff and a point downstream before or at the first customer.
3)
At least 99 percent (2-log) removal of Cryptosporidium between a point where the
raw water is not subject to recontamination by surface water runoff and a point
downstream before or at the first customer for filtered systems, or
Cryptosporidium control under the watershed control plan for unfiltered systems.
This requirement applies to all systems except those that serve fewer than 10,000
people; they must meet this requirement beginning January 1, 2005.
4)
Compliance with the profiling and benchmark requirements under the provisions
of Section 5.3.7.
5.1.2 A Section 5.0 system is considered to be in compliance with the requirements of Section
5.1.1 if:
1)
It meets the requirements for avoiding filtration in Section 5.2 below and the
disinfection requirements in Section 5.3 OR
2)
It meets the filtration requirements in Section 5.4 and the disinfection
requirements in Section 5.3.
19
Each Section 5.0 system must be operated by qualified personnel who meet the
requirements of the Rules and Regulations Pertaining to the Certification of
Public Drinking Water Treatment and Transmission and Distribution Operators
promulgated pursuant to the authority set forth in Chapter 23-65 of the General
Laws of Rhode Island, as amended.
5.1.4 Section 5.0 systems that served fewer than 10,000 people beginning January 1, 2002 but
currently serve or will serve at least 10,000 people before January 1, 2005 must comply
with all the requirements listed in this Filtration and Disinfection Document for systems
serving at least 10,000 people as soon as those systems begin serving at least 10,000
people. These systems must also consult with the Director to establish a disinfection
benchmark. If a significant change is made to the disinfection practice these systems must
consult with the Director prior to making such change as stated in Section 5.3.7 (4)
including, but not limited to, 5.3.7(4)(a)(i—iv).
5.1.5 Recycle Provisions: All Section 5.0 systems that employ conventional filtration or direct
filtration treatment and that recycle spent filter backwash water, thickener supernatant, or
liquids from dewatering processes must meet the requirements in 5.1.5 (1) and Section
5.8.4.
1)
Treatment technique requirement. Any system that recycles spent filter backwash
water, thickener supernatant, or liquids from dewatering processes must return
these flows through the processes of a system's existing conventional or direct
filtration system as defined in Section 1.0 or at an alternate location approved by
the Director by June 8, 2004. If capital improvements are required to modify the
recycle location to meet this requirement, all capital improvements must be
completed no later than June 8, 2006.
5.2
Criteria for avoiding filtration:
5.2.1 A PWS that uses a surface water source must meet all of the conditions of Sections 5.2.5
and 5.2.6 and is subject to 5.2.7 of this Section beginning December 30, 1991, unless the
Director has determined in writing that filtration is required.
5.2.2 A PWS that uses a ground water source under the direct influence of surface water must
meet all of the conditions of 5.2.5 and 5.2.6 of this Section and is subject to Section 5.2.7
eighteen (18) months after the Director determines that it is under the direct influence of
surface water, unless the Director has determined in writing that filtration is required.
5.2.3 If the Director determines in writing before December 30, 1991 that filtration is required,
the system must have installed filtration and meet the criteria for filtered systems
specified in these regulations by June 29, 1993.
5.2.4 Within 18 months of the failure of a system using surface water or a ground water source
under the direct influence of surface water to meet any one (1) of the requirements of
20
5.2.5 or 5.2.6 of this Section or after June 29, 1993, whichever is later, the system must
have installed filtration and meet the criteria for filtered systems specified in Section 5.4.
5.2.5 Source Water Quality Conditions:
1)
The fecal coliform concentration must be equal to or less than 20/100ml or the
total coliform concentration must be equal to or less than 100/100 ml (measured
as specified in Appendix 1) in representative samples of the source water
immediately prior to the first or only point of disinfectant application in at least
ninety (90) percent of the samples taken for the six (6) previous months that the
system served water to the public on an ongoing basis.
If a system measures both fecal and total coliforms, the fecal coliform criterion,
but not the total coliform criterion, must be met.
2)
The turbidity level cannot exceed 5 NTU (measured as specified in appendix 1) in
representative samples of the source water immediately prior to the first or only
point of disinfectant application.
5.2.6 Site Specific Conditions:
1)
Compliance
a)
The PWS must meet the requirements of 5.3.5(1) at least eleven (11) of
the twelve (12) previous months that the system served water to the public
on an ongoing basis.
b)
The PWS must meet the requirements of 5.3.5(2) and 5.3.5(3) at all times
the system serves water to the public.
c)
The PWS must meet the requirements of 5.3.5(4) on an ongoing basis.
2)
The PWS must maintain a watershed control program which minimizes the
potential for contamination by Giardia lamblia cysts, Cryptosporidium oocysts
(Cryptosporidum requirements do not apply to systems serving fewer than 10,000
until January 1, 2005), and viruses in the source water. During the onsite
inspection (discussed in 5.2.6 (3)), the adequacy of a watershed control program
will be determined by the Director. The adequacy of a program to limit potential
contamination by Giardia lamblia cysts, Cryptosporidium oocysts, and viruses
must include, but not be limited to, the following measures:
a)
The comprehensiveness of the watershed review;
b)
The effectiveness of the system's program to monitor and control
detrimental activities occurring in the watershed; and
21
c)
The extent to which the water system has maximized land ownership
and/or controlled land use within the watershed. At a minimum, the
watershed control program must:
i)
characterize the watershed hydrology and land ownership;
ii)
identify watershed characteristics and activities which may have an
adverse effect on source water quality; and
iii)
monitor the occurrence of activities which may have an adverse
effect on source water quality.
The PWS must demonstrate through ownership and/or written
agreements with landowners within the watershed that it can
control all human activities which may have an adverse impact on
the microbiological quality of the source water.
The PWS must submit an annual report to the Director that
identifies any special concerns about the watershed and how they
are being handled; describes activities in the watershed that affect
water quality; and projects what adverse activities are expected to
occur in the future and describes how the PWS expects to address
them. Approved watershed protection plans or wellhead protection
plans may be used to the extent that they are applicable.
3)
The PWS must be subject to an annual on-site inspection to assess the watershed
control program and disinfection treatment process.
A report of the on-site inspection summarizing all findings must be prepared
every year. The on-site inspection must indicate to the Director's satisfaction that
the watershed control program and disinfection treatment process are adequately
designed and maintained. The on-site inspection will include but not be limited to:
a)
A review of the effectiveness of the watershed control program;
b)
A review of the physical condition of the source intake and how well it is
protected;
c)
A review of the system's equipment maintenance program to ensure there
is low probability for failure of the disinfection process;
d)
An inspection of the disinfection equipment for physical deterioration;
e)
A review of operating procedures;
22
f)
A review of data records to ensure that all required tests are being
conducted and recorded and disinfection is effectively practiced; and
g)
Identification of any improvements which are needed in the equipment,
system maintenance and operation, or data collection.
4)
The PWS must not have been identified as a source of a waterborne disease
outbreak, or if it has been so identified, the system must have been modified
sufficiently to prevent another such occurrence as determined by the Director.
5)
The PWS must comply with the maximum contaminant level (MCL) for total
coliforms in Section 16.4 (c) at least eleven (11) of the twelve (12) previous
months that the system served water to the public on an ongoing basis, unless the
Director determines that failure to meet this requirement was not caused by a
deficiency in treatment of the source water.
6)
All Section 5.0 systems serving at least 10,000 people and, beginning January 1,
2004, systems serving fewer than 10,000 people, must comply with the
requirements for total trihalomethanes, haloacetic acids (five), bromate, chlorite,
chlorine, chloramines, and chlorine dioxide in Section 7.0.
5.2.7 Treatment Technique Violations:
1)
A system that fails to meet any one (1) of the criteria in Paragraphs 5.2.5 or 5.2.6
of this Section or for which the Director has determined that filtration is required
in writing and fails to install filtration by the date specified is in violation.
2)
A system that has not installed filtration is in violation of a treatment technique
requirement if:
a)
the turbidity level in a representative sample of the source water
immediately prior to the first or only point of disinfection application
exceeds 5 NTU; or
b)
the system is identified as a source of a waterborne disease outbreak.
5.3
Disinfection:
5.3.1 A PWS that uses a surface water source and does not provide filtration treatment must
provide the disinfection treatment specified in 5.3.5 beginning December 30, 1991 unless
the Director determines that filtration is required in writing.
5.3.2 A PWS that uses a ground water source under the direct influence of surface water and
does not provide filtration treatment must provide disinfection treatment specified in
Paragraph 5.3.5 eighteen (18) months after the Director determines that the ground water
23
source is under the influence of surface water, unless the Director has determined that
filtration is required in writing.
5.3.3 If the Director has determined that filtration is required, the system must comply with any
interim disinfection requirements the Director deems necessary before filtration is
installed. A system that uses a surface water source that provides filtration treatment must
provide the disinfection treatment specified in 5.3.6 beginning June 29, 1993 or
beginning when filtration is installed, whichever is later.
5.3.4 A system that uses a ground water source under the direct influence of surface water and
provides filtration treatment must provide disinfection treatment as specified in Section
5.3.6, beginning when filtration is installed. Failure to meet any requirement of this
Section is a treatment technique violation.
5.3.5 Disinfection requirements for PWSs that do not provide filtration:
1)
The disinfection treatment must be sufficient to ensure at least 99.9 percent (3-
log) inactivation of Giardia lamblia cysts and 99.99 percent (4-log) inactivation
of viruses, every day the system serves water to the public, except any one (1) day
each month. Each day a system serves water to the public, the PWS must
calculate the CT value(s) from the system's treatment parameters, using the
procedure specified in Section 5.6.1 (3) and 5.6.1 (4) and determine whether this
value is sufficient to achieve the specified inactivation rates for Giardia lamblia
cysts and viruses.
If a system uses a disinfectant other than chlorine, the system may demonstrate to
the Director, through the use of a protocol approved by the Director for on-site
disinfection challenge studies or other information satisfactory to the Director,
that the CT99.9 values other than those specified in Tables 2.1 and 3.1 or other
operational parameters are adequate to demonstrate that the system is achieving
minimum inactivation rates required by Paragraph 5.3.5(1) of this Section.
2)
The disinfection system must have either:
a)
redundant components, including an auxiliary power supply with
automatic start-up and alarm to ensure that disinfectant application is
maintained continuously while water is being delivered to the distribution
system; or
b)
automatic shut-off of delivery of water to the distribution system
whenever there is less than 0.2 mg/L of residual disinfectant concentration
in the water.
If the Director determines that automatic shut-off would cause
unreasonable risk to health or interfere with fire protection, the system
must comply with Paragraph 5.3.5 (2)(a) of this Section.
24
3)
The residual disinfectant concentration in the water entering the distribution
system measured as specified in Section 5.5 cannot be less than 0.2 mg/L for
more than 4 hours.
4)
The residual disinfectant concentration in the distribution system, measured as
total chlorine, combined chlorine or chlorine dioxide as specified in Section 5.5
cannot be undetectable in more than five (5) percent of the samples each month,
for any two (2) consecutive months that the system serves water to the public.
Water in the distribution system with a heterotrophic bacteria concentration less
than or equal to 500/ml measured as heterotrophic plate count (HPC) as specified
in Section 5.5 is deemed to have a detectable disinfectant residual for purposes of
determining compliance with this requirement. Thus, the value “V” in the
following formula cannot exceed 5 percent in one (1) month for any two (2)
consecutive months:
V =
c+d+e x 100
a+b
where:
a=
number of instances where the residual disinfectant concentration is
measured;
b=
number of instances where the residual disinfectant concentration is not
measured but the heterotrophic bacteria plate count (HPC) is measured;
c=
number of instances where the residual disinfectant concentration is
measured but not detected and no HPC is measured;
d=
number of instances where the residual disinfectant concentration is
measured but not detected and where the HPC is >500/ml; and
e=
number of instances where the residual disinfectant concentration is not
measured and HPC is >500/ml.
5.3.6 Disinfection requirements for PWSs which provide filtration:
Each PWS that provides filtration treatment must provide disinfection treatment as
follows:
1)
The disinfection treatment must be sufficient to ensure that the total treatment
processes of that system achieve at least 99.9 percent (3-log) inactivation and/or
removal of Giardia lamblia cysts and at least 99.99 percent (4-log) inactivation
and/or removal of viruses as determined by the Director.
25
2)
The residual disinfectant concentration in the water entering the distribution
system measured as specified in Section 5.5 cannot be less than 0.2 mg/L for
more than four (4) hours.
3)
The residual disinfectant concentration in the distribution system, measured as
total chlorine, combined chlorine or chlorine dioxide, as specified in Section 5.5
cannot be undetectable in more than five (5) percent of the samples each month,
for any two (2) consecutive months that the system serves water to the public.
Water in the distribution system with a heterotrophic bacteria concentration less
than or equal to 500/ml, measured as heterotrophic plate count (HPC) as specified
in Section 5.5, is deemed to have a detectable disinfectant residual for purposes of
determining compliance with this requirement. Thus the value of “V” cannot
exceed five (5) percent in one (1) month for any two (2) consecutive months. [See
formula in 5.3.5 (4)].
5.3.7 Disinfection Profiling and Benchmarking:
1)
A Section 5.0 community or non-transient, non-community water system that
serves fewer than 10,000 people must develop a disinfection profile, a graphical
representation of a system's level of Giardia lamblia or virus inactivation
measured during the course of a year, under the provisions of Paragraphs 5.3.7
(1)(a), (b), (c), (2) and (3) below unless the Director determines that it is
unnecessary. At the Director’s discretion, a Section 5.0 system that serves at least
10,000 people may also be required to develop a disinfection profile. If the
Director requires a system serving at least 10,000 people to develop a profile, the
Director shall specify procedures for developing that profile.
a)
The Director may only determine that a system's profile is unnecessary if a
system's TTHM and HAA5 levels are below 0.064 mg/L and 0.048 mg/L,
respectively.
b)
If TTHM and HAA5 levels are ≥ 0.064mg/L or ≥ 0.048 mg/L,
respectively, the system must comply with Paragraph (2)(a) of this
Section.
c)
To determine these levels, TTHM and HAA5 samples must be collected
after January 1, 1998, during the month with the warmest water
temperature, and at the point of maximum residence time in the
distribution system.
2)
Disinfection Profile Criteria
a)
Any Section 5.0 system serving fewer than 10,000 people that meets the
criteria in Paragraph (1)(b) of this Section must develop a disinfection
profile of its disinfection practice for a period of up to one (1) year. The
26
Director may approve the use of a more representative data set for
disinfection profiling than the data set required under 5.3.7 (2)(a)(i ) and
(3).
(i)
Systems must collect data for several parameters from the plant,
specified in Paragraphs (i)(A) through (D) below, once per week
on the same calendar day over twelve (12) consecutive calendar
months to determine the total logs of inactivation for each day of
operation, based on the CT99.9 values in Tables 1.1-1.6, 2.1 and
3.1 of these regulations, as appropriate, through the entire
treatment plant. Systems serving between 500 and 9,999 persons
must begin to collect data no later than July 1, 2003. Systems
serving fewer than 500 persons must begin to collect data no later
than January 1, 2004. The system must monitor the parameters
listed in Paragraphs (i)(A) through (D) below, necessary to
determine the total inactivation ratio, using analytical methods in
Section 5.5 and Appendix 1.
(A)
The temperature of the disinfected water at each residual
disinfectant concentration sampling point during peak
hourly flow;
(B)
If the system uses chlorine, the pH of the disinfected water
at each chlorine residual disinfectant concentration
sampling point during peak hourly flow;
(C)
The disinfectant contact time(s) (“T”) during peak hourly
flow; and
(D)
The residual disinfectant concentration(s) (“C”) of the
water before or at the first customer and prior to each
additional point of disinfection during peak hourly flow.
(ii)
Systems must use this data to calculate the inactivation ratios as
discussed in Sections 5.6.1 (4)(a) and (b). As a minimum, the
system with a single point of disinfectant application prior to
entrance to the distribution system must calculate the inactivation
ratio as discussed in Section 5.6.1 (4)(a). A system with more than
one (1) point of disinfectant application or measures disinfectant
residuals at more than one (1) location must calculate the
inactivation ratio as discussed in Section 5.6.1 (4)(b) for each
disinfection segment.
(iii)
Weekly log inactivations are calculated by multiplying the
CTcalc/CT99.9 ratio across the entire treatment train by 3.
27
(iv)
Systems must use these weekly log inactivations to develop a
disinfection profile as specified in Paragraph (3)(a) of this Section.
3)
Developing a Disinfection Profile
a)
Each log inactivation serves as a data point in your disinfection profile.
Systems serving fewer than 10,000 people will have obtained fifty-two
(52) measurements (one (1) for every week of the year). The system and
the Director will evaluate how microbial inactivation varied over the
course of the year by looking at all 52 measurements (the Disinfection
Profile). Systems must retain the Disinfection Profile data in graphic form,
such as a spreadsheet, which must be available for review by the Director
as part of a sanitary survey. Systems must use this data to calculate a
benchmark if the system is considering changes to disinfection practices.
b)
A system that uses chloramines, ozone or chlorine dioxide for primary
disinfection must also calculate the logs of inactivation for viruses and
develop an additional disinfection profile for viruses using a method
approved by the Director.
4)
Disinfection Benchmark
a)
A Section 5.0 system serving less than 10,000 people that is required to
develop a disinfection profile under the provisions of Section 5.3.7 (1),
must develop a Disinfection Benchmark as described in Paragraph 5.3.7
(4)(c) and (d) and provide the benchmark to the Director if the system
decides to make a significant change to its disinfection practice. A Section
5.0 system serving at least 10,000 people that is required to develop a
disinfection profile under the provisions of Section 5.3.7 (1), must develop
a Disinfection Benchmark using procedures specified by the Director if
the system decides to make a significant change to its disinfection
practice. All systems must consult with the Director for approval prior to
making such changes. Significant changes to disinfection practice are:
(i)
Changes to the point of disinfection;
(ii)
Changes to the disinfectant(s) used in the treatment plant;
(iii)
Changes to the disinfection process; and
(iv)
Any other modification identified by the Director.
b)
Systems must submit the following information to the State as part of the
consultation and approval process:
(i)
A description of the proposed change;
28
(ii)
The disinfection profile for Giardia lamblia (and, if necessary,
viruses) and disinfection benchmark;
(iii)
An analysis of how the proposed change will affect the current
levels of disinfection; and
(iv)
Any additional information requested by the Director.
c)
Any system that is modifying its disinfection practice must calculate its
disinfection benchmark using the procedures specified in Paragraphs (c)(i)
through (ii) below.
(i)
For one (1) year of profiling data collected weekly and calculated
under Paragraphs (2) and (3) of this Section, the system must
determine the lowest average monthly Giardia lamblia inactivation
for one (1) year. The system must determine the average Giardia
lamblia inactivation for each calendar month by dividing the sum
of all Giardia lamblia inactivations for that month by the number
of values calculated for that month.
(ii)
The disinfection benchmark value is the lowest monthly average
value out of twelve (12) values of Giardia lamblia inactivation in
one (1) year of profiling data.
d)
A system that uses chloramines, ozone or chlorine dioxide for primary
disinfection must calculate the disinfection benchmark from the data
collected for viruses to develop the disinfection profile in addition to the
Giardia lamblia disinfection benchmark calculated under Paragraph (4)(c)
of this Section. This viral benchmark must be approved by the Director
and must be calculated in the same manner used to calculate the Giardia
lamblia disinfection benchmark in Paragraph (4)(c) of this Section.
5.4
Filtration:
5.4.1 A Section 5.0 system that does not meet all of the criteria in Section 5.2 for avoiding
filtration, must provide treatment consisting of both disinfection, as specified in 5.3.6 and
filtration treatment which complies with the requirements of Section 5.4 by June 29,
1993, or within eighteen (18) months of the failure to meet any one (1) of the criteria for
avoiding filtration, whichever is later. Failure to meet any requirement of this Section by
the date specified in Section 5.4.1, shall constitute a treatment technique violation.
5.4.2 Conventional filtration treatment or direct filtration:
Table 5.1 – Summary of filtered effluent turbidity requirements for systems that use conventional or direct filtration
System Size
Effective Date
At least 95% of Turbidity
Maximum Turbidity
29
Measurements (NTU)
Measurement (NTU)
Systems serving at
least 10,000 people
January 1, 2002
≤ 0.3
1
Before
January 1, 2005
≤ 0.5 or as determined by the
Director
5
Systems serving fewer
than 10,000 people
Beginning January 1,
2005
≤ 0.3
1
1)
Systems serving at least 10,000 people that use conventional filtration or direct
filtration that do not meet all of the criteria listed in Section 5.2 for avoiding
filtration must meet the turbidity requirements listed in Table 5.1 and in (a), (b)
and (c) below:
a)
The turbidity level of representative samples of a system's filtered water
must be less than or equal to 0.3 NTU in at least ninety-five (95) percent
of the measurements taken each month, measured as specified in Sections
5.5 and 5.7. Monthly reporting must be completed according to Section
5.8.
b)
The turbidity level of representative samples of a system's filtered water
must at no time exceed one (1) NTU, measured as specified in Sections
5.5 and 5.7. Monthly reporting must be completed according to Section
5.8.
c)
A system that uses lime softening may acidify representative combined
filter effluent turbidity samples prior to analysis using a protocol approved
by the Director.
2)
Beginning January 1, 2005, systems serving fewer than 10,000 that use
conventional or direct filtration and do not meet all of the criteria listed in Section
5.2 for avoiding filtration, must meet the turbidity requirements listed in 5.4.2 (1).
Until then, they must meet the turbidity requirements listed in (a), (b) and (c)
below (refer to Table 5.1 for a summary of the requirements):
a)
The turbidity level of representative samples of a system's filtered water
must be less than or equal to 0.5 NTU in at least ninety-five (95) percent
of the measurements taken each month, measured as specified in Section
5.5. However, if the Director determines that the system is capable of
achieving at least 99.9 percent removal and/or inactivation of Giardia
lamblia cysts at some turbidity level higher than 0.5 NTU in at least
ninety-five (95) percent of the measurements taken each month, the
Director may substitute this higher turbidity limit for that system.
30
b)
In no case will a turbidity limit that allows more than one (1) NTU in
more than five (5) percent of the samples taken each month, measured as
specified in Section 5.5 be approved.
c)
The turbidity level of representative samples of a system's filtered water
must at no time exceed five (5) NTU measured as specified in Section 5.5.
5.4.3 Slow Sand Filtration:
1)
For systems using slow sand filtration, the turbidity level of representative
samples of a system's filtered water must be less than or equal to one (1) NTU in
at least ninety-five (95) percent of the measurements taken each month, measured
as specified in Section 5.5.
2)
The turbidity level of representative samples of a system's filtered water must at
no time exceed five (5) NTU measured as specified in Section 5.5.
5.4.4 Diatomaceous Earth Filtration:
1)
For systems using diatomaceous earth filtration, the turbidity level of
representative samples of a system's filtered water must be less than or equal to
one (1) NTU in at least ninety-five (95) percent of the measurements taken each
month, measured as specified in Section 5.5.
2)
The turbidity level of representative samples of a system's filtered water must at
no time exceed five (5) NTU, measured as specified in Section 5.5.
5.4.5 Other Filtration Technologies:
A PWS may use a filtration technology not listed in Sections 5.4.2, 5.4.3 or 5.4.4, if it
demonstrates to the Director, using pilot plant studies or other means, that the alternative
filtration technology, in combination with disinfection treatment that meets the
requirements of Section 5.3.6 and 5.3.7, consistently achieves ninety-nine (99) percent
removal of Cryptosporidium oocysts (systems serving fewer than 10,000 are not required
to comply with the Cryptosporidium oocysts requirement until January 1, 2005), 99.9
percent removal and/or inactivation of Giardia lamblia cysts and 99.99 percent removal
and/or inactivation of viruses. Upon completion of the demonstration, the Director will
determine the 95th percentile turbidity value (not to exceed one (1) NTU) and the
maximum turbidity value (not to exceed five (5) NTU) based on the demonstration.
5.5
Analytical Monitoring Requirements
5.5.1 Only the analytical method(s) specified in this Section, or otherwise approved by the
Director, may be used to demonstrate compliance with the requirements of Sections 5.2,
5.3, or 5.4.
31
Measurements for pH, temperature, turbidity and residual disinfectant concentrations
must be conducted by a party approved by the Director.
Measurements for total coliforms, fecal coliforms and HPC must be conducted by a
laboratory certified by the Director or EPA to do such analysis.
5.5.2 The following procedures shall be performed in accordance with the methods listed.
1)
Fecal Coliform/E. Coli Concentration Method, as set forth in Appendix 1.
2)
Total Coliform Concentration, as set forth in Appendix 1.
3)
Heterotrophic Plate Count, as set forth in Appendix 1.
4)
Turbidity, as set forth in Appendix 1.
5)
Residual Disinfectant Concentration, as set forth in Appendix 1.
6)
Temperature, Method 212, pp126-127, as set forth in Appendix 1.
7)
pH Method 423 (pH value) pp 429-437, as set forth in Appendix 1.
8)
Minimal Medium ONPG-MUG method for simultaneous enumeration of total
coliform and E. Coli as set forth in Appendix 1.
9)
Indigo Method for determination of Ozone in water, as set forth in Appendix 1.
5.6
Monitoring Requirements for Systems That Do Not Provide Filtration
5.6.1 A PWS that uses a surface water source and does not provide filtration treatment must
begin monitoring, as specified in this Section beginning December 31, 1990, unless the
Director has determined that filtration is required in writing, in which case the Director
may specify alternative monitoring requirements, until filtration is in place.
A PWS that uses a ground water source under the direct influence of surface water and
does not provide filtration treatment must begin monitoring as specified in this Section 6
months after the Director determines that the ground water source is under the direct
influence of surface water, unless the Director has determined that filtration is required in
writing.
1)
Fecal coliform or total coliform density measurements, as required by Section
5.2.5, must be performed on representative source water samples immediately
prior to the first or only point of disinfectant application. The system must sample
for fecal or total coliforms at the following minimum frequency each week the
system serves water to the public:
32
System Size (Persons Served)
Samples/Week*
<500
1
501 to 3,300
2
3,301 to 10,000
3
10,001 to 25,000
4
>25,000
5
*Samples must be taken on separate days
Also, one (1) fecal or total coliform density measurement must be made every day
the system serves water to the public and the turbidity of the source water exceeds
1 NTU (these samples count toward the weekly coliform sampling requirement,)
unless the Director determines that the system for logistical reasons outside the
system's control cannot have the sample analyzed within thirty (30) hours of
collection.
2)
Turbidity measurements as required by Section 5.2.5(2) must be performed on
representative grab samples of source water immediately prior to the first or only
point of disinfectant application every four (4) hours (or more frequently) that the
system serves water to the public. A PWS may substitute continuous turbidity
monitoring for grab sample monitoring if it validates the continuous measurement
for accuracy on a regular basis using a protocol approved by the Director.
3)
The total inactivation ratio for each day that the system is in operation must be
determined based on the CT99.9 values in Tables 1.1-1.6, 2.1 and 3.1 of this
Section, as appropriate. The parameters necessary to determine the total
inactivation ratio must be monitored as follows:
a)
The temperature of the disinfected water must be measured at least once
per day at each residual disinfectant concentration sampling point.
b)
If the system uses chlorine, the pH of the disinfected water must be
measured at least once per day at each chlorine residual disinfectant
concentration sampling point.
c)
The disinfectant contact time(s) (“T”) must be determined for each day
during peak hourly flow.
d)
The residual disinfectant concentration(s) (“C”) of the water before or at
the first customer must be measured each day during peak hourly flow.
e)
If a system uses a disinfectant other than chlorine, the system may
demonstrate to the Director, through the use of a protocol approved by the
Director, for on-site disinfection challenge studies or other information
satisfactory to the Director that CT99.9 values other than those specified
in Tables 2.1 and 3.1 in this Section or other operational parameters are
33
adequate to demonstrate that the system is achieving the minimum
inactivation rates required by Section 5.3.5(1).
4)
The total inactivation ratio must be calculated as follows:
a)
If the system uses only one (1) point of disinfectant application, the
system may determine the total inactivation ratio based on either of the
following two (2) methods:
i)
One (1) inactivation ratio (CTcalc/CT99.9) is determined before or at
the first customer during peak hourly flow and if the
CTcalc/CT99.9>1.0, the 99.9 percent Giardia lamblia inactivation
requirement has been achieved; OR
ii)
Successive CTcalc/CT99.9 values representing sequential inactivation
ratios are determined between the point of disinfectant application
and a point before or at the first customer during peak hourly flow.
Under this alternative, the following method must be used to
calculate the total inactivation ratio:
(A)
Determine (CTcalc/CT99.9) for each sequence
(B)
Add the (CTcalc/CT99.9) values together
(the sum of all CTcalc/CT99.9)
(C)
If the sum of (CTcalc/CT99.9) >1.0
Then the 99.9 percent Giardia lamblia inactivation
requirement has been achieved.
b)
If the system uses more than one (1) point of disinfectant application
before or at the first customer, the system must determine the CT value of
each disinfection sequence immediately prior to the next point of
disinfectant application during peak hourly flow. The CTcalc/CT99.9value of
each sequence and the sum of CTcalc/CT99.9must be calculated using the
method in Section 5.6.1(4)(a)(ii) of this Section to determine if the system
is in compliance with Section 5.3.5.
c)
Although not required, the total percent inactivation for a system with one
(1) or more points of residual disinfectant concentration monitoring may
be calculated by solving the following equation:
Percent inactivation = 100-(100/10Z)
where Z = 3 x the sum of (CTcalc/CT99.9)
34
5)
The residual disinfectant concentration of the water entering the distribution
system must be monitored continuously, and the lowest value must be recorded
each day. In the event of system monitoring failure, grab sampling may be
conducted every four (4) hours, for no more than five (5) working days.
Systems serving 3,300 or fewer persons may take grab samples in lieu of
continuous monitoring on an ongoing basis at the frequencies prescribed below:
System Size by Population
Samples/day*
<501
1
501 to 1,000
2
1,001 to 2,500
3
2,501 to 3,300
4
*The day's samples cannot be taken at the same time. The sampling intervals are subject to the Director's review and
approval
If at any time the residual disinfectant concentration falls below 0.2 mg/L in a
system using grab sampling in lieu of continuous monitoring, the system must
take a grab sample every four (4) hours until the residual concentration is equal to
or greater than 0.2 mg/L.
6)
The residual disinfectant concentration must be measured at least at the same
points in the distribution system and at the same time as total coliforms are
sampled, as specified in Section16.4, however, the Director may allow a PWS
which uses both a surface water source or a ground water source under direct
influence of surface water, and a ground water source to take disinfectant residual
samples at points other than the total coliform sampling points, if the Director
determines that such points are more representative of treated (disinfected) water
quality within the distribution system.
Heterotrophic bacteria, measured as HPC as specified in Section 5.5.2, may be
measured in lieu of residual disinfectant concentration.
5.7
Monitoring Requirements for Systems Using Filtration Treatment
5.7.1 A PWS that uses a surface water source or a ground water source under the influence of
surface water and provides filtration treatment must monitor in accordance with this
Section, beginning June 29, 1993, or when filtration is installed, whichever is later.
1)
Turbidity
a)
Representative Filtered Effluent Turbidity Requirements
i)
Turbidity measurements as required by Section 5.4 must be
performed on representative samples of the systems filtered water
35
every four (4) hours (or more frequently) that the system serves
water to the public. A PWS may substitute continuous turbidity
monitoring for grab sample monitoring if it validates the
continuous measurement for accuracy on a regular basis, using a
protocol approved by the Director.
ii)
For any systems using slow sand filtration or filtration treatment
other than conventional treatment, direct filtration or diatomaceous
earth filtration, the Director may reduce the sampling frequency to
once per day if it determines that less frequent monitoring is
sufficient to indicate effective filtration performance.
iii)
For systems serving 500 or fewer persons, the Director may reduce
the turbidity sampling frequency to once per day, regardless of the
type of filtration treatment used, if the Director determines that less
frequent monitoring is sufficient to indicate effective filtration
performance.
b)
Individual Filter Turbidity Requirements
i)
Section 5.0 systems that use conventional or direct filtration must
conduct continuous monitoring of turbidity for each individual
filter in the system, beginning January 1, 2002 for systems serving
at least 10,000 people and beginning January 1, 2005 for systems
serving fewer than 10,000 people. The following requirements
apply to continuous turbidity monitoring:
(A)
Continuous monitoring must be conducted using an
approved method in Section 5.5;
(B)
Calibration of turbidimeters must be conducted using
procedures specified by the manufacturer;
(C)
Results of turbidity monitoring must be recorded at least
every fifteen (15) minutes; and
(D)
Monthly reporting must be completed and records must be
maintained according to Section 5.8.
(ii)
If there is a failure in the continuous turbidity monitoring
equipment, the system must conduct grab sampling every four (4)
hours in lieu of continuous monitoring until the turbidimeter is
back on-line. Systems serving at least 10,000 people have no more
than five (5) working days, following equipment failure, and
systems serving fewer than 10,000 people have no more than
36
fourteen (14) total days, following equipment failure, to resume
continuous monitoring before a violation is incurred.
(iii)
For systems serving fewer than 10,000 people, systems that only
consist of two (2) or fewer filters may conduct continuous
monitoring of combined filter effluent turbidity in lieu of
individual filter effluent turbidity monitoring. Continuous
monitoring must meet the same requirements set forth in
Paragraphs (b)(i) and (b)(ii) of this Section.
2)
The residual disinfectant concentration of the water entering the distribution
system must be monitored as indicated in sections 5.6.1(5) and 5.6.1(6).
5.8
Reporting and Record Keeping Requirements
5.8.1 A PWS that uses a surface water source and does not provide filtration treatment must
report the following information monthly to the Director beginning December 31, 1990
unless the Director has determined that filtration is required in writing in which case the
Director may specify alternate reporting requirements as appropriate until filtration is in
place.
A PWS that uses a ground water source under the direct influence of surface water and
does not provide filtration treatment must report monthly to the Director, the following
information beginning no later than six (6) months after the Director determines that the
ground water source is under the direct influence of surface water.
1)
Source water quality information must be reported to the Director within ten (10)
days after the end of each month the system serves water to the public.
Information that must be reported:
a)
The cumulative number of months for which results are reported.
b)
The number of fecal and/or total coliform samples, whichever are
analyzed during the month (if a system monitors for both, only fecal
coliforms must be reported), the dates of sample collection and the dates
when the turbidity level exceeded one (1) NTU.
c)
The number of samples during the month that had equal to or less than
20/100 ml fecal coliforms and/or equal to or less than 100/100 ml total
coliforms, whichever are analyzed.
d)
The cumulative number of fecal or total coliform samples, whichever are
analyzed during the previous six (6) months the system served water to the
public.
37
e)
The cumulative number of samples that had equal to or less than 20/100
ml fecal coliforms or equal to or less than 100/100 ml total coliforms,
whichever are analyzed during the previous six (6) months the system
served water to the public.
f)
The percentage of samples that had equal to or less than 20/100 ml fecal
coliforms or equal to or less than 100/100 ml total coliforms, whichever
are analyzed during the previous six (6) months the system served water to
the public.
g)
The maximum turbidity level measured during the month, the date(s) of
occurrence for any measurement(s) which exceeded five (5) NTU, and the
date(s) the occurrence(s) was reported to the Director.
h)
For the first twelve (12) months of record-keeping, the dates and
cumulative number of events during which the turbidity exceeded five (5)
NTU and after one (1) year of record keeping for turbidity measurements,
the dates and cumulative number of events during which the turbidity
exceeded five (5) NTU in the previous twelve (12) months the system
served water to the public.
i)
For the first 120 months of record-keeping, the dates and cumulative
number of events during which the turbidity exceeded five (5) NTU and
after ten (10) years of record keeping for turbidity measurements, the dates
and cumulative number of events during which the turbidity exceeded five
(5) NTU in the previous 120 months they system service water to the
public.
2)
Disinfection information must be reported to the Director within ten (10) days
after the end of each month the system serves water to the public. Information that
must be reported:
a)
For each day, the lowest measurement of residual disinfectant
concentration in mg/L in water entering the distribution system.
b)
The date and duration of each period when the residual disinfectant
concentration in water entering the distribution system fell below 0.2
mg/L and when the Director was notified of the occurrence.
c)
The daily residual disinfectant concentration(s) (in mg/L) and disinfectant
contact time(s) (in minutes) used for calculating the CT value(s).
d)
If chlorine is used, the daily measurement(s) of pH of disinfected water
following each point of chlorine disinfection.
38
e)
The daily measurement(s) of water temperature in degrees centigrade
following each point of disinfection.
f)
The daily CTcalc and CTcalc/CT99.9 values for each disinfectant
measurement or sequence and the sum of all CTcalc/ CT99.9 values
(CTcalc/CT99.9) before or at the first customer.
g)
The daily determination of whether disinfection achieves adequate
Giardia cyst and virus inactivation, i.e. whether (CTcalc/ CT99.9) is at least
1.0 or where disinfectants other than chlorine are used, other indicator
conditions that the Director determines are appropriate, are met.
h)
The following information on the samples taken in the distribution system
in conjunction with total coliform monitoring specified in Section 5.3.
i)
number of instances where the residual disinfectant concentration is
measured;
ii)
number of instances where the residual disinfectant concentration is not
measured but HPC is measured;
iii)
number of instances where the residual disinfectant concentration is
measured, but not detected and no HPC is measured;
iv)
number of instances where the residual disinfectant concentration is
detected and where HPC is >500/ml;
v)
number of instances where the residual disinfectant concentration is not
measured and HPC is >500/ml;
vi)
for the current and previous month the system served water to the public,
the value of “V”, as defined in Section 5.3.5.
i)
A system need not report the data listed in Section 5.8.1(2)(a) and (c)-(f) if
all data listed in 5.8.1(2) (a)-(h) remain on file at the system and the
Director determines that:
i)
The system has submitted to the Director all the information required for
at least twelve (12) months; and
ii)
The Director has determined that the system is not required to provide
filtration treatment.
3)
No later than October 10 of each year, each system must provide to the Director a
report which summarizes its compliance with all watershed control program
requirements specified in 5.2.6(2).
39
4)
A report on the on-site inspection conducted during that year as specified in
5.2.6(3).
5)
a)
Each system, upon discovering that a waterborne disease outbreak
potentially
attributable to that water system has occurred, must report that occurrence
to
the Director as soon as possible, but no later than the end of the next
business
day.
b)
If at any time the turbidity exceeds five (5) NTU, the system must consult
with the Director as soon as practical, but no later than twenty-four (24)
hours after the exceedance is known, in accordance with the public
notification requirements under Section 16.8 (3).
c)
If at any time the residual falls below 0.2 mg/L in the water entering the
distribution system, the system must notify the Director as soon as
possible, but no later than by the end of the next business day. The system
must notify the Director by the end of the next business day whether or not
the residual was restored to at least 0.2 mg/L within four (4) hours.
5.8.2 Section 5.0 systems that provide filtration treatment must report monthly to the Director
the following information in 5.8.2 (1) through (4) unless otherwise stated.
1)
Turbidity requirements: Turbidity measurements as required by Section 5.4 and
5.7 must be reported within ten (10) days after the end of each month the system
serves water to the public. Information that must be reported includes:
a)
The total number of filtered water turbidity measurements taken during the
month.
b)
The number and percentage of filtered water turbidity measurements taken
during the month which are less than or equal to the turbidity limits
specified in section 5.4 for the filtration technology being used.
c)
The date and value of any turbidity measurements taken during the month
which exceed the maximum allowable turbidity specified in 5.4.
2)
Individual filter effluent reporting requirements (conventional and direct filtration
systems only):
Systems must maintain the results of individual filter monitoring taken under
Section 5.7.1 (1)(b) for at least three (3) years. Systems must report that they have
conducted individual filter turbidity monitoring under Section 5.7.1 (1)(b) within
ten (10) days after the end of each month the system serves water to the public.
40
Systems must report individual filter turbidity measurement results taken under
Section 5.7.1 (1)(b) within ten (10) days after the end of each month the system
serves water to the public only if measurements demonstrate one (1) or more of
the conditions in Paragraph (4) of this Section.
3)
Disinfection information must be reported to the Director within ten (10) days
after the end of each month and must include all items specified in sections
5.8.1(2)(a), (b), (h), and 5.8.1(5).
a)
Disinfection Profiling: By July 1, 2003, systems serving 500-9,999 people
and by January 1, 2004, systems serving fewer than 500 people must
report the results of optional monitoring which shows TTHM levels below
0.064 mg/L and HAA5 levels below 0.048 mg/L (only if the system
wishes to forgo profiling) or systems must report that they have begun
disinfection profiling. If profiling is required by the Director for systems
serving at least 10,000 people, the necessary reporting requirements will
be specified by the Director. Records of the profile, if required, must be
kept indefinitely including raw data and analysis and made available to the
Director as part of a sanitary survey.
b)
Disinfection Benchmarking: If a system serving fewer than 10,000 people
was required to produce a disinfection profile and is considering a
significant change to its disinfection practices, they must report a
description of the proposed change in disinfection, a disinfection profile
for Giardia lamblia (and, if necessary, viruses) and disinfection
benchmark, and an analysis of how the proposed change will affect the
current levels of disinfection. If benchmarking is required by the Director
for systems serving at least 10,000 people, the necessary reporting
requirements will be specified by the Director. Records of the benchmark
must be kept indefinitely including raw data and analysis and made
available to the Director as part of a sanitary survey.
4)
Individual filter effluent follow-up actions: For all systems, reporting to the
Director is required by the 10th of the following month for exceedance listed in
Paragraphs (a) through (d) of this Section unless otherwise stated. Systems that
use lime softening may apply to the Director for alternative exceedance levels for
the levels specified in (a) through (d) if they can demonstrate that higher turbidity
levels in individual filters are due to lime carryover only and not due to degraded
filter performance.
a)
If the individual filter effluent turbidity (or for systems serving fewer than
10,000, the turbidity of combined filter effluent (CFE) for systems with
two (2) filters that monitor CFE in lieu of individual filters) exceeded 1.0
NTU in two (2) consecutive measurements taken fifteen (15) minutes
apart, the system must report the filter number, the turbidity measurement,
the date(s) on which the exceedance occurred and the cause (if known) for
41
the exceedance. In addition, systems serving at least 10,000 people must
either produce a filter profile for the filter within seven (7) days of the
exceedance (if the system is not able to identify an obvious reason for the
abnormal filter performance) and report that the profile has been produced
or report the obvious reason for the exceedance.
b)
For systems serving at least 10,000 people, if the individual filter effluent
turbidity exceeded 0.5 NTU in two (2) consecutive measurements taken
fifteen (15) minutes apart at the end of the first four (4) hours of
continuous filter operation after the filter has been backwashed or
otherwise taken offline, the system must report the filter number, the
turbidity and the date(s) on which the exceedance occurred. In addition,
the system must either produce a filter profile for the filter within seven
(7) days of the exceedance (if the system is not able to identify an obvious
reason for the abnormal filter performance) and report that the profile has
been produced or report the obvious reason for the exceedance.
c)
If the individual filter effluent turbidity (or for systems serving fewer than
10,000, the CFE turbidity of systems with two (2) filters that monitor CFE
in lieu of individual filters) exceeded 1.0 NTU in two (2) consecutive 15-
minute readings for three (3) consecutive months, the system must report
the filter number, the turbidity measurement, and the date(s) on which the
exceedance occurred. In addition, the system must conduct a self-
assessment of the filter(s) within fourteen (14) days of the exceedance
unless a CPE as specified in (4)(d) of this Section was required. Systems
with two (2) filters that monitor combined filter effluent in lieu of
individual filters must conduct a self-assessment on both filters. The
system must report the date the filter self-assessment was triggered and the
date it was completed by the 10th of the following month or fourteen (14)
days after the self-assessment was triggered only if the self-assessment
was triggered during the last four (4) days of the month. The self
assessment must consist of at least the following components:
i)
assessment of filter performance;
ii)
development of a filter profile;
iii)
identification
and
prioritization
of
factors
limiting
filter
performance;
iv)
assessment of the applicability of corrections; and
v)
preparation of a filter self-assessment report.
d)
If the individual filter effluent turbidity (or for systems serving fewer than
10,000, the CFE for systems with two (2) filters that monitor combined
42
filter effluent in lieu of individual filters) exceeded 2.0 NTU in two (2)
consecutive recordings fifteen (15) minutes apart at the same filter for two
(2) consecutive months, the system must report the filter number, the
turbidity measurement, and the date(s) on which the exceedance occurred.
In addition, the system must arrange to have a comprehensive
performance evaluation (CPE) conducted by the Director or a third party
approved by Director not later than thirty (30) days (sixty (60) days for
systems serving fewer than 10,000) following the day the filter exceeded
2.0 NTU in two (2) consecutive measurements for the second straight
month. If a CPE has been completed by the Director or a third party
approved by the Director within the 12 prior months or the system and
Director are jointly participating in an ongoing Comprehensive Technical
Assistance (CTA) project at the system, a new CPE is not required. If
conducted, a CPE must be completed and submitted to the Director no
later than ninety (90) days (120 days for systems serving fewer than
10,000) following the day the filter exceeded 2.0 NTU in two (2)
consecutive measurements for the second straight month. The system must
report by the 10th of the following month that a CPE was required and the
date it was triggered.
5.8.3 For all filtration technologies, a Section 5.0 system that exceeds the maximum turbidity
as specified in Section 5.4 must inform the Director within twenty-four (24) hours.
5.8.4 Recycle Provisions:
1)
Reporting. A system must notify the Director in writing by December 8, 2003, if
the system recycles spent filter backwash water, thickener supernatant, or liquids
from dewatering processes. This notification must include, at a minimum, the
information specified in 5.8.4 (1)(a) and (b).
a)
A plant schematic showing the origin of all flows which are recycled
(including, but not limited to, spent filter backwash water, thickener
supernatant and liquids from dewatering processes), the hydraulic
conveyance used to transport them, and the location where they are re-
introduced back into the treatment plant.
b)
Typical recycle flow in gallons per minute (gpm) the highest observed
plant flow experienced in the previous year (gpm), design flow for the
treatment plant (gpm), and Director-approved operating capacity for the
plant where the Director has made such determinations.
2)
Recordkeeping. The system must collect and retain on file recycle flow
information specified in 5.8.4 (2)(a) through (f) for review and evaluation by the
Director beginning June 8, 2004.
43
a)
Copy of the recycle notification and information submitted to the Director
under 5.8.4 (1).
b)
List of all recycle flows and the frequency with which they are returned.
c)
Average and maximum backwash flow rate through the filters and the
average and maximum duration of the filter backwash process in minutes.
d)
Typical filter run length and a written summary of how filter run length is
determined.
e)
The type of treatment provided for the recycle flow.
f)
Data on the physical dimensions of the equalization and/or treatment units,
typical and maximum hydraulic loading rates, type of treatment chemicals
used and average dose and frequency of use, and frequency at which
solids are removed, if applicable.
Table 1.1 CT values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by free chlorine at 0.5°C or
lower1
Residual (mg/L)
pH
<6.0
6.5
7.0
7.5
8.0
8.5
<9.0
<0.4
137
163
195
237
277
329
390
0.6
141
168
200
239
286
342
407
0.8
145
172
205
246
295
354
422
1.0
148
176
210
253
304
365
437
1.2
152
180
215
259
313
376
451
1.4
155
184
221
266
321
387
464
1.6
157
189
226
273
329
397
477
1.8
162
193
231
279
338
407
489
2.0
165
197
236
286
346
417
500
2.2
169
201
242
297
353
426
511
2.4
172
205
247
298
361
435
522
2.6
175
209
252
304
368
444
533
2.8
178
213
257
310
375
452
543
3.0
181
217
261
316
382
460
552
1These CT values achieve greater than a 99.99 percent inactivation of viruses. CT values between the indicated pH
values may be determined by interpolation. CT values between the indicated temperatures of different tables may be
determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the lower temperature and at
the higher pH.
44
Table 1.2 CT values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by free chlorine at 0.5°C1
Free Residual (mg/L)
pH
<6.0
6.5
7.0
7.5
8.0
8.5
<9.0
<0.4
97
117
139
166
198
236
279
0.6
100
120
143
171
204
244
291
0.8
103
122
146
175
210
252
301
1.0
105
125
149
179
216
260
312
1.2
107
127
152
183
221
267
320
1.4
109
130
155
187
227
274
329
1.6
111
132
158
192
232
281
337
1.8
114
135
162
196
238
287
345
2.0
116
138
165
200
243
294
353
2.2
118
140
169
204
248
300
361
2.4
120
143
172
209
253
306
368
2.6
122
146
175
213
258
312
375
2.8
124
148
178
217
263
318
382
3.0
126
151
182
221
268
324
389
1These CT values achieve greater than a 99.99 percent inactivation of viruses. CT values between the indicated pH
values may be determined by interpolation. CT values between the indicated temperatures of different tables may be
determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the lower temperature and at
the higher pH
Table 1.3 CT values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by free chlorine at 10.0°C1
Free Residual (mg/L)
pH
<6.0
6.5
7.0
7.5
8.0
8.5
<9.0
<0.4
73
88
104
125
149
177
209
0.6
75
90
107
128
153
183
218
0.8
78
92
110
131
158
189
226
1.0
79
94
112
134
162
195
234
1.2
80
95
114
137
166
200
240
1.4
82
98
116
140
170
206
247
1.6
83
99
119
144
174
211
253
1.8
86
101
122
147
179
215
259
2.0
87
104
124
150
182
221
265
2.2
89
105
127
153
186
225
271
2.4
90
107
129
157
190
230
276
2.6
92
110
131
160
194
234
281
2.8
93
111
134
163
197
239
287
3.0
95
113
137
166
201
243
292
1These CT values achieve greater than a 99.99 percent inactivation of viruses. CT values between the indicated pH
values may be determined by interpolation. CT values between the indicated temperatures of different tables may be
determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the lower temperature and at
the higher pH.
45
TABLE 1.4 values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by free chlorine at 15.0°C1
Free Residual (mg/L)
pH
<6.0
6.5
7.0
7.5
8.0
8.5
<9.0
<0.4
49
59
70
83
99
118
140
0.6
50
60
72
86
102
122
146
0.8
52
61
73
88
105
126
151
1.0
53
63
75
90
108
130
156
1.2
54
64
76
92
111
134
160
1.4
55
65
78
94
114
137
165
1.6
56
66
79
96
116
141
169
1.8
57
68
81
98
119
144
173
2.0
58
69
83
100
122
147
177
2.2
59
70
85
102
124
150
181
2.4
60
72
86
105
127
153
184
2.6
61
73
88
107
129
156
188
2.8
62
74
89
109
132
159
191
3.0
63
76
91
111
134
162
195
1These CT values achieve greater than a 99.99 percent inactivation of viruses. CT values between the indicated pH
values may be determined by interpolation. CT values between the indicated temperatures of different tables may be
determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the lower temperature and at
the higher pH.
Table 1.5 CT values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by free chlorine at 20.0˚C1
Free Residual (mg/L)
pH
<6.0
6.5
7.0
7.5
8.0
8.5
<9.0
<0.4
36
44
52
62
74
89
105
0.6
38
45
54
64
77
92
109
0.8
39
46
55
66
79
95
113
1.0
39
47
56
67
81
98
117
1.2
40
48
57
69
83
100
120
1.4
41
49
58
70
85
103
123
1.6
42
50
59
72
87
105
126
1.8
43
51
61
74
89
108
129
2.0
44
52
62
75
91
110
132
2.2
44
53
63
77
93
113
135
2.4
45
54
65
78
95
115
138
2.6
46
55
66
80
97
117
141
2.8
47
56
67
81
99
119
143
3.0
47
57
68
83
101
122
146
1These CT values achieve greater than a 99.99 percent inactivation of viruses. CT values between the indicated pH
values may be determined by interpolation. CT values between the indicated temperatures of different tables may be
determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the lower temperature and at
the higher pH.
46
Table 1.6 CT values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by free chlorine at 25°C1 and
higher
Free Residual (mg/L)
pH
<6.0
6.5
7.0
7.5
8.0
8.5
<9.0
<0.4
24
29
35
42
50
59
70
0.6
25
30
36
43
51
61
73
0.8
26
31
37
44
53
63
75
1.0
26
31
37
45
54
65
78
1.2
27
32
38
46
55
67
80
1.4
27
33
39
47
57
69
82
1.6
28
33
40
48
58
70
84
1.8
29
34
41
49
60
72
86
2.0
29
35
41
50
61
74
88
2.2
30
35
42
51
62
75
90
2.4
30
36
43
52
63
77
92
2.6
31
37
44
53
65
78
94
2.8
31
37
45
54
66
80
96
3.0
32
38
46
55
67
81
97
1These CT values achieve greater than a 99.99 percent inactivation of viruses. CT values between the indicated pH
values may be determined by interpolation. CT values between the indicated temperatures of different tables may be
determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the lower temperature and at
the higher pH.
Table 2.1 CT values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by chlorine dioxide and ozone1
Free Residual (mg/L)
Temperature
1 °C
5°C
10°C
15°C
20°C
>25°C
Chlorine dioxide
63
26
23
19
15
11
Ozone
2.0
1.9
1.4
0.95
0.72
0.46
1These CT values achieve greater than 99.99 percent inactivation of viruses. CT values between the indicated
temperatures may be determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the
lower temperature for determining CT99.9 values between indicated temperatures
Table 3.1 CT values (CT99.9) for 99.9 percent inactivation of giardia lamblia cysts by chloramines1
Temperature
<1 °C
5 °C
10 °C
15 °C
20 °C
>25 °C
2.0
1.9
1.4
0.95
0.72
0.46
1These values are for pH values of 6 to 9. These CT values may be assumed to achieve greater than 99.99 percent
inactivation of viruses only if chlorine is added and mixed in the water prior to the addition of ammonia. If this
condition is not met, the system must demonstrate, based on on-site studies or other information, as approved by the
State, that the system is achieving at least 99.99 percent inactivation of viruses. CT values between the indicated
temperatures may be determined by linear interpolation. If no interpolation is used, use the CT99.9 value at the
lower temperature for determining CT99.9 values between indicated temperatures.
47
SECTION 6.0 CONTROL OF LEAD AND COPPER
6.80
General requirements
6.81
Applicability of corrosion control treatment steps to small, medium-size and large
water systems
6.82
Description of corrosion control treatment requirements
6.83
Source water treatment requirements
6.84
Lead service line replacement requirements
6.85
Public education and supplemental monitoring requirements
6.86
Monitoring requirements for lead and copper in tap water
6.87
Monitoring requirements for water quality parameters
6.88
Source monitoring requirements for lead and copper in water
6.89
Analytical methods
6.90
Reporting requirements
6.91
Record keeping requirements
6.80 General Requirements
(a)
Applicability and Effective Dates
(1)
The requirements of Section 6 constitute the national primary drinking
water regulations for lead and copper. Unless otherwise indicated, each of
the provisions of this Section applies to community water systems and
non-transient, non-community water systems (hereinafter referred to as
“water systems” or “systems”).
(2)
The requirements set forth in Sections 6.86-6.91 shall take effect July 7,
1991. The requirements in Sections 6.80-6.85 shall take effect December
7, 1992.
(b)
Scope
These regulations establish a treatment technique that includes requirements for
corrosion control treatment, source water treatment, lead service line replacement,
and public education. These requirements are triggered, in some cases, by lead
and copper action levels measured in samples collected at consumers' taps.
(c)
Lead and Copper Action Levels
(1)
The lead action level is exceeded if the concentration of lead in more than
ten (10) percent of tap water samples collected during any monitoring
period conducted in accordance with Section 6.86 is greater than 0.015
mg/L (i.e., if the “90th percentile” lead level is greater than 0.015 mg/L).
(2)
The copper action level is exceeded if the concentration of copper in more
than ten (10) percent of tap water samples collected during any monitoring
48
period conducted in accordance with Section 6.86 is greater than 1.3 mg/L
(i.e., if the “90th percentile” copper level is greater than 1.3 mg/L).
(3)
The 90th percentile lead and copper levels shall be computed as follows:
(i)
The results of all lead or copper samples taken during a monitoring
period shall be placed in ascending order from the sample with the
lowest concentration to the sample with the highest concentration.
Each sampling result shall be assigned a number, ascending by
single integers beginning with the number 1 for the sample with
the lowest contaminant level. The number assigned to the sample
with the highest contaminant level shall be equal to the total
number of samples taken.
(ii)
The number of samples taken during the monitoring period shall be
multiplied by 0.9.
(iii)
The contaminant concentration in the numbered sample yielded by
the calculation in Paragraph (c)(3)(ii) is the 90th percentile
contaminant level.
(iv)
For water systems serving fewer than 100 people that collect five
(5) samples per monitoring period, the 90th percentile is computed
by taking the average of the highest and second highest
concentrations.
(d)
Corrosion Control Treatment Requirements
(1)
All water systems shall install and operate optimal corrosion control
treatment as defined in Section 1.
(2)
Any water system that complies with the applicable corrosion control
treatment requirements specified by the Director under Sections 6.81 and
6.82 shall be deemed in compliance with the treatment requirement
contained in Paragraph (d)(1) of this Section.
(e)
Source Water Treatment Requirements
Any system exceeding the lead or copper action level shall implement all
applicable source water treatment requirements specified by the Director under
Section 6.83.
(f)
Lead Service Line Replacement Requirements
49
Any system exceeding the lead action level after implementation of applicable
corrosion control and source water treatment requirements shall complete the lead
service line replacement requirements contained in Section 6.84.
(g)
Public Education Requirements
Any system exceeding the lead action level shall implement the public educatio
requirements contained in Section 6.85.
(h)
Monitoring and Analytical Requirements
Tap water monitoring for lead and copper, monitoring for water quality
parameters, source water monitoring for lead and copper, and analyses of the
monitoring results under this subpart shall be completed in compliance with
Sections 6.86, 6.87, 6.88 and 6.89.
(i)
Reporting Requirements
Systems shall report to the Director any information required by the treatment
provisions of this subpart and Section 6.90.
(j)
Record-Keeping Requirements
Systems shall maintain records in accordance with Section 6.91.
(k)
Failure to comply with the applicable requirements of Sections 6.80-6.91,
including requirements established by the Director pursuant to these provisions,
shall constitute a violation of these regulations.
6.81
Applicability of Corrosion Control Treatment Steps to Small, Medium-size and Large
Water Systems
(a)
Systems shall complete the applicable corrosion control treatment requirements
described in Section 6.82 by the deadlines established in this Section.
(1)
A large system (serving >50,000 persons) shall complete the corrosion
control treatment steps specified in Paragraph (d) of this Section, unless it
is deemed to have optimized corrosion control under Paragraph (b)(2) or
(b)(3) of this Section.
(2)
A small system (serving <3,300 persons) and a medium-size system
(serving >3,300 and <50,000 persons) shall complete the corrosion control
treatment steps specified in Paragraph (e) of this Section, unless it is
deemed to have optimized corrosion control under Paragraph (b)(1),
(b)(2), or (b)(3) of this Section.
50
(b)
A system is deemed to have optimized corrosion control and is not required to
complete the applicable corrosion control treatment steps identified in this Section
if the system satisfies one (1) of the criteria specified in Paragraphs (b)(1) through
(b)(3) of this Section. Any such system deemed to have optimized corrosion
control under this Paragraph, and which has treatment in place, shall continue to
operate and maintain optimal corrosion control treatment and meet any
requirements that the Director determines appropriate to ensure optimal corrosion
control treatment is maintained.
(1)
A small or medium-size water system is deemed to have optimized
corrosion control if the system meets the lead and copper action levels
during each of two (2) consecutive six-month monitoring periods
conducted in accordance with Section 6.86.
(2)
Any water system may be deemed by the Director to have optimized
corrosion control treatment if the system demonstrates to the satisfaction
of the Director that it has conducted activities equivalent to the corrosion
control steps applicable to such system under this Section. If the Director
makes this determination, the Director shall provide the system with
written notice explaining the basis for his decision and shall specify the
water quality control parameters representing optimal corrosion control in
accordance with Section 6.82(f). Water systems deemed to have optimized
corrosion control under this Paragraph shall operate in compliance with
the Director-designated optimal water quality control parameters in
accordance with Section 6.82(g) and continue to conduct lead and copper
tap and water quality parameter sampling in accordance with Section
6.86(d)(3) and Section 6.87(d), respectively. A system shall provide the
Director with the following information in order to support a
determination under this Paragraph:
(i)
the results of all test samples collected for each of the water quality
parameters in Section 6.82(c)(3).
(ii)
a report explaining the test methods used by the water system to
evaluate the corrosion control treatments listed in Section
6.82(c)(1), the results of all tests conducted, and the basis for the
system's selection of optimal corrosion control treatment;
(iii)
a report explaining how corrosion control has been installed and
how it is being maintained to insure minimal lead and copper
concentrations at consumers' taps; and
(iv)
the results of tap water samples collected in accordance with
Section 6.86 at least once every six (6) months for one (1) year
after corrosion control has been installed.
51
(3)
Any water system is deemed to have optimized corrosion control if it
submits results of tap water monitoring conducted in accordance with
Section 6.86 and source water monitoring conducted in accordance with
Section 6.88 that demonstrates for two (2) consecutive six-month
monitoring periods that the difference between the 90th percentile tap
water lead level computed under Section 6.80(c)(3), and the highest source
water lead concentration, is less than the Practical Quantitation Level for
lead specified in Section 6.89(a)(1)(ii).
(i)
Those systems whose highest source water lead level is below the
Method Detection Limit may also be deemed to have optimized
corrosion control under this Paragraph if the 90th percentile tap
water lead level is less than or equal to the Practical Quantitation
Level for lead for two (2) consecutive 6-month monitoring periods.
(ii)
Any water system deemed to have optimized corrosion control in
accordance with this Paragraph shall continue monitoring for lead
and copper at the tap no less frequently than once every three (3)
calendar years using the reduced number of sites specified in
Section 6.86(c) and collecting the samples at times and locations
specified in Section 6.86(d)(4)(iv). Any such system that has not
conducted a round of monitoring pursuant to Section 6.86(d) since
September 30, 1997, shall complete a round of monitoring
pursuant to this Paragraph no later than September 30, 2000.
(iii)
Any water system deemed to have optimized corrosion control
pursuant to this Paragraph shall notify the Director in writing
pursuant to Section 6.90(a)(3) of any change in treatment or the
addition of a new source. The Director may require any such
system to conduct additional monitoring or to take other action the
Director deems appropriate to ensure that such systems maintain
minimal levels of corrosion in the distribution system.
(iv)
As of July 12, 2001, a system is not deemed to have optimized
corrosion control under this Paragraph, and shall implement
corrosion control treatment pursuant to Paragraph (b)(3)(v) of this
Section unless it meets the copper action level.
(v)
Any system triggered into corrosion control because it is no longer
deemed to have optimized corrosion control under this Paragraph
shall implement corrosion control treatment in accordance with the
deadlines in Paragraph (e) of this Section. Any such large system
shall adhere to the schedule specified in that Paragraph for
medium-size systems, with the time periods for completing each
step being triggered by the date the system is no longer deemed to
have optimized corrosion control under this Paragraph.
52
(c)
Any small or medium-size water system that is required to complete the corrosion
control steps due to its exceedance of the lead or copper action level may cease
completing the treatment steps whenever the system meets both action levels
during each of two (2) consecutive monitoring periods conducted pursuant to
Section 6.86, and submits the results to the Director. If any such water system
thereafter exceeds the lead or copper action level during any monitoring period,
the system shall recommence completion of the applicable treatment steps,
beginning with the first treatment step which was not previously completed in its
entirety. The Director may require a system to repeat treatment steps previously
completed by the system where the Director determines that this is necessary to
implement properly the treatment requirements of this Section. The Director shall
notify the system in writing of such a determination and explain the basis for its
decision. The requirement for any small or medium-size system to implement
corrosion control treatment steps in accordance with Paragraph (e) of this Section
(including systems deemed to have optimized corrosion control under Paragraph
(b)(1) of this Section) is triggered whenever any small or medium-size system
exceeds the lead or copper action level.
(d)
Treatment Steps and Deadlines for Large Systems
Except as provided in Paragraph (b)(2) and (3) of this Section, large systems shall
complete the following corrosion control treatment steps (described in the
referenced portions of Sections 6.82, 6.86, and 6.87) by the indicated dates.
(1)
Step 1: The system shall conduct initial monitoring (Section 6.86(d)(1)
and Section 6.87(b)) during two (2) consecutive six-month monitoring
periods by January 1, 1993.
(2)
Step 2: The system shall complete corrosion control studies (Section
6.82(c)) by July 1, 1994.
(3)
Step 3: The Director shall designate optimal corrosion control treatment
(Section 6.82(d)) by January 1, 1995.
(4)
Step 4: The system shall install optimal corrosion control treatment
(Section 6.82(e)) by January 1, 1997.
(5)
Step 5: The system shall complete follow-up sampling (Section 6.86(d)(2)
and Section 6.87(c)) by January 1, 1998.
(6)
Step 6: The Director shall review installation of treatment and designate
optimal water quality control parameters (Section 6.82(f)) by July 1, 1998.
53
(7)
Step 7: The system shall operate in compliance with the Director-specified
optimal water quality control parameters (Section 6.82(g)) and continue to
conduct tap sampling (Section 6.86(d)(3) and Section 6.87(d)).
(e)
Treatment Steps and Deadlines for Small and Medium-size Systems
Except as provided in Paragraph (b) of this Section, small and medium-size
systems shall complete the following corrosion control treatment steps (described
in the referenced portions of Sections 6.82, 6.86 and 6.87) by the indicated time
periods.
(1)
Step 1: The system shall conduct initial tap sampling (Section 6.86(d)(1)
and Section 6.87(b)) until the system either exceeds the lead or copper
action level or becomes eligible for reduced monitoring under Section
6.86(d)(4). A system exceeding the lead or copper action level shall
recommend optimal corrosion control treatment (Section 6.82(a)) within
six (6) months after it exceeds one (1) of the action levels.
(2)
Step 2: Within twelve (12) months after a system exceeds the lead or
copper action level, the Director may require the system to perform
corrosion control studies (Section 6.82(b)). If the Director does not require
the system to perform such studies, the Director shall specify optimal
corrosion control treatment (Section 6.82(d)) within the following time
frames:
(i)
for medium-size systems, within eighteen (18) months after such
system exceeds the lead or copper action level,
(ii)
for small systems, within twenty-four (24) months after such
system exceeds the lead or copper action level.
(3)
Step 3: If the Director requires a system to perform corrosion control
studies under step 2, the system shall complete the studies (Section
6.82(c)) within 18 months after the Director requires that such studies be
conducted.
(4)
Step 4: If the system has performed corrosion control studies under step 2,
the Director shall designate optimal corrosion control treatment (Section
6.82(d)) within 6 months after completion of Step 3.
(5)
Step 5: The system shall install optimal corrosion control treatment
(Section 6.82(e)) within twenty-four (24) months after the Director
designates such treatment.
54
(6)
Step 6: The system shall complete follow-up sampling (Section 6.86(d)(2)
and Section 6.87(c)) within thirty-six (36) months after the Director
designates optimal corrosion control treatment.
(7)
Step 7: The Director shall review the system's installation of treatment and
designate optimal water quality control parameters (Section 6.82(f))
within six (6) months after completion of Step 6.
(8)
Step 8: The system shall operate in compliance with the Director-
designated optimal water quality control parameters (Section 6.82(g)) and
continue to conduct tap sampling (Section 6.86(d)(3) and Section 6.87(d)).
6.82 Description of Corrosion Control Treatment Requirements
Each system shall complete the corrosion control treatment requirements described below
which are applicable to such system under Section 6.81.
(a)
System Recommendation Regarding Corrosion Control Treatment
Based upon the results of lead and copper tap monitoring and water quality
parameter monitoring, small and medium-size water systems exceeding the lead
or copper action level shall recommend installation of one (1) or more of the
corrosion control treatments listed in Paragraph (c)(1) of this Section which the
system believes constitutes optimal corrosion control for that system. The
Director may require the system to conduct additional water quality parameter
monitoring in accordance with Section 6.87(b) to assist the Director in reviewing
the system's recommendation.
(b)
Decision to Require Studies of Corrosion Control Treatment (Applicable to Small
and Medium-size Systems)
The Director may require any small or medium-size system that exceeds the lead
or copper action level to perform corrosion control studies under Paragraph (c) of
this Section to identify optimal corrosion control treatment for the system.
(c)
Performance of Corrosion Control Studies
(1)
Any PWS performing corrosion control studies shall evaluate the
effectiveness of each of the following treatments, and, if appropriate,
combinations of the following treatments to identify the optimal corrosion
control treatment for that system:
(i)
alkalinity and pH adjustment;
(ii)
calcium hardness adjustment; and
55
(iii)
the addition of a phosphate or silicate based corrosion inhibitor at a
concentration sufficient to maintain an effective residual
concentration in all test tap samples.
(2)
The water system shall evaluate each of the corrosion control treatments
using either pipe rig/loop tests, metal coupon tests, partial-system tests, or
analyses based on documented analogous treatments with other systems of
similar size, water chemistry and distribution system configuration.
(3)
The water system shall measure the following water quality parameters in
any tests conducted under this Paragraph before and after evaluating the
corrosion control treatments listed above:
(i)
lead;
(ii)
copper;
(iii)
pH;
(iv)
alkalinity;
(v)
calcium;
(vi)
conductivity;
(vii)
orthophosphate (when an inhibitor containing a phosphate
compound is used);
(viii) silicate (when an inhibitor containing a silicate compound is used);
and
(ix)
water temperature.
(4)
The water system shall identify all chemical or physical constraints that
limit or prohibit the use of a particular corrosion control treatment and
document such constraints with at least one (1) of the following:
(i)
data and documentation showing that a particular corrosion control
treatment has adversely affected other water treatment processes
when used by another water system with comparable water quality
characteristics; and/or
(ii)
data and documentation demonstrating that the water system has
previously attempted to evaluate a particular corrosion control
treatment and has found that the treatment is ineffective or
adversely affects other water quality treatment processes.
56
(5)
The water system shall evaluate the effect of the chemicals used for
corrosion control treatment on other water quality treatment processes.
(6)
On the basis of an analysis of the data generated during each evaluation,
the water system shall recommend to the Director in writing the treatment
option that the corrosion control studies indicate constitutes optimal
corrosion control treatment for that system. The water system shall
provide a rationale for its recommendation along with all supporting
documentation specified in Paragraphs (c)(1) through (5) of this Section.
(d)
Designation of Optimal Corrosion Control Treatment
(1)
Based upon consideration of available information including, where
applicable, studies performed under Paragraph (c) of this Section and a
system's recommended treatment alternative, the Director shall either
approve the corrosion control treatment option recommended by the
system, or designate alternative corrosion control treatment(s) from among
those listed in Paragraph (c)(1) of this Section. When designating optimal
treatment the Director shall consider the effects that additional corrosion
control treatment will have on water quality parameters and on other water
quality treatment processes.
(2)
The Director shall notify the system of his decision on optimal corrosion
control treatment in writing and explain the basis for this determination. If
the Director requests additional information to aid his review, the water
system shall provide the information.
(e)
Installation of Optimal Corrosion Control
Each system shall properly install and operate throughout its distribution system
the optimal corrosion control treatment designated by the Director under
Paragraph (d) of this Section.
(f)
Review of Treatment and Specification of Optimal Water Quality Control
Parameters
The Director shall evaluate the results of all lead and copper tap samples and
water quality parameter samples submitted by the water system and determine
whether the system has properly installed and operated the optimal corrosion
control treatment designated by the Director in Paragraph (d) of this Section.
Upon reviewing the results of tap water and water quality parameter monitoring
by the system, both before and after the system installs optimal corrosion control
treatment, the Director shall designate:
57
(1)
a minimum value or a range of values for pH measured at each entry point
to the distribution system;
(2)
a minimum pH value, measured in all tap samples. Such value shall be
equal to or greater than 7.0, unless the Director determines that meeting a
pH level of 7.0 is not technologically feasible or is not necessary for the
system to optimize corrosion control;
(3)
if a corrosion inhibitor is used, a minimum concentration or a range of
concentrations for the inhibitor, measured at each entry point to the
distribution system and in all tap samples, that the Director determines is
necessary to form a passivating film on the interior walls of the pipes of
the distribution system;
(4)
if alkalinity is adjusted as part of optimal corrosion control treatment, a
minimum concentration or a range of concentrations for alkalinity,
measured at each entry point to the distribution system and in all tap
samples;
(5)
if calcium carbonate stabilization is used as part of corrosion control, a
minimum concentration or a range of concentrations for calcium,
measured in all tap samples.
The values for the applicable water quality control parameters listed above
shall be those that the Director determines to reflect optimal corrosion
control treatment for the system. The Director may designate values for
additional water quality control parameters determined by the Director to
reflect optimal corrosion control for the system. The Director shall notify
the system in writing of these determinations and explain the basis for his
decisions.
(g)
Continued Operation and Monitoring
All systems optimizing corrosion control shall continue to operate and maintain
optimal corrosion control treatment, including maintaining water quality
parameters at or above minimum values or within ranges designated by the
Director under Paragraph (f) of this Section, in accordance with this Paragraph for
all samples collected under Sections 6.87(d)-(f). Compliance with the
requirements of this Paragraph shall be determined every six (6) months, as
specified under Section 6.87(d). A water system is out of compliance with the
requirements of this Paragraph for a six-month period if it has excursions for any
Director-specified parameter on more than nine (9) days during the period. An
excursion occurs whenever the daily value for one or more of the water quality
parameters measured at a sampling location is below the minimum value or
outside the range designated by the Director. Daily values are calculated as
58
follows. The Director has the discretion to delete results of obvious sampling
errors from this calculation.
(1)
On days when more than one (1) measurement for the water quality
parameter is collected at the sampling location, the daily value shall be the
average of all results collected during the day regardless of whether they
are collected through continuous monitoring, grab sampling, or a
combination of both.
(2)
On days when only one (1) measurement for the water quality parameter is
collected at the sampling location, the daily value shall be the result of that
measurement.
(3)
On days when no measurement is collected for the water quality parameter
at the sampling location, the daily value shall be the daily value calculated
on the most recent day on which the water quality parameter was
measured at the sample site.
(h)
Modification of the Director's Treatment Decisions
Upon his own initiative or in response to a request by a water system or other
interested party, the Director may modify his determination of the optimal
corrosion control treatment under Paragraph (d) of this Section or optimal water
quality control parameters under Paragraph (f) of this Section. A request for
modification by a system or other interested party shall be in writing, explain why
the modification is appropriate and provide supporting documentation. The
Director may modify his determination where he concludes that such change is
necessary to ensure that the system continues to optimize corrosion control
treatment. A revised determination shall be made in writing, set forth the new
treatment requirements, explain the basis for the Director's decision and provide
an implementation schedule for completing the treatment modifications.
6.83
Source Water Treatment Requirements
Systems shall complete the applicable source water monitoring and treatment
requirements (described in the referenced portions of Paragraph (b) of this Section, and in
Sections 6.86, and 6.88) by the following deadlines.
(a)
Deadlines for Completing Source Water Treatment Steps
(1)
Step 1: A system exceeding the lead or copper action level shall complete
lead and copper source water monitoring (Section 6.88(b)) and make a
treatment recommendation to the Director (Section 6.83(b)(1)) within 6
months after exceeding the lead or copper action level.
59
(2)
Step 2: The Director shall make a determination regarding source water
treatment (Section 6.83(b)(2)) within six (6) months after submission of
monitoring results under Step 1.
(3)
Step 3: If the Director requires installation of source water treatment, the
system shall install the treatment (Section 6.83(b)(3)) within twenty-four
(24) months after completion of Step 2.
(4)
Step 4: The system shall complete follow-up tap water monitoring
(Section 6.86(d)(2) and source water monitoring (Section 6.88(c)) within
thirty-six (36) months after completion of Step 2.
(5)
Step 5: The Director shall review the system's installation and operation of
source water treatment and specify maximum permissible source water
levels (Section 6.83(b)(4)) within six (6) months after completion of Step
4.
(6)
Step 6: The system shall operate in compliance with the Director-specified
maximum permissible lead and copper source water levels (Section
6.83(b)(4)) and continue source water monitoring (Section 6.88(d)).
(b)
Description of Source Water Treatment Requirements
(1)
System treatment recommendation
Any system which exceeds the lead or copper action level shall
recommend in writing to the Director the installation and operation of one
(1) of the source water treatments listed in Paragraph (b)(2) of this
Section. A system may recommend that no treatment be installed based
upon a demonstration that source water treatment is not necessary to
minimize lead and copper levels at users' taps.
(2)
The Director shall complete an evaluation of the results of all source water
samples submitted by the water system to determine whether source water
treatment is necessary to minimize lead or copper levels in water delivered
to users' taps. If the Director determines that treatment is needed, the
Director shall either require installation and operation of the source water
treatment recommended by the system (if any) or require the installation
and operation of another source water treatment from among the
following:
ion
exchange,
reverse
osmosis,
lime
softening
or
coagulation/filtration. If the Director requests additional information to aid
in his review, the water system shall provide the information by the date
specified by the Director in his request. The Director shall notify the
system in writing of its determination and set forth the basis for its
decision.
60
(3)
Installation of Source Water Treatment
Each system shall properly install and operate the source water treatment
designated by the Director under Paragraph (b)(2) of this Section.
(4)
The Director shall review the source water samples taken by the water
system both before and after the system installs source water treatment,
and determine whether the system has properly installed and operated the
source water treatment designated by the Director. Based upon his review,
the Director shall designate the maximum permissible lead and copper
concentrations for finished water entering the distribution system. Such
levels shall reflect the contaminant removal capability of the treatment
properly operated and maintained. The Director shall notify the system in
writing and explain the basis for his decision.
(5)
Continued Operation and Maintenance
Each water system shall maintain lead and copper levels below the
maximum permissible concentrations designated by the Director at each
sampling point monitored in accordance with Section 6.88. The system is
out of compliance with this Paragraph if the level of lead or copper at any
sampling point is greater than the maximum permissible concentration
designated by the Director.
(6)
Modification of Treatment Decisions
Upon his own initiative or in response to a request by a water system or
other interested party, the Director may modify his determination of the
source water treatment under Paragraph (2) of this Section, or maximum
permissible lead and copper concentrations for finished water entering the
distribution system under Paragraph (4) of this Section. A request for
modification by a system or other interested party shall be in writing,
explain why the modification is appropriate and provide supporting
documentation. The Director may modify his determination where he
concludes that such change is necessary to ensure that the system
continues to minimize lead and copper concentrations in source water. A
revised determination shall be made in writing, set forth the new treatment
requirements, explain the basis for the Director's decision, and provide an
implementation schedule for completing the treatment modifications.
6.84
Lead Service Line Replacement Requirements
(a)
Systems that fail to meet the lead action level in tap samples taken pursuant to
Section 6.86(d)(2), after installing corrosion control and/or source water treatment
(whichever sampling occurs later), shall replace lead service lines in accordance
with the requirements of this Section. If a system is in violation of Section 6.81 or
61
Section 6.83 for failure to install source water or corrosion control treatment, the
Director may require the system to commence lead service line replacement under
this Section after the date by which the system was required to conduct
monitoring under Section 6.86(d)(2) has passed.
(b)
A water system shall replace annually at least seven (7) percent of the initial
number of lead service lines in its distribution system. The initial number of lead
service lines is the number of lead lines in place at the time the replacement
program begins. The system shall identify the initial number of lead service lines
in its distribution system, including an identification of the portion(s) owned by
the system, based upon a materials evaluation, including the evaluation required
under Section 6.86(a) and relevant legal authorities (e.g., contracts, local
ordinances) regarding the portion owned by the system. The first year of lead
service line replacement shall begin on the date the action level was exceeded in
tap sampling referenced in Paragraph (a) of this Section.
(c)
A system is not required to replace an individual lead service line if the lead
concentration in all service line samples from that line, taken pursuant to Section
6.86(b)(3), is less than or equal to 0.015 mg/L.
(d)
A water system shall replace that portion of the lead service line that it owns. In
cases where the system does not own the entire lead service line, the system shall
notify the owner of the line, or the owner's authorized agent, that the system will
replace the portion of the service line that it owns and shall offer to replace the
owner’s portion of the line. A system is not required to bear the cost of replacing
the privately-owned portion of the line, nor is it required to replace the privately-
owned portion where the owner chooses not to pay the cost of replacing the
privately-owned portion of the line, or where replacing the privately-owned
portion would be precluded by the State, local or common law. A water system
that does not replace the entire length of the service line also shall complete the
following tasks.
(1)
At least forty-five (45) days prior to commencing with the partial
replacement of a lead service line, the water system shall provide notice to
the resident(s) of all buildings served by the line explaining that they may
experience a temporary increase of lead levels in their drinking water,
along with guidance on measures consumers can take to minimize their
exposure to lead. The Director may allow the water system to provide
notice under the previous sentence less than forty-five (45) days prior to
commencing partial lead service line replacement where such replacement
is in conjunction with emergency repairs. In addition, the water system
shall inform the resident(s) served by the line that the system will, at the
system’s expense, collect a sample from each partially-replaced lead
service line that is representative of the water in the service line for
analysis of lead content, as prescribed under Section 6.86(b)(3), within
seventy-two (72) hours after the completion of the partial replacement of
62
the service line. The system shall collect the sample and report the results
of the analysis to the owner and the resident(s) served by the line within
three (3) business days of receiving the results. Mailed notices post-
marked within three (3) business days of receiving the results shall be
considered “on time.”
(2)
The water system shall provide the information required by Paragraph (1)
of this Section to the residents of individual dwellings by mail or by other
methods approved by the Director. In instances where multi-family
dwellings are served by the line, the water system shall have the option to
post the information at a conspicuous location.
(e)
The Director shall require a system to replace lead service lines on a shorter
schedule than that required by this Section, taking into account the number of lead
service lines in the system, where such a shorter replacement schedule is feasible.
The Director shall make this determination in writing and notify the system of its
finding within six (6) months after the system is triggered into lead service line
replacement based on monitoring referenced in Paragraph (a) of this Section.
(f)
Any system may cease replacing lead service lines whenever first draw samples
collected pursuant to Section 6.86(b)(2) meet the lead action level during each of
two (2) consecutive monitoring periods and the system submits the results to the
Director. If the first draw tap samples collected in any such system thereafter
exceeds the lead action level, the system shall recommence replacing lead service
lines, pursuant to Paragraph (b) of this Section.
(g)
To demonstrate compliance with Paragraphs (a)-(d) of this Section, a system shall
report to the Director the information specified in Section 6.90(e).
6.85
Public Education and Supplemental Monitoring Requirements
A water system that exceeds the lead action level based on tap water samples collected in
accordance with Section 6.86 shall deliver the public education materials contained in
Paragraphs (a) and (b) of this Section in accordance with the requirements in Paragraph
(c) of this Section.
(a)
Content of Written Public Education Materials
(1)
Community Water Systems
A community water system shall include the following text in all of the
printed materials it distributes through its lead public education program.
Systems may delete information pertaining to lead service lines, upon
approval by the Director, if no lead service lines exist anywhere in the
water system service area. Public education language at Paragraphs
(a)(1)(iv)(B)(5) and (a)(1)(iv)(D)(2) of this Section may be modified
63
regarding building permit record availability and consumer access to these
records, if approved by the Director. Systems may also continue to utilize
pre-printed materials that meet the public education language requirements
in Section 6.85, effective December 7, 1991. Any additional information
presented by a system shall be consistent with the information below and
be in plain English that can be understood by lay people.
(i)
Introduction
The United States Environmental Protection Agency (EPA)
and [insert name of water supplier] are concerned about
lead in your drinking water. Although most homes have
very low levels of lead in their drinking water, some homes
in the community have lead levels above the EPA action
level of fifteen (15) parts per billion (ppb), or 0.015
milligrams of lead per liter of water (mg/L). Under Federal
law we are required to have a program in place to minimize
lead in your drinking water by [insert date when corrosion
control will be completed for your system]. This program
includes corrosion control treatment, source water
treatment and public education. We are also required to
replace the portion of each lead service line that we own if
the line contributes lead concentrations of more than fifteen
(15) ppb after we have completed the comprehensive
treatment program. If you have any questions about how
we are carrying out the requirements of the lead regulation
please give us a call at [insert water system's phone
number]. This brochure explains the simple steps you can
take to protect you and your family by reducing your
exposure to lead in drinking water.
(ii)
Health Effects of Lead
Lead is a common metal found throughout the environment
in lead-based paint, air, soil, household dust, food, certain
types of pottery porcelain and pewter and water. Lead can
pose a significant risk to your health if too much of it enters
your body. Lead builds up in the body over many years and
can cause damage to the brain, red blood cells and kidneys.
The greatest risk is to young children and pregnant women.
Amounts of lead that won't hurt adults can slow down
normal mental and physical development of growing
bodies. In addition, a child at play often comes into contact
with sources of lead contamination–like dirt and dust–that
rarely affect an adult. It is important to wash children's
64
hands and toys often, and to try to make sure they only put
food in their mouths.
(iii)
Lead in Drinking Water
(A)
Lead in drinking water, although rarely the sole
cause of lead poisoning, can significantly increase a
person's total lead exposure, particularly the
exposure of infants who drink baby formulas and
concentrated juices that are mixed with water. The
EPA estimates that drinking water can make up
twenty (20) percent or more of a person's total
exposure to lead.
(B)
Lead is unusual among drinking water contaminants
in that it seldom occurs naturally in water supplies
like rivers and lakes. Lead enters drinking water
primarily as a result of the corrosion, or wearing
away, of materials containing lead in the water
distribution system and household plumbing. These
materials include lead-based solder used to join
copper pipe, brass and chrome plated brass faucets,
and in some cases, pipes made of lead that connect
your house to the water main (service lines). In
1986, Congress banned the use of lead solder
containing greater than 0.2% lead, and restricted the
lead content of faucets, pipes and other plumbing
materials to 8.0%.
(C)
When water stands in lead pipes or plumbing
systems containing lead for several hours or more,
the lead may dissolve into your drinking water. This
means the first water drawn from the tap in the
morning, or later in the afternoon after returning
from work or school, can contain fairly high levels
of lead.
(iv)
Steps You Can Take in The Home to Reduce Exposure to
Lead in Drinking Water
(A)
Despite our best efforts mentioned earlier to control
water corrosivity and remove lead from the water
supply, lead levels in some homes or buildings can
be high. To find out whether you need to take action
in your own home, have your drinking water tested
to determine if it contains excessive concentrations
65
of lead. Testing the water is essential because you
cannot see, taste, or smell lead in drinking water.
Some local laboratories that can provide this service
are listed at the end of this booklet. For more
information on having your water tested, please call
[insert phone number of water system].
(B)
If a water test indicates that the drinking water
drawn from a tap in your home contains lead above
fifteen (15) ppb, then you should take the following
precautions:
(1)
Let the water run from the tap before using
it for drinking or cooking any time the water
in a faucet has gone unused for more than
six (6) hours. The longer water resides in
your home's plumbing the more lead it may
contain. Flushing the tap means running the
cold water faucet until the water gets
noticeably colder, usually about 15-30
seconds. If your house has a lead service
line to the water main, you may have to
flush the water for a longer time, perhaps
one (1) minute, before drinking. Although
toilet flushing or showering flushes water
through a portion of your home's plumbing
system, you still need to flush the water in
each faucet before using it for drinking or
cooking. Flushing tap water is a simple and
inexpensive measure you can take to protect
your family's health. It usually uses less than
one (1) or two (2) gallons of water and costs
less than [insert a cost estimate based on
flushing two (2) times a day for 30 days] per
month. To conserve water, fill a couple of
bottles for drinking water after flushing the
tap, and whenever possible use the first flush
water to wash the dishes or water the plants.
If you live in a high-rise building, letting the
water flow before using it may not work to
lessen your risk from lead. The plumbing
systems have more, and sometimes larger
pipes than smaller buildings. Ask your
landlord for help in locating the source of
the lead and for advice on reducing the lead
level.
66
(2)
Try not to cook with, or drink water from
the hot water tap. Hot water can dissolve
lead more quickly than cold water. If you
need hot water, draw water from the cold tap
and heat it on the stove.
(3)
Remove loose lead solder and debris from
the plumbing materials installed in newly
constructed homes, or homes in which the
plumbing has recently been replaced, by
removing the faucet strainers from all taps
and running the water from three (3) to five
(5) minutes. Thereafter, periodically remove
the strainers and flush out any debris that
has accumulated over time.
(4)
If your copper pipes are joined with lead
solder that has been installed illegally since
it was banned in 1986, notify the plumber
who did the work and request that he or she
replace the lead solder with lead-free solder.
Lead solder looks dull gray, and when
scratched with a key looks shiny. In
addition, notify the Division of Drinking
Water Quality, RI Department of Health
about the violation.
(5)
Determine whether or not the service line
that connects your home or apartment to the
water main is made of lead. The best way to
determine if your service line is made of
lead is by either hiring a licensed plumber to
inspect the line or by contacting the
plumbing contractor who installed the line.
You can identify the plumbing contractor by
checking the city's record of building
permits which should be maintained in the
files of the [insert name of department that
issues building permits]. A licensed plumber
can at the same time check to see if your
home's plumbing contains lead solder, lead
pipes, or pipe fittings that contain lead. The
PWS that delivers water to your home
should also maintain records of the materials
located in the distribution system. If the
67
service line that connects your dwelling to
the water main contributes more than fifteen
(15) ppb to drinking water, after our
comprehensive treatment program is in
place, we are required to replace the portion
of the line we own. If the line is only
partially owned by the [insert the name of
the city, county, or water system that owns
the line], we are required to provide the
owner of the privately-owned portion of the
line with information on how to replace the
privately-owned portion of the service line,
and offer to replace that portion of the line at
the owner's expense. If we replace only the
portion of the line that we own, we also are
required to notify you in advance and
provide you with information on the steps
you can take to minimize exposure to any
temporary increase in lead levels that may
result from the partial replacement, to take a
follow-up sample at our expense from the
line within seventy-two (72) hours after the
partial replacement and to mail or otherwise
provide you with the results of that sample
within three (3) business days of receiving
the
results.
Acceptable
replacement
alternatives include copper, steel, iron and
plastic pipes.
(6)
Have an electrician check your wiring. If
grounding wires from the electrical system
are attached to your pipes, corrosion may be
greater. Check with a licensed electrician or
your local electrical code to determine if
your wiring can be grounded elsewhere. DO
NOT attempt to change the wiring yourself
because improper grounding can cause
electrical shock and fire hazards.
(C)
The steps described above will reduce the lead
concentrations in your drinking water. However, if
a water test indicates that the drinking water coming
from your tap contains lead concentrations in excess
of fifteen (15) ppb after flushing, or after we have
completed our actions to minimize lead levels, then
68
you may want to take the following additional
measures:
(1)
Purchase or lease a home treatment device.
Home treatment devices are limited in that
each unit treats only the water that flows
from the faucet to which it is connected, and
all
of
the
devices
require
periodic
maintenance and replacement. Devices such
as reverse osmosis systems or distillers can
effectively remove lead from your drinking
water. Some activated carbon filters may
reduce lead levels at the tap, however all
lead reduction claims should be investigated.
Be sure to check the actual performance of a
specific home treatment device before and
after installing the unit.
(2)
Purchase bottled water for drinking and
cooking.
(D)
You can consult a variety of sources for additional
information. Your family doctor or pediatrician can
perform a blood test for lead and provide you with
information about the health effects of lead. State
and local government agencies that can be
contacted include:
(1)
The Office of Drinking Water Quality
within the Rhode Island Department of
Health at 401-222-6867 can provide you
with information about your community's
water supply, and a list of local laboratories
that have been certified by the Health
Department for testing water quality;
(2)
[insert the name of city of county
department that issues building permits] at
[insert phone number] can provide you with
information about building permit records
that should contain the names of plumbing
contractors that plumbed your home; and
(3)
The Division of Family Health within the
Rhode Island Department of Health at 401-
222-2312 can provide you with information
69
about the health effects of lead and how you
can have your child's blood tested.
(E)
The following is a list of some State of Rhode
Island approved laboratories in your area that you
can call to have your water tested for lead. [Insert
names and phone numbers of at least two (2)
laboratories].
(2)
Non-transient, Non-community Water Systems
A non-transient non-community water system shall either include the text
specified in Paragraph (a)(1) of this Section or shall include the following
text in all of the printed materials it distributes through its lead public
education program. Water systems may delete information pertaining to
lead service lines upon approval by the Director if no lead service lines
exist anywhere in the water system service area. Any additional
information presented by a system shall be consistent with the information
below and be in plain English that can be understood by lay people.
(i)
Introduction
The United States Environmental Protection Agency (EPA) and
[insert name of water supplier] are concerned about lead in your
drinking water. Some drinking water samples taken from this
facility have lead levels above the EPA action level of fifteen (15)
parts per billion (ppb), or 0.015 milligrams of lead per liter of
water (mg/L). Under Federal law we are required to have a
program in place to minimize lead in your drinking water by
[insert date when corrosion control will be completed for your
system]. This program includes corrosion control treatment, source
water treatment and public education. We are also required to
replace the portion of each lead service line that we own if the line
contributes lead concentrations of more than 15 ppb after we have
completed the comprehensive treatment program. If you have any
questions about how we are carrying out the requirements of the
lead regulation please give us a call at [insert water system's phone
number]. This brochure explains the simple steps you can take to
protect yourself by reducing your exposure to lead in drinking
water.
(ii)
Health Effects of Lead
Lead is found throughout the environment in lead-based paint, air,
soil, household dust, food, certain types of pottery, porcelain and
pewter, and water. Lead can pose a significant risk to your health if
70
too much of it enters your body. Lead builds up in the body over
many years and can cause damage to the brain, red blood cells and
kidneys. The greatest risk is to young children and pregnant
women. Amounts of lead that won't hurt adults can slow down
normal mental and physical development of growing bodies. In
addition, a child at play often comes into contact with sources of
lead contamination–like dirt and dust–that rarely affect an adult. It
is important to wash children's hands and toys often, and to try to
make sure they only put food in their mouths.
(iii)
Lead in Drinking Water
(A)
Lead in drinking water, although rarely the sole cause of
lead poisoning, can significantly increase a person's total
lead exposure, particularly the exposure of infants who
drink baby formulas and concentrated juices that are mixed
with water. The EPA estimates that drinking water can
make up twenty (20) percent or more of a person's total
exposure to lead.
(B)
Lead is unusual among drinking water contaminants in that
it seldom occurs naturally in water supplies like rivers and
lakes. Lead enters drinking water primarily as a result of
the corrosion, or wearing away, of materials containing
lead in the water distribution system and household
plumbing. These materials include lead-based solder used
to join copper pipe, brass and chrome-plated brass faucets,
and in some cases, pipes made of lead that connect houses
and buildings to water mains (service lines). In 1986,
Congress banned the use of lead solder containing greater
than 0.2% lead, and restricted the lead content of faucets,
pipes and other plumbing materials to 8.0%.
(C)
When water stands in lead pipes or plumbing systems
containing lead for several hours or more, the lead may
dissolve into your drinking water. This means the first
water drawn from the tap in the morning, or later in the
afternoon if the water has not been used all day, can contain
fairly high levels of lead.
(iv)
Steps You Can Take to Reduce Exposure to Lead in Drinking
Water
(A)
Let the water run from the tap before using it for drinking
or cooking any time the water in a faucet has gone unused
for more than six (6) hours. The longer water resides in
71
plumbing the more lead it may contain. Flushing the tap
means running the cold water faucet for about 15-30
seconds. Although toilet flushing or showering flushes
water through a portion of the plumbing system, you still
need to flush the water in each faucet before using it for
drinking or cooking. Flushing tap water is a simple and
inexpensive measure you can take to protect your health. It
usually uses less than one (1) gallon of water.
(B)
Do not cook with, or drink water from the hot water tap.
Hot water can dissolve lead more quickly than cold water.
If you need hot water, draw water from the cold tap and
then heat it.
(C)
The steps described above will reduce the lead
concentrations in your drinking water. However, if you are
still concerned, you may wish to use bottled water for
drinking and cooking.
(D)
You can consult a variety of sources for additional
information. Your family doctor or pediatrician can
perform a blood test for lead and provide you with
information about the health effects of lead. State and local
government agencies that can be contacted include:
(1)
[insert the name or title of facility official if
appropriate] at [insert phone number] can provide
you with information about your facility's water
supply; and
(2)
The Office of Drinking Water Quality within the
Rhode Island Department of Health at 401-222-
6867 can provide you with information about the
health effects of lead.
(b)
Content of Broadcast Materials: A water system shall include the following
information in all public service announcements submitted under its lead public
education program to television and radio stations for broadcasting:
(1)
Why should everyone want to know the facts about lead and drinking
water? Because unhealthy amounts of lead can enter drinking water
through the plumbing in your home. That's why I urge you to do what I
did. I had my water tested for [insert free or $ per sample]. You can
contact the [insert the name of the city or water system] for information on
testing and on simple ways to reduce your exposure to lead in drinking
water.
72
(2)
To have your water tested for lead, or to get more information about this
public health concern, please call [insert the phone number of the city or
water system].
(c)
Delivery of a Public Education Program
(1)
In communities where a significant proportion of the population speaks a
language other than English, public education materials shall be
communicated in the appropriate language(s).
(2)
A community water system that exceeds the lead action level on the basis
of tap water samples collected in accordance with Section 6.86, and that is
not already repeating public education tasks pursuant to Paragraph (c)(3),
(c)(7), or (c)(8) of this Section, shall, within 60 days:
(i)
insert notices in each customer's water utility bill or do a special
mailing containing the information in Paragraph (a) of this Section,
along with the following alert on the water bill itself in large print:
“SOME HOMES IN THIS COMMUNITY HAVE ELEVATED
LEAD LEVELS IN THEIR DRINKING WATER. LEAD CAN
POSE A SIGNIFICANT RISK TO YOUR HEALTH. PLEASE
READ
THE
ENCLOSED
NOTICE
FOR
FURTHER
INFORMATION.” A community water system having a billing
cycle that does not include a billing within 60 days of exceeding
the action level, or that cannot insert information in the water
utility bill without making major changes to its billing system, may
use a separate mailing to deliver the information in Paragraph
(a)(1) of this Section as long as the information is delivered to each
customer within 60 days of exceeding the action level. Such water
systems shall also include the “alert” language specified in this
Paragraph.
(ii)
submit the information in Paragraph (a)(1) to the editorial
departments of the major daily and weekly newspapers circulated
throughout the community.
(iii)
deliver pamphlets and/or brochures that contain the public
education materials in Paragraphs (a)(1)(ii) and (a)(1)(iv) of this
Section to facilities and organizations, including the following:
(A)
public schools and/or local school boards;
(B)
health department;
73
(C)
Women, Infants, and Children and/or Head Start
Program(s) whenever available;
(D)
public and private hospitals and/or clinics;
(E)
pediatricians;
(F)
family planning clinics; and
(G)
local welfare agencies.
(iv)
submit the public service announcement in Paragraph (b) of this
Section to at least five (5) of the radio and television stations with
the largest audiences that broadcast to the community served by
the water system.
(3)
A community water system shall repeat the tasks contained in Paragraphs
(c)(2)(I), (ii) and (iii) of this Section every twelve (12) months, and the
tasks contained in Paragraphs (c)(2)(iv) of this Section every six (6)
months for as long as the system exceeds the lead action level.
(4)
Within 60 days after it exceeds the lead action level (unless it already is
repeating public education tasks pursuant to Paragraph (c)(5) of this
Section), a non-transient, non-community water system shall deliver the
public education materials specified by Paragraphs (a)(1) of this Section or
the public education materials specified by Paragraph (a)(2) of this Section
as follows:
(i)
post informational posters on lead in drinking water in a public
place or common area in each of the buildings served by the
system; and
(ii)
distribute informational pamphlets and/or brochures on lead in
drinking water to each person served by the non-transient non-
community water system. The Director may allow the system to
utilize electronic transmission in lieu of or combined with printed
materials as long as it achieves at least the same coverage.
(5)
A non-transient, non-community water system shall repeat the tasks
contained in Paragraph (c)(4) of this Section at least once during each
calendar year in which the system exceeds the lead action level.
(6)
A water system may discontinue delivery of public education materials if
the system has met the lead action level during the most recent six-month
monitoring period conducted pursuant to Section 6.86. Such a system shall
74
recommence public education in accordance with this Section if it
subsequently exceeds the lead action level during any monitoring period.
(7)
A community water system may apply to the Director, in writing, to use
the text specified in Paragraph (a)(2) of this Section in lieu of the text in
Paragraph (a)(1) of this Section and to perform the tasks listed in
Paragraphs (c)(4) and (c)(5) of this Section in lieu of the tasks in
Paragraphs (c)(2) and (c)(3) of this Section if:
(i)
The system is a facility, such as a prison or a hospital, where the
population served is not capable of or is prevented from making
improvements to plumbing or installing point of use treatment
devices; and
(ii)
The system provides water as part of the cost of services provided
and does not separately charge for water consumption.
(8)(i) A community water system serving 3,300 or fewer people may omit the
task contained in Paragraph (c)(2)(iv) of this Section. As long as it
distributes notices containing the information contained in Paragraph
(a)(1) of this Section to every household served by the system, such
systems may further limit their public education programs as follows:
(A)
Systems serving 500 or fewer people may forego the task
contained in Paragraph (c)(2)(ii) of this Section. Such a system
may limit the distribution of the public education materials
required under Paragraph (c)(2)(iii) of this Section to facilities and
organizations served by the system that are most likely to be
visited regularly by pregnant women and children, unless it is
notified by the Director in writing that it must make a broader
distribution.
(B)
If approved by the Director in writing, a system serving 501 to
3,300 people may omit the task in Paragraph (c)(2)(ii) of this
Section and/or limit the distribution of the public education
materials required under Paragraph (c)(2)(iii) of this Section to
facilities and organizations served by the system that are most
likely to be visited regularly by pregnant women and children.
(ii)
A community water system serving 3,300 or fewer people that
delivers public education in accordance with Paragraph (c)(8)(i) of
this Section shall repeat the required public education tasks at least
once during each calendar year in which the system exceeds the
lead action level.
(d)
Supplemental Monitoring and Notification of Results
75
A water system that fails to meet the lead action level on the basis of tap samples
collected in accordance with Section 6.86 shall offer to sample the tap water of
any customer who requests it. The system is not required to pay for collecting or
analyzing the sample, nor is the system required to collect and analyze the sample
itself.
6.86 Monitoring Requirements for Lead and Copper in Tap Water
(a)
Sample Site Location
(1)
By the applicable date for commencement of monitoring under Paragraph
(d)(1) of this Section, each water system shall complete a materials
evaluation of its distribution system in order to identify a pool of targeted
sampling sites that meets the requirements of this Section, and which is
sufficiently large to ensure that the water system can collect the number of
lead and copper tap samples required in Paragraph (c) of this Section. All
sites from which first draw samples are collected shall be selected from
this pool of targeted sampling sites. Sampling sites may not include
faucets that have point-of-use or point-of-entry treatment devices designed
to remove inorganic contaminants.
(2)
A water system shall use the information on lead, copper and galvanized
steel that is required when conducting a materials evaluation (presence of
lead from piping, solder, caulking, interior home plumbing, copper from
piping and alloys, service lines, and home plumbing, and galvanized
piping, service lines and home plumbing within the distribution system.)
When an evaluation of the information collected pursuant to the above is
insufficient to locate the requisite number of lead and copper sampling
sites that meet the targeting criteria in Paragraph (a) of this Section, the
water system shall review the sources of information listed below in order
to identify a sufficient number of sampling sites. In addition, the system
shall seek to collect such information where possible in the course of its
normal operations (e.g., checking service line materials when reading
water meters or performing maintenance activities):
(i)
all plumbing codes, permits and records in the files of the building
department(s) which indicate the plumbing materials that are
installed within publicly and privately owned structures connected
to the distribution system;
(ii)
all inspections and records of the distribution system that indicate
the material composition of the service connections that connect a
structure to the distribution system; and
76
(iii)
all existing water quality information, which includes the results of
all prior analyses of the system or individual structures connected
to the system, indicating locations that may be particularly
susceptible to high lead or copper concentrations.
(3)
The sampling sites selected for a community water system's sampling pool
(“tier 1 sampling sites”) shall consist of single family structures that:
(i)
contain copper pipes with lead solder installed after 1982 or
contain lead pipes; and/or
(ii)
are served by a lead service line.
When multiple-family residences comprise at least twenty (20)
percent of the structures served by a water system, the system may
include these types of structures in its sampling pool.
(4)
Any community water system with insufficient tier 1 sampling sites shall
complete its sampling pool with “tier 2 sampling sites”, consisting of
buildings, including multiple-family residences that:
(i)
contain copper pipes with lead solder installed after 1982 or
contain lead pipes; and/or
(ii)
are served by a lead service line.
(5)
Any community water system with insufficient tier 1 and tier 2 sampling
sites shall complete its sampling pool with “tier 3 sampling sites”,
consisting of single family structures that contain copper pipes with lead
solder installed before 1983. A community water system with insufficient
tier 1, tier 2 and tier 3 sampling sites shall complete its sampling pool with
representative sites throughout the distribution system. For the purpose of
this Paragraph, a representative site is a site in which the plumbing
materials used at that site would be commonly found at other sites served
by the water system.
(6)
The sampling sites selected for a non-transient, non-community water
system (“tier 1 sampling sites”) shall consist of buildings that:
(i)
contain copper pipes with lead solder installed after 1982 or
contain lead pipes; and/or
(ii)
are served by a lead service line.
(7)
A non-transient, non-community water system with insufficient tier 1 sites
that meet the targeting criteria in Paragraph (a)(6) of this Section shall
77
complete its sampling pool with sampling sites that contain copper pipes
with lead solder installed before 1983. If additional sites are needed to
complete the sampling pool, the non-transient non-community water
system shall use representative sites throughout the distribution system.
For the purpose of this Paragraph, a representative site is a site in which
the plumbing materials used at that site would be commonly found at other
sites served by the water system.
(8)
Any water system whose distribution system contains lead service lines
shall draw fifty (50) percent of the samples it collects during each
monitoring period from sites that contain lead pipes, or copper pipes with
lead solder, and fifty (50) percent of the samples from sites served by a
lead service line. A water system that cannot identify a sufficient number
of sampling sites served by a lead service line shall collect first draw
samples from all of the sites identified as being served by such lines.
(b)
Sample Collection Methods
(1)
All tap samples for lead and copper collected in accordance with this
subpart, with the exception of lead service line samples collected under
Section 6.84(c) and samples collected under Paragraph (b)(5) of this
Section, shall be first draw samples.
(2)
Each first-draw tap sample for lead and copper shall be one (1) liter in
volume and have stood motionless in the plumbing system of each
sampling site for at least six (6) hours. First draw samples from residential
housing shall be collected from the cold water kitchen tap or bathroom
sink tap. First-draw samples from a non-residential building shall be one
(1) liter in volume and shall be collected at an interior tap from which
water is typically drawn for consumption. Non-first-draw samples
collected in lieu of first-draw samples pursuant to Paragraph (b)(5) of this
Section shall be one (1) liter in volume and shall be collected at an interior
tap from which water is typically drawn for consumption. First draw
samples may be collected by the system or the system may allow residents
to collect first draw samples after instructing the residents of the sampling
procedures specified in this Paragraph. To avoid problems of residents
handling nitric acid, acidification of first draw samples may be done up to
fourteen (14) days after the sample is collected. After acidification to
resolubilize the metals, the sample must stand in the original container for
the time specified in the approved EPA method before the sample can be
analyzed. If a system allows residents to perform sampling, the system
may not challenge, based on alleged errors in sample collection, the
accuracy of sampling results.
78
(3)
Each service line sample shall be one (1) liter in volume and have stood
motionless in the lead service line for at least six (6) hours. Lead service
line samples shall be collected in one (1) of the following three (3) ways:
(i)
at the tap after flushing the volume of water between the tap and
the lead service line. The volume of water shall be calculated based
on the interior diameter and length of the pipe between the tap and
the lead service line;
(ii)
tapping directly into the lead service line; or
(iii)
if the sampling site is a building constructed as a single-family
residence, allowing the water to run until there is a significant
change in temperature which would be indicative of water that has
been standing in the lead service line.
(4)
A water system shall collect each first draw tap sample from the same
sampling site from which it collected a previous sample. If, for any reason,
the water system cannot gain entry to a sampling site in order to collect a
follow-up tap sample, the system may collect the follow-up tap sample
from another sampling site in its sampling pool as long as the new site
meets the same targeting criteria, and is within reasonable proximity of the
original site.
(5)
A non-transient non-community water system, or a community water
system that meets the criteria of Sections 6.85(c)(7)(i) and (ii), that does
not have enough taps that can supply first-draw samples, as defined in
Section 1.24, may apply to the Director in writing to substitute non-first
draw samples. Such systems must collect as many first draw samples from
appropriate taps as possible and identify sampling times and locations that
would likely result in the longest standing time for the remaining sites.
The Director has the discretion to waive the requirement for prior Director
approval of non-first draw sample sites selected by the system, either
through State regulation or written notification to the system.
(c)
Number of Samples
Water systems shall collect at least one (1) sample during each monitoring period
specified in Paragraph (d) of this Section from the number of sites listed in the
first column (“standard monitoring”) of the table in this Paragraph. A system
conducting reduced monitoring under Paragraph (d)(4) of this Section shall
collect at least one (1) sample from the number of sites specified in the second
column (“reduced monitoring”) of the table in this Paragraph during each
monitoring period specified in Paragraph (d)(4) of this Section. Such reduced
monitoring sites shall be representative of the sites required for standard
79
monitoring. The Director may specify sampling locations when a system is
conducting reduced monitoring. The table is as follows:
System Size (# of People Served)
Number of Sites (Standard
Monitoring)
Number of Sites (Reduced
Monitoring)
>100,000
100
50
10,001-100,000
60
30
3,301 to 10,000
40
20
501 to 3,300
20
10
101 to 500
10
5
<100
5
5
(d)
Timing of Monitoring
(1)
Initial Tap Sampling
The first six-month monitoring period for small, medium-size and large
systems shall begin on the following dates:
System Size (# People Served)
First Six-Month Monitoring Period Begins On
>50,000
January 1, 1992
3,301 to 50,000
July 1, 1992
<3,300
July 1, 1993
(i)
All large systems shall monitor during two (2) consecutive six-
month periods.
(ii)
All small and medium-size systems shall monitor during each six-
month monitoring period until:
(A)
the system exceeds the lead or copper action level and is
therefore required to implement the corrosion control
treatment requirements under Section 6.81, in which case
the system shall continue monitoring in accordance with
Paragraph (d)(2) of this Section, or
(B)
the system meets the lead and copper action levels during
two (2) consecutive six-month monitoring periods, in
which case the system may reduce monitoring in
accordance with Paragraph (d)(4) of this Section.
(2)
Monitoring after Installation of Corrosion Control and Source Water
Treatment
80
(i)
Any large system which installs optimal corrosion control
treatment pursuant to Section 6.81(d)(4) shall monitor during two
(2) consecutive six-month monitoring periods by the date specified
in Section 6.81(d)(5).
(ii)
Any small or medium-size system which installs optimal corrosion
control treatment pursuant to Section 6.81(e)(5) shall monitor
during two (2) consecutive six-month monitoring periods by the
date specified in Section 6.81(e)(6).
(iii)
Any system which installs source water treatment pursuant to
Section 6.83(a)(3) shall monitor during two (2) consecutive six-
month monitoring periods by the date specified in Section
6.83(a)(4).
(3)
Monitoring after the Director Specifies Water Quality Parameter Values
for Optimal Corrosion Control
After the Director specifies the values for water quality control parameters
under Section 6.82(f), the system shall monitor during each subsequent
six-month monitoring period, with the first monitoring period to begin on
the date the Director specifies the optimal values under Section 6.82(f).
(4)
Reduced Monitoring
(i)
A small or medium-size water system that meets the lead and
copper action levels during each of two (2) consecutive six-month
monitoring periods may reduce the number of samples in
accordance with Paragraph (c) of this Section, and reduce the
frequency of sampling to once per year.
(ii)
Any water system that maintains the range of values for the water
quality control parameters reflecting optimal corrosion control
treatment specified by the Director under Section 6.82(f) during
each of two (2) consecutive six-month monitoring periods may
reduce the frequency of monitoring to once per year and reduce the
number of lead and copper samples in accordance with Paragraph
(c) of this Section if it receives written approval from the Director.
The Director shall review monitoring, treatment and other relevant
information submitted by the water system in accordance with
Section 6.90, and shall notify the system in writing when the
Director determines the system is eligible to commence reduced
monitoring pursuant to this Paragraph. The Director shall review,
and where appropriate, revise its determination when the system
submits new monitoring or treatment data, or when other data
81
relevant to the number and frequency of tap sampling becomes
available.
(iii)
A small or medium-size water system that meets the lead and
copper action levels during three (3) consecutive years of
monitoring may reduce the frequency of monitoring for lead and
copper from annually to once every three (3) years. Any water
system that maintains the range of values for the water quality
control parameters reflecting optimal corrosion control treatment
specified by the Director under Section 6.82(f) during three (3)
consecutive years of monitoring may reduce the frequency of
monitoring from annually to once every three (3) years if it
receives written approval from the Director. The Director shall
review monitoring, treatment, and other relevant information
submitted by the water system in accordance with Section 6.90,
and shall notify the system in writing, when the Director
determines the system is eligible to reduce the frequency of
monitoring to once every three (3) years. The Director shall
review, and where appropriate, revise his determination when the
system submits new monitoring or treatment data, or when other
data relevant to the number and frequency of tap sampling
becomes available.
(iv)
A water system that reduces the number and frequency of sampling
shall collect these samples from representative sites included in the
pool of targeted sampling sites identified in Paragraph (a) of this
Section. Systems sampling annually or less frequently shall
conduct the lead and copper tap sampling during the months of
June, July, August or September unless the Director has approved
a different sampling period in accordance with Paragraph
(d)(4)(iv)(A) of this Section.
(A)
The Director, at his or her discretion, may approve a
different period for conducting the lead and copper tap
sampling for systems collecting a reduced number of
samples. Such a period shall be no longer than four (4)
consecutive months and must represent a time of normal
operation where the highest levels of lead are most likely to
occur. For a non-transient, non-community water system
that does not operate during the months of June through
September, and for which the period of normal operation
where the highest levels of lead are most likely to occur is
not known, the Director shall designate a period that
represents a time of normal operation for the system.
82
Systems monitoring annually, that have been collecting
samples during the months of June through September and
that receive the Director’s approval to alter their sample
collection period under Paragraph (d)(4)(iv)(A) of this
Section, must collect their next round of samples during a
time period that ends no later than 21 months after the
previous round of sampling. Systems monitoring triennially
that have been collecting samples during the months of
June through September, and receive the Director’s
approval to alter the sampling collection period as per
Paragraph (d)(4)(iv)(A) of this Section, must collect their
next round of samples during a time period that ends no
later than 45 months after the previous round of sampling.
Subsequent rounds of sampling must be collected annually
or triennially, as required by this Section.
(v)
Any water system that demonstrates for two (2) consecutive 6-
month monitoring periods that the tap water lead level computed
under Section 6.80(c)(3) is less than or equal to 0.005 mg/L and
the tap water copper level computed under Section 6.80(c)(3) is
less than or equal to 0.65 mg/L may reduce the number of samples
in accordance with Paragraph (c) of this Section and reduce the
frequency of sampling to once every three (3) calendar years.
(vi)(A)
A small or medium-size water system subject to reduced
monitoring that exceeds the lead or copper action level
shall resume sampling in accordance Paragraph (c) of this
Section and collect the number of samples specified for
standard monitoring under Paragraph (c) of this Section.
Such a system shall also conduct water quality parameter
monitoring in accordance with 6.87(b), (c) or (d) (as
appropriate) during the monitoring period in which it
exceeded the action level. Any such system may resume
annual monitoring for lead and copper at the tap at the
reduced number of sites specified in Paragraph (c) of this
Section after it has completed two (2) subsequent
consecutive six-month rounds of monitoring that meet the
criteria of Paragraph (d)(4)(i) of this Section and/or may
resume triennial monitoring for lead and copper at the
reduced number of sites after it demonstrates through
subsequent rounds of monitoring that it meets the criteria of
either Paragraph (d)(4)(iii) or (d)(4)(v) of this Section.
(B)
Any water system subject to the reduced monitoring
frequency that fails to operate at or above the minimum
83
value or within the range of values for the water quality
parameters specified by the Director under Section 6.82(f)
for more than nine (9) days in any six-month period
specified in Section 6.87(d) shall conduct tap water
sampling for lead and copper at the frequency specified in
Paragraph (d)(3) of this Section, collect the number of
samples specified for standard monitoring under Paragraph
(c) of this Section, and shall resume monitoring for water
quality parameters within the distribution system in
accordance with Section 6.87(d). Such a system may
resume reduced monitoring for lead and copper at the tap
and for water quality parameters within the distribution
system under the following conditions:
(1)
The system may resume annual monitoring for lead
and copper at the tap at the reduced number of sites
specified in Paragraph (c) of this Section after it has
completed two (2) subsequent six-month rounds of
monitoring that meet the criteria of Paragraph
(d)(4)(ii) of this Section and the system has received
written approval from the Director that it is
appropriate to resume reduced monitoring on an
annual frequency.
(2)
The system may resume triennial monitoring for
lead and copper at the tap at the reduced number of
sites after it demonstrates through subsequent
rounds of monitoring that it meets the criteria of
either Paragraph (d)(4)(iii) or (d)(4)(v) of this
Section and the system has received written
approval from the Director that it is appropriate to
resume triennial monitoring.
(3)
The system may reduce the number of water quality
parameter tap water samples required in accordance
with Section 6.87(e)(1) and the frequency with
which it collects such samples in accordance with
Section 6.87(e)(2). Such a system may not resume
triennial monitoring for water quality parameters at
the tap until it demonstrates, in accordance with the
requirements of Section 6.87(e)(2), that it has re-
qualified for triennial monitoring.
(vii) Any water system subject to a reduced monitoring frequency under
Paragraph (d)(4) of this Section that either adds a new source of
water or changes any water treatment shall inform the Director in
84
writing in accordance with Section 6.90(a)(3). The Director may
require the system to resume sampling in accordance with
Paragraph (d)(3) of this Section and collect the number of samples
specified for standard monitoring under Paragraph (c) of this
Section or take other appropriate steps such as increased water
quality parameter monitoring or re-evaluation of its corrosion
control treatment given the potentially different water quality
considerations.
(e)
Additional Monitoring by Systems
The results of any monitoring conducted in addition to the minimum requirements
of this Section shall be considered by the system and the Director in making any
determinations (i.e., calculating the 90th percentile lead or copper level) under this
subpart.
(f)
Invalidation of Lead or Copper Tap Water Samples
A sample invalidated under this Paragraph does not count toward determining
lead or copper 90th percentile levels under Section 6.80(c)(3) or toward meeting
the minimum monitoring requirements of Paragraph (c) of this Section.
(1)
The Director may invalidate a lead or copper tap water sample if at least
one (1) of the following conditions is met:
(i)
The laboratory establishes that improper sample analysis caused
erroneous results;
(ii)
The Director determines that the sample was taken from a site that
did not meet the site selection criteria of this Section;
(iii)
The sample container was damaged in transit; or
(iv)
There is substantial reason to believe that the sample was subject
to tampering.
(2)
The system must report the results of all samples to the Director and all
supporting documentation for samples the system believes should be
invalidated.
(3)
To invalidate a sample under Paragraph (f)(l) of this Section, the decision
and the rationale for the decision must be documented in writing. The
Director shall not invalidate a sample solely on the grounds that a follow-
up sample result is higher or lower than that of the original sample.
85
(4)
The water system must collect replacement samples for any samples
invalidated under this Section if, after the invalidation of one (1) or more
samples, the system has too few samples to meet the minimum
requirements of Paragraph (c) of this Section. Any such replacement
samples must be taken as soon as possible, but no later than 20 days after
the date the Director invalidates the sample or by the end of the applicable
monitoring period, whichever occurs later. Replacement samples taken
after the end of the applicable monitoring period shall not also be used to
meet the monitoring requirements of a subsequent monitoring period. The
replacement samples shall be taken at the same locations as the invalidated
samples or, if that is not possible, at locations other than those already
used for sampling during the monitoring period.
6.87
Monitoring Requirements for Water Quality Parameters
All large water systems, and all small and medium-size systems that exceed the lead or copper
action level shall monitor water quality parameters in addition to lead and copper in accordance
with this Section. The requirements of this Section are summarized in the table at the end of this
Section.
(a)
General Requirements
(1)
Sample Collection Methods
(i)
Tap samples shall be representative of water quality throughout the
distribution system taking into account the number of persons
served, the different sources of water, the different treatment
methods employed by the system and seasonal variability. Tap
sampling under this Section is not required to be conducted at taps
targeted for lead and copper sampling under Section 6.86(a).
[Note: Systems may find it convenient to conduct tap sampling for
water quality parameters at sites used for coliform sampling.]
(ii)
Samples collected at the entry point(s) to the distribution system
shall be from locations representative of each source after
treatment. If a system draws water from more than one (1) source
and the sources are combined before distribution, the system must
sample at an entry point to the distribution system during periods
of normal operating conditions (i.e., when water is representative
of all sources being used).
(2)
Number of Samples
(i)
Systems shall collect two (2) tap samples for applicable water
quality parameters during each monitoring period specified under
86
Paragraphs (b) - (e) of this Section from the following number of
sites.
System Size
(# People Served)
# of sites for Water Quality Parameters
>100,000
25
10,001-100,000
10
3,301 to 10,000
3
501 to 3,300
2
101 to 500
1
<100
1
(ii)
Except as provided in Paragraph (c)(3) of this Section, systems
shall collect two (2) samples for each applicable water quality
parameter at each entry point to the distribution system during each
monitoring period specified in Paragraph (b) of this Section.
During each monitoring period specified in Paragraphs (c)-(e) of
this Section, systems shall collect one (1) sample for each
applicable water quality parameter at each entry point to the
distribution system.
(b)
Initial Sampling
All large water systems shall measure the applicable water quality parameters as
specified below at taps and at each entry point to the distribution system during each six-
month monitoring period specified in Section 6.86(d)(1). All small and medium-size
systems shall measure the applicable water quality parameters at the locations specified
below during each six-month monitoring period specified in Section 6.86(d)(1) during
which the system exceeds the lead or copper action level.
(1)
At taps:
(i)
pH;
(ii)
alkalinity;
(iii)
orthophosphate, when an inhibitor containing a phosphate
compound is used;
(iv)
silica, when an inhibitor containing a silicate compound is used;
(v)
calcium;
(vi)
conductivity; and
87
(vii)
water temperature.
(2)
At each entry point to the distribution system: all of the applicable
parameters listed in Paragraph (b)(1) above.
(c)
Monitoring after Installation of Corrosion Control
Any large system which installs optimal corrosion control treatment pursuant to Section
6.81(d)(4) shall measure the water quality parameters at the locations and frequencies
specified below during each six-month monitoring period specified in Section
6.86(d)(2)(i). Any small or medium-size system which installs optimal corrosion control
treatment shall conduct such monitoring during each six-month monitoring period
specified in Section 6.86(d)(2)(ii) in which the system exceeds the lead or copper action
level.
(1)
At taps, two (2) samples for:
(i)
pH;
(ii)
alkalinity;
(iii)
orthophosphate, when an inhibitor containing a phosphate
compound is used;
(iv)
silica, when an inhibitor containing a silicate compound is used;
and
(v)
calcium, when calcium carbonate stabilization is used as part of
corrosion control.
(2)
Except as provided in Paragraph (c)(3) of this Section, at each entry point
to the distribution system, at least one (1) sample no less frequently than
every two (2) weeks (bi-weekly) for:
(i)
pH;
(ii)
when alkalinity is adjusted as part of optimal corrosion control, a
reading of the dosage rate of the chemical used to adjust alkalinity,
and the alkalinity concentration; and
(iii)
when a corrosion inhibitor is used as part of optimal corrosion
control, a reading of the dosage rate of the inhibitor used, and the
concentration of orthophosphate or silica (whichever is applicable).
(3)
Any ground water system can limit entry point sampling described in
Paragraph (c)(2) of this Section to those entry points that are
88
representative of water quality and treatment conditions throughout the
system. If water from untreated ground water sources mixes with water
from treated ground water sources, the system must monitor for water
quality parameters both at representative entry points receiving treatment
and representative entry points receiving no treatment. Prior to the start of
any monitoring under this Paragraph, the system shall provide to the
Director written information identifying the selected entry points and
documentation, including information on seasonal variability, sufficient to
demonstrate that the sites are representative of water quality and treatment
conditions throughout the system.
(d)
Monitoring after the Director Specifies Water Quality Parameter Values for
Optimal Corrosion Control
After the Director specifies the values for applicable water quality control
parameters reflecting optimal corrosion control treatment under Section 6.82(f),
all large systems shall measure the applicable water quality parameters in
accordance with Paragraph (c) of this Section and determine compliance with the
requirements of Section 6.82(g) every six (6) months with the first six-month
period to begin on the date the Director specifies the optimal values under Section
6.82(f). Any small or medium-size system shall conduct such monitoring during
each six-month period specified in this Paragraph in which the system exceeds the
lead or copper action level. For any such small and medium-size system that is
subject to a reduced monitoring frequency pursuant to Section 6.86(d)(4) at the
time of the action level exceedance, the end of the applicable six-month period
under this Paragraph shall coincide with the end of the applicable monitoring
period under Section 6.86(d)(4). Compliance with Director-designated optimal
water quality parameter values shall be determined as specified under Section
6.82(g).
(e)
Reduced Monitoring
Any water system that maintains the range of values for the water quality
parameters reflecting optimal corrosion control treatment during each of two (2)
consecutive six-month monitoring periods under Paragraph (d) of this Section
shall continue monitoring at the entry point(s) to the distribution system as
specified in Paragraph (c)(2) of this Section. Such system may collect two (2) tap
samples for applicable water quality parameters from the following reduced
number of sites during each six-month monitoring period.
System Size (# People Served)
Reduced # of Sites for Water Quality Parameters
>100,000
10
10,001 to 100,000
7
3,301 to 10,000
3
501 to 3,300
2
101 to 500
1
89
<100
1
(2)
(i)
Any water system that maintains the range of values for the water quality
parameters reflecting optimal corrosion control treatment specified by the
Director under Section 6.82(f) during three (3) consecutive years of
monitoring may reduce the frequency with which it collects the number of
tap samples for applicable water quality parameters specified in Paragraph
(e)(1) from every six (6) months to annually. Any water system that
maintains the range of values for the water quality parameters reflecting
optimal corrosion control treatment specified by the State under Section
6.82(f) during three (3) consecutive years of annual monitoring under this
Paragraph may reduce the frequency with which it collects the number of
tap samples for applicable water quality parameters specified in Paragraph
(e)(1) from annually to every three (3) years.
(ii)
A water system may reduce the frequency with which it collects tap
samples for applicable water quality parameters specified in Paragraph
(e)(1) of this Section to every three (3) years if it demonstrates during two
(2) consecutive monitoring periods that its tap water lead level at the 90th
percentile is less than or equal to the PQL for lead specified in Section
6.89 (a)(1)(ii), that its tap water copper level at the 90th percentile is less
than or equal to 0.65 mg/L for copper in Section 6.80(c)(2), and that it also
has maintained the range of values for the water quality parameters
reflecting optimal corrosion control treatment specified by the Director
under Section 6.82(f).
(3)
A water system that conducts sampling annually shall collect these samples
evenly throughout the year so as to reflect seasonal variability.
(4)
Any water system subject to reduced monitoring frequency that fails to operate at
or above the minimum value or within the range of values for the water quality
parameters specified by the Director under Section 6.82(f) for more than nine (9)
days in any six-month period specified in Section 6.82(g) shall resume
distribution system tap water sampling in accordance with the number and
frequency requirements in Paragraph (d) of this Section. Such a system may
resume annual monitoring for water quality parameters at the tap at the reduced
number of sites specified in Paragraph (e)(1) of this Section after it has completed
two (2) subsequent consecutive six-month rounds of monitoring that meet the
criteria of that Paragraph and/or may resume triennial monitoring for water
quality parameters at the tap at the reduced number of sites after it demonstrates
through subsequent rounds of monitoring that it meets the criteria of either
Paragraph (e)(2)(i) or (e)(2)(ii) of this Section.
(f)
Additional Monitoring by Systems
90
The results of any monitoring conducted in addition to the minimum requirements of this
Section shall be considered by the system and the Director in making any determinations
(i.e., determining concentrations of water quality parameters) under this Section or
Section 6.82.
91
Summary of monitoring requirements for water quality parameters1
Monitoring Period
Parameters2
Location
Frequency
Initial Monitoring
pH, alkalinity, orthophosphate or silica3, calcium,
conductivity, temperature
Taps and at entry
point(s) to
distribution
system
Every 6
months
After Installation of
Corrosion Control
pH, alkalinity, orthophosphate or silica3, calcium4
Taps
Every 6
months
pH, alkalinity dosage rate and concentration (if
alkalinity adjusted as part of corrosion control),
inhibitor dosage rate and inhibitor residual5
Entry point(s) to
distribution
system
No less
frequently
than every
two (2)
weeks
After Director
Specifies Parameter
Values For Optimal
Corrosion
Control
pH, alkalinity, orthophosphate or silica3, calcium4
Taps
Every 6
months
pH, alkalinity dosage rate and concentration (if
alkalinity adjusted as part of corrosion control),
inhibitor dosage rate and inhibitor residual5
Entry point(s) to
distribution
system
No less
frequently
than every
two (2)
weeks
Reduced Monitoring
pH, alkalinity, orthophosphate or silica3, calcium4
Taps
Every 6
months,
annually7 or
every 3
years8
reduced
number of
sites
pH, alkalinity dosage rate and concentration (if
alkalinity adjusted as part of corrosion control),
inhibitor dosage rate and inhibitor residual5
Entry point(s) to
distribution
system
No less
frequently
than every
two (2)
weeks
1 Table is for illustrative purposes; consult the text of this Section for precise regulatory requirements.
2 Small and medium-size systems have to monitor for water quality parameters only during monitoring periods in
which the system exceeds the lead or copper action level.
3 Orthophosphate must be measured only when an inhibitor containing a phosphate compound is used. Silica must
be measured only when an inhibitor containing silicate compound is used.
4 Calcium must be measured only when calcium carbonate stabilization is used as part of corrosion control.
5 Inhibitor dosage rates and inhibitor residual concentrations (orthophosphate or silica) must be measured only when
an inhibitor is used.
6 Ground water systems may limit monitoring to representative locations throughout the system.
92
7 Water systems may reduce frequency of monitoring for water quality parameters at the tap from every six (6)
months to annually if they have maintained the range of values for water quality parameters reflecting optimal
corrosion control during 3 consecutive years of monitoring.
8 Water systems may further reduce the frequency of monitoring for water quality parameters at the tap from
annually to once every 3 years if they have maintained the range of values for water quality parameters reflecting
optimal corrosion control during 3 consecutive years of annual monitoring. Water systems may accelerate to
triennial monitoring for water quality parameters at the tap if they have maintained 90th percentile lead levels less
than or equal to 0.005 mg/L, 90th percentile copper levels less than or equal to 0.65 mg/L, and the range of water
quality parameters designated by the Director under Section 6.82(f) as representing optimal corrosion control during
two (2) consecutive six-month monitoring periods.
6.88 Monitoring Requirements for Lead and Copper in Source Water
(a)
Sample Location, Collection Methods and Number of Samples
(1)
A water system that fails to meet the lead or copper action level on the
basis of tap samples collected in accordance with Section 6.86 shall collect
lead and copper source water samples in accordance with the following
requirements regarding sample location, number of samples, and
collection methods:
(i)
Groundwater systems shall take a minimum of one (1) sample at
every entry point to the distribution system which is representative
of each well after treatment (hereafter called a sampling point).
The system shall take one (1) sample at the same sampling point
unless conditions make another sampling point more representative
of each source or treatment plant.
(ii)
Surface water systems shall take a minimum of one (1) sample at
every entry point to the distribution system after any application of
treatment, or in the distribution system at a point which is
representative of each source after treatment (hereafter called a
sampling point). The system shall take each sample at the same
sampling point unless conditions make another sampling point
more representative of each source or treatment plant. NOTE: For
the purposes of this Paragraph, surface water systems include
systems with a combination of surface and ground sources.
(iii)
If a system draws water from more than one (1) source and the
sources are combined before distribution, the system must sample
at an entry point to the distribution system during periods of
normal operating conditions (i.e., when water is representative of
all sources being used).
(iv)
The Director may reduce the total number of samples which must
be analyzed by allowing the use of compositing. Compositing of
93
samples must be done by certified laboratory personnel. Composite
samples from a maximum of five (5) samples are allowed,
provided that if the lead concentration in the composite sample is
greater than or equal to 0.001 mg/L or the copper concentration is
greater than or equal to 0.160 mg/L, then either:
(A)
A follow-up sample shall be taken and analyzed within 14
days at each sampling point included in the composite; or
(B)
If duplicates of, or sufficient quantities from, the original
samples from each sampling point used in the composite
are available, the system may use these instead of
resampling.
(2)
Where the results of sampling indicate an exceedance of maximum
permissible source water levels established under Section 6.83(b)(4), the
Director may require that one (1) additional sample be collected as soon as
possible after the initial sample was taken (but not to exceed two (2)
weeks) at the same sampling point. If a Director-required confirmation
sample is taken for lead or copper, then the results of the initial and
confirmation sample shall be averaged in determining compliance with the
Director-specified maximum permissible levels. Any sample value below
the detection limit shall be considered to be zero. Any value above the
detection limit but below the PQL shall either be considered as the
measured value or be considered one-half the PQL.
(b)
Monitoring Frequency after System Exceeds Tap Water Action Level
Any system which exceeds the lead or copper action level at the tap shall collect
one (1) source water sample from each entry point to the distribution system
within six (6) months after the exceedance.
(c)
Monitoring Frequency after Installation of Source Water Treatment
Any system which installs source water treatment pursuant to Section 6.83(a)(3)
shall collect an additional source water sample from each entry point to the
distribution system during two (2) consecutive six-month monitoring periods by
the deadline specified in Section 6.83(a)(4).
(d)
Monitoring Frequency after the Director Specifies Maximum Permissible Source
Water Levels or Determines That Source Water Treatment Is Not Needed
(1)
A system shall monitor at the frequency specified below in cases where
the Director specifies maximum permissible source water levels under
Section 6.83(b)(4), or determines that the system is not required to install
source water treatment under Section 6.83(b)(2).
94
(i)
A water system using only groundwater shall collect samples once
during the three-year compliance period (as that term is defined in
Section 1) in effect when the applicable Director determination
under Paragraph (d)(1) of this Section is made. Such systems shall
collect samples once during each subsequent compliance period.
(ii)
A water system using surface water (or a combination of surface
and groundwater) shall collect samples once during each year, the
first annual monitoring period to begin on the date on which the
applicable Director determination is made under Paragraph (d)(1)
of this Section.
(2)
A system is not required to conduct source water sampling for lead and/or
copper if the system meets the action level for the specific contaminant in
tap water samples during the entire source water sampling period
applicable to the system under Paragraph (d)(1)(i) or (d)(1)(ii) of this
Section.
(e)
Reduced Monitoring Frequency
(1)
A water system using only groundwater may reduce the monitoring
frequency for lead and/or copper in source water to once during each nine-
year compliance cycle (as that term is defined in Section 1) if the system
meets one (1) of the following criteria:
(i)
The system demonstrates that finished drinking water entering the
distribution system has been maintained below the maximum
permissible lead and copper concentrations specified by the
Director in Section 6.83(b)(4) during at least three (3) consecutive
compliance periods under Paragraph (d)(1) of this Section; or
(ii)
The Director has determined that source water treatment is not
needed and the system demonstrates that, during at least three (3)
consecutive compliance periods in which sampling was conducted
under Paragraph (d)(1) of this Section, the concentration of lead in
source water was less than or equal to 0.005 mg/L and the
concentration of copper in source water was less than or equal to
0.65 mg/L.
(2)
A water system using surface water (or a combination of surface water and
ground water)may reduce the monitoring frequency in Paragraph (d)(1) of
this Section to once during each nine-year compliance cycle (as that term
is defined in Section 1) if the system meets one (1) of the following
criteria:
95
(i)
The system demonstrates that finished drinking water entering the
distribution system has been maintained below the maximum
permissible lead and copper concentrations specified by the
Director in Section 6.83(b)(4) for at least three (3) consecutive
years; or
(ii)
The Director has determined that source water treatment is not
needed and the system demonstrates that, during at least three (3)
consecutive years, the concentration of lead in source water was
less than or equal to 0.005 mg/L and the concentration of copper in
source water was less than or equal to 0.65 mg/L.
(3)
A water system that uses a new source of water is not eligible for reduced
monitoring for lead and/or copper until concentrations in samples
collected from the new source during three (3) consecutive monitoring
periods are below the maximum permissible lead and copper
concentrations specified by the Director in Section 6.83(a)(5).
6.89 Analytical Methods
(a)
Analyses for lead, copper, pH, conductivity, calcium, alkalinity, orthophosphate,
silica and temperature shall be conducted with the methods in Appendix 1.
(1)
Analyses under this Section shall only be conducted by certified
laboratories using the methods specified in Appendix 1. Analyses for pH
and temperature shall be conducted using methods specified in Appendix
1 by a party approved by the Director. To obtain certification to conduct
analyses for lead and copper, laboratories must:
(i)
Analyze performance evaluation samples which include lead and
copper provided by a third party accredited provider or equivalent
samples provided by the State; and
(ii)
Achieve quantitative acceptance limits as follows:
(A)
For lead: + 30 percent of the actual amount in the
Performance Evaluation sample when the actual amount is
greater than or equal to 0.005 mg/L. The Practical
Quantitation Level, or PQL, for lead is 0.005 mg/L.
(B)
For copper: + 10 percent of the actual amount in the
Performance Evaluation sample when the actual amount is
greater than or equal to 0.050 mg/L. The PQL for copper is
0.050 mg/L.
96
(iii)
Achieve the method detection limit for lead of 0.001 mg/L
according to the procedures in appendix B of part 136 of 40 CFR.
This need only be accomplished if the laboratory will be
processing source water composite samples under Section
6.88(a)(1)(iii).
(iv)
Be currently certified by EPA or the State to perform analyses to
the specifications described in Paragraph (a)(2) of this Section.
(2)
The Director may allow the use of previously collected monitoring data
for purposes of monitoring, if the data were collected and analyzed in
accordance with the requirements of this Section.
(3)
All lead and copper levels measured between the PQL and the MDL must
be either reported as measured or they can be reported as one-half the PQL
specified for lead and copper in Appendix 1. All levels below the lead and
copper MDLs must be reported as zero.
(4)
All copper levels measured between the PQL and the MDL must be either
reported as measured or they can be reported as one-half the PQL (0.025
mg/L). All levels below the copper MDL must be reported as zero.
(b)
[Reserved]
6.90 Reporting Requirements
All water systems shall report all of the following information to the Director in accordance with
this Section.
(a)
Reporting Requirements for Tap Water Monitoring for Lead and Copper and for
Water Quality Parameter Monitoring
(1)
Except as provided in Paragraph (a)(1)(viii) of this Section, a water system
shall report the information specified below for all tap water samples
specified in Section 6.86 and for all water quality parameter samples
specified in Section 6.87 within the first 10 days following the end of each
applicable monitoring period specified in Section 6.86 and Section 6.87
(i.e., every six-months, annually, every 3 years, or every 9 years).
(i)
the results of all tap samples for lead and copper including the
location
of
each
site
and
the
criteria
under
Section
6.86(a)(3),(4),(5),(6) and/or (7) under which the site was selected
for the system's sampling pool;
97
(ii)
documentation for each tap water lead or copper sample for which
the water system requests invalidation pursuant to Section
6.86(f)(2);
(iii)
[Reserved];
(iv)
the 90th percentile lead and copper concentrations measured from
among all lead and copper tap water samples collected during each
monitoring period (calculated in accordance with Section
6.80(c)(3)) unless the Director calculates the system’s 90th
percentile lead and copper levels under Paragraph (h) of this
Section;
(v)
with the exception of initial tap sampling conducted pursuant to
Section 6.86(d)(1), the system shall designate any site which was
not sampled during previous monitoring periods, and include an
explanation of why sampling sites have changed;
(vi)
the results of all tap samples for pH, and where applicable,
alkalinity, calcium, conductivity, temperature, and orthophosphate
or silica collected under Section 6.87(b)-(e);
the results of all samples collected at the entry point(s) to the
distribution system for applicable water quality parameters under
Section 6.87(b)-(e).
(viii) A water system shall report the results of all water quality
parameter samples collected under Section 6.87(c)-(f) during each
six-month monitoring period specified in Section 6.87(d) within
the first 10 days following the end of the monitoring period unless
the Director has specified a more frequent reporting requirement.
(2)
For a non-transient non-community water system, or a community water
system meeting the criteria of Sections 6.85(c)(7)(i) and (ii), that does not
have enough taps that can provide first-draw samples, the system must
either:
(i)
Provide written documentation to the Director identifying standing
times and locations for enough non-first draw samples to make up
its sampling pool under Section 6.86(b)(5) by the start of the first
applicable monitoring period under Section 6.86(d) that
commences after April 11, 2000, unless the Director has waived
prior Director’s approval of non-first-draw sample sites selected by
the system pursuant to Section 6.86(b)(5); or
98
(ii)
If the Director has waived prior approval of non-first-draw sample
sites selected by the system, identify, in writing, each site that did
not meet the six-hour minimum standing time and the length of
standing time for that particular substitute sample collected
pursuant to Section 6.86(b)(5) and include this information with
the lead and copper tap sample results required to be submitted
pursuant to Paragraph (a)(1)(i) of this Section.
(3)
No later than 60 days after the addition of a new source or any change in
water treatment, unless the Director requires earlier notification, a water
system deemed to have optimized corrosion control under Section
6.81(b)(3), a water system subject to reduced monitoring pursuant to
Section 6.86(d)(4), or a water system subject to a monitoring waiver
pursuant to Section 6.86(g), shall send written documentation to the
Director describing the change. In those instances where prior Director’s
approval of the treatment change or new source is not required, water
systems are encouraged to provide the notification to the Director
beforehand to minimize the risk, the treatment change, or new source will
adversely affect optimal corrosion control.
(4)
Each ground water system that limits water quality parameter monitoring
to a subset of entry points under Section 6.87(c)(3) shall provide, by the
commencement of such monitoring, written correspondence to the
Director that identifies the selected entry points and includes information
sufficient to demonstrate that the sites are representative of water quality
and treatment conditions throughout the system.
(b)
Source Water Monitoring Reporting Requirements
(1)
A water system shall report the sampling results for all source water
samples collected in accordance with Section 6.88 within the first 10 days
following the end of each source water monitoring period (i.e., annually,
per compliance period, per compliance cycle) specified in Section 6.88.
(2)
With the exception of the first round of source water sampling conducted
pursuant to Section 6.88(b), the system shall specify any site which was
not sampled during previous monitoring periods, and include an
explanation of why the sampling point has changed.
(c)
Corrosion Control Treatment Reporting Requirements
By the applicable dates under Section 6.81, systems shall report the following
information:
(1)
for systems demonstrating that they have already optimized corrosion
control, information required in Section 6.81(b) (2) or (3).
99
(2)
for systems required to optimize corrosion control, their recommendation
regarding optimal corrosion control treatment under Section 6.82(a).
(3)
for systems required to evaluate the effectiveness of corrosion control
treatments under Section 6.82(c), the information required by that
Paragraph.
(4)
for systems required to install optimal corrosion control designated by the
Director under Section 6.82(d), a letter certifying that the system has
completed installing that treatment.
(d)
Source Water Treatment Reporting Requirements
By the applicable dates in Section 6.83, systems shall provide the following
information to the Director:
(1)
if required under Section 6.83(b)(1), their recommendation regarding
source water treatment;
(2)
for systems required to install source water treatment under Section
6.83(b)(2), a letter certifying that the system has completed installing the
treatment designated by the Director within 24 months after the Director
designated the treatment.
(e)
Lead Service Line Replacement Reporting Requirements
Systems shall report the following information to the Director to demonstrate
compliance with the requirements of Section 6.84:
(1)
Within 12 months after a system exceeds the lead action level in sampling
referred to in Section 6.84(a), the system shall demonstrate in writing to
the Director that it has conducted a materials evaluation, including the
evaluation in Section 6.86(a), to identify the initial number of lead service
lines in its distribution system, and shall provide the Director with the
system's schedule for replacing annually at least 7 percent of the initial
number of lead service lines in its distribution system.
(2)
Within 12 months after a system exceeds the lead action level in sampling
referred to in Section 6.84(a), and every 12 months thereafter, the system
shall demonstrate to the Director in writing that the system has either:
(i)
replaced in the previous 12 months at least 7 percent of the initial
lead service lines (or a greater number of lines specified by the
Director under Section 6.84(f)) in its distribution system, or
100
(ii)
conducted
sampling
which
demonstrates
that
the
lead
concentration in all service line samples from an individual line(s),
taken pursuant to Section 6.86(b)(3), is less than or equal to 0.015
mg/L. In such cases, the total number of lines replaced and/or
which meet the criteria in Section 6.84(c) shall equal at least 7
percent of the initial number of lead lines identified under
Paragraph (a) of this Section (or the percentage specified by the
Director under Section 6.84(f)).
(3)
The annual letter submitted to the Director under Paragraph (e)(2) of this
Section shall contain the following information:
(i)
the number of lead service lines scheduled to be replaced during
the previous year of the system's replacement schedule;
(ii)
the number and location of each lead service line replaced during
the previous year of the system's replacement schedule;
(iii)
if measured, the water lead concentration and location of each lead
service line sampled, the sampling method, and the date of
sampling.
(4)
Any system which collects lead service line samples following partial lead
service line replacement required by Section 6.84 shall report the results to
the Director within the first ten (10) days of the month following the
month in which the system receives the laboratory results, or as specified
by the Director. The Director, at his or her discretion may eliminate this
requirement to report these monitoring results. Systems shall also report
any additional information as specified by the Director, and in a time and
manner prescribed by the Director, to verify that all partial lead service
line replacement activities have taken place.
(f)
Public Education Program Reporting Requirements
(1)
Any water system that is subject to the public education requirements in
Section 6.85 shall, within ten (10) days after the end of each period in
which the system is required to perform public education tasks in
accordance with Section 6.85(c), send written documentation to the
Director that contains:
(i)
A demonstration that the system has delivered the public education
materials that meet the content requirements in Section 6.85(a) and
(b) and the delivery requirements in Section 6.85(c); and
(ii)
A list of all the newspapers, radio stations, television stations, and
facilities and organizations to which the system delivered public
101
education materials during the period in which the system was
required to perform public education tasks.
(2)
Unless required by the Director, a system that previously has submitted
the information required by Paragraph (f)(1)(ii) of this Section need not
resubmit the information required by Paragraph (f)(1)(ii) of this Section,
as long as there have been no changes in the distribution list and the
system certifies that the public education materials were distributed to the
same list submitted previously.
(g)
Reporting of Additional Monitoring Data
Any system which collects sampling data in addition to that required by this
subpart shall report the results to the Director within the first ten (10) days
following the end of the applicable monitoring period under Sections 6.86, 6.87
and Section 6.88 during which the samples are collected.
(h)
Reporting of 90th Percentile Lead and Copper Concentrations Where the Director
Calculates a System’s 90th Percentile Concentration
A water system is not required to report the 90th percentile lead and copper
concentrations measured from among all lead and copper tap water samples
collected during each monitoring period, as required by Paragraph (a)(1)(iv) of
this Section if:
(1)
The Director has previously notified the water system that it will calculate
the water system’s 90th percentile lead and copper concentrations, based
on the lead and copper tap results submitted pursuant to Paragraph
(h)(2)(i) of this Section, and has specified a date before the end of the
applicable monitoring period by which the system must provide the results
of lead and copper tap water samples;
(2)
The system has provided the following information to the Director by the
date specified in Paragraph (h)(1) of this Section:
(i)
The results of all tap samples for lead and copper including the
location of each site and the criteria under Section 6.86(a)(3), (4),
(5), (6), and/or (7) under which the site was selected for the
system’s sampling pool, pursuant to Paragraph (a)(1)(i) of this
Section; and
(ii)
An identification of sampling sites utilized during the current
monitoring period that were not sampled during previous
monitoring periods, and an explanation why sampling sites have
changed; and
102
The Director has provided the results of the 90th percentile lead and
copper calculations, in writing, to the water system before the end
of the monitoring period.
6.91 Record Keeping Requirements
Any system subject to the requirements of this Section shall retain on its premises original
records of all sampling data and analyses, reports, surveys, letters, evaluations, schedules,
Director determinations and any other information required by Section 6.81 through Section
6.88. Each water system shall retain the records required by this Section for no fewer than 12
years.
103
SECTION
7.0
DISINFECTANT
RESIDUALS,
DISINFECTION
BYPRODUCTS, AND DISINFECTION BYPRODUCT PRECURSORS
7.1
Maximum Contaminant Levels (MCLs) for Disinfection Byproducts
(a) The maximum contaminant levels (MCLs) for disinfection byproducts are as follows:
Disinfection byproduct
MCL (mg/L)
Total trihalomethanes (TTHM)
0.080
Haloacetic acids (five) (HAA5)
0.060
Bromate
0.010
Chlorite
1.0
(b)
Compliance dates.
(1)
CWSs and NTNCWSs. Subpart H systems serving 10,000 or more persons
must comply with this Section beginning January 1, 2002. Subpart H
systems serving fewer than 10,000 persons and systems using only ground
water not under the direct influence of surface water must comply with
this Section beginning January 1, 2004.
(2)
A system that is installing GAC or membrane technology to comply with
this Section may apply to the Director for an extension of up to 24 months
past the dates in Paragraphs (b)(1) of this Section, but not beyond
December 31, 2003. In granting the extension, the Director must set a
schedule for compliance and may specify any interim measures that the
system must take. Failure to meet the schedule or interim treatment
requirements constitutes a violation of these regulations.
(c)
The following are identified as the best technology, treatment techniques, or other
means available for achieving compliance with the maximum contaminant levels
for disinfection byproducts identified in Paragraph (a) of this Section:
Disinfection byproduct
Best available technology
TTHM
Enhanced coagulation or enhanced softening or GAC10, with chlorine as the primary
and residual disinfectant
HAA5
Enhanced coagulation or enhanced softening or GAC10, with chlorine as the primary
and residual disinfectant.
Bromate
Control of ozone treatment process to reduce production of bromate.
Chlorite
Control of treatment processes to reduce disinfectant demand and control of
disinfection treatment processes to reduce disinfectant levels.
104
7.2
Maximum Residual Disinfectant Levels (MRDLs)
(a)
Maximum residual disinfectant levels (MRDLs) are as follows:
Disinfectant residual
MRDL (mg/L)
Chlorine
4.0 (as Cl2)
Chloramines
4.0 (as Cl2)
Chlorine dioxide
0.8 (as ClO2)
(b)
Compliance dates.
(1)
CWSs and NTNCWSs. Subpart H systems serving 10,000 or more persons
must comply with this Section beginning January 1, 2002. Subpart H
systems serving fewer than 10,000 persons and systems using only ground
water not under the direct influence of surface water must comply with
this subpart beginning January 1, 2004.
(2)
Transient NCWSs. Subpart H systems serving 10,000 or more persons and
using chlorine dioxide as a disinfectant or oxidant must comply with the
chlorine dioxide MRDL beginning January 1, 2002. Subpart H systems
serving fewer than 10,000 persons and using chlorine dioxide as a
disinfectant or oxidant and systems using only ground water not under the
direct influence of surface water and using chlorine dioxide as a
disinfectant or oxidant must comply with the chlorine dioxide MRDL
beginning January 1, 2004.
(c)
The following are identified as the best technology, treatment techniques, or other
means available for achieving compliance with the maximum residual disinfectant
levels identified in Paragraph (a) of this Section: control of treatment processes to
reduce disinfectant demand and control of disinfection treatment processes to
reduce disinfectant levels.
7.3
General Requirements
(a)
The requirements of Section 7.0 constitute the State of Rhode Island primary
drinking water regulations.
(1)
The regulations in this Section establish criteria under which community
water systems (CWSs) and non-transient, non-community water systems
(NTNCWSs) which add a chemical disinfectant to the water in any part of
the drinking water treatment process must modify their practices to meet
MCLs and MRDLs in sections 7.1 and 7.2, respectively, and must meet
the treatment technique requirements for disinfection byproduct precursors
in Section 7.8.
105
(2)
The regulations in this Section establish criteria under which transient
NCWSs that use chlorine dioxide as a disinfectant or oxidant must modify
their practices to meet the MRDL for chlorine dioxide in Section 7.2.
(3)
The EPA has established MCLs for TTHM and HAA5 and treatment
technique requirements for disinfection byproduct precursors to limit the
levels of known and unknown disinfection byproducts which may have
adverse health effects. These disinfection byproducts may include
chloroform, bromodichloromethane, dibromochloromethane, bromoform,
dichloroacetic acid, and trichloroacetic acid.
(b)
Compliance Dates.
(1)
CWSs and NTNCWSs
Unless otherwise noted, systems must comply with the requirements of
this subpart as follows. Subpart H systems serving 10,000 or more persons
must comply with this subpart beginning January 1, 2002. Subpart H
systems serving fewer than 10,000 persons and systems using only ground
water not under the direct influence of surface water must comply with
this subpart beginning January 1, 2004.
(2)
Transient NCWSs
Subpart H systems serving 10,000 or more persons and using chlorine
dioxide as a disinfectant or oxidant must comply with any requirements
for chlorine dioxide in this subpart beginning January 1, 2002. Subpart H
systems serving fewer than 10,000 persons and using chlorine dioxide as a
disinfectant or oxidant and systems using only ground water not under the
direct influence of surface water and using chlorine dioxide as a
disinfectant or oxidant must comply with any requirements for chlorine
dioxide in this subpart beginning January 1, 2004.
(c)
Each CWS and NTNCWS regulated under Paragraph (a) of this Section must be
operated by qualified personnel who meet the requirements specified by the
Director and be certified in accordance with the Rules and Regulations Pertaining
to the Certification of Public Drinking Water Supply Transmission and
Distribution Operators (R23-65-DWQ), as amended.
(d)
Control Of Disinfectant Residuals
Notwithstanding the MRDLs in Section 7.2, systems may increase residual
disinfectant levels in the distribution system of chlorine or chloramines (but not
chlorine dioxide) to a level and for a time necessary to protect public health, to
address
specific
microbiological
contamination
problems
caused
by
106
circumstances such as, but not limited to, distribution line breaks, storm run-off
events, source water contamination events, or cross-connection events.
7.4
Analytical Requirements
Analytical requirements
(a)
General.
(1)
Systems must use only the analytical method(s) specified in this Section,
or otherwise approved by EPA for monitoring under this subpart, to
demonstrate compliance with the requirements of this Section. These
methods are effective for compliance monitoring February 16, 1999.
(2)
The following documents are incorporated by reference: The Director of
the Federal Register approves this incorporation by reference in
accordance with 5 U.S.C. 552(a) and 1 CFR part 51. Copies may be
inspected at EPA's Drinking Water Docket, 401 M St., SW., Washington,
DC 20460, or at the Office of the Federal Register, 800 North Capitol
Street, NW, Suite 700, Washington DC 20401. EPA Method 552.1 is in
Methods for the Determination of Organic Compounds in Drinking Water-
Supplement II, USEPA, August 1992, EPA/600/R-92/129 (available
through National Information Technical Service (NTIS), PB92-207703).
EPA Methods 502.2, 524.2, 551.1, and 552.2 are in Methods for the
Determination of Organic Compounds in Drinking Water-Supplement III,
USEPA, August 1995, EPA/600/R-95/131. (available through NTIS,
PB95-261616). EPA Method 300.0 is in Methods for the Determination of
Inorganic Substances in Environmental Samples, USEPA, August 1993,
EPA/600/R-93/100. (available through NTIS, PB94-121811). EPA
Method 300.1 is titled USEPA Method 300.1, Determination of Inorganic
Anions in Drinking Water by Ion Chromatography, Revision 1.0, USEPA,
1997, EPA/600/R-98/118 (available through NTIS, PB98-169196); also
available from: Chemical Exposure Research Branch, Microbiological &
Chemical Exposure Assessment Research Division, National Exposure
Research Laboratory, U.S. Environmental Protection Agency, Cincinnati,
OH 45268, Fax Number: 513-569-7757, Phone number: 513-569-7586.
Standard Methods 4500-Cl D, 4500-Cl E, 4500-Cl F, 4500-Cl G, 4500-Cl
H, 4500-Cl I, 4500-ClO2 D, 4500-ClO2 E, 6251 B, and 5910 B shall be
followed in accordance with Standard Methods for the Examination of
Water and Wastewater, 19th Edition, American Public Health Association,
1995; copies may be obtained from the American Public Health
Association, 1015 Fifteenth Street, NW, Washington, DC 20005. Standard
Methods 5310 B, 5310 C and 5310 D shall be followed in accordance with
the Supplement to the 19th Edition of Standard Methods for the
Examination of Water and Wastewater, American Public Health
Association, 1996; copies may be obtained from the American Public
107
Health Association, 1015 Fifteenth Street, NW, Washington, DC 20005.
ASTM Method D 1253-86 shall be followed in accordance with the
Annual Book of ASTM Standards, Volume 11.01, American Society for
Testing and Materials, 1996 edition; copies may be obtained from the
American Society for Testing and Materials, 100 Barr Harbor Drive, West
Conshohoken, PA 19428.
(b)
Disinfection Byproducts.
(1)
Systems must measure disinfection byproducts by the methods (as
modified by the footnotes) listed in the following table:
Approved Methods For Disinfection Byproduct Compliance Monitoring
Methodology2
EPA method
Standard
method
Byproduct measured1
TTHM
HAA5
Chlorite 4
Bromate
P&T/GC/ElCD & PID
3 502.2
X
P&T/GC/MS
524.2
X
LLE/GC/ECD
551.1
X
LLE/GC/ECD
6251 B
X
SPE/GC/ECD
552.1
X
LLE/GC/ECD
552.2
X
Amperometric Titration
4500-ClO2 E
X
IC
300.0
X
IC
300.1
X
X
1 X indicates method is approved for measuring specified disinfection byproduct.
2 P&T=purge and trap; GC=gas chromatography; ElCD=electrolytic conductivity detector; PID=photoionization
detector; MS=mass spectrometer; LLE=liquid/liquid extraction; ECD=electron capture detector; SPE=solid phase
extractor; IC=ion chromatography.
3 If TTHMs are the only analytes being measured in the sample, then a PID is not required.
4 Amperometric titration may be used for routine daily monitoring of chlorite at the entrance to the distribution
system, as prescribed in Section 7.5 (b)(2)(i)(A). Ion chromatography must be used for routine monthly monitoring
of chlorite and additional monitoring of chlorite in the distribution system, as prescribed in Section 7.5 (b)(2)(i)(B)
and (b)(2)(ii).
(2)
Analysis under this Section for disinfection byproducts must be conducted
by laboratories that have received certification by EPA or the Director,
except as specified under Paragraph (b)(3) of this Section. To receive
certification to conduct analyses for the contaminants in Section 7.1 (a),
the laboratory must carry out annual analyses of performance evaluation
(PE) samples approved by EPA or the Director. In these analyses of PE
samples, the laboratory must achieve quantitative results within the
acceptance limit on a minimum of 80% of the analytes included in each
PE sample. The acceptance limit is defined as the 95% confidence interval
calculated around the mean of the PE study data between a maximum and
minimum acceptance limit of + / - 50% and + / - 15% of the study mean.
108
(3)
A party approved by EPA or the Director must measure daily chlorite
samples at the entrance to the distribution system.
(c)
Disinfectant Residuals
(1)
Systems must measure residual disinfectant concentrations for free
chlorine, combined chlorine (chloramines), and chlorine dioxide by the
methods listed in the following table:
Approved Methods for Disinfectant Residual Compliance Monitoring
Methodology
Standard
Method
ASTM
Method
Residual Measured 1
Free chlorine
Combined
chlorine
Total
chlorine
Chlorine
dioxide
Amperometric Titration
4500-Cl D
D 1253-86
X
X
X
Low Level
Amperometric Titration
4500-Cl E
X
DPD Ferrous Titrimetric
4500-Cl F
X
X
X
DPD Colorimetric
4500-Cl G
X
X
X
Syringaldazin e
(FACTS)
4500-Cl H
X
Iodometric Electrode
4500-Cl I
X
DPD.
4500-ClO2 D
X
Amperometric Method II
4500-ClO2 E
X
1 X indicates method is approved for measuring specified disinfectant residual
(2)
If approved by the Director, systems may also measure residual
disinfectant concentrations for chlorine, chloramines and chlorine dioxide
by using DPD colorimetric test kits.
(3)
A party approved by EPA or the Director must measure residual
disinfectant concentration.
(d)
Additional Analytical Methods
Systems required to analyze parameters not included in Paragraphs (b) and (c) of
this Section must use the following methods. A party approved by EPA or the
Director must measure these parameters.
(1)
Alkalinity. All methods allowed in Appendix 1 for measuring alkalinity.
(2)
Bromide. EPA Method 300.0 or EPA Method 300.1.
(3)
Total Organic Carbon (TOC). Standard Method 5310 B (High-
Temperature Combustion Method) or Standard Method 5310 C
(Persulfate-Ultraviolet or Heated-Persulfate Oxidation Method) or
Standard Method 5310 D (Wet-Oxidation Method). TOC samples may not
109
be filtered prior to analysis. TOC samples must either be analyzed or must
be acidified to achieve pH less than 2.0 by minimal addition of phosphoric
or sulfuric acid as soon as practical after sampling, not to exceed 24 hours.
Acidified TOC samples must be analyzed within 28 days.
(4)
Specific Ultraviolet Absorbance (SUVA). SUVA is equal to the UV
absorption at 254nm (UV254) (measured in m-1) divided by the dissolved
organic carbon (DOC) concentration (measured as mg/L). In order to
determine SUVA, it is necessary to separately measure UV254 and DOC.
When determining SUVA, systems must use the methods stipulated in
Paragraph (d)(4)(i) of this Section to measure DOC and the method
stipulated in Paragraph (d)(4)(ii) of this Section to measure UV254. SUVA
must be determined on water prior to the addition of disinfectants/oxidants
by the system. DOC and UV254 samples used to determine a SUVA value
must be taken at the same time and at the same location.
(i)
Dissolved Organic Carbon (DOC). Standard Method 5310 B
(High-Temperature Combustion Method) or Standard Method
5310 C (Persulfate-Ultraviolet or Heated-Persulfate Oxidation
Method) or Standard Method 5310 D (Wet-Oxidation Method).
Prior to analysis, DOC samples must be filtered through a 0.45 m
pore-diameter filter. Water passed through the filter prior to
filtration of the sample must serve as the filtered blank. This
filtered blank must be analyzed using procedures identical to those
used for analysis of the samples and must meet the following
criteria: DOC < 0.5 mg/L. DOC samples must be filtered through
the 0.45 m pore-diameter filter prior to acidification. DOC samples
must either be analyzed or must be acidified to achieve pH less
than 2.0 by minimal addition of phosphoric or sulfuric acid as soon
as practical after sampling, not to exceed 48 hours. Acidified DOC
samples must be analyzed within 28 days.
(ii)
Ultraviolet Absorption at 254 nm (UV254). Method 5910 B
(Ultraviolet Absorption Method). UV absorption must be measured
at 253.7 nm (may be rounded off to 254 nm). Prior to analysis,
UV254 samples must be filtered through a 0.45 m pore-diameter
filter. The pH of UV254 samples may not be adjusted. Samples
must be analyzed as soon as practical after sampling, not to exceed
48 hours.
(5)
pH. All methods allowed in Appendix 1 for measuring pH.
7.5
Monitoring Requirements
(a)
General Requirements
110
(1)
Systems must take all samples during normal operating conditions.
(2)
Systems may consider multiple wells drawing water from a single aquifer
as one (1) treatment plant for determining the minimum number of TTHM
and HAA5 samples required, with approval of the Director in accordance
with criteria developed by the Director.
(3)
Failure to monitor in accordance with the monitoring plan required under
Paragraph (f) of this Section is a monitoring violation.
(4)
Failure to monitor will be treated as a violation for the entire period
covered by the annual average where compliance is based on a running
annual average of monthly or quarterly samples or averages and the
system's failure to monitor makes it impossible to determine compliance
with MCLs or MRDLs.
(5)
Systems may use only data collected under the provisions of this Section
subpart or the Information Collection Rule known as subpart M of 40 CFR
141 to qualify for reduced monitoring.
(b)
Monitoring requirements for disinfection byproducts
(1)
TTHMs and HAA5
(i)
Routine monitoring.
Systems must monitor at the frequency indicated in the following
table:
Routine Monitoring Frequency for TTHM and HAA5
Type Of System
Minimum Monitoring Frequency
Sample Location In The Distribution System
Subpart H system
serving at least 10,000
persons.
Four (4) water samples per
quarter per treatment plant.
At least 25 percent of all samples collected each
quarter at locations representing maximum
residence time. Remaining samples taken at
locations representative of at least average
residence time in the distribution system and
representing the entire distribution system, taking
into account number of persons served, different
sources of water, and different treatment methods.1
Subpart H system
serving from 500 to
9,999 persons.
One (1) water sample per quarter
per treatment plant.
Locations representing maximum residence time.1
Subpart H system
serving fewer than 500
persons.
One (1) sample per year per
treatment plant during month of
warmest water temperature.
Locations representing maximum residence time.1
If the sample (or average of annual samples, if
more than one (1) sample is taken) exceeds the
MCL, the system must increase monitoring to one
(1) sample per treatment plant per quarter, taken at
a point reflecting the maximum residence time in
the distribution system, until the system meets
111
Routine Monitoring Frequency for TTHM and HAA5
Type Of System
Minimum Monitoring Frequency
Sample Location In The Distribution System
criteria in Paragraph (b)(1)(iv) of this Section.
System using only
ground water not under
direct influence of
surface water using
chemical disinfectant
and serving at least
10,000 persons.
One (1) water sample per quarter
per treatment plant2.
Locations representing maximum residence time.1
System using only
ground water not under
direct influence of
surface water using
chemical disinfectant
and serving fewer than
10,000 persons.
One (1) water sample per year per
treatment plant2 during month of
warmest water temperature.
Locations representing maximum residence time.1
If the sample (or average of annual samples, if
more than one (1) sample is taken) exceeds the
MCL, the system must increase monitoring to one
(1) sample per treatment plant per quarter, taken at
a point reflecting the maximum residence time in
the distribution system, until the system meets
criteria in Paragraph (b)(1)(iv) of this Section.
1 If a system elects to sample more frequently than the minimum required, at least 25 percent of all samples
collected each quarter (including those taken in excess of the required frequency) must be taken at locations that
represent the maximum residence time of the water in the distribution system. The remaining samples must be taken
at locations representative of at least average residence time in the distribution system.
2 Multiple wells drawing water from a single aquifer may be considered one (1) treatment plant for determining the
minimum number of samples required, with Director approval in accordance with criteria developed by the Director.
(ii)
Systems may reduce monitoring, except as otherwise provided, in
accordance with the following table:
Reduced Monitoring Frequency for TTHM and HAA5
If You Are A…
You May Reduce Monitoring
If You Have Monitored At
Least One (1) Year And
Your…
To This Level
Subpart H system serving at
least 10,000 persons which
has a source water annual
average TOC level, before
any treatment, ≤4.0 mg/L.
TTHM annual average ≤0.040
mg/L and HAA5 annual
average ≤0.030 mg/L.
One (1) sample per treatment plant per quarter at
distribution system location reflecting maximum
residence time.
Subpart H system serving
from 500 to 9,999 persons
which has a source water
annual average TOC level,
before any treatment, ≤4.0
mg/L.
TTHM annual average ≤ 0.040
mg/L and HAA5 ≤0.030
mg/L.
One (1) sample per treatment plant per year at
distribution system location reflecting maximum
residence time during month of warmest water
temperature. NOTE: Any Subpart H system serving
fewer than 500 persons may not reduce its
monitoring to less than one (1) sample per
treatment plant per year.
System using only ground
water not under direct
influence of surface water
using chemical disinfectant
and serving at least 10,000
persons.
TTHM annual average ≤0.040
mg/L and HAA5 annual
average ≤0.030 mg/L.
One (1) sample per treatment plant per year
at distribution system location reflecting maximum
residence time during month of warmest water
temperature.
112
Reduced Monitoring Frequency for TTHM and HAA5
If You Are A…
You May Reduce Monitoring
If You Have Monitored At
Least One (1) Year And
Your…
To This Level
System using only ground
water not under direct
influence of surface water
using chemical disinfectant
and serving fewer than
10,000 persons.
TTHM annual average ≤0.040
mg/L and HAA5 distribution
annual average ≤0.030 mg/L
for two (2) consecutive years
OR TTHM annual average
≤0.020 mg/L and HAA5
annual average ≤0.015 mg/L
for one (1) year.
One (1) sample per treatment plant per three-year
monitoring cycle at system location reflecting
maximum residence time during month of warmest
water temperature, with the three-year cycle
beginning on January 1 following quarter in which
system qualifies for reduced monitoring.
(iii)
Systems on a reduced monitoring schedule may remain on that
reduced schedule as long as the average of all samples taken in the
year (for systems which must monitor quarterly) or the result of the
sample (for systems which must monitor no more frequently than
annually) is no more than 0.060 mg/L and 0.045 mg/L for TTHMs
and HAA5, respectively. Systems that do not meet these levels
must resume monitoring at the frequency identified in Paragraph
(b)(1)(i) of this Section in the quarter immediately following the
quarter in which the system exceeds 0.060 mg/L and 0.045 mg/L
for TTHMs and HAA5, respectively. Systems that do not meet
these levels must resume monitoring at the frequency identified in
Paragraph (b)(1)(i) of this Section (minimum monitoring
frequency column) in the quarter immediately following the
monitoring period in which the system exceeds 0.060 mg/L or
0.045 mg/L for TTHM or HAA5 respectively. For systems using
only ground water not under the direct influence of surface water
and serving fewer than 10,000 persons, if either the TTHM annual
average is >0.080 mg/L or the HAA5 annual average is >0.060
mg/L, the system must go to the increased monitoring identified in
Paragraph (b)(1)(i) of this Section (sample location column) in the
quarter immediately following the monitoring period in which the
system exceeds 0.080 mg/L or 0.060 mg/L for TTHMs or HAA5,
respectively.
(iv)
Systems on increased monitoring may return to routine monitoring
if, after at least one (1) year of monitoring, their TTHM annual
average is ≤0.060 mg/L and their HAA5 annual average is ≤0.045
mg/L.
(v)
The Director may return a system to routine monitoring at the
Director’s discretion.
(2)
Chlorite
113
Community and non-transient, non-community water systems using
chlorine dioxide, for disinfection or oxidation, must conduct monitoring
for chlorite.
(i)
Routine Monitoring
(A)
Daily monitoring
Systems must take daily samples at the entrance to the
distribution system. For any daily sample that exceeds the
chlorite MCL, the system must take additional samples in
the distribution system the following day at the locations
required by Paragraph (b)(2)(ii) of this Section, in addition
to the sample required at the entrance to the distribution
system.
(B)
Monthly monitoring
Systems must take a three-sample set each month in the
distribution system. The system must take one (1) sample at
each of the following locations: near the first customer, at a
location representative of average residence time, and at a
location reflecting maximum residence time in the
distribution system. Any additional routine sampling must
be conducted in the same manner (as three-sample sets, at
the specified locations). The system may use the results of
additional monitoring conducted under Paragraph (b)(2)(ii)
of this Section to meet the requirement for monitoring in
this Paragraph.
(ii)
Additional monitoring
On each day following a routine sample monitoring result that
exceeds the chlorite MCL at the entrance to the distribution
system, the system is required to take three (3) chlorite distribution
system samples at the following locations: as close to the first
customer as possible, in a location representative of average
residence time, and as close to the end of the distribution system as
possible (reflecting maximum residence time in the distribution
system).
(iii)
Reduced monitoring
114
(A)
Chlorite monitoring at the entrance to the distribution
system required by Paragraph (b)(2)(i)(A) of this Section
may not be reduced.
(B)
Chlorite monitoring in the distribution system required by
Paragraph (b)(2)(i)(B) of this Section may be reduced to
one (1) three-sample set per quarter after one (1) year of
monitoring where no individual chlorite sample taken in the
distribution system under Paragraph (b)(2)(i)(B) of this
Section has exceeded the chlorite MCL and the system has
not been required to conduct monitoring under Paragraph
(b)(2)(ii) of this Section. The system may remain on the
reduced monitoring schedule until either any of the three
(3) individual chlorite samples taken quarterly in the
distribution system under Paragraph (b)(2)(i)(B) of this
Section exceeds the chlorite MCL or the system is required
to conduct monitoring under Paragraph (b)(2)(ii) of this
Section, at which time the system must revert to routine
monitoring.
(3)
Bromate
(i)
Routine monitoring. Community and nontransient noncommunity
systems using ozone, for disinfection or oxidation, must take one
(1) sample per month for each treatment plant in the system using
ozone. Systems must take samples monthly at the entrance to the
distribution system while the ozonation system is operating under
normal conditions.
(ii)
Reduced monitoring. Systems required to analyze for bromate may
reduce monitoring from monthly to once per quarter, if the system
demonstrates that the average source water bromide concentration
is less than 0.05 mg/L based upon representative monthly bromide
measurements for one (1) year. The system may remain on reduced
bromate monitoring until the running annual average source water
bromide concentration, computed quarterly, is ≥0.05 mg/L based
upon representative monthly measurements. If the running annual
average source water bromide concentration is ≥0.05 mg/L, the
system must resume routine monitoring required by Paragraph
(b)(3)(i) of this Section.
(c)
Monitoring Requirements For Disinfectant Residuals
(1)
Chlorine And Chloramines
(i)
Routine monitoring.
115
Community and non-transient, non-community water systems that
use chlorine or chloramines must measure the residual disinfectant
level in the distribution system at the same point in the distribution
system and at the same time as total coliforms are sampled, as
specified in Sections 16.0 and 17.0 of these regulations. Subpart H
systems may use the results of residual disinfectant concentration
sampling conducted under sections 5.6.1 (6) and 5.7.1 (2), in lieu
of taking separate samples.
(ii)
Reduced monitoring
Monitoring may not be reduced.
(2)
Chlorine dioxide
(i)
Routine monitoring
Community,
nontransient
noncommunity,
and
transient
noncommunity water systems that use chlorine dioxide for
disinfection or oxidation must take daily samples at the entrance to
the distribution system. For any daily sample that exceeds the
MRDL, the system must take samples in the distribution system
the following day at the locations required by Paragraph (c)(2)(ii)
of this Section, in addition to the sample required at the entrance to
the distribution system.
(ii)
Additional monitoring
On each day following a routine sample monitoring result that
exceeds the MRDL, the system is required to take three (3)
chlorine dioxide distribution system samples. If chlorine dioxide or
chloramines are used to maintain a disinfectant residual in the
distribution system, or if chlorine is used to maintain a disinfectant
residual in the distribution system and there are no disinfection
addition points after the entrance to the distribution system (i.e., no
booster chlorination), the system must take three (3) samples as
close to the first customer as possible, at intervals of at least six (6)
hours. If chlorine is used to maintain a disinfectant residual in the
distribution system and there are one (1) or more disinfection
addition points after the entrance to the distribution system (i.e.,
booster chlorination), the system must take one (1) sample at each
of the following locations: as close to the first customer as
possible, in a location representative of average residence time,
and as close to the end of the distribution system as possible
(reflecting maximum residence time in the distribution system).
116
(iii)
Reduced monitoring
Chlorine dioxide monitoring may not be reduced.
(d)
Monitoring requirements for disinfection byproduct precursors (DBPP)
(1)
Routine monitoring
Subpart H systems which use conventional filtration treatment (as defined
in Section 1.0) must monitor each treatment plant for TOC no later than
the point of combined filter effluent turbidity monitoring and
representative of the treated water. All systems required to monitor under
this Paragraph (d)(1) must also monitor for TOC in the source water prior
to any treatment at the same time as monitoring for TOC in the treated
water. These samples (source water and treated water) are referred to as
paired samples. At the same time as the source water sample is taken, all
systems must monitor for alkalinity in the source water prior to any
treatment. Systems must take one (1) paired sample and one (1) source
water alkalinity sample per month per plant at a time representative of
normal operating conditions and influent water quality.
(2)
Reduced monitoring
Subpart H systems with an average treated water TOC of less than 2.0
mg/L for two (2) consecutive years, or less than 1.0 mg/L for one (1) year,
may reduce monitoring for both TOC and alkalinity to one (1) paired
sample and one (1) source water alkalinity sample per plant per quarter.
The system must revert to routine monitoring in the month following the
quarter when the annual average treated water TOC ≥2.0 mg/L.
(e)
Bromide
Systems required to analyze for bromate may reduce bromate monitoring from
monthly to once per quarter, if the system demonstrates that the average source
water bromide concentration is less than 0.05 mg/L based upon representative
monthly measurements for one (1) year. The system must continue bromide
monitoring to remain on reduced bromate monitoring.
(f)
Monitoring plans
Each system required to monitor under this subpart must develop and implement a
monitoring plan. The system must maintain the plan and make it available for
inspection by the Director and the general public no later than 30 days following
the applicable compliance dates in Section 7.3. All Subpart H systems serving
more than 3300 people must submit a copy of the monitoring plan to the Director
117
no later than the date of the first report required under Section 7.7. The Director
may also require the plan to be submitted by any other system. After review, the
Director may require changes in any plan elements. The plan must include at least
the following elements.
(1)
Specific locations and schedules for collecting samples for any parameters
included in this subpart.
(2)
How the system will calculate compliance with MCLs, MRDLs, and
treatment techniques.
(3)
If approved by the Director for monitoring as a consecutive system, or if
providing water to a consecutive system, the sampling plan must reflect
the entire distribution system.
7.6
Compliance Requirements
(a)
General Requirements
(1)
Where compliance is based on a running annual average of monthly or
quarterly samples or averages and the system fails to monitor for TTHM,
HAA5, or bromate, this failure to monitor will be treated as a monitoring
violation for the entire period covered by the annual average. Where
compliance is based on a running annual average of monthly or quarterly
samples or averages and the system failure to monitor makes it impossible
to determine compliance with MRDLs for chlorine and chloramines, this
failure to monitor will be treated as a monitoring violation for the entire
period covered by the annual average.
(2)
All samples taken and analyzed under the provisions of this Section must
be included in determining compliance, even if that number is greater than
the minimum required.
(3)
If, during the first year of monitoring under Section 7.5, any individual
quarter's average will cause the running annual average of that system to
exceed the MCL, the system is out of compliance at the end of that
quarter.
(b)
Disinfection byproducts
(1)
TTHMs and HAA5
(i)
For systems monitoring quarterly, compliance with MCLs in
Section 7.1 must be based on a running annual arithmetic average,
computed quarterly, of quarterly arithmetic averages of all samples
collected by the system as prescribed by Section 7.5(b)(1).
118
(ii)
For systems monitoring less frequently than quarterly, systems
demonstrate MCL compliance if the average of samples taken that
year under the provisions of Section 7.5 (b)(1) does not exceed the
MCLs in Section 7.1. If the average of these samples exceeds the
MCL, the system must increase monitoring to once per quarter per
treatment plant and such a system is not in violation of the MCL
until it has completed one (1) year of quarterly monitoring, unless
the result of fewer than four (4) quarters of monitoring will cause
the running annual average to exceed the MCL, in which case the
system is in violation at the end of that quarter. Systems required
to increase monitoring frequency to quarterly monitoring must
calculate compliance by including the sample which triggered the
increased monitoring plus the following three (3) quarters of
monitoring.
(iii)
If the running annual arithmetic average of quarterly averages
covering any consecutive four-quarter period exceeds the MCL,
the system is in violation of the MCL and must notify the public
pursuant to Section 16.8, whichever is effective for your system, in
addition to reporting to the State pursuant to Section 7.7.
(iv)
If a PWS fails to complete four (4) consecutive quarters of
monitoring, compliance with the MCL for the last four-quarter
compliance period must be based on an average of the available
data.
(2)
Bromate
Compliance must be based on a running annual arithmetic average,
computed quarterly, of monthly samples (or, for months in which the
system takes more than one (1) sample, the average f all samples taken
during the month) collected by the system as prescribed by Section 7.5
(b)(3). If the average of samples covering any consecutive four-quarter
period exceeds the MCL, the system is in violation of the MCL and must
notify the public pursuant to Section 16.8 in addition to reporting to the
Director pursuant to Section 7.7. If a PWS fails to complete 12
consecutive months' monitoring, compliance with the MCL for the last
four-quarter compliance period must be based on an average of the
available data.
(3)
Chlorite
Compliance must be based on an arithmetic average of each three (3)
sample set taken in the distribution system as prescribed by Section 7.5
(b)(2)(i)(B) and Section 7.5 (b)(2)(ii). If the arithmetic average of any
119
three (3) sample set exceeds the MCL, the system is in violation of the
MCL and must notify the public pursuant to Section 16.8, in addition to
reporting to the Director pursuant to Section 7.7
(c)
Disinfectant residuals
(1)
Chlorine and chloramines
(i)
Compliance must be based on a running annual arithmetic average,
computed quarterly, of monthly averages of all samples collected
by the system under Section 7.5 (c)(1). If the average covering any
consecutive four-quarter period exceeds the MRDL, the system is
in violation of the MRDL and must notify the public pursuant to
Section 16.8, in addition to reporting to the Director pursuant to
Section 7.7.
(ii)
In cases where systems switch between the use of chlorine and
chloramines for residual disinfection during the year, compliance
must be determined by including together all monitoring results of
both chlorine and chloramines in calculating compliance. Reports
submitted pursuant to Section 7.7 must clearly indicate which
residual disinfectant was analyzed for each sample.
(2)
Chlorine dioxide
(i)
Acute violations. Compliance must be based on consecutive daily
samples collected by the system under Section 7.5 (c)(2). If any
daily sample taken at the entrance to the distribution system
exceeds the MRDL, and on the following day one (1) (or more) of
the three (3) samples taken in the distribution system exceed the
MRDL, the system is in violation of the MRDL and must take
immediate corrective action to lower the level of chlorine dioxide
below the MRDL and must notify the public pursuant to the
procedures for acute health risks in Section 16.8 in addition to
reporting to the Director pursuant to Section 7.7. Failure to take
samples in the distribution system the day following an exceedance
of the chlorine dioxide MRDL at the entrance to the distribution
system will also be considered an MRDL violation and the system
must notify the public of the violation in accordance with the
provisions for acute violations under Section 16.8 in addition to
reporting to the Director pursuant to Section 7.7.
(ii)
Nonacute violations. Compliance must be based on consecutive
daily samples collected by the system under Section 7.5 (c)(2). If
any two (2) consecutive daily samples taken at the entrance to the
distribution system exceed the MRDL and all distribution system
120
samples taken are below the MRDL, the system is in violation of
the MRDL and must take corrective action to lower the level of
chlorine dioxide below the MRDL at the point of sampling and
will notify the public pursuant to the procedures for nonacute
health risks in Section 16.8 in addition to reporting to the Director
pursuant to Section 7.7. Failure to monitor at the entrance to the
distribution system the day following an exceedance of the
chlorine dioxide MRDL at the entrance to the distribution system
is also an MRDL violation and the system must notify the public of
the violation in accordance with the provisions for nonacute
violations under Section 16.8 in addition to reporting to the
Director pursuant to Section 7.7.
(d)
Disinfection byproduct precursors (DBPP)
Compliance must be determined as specified by Section 7.8 (c). Systems may
begin monitoring to determine whether Step 1 TOC removals can be met 12
months prior to the compliance date for the system. This monitoring is not
required and failure to monitor during this period is not a violation. However, any
system that does not monitor during this period, and then determines in the first
12 months after the compliance date that it is not able to meet the Step 1
requirements in Section 7.8 (b)(2) and must therefore apply for alternate
minimum TOC removal (Step 2) requirements, is not eligible for retroactive
approval of alternate minimum TOC removal (Step 2) requirements as allowed
pursuant to Section 7.8 (b)(3) and is in violation. Systems may apply for alternate
minimum TOC removal (Step 2) requirements any time after the compliance date.
For systems required to meet Step 1 TOC removals, if the value calculated under
Section 7.8 is less than 1.00, the system is in violation of the treatment technique
requirements and must notify the public pursuant to Section 16.8, in addition to
reporting to the State pursuant to Section 7.7.
7.7
Reporting And Recordkeeping Requirements
(a)
Systems required to sample quarterly or more frequently must report to the
Director within 10 days after the end of each quarter in which samples were
collected. Systems required to sample less frequently than quarterly must report to
the Director within 10 days after the end of each monitoring period in which
samples were collected.
(b)
Disinfection byproducts. Systems must report the information specified in the
following table:
If you are a…
You must report1…
(1) System monitoring for TTHMs and HAA5 under
the requirements of Section 7.5 (b) on a quarterly or
more frequent basis.
(i) The number of samples taken during the last quarter.
(ii) The location, date, and result of each sample taken
during the last quarter.
121
(iii) The arithmetic average of all samples taken in the
last quarter.
(iv) The annual arithmetic average of the quarterly
arithmetic averages of this Section for the last four (4)
quarters.
(v) Whether, based on Section 7.6 (b)(1), the MCL was
violated
(2) System monitoring for TTHMs and HAA5 under
the requirements of Section 7.5 (b) less frequently than
quarterly (but as least annually).
(i) The number of samples taken during the last year.
(ii) The location, date, and result of each sample taken
during the last monitoring period.
(iii) The arithmetic average of all samples taken over the
last year.
(iv) Whether, based on Section 7.6 (b)(1), the MCL was
violated.
(3) System monitoring for TTHMs and HAA5 under
the requirements of Section 7.5 (b) less frequently than
annually.
(i) The location, date, and result of each sample taken.
(ii) Whether, based on Section 7.6 (b)(1), the MCL was
violated.
(4) System monitoring for chlorite under the
requirements of Section 7.5 (b).
(i) The number of entry point samples taken each month
for the last 3 months.
(ii) The location, date, and result of each sample (both
entry point and distribution system) taken during the last
quarter.
(iii) For each month in the reporting period, the
arithmetic average of all samples taken in each three (3)
samples set taken in the distribution system.
(iv) Whether, based on Section 7.6 (b)(3), the MCL was
violated, in which month, and how many times it was
violated each month.
(5) System monitoring for bromate under the
requirements of Section 7.5 (b).
(i) The number of samples taken during the last quarter.
(ii) The location, date, and result of each sample taken
during the last quarter.
(iii) The arithmetic average of the monthly arithmetic
averages of all samples taken in the last year.
(iv) Whether, based on Section 7.6 (b)(2), the MCL was
violated.
1 The State may choose to perform calculations and determine whether the MCL was exceeded, in lieu of having the
system report that information
(c)
Disinfectants. Systems must report the information specified in the following
table:
If you are a…
You must report1…
(1) System monitoring for chlorine or chloramines
under the requirements of Section 7.5(c).
(i) The number of samples taken during each month of
the last quarter.
(ii) The month arithmetic average of all samples taken in
each month for the last 12 months.
(iii) The arithmetic average of the monthly averages for
the last 12 months.
(iv) Whether, based on Section 7.6 (c)(1), the MRDL was
violated.
(2) System monitoring for chlorine dioxide under the
requirements of Section 7.5 (c).
(i) The dates, result, and locations of samples taken
during the last quarter.
122
(ii) Whether, based on Section 7.6 (c)(2), the MRDL was
violated.
(iii) Whether the MRDL was exceeded in any two (2)
consecutive daily samples and whether the resulting
violation was acute or nonacute.
1 The Director may choose to perform calculations and determine whether the MRDL was exceeded, in lieu of
having the system report that information.
(d)
Disinfection byproduct precursors and enhanced coagulation or enhanced
softening. Systems must report the information specified in the following table:
If you are a…
You must report1…
(1) System monitoring monthly or quarterly for TOC
under the requirements of Section 7.5 (d) and required
to meet the enhanced coagulation or enhanced softening
requirements in Section 7.8 (b) (2) or (3).
(i) The number of paired (source water and treated water)
samples taken during the last quarter.
(ii) The location, date, and results of each paired sample
and associated alkalinity taken during the last quarter.
(iii) For each month in the reporting period that paired
samples were taken, the arithmetic average of the percent
reduction of TOC for each paired sample and the required
TOC percent removal.
(iv) Calculations for determining compliance with the
TOC percent removal requirements, as provided in
Section 7.8 (c)(1).
(v) Whether the system is in compliance with the
enhanced coagulation or enhanced softening percent
removal requirements in Section 7.8(b) for the last four
(4) quarters.
(2) System monitoring monthly or quarterly for TOC
under the requirements of Section 7.5 (d) and meeting
one (1) or more of the alternative compliance criteria in
sections 7.8(a)(2) or (3).
(i) The alternative compliance criterion that the system is
using.
(ii) The number of paired samples taken during the last
quarter.
(iii) The location, date, and result of each paired sample
and associated alkalinity taken during the last quarter.
(iv) The running annual arithmetic average based on
monthly averages (or quarterly samples) of source water
TOC for systems meeting a criterion in Section
7.8(a)(2)(i) or (iii) or of treated water TOC for systems
meeting the criterion in Section 7.8 (a)(2)(ii).
(v) The running annual arithmetic average based on
monthly averages (or quarterly samples) of source water
SUVA for systems meeting the criterion in Section 7.8
(a)(2)(v) or of treated water SUVA for systems meeting
the criterion in Section 7.8 (a)(2)(vi).
(vi) The running annual average of source water
alkalinity for systems meeting the criterion in Section 7.8
(a)(2)(iii) and of treated water alkalinity for systems
meeting the criterion in Section 7.8 (a)(3)(i).
(vii) The running annual average for both TTHM and
HAA5 for systems meeting the criterion in Section
Section 7.8 (a)(2)(iii) or (iv).
(viii) The running annual average of the amount of
magnesium hardness removal (as CaCO3, in mg/L) for
123
systems meeting the criterion in Section 7.8 (a)(3)(ii).
(ix) Whether the system is in compliance with the
particular alternative compliance criterion in Section 7.8
(a)(2) or (3).
1 The Director may choose to perform calculations and determine whether the treatment technique was met, in lieu
of having the system report that information.
7.8
Treatment Technique For Control Of Disinfection Byproduct (DBP) Precursors
(a)
Applicability
(1)
Subpart H systems using conventional filtration treatment (as defined in
Section 1.0) must operate with enhanced coagulation or enhanced
softening to achieve the TOC percent removal levels specified in
Paragraph (b) of this Section unless the system meets at least one (1) of
the alternative compliance criteria listed in Paragraph (a)(2) or (a)(3) of
this Section.
(2)
Alternative compliance criteria for enhanced coagulation and enhanced
softening systems
Subpart H systems using conventional filtration treatment may use the
alternative compliance criteria in Paragraphs (a)(2)(i) through (vi) of this
Section to comply with this Section in lieu of complying with Paragraph
(b) of this Section. Systems must still comply with monitoring
requirements in Section 7.5 (d).
(i)
The system's source water TOC level, measured according to
Section 7.4 (d)(3), is less than 2.0 mg/L, calculated quarterly as a
running annual average.
(ii)
The system's treated water TOC level, measured according to
Section 7.4 (d)(3), is less than 2.0 mg/L, calculated quarterly as a
running annual average.
(iii)
The system's source water TOC level, measured according to
Section 7.4 (d)(3), is less than 4.0 mg/L, calculated quarterly as a
running annual average; the source water alkalinity, measured
according to Section 7.4 (d)(1), is greater than 60 mg/L (as
CaCO3), calculated quarterly as a running annual average; and
either the TTHM and HAA5 running annual averages are no
greater than 0.040 mg/L and 0.030 mg/L, respectively; or prior to
the effective date for compliance in Section 7.3 (b), the system has
made a clear and irrevocable financial commitment not later than
the effective date for compliance in Section 7.3 (b) to use of
technologies that will limit the levels of TTHMs and HAA5 to no
124
more than 0.040 mg/L and 0.030 mg/L, respectively. Systems must
submit evidence of a clear and irrevocable financial commitment,
in addition to a schedule containing milestones and periodic
progress reports for installation and operation of appropriate
technologies, to the Director for approval not later than the
effective date for compliance in Section 7.3 (b). These
technologies must be installed and operating not later than June 30,
2005. Failure to install and operate these technologies by the date
in the approved schedule will constitute a violation of these
regulations.
(iv)
The TTHM and HAA5 running annual averages are no greater than
0.040 mg/L and 0.030 mg/L, respectively, and the system uses
only chlorine for primary disinfection and maintenance of a
residual in the distribution system.
(v)
The system's source water SUVA, prior to any treatment and
measured monthly according to Section 7.4 (d)(4), is less than or
equal to 2.0 L/mg-m, calculated quarterly as a running annual
average.
(vi)
The system's finished water SUVA, measured monthly according
to Section 7.4 (d)(4), is less than or equal to 2.0 L/mg-m,
calculated quarterly as a running annual average.
(3)
Additional alternative compliance criteria for softening systems. Systems
practicing enhanced softening that cannot achieve the TOC removals
required by Paragraph (b)(2) of this Section may use the alternative
compliance criteria in Paragraphs (a)(3)(i) and (ii) of this Section in lieu of
complying with Paragraph (b) of this Section. Systems must still comply
with monitoring requirements in Section 7.5 (d).
(i)
Softening that results in lowering the treated water alkalinity to
less than 60 mg/L (as CaCO3), measured monthly according to
Section 7.4 (d)(1) and calculated quarterly as a running annual
average.
(ii)
Softening that results in removing at least 10 mg/L of magnesium
hardness (as CaCO3), measured monthly and calculated quarterly
as an annual running average.
(b)
Enhanced coagulation and enhanced softening performance requirements
(1)
Systems must achieve the percent reduction of TOC specified in
Paragraph (b)(2) of this Section between the source water and the
combined filter effluent, unless the Director approves a system's request
125
for alternate minimum TOC removal (Step 2) requirements under
Paragraph (b)(3) of this Section.
(2)
Required Step 1 TOC reductions, indicated in the following table, are
based upon specified source water parameters measured in accordance
with Section 7.4 (d). Systems practicing softening are required to meet the
Step 1 TOC reductions in the far-right column (Source water alkalinity
>120 mg/L) for the specified source water TOC.
Step 1 Required Removal of TOC by Enhanced Coagulation and Enhanced Softening for Subpart H Systems Using
Conventional Treatment 1 2
Source-water
TOC, mg/L
Source-water alkalinity, mg/L
as CaCO3 (in percentages)
0-60
>60-120
>1203
>2.0-4.0
35.0
25.0
15.0
>4.0-8.0
45.0
35.0
25.0
>8.0
50.0
40.0
30.0
1 Systems meeting at least one (1) of the conditions in Paragraph (a)(2)(i)- (vi) of this Section are not required to
operate with enhanced coagulation.
2 Softening systems meeting one (1) of the alternative compliance criteria in Paragraph (a)(3) of this Section are not
required to operate with enhanced softening.
3 Systems practicing softening must meet the TOC removal requirements in this column.
(3)
Subpart H conventional treatment systems that cannot achieve the Step 1
TOC removals required by Paragraph (b)(2) of this Section due to water
quality parameters or operational constraints must apply to the Director,
within three (3) months of failure to achieve the TOC removals required
by Paragraph (b)(2) of this Section, for approval of alternative minimum
TOC (Step 2) removal requirements submitted by the system. If the
Director approves the alternative minimum TOC removal (Step 2)
requirements, the Director may make those requirements retroactive for
the purposes of determining compliance. Until the State approves the
alternate minimum TOC removal (Step 2) requirements, the system must
meet the Step 1 TOC removals contained in Paragraph (b)(2) of this
Section.
(4)
Alternate minimum TOC removal (Step 2) requirements
Applications made to the Director by enhanced coagulation systems for
approval of alternate minimum TOC removal (Step 2) requirements under
Paragraph (b)(3) of this Section must include, at a minimum, results of
bench- or pilot-scale testing conducted under Paragraph (b)(4)(i) of this
Section. The submitted bench- or pilot-scale testing must be used to
determine the alternate enhanced coagulation level.
126
(i)
Alternate enhanced coagulation level is defined as coagulation at a
coagulant dose and pH as determined by the method described in
Paragraphs (b)(4)(i) through (v) of this Section such that an
incremental addition of 10 mg/L of alum (or equivalent amount of
ferric salt) results in a TOC removal of ≤0.3 mg/L. The percent
removal of TOC at this point on the “TOC removal versus
coagulant dose” curve is then defined as the minimum TOC
removal required for the system. Once approved by the Director,
this minimum requirement supersedes the minimum TOC removal
required by the table in Paragraph (b)(2) of this Section. This
requirement will be effective until such time as the Director
approves a new value based on the results of a new bench- and
pilot-scale test. Failure to achieve Director-set alternative
minimum TOC removal levels is a violation of these regulations.
(ii)
Bench- or pilot-scale testing of enhanced coagulation must be
conducted by using representative water samples and adding 10
mg/L increments of alum (or equivalent amounts of ferric salt)
until the pH is reduced to a level less than or equal to the enhanced
coagulation Step 2 target pH shown in the following table:
Enhanced Coagulation Step 2 Target pH
Alkalinity (mg/L as CaCO3)
Target pH
0-60
5.5
>60-120
6.3
>120-240
7.0
>240
7.5
(iii) For waters with alkalinities of less than 60 mg/L for which
addition of small amounts of alum or equivalent addition of iron
coagulant drives the pH below 5.5 before significant TOC removal
occurs, the system must add necessary chemicals to maintain the
pH between 5.3 and 5.7 in samples until the TOC removal of 0.3
mg/L per 10 mg/L alum added (or equivalant addition of iron
coagulant) is reached.
(iv)
The system may operate at any coagulant dose or pH necessary
(consistent with the other requirements of these regulations) to
achieve the minimum TOC percent removal approved under
Paragraph (b)(3) of this Section.
(v)
If the TOC removal is consistently less than 0.3 mg/L of TOC per
10 mg/L of incremental alum dose at all dosages of alum (or
equivalent addition of iron coagulant), the water is deemed to
127
contain TOC not amenable to enhanced coagulation. The system
may then apply to the Director for a waiver of enhanced
coagulation requirements.
(c)
Compliance Calculations
(1)
Subpart H systems other than those identified in Paragraph (a)(2) or (a)(3)
of this Section must comply with requirements contained in Paragraph
(b)(2) or (b)(3) of this Section. Systems must calculate compliance
quarterly, beginning after the system has collected 12 months of data, by
determining an annual average using the following method:
(i)
Determine actual monthly TOC percent removal, equal to: (1 –
(treated water TOC/source water TOC)) × 100
(ii)
Determine the required monthly TOC percent removal (from either
the table in Paragraph (b)(2) of this Section or from Paragraph
(b)(3) of this Section).
(iii)
Divide the value in Paragraph (c)(1)(i) of this Section by the value
in Paragraph (c)(1)(ii) of this Section.
(iv)
Add together the results of Paragraph (c)(1)(iii) of this Section for
the last 12 months and divide by 12.
(v)
If the value calculated in Paragraph (c)(1)(iv) of this Section is less
than 1.00, the system is not in compliance with the TOC percent
removal requirements.
(2)
Systems may use the provisions in Paragraphs (c)(2)(i) through (v) of this
Section in lieu of the calculations in Paragraph (c)(1)(i) through (v) of this
Section to determine compliance with TOC percent removal requirements.
(i)
In any month that the system's treated or source water TOC level,
measured according to Section 7.4 (d)(3), is less than 2.0 mg/L, the
system may assign a monthly value of 1.0 (in lieu of the value
calculated in Paragraph (c)(1)(iii) of this Section) when calculating
compliance under the provisions of Paragraph (c)(1) of this
Section.
(ii)
In any month that a system practicing softening removes at least 10
mg/L of magnesium hardness (as CaCO3), the system may assign a
monthly value of 1.0 (in lieu of the value calculated in Paragraph
(c)(1)(iii) of this Section) when calculating compliance under the
provisions of Paragraph (c)(1) of this Section.
128
(iii)
In any month that the system's source water SUVA, prior to any
treatment and measured according to Section 7.4 (d)(4), is ≤2.0
L/mg-m, the system may assign a monthly value of 1.0 (in lieu of
the value calculated in Paragraph (c)(1)(iii) of this Section) when
calculating compliance under the provisions of Paragraph (c)(1) of
this Section.
(iv)
In any month that the system's finished water SUVA, measured
according to Section 7.4 (d)(4), is ≤2.0 L/mg-m, the system may
assign a monthly value of 1.0 (in lieu of the value calculated in
Paragraph (c)(1)(iii) of this Section) when calculating compliance
under the provisions of Paragraph (c)(1) of this Section.
(v)
In any month that a system practicing enhanced softening lowers
alkalinity below 60 mg/L (as CaCO3), the system may assign a
monthly value of 1.0 (in lieu of the value calculated in Paragraph
(c)(1)(iii) of this Section) when calculating compliance under the
provisions of Paragraph (c)(1) of this Section.
(3)
Subpart H systems using conventional treatment may also comply with the
requirements of this Section by meeting the criteria in Paragraph (a)(2) or
(3) of this Section.
(d)
Treatment technique requirements for DBP precursors.
The following are identified as treatment techniques to control the level of
disinfection byproduct precursors in drinking water treatment and distribution
systems: For Subpart H systems using conventional treatment, enhanced
coagulation or enhanced softening.
129
SECTION 8.0 (RESERVED)
130
SECTION 9.0 ASSURANCE OF SAFETY IN PUBLIC SUPPLY
9.1
Any person maintaining a PWS shall operate and maintain the water supply facilities so
that the water furnished the public is safe and potable.
9.2
Contamination of Tanks
Connected to Unsafe Supplies
(a)
Any person who maintains a PWS connection to a tank which is also supplied
with water from a water system found by the Director to be unsafe shall maintain
the tank open to atmospheric pressure, and the public water supply pipe shall
terminate at least two (2) pipe diameters above the maximum level of water in the
tank. The tank overflow shall be of adequate size to fix definitely the maximum
level.
Avoidance of Contamination in Tanks
(b)
Any person who is furnished water from a PWS and maintains a tank supplied
only by such water shall have such tank so constructed and maintained to prevent
contaminants from gaining access to the tank interior.
9.3
Connections Between Distribution Systems
(a)
No person shall maintain a physical connection joining a PWS with any other
water system, unless such connection is approved by the Director.
(b)
It is the responsibility of the PWS to register all existing or proposed connections
between the PWS and any other water supply with the Director on or before
January 1, 1992 or as they are proposed or discovered, whichever is later.
131
SECTION 10.0 CORRECTION OF UNSAFE CONDITIONS
10.1
When the water from a PWS is not safe or is subject to contamination, as determined by
the Director, the person maintaining such PWS shall take immediate action to correct
sanitary defects, improve operation, provide necessary water treatment, or make any
other changes or additions deemed necessary by the Director to provide safe water.
10.2
Any person maintaining a water system who is aware of an unsafe condition, that the
water is not safe or is subject to contamination, shall notify the Director immediately.
132
SECTION 11.0 REPORTS AS TO PUBLIC SUPPLIES
11.1
Any person maintaining a PWS shall submit or cause to be submitted by operating
personnel such reports of operation pertaining to the sanitary quality, treatment and
output as may be required by the Director. Such operation reports shall be submitted
within ten (10) days after demand and shall be accurate and complete as required by the
Director. Violations of maximum contaminant levels shall be reported to the Director
within 48 hours after such a determination is made unless otherwise required for specific
contaminants.
11.2
It is the responsibility of the water system to collect, have analyzed, and report the results
of all water quality samples required by these regulations. Samples must be collected in
accordance with a written sample siting plan. These plans are subject to the Director's
review and revision.
133
SECTION 12.0 CERTIFIED LABORATORIES
12.1
For the purpose of determining compliance with these regulations, only analyses carried
out by the Department of Health or in a laboratory certified by the Department of Health,
EPA, or by reciprocity with another state will be considered with the exception of
turbidity pH, temperature, and residual disinfectant concentration determinations, which
must be carried out by a party approved by the Director.
134
SECTION 13.0 GROUND WATER MICROBIOLOGY
13.1
Ground water sources shall meet the stipulated microbiological standard prior to
disinfection where disinfection is practiced.
135
SECTION 14.0 CONSECUTIVE WATER SYSTEM MONITORING
14.1
These regulations shall also pertain to a PWS which is supplied by another PWS except
as specifically modified by the Director and agreed upon by the EPA Administrator.
136
SECTION 15.0 VARIANCES AND EXEMPTIONS
15.1
Variances and exemptions to these regulations may be granted by the Director in
accordance with Chapter 42-35 of the Rhode Island General Laws of 1956, as amended
and if deemed applicable by the Director the provisions of Sections 300g-4 and 300g-5 of
42 USC et seq. (Section 1415 variances to regulations promulgated pursuant to the
SDWA and Section 1416 exemptions to regulations promulgated pursuant to the SDWA
of Public Law 93-523 as amended).
15.1.1 Variances pursuant to Section 1415 may be granted as follows:
(a)
The Director may grant variances from an applicable national primary drinking
water regulation to a PWS which, because of characteristics of the raw water
sources which are reasonably available to the system, cannot meet the
requirements respecting the maximum contaminant levels of such drinking water
regulation. A variance may be issued to a system on condition that the system
install the best technology, treatment techniques, or other means, which the
Director finds are available (taking costs into consideration) and based upon an
evaluation satisfactory to the Director that indicates that alternative sources of
water are not reasonably available to the system.
Before the Director may grant a variance under this Subparagraph, the Director
must find that the variance will not result in an unreasonable risk to health. If the
Director grants a PWS a variance under this Subparagraph, the Director shall
prescribe at the time the variance is granted, a schedule for:
(i)
compliance (including increments of progress) by the PWS with each
contaminant level requirement with respect to which the variance was
granted, and
(ii)
implementation by the PWS of such additional control measures as the
State may require for each contaminant, subject to such contaminant level
requirement, during the period ending on the date compliance with such
requirement is required. Before a schedule prescribed pursuant to this
Subparagraph may take effect, the Director shall provide notice and
opportunity for a public hearing on the schedule. A schedule prescribed
pursuant to this Subparagraph for a PWS granted a variance shall require
compliance by the system with each contaminant level requirement with
respect to which the variance was granted as expeditiously as practicable.
(b)
The Director may grant variances from any provisions of a national primary
drinking water regulation which requires the use of a specified treatment
technique with respect to a contaminant if the PWS applying for the variance
demonstrates to the satisfaction of the Director that such treatment technique is
not necessary to protect the health of persons because of the nature of the raw
water source of such system. A variance granted under this Subparagraph shall be
137
conditioned on such monitoring and other requirements as the Director may
prescribe.
(c)
Before a variance proposed to be granted by the Director under Subparagraph (a)
or (b) may take effect, the Director shall provide notice and opportunity for public
hearing on the proposed variance. The Director shall promptly notify the
Administrator of all variances that are granted. Such notification shall contain the
reason for the variance [and in the case of a variance under Subparagraph (a), the
basis for the finding required by that Subparagraph before the granting of the
variance] and documentation of the need for the variance.
(d)
Each PWS’s variance granted under Subparagraph (a) shall be conditioned upon
compliance by the PWS with the schedule prescribed by the Director pursuant to
that Subparagraph.
(e)
For such variance issued under this Subparagraph, the Director
(1)
must document all findings that are required under Section 1415(a) of the
SDWA.
(2)
If the Director prescribes a schedule pursuant to Section 15.1.1(a)
requiring compliance with a contaminant level for which the variance is
granted later than five (5) years from the date of issuance of the variance
the Director must
(i)
document the rationale for the extended compliance schedule;
(ii)
discuss the rationale for the extended compliance schedule in the
required public notice and opportunity for public hearing; and
(iii)
provide the shortest practicable time schedule feasible under the
circumstances.
(f)
Variances for Small Systems
General Provisions
(1)
What is a small system variance?
Small system variances are variances from the requirement to comply with
a maximum contaminant level or treatment technique to systems serving
fewer than 10,000 persons. The purpose of this subpart is to provide the
procedures and criteria for obtaining these variances.
(2)
Who can issue a small system variance?
138
A small system variance under this subpart may only be issued by the
Director.
(3)
Which size PWSs can receive a small system variance?
(a)
The Director may grant a small system variance to PWSs serving
3,300 or fewer persons.
(b)
With the approval of the EPA Regional Administrator, the Director
may grant a small system variance to PWSs serving more than
3,300 persons but fewer than 10,000 persons.
(c)
In determining the number of persons served by the PWS, persons
served by consecutive systems must be included. A small system
variance granted to a PWS would also apply to any consecutive
system served by it.
(4)
For which of the regulatory requirements is a small system variance
available?
(a)
A small system variance is not available under this subpart for a
national primary drinking water regulation for a microbial
contaminant (including a bacterium, virus, or other organism) or an
indicator or treatment technique for a microbial contaminant.
(b)
A small system variance under this subpart is otherwise only
available for compliance with a requirement specifying a
maximum contaminant level or treatment technique for a
contaminant with respect to which:
(1)
a national primary drinking water regulation was
promulgated on or after January 1, 1986; and
(2)
the Administrator has published a small system variance
technology pursuant to Section 1412(b)(15) of the SDWA.
Note to Paragraph (b)(1): Small system variances are not
available for any PWS above the pre-1986 maximum
contaminant level even if subsequently revised. If the
agency revises a pre-1986 maximum contaminant level and
makes it more stringent, then a variance would be available
for that contaminant, but only up to the pre-1986 maximum
contaminant level.
(5)
When can a small system variance be granted by the Director?
139
No small system variance can be granted by the Director until the later of
the following:
(a)
90 days after the Director proposed to grant the small system
variance;
(b)
If the Director is proposing to grant a small system variance to a
PWS serving 3,300 or fewer persons and the Administrator objects
to the small system variance, the date on which the Director makes
the recommended modifications or responds in writing to each
objection; or
(c)
If the Director is proposing to grant a small system variance to a
PWS serving a population more than 3,300 and fewer than 10,000
persons, the date the Administrator approves the small system
variance. The Administrator must approve or disapprove the
variance within 90 days after it is submitted to the Administrator
for review.
Review of Small System Variance Application
(6)
What are the responsibilities of the PWS, Director, and the Administrator
in ensuring that sufficient information is available and for evaluation of a
small system variance application?
(a)
A PWS requesting a small system variance must provide accurate
and correct information to the Director to issue a small system
variance in accordance with this subpart.
(b)
Based upon an application for a small system variance and other
information, and before a small system variance may be proposed
under this subpart, the Director must find and document the
following:
(1)
The PWS is eligible for a small system variance pursuant to
15.1.1(f)(3) (i.e., the system serves a population of fewer
than 10,000 persons) and (f)(4) (i.e., the contaminant for
which the small system variance is sought is not excluded
from variance eligibility);
(2)
The PWS cannot afford to comply, in accordance with the
affordability criteria established by the Director, with the
national primary drinking water regulation for which a
small system variance is sought, including by:
(i)
Treatment;
140
(ii)
Alternative sources of water supply;
(iii)
Restructuring or consolidation changes, including
ownership change and/or physical consolidation
with another PWS; or
(iv)
Obtaining financial assistance;
(3)
The PWS meets the source water quality requirements for
installing the small system variance technology;
(4)
The PWS is financially and technically capable of
installing, operating and maintaining the applicable small
system variance technology; and
(5)
The terms and conditions of the small system variance, as
developed through compliance with (f)(7) ensure adequate
protection of human health, considering the following:
(i)
The quality of the source water for the PWS; and
(ii)
Removal efficiencies and expected useful life of the
small system variance technology.
(7)
What terms and conditions must be included in a small system variance?
(a)
The Director must clearly specify enforceable terms and conditions
of a small system variance.
(b)
The terms and conditions of a small system variance issued under
this subpart must include, at a minimum, the following
requirements:
(1)
Proper
and
effective
installation,
operation,
and
maintenance of the applicable small system variance
technology taking into consideration any relevant source
water characteristics and any other site-specific conditions
that may affect proper and effective operation and
maintenance of the technology;
(2)
Monitoring requirements, for the contaminant for which a
small system variance is sought; and
(3)
Any other terms or conditions that are necessary to ensure
adequate protection of public health, which may include:
141
(i)
Public education requirements; and
(ii)
Source water protection requirements.
(c)
The Director must establish a schedule for the PWS to comply
with the terms and conditions of the small system variance which
must include, at a minimum, the following requirements:
(1)
Increments of progress, such as milestone dates for the
PWS to apply for financial assistance and begin capital
improvements;
(2)
Quarterly reporting to the Director of the PWS’s
compliance with the terms and conditions of the small
system variance;
(3)
Schedule for the Director to review the small system
variance under Paragraph (d) of this Section; and
(4)
Compliance with the terms and conditions of the small
system variance as soon as practicable but not later than 3
years after the date on which the small system variance is
granted. The Director may allow up to 2 additional years if
the Director determines that additional time is necessary for
the PWS to:
(i)
Complete necessary capital improvements to
comply with the small system variance technology,
secure an alternative source of water, or restructure
or consolidate; or
(ii)
Obtain financial assistance.
(d)
The Director must review each small system variance granted not
less often than every 5 years after the compliance date established
in the small system variance to determine whether the PWS
continues to meet the eligibility criteria and remains eligible for the
small system variance and is complying with the terms and
conditions of the small system variance. If the PWS would no
longer be eligible for a small system variance, the Director must
determine whether continuing the variance is in the public interest.
If the Director finds that continuing the variance is not in the
public interest, the variance must be withdrawn.
Public Participation
142
(8)
What public notice is required before the Director proposes to issue a
small system variance?
(a)
At least fifteen (15) days before the date of proposal, and at least
thirty (30) days prior to a public meeting to discuss the proposed
small system variance, the Director, or PWS as directed by the
Director, must provide notice to all persons served by the PWS.
For billed customers, identified in Paragraph (a)(1) of this Section,
this notice must include the information listed in Paragraph (c) of
this Section.
For other persons regularly served by the system, identified in
Paragraph (a)(2) of this Section, the notice shall include the
information identified in Paragraph (d) of this Section. Notice must
be provided to all persons served by:
(1)
Direct mail or other home delivery to billed customers or
other service connections; and
(2)
Any other method reasonably calculated to notify, in a brief
and concise manner, other persons regularly served by the
system. Such methods may include publication in a local
newspaper, posting in public places, or delivery to
community organizations.
(b)
At the time of proposal, the Director must publish a notice in a
newspaper or newspapers of wide circulation in the State. This
notice shall include the information listed in Paragraph (c) of this
Section.
(c)
The notice in Paragraphs (a)(1) and (b) of this Section must
include, at a minimum, the following:
(1)
Identification of the contaminant(s) for which a small
system variance is sought;
(2)
A brief statement of the health effects associated with the
contaminant(s) for which a small system variance is sought
using language in Section 16.10 of these regulations;
(3)
The address and telephone number at which interested
persons may obtain further information concerning the
contaminant and the small system variance;
143
(4)
A brief summary, in easily understandable terms, of the
terms and conditions of the small system variance;
(5)
A description of the consumer petition process under
Section 15.1.1(f)(10) and information on contacting the
EPA Regional Office;
(6)
A brief statement announcing the public meeting required
under Section (f)(9)(a), including a statement of the
purpose of the meeting, information regarding the time and
location for the meeting and the address and telephone
number at which interested persons may obtain further
information concerning the meeting; and
(7)
In communities with a large proportion of non-English-
speaking residents, as determined by the Director,
information in the appropriate language regarding the
content and importance of the notice.
(d)
The notice in Paragraph (a)(2) of this Section must provide
sufficient information to alert readers to the proposed variance and
direct them where to receive additional information.
(e)
At his option, the Director or the Administrator may choose to
issue separate notices or additional notices related to the proposed
small system variance, provided that the requirements in
Paragraphs (a) through (d) of this Section are satisfied.
(f)
Prior to promulgating the final variance, the Director must respond
in writing to all significant public comments received relating to
the small system variance. Response to public comment and any
other documentation supporting the issuance of a variance must be
made available to the public after final promulgation.
(9)
What are the public meeting requirements associated with the proposal of
a small system variance?
(a)
The Director must provide for at least one (1) public meeting on
the small system variance no later than 15 days after the small
system variance is proposed.
(b)
At the time of the public meeting, the Director must prepare and
make publicly available, in addition to the information listed in
15.1.1(f)(8)(c) either:
144
(1)
The proposed small system variance, if the public meeting
occurs after proposal of the small system variance; or
(2)
A draft of the proposed small system variance, if the public
meeting occurs prior to proposal of the proposed small
system variance.
(c)
Notice of the public meeting must be provided in the manner
required under 15.1.1(f)(8) at least 30 days in advance of the
public meeting. This notice must be provided by the Director or the
PWS as directed by the Director.
(10)
How can a person served by the PWS obtain EPA review of a small
system variance proposed by the Director?
(a)
Any person served by the PWS may petition the Administrator to
object to the granting of a small system variance within 30 days
after the Director proposes to grant a small system variance for a
PWS.
(b)
The Administrator must respond to a petition filed by any person
served by the PWS and determine whether to object to the small
system variance no later than 60 days after the receipt of the
petition.
EPA Review and Approval of Small System Variances
(11)
What procedures allow the Administrator to object to a proposed small
system variance or overturn a granted small system variance for a PWS
serving 3,300 or fewer persons?
(a)
At the time the Director proposes to grant a small system variance
under this subpart, the Director must submit to the Administrator
the proposed small system variance and all supporting information,
including any written public comments received prior to proposal.
(b)
The Administrator may review and object to any proposed small
system variance within 90 days of receipt of the proposed small
system variance. The Administrator must notify the Director in
writing of each basis for the objection and propose a modification
to the small system variance to resolve the concerns of the
Administrator. The Director must make the recommended
modification, respond in writing to each objection, or withdraw the
proposal to grant the small system variance.
145
(c)
If the Director issues the small system variance without resolving
the concerns of the Administrator, the Administrator may overturn
the decision to grant the variance if the Administrator determines
that the Director’s decision does not comply with the Safe
Drinking Water Act or this rule.
(12)
What EPA action is necessary when the Director proposes to grant a small
system variance to a PWS serving a population of more than 3,300 and
fewer than 10,000 persons?
(a)
At the time the Director proposes to grant a small system variance
to a PWS serving a population of more than 3,300 and fewer than
10,000 persons, the Director must submit the proposed small
system variance and all supporting information, including public
comments received prior to proposal, to the Administrator.
(b)
The Administrator must approve or disapprove the small system
variance within 90 days of receipt of the proposed small system
variance and supporting information. The Administrator must
approve the small system variance if it meets each requirement
within the Act and this rule.
(c)
If the Administrator disapproves the small system variance, the
Administrator must notify the Director in writing of the reasons for
disapproval and the small system variance does not become
effective. The Director may resubmit the small system variance for
review and approval with modifications to address the objections
stated by the Administrator.
15.1.2 Exemptions Pursuant to Section 1416 of the Safe Drinking Water Act
(a)
Exemptions may be granted by the Director from any requirement respecting a
maximum contaminant level or any treatment technique requirement, or from
both, of an applicable national primary drinking water regulation upon a finding
that:
(1)
due to compelling factors (which may include economic factors, including
qualification of the PWS as a system serving a disadvantaged community),
the PWS is unable to comply with such contaminant level or treatment
technique requirement or to implement measures to develop an alternative
source of water supply;
(2)
the PWS was in operation on the effective date of such contaminant level
or treatment technique requirement or for a system that was not in
operation by that date, only if no reasonable alternative source of drinking
water is available to such new system;
146
(3)
the granting of the exemption will not result in an unreasonable risk to
health; and
(4)
management or restructuring changes (or both) cannot reasonably be made
that will result in compliance or, if compliance cannot be achieved,
improve the quality of the drinking water.
(b)
If the Director grants a PWS an exemption under subsection (a), the exemption
shall include a schedule which includes the items listed in this Paragraph. Before
a schedule prescribed by the Director pursuant to this subsection may take effect,
the Director shall provide notice and opportunity for a public hearing on the
schedule.
(1)
The Director shall prescribe, at the time the exemption is granted, a
schedule for:
(A)
compliance (including increments of progress or measures to
develop an alternative source of water supply) by the PWS with
each contaminant level requirement or treatment technique
requirement with respect to which the exemption was granted, and
(B)
implementation by the PWS of such control measures as the
Director may require for each contaminant, subject to such
contaminant level requirement or treatment technique requirement,
during the period ending on the date compliance with such
requirement is required.
(2)
A schedule prescribed pursuant to this subsection for a PWS granted an
exemption under subsection (a) shall require the following:
(A)
compliance by the system with each contaminant level and
treatment technique requirement with respect to which the
exemption was granted as expeditiously as practicable but not later
than 3 years after the otherwise applicable compliance date.
(B)
No exemption shall be granted unless the PWS establishes that the
system is taking all practicable steps to meet the standard; and
(i)
the system cannot meet the standard without capital
improvements which cannot be completed prior to the
otherwise applicable compliance date;
(ii)
in the case of a system which needs financial assistance for
the necessary improvements, the system has entered into an
agreement to obtain such financial assistance or assistance
147
is reasonably likely to be available within the period of the
exemption; or
(iii)
the system has entered into an enforceable agreement to
become a part of a regional PWS.
(C)
In the case of a system which does not serve more than a
population of 3,300 and which needs financial assistance for the
necessary improvements, an exemption granted under clause (i) or
(ii) of Subparagraph (B) may be renewed for one (1) or more
additional 2-year periods, but not to exceed a total of 6 years, if the
system establishes that it is taking all practicable steps to meet the
requirements of Subparagraph (B).
(D)
Limitation - A PWS may not receive an exemption under this
Section if the system was granted a variance under Section 15.1.1.
(3)
Each PWS’s exemption granted by the Director under subsection (a) shall
be conditioned upon compliance by the PWS with the schedule prescribed
pursuant to this subsection.
(c)
The Director shall promptly notify the Administrator of the granting of all
exemptions. Such notification shall contain the reasons for the exemption and
document the need for the exemption.
(d)
The Director must document all findings that are required under Section 1416 of
the Act:
(1)
Before finding that management and restructuring changes cannot be
made, the Director must consider the following measures, and the
availability of State Revolving Loan Fund assistance, or any other Federal
or State program, that is reasonably likely to be available within the period
of the exemption to implement these measures:
(A)
Consideration of rate increases, accounting changes, the
appointment of a State-certified operator under the State’s
Operator Certification program, contractual agreements for joint
operation with one (1) or more PWSs;
(B)
Activities consistent with the State’s Capacity Development
Strategy to help the PWS acquire and maintain technical, financial,
and managerial capacity to come into compliance; and
(C)
Ownership changes, physical consolidation with another PWS, or
other feasible and appropriate means of consolidation which would
result in compliance;
148
(2)
The Director must consider the availability of an alternative source of
water, including the feasibility of partnerships with neighboring PWSs, as
identified by the PWS or by the Director consistent with the Capacity
Development Strategy.
15.2
Variances or exemptions from Maximum Contaminant Level (MCL) to total coliforms or
from any of the treatment technique requirements of Section 5 contained herein will not
be granted.
15.2.1 Exceptions to Section 15.2 with respect to the MCL for total coliforms can be granted if
the system can demonstrate to the Director that:
A.
the violation of the total coliform MCL is due to a persistent growth of total
coliforms in the distribution system;
B.
no fecal or pathogenic contamination exists;
C.
no treatment lapse or deficiency has occurred;
D.
no problem in the operation or maintenance of the distribution system exists.
15.3
Variances and exemptions from the maximum contaminant levels for organic and
inorganic contaminants, radionuclides and the treatment technique for lead and copper.
a)
Community water systems and non-transient, non-community water systems shall
be required to install and/or use any treatment method identified in 16.1 (t), 16.2
(c) and 16.5(h) as a condition for granting a variance except as provided in
Paragraph 15.3(a)(1) of this Section. If, after the systems's installation of the
treatment method, the system cannot meet the MCL, that system shall be eligible
for a variance.
1)
If a system can demonstrate through comprehensive engineering
assessments, which may include pilot plant studies, that the treatment
methods identified in 16.1 (t), 16.2 (c) and 16.5(h) would only achieve a
de minimis reduction in contaminants, the Director may issue a schedule
of compliance that requires the system being granted the variance to
examine other treatment methods as a condition of obtaining the variance.
2)
If the Director determines that a treatment method identified in Paragraph
15.3(a)(1) of this Section is technically feasible, the system will be
required to install and/or use that treatment method in connection with a
compliance schedule. The Director's determination shall be based upon
studies by the system and other relevant information.
149
15.4
In addition to the requirements of 15.3, a PWS may be required to use bottled water,
point-of-use devices, point-of-entry devices or other means as a condition of granting a
variance or an exemption to avoid an unreasonable risk to health. The Director may
require a PWS to use bottled water and point-of-use devices or other means, but not
point-of-entry devices, as a condition for granting an exemption from corrosion control
treatment requirements for lead and copper in Section 6.81 and 6.82 to avoid an
unreasonable risk to health. The Director may require a PWS to use point-of-entry
devices as a condition for granting an exemption for the source water and lead service
line replacement requirements for lead and copper under Section 6.83 or 6.84 to avoid an
unreasonable risk to health.
(a)
A PWS that uses bottled water as a condition for receiving a variance or an
exemption from the requirements of sections 16.1, 16.2(a) 16.2(b) and 16.5 or an
exemption from the requirements of 6.81-6.84, must use bottled water that is
approved by the Director.
(b)
In requiring the use of a point-of-entry device as a condition for granting an
exemption from the treatment requirements for lead and copper under Section
6.83 or 6.84, the Director must be assured that use of the device will not cause
increased corrosion of lead and copper bearing materials located between the
device and the tap that could increase contaminant levels at the tap.
15.5
At the discretion of the Director, nitrate levels not to exceed 20 mg/l may be allowed in a
non-community water system if the supplier of water demonstrates to the satisfaction of
the Director that:
(a)
Such water will not be available to children under 6 months of age; and
(b)
The non-community water system is meeting the public notification requirements
under Section 16.8 (9), including continuous posting of the fact that nitrate levels
exceed 10 mg/l and the potential health effects of exposure; and
(c)
Local and state public health authorities will be notified annually of nitrate levels
that exceed 10 mg/l; and
(d)
No adverse health effects shall result.
150
SECTION 16.0 COMMUNITY WATER SYSTEM REQUIREMENTS
16.1
Inorganic Chemicals
Maximum Contaminant Levels (MCLs) For Certain Inorganic Chemicals
Contaminant
MCL1 (mg/L)
(1)
Fluoride
4.0
(2)
Asbestos
7 million Fibers/liter longer than 10 µm)
(3)
Barium
2
(4)
Cadmium
0.005
(5)
Chromium
0.1
(6)
Mercury
0.002
(7)
Nitrate
10 (as Nitrogen)
(8)
Nitrite
1 (as Nitrogen)
(9)
Total Nitrate and Nitrite
10 (as Nitrogen)
(10)
Selenium
0.05
(11)
Antimony
0.006
(12)
Beryllium
0.004
(13)
Cyanide (as free Cyanide)
0.2
(14)
Nickel
0.1
(15)
Thallium
0.002
(16)
Arsenic
20.010
1 The MCLs for antimony, asbestos, barium, beryllium, cadmium, chromium, cyanide, mercury, nickel, selenium,
and thallium apply to community and non-transient, non-community water systems.
2 The MCL for arsenic is effective January 23, 2006 and applies to community and non-transient, non-community
water systems. Until then the MCL is 0.05 mg/L and applies only to community water systems.
a)
Community water systems shall conduct monitoring to determine compliance
with the MCLs specified in this Section. Monitoring shall be conducted as
follows:
(1)
Groundwater systems shall take a minimum of one (1) sample at every
entry point to the distribution system which is representative of each well
after treatment (hereafter called a sampling point) beginning in the initial
compliance period. The system shall take each sample at the same
sampling point unless conditions make another sampling point more
representative of each source or treatment plant.
(2)
Surface water systems shall take a minimum of one (1) sample at every
entry point to the distribution system after any application of treatment or
in the distribution system at a point which is representative of each source
after treatment (hereafter called a sampling point) beginning in the initial
compliance period. The system shall take each sample at the same
sampling point unless conditions make another sampling point more
representative of each source or treatment plant.
151
Note: For purpose of this Paragraph, surface water systems include
systems with a combination of surface and ground sources.
(3)
If a system draws water from more than one (1) source and the sources are
combined before distribution, the system must sample at an entry point to
the distribution system during periods of normal operating conditions (i.e.,
when water is representative of all sources being used).
(4)
The Director may reduce the total number of samples which must be
analyzed by allowing the use of compositing. Composite samples from a
maximum of five (5) samples are allowed, provided that the detection
limit of the method used for analysis is less than one-fifth of the MCL.
Compositing of samples must be done in the laboratory.
(i)
If the concentration in the composite sample is greater than or
equal to one-fifth of the MCL of any inorganic chemical, then a
follow-up sample must be taken within 14 days at each sampling
point included in the composite. These samples must be analyzed
for the contaminants which exceeded one-fifth of the MCL in the
composite sample. Detection limits for each analytical method are
found in Appendix 1.
(ii)
If the population served by the system is > 3,300 persons, then
compositing may only be permitted by the Director at sampling
points within a single system. In systems serving < 3,300 persons,
the Director may permit compositing among different systems
provided the 5-sample limit is maintained.
(iii)
If duplicates of the original sample taken from each sampling point
used in the composite are available, the system may use these
instead of resampling. The duplicates must be analyzed and the
results reported to the State within 14 days of collection.
(5)
The frequency of monitoring for asbestos shall be in accordance with 16.1
(b); the frequency of monitoring for antimony, arsenic, barium, beryllium,
cadmium, chromium, cyanide, fluoride, mercury, nickel, selenium and
thallium shall be in accordance with 16.1 (c); the frequency of monitoring
for nitrate shall be in accordance with 16.1(d); and the frequency of
monitoring for nitrite shall be in accordance with 16.1(e).
(b)
The frequency of monitoring conducted to determine compliance with the
maximum contaminant level for asbestos shall be conducted as follows:
(1)
Each community and non-transient, non-community water system is
required to monitor for asbestos during the first three-year compliance
152
period of each nine-year compliance cycle beginning in the compliance
period starting January 1, 1993.
(2)
If the system believes it is not vulnerable to either asbestos contamination
in its source water or due to corrosion of asbestos-cement pipe, or both, it
may apply to the Director for a waiver of the monitoring requirement in
Paragraph (b)(1) of this Section. If the Director grants the waiver, the
system is not required to monitor.
(3)
The Director may grant a waiver based on a consideration of the following
factors:
(i)
Potential asbestos contamination of the water source, and
(ii)
The use of asbestos-cement pipe for finished water distribution and
the corrosive nature of the water.
(4)
A waiver remains in effect until the completion of the three-year
compliance period. Systems not receiving a waiver must monitor in
accordance with the provisions of Paragraph (b)(1) of this Section.
(5)
A system vulnerable to asbestos contamination due solely to corrosion of
asbestos-cement pipe shall take one (1) sample at a tap served by asbestos-
cement pipe and under conditions where asbestos contamination is most
likely to occur.
(6)
A system vulnerable to asbestos contamination due solely to source water
shall monitor in accordance with the provision of 16.1(a) of this Section.
(7)
A system vulnerable to asbestos contamination due both to its source
water supply and corrosion of asbestos-cement pipe shall take one (1)
sample at a tap served by asbestos-cement pipe and under conditions
where asbestos contamination is most likely to occur.
(8)
A system which exceeds the maximum contaminant levels shall monitor
quarterly beginning in the next quarter after the violation occurred.
(9)
The Director may decrease the quarterly monitoring requirement to the
frequency specified in Paragraph (b) (1) of this Section provided the
Director has determined that the system is reliably and consistently below
the maximum contaminant level. In no case can the Director make this
determination unless a groundwater system takes a minimum of two (2)
quarterly samples and a surface (or combined surface/ground) water
system takes a minimum of four (4) quarterly samples.
153
(10)
If monitoring data collected after January 1, 1990 is generally consistent
with the requirements of Appendix 1 then the Director may allow systems
to use that data to satisfy the monitoring requirement for the initial
compliance period beginning January 1, 1993.
(c)
The frequency of monitoring conducted to determine compliance with the
maximum contaminant levels in 16.1 for antimony, arsenic, barium, beryllium,
cadmium, chromium, cyanide, fluoride, mercury, nickel, thallium and selenium
shall be as follows:
(1)
Groundwater systems shall take one (1) sample at each sampling point
during each compliance period. Surface water systems (or combines
surface/ground) shall take one (1) sample annually at each sampling point.
(2)
The system may apply to the Director for a waiver from the monitoring
frequencies specified in Paragraph (c) (1) of this Section. The Director
may grant a PWS a waiver for monitoring of cyanide, provided that the
State determines that the system is not vulnerable due to lack of any
industrial source of cyanide.
(3)
A condition of the waiver shall require that a system shall take a minimum
of one (1) sample while the waiver is effective. The term during which the
waiver is effective shall not exceed one (1) compliance cycle (i.e., nine (9)
years).
(4)
The Director may grant a waiver provided surface water systems have
monitored annually for at least three (3) years and groundwater systems
have conducted a minimum of three (3) rounds of monitoring. (At least
one (1) sample shall have been taken since January 1, 1990). Both surface
and groundwater systems shall demonstrate that all previous analytical
results were less than the maximum contaminant level. Systems that use a
new water source are not eligible for a waiver until three (3) rounds of
monitoring from the new source have been completed.
(5)
In determining the appropriate reduced monitoring frequency, the Director
shall consider:
(i)
Reported concentrations from all previous monitoring;
(ii)
The degree of variation in reported concentrations; and
(iii)
Other factors which may affect contaminant concentration such as
changes in groundwater pumping rates, changes in the system's
configuration, changes in the system's operating procedures, or
changes in stream flows or characteristics.
154
(6)
A decision by the Director to grant a waiver shall be made in writing and
shall set forth the basis for the determination. The determination may be
initiated by the Director or upon an application by the PWS. The PWS
shall specify the basis for its request. The Director shall review and, where
appropriate, revise its determination of the appropriate monitoring
frequency when the system submits new monitoring data or when other
data relevant to the system's appropriate monitoring frequency become
available.
(7)
Systems which exceed the maximum contaminant levels as calculated in
16.1 (i ) of this Section shall monitor quarterly beginning in the next
quarter after the violation occurred.
The Director may decrease the quarterly monitoring requirement to the
frequencies specified in Paragraph (c)(1) and (c)(2) of this Section
provided it has determined that the system is reliably and consistently
below the maximum contaminant level. In no case can a Director make
this determination unless a groundwater system takes a minimum of two
(2) quarterly samples and a surface water system take a minimum of four
(4) quarterly samples.
(9)
All new water systems or systems that use a new source of water that
begin operation after January 22, 2004 must demonstrate compliance with
the MCL of all contaminants listed in Section 16.1 within a period of time
specified by the Director. The system must also comply with the initial
sampling frequencies specified by the Director to ensure a system can
demonstrate compliance with the MCL. Routine and increase monitoring
frequencies shall be conducted in accordance with the requirements in this
Section.
(d)
All PWSs (community; non-transient, non-community; and transient, non-
community systems) shall monitor to determine compliance with the maximum
contaminant level for nitrate in Section 16.1.
(1)
Community and non-transient, non-community water systems served by
groundwater systems shall monitor annually; systems served by surface
water shall monitor quarterly beginning January 1, 1993.
(2)
For community and non-transient, non-community water systems, the
repeat monitoring frequency for groundwater systems shall be quarterly
for at least one (1) year following any one (1) sample in which the
concentration is greater than or equal to 50 percent of the MCL. The
Director may allow a groundwater system to reduce the sampling
frequency to annually after four (4) consecutive quarterly samples are
reliably and consistently less than the MCL.
155
(3)
For community and non-transient, non-community water systems, the
Director may allow a surface water system to reduce the sampling
frequency to annually if all analytical results from four (4) consecutive
quarters are < 50 percent of the MCL. A surface water system shall return
to quarterly monitoring if any sample is greater than or equal to 50 percent
of the MCL.
(4)
After the initial round of quarterly sampling is completed, each
community and non-transient non-community system which is monitoring
annually shall take subsequent samples during the quarter(s) which
previously resulted in the highest analytical result.
(e)
All PWSs (community; non-transient, non-community; and transient, non-
community systems) shall monitor to determine compliance with the maximum
contaminant level for nitrite.
(1)
All PWSs shall take a minimum of one (1) sample at each sampling point
in each compliance period.
(2)
After the initial sample, systems where an analytical result for nitrite is <
50 percent of the MCL shall monitor at the frequency specified by the
Director.
(3)
For community, non-transient, non-community, and transient non-
community water systems, the repeat monitoring frequency for any water
system shall be quarterly for at least one (1) year following any one (1)
sample in which the concentration is > 50 percent of the MCL. The
Director may allow a system to reduce the sampling frequency to annually
after determining the system is reliably and consistently less than the
MCL.
(4)
Systems which are monitoring annually shall take each subsequent sample
during the quarter(s) which previously resulted in the highest analytical
result.
(f)
Confirmation Samples:
(1)
Where the results of sampling for asbestos, antimony, arsenic, barium,
beryllium, cadmium, chromium, cyanide, fluoride, mercury, nickel,
selenium or thallium indicate an exceeding of the maximum contaminant
level, the Director may require that one (1) additional sample be collected
as soon as possible after the initial sample was taken (but not to exceed
two (2) weeks) at the same sampling point.
(2)
Where nitrate or nitrite sampling results indicate an exceedance of the
maximum contaminant level, the system shall take a confirmation sample
156
within 24 hours of the system's receipt of notification of the analytical
results of the first sample. Systems unable to comply with the 24-hour
sampling requirement must immediately notify persons served by the
PWS in accordance with Section 16.8 (2) and meet other Tier 1 public
notification requirements under Section 16.8 or 17.6 of this part. Systems
exercising this option must take and analyze a confirmation sample within
two (2) weeks of notification of the analytical results of the first sample.
(3)
If a required confirmation sample is taken for any contaminant, then the
results of the initial and confirmation sample shall be averaged. The
resulting average shall be used to determine the system's compliance in
accordance with Paragraph (i) of this Section. The Director has the
discretion to delete results of obvious sampling errors.
(g)
The Director may require more frequent monitoring than specified in Paragraphs
(b), (c), (d) and (e) of this Section or may require confirmation samples for
positive and negative results at its discretion.
(h)
Systems may apply to the Director to conduct more frequent monitoring than the
minimum monitoring frequencies specified in this Section.
(i)
Compliance with 16.1 shall be determined based on the analytical result(s)
obtained at each sampling point.
(1)
For systems which are conducting monitoring at a frequency greater than
annual, compliance with the maximum contaminant levels for antimony,
arsenic, asbestos, barium, beryllium, cadmium, chromium, cyanide,
fluoride, mercury, nickel, selenium or thallium is determined by a running
annual average at any sampling point. If the average at any sampling point
is greater than the MCL, then the system is out of compliance. If any one
(1) sample would cause the annual average to be exceeded, then the
system is out of compliance immediately. Any sample below the method
detection limit shall be calculated at zero for the purpose of determining
the annual average. Beginning January 22, 2004, if a system fails to
collect
the
required
number
of
samples,
compliance
(average
concentration) will be based on the number of samples collected.
(2)
For systems which are monitoring annually, or less frequently, the system
is out of compliance with the maximum contaminant levels for antimony,
arsenic, asbestos, barium, beryllium, cadmium, chromium, cyanide,
fluoride, mercury, nickel, selenium or thallium if the level of a
contaminant at any sampling point is greater than the MCL. If a
confirmation sample is required by the Director, the determination of
compliance will be based on the annual average of the initial MCL
exceedance and any Director-required confirmation samples. Beginning
January 22, 2004, if a system fails to collect the required number of
157
samples, compliance (average concentration) will be based on the total
number of samples collected.
(3)
Compliance with the maximum contaminant levels for nitrate and nitrite is
determined based on one (1) sample if the levels of these contaminants are
below the MCLs. If the levels of nitrate and/or nitrite exceed the MCLs in
the initial sample, a confirmation sample is required in accordance with
Paragraph (f)(2) of this Section, and compliance shall be determined based
on the average of the initial and confirmation samples.
(j)
Sample collection and analyses for the purpose of determining compliance with
arsenic shall be conducted using the requirements specified in Appendix 1.
(1)
Analyses for all community water systems utilizing surface water sources
shall be repeated at yearly intervals.
(2)
Analyses for all community water systems utilizing only ground water
sources shall be repeated at three-year intervals.
(3)
The Director has the authority to determine compliance or initiate
enforcement action based upon analytical results and other information
compiled by their sanctioned representatives and agencies.
(4)
Until January 23, 2006, the maximum contaminant level for arsenic is 0.05
mg/L and applies to community water systems only. For analyses and
determination of compliance with the 0.05 mg/L maximum contaminant
level for arsenic, use the requirements of this Section 16.1. Beginning
January 23, 2006, the MCL for arsenic for community and non-transient,
non-community water systems is 0.010 mg/L.
(k)
If the result of an analysis made under Paragraph (j) of this Section indicates that
the arsenic concentration exceeds the maximum contaminant level, the supplier of
the water shall report to the Director within 7 days and initiate three (3) additional
analyses at the same sampling point within one (1) month.
(l)
When the average of four (4) analyses made pursuant to Paragraph (k) of this
Section, rounded to the same number of significant figures as the maximum
contaminant level for arsenic exceeds the maximum contaminant level, the
supplier of water shall notify the Director pursuant to 11.2 and give notice to the
public pursuant to 16.8. Monitoring after public notification shall be at a
frequency designated by the Director and shall continue until the maximum
contaminant level has not been exceeded in two (2) successive samples or until a
monitoring schedule as a condition to a variance, exemption or enforcement
action shall become effective.
(m)
Reserved
158
(n)
Reserved
(o)
If a PWS has a distribution system separable from other parts of the distribution
system with no interconnections, the Director may allow the system to give public
notice to only the area served by that portion of the system which is out of
compliance.
(p)
Each PWS shall monitor at the time designated by the Director during each
compliance period.
(q)
Mechanical Fluoride Adjustment - Monitoring Frequency and Reporting
Requirements
i)
For each source where the fluoride concentration is mechanically adjusted,
a fluoride determination of the treated water shall be made and recorded
daily by the water purveyor. Fluoride analysis shall be conducted in
accordance with Appendix 1. Results shall be reported monthly to the
Director within ten (10) days after the end of the month.
ii)
Failure to comply with the requirements of this Paragraph (q) is not
subject to the public notice requirements of Section 16.8.
(r)
Monitoring Protocol for Sodium
Each community system will sample each of its active sources at the entry point
of the source into the distribution system, following any treatment provided to one
(1) or more sources of water, as follows:
Surface water sources shall be sampled during the months of January, February,
and March of each calendar year:
Six (6) consecutive biweekly samples may be composited into a single sample.
Compositing must be done at the laboratory. (Groundwater sources shall be
sampled annually during the months of March or April.)
Samples shall be analyzed for sodium. Results shall be reported to the Director
within ten (10) days after determination. Sodium sampling requirements may be
modified or waived at the discretion of the Director.
(s)
Analytical Techniques - Inorganic chemical analyses shall be made in accordance
with Appendix 1 of these regulations.
(t)
BAT for Inorganic Contaminants
159
The following are hereby identified as the best technology, treatment technique,
or other means available for achieving compliance with the maximum
contaminant level for inorganic contaminants identified in this Section, except
fluoride:
BAT For Inorganic Contaminants Listed in Section 16.1
Chemical Name
BAT(s)
Antimony
2,7
Arsenic 4,5
1,2,5,6,7,9,126
Asbestos
2,3,8
Barium
5,6,7,9
Beryllium
1,2,5,6,7
Cadmium
2,5,6,7
Chromium
2,5,62,7
Cyanide
5,7,10
Mercury
21,4,61,71
Nickel
5,6,7
Nitrate
5,7,9
Nitrite
5,7
Selenium
1,23,6,7,9
Thallium
1,5
1 BAT only if influent Hg concentrations <10µg/L.
2 BAT for Chromium III only.
3 BAT for Selenium IV only.
4 BATs for Arsenic V. Pre-oxidation may be required to convert Arsenic III to Arsenic V.
5 BATs for arsenic become effective January 23, 2006.
6 To obtain high removals, iron to arsenic ratio must be at least 20:1.
Key to BATS in Table
1=Activated Alumina
2=Coagulation/Filtration (not BAT for systems <500 service connections)
3=Direct and Diatomite Filtration
4=Granular Activated Carbon
5=Ion Exchange
6=Lime Softening (not BAT for systems <500 service connections)
7=Reverse Osmosis
8=Corrosion Control
9=Electrodialysis
10=Chlorine
11=Ultraviolet
12=Oxidation/Filtration
(u)
The Director hereby identifies in the following table the affordable technology,
treatment technique, or other means available to systems serving 10,000 persons
or fewer for achieving compliance with the maximum contaminant level for
arsenic effective January 23, 2006:
160
Small System Compliance Technologies (SSCTS)1 for Arsenic2
Small System Compliance Technology
Affordable for listed small system
categories3
Activated Alumina (centralized)
All size categories.
Activated Alumina (Point-of-Use)4
All size categories.
Coagulation/Filtration5
501–3,300, 3,301–10,000.
Coagulation-assisted Microfiltration
501–3,300, 3,301–10,000.
Electrodialysis reversal6
501–3,300, 3,301–10,000.
Enhanced coagulation/filtration
All size categories
Enhanced lime softening (pH> 10.5)
All size categories
Ion Exchange
All size categories
Lime Softening5
501–3,300, 3,301–10,000.
Oxidation/Filtration7
All size categories
Reverse Osmosis (centralized)6
501–3,300, 3,301–10,000.
Reverse Osmosis (Point-of-Use)4
All size categories
1 Section 1412(b)(4)(E)(ii) of SDWA specifies that SSCTs must be affordable and technically feasible for small
systems.
2 SSCTs for Arsenic V. Pre-oxidation may be required to convert Arsenic III to Arsenic V.
3 The Act (ibid.) specifies three (3) categories of small systems: (i) those serving 25 or more, but fewer than 501, (ii)
those serving more than 500, but fewer than 3,301, and (iii) those serving more than 3,300, but fewer than 10,001.
4 When POU or POE devices are used for compliance, programs to ensure proper long-term operation, maintenance,
and monitoring must be provided by the water system to ensure adequate performance.
5 Unlikely to be installed solely for arsenic removal. May require pH adjustment to optimal range if high removals
are needed.
6 Technologies reject a large volume of water—may not be appropriate for areas where water quantity may be an
issue.
7 To obtain high removals, iron to arsenic ratio must be at least 20:1
16.2
Organic Chemicals
(a)
Maximum contaminant levels for organic contaminants
Contaminant
MCL (mg/L)
Alachlor
0.002
Aldicarb
reserved
Aldicarb sulfoxide
reserved
Aldicarb sulfone
reserved
Altrazine
0.003
Carbofuran
0.04
Chlordane
0.002
Dibromochloropropane
0.0002
2,4-D
0.07
Ethylene dibromide
0.00005
Heptachlor
0.0004
Heptachlor epoxide
0.0002
Lindane
0.0002
Methoxychlor
0.04
Polychlorinated biphenyls
0.0005
161
Contaminant
MCL (mg/L)
Pentachlorophenol
0.001
Toxaphene
0.003
2,4,5-TP
0.05
Benzo[a]pyrene
0.0002
Dalapon
0.2
Di(2-ethylhexyl) adipate
0.4
Di(2-ethylhexyl) phthalate
0.006
Dinoseb
0.007
Diquat
0.02
Endothall
0.1
Endrin
0.002
Glyphosate
0.7
Hexacholorbenzene
0.001
Hexachlorocyclopentadiene
0.05
Oxamyl (Vydate)
0.2
Picloram
0.5
Simazine
0.004
2,3,7,8-TCDD (Dioxin)
3x10-8
Total Trihalomethanes
0.10
The maximum contaminant level of 0.10 mg/L for total trihalomethanes (the sum
of the concentrations of bromodichloromethane, dibromochloromethane,
tribromomethane (bromoform) and trichloromethane (chloroform) applies to
subpart H community water systems which serve a population of 10,000 people or
more until December 31, 2001. After December 31, 2001, Section 7.0 will apply.
This level also applies to community water systems that use only ground water
not under the direct influence of surface water and serve a population of 10,000
people or more until December 31, 2003. Compliance with the maximum
contaminant level for total trihalomethanes is calculated pursuant to Section 16.2
(a)(19). After December 31, 2003, this Section is no longer applicable and Section
7.0 will apply.
Analysis of the contaminants listed in 16.2 (a) for the purposes of determining
compliance with the maximum contaminant level shall be conducted as follows
except that monitoring for the contaminants aldicarb, aldicarb sulfoxide and
aldicarb sulfone shall be conducted in accordance with Section 16.7:
(1)
Groundwater systems shall take a minimum of one (1) sample at every
entry point to the distribution system which is representative of each well
after treatment (hereafter called a sampling point). Each sample must be
taken at the same sampling point unless conditions make another sampling
point more representative of each source or treatment plant.
(2)
Surface water systems shall take a minimum of one (1) sample at points in
the distribution system that are representative of each source or at each
entry point to the distribution system after treatment (hereafter called a
sampling point.) Each sample must be taken at the same sampling point
162
unless conditions make another sampling point more representative of
each source or treatment plant.
Note: For purposes of this Paragraph, surface water systems include
systems with a combination of surface and ground sources.
(3)
If the system draws water from more than one (1) source and the sources
are combined before distribution, the system must sample at an entry point
to the distribution system during periods of normal operating conditions
(i.e., when water representative of all sources if being used).
(4)
Monitoring frequency:
(i)
Each community and non-transient non-community water system
shall take four (4) consecutive quarterly samples for each
contaminant listed in 16.2 (a) during each compliance period
beginning with the initial compliance period.
(ii)
Systems serving more than 3,300 persons which do not detect a
contaminant in the initial compliance period may reduce the
sampling frequency to a minimum of two (2) quarterly samples in
one (1) year during each repeat compliance period.
(iii)
Systems serving less than or equal to 3,300 persons which do not
detect a contaminant in the initial compliance period may reduce
the sampling frequency to a minimum of one (1) sample during
each repeat compliance period.
(5)
Each community and non-transient non-community water system may
apply to the Director for a waiver from the requirement of Paragraph (h)
(4) of this Section. A system must reapply for a waiver for each
compliance period.
(6)
The Director may grant a waiver after evaluating the following factor(s):
Knowledge of previous use (including transport, storage, or disposal) of
the contaminant within the watershed or zone of influence of the system.
If a determination by the Director reveals no previous use of the
contaminant within the watershed or zone of influence, a waiver may be
granted. If previous use of the contaminant is unknown or it has been used
previously, then the following factors shall be used to determine whether a
waiver is granted.
(i)
Previous analytical results.
(ii)
The proximity of the system to a potential point or non-point
source of contamination. Point sources include spills and leaks of
163
chemicals at or near a water treatment facility or at manufacturing,
distribution, or storage facilities, or from hazardous and municipal
waste landfills and other waste handling or treatment facilities.
Non-point sources include the use of pesticides to control insect
and weed pests on agricultural areas, forest lands, home and
gardens, and other land application uses.
(iii)
The environmental persistence and transport of the pesticide or
PCBs.
(iv)
How well the water source is protected against contamination due
to such factors as depth of the well and the type of soil and the
integrity of the well casing.
(v)
Elevated nitrate levels at the water supply source.
(vi)
Use of PCBs in equipment used in the production, storage, or
distribution of water (i.e., PCBs used in pumps, transformers, etc.).
(7)
If an organic contaminant listed in 16.2 (a) is detected (as defined by
Paragraph (a) (17) of this Section) in any sample, then:
(i)
Each system must monitor quarterly at each sampling point which
resulted in a detection.
(ii)
The Director may decrease the quarterly monitoring requirement
specified in Paragraph (a) (7) (i) of this Section provided it has
determined that the system is reliably and consistently below the
maximum contaminant level. In no case shall the Director make
this determination unless a groundwater system takes a minimum
of two (2) quarterly samples and a surface water system takes a
minimum of four (4) quarterly samples.
(iii)
After the Director determines the system is reliably and
consistently below the maximum contaminant level the Director
may allow the system to monitor annually. Systems which monitor
annually must monitor during the quarter that previously yielded
the highest analytical result.
(iv)
Systems which have 3 consecutive annual samples with no
detection of a contaminant may apply to the Director for a waiver
as specified in Paragraph (a) (6) of this Section.
(v)
If monitoring results in detection of one (1) or more of certain
related contaminants (aldicarb, aldicarb sulfone, aldicarb sulfoxide
164
and heptachlor, heptachlor epoxide), than subsequent monitoring
shall analyze for all related contaminants.
(8)
Systems which violate the requirements of 16.2 (a) as determined by
Paragraph (a) (11) of this Section must monitor quarterly. After a
minimum of four (4) quarterly samples show the system is in compliance
and the Director determines the system is reliably and consistently below
the MCL, as specified in Paragraph (a) (11) of this Section, the system
shall monitor at the frequency specified in Paragraph (a) (7) (iii) of this
Section.
(9)
The Director may require a confirmation sample for positive or negative
results. If a confirmation sample is required by the Director, the result
must be averaged with the first sampling result and the average used for
the compliance determination as specified by Paragraph (a) (11) of this
Section. The Director has discretion to delete results of obvious sampling
errors from this calculation.
(10)
The Director may reduce the total number of samples a system must
analyze by allowing the use of compositing. Composite samples from a
maximum of five (5) sampling points are allowed, provided that the
detection limit of the method used for analysis is less than one-fifth of the
MCL. Compositing of samples must be done in the laboratory and
analyzed within 14 days of sample collection.
(i)
If the concentration in the composite sample is greater than or
equal to 0.0005 mg/L for any contaminant listed in Section 16.2(a),
then a follow-up sample must be taken within 14 days at each
sampling point included in the composite and be analyzed for that
contaminant.
(ii)
If duplicates of the original sample taken from each sampling point
used in the composite are available, the system may use these
duplicates instead of resampling. The duplicate must be analyzed
and the results reported to the Director within 14 days of
collection.
(iii)
If the population served by the system is > 3,300 persons, then
compositing may only be permitted by the Director at sampling
points within a single system. In systems serving less than or equal
to 3,300 persons, the Director may permit compositing among
different systems provided the 5-sample limit is maintained.
(11)
Compliance with 16.2 (a) shall be determined based on the analytical
results obtained at each sampling point. If one (1) sampling point is in
violation of an MCL, the system is in violation of the MCL.
165
(i)
For systems which are conducting monitoring at a frequency
greater than annual, compliance is determined by a running annual
average of all samples taken at each sampling point. If the annual
average of any sampling point is greater than the MCL, then the
system is out of compliance. If the initial sample or a subsequent
sample would cause the annual average to be exceeded, then the
system is out of compliance immediately.
(ii)
Systems monitoring annually or less frequently whose sample
result exceeds the regulatory detection level as defined by 16.2
(a)(17) must begin quarterly sampling. The system will not be
considered in violation of the MCL until it has completed one (1)
year of quarterly sampling. Effective January 22, 2004, the
following statements no longer apply: If monitoring is conducted
annually, or less frequently, the system is out of compliance if the
level of a contaminant at any sampling point is greater than the
MCL. If a confirmation sample is required by the Director, the
determination of compliance will be based on the average of two
(2) samples.
(iii)
If any sample result will cause the running annual average to
exceed the MCL at any sampling point, the system is out of
compliance with the MCL immediately.
(iv)
If a system fails to collect the required number of samples,
compliance will be based on the total number of samples collected.
(v)
If a sample result is less than the detection limit, zero will be used
to calculate the annual average.
(vi)
If a PWS has a distribution system separable from other parts of
the distribution system with no interconnections, the Director may
allow the system to give public notice to only that area served by
that portion of the system which is out of compliance.
(12)
Analysis for the contaminants listed in 16.2 (a) shall be conducted using
the EPA methods or their equivalent as approved by EPA and as described
in Appendix 1.
(13)
If monitoring data collected after January 1, 1990, is generally consistent
with the requirements of 16.2 (a) then the Director may allow systems to
use that data to satisfy the monitoring requirement for the initial
compliance period.
166
(14)
The Director may increase the required monitoring frequency, where
necessary, to detect variations within the system (e.g., fluctuations in
concentration due to seasonal use, changes in water source).
(15)
The Director has the authority to determine compliance or initiate
enforcement action based upon analytical results and other information
compiled by their sanctioned representatives and agencies.
(16)
Each PWS shall monitor at the time designated by the Director within
each compliance period.
Detection as used in this Paragraph shall be defined as greater than or
equal to the following concentrations for each contaminant.
Contaminant
Detection limit (mg/L)
Alachlor
0.0002
Aldicarb
0.0005
Aldicarb sulfoxide
0.0005
Aldicarb sulfone
0.0008
Atrazine
0.0001
Benzo[a]pyrene
0.00002
Carbofuran
0.0009
Chlordane
0.0002
Dalapon
0.001
Dibromochloropropane (DBCP)
0.00002
Di (2-ethylhexyl) adipate
0.0006
Di (2-ethylhexyl) phthalate
0.0006
Dinoseb
0.0002
Diquat
0.0004
2,4-D
0.0001
Endothall
0.009
Endrin
0.00001
Ethylene dibromide (EDB)
0.00001
Glyphosate
0.006
Heptachlor
0.00004
Heptachlor epoxide
0.00002
Hexachlorobenzene
0.0001
Hexachlorocyclopentadiene
0.0001
Lindane
0.00002
Methoxychlor
0.0001
Oxamyl
0.002
Picloram
0.0001
Polychlorinated biphenyls (PCBs) (as decachlorobiphenyl)
0.0001
Pentachlorophenol
0.00004
Simazine
0.00007
Toxaphene
0.001
2,3,7,8-TCDD (Dixon)
0.000000005
2,4,5-TP (Silvex)
0.0002
167
(18)
Notwithstanding Paragraphs (1) through (17) of this Section, monitoring
for endrin shall be as stipulated here.
Monitoring Frequency - Each active drinking water source maintained by
a water purveyor shall be analyzed for endrin at least once every three (3)
years.
(19)
Total Trihalomethanes Sampling, Analytical and Other Requirements
Water systems which practice disinfection of the water shall be monitored
for total trihalomethanes.
Notwithstanding Paragraphs (1) through (17) of this Section, monitoring
for total trihalomethanes shall be as stipulated here.
Water systems serving less than 10,000 people shall be monitored for total
trihalomethanes at the discretion of the Director. Water systems serving at
least 10,000 people shall be monitored in the following manner:
i)
a minimum of four (4) samples per quarter per treatment plant
collected on the same day shall be analyzed. Twenty-five percent
(25%) of the samples shall reflect maximum storage time of the
water in the distribution system while seventy-five (75%) percent
shall be collected at representative points in the system. The results
of all samples analyzed in any calendar quarter shall be averaged
and compliance with the TTHM MCL listed Table 16.2 (a) shall be
determined based on this running annual average of quarterly
samples collected by the system;
ii)
compliance with the maximum contaminant level for total
trihalomethanes shall be based on a running annual average of the
findings in any four (4) consecutive calendar quarters. Based on a
history of low trihalomethanes, the Director may grant a reduction
in monitoring frequency to one (1) sample per treatment plant per
quarter collected at a point which reflects maximum storage time
of the water in the distribution system;
iii)
additional monitoring shall be required whenever there is reason to
believe an organic chemical maximum contaminant level is or may
be exceeded.
iv)
Upon written request to the Director, a community water system
utilizing only ground water sources may seek to have the
monitoring frequency required by Paragraph (19)(i) of this Section
reduced to a minimum of one (1) sample for maximum TTHM
168
potential per year for each treatment plant used by the system
taken at a point in the distribution system reflecting maximum
residence time of the water in the system. The system shall submit
the results of at least one (1) sample for maximum TTHM potential
using the procedure specified in Paragraph (19)(vi) of this Section.
A sample must be analyzed from each treatment plant used by the
system and be taken at a point in the distribution system reflecting
the maximum residence time of the water in the system. The
system's monitoring frequency may only be reduced upon a written
determination by the Director that, based upon the data submitted
by the system, the system has a maximum TTHM potential of less
than 0.10 mg/L and that, based upon an assessment of the local
conditions of the system, the system is not likely to approach or
exceed the maximum contaminant level for total TTHMs. The
results of all analyses shall be reported to the Director within 30
days of the system's receipt of such results. Results shall also be
reported to EPA until such monitoring requirements have been
adopted by the Director. All samples collected shall be used for
determining whether the system must comply with the monitoring
requirements of Paragraph (19)(i) of this Section, unless the
analytical results are invalidated for technical reasons. Sampling
and analyses shall be conducted in accordance with the methods
listed in Paragraph (19)(viii) of this Section. If at any time the
results from any analysis taken by the system for maximum TTHM
potential are equal to or greater than 0.10 mg/L, and such results
are confirmed by at least one (1) check sample taken promptly
after such results are received, the system shall immediately begin
monitoring in accordance with the requirements of Paragraph
(19)(i) of this Section and such monitoring shall continue for at
least one (1) year before the frequency may be reduced again. In
the event of any significant change to the system's raw water or
treatment program, the system shall immediately analyze an
additional sample for maximum TTHM potential taken at a point
in the distribution system reflecting maximum residence time of
the water in the system for the purpose of determining whether the
system must comply with the monitoring requirements of
Paragraph (19)(i) of this Section. At the option of the Director,
monitoring frequencies may and should be increased above the
minimum in those cases where this is necessary to detect variation
of TTHM levels within the distribution system.
Before a community water system makes any significant
modifications to its existing treatment process for the purposes of
achieving compliance with Section 16.2 (a), such system must
submit and obtain the Director's approval of a detailed plan setting
forth its proposed modification and those safeguards that it will
169
implement to ensure that the bacteriological quality of the drinking
water served by such system will not be adversely affected by such
modification.
vi)
The water sample for determination of maximum total
trihalomethane potential is taken from a point in the distribution
system that reflects maximum residence time. Procedures for
sample collection and handling are given in the methods. No
reducing agent is added to “quench” the chemical reaction
producing THMs at the time of sample collection. The intent is to
permit the level of THM precursors to be depleted and the
concentration of THMs to be maximized for the supply being
tested. Four (4) experimental parameters affecting maximum THM
production are pH, temperature, reaction time and the presence of a
disinfectant residual. These parameters are dealt with as follows:
Measure the disinfectant residual at the selected sampling point.
Proceed only if a measurable disinfectant residual is present.
Collect triplicate 40 ml water samples at the pH prevailing at the
time of sampling, and prepare a method blank according to the
methods. Seal and store these samples together for seven (7) days
at 25 °C or above. After this time period, open one (1) of the
sample containers and check for disinfectant residual. Absence of a
disinfectant residual invalidates the sample for further analysis.
Once a disinfectant residual has been demonstrated, open another
of the sealed samples and determine total THM concentration
using an approved analytical method.
vii)
The requirements in Paragraph (19) of this Section apply to
subpart H community water systems which serve a population of
10,000 or more until December 31, 2001. After December 31,
2001, Section 7.0 will apply. The requirements in Paragraph 19 of
this Section apply to community water systems which use only
ground water not under the direct influence of surface water that
add a disinfectant (oxidant) in any part of the treatment process
and serve a population of 10,000 or more until December 31, 2003.
After December 31, 2003, Paragraph 16.2(a)(19) is no longer
applicable and Section 7.0 will apply.
viii)
Analytical Techniques -Sampling and analyses made pursuant to
this Section shall be conducted by one (1) of the total
trihalomethanes methods as directed in Appendix 1 and the
Technical Notes on Drinking Water Methods, EPA-600/R-94-173,
October 1994, which is available from NTIS, PB-104766, or in
Section 7.4 (b).
170
(20)
All new systems or systems that use a new source of water that begin
operation after January 22, 2004 must demonstrate compliance with the
MCL in 16.2(a) within a period of time specified by the Director. The
system must also comply with the initial sampling frequencies specified
by the Director to ensure a system can demonstrate compliance with the
MCL. Routine and increased monitoring frequencies shall be conducted in
accordance with the requirements in this Section beginning January 22,
2004.
16.2 (b) Volatile Organic Chemicals
Maximum contaminant levels for certain volatile organic chemicals:
Contaminant
MCL (mg/L)
(1) Vinyl Chloride
0.002
(2) Benzene
0.005
(3) Carbon Tetrachloride
0.005
(4) 1,2-Dichloroethane
0.005
(5) Trichloroethylene
0.005
(6) p-Dichlorobenzene
0.075
(7) 1,1-Dichloroethylene
0.007
(8) 1,1,1-Trichloroethane
0.2
(9) cis-1,2-Dichloroethylene
0.07
(10) 1,2-Dichloropropane
0.005
(11) Ethylbenzene
0.7
(12) Monochlorobenzene
0.1
(13) o-Dichlorobenzene
0.6
(14) Styrene
0.1
(15) Tetrachloroethylene
0.005
(16) Toluene
1
(17) trans-1,2-Dichloroethylene
0.1
(18) Xylenes (total)
10
(19) Dichloromethane
0.005
(20) 1,2,4-Trichlorobenzene
0.07
(21) 1,1,2-Trichloroethane
0.005
Beginning with the initial compliance period, analysis of the contaminants listed in 16.2 (b) (1)
through (21) for the purpose of determining compliance with the maximum contaminant level
shall be conducted as follows:
(22)
Groundwater systems shall take a minimum of one (1) sample at every
entry point to the distribution system which is representative of each well
after treatment (hereafter called a sampling point). Each sample must be
taken at the same sampling point unless conditions make another sampling
point more representative of each source, treatment plant, or within the
distribution system.
171
(23)
Surface water systems (or combined surface/ground) shall take a
minimum of one (1) sample at points in the distribution system that are
representative of each source or at each entry point to the distribution
system after treatment (hereafter called a sampling point). Each sample
must be taken at the same sampling point unless conditions make another
sampling point more representative of each source, treatment plant, or
within the distribution system.
(24)
If the system draws water from more than one (1) source and the sources
are combined before distribution, the system must sample at an entry point
to the distribution system during periods of normal operating conditions
(i.e., when water representative of all sources if being used).
(25)
Each community and non-transient, non-community water system shall
take four (4) consecutive quarterly samples for each contaminant listed in
16.2 (b) (2) through (21) during each compliance period, beginning in the
initial compliance period.
(26)
If the initial monitoring for contaminants listed in 16.2 (b) (1) through (8)
and the monitoring for the contaminants listed in 16.2 (b) (9) through (21)
as allowed in Paragraph 16.2 (b) (37) has been completed by December
31, 1992, and the system did not detect any contaminant listed in 16.2 (b)
(1) through (21), than each ground and surface water system shall take one
(1) sample annually beginning with the initial compliance period.
(27)
After a minimum of three (3) years of annual sampling, the Director may
allow groundwater systems with no previous detection of any contaminant
listed in 16.2 (b) to take one (1) sample during each compliance period.
(28)
Each community and non-transient groundwater system which does not
detect a contaminant listed in 16.2 (b) (1) through (21) may apply to the
Director for a waiver from the requirements of Paragraphs (26) and (27) of
this Section after completing the initial monitoring. (For the purposes of
this Section, detection is defined as greater than or equal to 0.0005 mg/L).
A waiver shall be effective for no more than six (6) years (two (2)
compliance periods). The Director may also issue waivers to small
systems for the initial round of monitoring for 1,2,4-trichlorobenzene.
(29)
The Director may grant a waiver after evaluating the following factor(s):
(i)
Knowledge of previous use (including transport, storage, or
disposal) of the contaminant within the watershed or zone
influence of the system. If a determination by the Director reveals
no previous use of the contaminant within the watershed or zone of
influence, a waiver may be granted.
172
(ii)
If previous use of the contaminant is unknown or it has been used
previously, then the factors below shall be used to determine
whether a waiver is granted.
(A)
Previous analytical results;
(B)
The proximity of the system to a potential point or non-
point source of contamination. Point sources include spills
and leaks of chemicals at or near a water treatment facility
or at manufacturing, distribution, or storage facilities, or
from hazardous and municipal waste landfills and other
waste handling or treatment facilities;
(C)
The environmental persistence and transport of the
contaminants;
(D)
The number of persons served by the PWS and the
proximity of a smaller system to a larger system; and
(E)
How well the water source is protected against
contamination, such as whether it is a surface or
groundwater system. Groundwater systems must consider
factors such as depth of the well, the type of soil and
wellhead protection. Surface water systems must consider
watershed protection;
(30)
As a condition of the waiver a groundwater system must take one (1)
sample at each sampling point during the time the waiver is effective (i.e.,
one (1) sample during two (2) compliance periods or six (6) years) and
update its vulnerability assessment considering the factors listed in
Paragraph (29) of this Section. Based on this vulnerability assessment the
Director must reconfirm that the system is non-vulnerable. If the Director
does not make this reconfirmation within three (3) years of the initial
determination, then the waiver is invalidated and the system is required to
sample annually as specified in Paragraph (26) of this Section.
(31)
Each community and non-transient surface water system which does not
detect a contaminant listed in 16.2 (b) (1) through (21) may apply to the
Director for a waiver from the requirements of (26) of this Section after
completing the initial monitoring. Composite samples from a maximum of
five (5) sampling points are allowed, provided that the detection limit of
the method used for analysis is less than one-fifth of the MCL. Systems
meeting this criterion must be determined by the Director to be non-
vulnerable based on a vulnerability assessment during each compliance
173
period. Each system receiving a waiver shall sample at the frequency
specified by the Director (if any).
(32)
If a contaminant listed in 16.2 (b) (2) through (21) is detected at a level
exceeding 0.0005 mg/L in any sample, then:
(i)
The system must monitor quarterly at each sampling point which
resulted in a detection.
(ii)
The Director may decrease the quarterly monitoring requirement
specified in Paragraph (32) (i) of this Section provided it has
determined that the system is reliably and consistently below the
maximum contaminant level. In no case shall the Director make
this determination unless a groundwater system takes a minimum
of two (2) quarterly samples and a surface water system takes a
minimum of four (4) quarterly samples.
(iii)
If the Director determines that the system is reliably and
consistently below the MCL, the Director may allow the system to
monitor annually. Systems which monitor annually must monitor
during the quarter(s) which previously yielded the highest
analytical result.
(iv)
Systems which have three (3) consecutive annual samples with no
detection of a contaminant may apply to the Director for a waiver
as specified in Paragraph (28) of this Section.
(v)
Groundwater systems which have detected one (1) or more of the
following two-carbon organic compounds: trichloroethylene,
tetrachloroethylene, 1,2-dichloroethane, 1,1,1-trichloroethane, cis-
1,2-dichloroethylene,
trans-1,2-dichloroethylene,
or
1,1-
dichloroethylene shall monitor quarterly for vinyl chloride. A vinyl
chloride sample shall be taken at each sampling point at which one
(1) or more of the two (2) carbon organic compounds was detected.
If the results of the first analysis do not detect vinyl chloride, the
Director may reduce the quarterly monitoring frequency of vinyl
chloride monitoring to one (1) sample during each compliance
period. Surface water systems are required to monitor for vinyl
chloride as specified by the Director.
(33)
Systems which violate the requirements of 16.2 (b) (1) through (21), as
determined by Paragraph (36) of this Section, must monitor quarterly.
After a minimum of four (4) consecutive quarterly samples which show
the system is in compliance as specified in Paragraph (36) of this Section
the system and the Director determines the system and the Director
determines that the system is reliably and consistently below the
174
maximum contaminant level, the system may monitor at the frequency and
time specified in Paragraph (32) (iii) of this Section.
(34)
The Director may require a confirmation sample for positive or negative
results. If a confirmation sample is required by the Director, the result
must be average with the first sampling result and the average is used for
the compliance determination as specified by Paragraph (36). The Director
has discretion to delete results of obvious sampling errors from this
calculation.
(35)
The Director may reduce the total number of samples a system must
analyze by allowing the use of compositing. Composite samples from a
maximum of five (5) sampling points are allowed provided that the
detection limit of the method used for analysis is less than one-fifth of the
MCL. Compositing of samples must be done in the laboratory and
analyzed within 14 days of sample collection.
(i)
If the concentration in the composite sample detects one (1) or
more contaminants listed in Section 16.2 (b), then a follow-up
sample must be taken within 14 days at each sampling point
included in the composite, and be analyzed for that contaminant.
(ii)
If duplicates of the original sample taken from each sampling point
used in the composite are available, the system may use these
duplicates instead of resampling. The duplicate must be analyzed
and the results reported to the State within 14 days of collection.
(iii)
If the populations served by the system is >3,300 persons, then
compositing may only be permitted by the Director at sampling
points within a single system. In systems serving 3,300 persons,
the Director may permit compositing among different systems
provided the 5-sample limit is maintained.
(36)
Compliance with 16.2 (b) (1) through (21) shall be determined based on
the analytical results obtained at each sampling point. If one (1) sampling
point is in violation of an MCL, the system is in violation of the MCL.
(i)
For systems which are conducting monitoring at a frequency
greater than annual, compliance is determined by a running annual
average of all samples taken at each sampling point. If the annual
average of any sampling point is greater than the MCL, then the
system is out of compliance. If the initial sample or a subsequent
sample would cause the annual average to be exceeded, then the
system is out of compliance immediately.
175
(ii)
Systems monitoring annually or less frequently whose sample
result exceeds the MCL must begin quarterly sampling. The
system will not be considered in violation of the MCL until it has
completed one (1) year of quarterly sampling. Effective January
22, 2004, the following statements no longer apply: If monitoring
is conducted annually, or less frequently, the system is out of
compliance if the level of a contaminant at any sampling point is
greater than the MCL. If a confirmation sample is required by the
Director, the determination of compliance will be based on the
average of two (2) samples.
(iii)
If any sample result will cause the running annual average to
exceed the MCL at any sampling point, the system is out of
compliance with the MCL immediately.
(iv)
If a system fails to collect the required number of samples,
compliance will be based on the total number of samples collected.
(v)
If a sample result is less than the detection limit, zero will be used
to calculate the annual average.
(vi)
If a PWS has a distribution system separable from other parts of
the distribution system with no interconnections, the Director may
allow the system to give public notice to only that area served by
that portion of the system which is out of compliance.
(37)
Analysis for the contaminants listed in 16.2 (b) (1) through (21) shall be
conducted using EPA methods or their equivalent as approved by EPA
and as specified in Appendix 1.
(38)
The Director may allow the use of monitoring data collected after January
1, 1988, for purposes of initial monitoring compliance. If the data is
generally consistent with the other requirements in this Section, the
Director may use this data (i.e., a single sample rather than four (4)
quarterly samples) to satisfy the initial monitoring requirement of
Paragraph (4) of this Section. Systems which use grand fathered samples
and did not detect any contaminant listed in 16.2 (b) (1) through (21) shall
begin monitoring annually in accordance with Paragraph (26) of this
Section beginning with the initial compliance period.
(39)
The Director may increase required monitoring where necessary to detect
variations within the system.
(40)
Each PWS shall monitor at the time designated by the Director within
each compliance period.
176
(41)
All new systems or systems that use a new source of water that begin
operation after January 22, 2004 must demonstrate compliance with the
MCL in 16(b)(1) through (21) within a period of time specified by the
Director. The system must also comply with the initial sampling
frequencies specified by the Director to ensure a system can demonstrate
compliance with the MCL. Routine and increased monitoring frequencies
shall be conducted in accordance with the requirements in this Section
beginning January 22, 2004.
(42)
Reserved.
(43)
Bottled water may be used on a temporary basis to avoid an unreasonable
risk to health. If bottled water is used, it must be obtained from an
approved source. A PWS shall not use bottled water to achieve
compliance with a maximum contaminant level listed in Section (b) unless
required by the Director as a condition for granting an exemption and
providing there are reasonable assurances that the bottled water will not
exceed maximum contaminant levels.
(44)
Compliance with a maximum contaminant level shall be achieved by
installation of central treatment using BAT as stipulated in Section (c).
Point-of-use or point of entry devices may be used only as a condition for
obtaining a variance from the requirement for adoption of central
treatment providing the devices and a monitoring plan for their
maintenance are approved by the Director prior to their installation, and
that every building connected to the water system has a device installed,
maintained and adequately monitored by the PWS.
(45)
Reserved.
16.2 (c)
BAT for Organic Contaminants
The following table identifies granular activated carbon (GAC), packed tower aeration
(PTA), or oxidation (OX) as the best technology, treatment technique, or other means
available for achieving compliance with the maximum contaminant level for organic
contaminants identified in Paragraphs (a) and (b) of this Section:
177
BAT for Organic Contaminants Listed in Sections 16.2(a) and (b)
CAS No.
CONTAMINANT
GAC
PTA
OX
15972-60-8
Alachlor
X
116-06-3
Aldicarb
X
1646-88-4
Aldicarb sulfone
X
1646-87-3
Aldicarb sulfoxide
X
1912-24-9
Atrazine
X
71-43-2
Benzene
X
X
50-32-8
Benzo[a]pyrene
X
1563-66-2
Carbofuran
X
56-23-5
Carbon tetrachloride
X
X
57-74-9
Chlordane
X
75-99-0
Dalapon
X
94-75-7
2,4-D
X
103-23-1
Di(2-ethylhexyl) adipate
X
X
117-81-7
Di (2-ethylhexyl) phthalate
X
96-12-8
Dibromochloropropane (DBCP)
X
X
95-50-1
o-Dichlorobenzene
X
X
106-46-7
para-Dichlorobenzene
X
X
107-06-2
1,2-Dichloroethane
X
X
75-35-4
1,1-Dichloroethylene
X
X
156-59-2
cis-1,2-Dichloroethylene
X
X
156-60-5
trans-1,2-Dichloroethylene
X
X
75-09-2
Dichloromethane
X
78-87-5
1,2-Dichloropropane
X
X
88-85-7
Dinoseb
X
85-00-7
Diquat
X
145-73-3
Endothall
X
72-20-8
Endrin
X
100-41-4
Ethylbenzene
X
X
106-93-4
Ethylene Dibromide (EDB)
X
X
1071-83-6
Gylphosate
X
76-44-8
Heptachlor
X
1024-57-3
Heptachlor epoxide
X
118-74-1
Hexachlorobenzene
X
77-47-3
Hexachlorocyclopentadiene
X
X
58-89-9
Lindane
X
72-43-5
Methoxychlor
X
108-90-7
Monochlorobenzene
X
X
23135-22-0
Oxamyl (Vydate)
X
87-86-5
Pentachlorophenol
X
1918-02-1
Picloram
X
1336-36-3
Polychlorinated biphenyls (PCB)
X
122-34-9
Simazine
X
100-42-5
Styrene
X
X
1746-01-6
2,3,7,8-TCDD (Dioxin)
X
127-18-4
Tetrachloroethylene
X
X
108-88-3
Toluene
X
X
8001-35-2
Toxaphene
X
93-72-1
2,4,5-TP (Silvex)
X
178
BAT for Organic Contaminants Listed in Sections 16.2(a) and (b)
CAS No.
CONTAMINANT
GAC
PTA
OX
120-82-1
1,2,4-Trichlorobenzene
X
X
71-55-6
1,1,1-Trichloroethane
X
X
79-00-5
1,1,2-Trichloroethane
X
X
79-01-6
Trichloroethylene
X
X
75-01-4
Vinyl chloride
X
1330-20-7
Xylene
X
X
d)
Treatment techniques for acrylamide and epichlorohydrin.
Each PWS must certify annually in writing to the Director (using third party or
manufacturer's certification) that when acrylamide and epichlorohydrin are used
in drinking water systems, the combination (or product) of dose and monomer
level does not exceed the levels specified as follows:
Acrylamide = 0.05% dosed at 1 ppm (or equivalent)
Epichlorohydrin = 0.01% dosed at 20 ppm (or equivalent)
Certifications can rely on manufacturers or third parties, as approved by the
Director.
16.3 Turbidity
a)
Applicability
The maximum contaminant level for turbidity applies only to surface water
sources. The turbidity of the water shall be determined and recorded daily by the
water purveyor and measured at a representative entry point into the distribution
system.
The requirements in Section 16.3(a) and (b) apply to unfiltered systems until
December 30, 1991 unless the Director has determined prior to that date, in
writing pursuant to Section 5 that filtration is required. The requirements in this
Section apply to filtered systems until June 29, 1993. The requirements in this
Section apply to unfiltered systems that the Director has determined in writing
pursuant to Section 5 must install filtration, until June 29, 1993 or until filtration
is installed, whichever is later.
b)
Maximum Contaminant Level for Turbidity
The maximum contaminant level for turbidity shall not exceed a monthly average
of 1 turbidity unit (TU). A turbidity monthly average of two (2) turbidity units
may be acceptable provided it is demonstrated the higher turbidities did not
interfere with disinfection, and a residual disinfection was maintained throughout
179
the distribution system and did not interfere with microbiological determinations.
An average of five (5) turbidity units shall not be exceeded for any 2 consecutive
days.
c)
Analytical Techniques
Turbidity measurements shall be made in accordance with Appendix 1.
d)
A PWS that uses surface water or ground water under the direct influence of
surface water, as defined in Section 1 and does not practice filtration in
compliance with Section 5.4, must collect at least one (1) sample near the first
service connection each day the turbidity level of the source water measured as
specified in Section 5, exceeds 1 NTU. This sample must be analyzed for the
presence of total coliforms. When one (1) or more turbidity measurements in any
day exceed 1 NTU, the system must collect this coliform sample within 24 hours
of the first excedence, unless the Director determines that the system, for
logistical reasons outside of the system's control cannot have the sample analyzed
within 30 hours of collection. Sample results from this coliform monitoring must
be included in determining compliance with the MCL for total coliforms in
Section 16.4 c).
16.4
Microbiological
a)
Routine Monitoring
Community water systems must collect total coliform samples at sites which are
representative of water throughout the distribution system according to a written
sample siting plan. These plans are subject to the Director's review and revision.
The monitoring frequency for total coliforms for community water systems is
based on the population served by the system as follows:
Total Coliform Monitoring Frequency For Community Water Systems
Population Served
Minimum Number of Samples per
Month
25 to 1,0001
1
1,001 to 2,500
2
2,501 to 3,300
3
3,301 to 4,100
4
4,101 to 4,900
5
4,901 to 5,800
6
5,801 to 6,700
7
6,701 to 7,600
8
7,601 to 8,500
9
8,501 to 12,900
10
12,901 to 17,200
15
17,201 to 21,500
20
180
Total Coliform Monitoring Frequency For Community Water Systems
Population Served
Minimum Number of Samples per
Month
21,501 to 25,000
25
25,001 to 33,000
30
33,001 to 41,000
40
41,001 to 50,000
50
50,001 to 59,000
60
59,001 to 70,000
70
70,001 to 83,000
80
83,001 to 96,000
90
96,001 to 130,000
100
130,001 to 220,000
120
220,001 to 320,000
150
320,001 to 450,000
180
450,001 to 600,000
210
600,001 to 780,000
240
780,001 to 970,000
270
970,001 to 1,230,000
300
1,230,001 to 1,520,000
330
1,520,001 to 1,850,000
360
1,850,001 to 2,270,000
390
2,270,001 to 3,020,000
420
3,020,001 to 3,960,000
450
3,960,001 or more
480
1 Includes PWSs which have at least 15 service connections, but serve fewer than 25 persons.
1)
The Director may reduce the monitoring frequency of a community water
system serving 25-1000 persons in a written directive to not less than one
(1) sample per quarter if:
i)
A sanitary survey conducted in the past five (5) years shows that
the system is supplied solely by a protected ground water source
and is free of sanitary defects, and
ii)
Said water system has no history of total coliform contamination in
its current configuration.
2)
The PWS must collect samples at regular time intervals throughout the
month, EXCEPT: a system which uses groundwater not under the direct
influence of surface water as determined by the Director, and serves 4,900
persons or less, may collect all required samples on a single day, if they
are taken from different sites.
3)
A PWS that uses surface water or ground water under the direct influence
of surface water, as determined by the Director, and does not practice
filtration in compliance with Section 5 must:
181
i)
Collect at least one (1) sample near the first service connection
each day the turbidity level exceeds 1 NTU. This sample must be
analyzed for the presence of total coliforms.
ii)
When one (1) or more turbidity measurements exceed 1 NTU, the
system must collect the coliform sample within 24 hours of the
first exceedence unless the Director determines that the system for
logistical reasons outside the system's control cannot have the
sample analyzed within 30 hours of collection. Sample results must
be included in determining compliance with the MCL for total
coliforms, as indicated in Section 16.4 (c).
b)
Analytical Methodology
Coliform organism examinations shall be made in accordance with Appendix 1.
1)
The standard sample volume required for total coliform analysis,
regardless of analytical method used, is 100 ml.
2)
A PWS need only determine the presence or absence of total coliforms: a
determination of total coliform density is not required.
3)
If any routine or repeat sample is total coliform positive, the system must
analyze the culture medium to determine if fecal coliforms are present.
The system may test for E. Coli in lieu of fecal coliforms.
The Director has the discretion to allow a PWS, on a case-by-case basis, to
forgo fecal coliform or E. Coli testing on a total coliform positive sample
if that system assumes that the total coliform-positive sample is fecal
coliform-positive or E. Coli-positive. Accordingly, the system must notify
the Director as specified in Paragraph (e) of this Section and the
provisions of 16.4 c) 6) c) apply.
c)
Maximum Contaminant Levels (MCLs) for Microbiological Contaminants
1)
The goal for total coliforms (including fecal coliforms and Escherichia
Coli) is ZERO.
2)
The following constitute the best technology treatment techniques, or
other means available for achieving compliance with the MCL:
A)
Protection of wells from contamination by coliforms by
appropriate placement and construction;
B)
Maintenance of a disinfectant residual throughout the distribution
system;
182
C)
Proper maintenance of the distribution system;
D)
Filtration and/or disinfection of surface water as described in
Section 5; and
E)
The development and implementation of an approved wellhead
protection program, or watershed protection plan, if applicable.
3)
The MCL is based on the presence or absence of total coliforms in a
sample, rather than coliform density.
4)
A PWS must determine compliance with the MCL for total coliforms for
each month in which it is required to monitor for total coliforms.
5)
Special purpose samples such as those taken to determine disinfection
practices, shall not be used to determine compliance with the MCL for
total coliforms. Repeat samples must be used in determination of the
monthly MCL compliance.
6)
The following constitutes a violation of the MCL for total coliforms:
A)
If a system collects at least 40 samples per month, the system is not
in compliance with the MCL for total coliforms if more than 5.0
percent of the samples collected during the month are total
coliform positive.
B)
If a system collects less than 40 samples per month, the system is
not in compliance with the MCL for total coliforms if more than
one (1) sample is total coliform positive.
C)
*If any repeat sample is fecal coliform or E. Coli positive.
D)
* If any repeat sample is total coliform positive following a fecal
coliform or E. Coli positive routine sample.
* For purposes of public notification requirements discussed in Section
16.8, this is a violation that may pose an acute health risk.
7)
The Director must be notified of any total coliform MCL violation by the
end of the day on which the system learns of the violation or by the end of
the next business day if state offices are closed.
d)
Repeat monitoring
183
If a routine sample is total coliform-positive, the PWS must collect a set of repeat
samples for each total coliform positive within 24 hours of being notified of the
positive result as follows:
1)
> 1 routine sample/month: 3 repeat samples
(Total volume collected must be at least 300 ml)
2)
1 or < 1 routine sample/month: 4 repeat samples
(Total volume collected must be at least 400 ml)
3)
The Director may extend the 24-hour limit on a case-by-case basis if the
system has a logistical problem in collecting the repeat samples within 24
hours that is beyond its control. In the case of an extension, the Director
will specify how much time the system has to collect the repeat samples.
4)
At least one (1) repeat sample must be collected from the sampling tap
where the original positive sample was taken; at least one (1) repeat
sample within 5 service connections upstream; and one (1) repeat sample
within 5 service connections downstream. If a total coliform-positive
sample is at the end of the distribution system, or one (1) away from the
end of the distribution system, the Director may waive the requirement to
collect at least one (1) repeat sample upstream or downstream of the
original sampling site.
5)
These repeat samples must be collected on the same day, except that the
Director may allow a system with a single service connection to collect the
required set of repeat samples over a four-day period or to collect a larger
volume repeat sample(s) in one (1) or more sample containers of any size,
as long as the total volume collected is at least 400 ml(300 ml for systems
which collect more than one (1) routine sample/month).
6)
If one (1) or more repeat sample in the set is total coliform-positive, the
PWS must collect an additional set of repeat samples as described in this
Section. The system must repeat this process until either total coliforms
are not detected in one (1) complete set of repeat samples, or the system
determines that the MCL for total coliforms has been exceeded and
notifies the Director.
7)
Results of all routine and repeat samples not invalidated by the Director
must be included in determining compliance with the MCL for total
coliforms in Section 16.4 c).
8)
Any system collecting fewer than five (5) routine samples per month must
collect at least five (5) samples during the month following repeat
184
sampling for total coliform positive samples. The Director can waive this
requirement on a case by case basis.
A)
The Director may waive the requirement to collect five (5) routine
samples the next month the system provides water to the public if
the Director performs a site visit before the end of the next month
the system provides water to the public. Although a sanitary survey
need not be performed, the site visit must be sufficiently detailed to
allow the Director to determine whether additional monitoring
and/or any corrective action is needed.
B)
The Director may waive the requirement to collect five (5) routine
samples the next month the system provides water to the public if
the Director has determined in writing why the sample was total
coliform positive and establishes that the system has corrected the
problem or will correct the problem before the end of the next
month the system serves water to the public. The written
documentation must describe the specific cause of the total
coliform-positive sample and what action the system has taken
and/or will take to correct this problem. The Director will not
waive the requirement to collect five (5) routine samples the next
month the system provides water to the public solely on the
grounds that all repeat samples are total coliform-negative. Under
this Paragraph, a system must still take at least one (1) routine
sample before the end of the next month it serves water to the
public and use it to determine compliance with the MCL for total
coliforms in Section 16.4.
e)
Fecal Coliforms/E. Coli Testing
If any routine or repeat sample is total coliform positive, the system must analyze
the culture medium to determine if fecal coliforms are present. The system may
test for E. Coli in lieu of fecal coliforms. If either are present, the system must
notify the Director by the end of that day or the next business day if state offices
are closed.
f)
Invalidation of Samples
1)
A total coliform sample invalidated under this Paragraph does not count
towards meeting the minimum monitoring requirements of this Section.
2)
The Director will invalidate a total coliform-positive sample and
document same in writing only if:
i)
The laboratory establishes that improper sample analysis caused
the total coliform-positive result.
185
ii)
The Director determines that the total coliform-positive sample
resulted from a domestic or other non-distribution system
plumbing problem.
iii)
The Director has substantial grounds to believe that a total
coliform-positive result is due to a circumstance or condition
which does not reflect water quality in the distribution system. (In
this case, the system must still collect all repeat samples required.)
3)
A total coliform-positive sample will not be invalidated solely on the
grounds that all repeat samples are total coliform negative.
4)
A laboratory must invalidate a total coliform sample, unless total
coliforms are detected, only if:
i)
The sample produces a turbid culture in the absence of gas
production using the method cited in Section 16.4 (b) (4) (a);
ii)
The sample produces a turbid culture in the absence of an acid
reaction; using the method cited in Section 16.4 (b) (4) (c).
iii)
It exhibits confluent growth , or produces colonies too numerous to
count, using the method cited in Section 16.4 (b) (4) (b).
5)
If a laboratory invalidates a sample for the above reasons, the system must
collect another sample from the same location as the original sample
within 24 hours of being notified of the result. The system must continue
to re-sample within 24 hours and have the samples analyzed until it
obtains a valid result. The Director may extend the 24-hour limit on a
case-by-case basis if the system has a logistical problem in collecting the
repeat samples within 24 hours that is beyond its control. In the case of a
extension, the Director will specify how much time the system has to
collect the repeat samples.
g) Sanitary Surveys:
1)
A PWS which does not collect five (5) or more routine samples/month
must undergo an initial sanitary survey by June 29, 1994 for community
PWSs and June 29, 1999 for non-community water systems. Thereafter,
systems must undergo another sanitary survey every five (5) years, except
that non-community water systems using only protected and disinfected
ground water as defined by the Director must undergo subsequent sanitary
surveys at least every ten (10) years after the initial sanitary survey.
186
2)
A PWS is responsible for making all necessary facilities, personnel and
records available so that a sanitary survey may be completed.
3)
Deficiencies listed in a sanitary survey are considered to be unsafe
conditions and must be addressed as provided for in Section 10 of these
regulations.
h)
Reporting Requirements:
1)
The supplier of water must report to the Director any failure to comply
with any drinking water regulation within 48 hours, except where different
reporting is specified in these regulations.
2)
A PWS which has exceeded the MCL for total coliforms must report the
violation to the Director no later than the end of the next business day, and
notify the public in accordance with Section 16.8.
3)
A PWS which has failed to comply with a coliform monitoring
requirement, including the sanitary survey must report the monitoring
violation to the Director within ten (10) days after the system discovers the
violation, and notify the public in accordance with Section 16.8.
16.5 Radioactivity
(a)
Monitoring and Compliance Requirements for Gross Alpha Particle Activity,
Radium-226, Radium-228 and Uranium.
(1)
Community water systems must conduct initial monitoring to determine
compliance with 16.5 (b) and (c) by December 31, 2007. For the purposes
of monitoring for gross alpha particle activity, radium-226, radium-228,
and uranium in drinking water, “detection limit” is defined as in Appendix
1 Section II D (3).
(i)
Applicability and sampling location for existing community water
systems or sources - All existing community water systems using
ground water, surface water or systems using both ground and
surface water (for the purpose of this Section hereafter referred to
as systems) must sample at every entry point to the distribution
system that is representative of all sources being used (hereafter
called a sampling point) under normal operating conditions. The
system must take each sample at the same sampling point unless
conditions make another sampling point more representative of
each source or the Director has designated a distribution system
location, in accordance with 16.5 (a)(2)(ii)(C).
187
(ii)
Applicability and sampling location for new community water
systems or sources - All new community water systems or
community water systems that use a new source of water must
begin to conduct initial monitoring for the new source within the
first quarter after initiating use of the source. Community water
systems must conduct more frequent monitoring when ordered by
the Director in the event of possible contamination or when
changes in the distribution system or treatment processes occur
which may increase the concentration of radioactivity in finished
water.
(2)
Initial monitoring
Systems must conduct initial monitoring for gross alpha particle activity,
radium-226, radium-228, and uranium as follows:
(i)
Systems without acceptable historical data, as defined in 16.5
(a)(2)(ii), must collect four (4) consecutive quarterly samples at all
sampling points before December 31, 2007.
(ii)
Grandfathering of data: The Director may allow historical
monitoring data collected at a sampling point to satisfy the initial
monitoring requirements for that sampling point, for the following
situations:
To satisfy initial monitoring requirements, a community water
system having only one (1) entry point to the distribution system
may use the monitoring data from the last compliance monitoring
period that began between June 2000 and December 8, 2003.
To satisfy initial monitoring requirements, a community water
system with multiple entry points and having appropriate historical
monitoring data for each entry point to the distribution system may
use the monitoring data from the last compliance monitoring
period that began between June 2000 and December 8, 2003.
To satisfy initial monitoring requirements, a community water
system with appropriate historical data for a representative point in
the distribution system may use the monitoring data from the last
compliance monitoring period that began between June 2000 and
December 8, 2003, provided that the Director finds that the
historical data satisfactorily demonstrate that each entry point to
the distribution system is expected to be in compliance based upon
the historical data and reasonable assumptions about the variability
of contaminant levels between entry points. The Director must
188
make a written finding indicating how the data conforms to these
requirements.
(iii)
For gross alpha particle activity, uranium, radium-226, and
radium-228 monitoring, the Director may waive the final two (2)
quarters of initial monitoring for a sampling point if the results of
the samples from the previous two (2) quarters are below the
detection limit.
(iv)
If the average of the initial monitoring results for a sampling point
is above the MCL, the system must collect and analyze quarterly
samples at that sampling point until the system has results from
four (4) consecutive quarters that are at or below the MCL, unless
the system enters into another schedule as part of a formal
compliance agreement with the Director.
(3)
Reduced monitoring
Beginning January 1, 2008, the Director may allow community water
systems to reduce the future frequency of monitoring from once every
three (3) years to once every six (6) or nine (9) years at each sampling
point, based on the following criteria.
(i)
If the average of the initial monitoring results for each contaminant
(i.e., gross alpha particle activity, uranium, radium-226, or radium-
228) is below the detection limit specified in Table B in Appendix
1, the system must collect and analyze for that contaminant using
at least one (1) sample at that sampling point every nine (9) years.
(ii)
For gross alpha particle activity and uranium, if the average of the
initial monitoring results for each contaminant is at or above the
detection limit but at or below 1/2 the MCL, the system must
collect and analyze for that contaminant using at least one (1)
sample at that sampling point every six (6) years. For combined
radium-226 and radium-228, the analytical results must be
combined. If the average of the combined initial monitoring results
for radium-226 and radium-228 is at or above the detection limit
but at or below 1/2 the MCL, the system must collect and analyze
for that contaminant using at least one (1) sample at that sampling
point every six (6) years.
(iii)
For gross alpha particle activity and uranium, if the average of the
initial monitoring results for each contaminant is above 1/2 the
MCL but at or below the MCL, the system must collect and
analyze at least one (1) sample at that sampling point every three
(3) years. For combined radium-226 and radium-228, the analytical
189
results must be combined. If the average of the combined initial
monitoring results for radium-226 and radium-228 is above 1/2 the
MCL but at or below the MCL, the system must collect and
analyze at least one (1) sample at that sampling point every three
(3) years.
(iv)
Systems must use the samples collected during the reduced
monitoring period to determine the monitoring frequency for
subsequent monitoring periods (e.g., if a system's sampling point is
on a nine (9) year monitoring period, and the sample result is
above 1/2 MCL, then the next monitoring period for that sampling
point is three (3) years).
(v)
If a system has a monitoring result that exceeds the MCL while on
reduced monitoring, the system must collect and analyze quarterly
samples at that sampling point until the system has results from
four (4) consecutive quarters that are below the MCL, unless the
system enters into another schedule as part of a formal compliance
agreement with the Director.
(4)
Compositing
To fulfill quarterly monitoring requirements for gross alpha particle
activity, radium-226, radium-228, or uranium, a system may composite up
to four (4) consecutive quarterly samples from a single entry point if
analysis is done within a year of the first sample. The Director will treat
analytical results from the composited sample as the average analytical
result to determine compliance with the MCLs and the future monitoring
frequency. If the analytical result from the composited sample is greater
than 1/2 MCL, the Director may direct the system to take additional
quarterly samples before allowing the system to sample under a reduced
monitoring schedule.
(5)
A gross alpha particle activity measurement may be substituted for the
required radium-226 measurement provided that the measured gross alpha
particle activity does not exceed 5 pCi/l. A gross alpha particle activity
measurement may be substituted for the required uranium measurement
provided that the measured gross alpha particle activity does not exceed
15 pCi/l. The gross alpha measurement shall have a confidence interval of
95% (1.65σ, where σ is the standard deviation of the net counting rate of
the sample) for radium-226 and uranium. When a system uses a gross
alpha particle activity measurement in lieu of a radium-226 and/or
uranium measurement, the gross alpha particle activity analytical result
will be used to determine the future monitoring frequency for radium-226
and/or uranium. If the gross alpha particle activity result is less than
190
detection, 1/2 the detection limit will be used to determine compliance and
the future monitoring frequency.
(b)
Maximum Contaminant Level for Gross Alpha Particle Activity and Radium-226
and Radium-228
Contaminant
Picocuries per Liter(pCi/l)
Radium-226 and Radium-228 Combined1
5
Gross alpha particle activity2
15
1 The combined radium-226 and radium-228 value is determined by the addition of the results of the analysis for
radium-226 and the analysis for radium-228.
2 Including radium-226 but excluding radon and uranium
(c)
Maximum Contaminant Level for Uranium
The maximum contaminant level for uranium is 30 µg/L.
(d)
Maximum Contaminant Level for Manmade Beta Particle and Photon Emitters
The average annual concentration of manmade beta particle and photon emitters
shall not meet or exceed an annual dose equivalent of 4 millirems/year.
(1)
Except for the radionuclides listed in Table A, the concentration of man-
made radionuclides causing 4 mrem total body or organ dose equivalents
must be calculated on the basis of 2 liter per day drinking water intake
using the 168 hour data list in “Maximum Permissible Body Burdens and
Maximum Permissible Concentrations of Radionuclides in Air and in
Water for Occupational Exposure,” NBS (National Bureau of Standards)
Handbook 69 as amended August 1963, U.S. Department of Commerce.
This incorporation by reference was approved by the Director of the
Federal Register in accordance with 5 U.S.C. 552(a) and 1 CFR part 51.
Copies of this document are available from the National Technical
Information Service, NTIS ADA 280 282, U.S. Department of Commerce,
5285 Port Royal Road, Springfield, Virginia 22161. The toll-free number
is 800-553-6847. Copies may be inspected at EPA's Drinking Water
Docket, 401 M Street, SW., Washington, DC 20460; or at the Office of the
Federal Register, 800 North Capitol Street, NW., Suite 700, Washington,
DC.
(2)
If two (2) or more radionuclides are present, the sum of their annual dose
equivalent to the total body or to any organ shall not exceed 4 mrem/year.
191
TABLE A. Average Annual Concentrations Assumed to Produce a Total Body or Organ Dose of 4 millirem/year
Radionuclide
Critical Organ
pCi/l
tritium
total body
20,000
Strontium-90
bone marrow
8
(e)
Monitoring and compliance requirements for beta particle and photon
radioactivity.
For the purposes of monitoring for beta particle and photon radioactivity in
drinking water, “detection limit” is defined as in Appendix 1 Section II D (3). To
determine compliance with the maximum contaminant levels in 16.5 (d) for beta
particle and photon radioactivity, a system must monitor at a frequency as
follows:
Community water systems (both surface and ground water) designated by the
Director as vulnerable must sample for beta particle and photon radioactivity.
Systems must collect quarterly samples for beta emitters and annual samples for
tritium and strontium-90 at each entry point to the distribution system (hereafter
called a sampling point), beginning within one (1) quarter after being notified by
the Director. Systems already designated by the Director must continue to sample
until the Director reviews and either reaffirms or removes the designation.
(i)
If the gross beta particle activity minus the naturally occurring potassium-
40 beta particle activity at a sampling point has a running annual average
(computed quarterly) less than or equal to 50 pCi/L (screening level), the
Director may reduce the frequency of monitoring at that sampling point to
once every 3 years. Systems must collect all samples required in 16.5
(e)(1) during the reduced monitoring period.
(ii)
For systems in the vicinity of a nuclear facility, the Director may allow the
community water system to utilize environmental surveillance data
collected by the nuclear facility in lieu of monitoring at the system's entry
point(s), where the Director determines if such data is applicable to a
particular water system. In the event that there is a release from a nuclear
facility, systems which are using surveillance data must begin monitoring
at the community water system's entry point(s) in accordance with 16.5
(e)(1).
Community water systems (both surface and ground water) designated by
the Director as utilizing waters contaminated by effluents from nuclear
facilities must sample for beta particle and photon radioactivity. Systems
must collect quarterly samples for beta emitters and iodine-131 and annual
samples for tritium and strontium-90 at each entry point to the distribution
system (hereafter called a sampling point), beginning within one (1)
quarter after being notified by the Director. Systems already designated by
192
the Director as systems using waters contaminated by effluents from
nuclear facilities must continue to sample until the Director reviews and
either reaffirms or removes the designation.
Quarterly monitoring for gross beta particle activity shall be based on the
analysis of monthly samples or the analysis of a composite of three (3)
monthly samples. The former is recommended.
For iodine-131, a composite of five (5) consecutive daily samples shall be
analyzed once each quarter. As ordered by the Director, more frequent
monitoring shall be conducted when iodine-131 is identified in the
finished water.
Annual monitoring for strontium-90 and tritium shall be conducted by
means of the analysis of a composite of four (4) consecutive quarterly
samples or analysis of four (4) quarterly samples. The latter procedure is
recommended.
If the gross beta particle activity minus the naturally occurring potassium-
40 beta particle activity at a sampling point has a running annual average
(computed quarterly) less than or equal to 15 pCi/L (screening level), the
Director may reduce the frequency of monitoring at that sampling point to
every 3 years. Systems must collect all samples required in 16.5 (e)(2)
during the reduced monitoring period.
For systems in the vicinity of a nuclear facility, the Director may allow the
community water system to utilize environmental surveillance data
collected by the nuclear facility in lieu of monitoring at the system's entry
point(s), where the Director determines if such data is applicable to a
particular water system. In the event that there is a release from a nuclear
facility, systems which are using surveillance data must begin monitoring
at the community water system's entry point(s) in accordance with 16.5
(e)(2).
Community water systems designated by the Director to monitor for beta
particle and photon radioactivity can not apply to the Director for a waiver
from the monitoring frequencies specified in 16.5 (e)(1) or (e)(2).
Community water systems may analyze for naturally occurring potassium-
40 beta particle activity from the same or equivalent sample used for the
gross beta particle activity analysis. Systems are allowed to subtract the
potassium-40 beta particle activity value from the total gross beta particle
activity value to determine if the screening level is exceeded. The
potassium-40 beta particle activity must be calculated by multiplying
elemental potassium concentrations (in mg/L) by a factor of 0.82.
193
If the gross beta particle activity minus the naturally occurring potassium-
40 beta particle activity exceeds the appropriate screening level, an
analysis of the sample must be performed to identify the major radioactive
constituents present in the sample and the appropriate doses must be
calculated and summed to determine compliance with 16.5 (d), using the
formula in 16.5 (d)(1). Doses must also be calculated and combined for
measured levels of tritium and strontium to determine compliance.
Systems must monitor monthly at the sampling point(s) which exceed the
maximum contaminant level in 16.5 (d) beginning the month after the
exceedance occurs. Systems must continue monthly monitoring until the
system has established, by a rolling average of 3 monthly samples, that the
MCL is being met. Systems who establish that the MCL is being met must
return to quarterly monitoring until they meet the requirements set forth in
16.5 (e)(1)(i) or (e)(2)(iv).
(f)
General monitoring and compliance requirements for radionuclides.
(1)
The Director may require more frequent monitoring than specified in 16.5
(a) and (e), or may require confirmation samples at its discretion. The
results of the initial and confirmation samples will be averaged for use in
compliance determinations.
(2)
Each PWS shall monitor at the time designated by the Director during
each compliance period.
(3)
Compliance: Compliance with 16.5 (b) through (d) will be determined
based on the analytical result(s) obtained at each sampling point. If one (1)
sampling point is in violation of an MCL, the system is in violation of the
MCL. To judge compliance with the maximum contaminant levels listed
in 16.5 (b), (c) and (d), averages of data shall be used and shall be rounded
to the same number of significant figures as the maximum contaminant
level for the substance in question.
(i)
For systems monitoring more than once per year, compliance with
the MCL is determined by a running annual average at each
sampling point. If the average of any sampling point is greater than
the MCL, then the system is out of compliance with the MCL.
(ii)
For systems monitoring more than once per year, if any sample
result will cause the running average to exceed the MCL at any
sample point, the system is out of compliance with the MCL
immediately.
194
(iii)
Systems must include all samples taken and analyzed under the
provisions of this Section in determining compliance, even if that
number is greater than the minimum required.
(iv)
If a system does not collect all required samples when compliance
is based on a running annual average of quarterly samples,
compliance will be based on the running average of the samples
collected.
(v)
If a sample result is less than the detection limit, zero will be used
to calculate the annual average, unless a gross alpha particle
activity is being used in lieu of radium-226 and/or uranium. If the
gross alpha particle activity result is less than detection, 1/2 the
detection limit will be used to calculate the annual average.
(vi)
To judge compliance with the maximum contaminant levels listed
in 16.5 (b), (c), and (d), averages of data shall be used and shall be
rounded to the same number of significant figures as the maximum
contaminant level for the substance in question.
(4)
The Director has the discretion to delete results of obvious sampling or
analytic errors.
(5)
If the MCL for radioactivity set forth in 16.5 (b), (c) or (d) is exceeded, the
operator of a community water system must give notice to the Director
pursuant to Section 11.0 and to the public as required by Section 16.8.
(6)
Where monitoring results exceed the MCLs specified in Paragraphs
16.5(b), (c), or (d) of this Section, compliance shall be achieved by
installation of central treatment which is approved by the Director. Point
of use or point of entry devices may be used only as a condition of a
variance from this Paragraph and only if a plan for their maintenance and
operation is approved by the Director and every building connected to the
water system has a device installed and adequately monitored by the PWS.
Requirements for approval of use of non-centralized treatment are
contained in Sections 4.4 and 4.5 of these rules.
(g)
Compliance dates
Compliance dates for combined radium-226 and -228, gross alpha particle
activity, gross beta particle and photon radioactivity, and uranium: Community
water systems must comply with the MCLs listed in Section 16.5 (b), (c) and (d)
beginning December 8, 2003 and compliance shall be determined in accordance
with the requirements of Section 16.5 and Appendix 1. Compliance with reporting
requirements for the radionuclides under Section 11.0 and 16.8 is required on
December 8, 2003.
195
(h)
Best Available Technologies (BATs) for Radionuclides.
The Administrator, pursuant to Section 1412 of the Act, hereby identifies as
indicated in the following table the best technology available for achieving
compliance with the maximum contaminant levels for combined radium-226 and -
228, uranium, gross alpha particle activity and beta particle and photon
radioactivity.
196
Table B. BAT for Combined Radium-226 and Radium-228, Uranium, Gross Alpha Particle Activity, and Beta
Particle and Photon Radioactivity
Contaminant
BAT
1. Combined radium-226 and radium-228
2. Uranium
3. Gross alpha particle activity (excluding Radon and
Uranium)
4. Beta particle and photon radioactivity
Ion exchange, reverse osmosis, lime softening.
Ion exchange, reverse osmosis, lime softening,
coagulation/filtration.
Reverse osmosis.
Ion exchange, reverse osmosis.
(i)
Small systems compliance technologies list for radionuclides
Table C. List of Small Systems Compliance Technologies for Radionuclides and Limitations to Use.
Unit Technologies
Limitations
(see
footnotes)
Operator Skill Level
Required1
Raw Water Quality Range and
Considerations1
1. Ion exchange (IE)
(a)
Intermediate
All ground waters.
2. Point of use (POU2) IE
(b)
Basic
All ground waters.
3. Reverse osmosis (RO)
(c)
Advanced
Surface waters usually require pre-
filtration.
4. POU2 RO
(b)
Basic
Surface waters usually require pre-
filtration.
5. Lime softening
(d)
Advanced
All waters.
6. Green sand filtration
(e)
Basic
7. Co-precipitation with
Barium sulfate
(f)
Intermediate to Advanced
Ground waters with suitable water
quality.
8.
Electrodialysis/electrodialysis
reversal
Basic to Intermediate
All ground waters.
9. Pre-formed hydrous
Manganese oxide filtration.
(g)
Intermediate
All ground waters.
10. Activated alumina
(a), (h)
Advanced
All ground waters; competing anion
concentrations may affect
regeneration frequency.
11. Enhanced
coagulation/filtration
(i)
Advanced
Can treat a wide range of water
qualities.
1 National Research Council (NRC). Safe Water from Every Tap: Improving Water Service to Small Communities.
National Academy Press. Washington, D.C. 1997.
2 A POU, or ‘‘point-of-use’’ technology is a treatment device installed at a single tap used for the purpose of
reducing contaminants in drinking water at that one (1) tap. POU devices are typically installed at the kitchen tap.
See the April 21, 2000 NODA for more details.
Limitations Footnotes:
Technologies for Radionuclides:
a The regeneration solution contains high concentrations of the contaminant ions. Disposal options should be
carefully considered before choosing this technology.
197
b When POU devices are used for compliance, programs for long-term operation, maintenance and monitoring must
be provided by water utility to ensure proper performance.
c Reject water disposal options should be carefully considered before choosing this technology. See other RO
limitations described in the SWTR Compliance Technologies Table.
d The combination of variable source water quality and the complexity of the water chemistry involved may make
this technology too complex for small surface water systems.
e Removal efficiencies can vary depending on water quality.
f This technology may be very limited in application to small systems. Since the process requires static mixing,
detention basins, and filtration, it is most applicable to systems with sufficiently high sulfate levels that already have
a suitable filtration treatment train in place.
g This technology is most applicable to small systems that already have filtration in place.
h Handling of chemicals required during regeneration and pH adjustment may be too difficult for small systems
without an adequately trained operator.
i Assumes modification to a coagulation/filtration process already in place.
Table D. Compliance Technologies by System Size Category for Radionuclide NPDWR’s
Compliance Technologies1 for System Size Categories
(population served)
Contaminant
25-500
501-3,300
3,300-10,000
1. Combined radium-226 and radium-228
2. Gross alpha particle activity
3. Beta particle activity and photon activity
4. Uranium
1, 2, 3, 4, 5, 6, 7, 8,
9.
3, 4.
1, 2, 3, 4.
1, 2, 4, 10, 11.
1, 2, 3, 4, 5, 6, 7, 8, 9.
3, 4.
1, 2, 3, 4.
1, 2, 3, 4, 5, 10, 11.
1, 2, 3, 4, 5, 6, 7, 8,
9.
3, 4.
1, 2, 3, 4.
1, 2, 3, 4, 5, 10, 11.
Note: 1 Numbers correspond to those technologies found listed in Table C of 16.5 (i) above.
16.6
Unregulated Contaminants: Community water systems shall monitor for the following
volatile organic contaminants:
(1)
Chloroform
(2)
Bromodichloromethane
(3)
Chlorodibromomethane
(4)
Bromoform
(5)
Dibromomethane
(6)
m-Dichlorobenzene
(7)
1,1-Dichloropropene
(8)
1,1-Dichloroethane
(9)
1,1,2,2-Tetrachloroethane
(10)
1,3-Dichloropropane
(11)
Chloromethane
(12)
Bromomethane
198
(13)
1,2,3-Trichloropropane
(14)
1,1,1,2-Tetrachloroethane
(15)
Chloroethane
(16)
2,2-Dichloropropane
(17)
o-Chlorotoluene
(18)
p-Chlorotoluene
(19)
Bromobenzene
(20)
1,3-Dichloropropene
(a)
Monitoring Frequency - Effective January 8, 1999 only community systems
serving more than 10,000 persons must comply with Section 16.6 herein. Each
active drinking water source maintained by a water purveyor shall be analyzed for
the unregulated contaminants listed in this Section. Systems must sample at each
entry point to the distribution system and after any treatment provided to the
sources of water. Each entry point must be sampled for four (4) consecutive
quarters at least once every five (5) years for each surface water source and at
least one (1) sample per entry point to the distribution system every five (5) years
for each groundwater source beginning no later than January 1, 1989 for water
systems serving 3,300 or more persons, and no later than January 1, 1991 for
water systems serving less than 3300 persons.
(b)
Notification - Upon completion of the sampling required under this Section, the
water purveyor shall notify persons served by the system of the availability of the
analytical results and shall identify a person and telephone number to contact for
information regarding these results. The notification shall be performed by either
a notice in the first set of water bills issued by the system after the receipt of the
results or written notice within three (3) months.
(c)
Analytical Techniques - Analyses shall be conducted in accordance with
Appendix 1 and only by a laboratory certified by EPA or the Department of
Health to perform volatile organic chemical analyses by the appropriate method.
16.7
Special monitoring for inorganic and organic contaminants*
* Note: Monitoring requirements of Section 16.7 were completed as of December 31,
1995.
(a)
Monitoring of the contaminants listed in 16.7 (a) (11) and (12) shall be conducted
as follows:
(1)
Each community and non-transient, non-community water system shall
take four (4) consecutive quarterly samples at each sampling point for
each contaminant listed in Paragraph (a) (11) of this Section and report the
results to the Director. Monitoring must be completed by December 31,
1995.
199
(2)
Each community and non-transient, non-community water system shall
take one (1) sample at each sampling point for each contaminant listed in
Paragraph (a) (12) of this Section and report the results to the Director.
Monitoring must be completed by December 31, 1995.
(3)
Each community and non-transient non-community water system may
apply to the Director for a waiver from the requirements of Paragraph (a)
(1) and (2) of this Section.
(4)
The Director may grant a waiver for the requirement of Paragraph (a) (1)
of this Section based on the criteria specified in 16.2(a)(6). The Director
may grant a waiver from the requirement of Paragraph (a) (2) of this
Section if previous analytical results indicate contamination would not
occur, provided this data was collected after January 1, 1990.
(5)
Groundwater systems shall take a minimum of one (1) sample at every
entry point to the distribution system which is representative of each well
after treatment (hereafter called a sampling point). Each sample must be
taken at the same sampling point unless conditions make another sampling
point more representative of each source or treatment plant.
(6)
Surface water systems shall take a minimum of one (1) sample at points in
the distribution system that are representative of each source or at each
entry point to the distribution system after treatment (hereafter called a
sampling point). Each sample must be taken at the same sampling point
unless conditions make another sampling point more representative of
each source or treatment plant.
Note: For purposes of this Paragraph, surface water systems include
systems with a combination of surface and ground sources.
(7)
If the system draws water from more than one (1) source and the sources
are combined before distribution, the system must sample at an entry point
to the distribution system during periods of normal operating conditions
(i.e., when water representative of all sources is being used).
(8)
The Director may require a confirmation sample for positive or negative
results.
(9)
The Director may reduce the total number of samples a system must
analyze by allowing the use of compositing. Composite samples from a
maximum of five (5) sampling points are allowed. Compositing of
samples must be done in the laboratory and the composite sample must be
analyzed within 14 days of collection. If the population served by the
system is > 3,300 persons, then compositing may only be permitted by the
Director at sampling points within a single system. In systems serving less
200
than or equal to 3,300 persons, the Director may permit compositing
among different systems provided the 5-sample limit is maintained.
(10)
Instead of performing the monitoring required by this Section, a
community water system or non-transient non-community water system
serving fewer than 150 service connections may send a letter to the
Director stating that the system is available for sampling. This letter must
be sent to the Director by January 1, 1994. The system shall not send such
samples to the Director, unless requested to do so by the Director.
(11)
List of Unregulated Organic Contaminants:
Aldrin
Butachlor
Carbaryl
Dicamba
Dieldrin
3-Hydroxycarbofuran
Methomyl
Metolachlor
Metribuzin
Propachlor
(12)
List of Unregulated Inorganic Contaminants:
Sulfate
16.8
Public Notification of Drinking Water Violations
Section (1) General public notification requirements.
Section (2) Tier 1 Public Notice–Form, manner and frequency of notice.
Section (3) Tier 2 Public Notice–Form, manner and frequency of notice.
Section (4) Tier 3 Public Notice–Form, manner and frequency of notice.
Section (5) Content of the public notice.
Section (6) Notice to new billing units or new customers.
Section (7) Special notice of the availability of unregulated contaminant monitoring results.
Section (8) Special notice for exceedance of the SMCL for fluoride.
Section (9) Special notice for nitrate exceedances above MCL by non-community water systems
(NCWS), where granted permission by the Director under 141.11(d)
Section (10) Notice by the Director on behalf of the PWS.
Section (11) Reporting.
Appendix A to Section 16.8 –NPDWR Violations and Other Situations Requiring Public Notice
Appendix B to Section 16.8 –Standard Health Effects Language for Public Notification
Appendix C to Section 16.8 –List of Acronyms Used in Public Notification Regulation
(1)
General Public Notification Requirements
201
(a)
Who must give public notice? Each owner or operator of a PWS
(community water systems, non-transient, non-community water systems
and transient non-community water systems) must give notice for all
violations of National Primary Drinking Water Regulations (NPDWR) and
for other situations, as listed in Table 1. The term “NPDWR violations” is
used in this subpart to include violations of the maximum contaminant
level (MCL), maximum residual disinfection level (MRDL), treatment
technique (TT), monitoring requirements and testing procedures in these
regulations. Appendix A to this Section identifies the tier assignment for
each specific violation or situation requiring a public notice.
Table 1 to Section 16.8(1) –Violation Categories and Other Situations Requiring a Public Notice
(1)
NPDWR violations:
(i)
Failure to comply with an applicable maximum contaminant level (MCL) or maximum residual
disinfectant level (MRDL).
(ii)
Failure to comply with a prescribed treatment technique (TT).
(iii)
Failure to perform water quality monitoring, as required by the drinking water regulations.
(iv)
Failure to comply with testing procedures as prescribed by a drinking water regulation.
(2)
Variance and exemptions under sections 1415 and 1416 of SDWA:
(i)
Operation under a variance or an exemption.
(ii)
Failure to comply with the requirements of any schedule that has been set under a variance or
exemption.
(3)
Special public notices:
(i)
Occurrence of a waterborne disease outbreak or other waterborne emergency.
(ii)
Exceedance of the nitrate MCL by non-community water systems (NCWS), where granted
permission by the Director under Section 15.5 of this part.
(iii)
Exceedance of the secondary maximum contaminant level (SMCL) for fluoride.
(iv)
Availability of unregulated contaminant monitoring data.
(v)
Other violations and situations determined by the Director to require a public notice under this
subpart, not already listed in Appendix A.
(b)
What type of public notice is required for each violation or situation?
Public notice requirements are divided into three (3) tiers, to take into
account the seriousness of the violation or situation and of any potential
adverse health effects that may be involved. The public notice
requirements for each violation or situation listed in Table 1 of this
Section are determined by the tier to which it is assigned. Table 2 of this
Section provides the definition of each tier. Appendix A of this part
identifies the tier assignment for each specific violation or situation.
Table 2 to Section 16.8(1)–Definition of Public Notice Tiers
(1)
Tier 1 public notice–required for NPDWR violations and situations with significant potential to have
serious adverse effects on human health as a result of short-term exposure.
(2)
Tier 2 public notice–required for all other NPDWR violations and situations with potential to have serious
adverse effects on human health.
202
(3)
Tier 3 public notice–required for all other NPDWR violations and situations not included in Tier 1 and
Tier 2.
(c)
Who must be notified?
(1)
Each PWS must provide public notice to persons served by the
water system, in accordance with this subpart. PWSs that sell or
otherwise provide drinking water to other PWSs (i.e., to
consecutive systems) are required to give public notice to the
owner or operator of the consecutive system; the consecutive
system is responsible for providing public notice to the persons it
serves.
(2)
If a PWS has a violation in a portion of the distribution system that
is physically or hydraulically isolated from other parts of the
distribution system, the Director may allow the system to limit
distribution of the public notice to only persons served by that
portion of the system which is out of compliance. Permission by
the Director for limiting distribution of the notice must be granted
in writing.
(3)
A copy of the notice must also be sent to the Director, in
accordance with the requirements under Section 16.8 (11).
(2)
Tier 1 Public Notice–Form, Manner and Frequency of Notice
(a)
Which violations or situations require a Tier 1 public notice? Table 1 of
this Section lists the violation categories and other situations requiring a
Tier 1 public notice. Appendix A to this subpart identifies the tier
assignment for each specific violation or situation.
Table 1 to Section 16.8(2)–Violation Categories and Other Situations Requiring a Tier 1 Public Notice
(1)
Violation of the MCL for total coliforms when fecal coliform or E. Coli are present in the water
distribution system (as specified in Section 16.4(c)), or when the water system fails to test for fecal
coliforms or E. Coli when any repeat sample tests positive for coliform (as specified in 16.4(e));
(2)
Violation of the MCL for nitrate, nitrite, or total nitrate and nitrite, as defined in Section 16.1, or when the
water system fails to take a confirmation sample within 24 hours of the system's receipt of the first sample
showing an exceedance of the nitrate or nitrite MCL, as specified in Section 16.1(f)(2);
(3)
Exceedance of the nitrate MCL by non-community water systems, where permitted to exceed the MCL by
the Director under Section 16.8 (9), as required under Section 15.5;
(4)
Violation of the MRDL for chlorine dioxide, as defined in Section 7.2(a), when one (1) or more samples
taken in the distribution system the day following an exceedance of the MRDL at the entrance of the
distribution system exceed the MRDL, or when the water system does not take the required samples in the
distribution system, as specified in Section 7.6 (c)(2)(i);
(5)
Violation of the turbidity MCL under Section 16.3, where the Director determines after consultation that a
Tier 1 notice is required or where consultation does not take place within 24 hours after the system learns
of the violation;
203
(6)
Violation of the Surface Water Treatment Rule (SWTR), Interim Enhanced Surface Water Treatment Rule
(IESWTR), or Long Term 1 Interim Enhanced Surface Water Treatment Rule (LT1ESWTR), Section 5 of
these regulations, treatment technique requirements resulting from a single exceedance of the maximum
allowable turbidity limit (as identified in Appendix A), where the Director determines after consultation
that a Tier 1 notice is required or where consultation does not take place within 24 hours after the system
learns of the violation;
(7)
Occurrence of a waterborne disease outbreak, as defined in Section 1, or other waterborne emergency (such
as a failure or significant interruption in key water treatment processes, a natural disaster that disrupts the
water supply or distribution system, or a chemical spill or unexpected loading of possible pathogens into
the source water that significantly increases the potential for drinking water contamination);
(8)
Other violations or situations with significant potential to have serious adverse effects on human health as a
result of short-term exposure, as determined by the Director either in its regulations or on a case-by-case
basis.
(b)
When is the Tier 1 public notice to be provided? What additional steps are
required? PWSs must:
(1)
Provide a public notice as soon as practical but no later than 24
hours after the system learns of the violation;
(2)
Initiate consultation with the Director as soon as practical, but no
later than 24 hours after the PWS learns of the violation or
situation, to determine additional public notice requirements; and
(3)
Comply with any additional public notification requirements
(including any repeat notices or direction on the duration of the
posted notices) that are established as a result of the consultation
with the Director. Such requirements may include the timing, form,
manner, frequency and content of repeat notices (if any) and other
actions designed to reach all persons served.
(c)
What is the form and manner of the public notice? PWSs must provide the
notice within 24 hours in a form and manner reasonably calculated to
reach all persons served. The form and manner used by the PWS are to fit
the specific situation, but must be designed to reach residential, transient
and non-transient users of the water system. In order to reach all persons
served, water systems are to use, at a minimum, one (1) or more of the
following forms of delivery:
(1)
Appropriate broadcast media (such as radio and television);
(2)
Posting of the notice in conspicuous locations throughout the area
served by the water system;
(3)
Hand delivery of the notice to persons served by the water system;
or
204
(4)
Another delivery method approved in writing by the Director.
(3)
Tier 2 Public Notice–Form, Manner, and Frequency of Notice
(a)
Which violations or situations require a Tier 2 public notice? Table 1 of
this Section lists the violation categories and other situations requiring a
Tier 2 public notice. Appendix A to this subpart identifies the tier
assignment for each specific violation or situation.
Table 1 to Section 16.8 (3)–Violation Categories and Other Situations Requiring a Tier 2 Public Notice
(1)
All violations of the MCL, MRDL and treatment technique requirements, except where a Tier 1
notice is required under Section 16.8(2)(a) or where the Director determines that a Tier 1 notice is
required;
(2)
Violations of the monitoring and testing procedure requirements, where the Director determines that a
Tier 2 rather than a Tier 3 public notice is required, taking into account potential health impacts and
persistence of the violation; and
(3)
Failure to comply with the terms and conditions of any variance or exemption in place.
(b)
When is the Tier 2 public notice to be provided?
(1)
A PWS must provide the public notice as soon as practical, but no
later than 30 days after the system learns of the violation. If the
public notice is posted, the notice must remain in place for as long
as the violation or situation persists, but in no case for less than
seven (7) days, even if the violation or situation is resolved. The
Director may, in appropriate circumstances, allow additional time
for the initial notice of up to three (3) months from the date the
system learns of the violation. It is not appropriate for the Director
to grant an extension to the 30-day deadline for any unresolved
violation or to allow across-the-board extensions by rule or policy
for other violations or situations requiring a Tier 2 public notice.
Extensions granted by the Director must be in writing.
(2)
The PWS must repeat the notice every three (3) months as long as
the violation or situation persists, unless the Director determines
that appropriate circumstances warrant a different repeat notice
frequency. In no circumstance may the repeat notice be given less
frequently than once per year. It is not appropriate for the Director
to allow less frequent repeat notice for an MCL violation under the
Total Coliform Rule or a treatment technique violation under the
Surface Water Treatment Rule or Interim Enhanced Surface Water
Treatment Rule (Section 5). It is also not appropriate for the
Director to allow through its rules or policies across-the-board
reductions in the repeat notice frequency for other ongoing
violations requiring a Tier 2 repeat notice. The Director’s
205
determinations allowing repeat notices to be given less frequently
than once every three (3) months must be in writing.
(3)
For the turbidity violations specified in this Paragraph, PWSs must
consult with the Director as soon as practical but no later than 24
hours after the PWS learns of the violation, to determine whether a
Tier 1 public notice under Section 16.8 (2)(a) is required to protect
public health. When consultation does not take place within the 24-
hour period, the water system must distribute a Tier 1 notice of the
violation within the next 24 hours (i.e., no later than 48 hours after
the system learns of the violation), following the requirements
under sections 16.8(2)(b) and (c). Consultation with the Director is
required for:
(i)
Violation of the turbidity MCL under Section 16.3; or
(ii)
Violation of the SWTR, IESWTR or LT1ESWTR (Section
5) treatment technique requirement resulting from a single
exceedance of the maximum allowable turbidity limit.
(c)
What is the form and manner of the Tier 2 public notice? PWSs must
provide the initial public notice and any repeat notices in a form and
manner that is reasonably calculated to reach persons served in the
required time period. The form and manner of the public notice may vary
based on the specific situation and type of water system, but it must at a
minimum meet the following requirements:
(1)
Unless directed otherwise by the Director in writing, community
water systems must provide notice by:
(i)
Mail or other direct delivery to each customer receiving a
bill and to other service connections to which water is
delivered by the PWS; and
(ii)
Any other method reasonably calculated to reach other
persons regularly served by the system, if they would not
normally be reached by the notice required in Paragraph
(c)(1)(i) of this Section. Such persons may include those
who do not pay water bills or do not have service
connection addresses (e.g., house renters, apartment
dwellers, university students, nursing home patients, prison
inmates, etc.). Other methods may include: Publication in a
local
newspaper;
delivery
of
multiple
copies
for
distribution by customers that provide their drinking water
to others (e.g., apartment building owners or large private
206
employers); posting in public places served by the system
or on the Internet; or delivery to community organizations.
(2)
Unless directed otherwise by the Director in writing, non-
community water systems must provide notice by:
(i)
Posting the notice in conspicuous locations throughout the
distribution system frequented by persons served by the
system, or by mail or direct delivery to each customer and
service connection (where known); and
(ii)
Any other method reasonably calculated to reach other
persons served by the system if they would not normally be
reached by the notice required in Paragraph (c)(2)(i) of this
Section. Such persons may include those served who may
not see a posted notice because the posted notice is not in a
location they routinely pass by. Other methods may
include: Publication in a local newspaper or newsletter
distributed to customers; use of E-mail to notify employees
or students; or, delivery of multiple copies in central
locations (e.g., community centers).
(4)
Tier 3 Public Notice–Form, Manner, and Frequency of Notice
(a)
Which violations or situations require a Tier 3 public notice? Table 1 of
this Section lists the violation categories and other situations requiring a
Tier 3 public notice. Appendix A to this subpart identifies the tier
assignment for each specific violation or situation.
207
Table 1 to Section 16.8(4).–Violation Categories and Other Situations Requiring a Tier 3 Public Notice
(1)
Monitoring violations, except where a Tier 1 notice is required under Section 16.8(2) or where the
Director determines that a Tier 2 notice is required;
(2)
Failure to comply with a testing procedure established in these regulations, except where a Tier 1
notice is required under Section 16.8(2)(a) or where the Director determines that a Tier 2 notice is
required;
(3)
Operation under a variance granted under Section 1415 or an exemption granted under Section 1416
of the Safe Drinking Water Act (Section 15);
(4)
Availability of unregulated contaminant monitoring results, as required under Section 16.8(7); and
(5)
Exceedance of the fluoride secondary maximum contaminant level (SMCL), as required under
Section 16.8(8).
(b)
When is the Tier 3 public notice to be provided?
(1)
A PWS must provide the public notice not later than one (1) year
after the PWS learns of the violation or situation or begins
operating under a variance or exemption. Following the initial
notice, the PWS must repeat the notice annually for as long as the
violation, variance, exemption, or other situation persists. If the
public notice is posted, the notice must remain in place for as long
as the violation, variance, exemption, or other situation persists,
but in no case less than seven (7) days (even if the violation or
situation is resolved).
(2)
Instead of individual Tier 3 public notices, a PWS may use an
annual report detailing all violations and situations that occurred
during the previous twelve (12) months, as long as the timing
requirements of Paragraph (b)(1) of this Section are met.
(c)
What is the form and manner of the Tier 3 public notice? A PWS must
provide the initial notice and any repeat notices in a form and manner that
is reasonably calculated to reach persons served in the required time
period. The form and manner of the public notice may vary based on the
specific situation and type of water system, but it must at a minimum meet
the following requirements:
(1)
Unless directed otherwise by the Director in writing, community
water systems must provide notice by:
(i)
Mail or other direct delivery to each customer receiving a
bill and to other service connections to which water is
delivered by the PWS; and
(ii)
Any other method reasonably calculated to reach other
persons regularly served by the system, if they would not
208
normally be reached by the notice required in Paragraph
(c)(1)(i) of this Section. Such persons may include those
who do not pay water bills or do not have service
connection addresses (e.g., house renters, apartment
dwellers, university students, nursing home patients, prison
inmates, etc.). Other methods may include: Publication in a
local
newspaper;
delivery
of
multiple
copies
for
distribution by customers that provide their drinking water
to others (e.g., apartment building owners or large private
employers); posting in public places or on the Internet; or
delivery to community organizations.
(2)
Unless directed otherwise by the Director in writing, non-
community water systems must provide notice by:
(i)
Posting the notice in conspicuous locations throughout the
distribution system frequented by persons served by the
system, or by mail or direct delivery to each customer and
service connection (where known); and
(ii)
Any other method reasonably calculated to reach other
persons served by the system, if they would not normally
be reached by the notice required in Paragraph (c)(2)(i) of
this Section. Such persons may include those who may not
see a posted notice because the notice is not in a location
they routinely pass by. Other methods may include:
Publication in a local newspaper or newsletter distributed to
customers; use of E-mail to notify employees or students;
or, delivery of multiple copies in central locations (e.g.,
community centers).
(d)
In what situations may the Consumer Confidence Report be used to meet
the Tier 3 public notice requirements? For community water systems, the
Consumer Confidence Report (CCR) required under Section 16.10 of this
part may be used as a vehicle for the initial Tier 3 public notice and all
required repeat notices, as long as:
(1)
The CCR is provided to persons served no later than 12 months
after the system learns of the violation or situation as required
under Section 16.8(4)(b);
(2)
The Tier 3 notice contained in the CCR follows the content
requirements under Section 16.8(5); and
(3)
The CCR is distributed following the delivery requirements under
Section 16.8(4)(c).
209
(5)
Content of the Public Notice
(a)
What elements must be included in the public notice for violations of
National Primary Drinking Water Regulations (NPDWR) or other
situations requiring a public notice? When a PWS violates a NPDWR or
has a situation requiring public notification, each public notice must
include the following elements:
(1)
A description of the violation or situation, including the
contaminant(s) of concern, and (as applicable) the contaminant
level(s);
(2)
When the violation or situation occurred;
(3)
Any potential adverse health effects from the violation or situation,
including the standard language under Paragraph (d)(1) or (d)(2) of
this Section, whichever is applicable;
(4)
The population at risk, including subpopulations particularly
vulnerable if exposed to the contaminant in their drinking water;
(5)
Whether alternative water supplies should be used;
(6)
What actions consumers should take, including when they should
seek medical help, if known;
(7)
What the system is doing to correct the violation or situation;
(8)
When the water system expects to return to compliance or resolve
the situation;
(9)
The name, business address and phone number of the water system
owner, operator, or designee of the PWS as a source of additional
information concerning the notice; and
(10)
A statement to encourage the notice recipient to distribute the
public notice to other persons served, using the standard language
under Paragraph (d)(3) of this Section, where applicable.
(b)
What elements must be included in the public notice for a PWS operating
under a variance or exemption?
(1)
If a PWS has been granted a variance or an exemption, the public
notice must contain:
210
(i)
An explanation of the reasons for the variance or
exemption;
(ii)
The date on which the variance or exemption was issued;
(iii)
A brief status report on the steps the system is taking to
install treatment, find alternative sources of water, or
otherwise comply with the terms and schedules of the
variance or exemption; and
(iv)
A notice of any opportunity for public input in the review
of the variance or exemption.
(2)
If a PWS violates the conditions of a variance or exemption, the
public notice must contain the ten (10) elements listed in Paragraph
(a) of this Section.
(c)
How is the public notice to be presented?
(1)
Each public notice required by this Section:
(i)
Must be displayed in a conspicuous way
when printed or posted;
(ii)
Must not contain overly technical language
or very small print;
(iii)
Must not be formatted in a way that defeats
the purpose of the notice;
(iv)
Must not contain language which nullifies
the purpose of the notice.
(2)
Each public notice required by this Section must
comply with multilingual requirements, as follows:
(i)
For a PWS serving a large proportion of
non-English
speaking
consumers,
as
determined by the Director, the public notice
must contain information in the appropriate
language(s) regarding the importance of the
notice or contain a telephone number or
address where persons served may contact
the water system to obtain a translated copy
of the notice or to request assistance in the
appropriate language.
211
(ii)
In cases where the Director has not
determined
what
constitutes
a
large
proportion
of
non-English
speaking
consumers, the PWS must include in the
public notice the same information as in
Paragraph (c)(2)(i) of this Section, where
appropriate to reach a large proportion of
non-English speaking persons served by the
water system.
(d)
What standard language must a PWS include in their public
notice? PWSs are required to include the following
standard language in their public notice:
(1)
Standard health effects language for MCL or
MRDL violations, treatment technique violations,
and violations of the condition of a variance or
exemption. PWSs must include in each public
notice the health effects language specified in
Appendix B to this subpart corresponding to each
MCL, MRDL, and treatment technique violation
listed in Appendix A to this subpart, and for each
violation of a condition of a variance or exemption.
(2)
Standard language for monitoring and testing
procedure violations. PWSs must include the
following language in their notice, including the
language necessary to fill in the blanks, for all
monitoring and testing procedure violations listed in
Appendix A to this subpart:
We are required to monitor your drinking water for
specific contaminants on a regular basis. Results of
regular monitoring are an indicator of whether or
not your drinking water meets health standards.
During[compliance period], we “did not monitor or
test” or “did not complete all monitoring or testing”
for [contaminant(s)], and therefore cannot be sure of
the quality of your drinking water during that time.
(3)
Standard language to encourage the distribution of
the public notice to all persons served. PWSs must
include in their notice the following language
(where applicable):
212
Please share this information with all the other
people who drink this water, especially those who
may not have received this notice directly (for
example, people in apartments, nursing homes,
schools, and businesses). You can do this by posting
this notice in a public place or distributing copies by
hand or mail
(6)
Notice to New Billing Units or New Customers
(a)
What is the requirement for community water systems? Community water
systems must give a copy of the most recent public notice for any
continuing violation, the existence of a variance or exemption, or other
ongoing situations requiring a public notice to all new billing units or new
customers prior to or at the time service begins.
(b)
What is the requirement for non-community water systems? Non-
community water systems must continuously post the public notice in
conspicuous locations in order to inform new consumers of any continuing
violation, variance or exemption, or other situation requiring a public
notice for as long as the violation, variance, exemption, or other situation
persists.
(7)
Special Notice of the Availability of Unregulated Contaminant Monitoring
Results
(a)
When is the special notice to be given? The owner or operator of a
community water system or non-transient, non-community water system is
required to monitor under Section 141.40 of the Code of Federal
Regulations. They must notify persons served by the system of the
availability of the results of such sampling no later than 12 months after
the monitoring results are known.
(b)
What is the form and manner of the special notice? The form and manner
of the public notice must follow the requirements for a Tier 3 public notice
prescribed in Section 16.8 (4)(c), (d)(1) and (d)(3). The notice must also
identify a person and provide the telephone number to contact for
information on the monitoring results.
(8)
Special Notice for Exceedance of 2 mg/L Fluoride
(a)
When is the special notice to be given? Community water systems that
exceed the fluoride secondary maximum contaminant level (SMCL) of 2
mg/L (determined by the last single sample taken in accordance with
Section 16.1), but do not exceed the maximum contaminant level (MCL)
of 4 mg/L for fluoride (as specified in Section 16.1), must provide the
213
public notice in Paragraph (c) of this Section to persons served. Public
notice must be provided as soon as practical but no later than 12 months
from the day the water system learns of the exceedance. A copy of the
notice must also be sent to all new billing units and new customers at the
time service begins, and to the Director. The PWS must repeat the notice
at least annually for as long as the SMCL is exceeded. If the public notice
is posted, the notice must remain in place for as long as the SMCL is
exceeded, but in no case less than seven (7) days (even if the exceedance
is eliminated). On a case-by-case basis, the Director may require an initial
notice sooner than 12 months and repeat notices more frequently than
annually.
(b)
What is the form and manner of the special notice? The form and manner
of the public notice (including repeat notices) must follow the
requirements for a Tier 3 public notice in Section 16.8(c) and (d)(1) and
(d)(3).
(c)
What mandatory language must be contained in the special notice? The
notice must contain the following language, including the language
necessary to fill in the blanks:
This is an alert about your drinking water and a cosmetic dental problem
that might affect children under nine (9) years of age. At low levels,
fluoride can help prevent cavities, but children drinking water containing
more than 2 milligrams per liter (mg/L) of fluoride may develop cosmetic
discoloration of their permanent teeth (dental fluorosis). The drinking
water provided by your community water system [name] has a fluoride
concentration of [insert value] mg/L.
Dental fluorosis, in its moderate or severe forms, may result in a brown
staining and/or pitting of the permanent teeth. This problem occurs only in
developing teeth, before they erupt from the gums. Children under nine (9)
should be provided with alternative sources of drinking water or water that
has been treated to remove the fluoride to avoid the possibility of staining
and pitting of their permanent teeth. You may also want to contact your
dentist about proper use by young children of fluoride-containing
products. Older children and adults may safely drink the water.
Drinking water containing more than 4 mg/L of fluoride (the U.S.
Environmental Protection Agency's drinking water standard) can increase
your risk of developing bone disease. Your drinking water does not
contain more than 4 mg/L of fluoride, but we're required to notify you
when we discover that the fluoride levels in your drinking water exceed 2
mg/L because of this cosmetic dental problem.
214
For more information, please call [name of water system contact] of [name
of community water system] at [phone number]. Some home water
treatment units are also available to remove fluoride from drinking water.
To learn more about available home water treatment units, you may call
NSF International at 1-877-8-NSF-HELP.”
(9)
Special Notice for Nitrate Exceedances Above MCL by Non-community Water
Systems (NCWS), where Granted Permission by the Director Under Section 15.5.
(a)
When is the special notice to be given? The owner or operator of a non-
community water system granted permission by the Director under
Section 15.5 to exceed the nitrate MCL must provide notice to persons
served according to the requirements for a Tier 1 notice under Section
16.8(2)(a) and (b).
(b)
What is the form and manner of the special notice? Non-community water
systems granted permission by the Director to exceed the nitrate MCL
under Section 15.5 must provide continuous posting of the fact that nitrate
levels exceed 10 mg/L and the potential health effects of exposure,
according to the requirements for Tier 1 notice delivery under Section
16.8(2(c) and the content requirements under Section 16.8 (5).
(10)
Notice by Director on Behalf of the PWS
(a)
May the Director give the notice on behalf of the PWS? The Director may
give the notice required by this subpart on behalf of the owner and
operator of the PWS if the Director complies with the requirements of this
subpart.
(b)
What is the responsibility of the PWS when notice is given by the
Director? The owner or operator of the PWS remains responsible for
ensuring that the requirements of this subpart are met.
(11)
The PWS, within 10 days of completing the public notification requirements
under this part for the initial public notice and any repeat notices, must submit to
the Director a certification that it has fully complied with the public notification
regulations. The PWS must include with this certification a representative copy of
each type of notice distributed, published, posted, and made available to the
persons served by the system and to the media.
Copies of public notices issued pursuant to this Section and certifications made to
the Director pursuant to this Section must be kept for three (3) years after
issuance.
215
APPENDIX “A” TO SECTION 16.8 – NPDWR VIOLATIONS AND OTHER
SITUATIONS REQUIRING PUBLIC NOTICE 1
MCL/MRDL/TT violations2
Monitoring & testing procedure
violations
Contaminant
Tier of
public
notice
required
Citation
Tier of public
notice required
Citation
I. Violations of National Primary Drinking
Water Regulations (NPDWR) 3
A. Microbiological Contaminants
1. Total coliform
2
16.4(c)
3
16.4
2. Fecal coliform/E. Coli
1
16.4(c)
1, 34
16.4
3. Turbidity MCL
2
16.3
3
16.3
4. Turbidity MCL (average of 2 days'
samples >5 NTU)
2, 15
16.3
3
16.3
5. Turbidity (for TT violations resulting from
a single exceedance of maximum allowable
turbidity level)
2, 16
5.2.5 (2)
5.2.7
5.4.2 (1)(b)
5.4.2 (2)(c)
5.4.3 (2)
5.4.4 (2)
5.4.5
3
5.6
5.7
6. Surface Water Treatment Rule violations,
other than violations resulting from single
exceedance of max. allowable turbidity level
(TT).
2
5.0–5.8
3
5.5
7. Interim Enhanced Surface Water
Treatment Rule violations, other than
violations resulting from single exceedance
of max. turbidity level (TT)
2
5.1
5.2
5.3.7
5.4.2 (1)
5.4.2 (2)
5.4.5
3
5.3.7
5.7.1 (1)(b)
5.8.2 (4)
8. Filter Backwash Recycling Rule violations
2
5.1.5
5.8.4
3
5.1.5
5.8.4
9. Long Term 1 Enhanced Surface Water
Treatment Rule violations
2
5.1
5.2
5.3.7
5.4.2 (1)
5.4.2 (2)
5.4.5
3
5.3.7
5.7.1 (1)(b)
5.8.2 (4)
B. Inorganic Chemicals (IOCs)
1. Antimony
2
16.1
3
16.1
2. Arsenic
2
16.1 (j)(4)8
3
16.1(a) and
(c)11
3. Asbestos (fibers >10 µm)
2
16.1
3
16.1
4. Barium
2
16.1
3
16.1
5. Beryllium
2
16.1
3
16.1
6. Cadmium
2
16.1
3
16.1
7. Chromium (total)
2
16.1
3
16.1
216
MCL/MRDL/TT violations2
Monitoring & testing procedure
violations
Contaminant
Tier of
public
notice
required
Citation
Tier of public
notice required
Citation
8. Cyanide
2
16.1
3
16.1
9. Fluoride
2
16.1
3
16.1
10. Mercury (inorganic)
2
16.1
3
16.1
11. Nitrate
1
16.1
1, 312
16.1
12. Nitrite
1
16.1
1, 312
16.1
13. Total Nitrate and Nitrite
1
16.1
3
16.1
14. Selenium
2
16.1
3
16.1
15. Thallium
2
16.1
3
16.1
C. Lead and Copper Rule (Action Level for
lead is 0.015 mg/L, for copper is 1.3 mg/L)
1. Lead and Copper Rule (TT)
2
6.8–6.85
3
6.86–6.89
D. Synthetic Organic Chemicals (SOCs)
1. 2,4-D
2
16.2(a)
3
16.2(a)
2. 2,4,5-TP (Silvex)
2
16.2(a)
3
16.2(a)
3. Alachlor
2
16.2(a)
3
16.2(a)
4. Atrazine
2
16.2(a)
3
16.2(a)
5. Benzo(a)pyrene (PAHs)
2
16.2(a)
3
16.2(a)
6. Carbofuran
2
16.2(a)
3
16.2(a)
7. Chlordane
2
16.2(a)
3
16.2(a)
8. Dalapon
2
16.2(a)
3
16.2(a)
9. Di (2-ethylhexyl) adipate
2
16.2(a)
3
16.2(a)
10. Di (2-ethylhexyl) phthalate
2
16.2(a)
3
16.2(a)
11. Dibromochloropropane
2
16.2(a)
3
16.2(a)
12. Dinoseb
2
16.2(a)
3
16.2(a)
13. Dioxin (2,3,7,8-TCDD)
2
16.2(a)
3
16.2(a)
14. Diquat
2
16.2(a)
3
16.2(a)
15. Endothall
2
16.2(a)
3
16.2(a)
16. Endrin
2
16.2(a)
3
16.2(a)
17. Ethylene dibromide
2
16.2(a)
3
16.2(a)
18. Glyphosate
2
16.2(a)
3
16.2(a)
19. Heptachlor
2
16.2(a)
3
16.2(a)
20. Heptachlor epoxide
2
16.2(a)
3
16.2(a)
21. Hexachlorobenzene
2
16.2(a)
3
16.2(a)
22. Hexachlorocyclo-pentadiene
2
16.2(a)
3
16.2(a)
23. Lindane
2
16.2(a)
3
16.2(a)
24. Methoxychlor
2
16.2(a)
3
16.2(a)
25. Oxamyl (Vydate)
2
16.2(a)
3
16.2(a)
26. Pentachlorophenol
2
16.2(a)
3
16.2(a)
27. Picloram
2
16.2(a)
3
16.2(a)
28. Polychlorinated biphenyls (PCBs)
2
16.2(a)
3
16.2(a)
29. Simazine
2
16.2(a)
3
16.2(a)
30. Toxaphene
2
16.2(a)
3
16.2(a)
E. Volatile Organic Chemicals (VOCs)
1. Benzene
2
16.2(b)
3
16.2(b)
2. Carbon tetrachloride
2
16.2(b)
3
16.2(b)
3. Chlorobenzene (monochlorobenzene)
2
16.2(b)
3
16.2(b)
217
MCL/MRDL/TT violations2
Monitoring & testing procedure
violations
Contaminant
Tier of
public
notice
required
Citation
Tier of public
notice required
Citation
4. o-Dichlorobenzene
2
16.2(b)
3
16.2(b)
5. p-Dichlorobenzene
2
16.2(b)
3
16.2(b)
6. 1,2-Dichloroethane
2
16.2(b)
3
16.2(b)
7. 1,1-Dichloroethylene
2
16.2(b)
3
16.2(b)
8. cis-1,2-Dichloroethylene
2
16.2(b)
3
16.2(b)
9. trans-1,2-Dichloroethylene
2
16.2(b)
3
16.2(b)
10. Dichloromethane
2
16.2(b)
3
16.2(b)
11. 1,2-Dichloropropane
2
16.2(b)
3
16.2(b)
12. Ethylbenzene
2
16.2(b)
3
16.2(b)
13. Styrene
2
16.2(b)
3
16.2(b)
14. Tetrachloroethylene
2
16.2(b)
3
16.2(b)
15. Toluene
2
16.2(b)
3
16.2(b)
16. 1,2,4-Trichlorobenzene
2
16.2(b)
3
16.2(b)
17. 1,1,1-Trichloroethane
2
16.2(b)
3
16.2(b)
18. 1,1,2-Trichloroethane
2
16.2(b)
3
16.2(b)
19. Trichloroethylene
2
16.2(b)
3
16.2(b)
20. Vinyl chloride
2
16.2(b)
3
16.2(b)
21. Xylenes (total)
2
16.2(b)
3
16.2(b)
F. Radioactive Contaminants
1. Beta/photon emitters
2
16.5(c)
3
16.5
2. Alpha emitters
2
16.5(b)
3
16.5
3. Combined radium (226 & 228)
2
16.5(b)
3
16.5
4. Uranium
29
16.5(c)
310
16.5(a) and
Appendix 1,
Section D
G. Disinfection Byproducts (DBPs),
Byproduct Precursors, Disinfectant
Residuals. Where disinfection is used in the
treatment of drinking water, disinfectants
combine with organic and inorganic matter
present in water to form chemicals called
disinfection byproducts
(DBPs). EPA sets standards for controlling
the levels of disinfectants and DBPs in
drinking water, including trihalomethanes
(THMs) and haloacetic acids (HAAs). 13
2
1. Total trihalomethanes (TTHMs)
2
16.2(a)14
7.1(a)
3
16.2(a) (19)
2. Haloacetic Acids (HAA5)
2
7.1(a)
3
7.5 (a)–(b)
3. Bromate
2
7.1(a)
3
7.5 (a)–(b)
4. Chlorite
2
7.1(a)
3
7.5 (a)–(b)
5. Chlorine (MRDL)
2
7.2(a)
3
7.5 (a)–(b)
6. Chloramine (MRDL)
2
7.2(a)
3
7.5 (a), (c)
7. Chlorine dioxide (MRDL), where any 2
consecutive daily samples at entrance to
distribution system only are above MRDL
2
7.2(a)
7.6(c)(3)
2, 315
7.5 (a), (c)
7.6 (c)(2)
8. Chlorine dioxide (MRDL), where
116
7.2(a)
1
7.5 (a), (c)
218
MCL/MRDL/TT violations2
Monitoring & testing procedure
violations
Contaminant
Tier of
public
notice
required
Citation
Tier of public
notice required
Citation
sample(s) in distribution system the next day
are also above MRDL
7.6(c)(3)
7.6(c)(2)
9. Control of DBP precursors– TOC (TT)
2
7.8(a)–(b)
3
7.5 (9)(d)
10. Bench marking and disinfection profiling
N/A
N/A
3
5.3.7
11. Development of monitoring plan
N/A
N/A
3
7.5(8)
H. Other Treatment Techniques
1. Acrylamide (TT)
2
16.2(d)
N/A
N/A
2. Epichlorohydrin (TT)
2
16.2(d)
N/A
N/A
II. Unregulated Contaminant Monitoring 17
A. Unregulated contaminants
N/A
N/A
3
40 CFR 141-40
B. Nickel
N/A
N/A
3
16.1
III. Public Notification for Variances and
Exemptions:
A. Operation under a variance or exemption
3
1415, 141618
N/A
N/A
B. Violation of conditions of a variance or
exemption
2
1415, 1416
142.30719
N/A
N/A
IV. Other Situations Requiring Public
Notification:
A. Fluoride secondary maximum
contaminant level (SMCL) exceedance
3
16.8(8)
N/A
N/A
B. Exceedance of nitrate MCL for non-
community systems, as allowed by Director
1
15.5
N/A
N/A
C. Availability of unregulated contaminant
monitoring data
3
16.8(7)
N/A
N/A
D. Waterborne disease outbreak
1
Section 1
5.2.7 B.2
N/A
N/A
E. Other waterborne emergency20
1
N/A
N/A
N/A
F. Other situations as determined by the
Director
1, 2, 321
N/A
N/A
N/A
Appendix A – Endnotes
1.
Violations and other situations not listed in this table (e.g., reporting violations and
failure to prepare Consumer Confidence Reports), do not require notice, unless otherwise
determined by the Director. The Director may, at their option, also require a more
stringent public notice tier (e.g., Tier 1 instead of Tier 2 or Tier 2 instead of Tier 3) for
specific violations and situations listed in this Appendix, as authorized under sections
16.8(2)(a) and 16.8(3)(a).
2.
MCL – Maximum contaminant level, MRDL – Maximum residual disinfectant level, TT
– Treatment technique
3.
The term Violations of National Primary Drinking Water Regulations (NPDWR) is used
here to include violations of MCL, MRDL,
219
treatment technique, monitoring, and testing procedure requirements.
4.
Failure to test for fecal coliform or E. Coli is a Tier 1 violation if testing is not done after
any repeat sample tests positive for coliform. All other total coliform monitoring and
testing procedure violations are Tier 3.
5.
Systems that violate the turbidity MCL of 5 NTU based on an average of measurements
over two (2) consecutive days must consult with the Director within 24 hours after
learning of the violation. Based on this consultation, the Director may subsequently
decide to elevate the violation to Tier 1. If a system is unable to make contact with the
Director in the 24-hour period, the violation is automatically elevated to Tier 1.
6.
Systems with treatment technique violations involving a single exceedance of a
maximum turbidity limit under the Surface Water Treatment Rule (SWTR), the Interim
Enhanced Surface Water Treatment Rule (IESWTR), or the Long Term 1 Enhanced
Surface Water Treatment Rule (LT1ESWTR) are required to consult with the Director
within 24 hours after learning of the violation. Based on this consultation, the Director
may subsequently decide to elevate the violation to Tier 1. If a system is unable to make
contact with the Director in the 24-hour period, the violation is automatically elevated to
Tier 1.
7.
Most of the requirements of the Interim Enhanced Surface Water Treatment Rule (63 FR
69477) (Sections141.170-141.171, 141.173-141.174) become effective January 1, 2002
for Subpart H systems (surface water systems and ground water systems under the direct
influence of surface water) serving at least 10,000 persons. However, Section 141.172
has some requirements that become effective as early as April 16, 1999.
The Surface Water Treatment Rule remains in effect for systems serving at least 10,000
persons even after 2002; the Interim Enhanced
Surface Water Treatment Rule adds additional requirements and does not in many cases
supercede the SWTR.
8.
The arsenic MCL citations are effective January 23, 2006. Until then, refer to 16.1 (j)(4)
and (l).
9.
The uranium MCL Tier 2 violation citations are effective December 8, 2003 for all
community water systems.
10.
The uranium MCL Tier 3 violation citations are effective December 8, 2003 for all
community water systems.
11.
The arsenic Tier 3 violation MCL citations are effective January 23, 2006. Until then,
refer to 16.1 (a) and (j).
220
12.
Failure to take a confirmation sample within 24 hours for nitrate or nitrite after an initial
sample exceeds the MCL is a Tier 1 violation.
Other monitoring violations for nitrate are Tier 3.
13
Subpart H community and non-transient, non-community systems serving ≥ 10,000 must
comply with new DBP MCLs, disinfectant MRDLs, and related monitoring requirements
beginning January 1, 2002. All other community and non-transient non-community
systems must meet the MCLs and MRDLs beginning January 1, 2004. Subpart H
transient non-community systems serving 10,000 or more persons and using chlorine
dioxide as a disinfectant or oxidant must comply with the chlorine dioxide MRDL
beginning January 1, 2002. Subpart H transient non-community systems serving fewer
than 10,000 persons and using only ground water not under the direct influence of surface
water and using chlorine dioxide as a disinfectant or oxidant must comply with the
chlorine dioxide MRDL beginning January 1, 2004.
14.
Section 141.12 will no longer apply after January 1, 2004.
15.
Failure to monitor for chlorine dioxide at the entrance to the distribution system the day
after exceeding the MRDL at the entrance to the
distribution system is a Tier 2 violation.
16.
If any daily sample taken at the entrance to the distribution system exceeds the MRDL
for chlorine dioxide and one (1) or more samples taken in the distribution system the next
day exceed the MRDL, Tier 1 notification is required. Failure to take the required
samples in the distribution system after the MRDL is exceeded at the entry point also
triggers Tier 1 notification.
17.
Some water systems must monitor for certain unregulated contaminants listed in 40 CFR
141.40.
18.
This citation refers to Sections1415 and 1416 of the Safe Drinking Water Act.
Sections1415 and 1416 require that “a schedule prescribed. . . for a
PWS granted a variance [or exemption] shall require compliance by the system...”
19.
In addition to Sections1415 and 1416 of the Safe Drinking Water Act, 40 CFR 142.307
specifies the items and schedule milestones that must be
included in a variance for small systems.
20.
Other waterborne emergencies require a Tier 1 public notice under Section 16.8(2)(a) for
situations that do not meet the definition of a waterborne disease outbreak given in
Section 1 but that still have the potential to have serious adverse effects on health as a
result of short-term exposure. These could include outbreaks not related to treatment
deficiencies, as well as situations that have the potential to cause outbreaks, such as
failures or significant interruption in water treatment processes, natural disasters that
disrupt the water supply or distribution system, chemical spills, or unexpected loading of
possible pathogens into the source water.
221
21.
The Director may place other situations in any tier they believe appropriate, based on
threat to public health.
222
APPENDIX B TO SECTION 16.8 – STANDARD HEALTH EFFECTS
LANGUAGE FOR PUBLIC NOTIFICATION
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
National Primary Drinking Water Regulations (NPDWR)
A. Microbiological Contaminants
1a. Total coliform
Zero
See
footnote 3
Coliforms are bacteria that are naturally
present in the environment and are used as an
indicator that other, potentially harmful,
bacteria may be present. Coliforms were
found in more samples than allowed and this
was a warning of potential problems.
1b. Fecal coliform/E. Coli
Zero
Zero
Fecal coliforms and E. Coli are bacteria
whose presence indicates that the water may
be contaminated with human or animal
wastes. Microbes in these wastes can cause
short- term effects, such as diarrhea, cramps,
nausea, headaches, or other symptoms. They
may pose a special health risk for infants,
young children and people with severely
compromised immune systems.
2a. Turbidity (MCL4)
None
1 NTU5
5 NTU
Turbidity has no health effects. However,
turbidity can interfere with disinfection and
provide a medium for microbial growth.
Turbidity may indicate the presence of
disease-causing organisms. These organisms
include bacteria, viruses, and parasites that
can cause symptoms such as nausea, cramps,
diarrhea and associated headaches.
2b. Turbidity (SWTR TT)6
None
TT7
Turbidity has no health effects. However,
turbidity can interfere with disinfection and
provide a medium for microbial growth.
Turbidity may indicate the presence of
disease-causing organisms. These organisms
include bacteria, viruses, and parasites that
can cause symptoms such as nausea, cramps,
diarrhea and associated headaches.
2c. Turbidity (IESWTR
TT and LT1ESWTR TT)8
None
TT
Turbidity has no health effects. However,
turbidity can interfere with disinfection and
provide a medium for microbial growth.
Turbidity may indicate the presence of
disease-causing organisms. These organisms
include bacteria, viruses, and parasites that
can cause symptoms such as nausea, cramps,
diarrhea and associated headaches.
B. Surface Water Treatment Rule (SWTR) and Interim Enhanced Surface Water Treatment Rule (IESWTR), Long
Term 1 Enhanced Surface Water Treatment Rule (LT1ESWTR) and the Filter Backwash Recycling Rule (FBRR)
violations
3. Giardia lamblia
(SWTR/IESWTR/LT1ESWTR)
4. Viruses
(SWTR/IESWTR/LT1ESWTR)
5. Heterotrophic plate count (HPC)
Zero
TT10
Inadequately treated water may contain
disease-causing organisms. These organisms
include bacteria, viruses and parasites which
can cause symptoms such as nausea, cramps,
diarrhea, and associated headaches.
223
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
bacteria 9
(SWTR/IESWTR/LT1ESWTR)
6. Legionella
(SWTR/IESWTR/LT1ESWTR)
7. Cryptosporidium
(IESWTR/FBRR/LT1ESWTR)
C. Inorganic Chemicals (IOCs)
8. Antimony
0.006
0.006
Some people who drink water containing
antimony well in excess of the MCL over
many years could experience increases in
blood cholesterol and decreases in blood
sugar.
9. Arsenic11
Zero
0.010
Some people who drink water containing
arsenic in excess of the MCL over many years
could experience skin damage or problems
with their circulatory system, and may have
an increased risk of getting cancer.
10. Asbestos (>10 µm)
7 MFL12
7 MFL
Some people who drink water containing
asbestos in excess of the MCL over many
years may have an increased risk of
developing benign intestinal polyps.
11. Barium
2
2
Some people who drink water containing
barium in excess of the MCL over many years
could experience an increase in their blood
pressure.
12. Beryllium
0.004
0.004
Some people who drink water containing
beryllium well in excess of the MCL over
many years could develop intestinal lesions.
13. Cadmium
0.005
0.005
Some people who drink water containing
cadmium in excess of the MCL over many
years could experience kidney damage.
14. Chromium (total)
0.1
0.1
Some people who use water containing
chromium well in excess of the MCL over
many years could experience allergic
dermatitis.
15. Cyanide
0.2
0.2
Some people who drink water containing
cyanide well in excess of the MCL over many
years could experience nerve damage or
problems with their thyroid.
16. Fluoride
4.0
4.0
Some people who drink water containing
fluoride in excess of the MCL over many
years could get bone disease, including pain
and tenderness of the bones. Fluoride in
drinking water at half the MCL or more may
cause mottling of children's teeth, usually in
children less than nine (9) years old. Mottling,
also known as dental flurosis, may include
brown staining and/or pitting of the teeth, and
occurs only in developing teeth, before they
erupt from the gums.
17. Mercury (inorganic)
0.002
0.002
Some people who drink water containing
inorganic mercury well in excess of the MCL
over many years could experience kidney
224
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
damage.
18. Nitrate
10
10
Infants below the age of six (6) months who
drink water containing nitrate in excess of the
MCL could become seriously ill and, if
untreated, may die. Symptoms include
shortness of breath and blue-baby syndrome.
19. Nitrite
1
1
Infants below the age of six (6) months who
drink water containing nitrite in excess of the
MCL could become seriously ill and, if
untreated, may die. Symptoms include
shortness of breath and blue-baby syndrome.
20. Total Nitrate and Nitrite
10
10
Infants below the age of six (6) months who
drink water containing nitrate and nitrite in
excess of the MCL could become seriously ill
and, if untreated, may die. Symptoms include
shortness of breath and blue baby syndrome.
21. Selenium
0.05
0.05
Selenium is an essential nutrient. However
some people who drink water containing
selenium in excess of the MCL over many
years could experience hair or fingernail
losses, numbness in fingers or toes, or
problems with their circulation.
22. Thallium
0.0005
0.002
Some people who drink water containing
thallium in excess of the MCL over many
years could experience hair loss, changes in
their blood, or problems with their kidneys,
intestines, or liver.
D. Lead and Copper Rule
23. Lead
Zero
TT13
Infants and children who drink water
containing lead in excess of the action level
could experience delays in their physical or
mental development. Children could show
slight deficits in attention span and learning
abilities. Adults who drink this water over
many years could develop kidney problems or
high blood pressure.
24. Copper
1.3
TT14
Copper is an essential nutrient, but some
people who drink water containing copper in
excess of the action level over a relatively
short amount of time could experience
gastrointestinal distress. Some people who
drink water containing copper in excess of the
action level over many years could suffer liver
or kidney damage. People with Wilson's
Disease should consult their personal doctor.
E. Synthetic Organic Compounds (SOCs)
25. 2,4-D
0.07
0.07
Some people who drink water containing the
weed killer 2,4-D well in excess of the MCL
over many years could experience problems
with their kidneys, liver, or adrenal glands.
26. 2,4,5-TP (Silvex)
0.05
0.05
Some people who drink water containing
silvex in excess of the MCL over many years
could experience liver problems.
225
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
27. Alachlor
Zero
0.002
Some people who drink water containing
alachlor in excess of the MCL over many
years could have problems with their eyes,
liver, kidneys, or spleen, experience anemia,
or may have an increased risk of getting
cancer.
28. Atrazine
0.003
0.003
Some people who drink water containing
atrazine well in excess of the MCL over many
years could experience problems with their
cardiovascular system or reproductive
difficulties.
29. Benzo(a)pyrene (PAHs)
Zero
0.0002
Some people who drink water containing
benzo(a)pyrene in excess of the MCL over
many years may experience reproductive
difficulties or may have an increased risk of
getting cancer.
30. Carbofuran
0.04
0.04
Some people who drink water containing
carbofuran in excess of the MCL over many
years could experience problems with their
blood, or nervous or reproductive systems.
31. Chlordane
Zero
0.002
Some people who drink water containing
chlordane in excess of the MCL over many
years could experience problems with their
liver, or nervous system, and may have an
increased risk of getting cancer.
32. Dalapon
0.2
0.2
Some people who drink water containing
dalapon well in excess of the MCL over many
years could experience minor kidney changes.
33. Di (2-ethylhexyl) adipate
0.4
0.4
Some people who drink water containing di
(2- ethylhexyl) adipate well in excess of the
MCL over many years could experience
general toxic effects or reproductive
difficulties.
34. Di(2-ethylhexyl) phthalate
Zero
0.006
Some people who drink water containing di
(2- ethylhexyl) phthalate in excess of the
MCL over many years may have problems
with their liver, or experience reproductive
difficulties, and may have an increased risk of
getting cancer.
35. Dibromochloropropane
(DBCP)
Zero
0.0002
Some people who drink water containing
DBCP in excess of the MCL over many years
could experience reproductive difficulties and
may have an increased risk of getting cancer.
36. Dinoseb
0.007
0.007
Some people who drink water containing
dinoseb well in excess of the MCL over many
years could experience reproductive
difficulties.
37. Dioxin (2,3,7,8-TCDD)
Zero
3 x 10-8
Some people who drink water containing
dioxin in excess of the MCL over many years
could experience reproductive difficulties and
may have an increased risk of getting cancer.
38. Diquat
0.02
0.02
Some people who drink water containing
diquat in excess of the MCL over many years
226
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
could get cataracts.
39. Endothall
0.1
0.1
Some people who drink water containing
endothall in excess of the MCL over many
years could experience problems with their
stomach or intestines.
40. Endrin
0.002
0.002
Some people who drink water containing
endrin in excess of the MCL over many years
could experience liver problems.
41. Ethylene dibromide
Zero
0.00005
Some people who drink water containing
ethylene dibromide in excess of the MCL over
many years could experience problems with
their liver, stomach, reproductive system, or
kidneys, and may have an increased risk of
getting cancer.
42. Glyphosate
0.7
0.7
Some people who drink water containing
glyphosate in excess of the MCL over many
years could experience problems with their
kidneys or reproductive difficulties.
43. Heptachlor
Zero
0.0004
Some people who drink water containing
heptachlor in excess of the MCL over many
years could experience liver damage and may
have an increased risk of getting cancer.
44. Heptachlor epoxide
Zero
0.0002
Some people who drink water containing
heptachlor epoxide in excess of the MCL over
many years could experience liver damage,
and may have an increased risk of getting
cancer.
45. Hexachlorobenzene
Zero
0.001
Some people who drink water containing
hexachlorobenzene in excess of the MCL over
many years could experience problems with
their liver or kidneys, or adverse reproductive
effects, and may have an increased risk of
getting cancer.
46. Hexachlorocyclopentadiene
0.05
0.05
Some people who drink water containing
hexachlorocyclopentadiene well in excess of
the MCL over many years could experience
problems with their kidneys or stomach.
47. Lindane
0.0002
0.0002
Some people who drink water containing
lindane in excess of the MCL over many years
could experience problems with their kidneys
or liver.
48. Methoxychlor
0.04
0.04
Some people who drink water containing
methoxychlor in excess of the MCL over
many years could experience reproductive
difficulties.
49. Oxamyl (Vydate)
0.2
0.2
Some people who drink water containing
oxamyl in excess of the MCL over many
years could experience slight nervous system
effects.
50. Pentachlorophenol
Zero
0.001
Some people who drink water containing
pentachlorophenol in excess of the MCL over
many years could experience problems with
their liver or kidneys, and may have an
227
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
increased risk of getting cancer.
51. Picloram
0.5
0.5
Some people who drink water containing
picloram in excess of the MCL over many
years could experience problems with their
liver.
52. Polychlorinated biphenyls
(PCBs)
Zero
0.0005
Some people who drink water containing
PCBs in excess of the MCL over many years
could experience changes in their skin,
problems with their thymus gland, immune
deficiencies, or reproductive or nervous
system difficulties, and may have an increased
risk of getting cancer.
53. Simazine
0.004
0.004
Some people who drink water containing
simazine in excess of the MCL over many
years could experience problems with their
blood.
54. Toxaphene
Zero
0.003
Some people who drink water containing
toxaphene in excess of the MCL over many
years could have problems with their kidneys,
liver, or thyroid, and may have an increased
risk of getting cancer.
F. Volatile Organic Chemicals (VOCs)
55. Benzene
Zero
0.005
Some people who drink water containing
benzene in excess of the MCL over many
years could experience anemia or a decrease
in blood platelets, and may have an increased
risk of getting cancer.
56. Carbon tetrachloride
Zero
0.005
Some people who drink water containing
carbon tetrachloride in excess of the MCL
over many years could experience problems
with their liver and may have an increased
risk of getting cancer.
57. Chlorobenzene
(monochlorobenzene)
0.1
0.1
Some people who drink water containing
chlorobenzene in excess of the MCL over
many years could experience problems with
their liver or kidneys
58. o-Dichlorobenzene
0.6
0.6
Some people who drink water containing o-
dichlorobenzene well in excess of the MCL
over many years could experience problems
with their liver, kidneys, or circulatory
systems.
59. p-Dichlorobenzene
0.075
0.075
Some people who drink water containing p-
dichlorobenzene in excess of the MCL over
many years could experience anemia, damage
to their liver, kidneys, or spleen, or changes in
their blood.
60. 1,2-Dichloroethane
Zero
0.005
Some people who drink water containing 1,2-
dichloroethane in excess of the MCL over
many years may have an increased risk of
getting cancer.
61. 1,1-Dichloroethylene
0.007
0.007
Some people who drink water containing 1,1-
dichloroethylene in excess of the MCL over
many years could experience problems with
228
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
their liver.
62. cis-1,2-Dichloroethylene
0.07
0.07
Some people who drink water containing cis-
1,2-dichloroethylene in excess of the MCL
over many years could experience problems
with their liver.
63. trans-1,2-Dichloroethylene
0.1
0.1
Some people who drink water containing
trans-1,2-dichloroethylene well in excess of
the MCL over many years could experience
problems with their liver.
64. Dichloromethane
Zero
0.005
Some people who drink water containing
dichloromethane in excess of the MCL over
many years could have liver problems and
may have an increased risk of getting cancer.
65. 1,2-Dichloropropane
Zero
0.005
Some people who drink water containing 1,2-
dichloropropane in excess of the MCL over
many years may have an increased risk of
getting cancer.
66. Ethylbenzene
0.7
0.7
Some people who drink water containing
ethylbenzene well in excess of the MCL over
many years could experience problems with
their liver or kidneys.
67. Styrene
0.1
0.1
Some people who drink water containing
styrene well in excess of the MCL over many
years could have problems with their liver,
kidneys, or circulatory system.
68. Tetrachloroethylene
Zero
0.005
Some people who drink water containing
tetrachloroethylene in excess of the MCL over
many years could have problems with their
liver, and may have an increased risk of
getting cancer.
69. Toluene
1
1
Some people who drink water containing
toluene well in excess of the MCL over many
years could have problems with their nervous
system, kidneys, or liver.
70. 1,2,4-Trichlorobenzene
0.07
0.07
Some people who drink water containing
1,2,4-trichlorobenzene well in excess of the
MCL over many years could experience
changes in their adrenal glands.
71. 1,1,1-Trichloroethane
0.2
0.2
Some people who drink water containing
1,1,1- trichloroethane in excess of the MCL
over many years could experience problems
with their liver, nervous system, or circulatory
system.
72. 1,1,2-Trichloroethane
0.003
0.005
Some people who drink water containing
1,1,2- trichloroethane well in excess of the
MCL over many years could have problems
with their liver, kidneys, or immune systems.
73. Trichloroethylene
Zero
0.005
Some people who drink water containing
trichloroethylene in excess of the MCL over
many years could experience problems with
their liver and may have an increased risk of
getting cancer.
74. Vinyl chloride
Zero
0.002
Some people who drink water containing
229
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
vinyl chloride in excess of the MCL over
many years may have an increased risk of
getting cancer.
75. Xylenes (total)
10
10
Some people who drink water containing
xylenes in excess of the MCL over many
years could experience damage to their
nervous system.
G. Radioactive Contaminants
76. Beta/photon emitters
Zero
4
mrem/yr15
Certain minerals are radioactive and may emit
forms of radiation known as photons and beta
radiation. Some people who drink water
containing beta and photon emitters in excess
of the MCL over many years may have an
increased risk of getting cancer.
77. Alpha emitters
(Gross alpha)
Zero
15 pCi/L16
Certain minerals are radioactive and may emit
a form of radiation known as alpha radiation.
Some people who drink water containing
alpha emitters in excess of the MCL over
many years may have an increased risk of
getting cancer.
78. Combined radium (226 & 228)
Zero
5 pCi/L
Some people who drink water containing
radium 226 or 228 in excess of the MCL over
many years may have an increased risk of
getting cancer.
79. Uranium17
Zero
30 µg/L
Some people who drink water containing
uranium in excess of the MCL over many
years may have an increased risk of getting
cancer and kidney toxicity.
H. Disinfection Byproducts (DBPs), Byproduct Precursors, and Disinfectant Residuals: Where disinfection is used
in the treatment of drinking water, disinfectants combine with organic and inorganic matter present in water to form
chemicals called disinfection byproducts (DBPs). EPA also sets standards for controlling the levels of disinfectants
and DBPs in drinking water, which include trihalomethanes (THMs) and haloacetic acids (HAAs).18
80. Total trihalomethanes
(TTHMs)
N/A
0.10/
0.08019,20
Some people who drink water containing
trihalomethanes in excess of the MCL over
many years may experience problems with
their liver, kidneys, or central nervous system,
and may have an increased risk of getting
cancer.
81. Haloacetic Acids (HAA5)
N/A
0.06021
Some people who drink water containing
HAAs in excess of the MCL over many years
may have an increased risk of developing
cancer.
82. Bromate
Zero
0.010
Some people who drink water containing
bromate in excess of the MCL over many
years may have an increased risk of
developing cancer.
83. Chlorite
0.8
1.0
Some infants and young children who drink
water containing chlorite in excess of the
MCL could experience nervous system
effects. Similar effects may occur in fetuses of
pregnant mothers who drink water containing
chlorite in excess of the MCL. Some people
may experience anemia.
230
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
84. Chlorine
4 (MRDLG)21
4.0
(MRDL)23
Some people who contact drinking water
containing chlorine well in excess of the
MRDL could experience irritating effects to
their eyes and nose. Some people who drink
water containing chlorine well in excess of the
MRDL could experience stomach discomfort.
85. Chloramines
4 (MRDLG)
4.0
(MRDL)
Some people who contact drinking water
containing chloramines well in excess of the
MRDL could experience irritating effects to
their eyes and nose. Some people who drink
water containing chloramines well in excess
of the MRDL could experience stomach
discomfort or anemia.
86a. Chlorine dioxide, where any 2
consecutive daily samples taken at
the entrance to the distribution
system are above the MRDL
0.8 (MRDLG)
0.8
(MRDL)
Some infants and young children who drink
water containing chlorine dioxide in excess of
the MRDL could experience nervous system
effects. Similar effects may occur in fetuses of
pregnant mothers who drink water containing
chlorine dioxide in excess of the MRDL.
Some people may experience anemia.
Add for public notification only: The chlorine
dioxide violations reported today are the result
of exceedances at the treatment facility only,
not within the distribution system which
delivers water to consumers. Continued
compliance with chlorine dioxide levels
within the distribution system minimizes the
potential risk of these violations to consumers.
86b. Chlorine dioxide, where one
(1) or more distribution system
sample(s) are above the MRDL
0.8 (MRDLG)
0.8
(MRDL)
Some infants and young children who drink
water containing chlorine dioxide in excess of
the MRDL could experience nervous system
effects. Similar effects may occur in fetuses of
pregnant mothers who drink water containing
chlorine dioxide in excess of the MRDL.
Some people may experience anemia.
Add for public notification only: The chlorine
dioxide violations reported today include
exceedances of the EPA standard within the
distribution system which delivers water to
consumers. Violations of the chlorine dioxide
standard within the distribution system may
harm human health based on short-term
exposures. Certain groups, including fetuses,
infants and young children, may be especially
susceptible to nervous system effects from
excessive chlorine dioxide exposure.
87. Control of DBP precursors
(TOC)
None
TT
Total organic carbon (TOC) has no health
effects. However, total organic carbon
provides a medium for the formation of
disinfection by products. These byproducts
include trihalomethanes (THMs) and
haloacetic acids (HAAs), which may lead to
adverse health effects, liver or kidney
problems, or nervous system effects, and may
231
Contaminant
MCLG1 mg/L
MCL2
mg/L
Standard health effects language for public
notification
lead to an increased risk of getting cancer.
I. Other Treatment Techniques
88. Acrylamide
Zero
TT
Some people who drink water containing high
levels of acrylamide over a long period of
time could have problems with their nervous
system or blood, and may have an increased
risk of getting cancer.
89. Epichlorohydrin
Zero
TT
Some people who drink water containing high
levels of epichlorohydrin over a long period
of time could experience stomach problems,
and may have an increased risk of getting
cancer.
Appendix B – Endnotes
1.
MCLG–Maximum contaminant level goal.
2.
MCL–Maximum contaminant level.
3.
For water systems analyzing at least 40 samples per month, no more than 5.0
percent of the monthly samples may be positive for total coliforms. For systems
analyzing fewer than 40 samples per month, no more than one (1) sample per
month may be positive for total coliforms.
4.
There are various regulations that set turbidity standards for different types of
systems, including 40 CFR 141.13, the 1989 Surface Water Treatment Rule, the
1998 Interim Enhanced Surface Water Treatment Rule, and the 2001 Long Term
1 Enhanced Surface Water Treatment Rule. The MCL for the monthly turbidity
average is 1 NTU; the MCL for the 2-day average is 5 NTU for systems that are
required to filter but have not yet installed filtration (40 CFR 141.13).
5.
NTU–Nephelometric turbidity unit.
6.
There are various regulations that set turbidity standards for different types of
systems, including 40 CFR 141.13, the 1989 Surface Water Treatment Rule, the
1998 Interim Enhanced Surface Water Treatment Rule, and the 2001 Long Term
1 Enhanced Surface Water Treatment Rule. Systems subject to the Surface Water
Treatment Rule (both filtered and unfiltered) may not exceed 5 NTU. In addition,
in filtered systems, 95 percent of samples each month must not exceed 0.5 NTU
in systems using conventional or direct filtration and must not exceed 1 NTU in
systems using slow sand or diatomaceous earth filtration or other filtration
technologies approved by the Director.
7.
TT–Treatment technique.
232
8.
There are various regulations that set turbidity standards for different types of
systems, including 40 CFR 141.13, the 1989 Surface Water Treatment Rule
(SWTR), the 1998 Interim Enhanced Surface Water Treatment Rule (IESWTR)
and the 2001 Long Term 1 Enhanced Surface Water Treatment Rule
(LT1ESWTR). For systems subject to the IESWTR (systems serving at least
10,000 people, using surface water or ground water under the direct influence of
surface water), that use conventional filtration or direct filtration, after January 1,
2002, the turbidity level of a system's combined filter effluent may not exceed 0.3
NTU in at least 95 percent of monthly measurements, and the turbidity level of a
system's combined filter effluent must not exceed 1 NTU at any time. Systems
subject to the IESWTR using technologies other than conventional, direct, slow
sand, or diatomaceous earth filtration must meet turbidity limits set by the
Director. For systems subject to the LT1ESWTR (systems serving fewer than
10,000 people, using surface water or ground water under the influence of surface
water) that use conventional or direct filtration, after January 1, 2005 the turbidity
level of a system’s combined filter effluent may not exceed 0.3 NTU in at least 95
percent of monthly measurements, and the turbidity level of a system’s combined
filter effluent must not exceed 1 NTU at any time. Systems subject to the
LT1ESWTR using technologies other than conventional, direct, slow sand, or
diatomaceous earth filtration must meet turbidity limits set by the Director.
9.
The bacteria detected by heterotrophic plate count (HPC) are not necessarily
harmful. HPC is simply an alternative method of determining disinfectant residual
levels. The number of such bacteria is an indicator of whether there is enough
disinfectant in the distribution system.
10.
SWTR, IESWTR and LT1ESWTR treatment technique violations that involve
turbidity exceedances may use the health effects language for turbidity instead.
11.
These arsenic values (MCL, MCLG) are effective January 23, 2006. Until then,
the MCL is 0.05 mg/L and there is no MCLG.
12.
Millions of fibers per liter.
13.
Action Level=0.015 mg/L.
14.
Action Level=1.3 mg/L.
15.
Millirems per year.
16.
Picocuries per liter.
17.
The uranium MCL is effective December 8, 2003 for all community water
systems.
233
18.
Surface water systems and ground water systems under the direct influence of
surface water are regulated under Subpart H of 40 CFR part 141. Subpart H
community and non-transient non-community systems serving ≥10,000 must
comply with DBP MCLs and disinfectant maximum residual disinfectant levels
(MRDLs) beginning January 1, 2002. All other community and non-transient
noncommunity systems must meet the MCLs and MRDLs beginning January 1,
2004. Subpart H transient non-community systems serving 10,000 or more
persons and using chlorine dioxide as a disinfectant or oxidant must comply with
the chlorine dioxide MRDL beginning January 1, 2002. Subpart H transient non-
community systems serving fewer than 10,000 persons and systems using only
ground water not under the direct influence of surface water and using chlorine
dioxide as a disinfectant or oxidant must comply with the chlorine dioxide MRDL
beginning January 1, 2004.
19.
The MCL of 0.10 mg/L for TTHMs is in effect until January 1, 2002 for Subpart
H community water systems serving 10,000 or more. This MCL is in effect until
January 1, 2004 for community water systems with a population of 10,000 or
more using only ground water not under the direct influence of surface water.
After these deadlines, the MCL will be 0.080 mg/L. On January 1, 2004, all
systems serving less than 10,000 will have to comply with the new MCL as well.
20.
The MCL for total trihalomethanes is the sum of the concentrations of the
individual trihalomethanes.
21.
The MCL for haloacetic acids is the sum of the concentrations of the individual
haloacetic acids.
22.
MRDLG–Maximum residual disinfectant level goal.
23.
MRDL–Maximum residual disinfectant level.
Appendix C To Section 16.8
List of Acronyms Used in Public Notification Regulation
CCR Consumer Confidence Report
CWS Community Water System
DBP Disinfection Byproduct
EPA Environmental Protection Agency
FBRR Filter Backwash Recycling Rule
HPC Heterotrophic Plate Count
234
IESWTR Interim Enhanced Surface Water Treatment Rule
IOC Inorganic Chemical
LCR Lead and Copper Rule
LT1ESWTR Long Term 1 Enhanced Surface Water Treatment Rule
MCL Maximum Contaminant Level
MCLG Maximum Contaminant Level Goal
MRDL Maximum Residual Disinfectant Level
MRDLG Maximum Residual Disinfectant Level Goal
NCWS Non-Community Water System
NPDWR National Primary Drinking Water Regulation
NTNCWS Non-Transient Non-Community Water System
NTU Nephelometric Turbidity Unit
OGWDW Office of Ground Water and Drinking Water
OW Office of Water
PN Public Notification
PWS PWS
SDWA Safe Drinking Water Act
SMCL Secondary Maximum Contaminant Level
SOC Synthetic Organic Chemical
SWTR Surface Water Treatment Rule
TCR Total Coliform Rule
TT Treatment Technique
TWS Transient Non-Community Water System
235
VOC Volatile Organic Chemical
16.9 Records
a)
Records of analyses shall be maintained by the water purveyor. The records of
each sample analyzed to comply with these regulations shall contain the following
information:
1.
The time, date and place of sampling and the name of the sample
collector;
2.
The sampling point and the reason for collection;
3.
Date analysis started and completion date if more than one (1) day is
needed;
4.
Name of laboratory and person responsible for performing the analysis;
5.
The analytical technique or method used;
6.
The results of the analysis.
b)
Records of microbiological examinations shall be readily available for at least 5
years.
c)
Records of organic and inorganic chemical, radiological and turbidity analyses
shall be readily available for at least 10 years.
d)
Any written document relating to a sanitary survey of a PWS shall be kept for at
least 10 years. Records of action taken to correct a violation of these regulations
shall be kept for at least 3 years after the last action taken with respect to the
particular violation involved.
e)
Records concerning a variance or exemption granted to a system shall be kept for
at least 5 years following the expiration date of such variance or exemption.
16.10 Consumer Confidence Reports
(1)
Purpose and Applicability of this Subpart
(a)
This subpart establishes the minimum requirements for the content of
annual reports that community water systems must deliver to their
customers. These reports must contain information on the quality of the
water delivered by the systems and characterize the risks (if any) from
236
exposure to contaminants detected in the drinking water in an accurate and
understandable manner.
(b)
Notwithstanding the provisions of Section 2, this subpart applies only to
community water systems.
(c)
For the purpose of this subpart, customers are defined as billing units or
service connections to which water is delivered by a community water
system.
(d)
For the purpose of this subpart, detected means: at or above the levels
prescribed by Appendix 1for the inorganic contaminants listed at
16.1(1)—(15), for the synthetic organic contaminants listed in 16.2(a) or
the volatile organic contaminants listed in 16.2(b)(1)—(21) or the
radioactive contaminants listed at 16.5(b), (c) and (d).
(2)
Effective Dates
(a)
The regulations in this subpart shall take effect on January 1, 2000.
(b)
Each existing community water system must deliver a consumer
confidence report by July 1, 2000, and subsequent reports by July 1
annually thereafter. The first report must contain data collected during, or
prior to, calendar year 1999 as prescribed in 16.10(3)(d)(3). Each report
thereafter must contain data collected during, or prior to, the previous
calendar year.
(c)
A new community water system must deliver its first report by July 1 of
the year after its first full calendar year in operation and annually
thereafter.
(d)
A community water system that sells water to another community water
system must deliver the applicable information required in 16.10(3) to the
buyer system:
(1)
No later than April 1, 2000, and by April 1 annually thereafter or
(2)
On a date mutually agreed upon by the seller and the purchaser,
and specifically included in a contract between the parties.
(3)
Content of the Reports
(a)
Each community water system must provide to its customers an annual
report that contains the information specified in this Section and Section
16.10(4).
237
(b)
Information on the source of the water delivered:
(1)
Each report must identify the source(s) of the water delivered by
the community water system by providing information on:
(i)
The type of the water: e.g., surface water, ground water;
and
(ii)
The commonly used name (if any) and location of the body
(or bodies) of water.
(2)
If a source water assessment has been completed, the report must
notify consumers of the availability of this information and the
means to obtain it. In addition, systems are encouraged to highlight
in the report significant sources of contamination in the source
water area if they have readily available information. Where a
system has received a source water assessment from the
Department, the report must include a brief summary of the
system's susceptibility to potential sources of contamination, using
language provided by the Department or written by the operator.
(c)
Definitions
(1)
Each report must include the following definitions:
(i)
Maximum Contaminant Level Goal or MCLG: The level of
a contaminant in drinking water below which there is no
known or expected risk to health. MCLGs allow for a
margin of safety.
(ii)
Maximum Contaminant Level or MCL: The highest level
of a contaminant that is allowed in drinking water. MCLs
are set as close to the MCLGs as feasible using the best
available treatment technology.
(2)
A report for a community water system operating under a variance
or an exemption issued under Section 15 of these regulations
(excepting a variance pursuant to the requirements of Section 3
New Water Sources) must include the following definition:
Variances and Exemptions: State or EPA permission not to meet
an MCL or a treatment technique under certain conditions.
(3)
A report that contains data on contaminants that EPA regulates
using any of the following terms must include the applicable
definitions:
238
(i)
Treatment Technique: A required process intended to
reduce the level of a contaminant in drinking water.
(ii)
Action Level: The concentration of a contaminant which, if
exceeded, triggers treatment or other requirements which a
water system must follow.
(iii)
Maximum residual disinfectant level goal or MRDLG: The
level of a drinking water disinfectant below which there is
no known or expected risk to health. MRDLGs do not
reflect the benefits of the use of disinfectants to control
microbial contaminants.
(iv)
Maximum residual disinfectant level or MRDL: The
highest level of a disinfectant allowed in drinking water.
There is convincing evidence that the addition of a
disinfectant is necessary for control of microbial
contaminants.
(d)
Information on Detected Contaminants
(1)
This subsection specifies the requirements for information to be
included in each report for contaminants subject to mandatory
monitoring (except Cryptosporidium). It applies to:
(i)
Contaminants subject to a MCL, action level, maximum
residual
disinfectant
level,
or
treatment
technique
(regulated contaminants).
(ii)
Contaminants for which monitoring is required by Section
16.6 (unregulated contaminants); and
(iii)
Disinfection byproducts or microbial contaminants for
which monitoring is required by the Information Collection
Rule, 40 CFR 141.142 and 141.143 except as provided
under Paragraph (e)(1) of this Section, and which are
detected in the finished water.
(2)
The data relating to these contaminants must be displayed in one
(1) table or in several adjacent tables. Any additional monitoring
results which a community water system chooses to include in its
report must be displayed separately.
(3)
The data must be derived from data collected to comply with EPA
and State monitoring, and analytical requirements during calendar
239
year 1999 for the first report and subsequent calendar years
thereafter except that:
(i)
Where a system is allowed to monitor for regulated
contaminants less often than once a year, the table(s) must
include the date and results of the most recent sampling and
the report must include a brief statement indicating that the
datum presented in the report are from the most recent
testing done in accordance with the regulations. No data
older than 5 years need be included.
(ii)
Results of monitoring in compliance with the Information
Collection Rule, 40 CFR 141.142 and 141.143 need only be
included for 5 years from the date of last sample or until
any of the detected contaminants becomes regulated and
subject to routine monitoring requirements, whichever
comes first.
(4)
For detected regulated contaminants (listed in appendix A to this
subpart), the table(s) must contain:
(i)
The MCL for that contaminant expressed as a number
equal to or greater than 1.0 (as provided in appendix A to
this subpart);
(ii)
The MCLG for that contaminant expressed in the same
units as the MCL;
(iii)
If there is no MCL for a detected contaminant, the table
must indicate that there is a treatment technique, or specify
the action level, applicable to that contaminant, and the
report must include the definitions for treatment technique
and/or
action
level,
as
appropriate,
specified
in
Paragraph(c)(3) of this Section;
(iv)
For contaminants subject to an MCL, except turbidity and
total coliforms, the highest contaminant level used to
determine compliance with the MCL and the range of
detected levels, as follows:
(A)
When compliance with the MCL is determined
annually or less frequently: The highest detected
level at any sampling point and the range of
detected levels expressed in the same units as the
MCL.
240
(B)
When compliance with the MCL is determined by
calculating a running annual average of all samples
taken at a sampling point: the highest average of
any of the sampling points and the range of all
sampling points expressed in the same units as the
MCL.
(C)
When compliance with the MCL is determined on a
systemwide basis by calculating a running annual
average of all samples at all sampling points: the
average and range of detection expressed in the
same units as the MCL.
Note to Paragraph (d)(4)(iv): When rounding of
results to determine compliance with the MCL is
allowed by the regulations, rounding should be done
prior to multiplying the results by the factor listed in
appendix A of this subpart;
(v)
For turbidity
(A)
When it is reported pursuant to Section 16.3: The
highest average monthly value.
(B)
When it is reported pursuant to the requirements of
Section 5.2: the highest monthly value. The report
should include an explanation of the reasons for
measuring turbidity.
(C)
When it is reported pursuant to Section 5.4: The
highest single measurement and the lowest monthly
percentage of samples meeting the turbidity limits
specified in Section 5.4 for the filtration technology
being used. The report should include an
explanation of the reasons for measuring turbidity.
(vi)
For lead and copper: the 90th percentile value of the most
recent round of sampling and the number of sampling sites
exceeding the action level;
(vii)
For total coliform:
(A)
The highest monthly number of positive samples for
systems collecting fewer than 40 samples per
month; or
241
(B)
The highest monthly percentage of positive samples
for systems collecting at least 40 samples per
month;
(viii) For fecal coliform:
(A)
The total number of positive samples; and
(B)
The likely source(s) of detected contaminants to the
best
of
the
operator's
knowledge.
Specific
information
regarding
contaminants
may
be
available in sanitary surveys and source water
assessments, and should be used when available to
the operator. If the operator lacks specific
information on the likely source, the report must
include one (1) or more of the typical sources for
that contaminant listed in appendix A to this subpart
which are most applicable to the system.
(5)
If a community water system distributes water to its customers
from multiple hydraulically independent distribution systems that
are fed by different raw water sources, the table should contain a
separate column for each service area and the report should
identify each separate distribution system. Alternatively, systems
could produce separate reports tailored to include data for each
service area.
(6)
The table(s) must clearly identify any data indicating violations of
MCLs, MRDLs, or treatment techniques and the report must
contain a clear and readily understandable explanation of the
violation including: the length of the violation, the potential
adverse health effects and actions taken by the system to address
the violation. To describe the potential health effects, the system
must use the relevant language of appendix A to this subpart.
(7)
For detected unregulated contaminants for which monitoring is
required (except Cryptosporidium), the table(s) must contain the
average and range at which the contaminant was detected. The
report may include a brief explanation of the reasons for
monitoring for unregulated contaminants.
(e)
Information on Cryptosporidium, Radon, and Other Contaminants:
(1)
If the system has performed any monitoring for Cryptosporidium,
including monitoring performed to satisfy the requirements of
[RESERVED for future rulemaking 141.143 equivalent], which
242
indicates that Cryptosporidium may be present in the source water
or the finished water, the report must include:
(i)
A summary of the results of the monitoring; and
(ii)
An explanation of the significance of the results.
(2)
If the system has performed any monitoring for radon which
indicates that radon may be present in the finished water, the report
must include:
(i)
The results of the monitoring; and
(ii)
An explanation of the significance of the results.
(3)
If the system has performed additional monitoring which indicates
the presence of other contaminants in the finished water, the
system is strongly encouraged to report any results which may
indicate a health concern. To determine if results may indicate a
health concern, it is recommended that systems find out if EPA has
proposed an NPDWR or issued a health advisory for that
contaminant by calling the Safe Drinking Water Hotline (800-426-
4791). Detects above a proposed MCL or health advisory level are
considered to indicate possible health concerns. For such
contaminants, it is recommended that the report include:
(i)
The results of the monitoring; and
(ii)
An explanation of the significance of the results noting the
existence of a health advisory or a proposed regulation.
(f)
Compliance with NPDWR
In addition to the requirements of Section 16.10(3)(d), the report must
note any violation that occurred during the year covered by the report of a
requirement listed below, and include a clear and readily understandable
explanation of the violation, any potential adverse health effects, and the
steps the system has taken to correct the violation.
(1)
Monitoring and reporting of compliance data;
(2)
Filtration and disinfection prescribed by Section 5 of these
regulations. For systems which have failed to install adequate
filtration or disinfection equipment or processes, or have had a
failure of such equipment or processes which constitutes a
violation, the report must include the following language as part of
243
the explanation of potential adverse health effects: Inadequately
treated water may contain disease causing organisms. These
organisms include bacteria, viruses and parasites which can cause
symptoms such as nausea, cramps, diarrhea and associated
headaches.
(3)
Lead and copper control requirements prescribed by Section 6 of
these regulations. For systems which fail to take one (1) or more
actions prescribed by sections 6.80(d), 6.81, 6.82, 6.83, or 6.84
herein, the report must include the applicable language of appendix
A to this subpart for lead, copper, or both.
(4)
Treatment techniques for Acrylamide and Epichlorohydrin
prescribed by Section 16.2(d) of these regulations. For systems
which violate the requirements of Section 16.2(d) herein, the report
must include the relevant language from appendix A to this
subpart.
(5)
Recordkeeping of Compliance Data
(6)
Special monitoring requirements prescribed by sections 16.6, 16.7
and 16.1(r); and
(7)
Violation of the terms of a variance, an exemption, or an
administrative or judicial order.
(g)
Variances and Exemptions
If a system is operating under the terms of a variance or an exemption
issued under Sec. 1415 or 1416 of SDWA, the report must contain:
(1)
An explanation of the reasons for the variance or exemption;
(2)
The date on which the variance or exemption was issued;
(3)
A brief status report on the steps the system is taking to install
treatment, find alternative sources of water, or otherwise comply
with the terms and schedules of the variance or exemption; and
(4)
A notice of any opportunity for public input in the review, or
renewal, of the variance or exemption.
(h)
Additional Information
244
(1)
The report must contain a brief explanation regarding contaminants
which may reasonably be expected to be found in drinking water
including bottled water. This explanation may include the language
of Paragraphs (h)(1)(i) through (iii) or systems may use their own
comparable language. The report also must include the language of
Paragraph (h)(1)(iv) of this Section.
(i)
The sources of drinking water (both tap water and bottled
water) include rivers, lakes, streams, ponds, reservoirs,
springs and wells. As water travels over the surface of the
land or through the ground, it dissolves naturally occurring
minerals and, in some cases, radioactive material, and can
pick up substances resulting from the presence of animals
or from human activity.
(ii)
Contaminants that may be present in source water include:
(A)
Microbial contaminants, such as viruses and
bacteria, which may come from sewage treatment
plants,
septic
systems,
agricultural
livestock
operations, and wildlife.
(B)
Inorganic contaminants, such as salts and metals,
which can be naturally occurring or result from
urban storm water runoff, industrial or domestic
wastewater discharges, oil and gas production,
mining, or farming.
(C)
Pesticides and herbicides, which may come from a
variety of sources such as agriculture, urban storm
water runoff and residential uses.
(D)
Organic chemical contaminants, including synthetic
and
volatile
organic
chemicals,
which
are
byproducts of industrial processes and petroleum
production, and can also come from gas stations,
urban storm water runoff and septic systems.
(E)
Radioactive contaminants, which can be naturally-
occurring or be the result of oil and gas production
and mining activities.
(iii)
In order to ensure that tap water is safe to drink, EPA
prescribes regulations which limit the amount of certain
contaminants in water provided by PWSs. FDA regulations
245
establish limits for contaminants in bottled water which
must provide the same protection for public health.
(iv)
Drinking water, including bottled water, may reasonably be
expected to contain at least small amounts of some
contaminants. The presence of contaminants does not
necessarily indicate that water poses a health risk. More
information about contaminants and potential health effects
can be obtained by calling the Environmental Protection
Agency's Safe Drinking Water Hotline (800-426-4791).
(2)
The report must include the telephone number of the owner,
operator, or designee of the community water system as a source of
additional information concerning the report.
(3)
In communities with a large proportion of non-English speaking
residents, as determined by the Director, the report must contain
information in the appropriate language(s) regarding the
importance of the report or contain a telephone number or address
where such residents may contact the system to obtain a translated
copy of the report or assistance in the appropriate language.
(4)
The report must include information (e.g., time and place of
regularly scheduled board meetings) about opportunities for public
participation in decisions that may affect the quality of the water.
(5)
The systems may include such additional information as they deem
necessary for public education consistent with, and not detracting
from, the purpose of the report.
(4)
Required Additional Health Information
(a)
All reports must prominently display the following language:
Some people may be more vulnerable to contaminants in drinking water
than the general population. Immuno-compromised persons such as
persons with cancer undergoing chemotherapy, persons who have
undergone organ transplants, people with HIV/AIDS or other immune
system disorders, some elderly, and infants can be particularly at risk from
infections. These people should seek advice about drinking water from
their health care providers. EPA/CDC guidelines on appropriate means to
lessen the risk of infection by Cryptosporidium and other microbial
contaminants are available from the Safe Drinking Water Hotline
(800-426-4791).
246
(b)
Ending in the report due by July 1, 2001, a system which detects arsenic at
levels above 0.025 mg/L, but below the 0.05 mg/L, and beginning in the
report due by July 1, 2002, a system that detects arsenic above 0.005 mg/L
and up to and including 0.010 mg/L;
(1)
Must include in its report a short informational statement about
arsenic, using language such as: While your drinking water meets
EPA’s standard for arsenic, it does contain low levels of arsenic.
EPA’s standard balances the current understanding of arsenic’s
possible health effects against the costs of removing arsenic from
drinking water. EPA continues to research the health effects of low
levels of arsenic, which is a mineral known to cause cancer in
humans at high concentrations and is linked to other health effects
such as skin damage and circulatory problems.
(2)
May write its own educational statement, but only in consultation
with the Director.
(c)
A system which detects nitrate at levels above 5 mg/L but below the
MCL:
(1)
Must include a short informational statement about the impacts of
nitrate on children using language such as: Nitrate in drinking
water at levels above 10 ppm is a health risk for infants of less than
six (6) months of age. High nitrate levels in drinking water can
cause blue baby syndrome. Nitrate levels may rise quickly for
short periods of time because of rainfall or agricultural activity. If
you are caring for an infant you should ask advice from your health
care provider.
(2)
May write its own educational statement, but only in consultation
with the Director.
(d)
Systems which detect lead above the action level in more than 5%, and up
to and including 10%, of homes sampled:
(1)
Must include a short informational statement about the special
impact of lead on children using language such as: Infants and
young children are typically more vulnerable to lead in drinking
water than the general population. It is possible that lead levels at
your home may be higher than at other homes in the community as
a result of materials used in your home's plumbing. If you are
concerned about elevated lead levels in your home's water, you
may wish to have your water tested and flush your tap for 30
seconds to 2 minutes before using tap water. Additional
247
information is available from the Safe Drinking Water Hotline
(800-426-4791).
(2)
May write its own educational statement, but only in consultation
with the Director.
(e)
Community water systems that detect TTHM above 0.080 mg/L, but
below the MCL in Paragraph 16.2 (a), as an annual average, monitored
and calculated under the provisions of Paragraph 16.2 (a)(49) must include
health effects language for TTHMs prescribed by Appendix A to Section
16.10.
(f)
Beginning in the report due by July 1, 2003 and ending January 22, 2006,
a community water system that detects arsenic above 0.010 mg/L and up
to and including 0.05 mg/L must include the arsenic health effects
language prescribed by Appendix B to Section 16.8.
(5)
Report Delivery and Recordkeeping
(a)
Except as provided in Paragraph (g) of this Section, each community
water system must mail or otherwise directly deliver one (1) copy of the
report to each customer.
(b)
The system must make a good faith effort to reach consumers who do not
get water bills, using means recommended by the Director. It is expected
that an adequate good faith effort will be tailored to the consumers who
are served by the system but are not bill-paying customers, such as renters
or workers. A good faith effort to reach consumers would include a mix of
methods appropriate to the particular system such as: Posting the reports
on the Internet; mailing to postal patrons in metropolitan areas; advertising
the availability of the report in the news media; publication in a local
newspaper; posting in public places such as cafeterias or lunch rooms of
public buildings; delivery of multiple copies for distribution by
single-biller customers such as apartment buildings or large private
employers; delivery to community organizations.
(c)
No later than the date the system is required to distribute the report to its
customers, each community water system must mail a copy of the report
to the Director, followed within 3 months by a certification that the report
has been distributed to customers, and that the information is correct and
consistent with the compliance monitoring data previously submitted to
the Director.
(d)
No later than the date the system is required to distribute the report to its
customers, each community water system must deliver the report to any
other agency or clearinghouse identified by the Director.
248
(e)
Each community water system must make its reports available to the
public upon request.
(f)
Each community water system serving 100,000 or more persons must post
its current year's report to a publicly-accessible web site on the Internet.
(g)
Any system subject to this Section must retain copies of its consumer
confidence report for no less than three (3) years.
(h)
Special Delivery Requirement for Community Water Systems Serving a
Population of 10,000 or More
Any community water system serving a population of 10,000 or more
shall directly deliver a full copy of the Consumer Confidence Report to
each household within the water system's service area that receives water
from that system. The method of delivery shall be determined by the water
system but can include delivery via either: (a) postal patron mailing; or (b)
a community newsletter that is directly delivered to each household; or (c)
a community calendar that is directly delivered to each household or (d)
any other method that will directly reach each household within the water
system's service area that receives water from that system. In the event
that within the service area there are buildings with 5 or more residential
units, the system will not be required to deliver directly to each of these
units. Instead, the water system shall mail multiple copies of the report to
building manager or other appropriate individual, noting that the reports
should be distributed to residents and/or posted in a common area.
Additionally, college and universities will be exempted from this
Paragraph, {16.10 (5)(h)}.
Appendix A to Section 16.10.–Regulated Contaminants
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
249
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
Microbiologi
cal
contaminants:
Total
Coliform
Bacteria
MCL: (systems
that collect >
40
samples/month)
5% of monthly
samples are
positive;
(systems that
collect < 40
samples/month)
1 positive
monthly
sample.
MCL: (systems
>40 samples
/month) 5% of
monthly samples
are positive;
(systems that
collect <40
samples/ month)
1 positive
monthly sample
0
Naturally
present in the
environment.
Coliforms are
bacteria that are
naturally
present in the
environment
and are used as
an indicator
that other,
potentially-
harmful,
bacteria may be
present.
Coliforms were
found in more
samples than
allowed and
this was a
warning of
potential
problems.
Fecal
coliform and
E. Coli
0
0
0
Human and
animal fecal
waste.
Fecal coliforms
and E. Coli are
bacteria whose
presence
indicates that
the water may
be
contaminated
with human or
animal wastes.
Microbes in
these wastes
can cause
short-term
effects, such as
diarrhea,
cramps, nausea,
headaches, or
other
symptoms.
They may pose
a special health
risk for infants,
young children,
some of the
elderly, and
people with
severely-
compromised
immune
systems.
Total organic
carbon (ppm)
TT
TT
N/A
Naturally
present in the
Total organic
carbon (TOC)
250
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
environment.
has no health
effects.
However, total
organic carbon
provides a
medium for the
formation of
disinfection by
products. These
byproducts
include
trihalomethanes
(THMs) and
haloacetic acids
(HAAs).
Drinking water
containing
these
byproducts in
excess of the
MCL may lead
to adverse
health effects,
liver or kidney
problems, or
nervous system
effects, and
may lead to an
increased risk
of getting
cancer.
Turbidity
(NTU)
TT
TT
N/A
Soil runoff.
Turbidity has
no health
effects.
However,
turbidity can
interfere with
disinfection and
provide a
medium for
microbial
growth.
Turbidity may
indicate the
presence of
disease-causing
organisms.
These
organisms
include
bacteria,
viruses, and
parasites that
251
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
can cause
symptoms such
as nausea,
cramps,
diarrhea and
associated
headaches.
Radioactive
contaminants:
Beta/photon
emitters
(mrem/ yr)
4 mrem/yr
4.
0
Decay of
natural and
man-made
deposits.
Certain
minerals are
radioactive and
may emit forms
of radiation
known as
photons and
beta radiation.
Some people
who drink
water
containing beta
and photon
radioactivity in
excess of the
MCL over
many years
may have an
increased risk
of getting
cancer.
Alpha
emitters
(pCi/l)
15 pCi/l
15
0
Erosion of
natural
deposits.
Certain
minerals are
radioactive and
may emit a
form of
radiation
known as alpha
radiation. Some
people who
drink water
containing
alpha emitters
in excess of the
MCL over
many years
may have an
increased risk
of getting
cancer.
Combined
radium
(pCi/l)
5 pCi/l
5
0
Erosion of
natural
deposits.
Some people
who drink
water
containing
radium 226 or
228 in excess
252
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
of the MCL
over many
years may have
an increased
risk of getting
cancer.
Uranium
(ug/L)
0.030
1000
30
0
Erosion of
natural
deposits.
Some people
who drink
water
containing
uranium in
excess of the
MCL over
many years
may have an
increased risk
of getting
cancer and
kidney toxicity.
Inorganic
contaminants:
Antimony
(ppb)
0.006
1000
6
6
Discharge
from
petroleum
refineries; fire
retardants;
ceramics;
electronics;
solder.
Some people
who drink
water
containing
antimony well
in excess of the
MCL over
many years
could
experience
increases in
blood
cholesterol and
decreases in
blood sugar.
Arsenic (ppb)
1 0.010
1000
101
01
Erosion of
natural
deposits;
Runoff from
orchards;
Runoff from
glass and
electronics
production
wastes.
Some people
who drink
water
containing
arsenic in
excess of the
MCL over
many years
could
experience skin
damage or
problems with
their circulatory
system, and
may have an
increased risk
of getting
253
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
cancer.
Asbestos
(MFL)
7 MFL
7
7
Decay of
asbestos
cement water
mains; Erosion
of natural
deposits.
Some people
who drink
water
containing
asbestos in
excess of the
MCL over
many years
may have an
increased risk
of developing
benign
intestinal
polyps.
Barium
(ppm)
2
2
2
Discharge of
drilling
wastes;
Discharge
from metal
refineries;
Erosion of
natural
deposits.
Some people
who drink
water
containing
barium in
excess of the
MCL over
many years
could
experience an
increase in their
blood pressure.
Beryllium
(ppb)
0.004
1000
4
4
Discharge
from metal
refineries and
coal- burning
factories;
Discharge
from
electrical,
aerospace, and
defense
industries.
Some people
who drink
water
containing
beryllium well
in excess of the
MCL over
many years
could develop
intestinal
lesions.
Cadmium
(ppb)
0.005
1000
5
5
Corrosion of
galvanized
pipes; Erosion
of natural
deposits;
Discharge
from metal
refineries;
Runoff from
waste batteries
and paints.
Some people
who drink
water
containing
cadmium in
excess of the
MCL over
many years
could
experience
kidney damage.
Chromium
(ppb)
0.1
1000
100
100
Discharge
from steel and
pulp mills;
Erosion of
Some people
who use water
containing
chromium well
254
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
natural
deposits.
in excess of the
MCL over
many years
could
experience
allergic
dermatitis.
Copper (ppm)
AL=1.3
AL=1.3
1.3
Corrosion of
household
plumbing
systems;
Erosion of
natural
deposits;
Leaching from
wood
preservatives.
Copper is an
essential
nutrient, but
some people
who drink
water
containing
copper in
excess of the
action level
over a
relatively short
amount of time
could
experience
gastrointestinal
distress. Some
people who
drink water
containing
copper in
excess of the
action level
over many
years could
suffer liver or
kidney damage.
People with
Wilson's
Disease should
consult their
personal
doctor.
Cyanide
(ppb)
0.2
1000
200
200
Discharge
from
steel/metal
factories
Discharge
from plastic
and fertilizer
factories.
Some people
who drink
water
containing
cyanide well in
excess of the
MCL over
many years
could
experience
nerve damage
or problems
with their
255
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
thyroid.
Fluoride
(ppm)
4.
4.
4.
Erosion of
natural
deposits;
Water additive
which
promotes
strong teeth;
Discharge
from fertilizer
and aluminum
factories.
Some people
who drink
water
containing
fluoride in
excess of the
MCL over
many years
could get bone
disease,
including pain
and tenderness
of the bones.
Fluoride in
drinking water
at half the MCL
or more may
cause mottling
of children's
teeth, usually in
children less
than nine (9)
years old.
Mottling, also
known as
dental
fluorosis, may
include brown
staining and/or
pitting of the
teeth, and
occurs only in
developing
teeth before
they erupt from
the gums.
Lead (ppb)
AL=0.015
1000
AL=15
0
Corrosion of
household
plumbing
systems;
Erosion of
natural
deposits.
Infants and
children who
drink water
containing lead
in excess of the
action level
could
experience
delays in their
physical or
mental
development.
Children could
show slight
256
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
deficits in
attention span
and learning
abilities. Adults
who drink this
water over
many years
could develop
kidney
problems or
high blood
pressure.
Mercury
[inorganic]
(ppb)
0.002
1000
2
2
Erosion of
natural
deposits;
Discharge
from refineries
and factories;
Runoff from
landfills;
Runoff from
cropland.
Some people
who drink
water
containing
inorganic
mercury well in
excess of the
MCL over
many years
could
experience
kidney damage.
Nitrate (ppm)
10
10
10
Runoff from
fertilizer use;
Leaching from
septic tanks,
sewage;
Erosion of
natural
deposits.
Infants below
the age of six
(6) months who
drink water
containing
nitrate in
excess of the
MCL could
become
seriously ill
and, if
untreated, may
die. Symptoms
include
shortness of
breath and blue
baby syndrome.
Nitrite (ppm)
1
1
1
Runoff from
fertilizer use;
Leaching from
septic tanks,
sewage;
Erosion of
natural
deposits.
Infants below
the age of six
(6) months who
drink water
containing
nitrite in excess
of the MCL
could become
seriously ill
and, if
untreated, may
die. Symptoms
257
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
include
shortness of
breath and blue
baby syndrome.
Selenium
(ppb)
0.05
1000
50
50
Discharge
from
petroleum and
metal
refineries;
Erosion of
natural
deposits;
Discharge
from mines.
Selenium is an
essential
nutrient.
However, some
people who
drink water
containing
selenium in
excess of the
MCL over
many years
could
experience hair
or fingernail
losses,
numbness in
fingers or toes,
or problems
with their
circulation.
Thallium
(ppb)
0.002
1000
2
0.5
Leaching from
ore-processing
sites;
Discharge
from
electronics,
glass, and drug
factories.
Some people
who drink
water
containing
thallium in
excess of the
MCL over
many years
could
experience hair
loss, changes in
their blood, or
problems with
their kidneys,
intestines, or
liver.
Synthetic
organic
contaminants
including
pesticides and
herbicides:
2,4-D (ppb)
0.07
1000
70
70
Runoff from
herbicide used
on row crops.
Some people
who drink
water
containing the
weed killer 2,4-
D well in
excess of the
MCL over
many years
could
experience
problems with
their kidneys,
258
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
liver, or adrenal
glands.
2,4,5-TP
[Silvex](ppb)
0.05
1000
50
50
Residue of
banned
herbicide.
Some people
who drink
water
containing
silvex in excess
of the MCL
over many
years could
experience liver
problems.
Acrylamide
TT
TT
0
Added to
water during
sewage/waste
water
treatment.
Some people
who drink
water
containing high
levels of
acrylamide
over a long
period of time
could have
problems with
their nervous
system or
blood, and may
have an
increased risk
of getting
cancer.
Alachlor
(ppb)
0.002
1000
2
0
Runoff from
herbicide used
on row crops.
Some people
who drink
water
containing
alachlor in
excess of the
MCL over
many years
could have
problems with
their eyes,
liver, kidneys,
or spleen, or
experience
anemia, and
may have an
increased risk
of getting
cancer.
Atrazine
(ppb)
0.003
1000
3
3
Runoff from
herbicide used
on row crops
Some people
who drink
water
containing
atrazine well in
259
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
excess of the
MCL over
many years
could
experience
problems with
their
cardiovascular
system or
reproductive
difficulties.
Benzo(a)pyre
ne [PAH]
(nanograms/l)
0.0002
1,000,000
200
0
Leaching from
linings of
water storage
tanks and
distribution
lines.
Some people
who drink
water
containing
benzo(a)pyrene
in excess of the
MCL over
many years
may experience
reproductive
difficulties and
may have an
increased risk
of getting
cancer.
Carbofuran
(ppb)
0.04
1000
40
40
Leaching of
soil fumigant
used on rice
and alfalfa.
Some people
who drink
water
containing
carbofuran in
excess of the
MCL over
many years
could
experience
problems with
their blood, or
nervous or
reproductive
systems.
Chlordane
(ppb)
0.002
1000
2
0
Residue of
banned
termiticide.
Some people
who drink
water
containing
chlordane in
excess of the
MCL over
many years
could
experience
problems with
their liver or
260
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
nervous
system, and
may have an
increased risk
of getting
cancer.
Dalapon
(ppb)
0.2
1000
200
200
Runoff from
herbicide used
on rights of
way.
Some people
who drink
water
containing
dalapon well in
excess of the
MCL over
many years
could
experience
minor kidney
changes.
Di(2-
ethylhexyl)
adipate (ppb)
0.4
1000
400
400
Discharge
from chemical
factories.
Some people
who drink
water
containing di
(2-ethylhexyl)
adipate well in
excess of the
MCL over
many years
could
experience
general toxic
effects or
reproductive
difficulties.
Di(2-
ethylhexyl)
phthalate
(ppb).
0.006
1000
6
0
Discharge
from rubber
and chemical
factories.
Some people
who drink
water
containing di
(2-ethylhexyl)
phthalate in
excess of the
MCL over
many years
may have
problems with
their liver, or
experience
reproductive
difficulties, and
may have an
increased risk
of getting
cancer.
261
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
Dibromochlor
opropane
(ppt)
0.0002
1,000,000
200
0
Runoff/leachin
g from soil
fumigant used
on soybeans,
cotton,
pineapples,
and orchards.
Some people
who drink
water
containing
DBCP in
excess of the
MCL over
many years
could
experience
reproductive
problems and
may have an
increased risk
of getting
cancer.
Dinoseb
(ppb)
0.007
1000
7
7
Runoff from
herbicide used
on soybeans
and
vegetables.
Some people
who drink
water
containing
dinoseb well in
excess of the
MCL over
many years
could
experience
reproductive
difficulties.
Diquat (ppb)
0.02
1000
20
20
Runoff from
herbicide use.
Some people
who drink
water
containing
diquat in excess
of the MCL
over many
years could get
cataracts.
Dioxin
[2,3,7,8-
TCDD]
(ppq).
0.00000003
1,000,000, 000
30
0
Emissions
from waste
incineration
and other
combustion;
Discharge
from chemical
factories.
Some people
who drink
water
containing
dioxin in
excess of the
MCL over
many years
could
experience
reproductive
difficulties and
may have an
increased risk
of getting
cancer.
262
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
Endothall
(ppb)
0.1
1000
100
100
Runoff from
herbicide use.
Some people
who drink
water
containing
endothall in
excess of the
MCL over
many years
could
experience
problems with
their stomach
or intestines.
Endrin (ppb)
0.002
1000
2
2
Residue of
banned
insecticide.
Some people
who drink
water
containing
endrin in
excess of the
MCL over
many years
could
experience liver
problems.
Epichlorohyd
rin
TT
TT
0
Discharge
from industrial
chemical
factories; An
impurity of
some water
treatment
chemicals.
Some people
who drink
water
containing high
levels of
epichlorohydrin
over a long
period of time
could
experience
stomach
problems, and
may have an
increased risk
of getting
cancer.
Ethylene
dibromide
(ppt)
0.00005
1,000,000
50
0
Discharge
from
petroleum
refineries.
Some people
who drink
water
containing
ethylene
dibromide in
excess of the
MCL over
many years
could
experience
problems with
their liver,
263
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
stomach,
reproductive
system, or
kidneys, and
may have an
increased risk
of getting
cancer.
Glyphosate
(ppb)
0.7
1000
700
700
Runoff from
herbicide use
Some people
who drink
water
containing
glyphosate in
excess of the
MCL over
many years
could
experience
problems with
their kidneys or
reproductive
difficulties.
Heptachlor
(ppt)
0.0004
1,000,000
400
0
Residue of
banned
pesticide.
Some people
who drink
water
containing
heptachlor in
excess of the
MCL over
many years
could
experience liver
damage and
may have an
increased risk
of getting
cancer.
Heptachlor
epoxide (ppt)
0.0002
1,000,000
200
0
Breakdown of
heptachlor.
Some people
who drink
water
containing
heptachlor
epoxide in
excess of the
MCL over
many years
could
experience liver
damage, and
may have an
increased risk
of getting
cancer.
264
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
Hexachlorobe
nzene (ppb)
0.001
1000
1
0
Discharge
from metal
refineries and
agricultural
chemical
factories.
Some people
who drink
water
containing
hexachlorobenz
ene in excess of
the MCL over
many years
could
experience
problems with
their liver or
kidneys, or
adverse
reproductive
effects, and
may have an
increased risk
of getting
cancer.
Hexachlorocy
clopentadiene
(ppb)
0.05
1000
50
50
Discharge
from chemical
factories.
Some people
who drink
water
containing
hexachlorocycl
opentadiene
well in excess
of the MCL
over many
years could
experience
problems with
their kidneys
or stomach.
Lindane (ppt)
0.0002
1,000,000
200
200
Runoff/
leaching from
insecticide
used on cattle,
lumber,
gardens.
Some people
who drink
water
containing
lindane in
excess of the
MCL over
many years
could
experience
problems with
their kidneys or
liver.
Methoxychlor
(ppb)
0.04
1000
40
40
Runoff/
leaching from
insecticide
used on fruits,
vegetables,
alfalfa,
Some people
who drink
water
containing
methoxychlor
in excess of the
265
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
livestock.
MCL over
many years
could
experience
reproductive
difficulties.
Oxamyl
[Vydate]
(ppb)
0.2
1000
200
200
Runoff/leachin
g from
insecticide
used on
apples,
potatoes and
tomatoes.
Some people
who drink
water
containing
oxamyl in
excess of the
MCL over
many years
could
experience
slight nervous
system effects.
PCBs
[Polychlorina
ted
biphenyls]
(ppt)
0.0005
1,000,000
500
0
Runoff from
landfills;
Discharge of
waste
chemicals.
Some people
who drink
water
containing
PCBs in excess
of the MCL
over many
years could
experience
changes in their
skin, problems
with their
thymus gland,
immune
deficiencies, or
reproductive or
nervous system
difficulties, and
may have an
increased risk
of getting
cancer.
Pentachlorop
henol (ppb)
0.001
1000
1
0
Discharge
from wood
preserving
factories.
Some people
who drink
water
containing
pentachlorophe
nol in excess of
the MCL over
many years
could
experience
problems with
their liver or
kidneys, and
266
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
may have an
increased risk
of getting
cancer.
Picloram
(ppb)
0.5
1000
500
500
Herbicide
runoff
Some people
who drink
water
containing
picloram in
excess of the
MCL over
many years
could
experience
problems with
their liver.
Simazine
(ppb)
0.004
1000
4
4
Herbicide
runoff
Some people
who drink
water
containing
simazine in
excess of the
MCL over
many years
could
experience
problems with
their blood
Toxaphene
(ppb)
0.003
1000
3
0
Runoff/leachin
g from
insecticide
used on cotton
and cattle.
Some people
who drink
water
containing
toxaphene in
excess of the
MCL over
many years
could have
problems with
their kidneys,
liver, or
thyroid, and
may have an
increased risk
of getting
cancer.
Volatile
organic
contaminants:
Benzene
(ppb)
0.005
1000
5
0
Discharge
from factories;
Leaching from
gas storage
tanks and
landfills
Some people
who drink
water
containing
benzene in
excess of the
MCL over
many years
267
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
could
experience
anemia or a
decrease in
blood platelets,
and may have
an increased
risk of getting
cancer.
Bromate
(ppb)
0.010
1000
10
0
By-product of
drinking water
chlorination.
Some people
who drink
water
containing
bromate in
excess of the
MCL over
many years
may have an
increased risk
of getting
cancer.
Carbon
tetrachloride
(ppb)
0.005
1000
5
0
Discharge
from chemical
plants and
other
industrial
activities.
Some people
who drink
water
containing
carbon
tetrachloride in
excess of the
MCL over
many years
could
experience
problems with
their liver and
may have an
increased risk
of getting
cancer.
Chloramines
(ppm)
MRDL = 4
MRDL = 4
MRD
LG =
4
Water additive
used to control
microbes.
Some people
who use water
containing
chloramines
well in excess
of the MRDL
could
experience
irritating effects
to their eyes
and nose. Some
who drink
water
containing
chloramines
268
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
well in excess
of the MRDL
could
experience
stomach
discomfort or
anemia.
Chlorine
(ppm)
MRDL = 4
MRDL = 4
MRD
LG =
4
Water additive
used to control
microbes.
Some people
who use water
containing
chlorine well in
excess of the
MRDL could
experience
irritating effects
to their eyes
and nose. Some
people who
drink water
containing
chlorine well in
excess of the
MRDL could
experience
stomach
discomfort.
Chlorite
(ppm)
1
1
0.8
By-product of
drinking water
chlorination.
Some infants
and young
children who
drink water
containing
chlorite in
excess of the
MCL could
experience
nervous system
effects. Similar
effects may
occur in fetuses
of pregnant
women who
drink water
containing
chlorite in
excess of the
MCL. Some
people may
experience
anemia.
Chloride
dioxide (ppb)
MRDL = .8
1000
MRDL = 800
MRD
LG =
800
Water additive
used to control
microbes.
Some infants
and young
children who
drink water
269
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
containing
chlorine
dioxide in
excess of the
MRDL could
experience
nervous system
effects. Similar
effects may
occur in fetuses
of pregnant
women who
drink water
containing
chlorine
dioxide in
excess of the
MRDL. Some
people may
experience
anemia.
Chlorobenzen
e (ppb)
.1
1000
100
100
Discharge
from chemical
and
agricultural
chemical
factories.
Some people
who drink
water
containing
chlorobenzene
in excess of the
MCL over
many years
could
experience
problems with
their liver or
kidneys.
o-
Dichlorobenz
ene (ppb)
0.6
1000
600
600
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing o-
dichlorobenzen
e well in excess
of the MCL
over many
years could
experience
problems with
their liver,
kidneys, or
circulatory
systems.
p-
Dichlorobenz
ene (ppb)
0.075
1000
75
75
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing p-
270
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
dichlorobenzen
e in excess of
the MCL over
many years
could
experience
anemia,
damage to their
liver, kidneys,
or spleen, or
changes in their
blood.
1,2-
Dichloroetha
ne (ppb)
0.005
1000
5
0
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing 1,2-
dichloroethane
in excess of the
MCL over
many years
may have an
increased risk
of getting
cancer.
1,1-
Dichloroethyl
ene (ppb)
0.007
1000
7
7
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing 1,1-
dichloroethylen
e in excess of
the MCL over
many years
could
experience
problems with
their liver.
cis-1,2-
Dichloroethyl
ene (ppb)
0.07
1000
70
70
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing cis-
1,2-
dichloroethylen
e in excess of
the MCL over
many years
could
experience
problems with
their liver.
trans-1,2-
Dichloroethyl
ene
(ppb)
0.1
1000
100
100
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing
271
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
trans-1,2-
dichloroethylen
e well in excess
of the MCL
over many
years could
experience
problems with
their liver.
Dichlorometh
ane (ppb)
0.005
1000
5
0
Discharge
from
pharmaceutica
l and chemical
factories.
Some people
who drink
water
containing
dichloromethan
e in excess of
the MCL over
many years
could have
liver problems
and may have
an increased
risk of getting
cancer.
1,2-
Dichloroprop
ane (ppb)
0.005
1000
5
0
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing 1,2-
dichloropropan
e in excess of
the MCL over
many years
may have an
increased risk
of getting
cancer.
Ethylbenzene
(ppb)
0.7
1000
700
700
Discharge
from
petroleum
refineries.
Some people
who drink
water
containing
ethylbenzene
well in excess
of the MCL
over many
years could
experience
problems with
their liver or
kidneys.
Haloacetic
Acids (HAA)
(ppb).
0.060
1000
60
N/A
By-product of
drinking water
disinfection.
Some people
who drink
water
containing
haloacetic acids
272
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
in excess of the
MCL over
many years
may have an
increased risk
of getting
cancer.
Styrene (ppb)
.1
1000
100
100
Discharge
from rubber
and plastic
factories;
Leaching from
landfills.
Some people
who drink
water
containing
styrene well in
excess of the
MCL over
many years
could have
problems with
their liver,
kidneys, or
circulatory
system.
Tetrachloroet
hylene (ppb)
0.005
1000
5
0
Discharge
from factories
and dry
cleaners.
Some people
who drink
water
containing
tetrachloroethyl
ene in excess of
the MCL over
many years
could have
problems with
their liver, and
may have an
increased risk
of getting
cancer.
1,2,4-
Trichlorobenz
ene (ppb)
0.07
1000
70
70
Discharge
from textile-
finishing
factories.
Some people
who drink
water
containing
1,2,4-
trichlorobenzen
e well in excess
of the MCL
over many
years could
experience
changes in their
adrenal glands.
1,1,1-
Trichloroetha
ne (ppb)
0.2
1000
200
200
Discharge
from metal
degreasing
sites and other
Some people
who drink
water
containing
273
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
factories.
1,1,1-
trichloroethane
in excess of the
MCL over
many years
could
experience
problems with
their liver,
nervous
system, or
circulatory
system.
1,1,2-
Trichloroetha
ne (ppb)
0.005
1000
5
3
Discharge
from industrial
chemical
factories.
Some people
who drink
water
containing
1,1,2-
trichloroethane
well in excess
of the MCL
over many
years could
have problems
with their liver,
kidneys, or
immune
systems.
Trichloroethy
lene (ppb)
0.005
1000
5
0
Discharge
from metal
degreasing
sites and other
factories.
Some people
who drink
water
containing
trichloroethylen
e in excess of
the MCL over
many years
could
experience
problems with
their liver and
may have an
increased risk
of getting
cancer.
TTHMs
[Total
trihalomethan
es] (ppb)
0.10/0.080
1000
100/80
N/A
By-product of
drinking water
chlorination.
Some people
who drink
water
containing
trihalomethanes
in excess of the
MCL over
many years
may experience
274
Contaminant
(units)
Traditional
MCL in mg/L
To convert for
CCR, multiply
by
MCL in CCR
units
MCL
G
Major sources
in drinking
water
Health effects
language
problems with
their liver,
kidneys, or
central nervous
systems, and
may have an
increased risk
of getting
cancer.
Toluene
(ppm)
1
1
1
Discharge
from
petroleum
factories.
Some people
who drink
water
containing
toluene well in
excess of the
MCL over
many years
could have
problems with
their nervous
system,
kidneys, or
liver.
Vinyl
Chloride
(ppb)
0.002
1000
2
0
Leaching from
PVC piping;
Discharge
from plastics
factories.
Some people
who drink
water
containing
vinyl chloride
in excess of the
MCL over
many years
may have an
increased risk
of getting
cancer.
Xylenes
(ppm)
10
10
10
Discharge
from
petroleum
factories;
Discharge
from chemical
factories.
Some people
who drink
water
containing
xylenes in
excess of the
MCL over
many years
could
experience
damage to their
nervous
system.
1 These arsenic values (MCL, MCLG) are effective January 23, 2006. Until then, the MCL is 0.05 mg/L and there is
no MCLG.
275
Key:
AL= Action Level
MCL= Maximum Contaminant Level
MCLG= Maximum Contaminant Level Goal
MFL= million fibers per liter
MRDL= Maximum Residual Disinfectant Level
MRDLG= Maximum Residual Disinfectant Level Goal
mrem/year= millirems per year (a measure of radiation absorbed by the body)
N/A= Not Applicable
NTU= Nephelometric Turbidity Units (a measure of water clarity)
pCi/l= picocuries per liter (a measure of radioactivity)
ppm= parts per million, or milligrams per liter (mg/l)
ppb= parts per billion, or micrograms per liter (µg/l)
ppt= parts per trillion, or nanograms per liter
ppq= parts per quadrillion, or picograms per liter
TT= Treatment Technique
276
SECTION 17.0 NON-COMMUNITY WATER SYSTEM REQUIREMENTS
17.1 Microbiological
a)
Routine monitoring: PWSs must collect total coliform samples at sites which are
representative of water throughout the distribution system according to a written
sample siting plan. At least one (1) representative sample shall be collected each
calendar quarter when the system is in operation. These plans are subject to
review and revision by the Director.
Monitoring Frequency
For total coliforms for non-community water systems is as follows:
i)
A non-community water system using only ground water and serving
1,000 persons or fewer must monitor each calendar quarter that the system
provides water to the public.
ii)
A non-community water systems using only ground water, and serving
more than 1000 persons during any month must monitor at the same
frequency as a like-sized community water system, as specified in Section
16.4 a of these regulations.
iii)
A non-community water system using surface water in total or in part,
must monitor at the same frequency as a like-sized community water
system as specified in Section 16.4 (a) of these regulations.
iv)
A non-community water system using ground water under the direct
influence of surface water, as determined by the Director, in total or in
part must monitor at the same frequency as a like-sized community water
system, as specified in Section 16.4 (a) of these regulations, within 6
months of said determination by the Director.
b)
The following requirements for PWSs found in Section 16.4 also apply to non-
community water systems. This includes Sections:
16.4(a)(2) and (3)
Routine Monitoring;
16.4(b)
Analytical Methodology;
16.4(c)
Maximum
Contaminant
Levels
for
Microbiological
Contaminants;
16.4(d)
Repeat Monitoring;
16.4(e)
Fecal Coliforms/E. Coli testing;
16.4(f)
Invalidation of Samples;
16.4(g)
Sanitary Surveys
16.4(h)
Reporting Requirements
277
17.2
Inorganic Chemicals
Non-transient non-community water systems shall be required to comply with the requirements
of Sections 6 and 16.1 with the following exceptions. (1) Monitoring and compliance with the
requirements for sodium shall not be required. (2) Monitoring and compliance requirements for
arsenic do not become effective until January 23, 2006.
a)
Nitrate and Nitrite
The maximum contaminant levels for nitrate, nitrite and combined nitrate and
nitrite are as follows:
Contaminant
MCL (mg/L)
Nitrate
10 (as Nitrogen)
Nitrite
1 (as Nitrogen)
Total Nitrate and Nitrite
10 (as Nitrogen)
When the nitrate or nitrite sampling results indicate an excess of the maximum
contaminant level, a second analyses shall be initiated within 24 hours, and if the
mean of the two (2) analyses exceeds the maximum contaminant level the
supplier shall notify the Director and initiate public notification.
Systems unable to comply with the 24-hour sampling requirement must
immediately notify the consumers served by the area served by the PWS in
accordance with Section 17.6. Systems exercising this option must take and
analyze a confirmation sample within two (2) weeks of notification of the
analytical results of the first sample.
b)
Monitoring Frequency
The nitrate concentration of each active drinking water source maintained by a
water purveyor shall be determined as required by Section 16.1(e) of these
regulations. Beginning January 1, 1993 the nitrite and total nitrate/nitrite
concentration shall also be determined annually.
c)
Analytical Techniques
Nitrate analyses shall be made in accordance with the methods specified in
Appendix 1.
17.3
Organic Chemicals
Non-transient, non-community water systems shall be required to comply with the requirements
of Section 16.2 with the following exceptions. Monitoring and compliance with the requirements
for total trihalomethanes shall not be required until January 1, 2004 for Section 5.0 systems
278
serving fewer than 10,000. Non-transient, non-community Section 5.0 systems serving at least
10,000 should currently be meeting the monitoring and compliance requirements for total
trihalomethanes.
17.4 Turbidity
Non-community water systems shall comply with the requirements of Section 16.3.
17.5 Unregulated Contaminants and Special Monitoring
Non-transient, non-community water systems that serve more than 10,000 persons (effective
January 8, 1999) shall be required to monitor for unregulated contaminants in conformance with
Section 16.6 and 16.7.
17.6 Public Notification
Non-community water systems shall comply with the requirements of Section 16.8 herein.
17.7 Records
a)
Records of analyses performed by the water purveyor shall be maintained by the
water purveyor. The records shall contain the following information:
1.
The time, date and place of sampling and the name of the sample
collector;
2.
The sampling point and the reason for collection;
3.
Date analysis started and completion date if more than one (1) day is
needed;
4.
Name of laboratory and person responsible for performing the analysis;
5.
The analytical technique or method used; and
6.
The results of the analysis.
b)
Records of microbiological examinations shall be readily available for at least 5
years and records of nitrate analyses and turbidity determinations shall be readily
available for 10 years. Any written document relating to a sanitary survey of a
PWS shall be kept for at least 10 years.
c)
Records of action taken to correct a violation of these regulations shall be kept for
at least 3 years after the last action taken with respect to the particular violation
involved.
279
d)
Records concerning a variance or exemption granted to a system shall be kept for
at least 5 years following the expiration date of such variance or grant.
280
Section 18.0 Fee Schedule
18.1
Pursuant to the amended Section 46-13-3 of the General Laws Chapter 46-13, entitled
“Public Drinking Water Supply”, the Director is authorized to charge fees to support the
collection and analysis of samples that are required to meet the minimum monitoring
requirements for public drinking water supplies.
18.2
Any Public Drinking Water Supply for which analytical and collection services are
provided by the R.I. Department of Health to meet the minimum monitoring requirements
for public drinking water is liable for payment of the fee for these services.
18.3
The fee for each chemical, radiological and microbiological test required and conducted
by the Division of Laboratories shall be reasonable and shall be determined on the basis
of current costs for conducting the analysis. Such cost shall include administrative,
personnel, equipment and such other related costs which may be incurred in the analysis.
The laboratory fee schedule is listed in table 18-1.
18.4
The fee for each collection of each sample by the Division of Drinking Water Quality
shall be reasonable and shall be determined on the basis of current costs for such service.
The current sampling fee is $21.00. Sampling fees will be assessed for each on-site visit
to the supply for the purpose of collecting samples. It is the responsibility of the purveyor
to make the necessary operational arrangement for sampling. Scheduled on-site visits
canceled in the field because of lack of proper operational arrangement will be assessed
the sampling fee for the visit and any subsequent visit.
18.5
Payment for scheduled services will be required on the due date. The Department of
Health will provide bills approximately six (6) weeks in advance of the due date. Billing
will be on a quarterly basis. Payment shall be made payable to the General Treasurer,
State of Rhode Island.
18.6
Services will be provided only if payment in full has been received. It remains the
responsibility of the purveyor to meet all compliance testing requirements.
18.7
A surcharge shall be placed on overdue sampling and analysis payments. The surcharge
shall be set at the rate of $5.00 per month.
TABLE 18-1 Laboratory Fee Schedule
Chemical Group
Analysis Code
Test
Fee
Inorganic Chemistry
WL 1
Turbidity
$ 3.00
WL 2
Sediment
3.00
WL 3
Odor
3.00
WL 4
Color
3.00
WL 5
Total Solids
15.00
WL 6
Ignition Solids
15.00
WL 7
Suspended Solids
15.00
281
TABLE 18-1 Laboratory Fee Schedule
Chemical Group
Analysis Code
Test
Fee
WL 8
Settleable Solids
10.00
WL 9
% Moisture
8.00
WL 10
BOD (5 day)
30.00
WL 11
Cyanide
35.00
WL 12
Phosphorous (total)
20.00
WL 13
pH
5.00
WL 14 (lab)
Residual Chlorine
15.00
WL 15
Ammonia Nitrogen
15.00
WL 16
Nitrate
10.00
WL 56
Nitrite
10.00
WL 17
Phosphate (ortho)
20.00
WL 18
Alkalinity
10.00
WL 19
Aluminum
12.00
WL 20
Chloride
10.00
WL 21
Fluoride
12.00
WL 22
Hardness
15.00
WL 23
Iron
12.00
WL 24
Manganese
12.00
WL 25
Sodium
12.00
WL 26
Potassium
12.00
Inorganic Chemistry
WL 27
Calcium
12.00
WL 28
Magnesium
12.00
WL 29
Sulfate
15.00
WL 30
Arsenic
17.00
WL 31
Barium
12.00
WL 32
Cadmium
17.00
WL 33a
Chromium (hex)
15.00
WL 33b
Chromium (total)
17.00
WL 34
Copper
12.00
WL 35
Lead
17.00
WL 36
Mercury
30.00
WL 37
Nickel
12.00
WL 38
Selenium
17.00
WL 39
Silver
17.00
WL 40
Zinc
12.00
WL 41
Specific Conductance
10.00
WL 42
Oil & Grease
30.00
WL 43
MBAS
35.00
WL 44
Antimony
17.00
WL 45
Beryllium
17.00
WL 46
Turbidity (screen)
3.00
WL 47
Thallium
17.00
WL 49
Total Phenol
30.00
WL 50
Foam Screen
3.00
WL A
Inorganic Testing
44.00
WL J
Limited Metals Testing
110.00
WL K
Limited IOC Testing
51.00
WL 55
Composited Sodium
22.00
WL F
Metals
110.00
282
TABLE 18-1 Laboratory Fee Schedule
Chemical Group
Analysis Code
Test
Fee
Volatile Organic Chemistry
TO 2
4 Trihalomethane (THM) and Total
Trihalomethane
100.00
TO 4
Petroleum Hydrocarbons
180.00
TO 12
Water Quality Volatile Organics
200.00
TO 14
Other Base Neutral Extractable
200.00
TO 17
Petroleum Hydrocarbons and TO 12
200.00
TO 25
Method 525 Organic Compounds by
Liquid- Solid Extraction
250.00
Organic Chemistry
PE 4
Carbamates
75.00
PE 18
Pesticides/PCB's, Method 508
173.00
PE 14
EDB and DBCP, Method 504
100.00
PE 19
Chlorinated Acid Herbicides
Method 515.2
200.00
Radiological
RA 1
Gross Alpha
27.00
RA 2
Gross Beta
27.00
Microbiology
SM 1
Total and Fecal Coliform (Non Potable)
25.00
SM 1a
Total and Fecal Coliform (Potable)
25.00
SM 2
Total Coliform MF
15.00
SM 3
Heterotrophic Plate Count
7.00
SM 35
Fecal Coliform (confirmation)
10.00
SM 36
Total Coliform PA method
15.00
SM 34
Total & Focal Coliform (MMO-MUG)
25.00
Sampling
CL 1
Sampling fee (per on site visit)
21.00
283
Section 19.0 Rules Governing Practices and Procedures
19.1
All hearings and reviews required under the provisions of Chapter 46-13 of the General
Laws of Rhode Island, 1956, as amended, shall be held in accordance with the provisions
of the rules and regulations promulgated by the Rhode Island Department of Health
entitled Rules and Regulations of the Rhode Island Department of Health Regarding
Practices and Procedures Before the Department of Health and Access to Public Records
of the Department of Health (R42-35-PP).
284
Section 20.0 Violations, Noncompliance, and Enforcement
20.1
In order to obtain and/or maintain any approval, permit, certification, and/or license,
compliance with the provisions of the Act and these regulations is required. Failure to
comply with these requirements of the Act and/or these regulations shall constitute
grounds to revoke, suspend, or otherwise limit or restrict any approval, permit,
certification, and/or license issued by the Director. In addition, any person who violates
the provisions of Chapter 46-13, or these regulations shall be subject to the penalties and
remedies set forth in Section 46-13-16 of the Act.
Furthermore, the Director is authorized pursuant to the provisions of sections 46-13-10
and 46-13-12 to issue orders requiring corrective action(s) necessary to “provide safe and
potable water.”
20.2
(a)
Purpose and Goals
(1)
To assure the protection of public health, safety and welfare by promoting
compliance and deterring noncompliance with the Act, and the rules,
regulations, approvals, permits, certification, license and orders adopted
pursuant to the Act and these regulations;
(2)
To assure that the Department assesses administrative penalties, and
otherwise implements the Act and these regulations, lawfully, fairly and
consistently;
(3)
To clarify the Department's authority to enforce the Act and the rules,
regulations, approvals, permits and orders adopted pursuant to the Act.
(b)
Policy
These regulations promote a policy of assuring the effective enforcement of the
Act as administered by the Director and to deter noncompliance with the rules,
regulations, approvals, permits, certification, license and orders adopted pursuant
to the Act and these regulations:
(1)
seeking any appropriate legal and equitable relief, including:
(i)
removal of any economic benefit or competitive advantage realized
as a direct or indirect result of the violation;
(2)
assessing administrative penalties, where appropriate, which:
(i)
reflect the nature and gravity of the violation and the potential for
harm to the public health;
285
(ii)
reflect the length of time during which the violation was repeated
or continued;
(iii)
will deter future noncompliance by the person in violation; and
(iv)
will encourage continued compliance by persons similarly
regulated;
(3)
seeking from any person found to be in violation, those additional or
extraordinary costs which are actually expended by the Director during the
course of the investigation and enforcement of noncompliance for which
the State of Rhode Island is not otherwise reimbursed other than non-
overtime personnel costs; and
(4)
pursuing any other lawful enforcement option necessary to achieve
compliance.
(c)
Application
(1)
These regulations shall be liberally construed to permit the Department to
effectuate the purposes of the Act.
(2)
These regulations shall apply to all persons subject to enforcement action
by the Department under the Act, and the rules, regulations, approvals,
permits, certification, license and orders adopted pursuant to the Director's
authority hereunder.
(3)
These regulations shall be applied in a manner that is consistent with or
more stringent than any applicable Federal program requirements for
delegated programs.
(d)
Enforcement Options
The Director may pursue any combination of administrative and judicial
enforcement actions depending upon the circumstances and gravity of each case.
The penalty and remedies prescribed by the Act (Section 46-13-16) shall be
deemed to be concurrent and the existence of an exercised remedy shall not
prevent the Director from exercising any other remedy.
(e)
Preconditions for Assessment of Administrative Penalty
An administrative penalty may be assessed only for a violation or a failure to
comply that, at the time it occurred, constituted noncompliance with a legal
requirement:
(1)
which was then in effect; and
286
(2)
to which the person was then subject; and
(3)
to which these regulations apply.
(f)
Assessment of Administrative Penalty - Penalty Ceiling
No penalty shall exceed the maximum penalty allowed by the Act. The maximum
administrative penalty which the Director has the authority to impose under the
Act is $5000.00 per violation per day.
(1)
A penalty may be assessed “per day,” multi-day violations are counted
from the initial day of noncompliance until compliance is achieved.
(2)
A penalty may be assessed “per violation”, multiple violations of the same
law, rule, regulation, permit approval, certification, license or order are
counted as separate violations if any violation:
(i)
involves a prohibited act which is distinguishable from any other
by the nature of the act itself; or
(ii)
involves a prohibited act which is distinct from any other by the
time or place of its commission; or
(iii)
involves a prohibited act which is distinct from any other by
definition; or
(iv)
presents a risk of harm to the public health, safety or welfare which
is distinguishable from the risk threatened by any other violation.
(3)
Each day following service of a Notice of Violation, or Immediate
Compliance Order or Cease and Desist Order, to which the Director is a
party, during which a violation is repeated, continued or remains in place,
constitutes a continuing violation. The Director may assess an additional
administrative penalty, not to exceed five thousand dollars ($5,000) for
each day the violation or failure to comply is repeated, continued or
remains in place.
(4)
The penalty imposed shall continue to accrue from the day the Notice of
Violation, Immediate Compliance Order or Cease and Desist Order is
issued until compliance is achieved.
(g)
Assessment of Administrative Penalty - Calculation
The amount of the penalty will be calculated based on the factors enumerated
below.
287
(1)
The penalty may be based on the gravity of the violation. That portion will
be calculated according to the “DWQ Penalty Matrix” (See Appendix 3).
The applicable penalty range is reached by first determining the “Type of
Violation” and the “Deviation from the Standard” of the alleged violation.
(i)
“Type of Violation” - refers to the nature of the legal requirement
allegedly violated.
(A)
Type I violations - Type I violations include violations of
legal requirements identified by the Director as directly
related to the protection of the public health. Such
violations include, but are not necessarily limited to,
exceeding any MCL, failure to adhere to new source
approval requirements or plan requirements, and/or any
failure to comply with an order of the Director which is
presently enforceable.
(B)
Type II violations also have a direct impact on public
health, but are mainly non-compliance with technical
safeguards. Such violations include but are not limited to
failure to monitor as required, failure to comply with
reporting requirements, and failure to make public notice.
(C)
Type III violations have an indirect impact on public health
and are generally related to poor record keeping. Such
violations include, but are not limited to failure to submit
monitoring reports, late submittal of monitoring reports,
and failure to keep records on file as required.
(ii)
“Deviation from the Standard” - refers to the degree to which the
violation is out of compliance with the legal requirement allegedly
violated. The Deviation from the Standard may be determined
without consideration of the factors enunciated below in cases of
strict liability. In all other cases, the Department's assessment of
whether a violation is a minor, moderate or major deviation from
the standard is based upon an evaluation of one (1) or more of the
following factors except to the extent already considered:
(A)
the degree to which the act or failure to act was from
compliance;
(B)
whether the person took reasonable and appropriate steps to
prevent and/or mitigate the non-compliance;
288
(C)
whether the person has previously failed to comply with
any regulations, order, permit or approval issued or adopted
by the Department;
(D)
the degree of willfulness or negligence, including but not
limited to, how much control the violator had over the
occurrence of the violation and whether the violation was
foreseeable; and
(E)
any other factor(s) that may be relevant in determining the
amount of a penalty, provided that said other factor(s) shall
be set forth in the Notice of Violation or other written
notice of the assessment of a penalty.
(2)
The Economic Benefit from Non-Compliance
The penalty shall include an amount intended to offset the economic
benefit of non-compliance.
(i)
Such an amount may include, but not be limited to:
(A)
the cost of complying;
(B)
the cost of equipment needed to comply;
(C)
any associated operation and maintenance costs;
(D)
the costs of studies needed to achieve compliance;
(E)
any other delayed or avoided costs including, interest, market or
competitive advantage over other regulated entities which are in
compliance.
(ii)
The economic benefit portion may not be included in the penalty
only if:
(A)
there is no identifiable benefit from non-compliance; or
(B)
the amount of economic benefit cannot be quantified.
(3)
The penalty shall include additional or extraordinary costs which are
incurred by the Director during the course of the investigation and
enforcement of noncompliance for which the State of Rhode Island is not
otherwise reimbursed other than non-overtime personnel costs.
289
(4)
Nothing herein shall preclude the Director from resolving the outstanding
penalty through a Consent Agreement at any time he or she deems
appropriate.
(h)
Assessment of Administrative Penalty–Hearing
(1)
Any person against whom the Director seeks to assess an administrative
penalty for a violation of a law, rule, regulation, approval, license,
certification, or order which is within the Director's authority and
responsibility to enforce, has the right to request a hearing thereon. The
request for a hearing must be filed with the Director within thirty (30)
days after service of the notice assessing said penalty.
(2)
If a timely request for a hearing is made, a hearing shall be conducted in
accordance with Section 42-35 of the General Laws of Rhode Island,
1956, as amended.
(3)
Judicial review of any final decision of the administrative hearing officer
shall be available in accordance with Section 42-35-15 of the General
Laws of Rhode Island, 1956, as amended.
(i)
Assessment of Administrative Penalty–Enforcement
The Department's proposal of an administrative penalty shall become a final order
of the Director upon the person's election to waive, or failure to timely request, an
administrative hearing on the violation and/or the penalty. Each day during which
the person fails to pay said penalty or otherwise fails to comply with a final order
of the Director constitutes a separate and distinct violation. An additional
administrative penalty, not to exceed five thousand dollars ($5,000) for each such
violation of a final order, may be assessed by the Director.
The Director may also, by summons and complaint, seek to enforce said final
order in the Superior court for Providence County.
(j)
These regulations shall not be construed to govern any enforcement action which
is commenced by the Director prior to the formal adoption of these regulations, or
any administrative appeal taken therefrom, except that they shall apply to all
unresolved monitoring and public notice violations as of the effective date of
these regulations.
290
Section 21.0 Severability
21.1
If any provision of the rules and regulations herein or the application thereof to any
facility or circumstances shall be held invalid, such invalidity shall not affect the
provisions or application of the rules and regulations which can be given effect, and to
this end the provisions of the rules and regulations are declared to be severable.
291
APPENDIX I
INDEX
SECTION I - Microbiology
A.
Microbiological Methods for Surface Water and Ground Water Under the
Influence of Surface Water - Regulation Section 5.0
Total Coliforms
Fecal Coliforms
Heterotrophic Bacteria
B.
Microbiological Methods for Distribution Samples including Storage Facilities
and Ground Water Sources Regulation Section 16.4 and 17.2.
Total Coliforms
Fecal Coliforms and E. Coli
C.
Invalidation of Samples
SECTION II - Chemistry
A.
Inorganic Chemistry:
References for Sections 5.0, 6.0,16.1, and 17.2 of the Regulations.
1.
Surface Water Treatment Rule Monitoring
2.
Residual Disinfectant Concentration
Free Chlorine
Total Chlorine
Chlorine Dioxide
Ozone
3.
Turbidity
4.
Regulated Inorganic Chemicals
Methodology and Detection Limits
Sampling Protocol
Acceptance Criteria
5.
Special Inorganic Chemicals
Methodology
B.
Volatile Organic Chemistry
References for Sections 16.2, 16.6, 17.3 and 17.5 of the Regulations
292
1.
Regulated Volatile Organic Chemistry
Methodology
Laboratory Criteria
Acceptance Criteria
Detection Criteria
2.
Trihalomethane Chemistry
Methodology
3.
Unregulated Volatile Contaminants
Methodology
4.
Compositing of Samples
C.
Synthetic Organic Chemistry
References for Sections 16.2, 16.7, 17.3 and 17.5 of the Regulations
1.
Regulated Synthetic Organic Chemistry
Methodology
Laboratory Criteria
Acceptance Criteria
Detection Criteira
2.
Special Monitoring Synthetic Organic Chemicals
Methodology
D.
Radiological Chemistry
References for Section 16.5 of the Regulations
1.
Radiological Chemistry
Methodology
Detection Criteria
293
SECTION I
A.
Microbiological Testing
Reference for Section 5.0 - Surface Water and Ground Water Under the Influence
of Surface Water - Section 5.0
1.
PWSs conducting analyses for total coliforms, fecal coliforms and
heterotrophic bacteria, on surface water sources or ground water under the
influence of surface water, as required in Section 5.0 of these regulations,
must perform these analyses in accordance with one (1) of the following
analytical methods and by using analytical test procedures contained in
Technical Notes on Drinking Water Methods, EPA-600\R-94-173,
October 1994, which is available at NTIS PB95-104766.
2.
Total Coliforms:
a)
Total Coliform Fermentation Technique3,4,5
Citation1 9221 A,B,C
b)
Total Coliform Membrane Filter Technique
Citation1 92222 A,B,C
c)
ONPG-MUG Test 6
Citation1 9223
3.
Fecal Coliforms:
a)
Fecal Coliform MPN Procedure7
Citation1 9221 E
b)
Fecal Coliforms Membrane Filter Procedure
Citation1 9222 D
4.
Hetrotrophic Bacteria:2
a)
Pour Plate Method
Citation1 9215B
B.
Microbiological Testing - for Distribution Samples including Storage Facilities
and Ground Water Sources Regulation Sections 16.4 and 17.1
1.
The presence or absence of total coliform need only be determined. The
total coliform density is not required.
294
2.
The standard sample volume for total coliform analysis, regardless of the
analytical method used is 100ml.
3.
PWSs must conduct total coliform analyses in accordance with one (1) of
the analytical methods in the following table.
i)
Total Coliforms:8
aa)
Total Coliform Fermentation Technique3,4,5
Citation1 9221 A,B
bb)
Total Coliform Membrane Filter Technique
Citation1 92222 A,B,C
cc)
Presence-Absence (P-A) Coliform Test5,9
Citation1 92221 D
dd)
ONPG-MUG Test 6
Citation1 9223
ee)
Colisure Test 10
4.
PWSs must conduct fecal coliform analysis in accordance with the
following procedure. When the MTF Technique or Presence-Absence
(PA) Coliform Test is used to test for total coliforms, shake the lactose-
positive presumptive tube or P-A vigorously and transfer the growth with
a sterile 3-mm loop or sterile applicator stick into brilliant green lactose
bile broth and EC medium to determine the presence of total and fecal
coliforms, respectively. For EPA-approved analytical methods which use a
membrane filter, transfer the total coliform-positive culture by one (1) of
the following methods: remove the membrane containing the total
coliform colonies from the substrate with a sterile forceps and carefully
curl and insert the membrane into a tube of EC medium (the laboratory
may first remove a small portion of selected colonies for verification),
swab the entire membrane filter surface with a sterile cotton swab and
transfer the inoculum to EC medium (do not leave the cotton swab in the
EC medium), or inoculate individual total coliform-positive colonies into
EC Medium. Gently shake the inoculated tubes of EC medium to insure
adequate mixing and incubate in a waterbath at 44.5 + 0.2 C for 24 + 2
hours. Gas production of any amount in the inner fermentation tube of the
EC medium indicates a positive fecal coliform test. The preparation of EC
medium is described in the 18th edition of Standard Methods for the
Examination of Water and Wastewater, 1992, Method 9221E-p. 9-52,
Paragraph la. PWSs need only determine the presence or absence of fecal
coliforms; a determination of fecal coliform density is not required.
295
5.
PWSs must conduct analysis of Escherichia Coli in accordance with one
(1) of the following analytical methods:
(i)
EC medium supplemented with 50 ug/ml of 4-methylumbelliferyl-
beta-D-glucuronide (MUG) (final concentration). EC medium is
described in the 18th edition of Standard Methods for the
Examination of Water and Wastewater, 1992, Method 9221E–p. 9-
52, Paragraph la. MUG may be added to EC medium before
autoclaving. EC medium supplemented with 50 ug/ml of MUG is
commercially available. At least 10 ml of EC medium
supplemented with MUG must be used. The inner inverted
fermentation tube may be omitted. The procedure for transferring a
total coliform-positive culture to EC medium supplemented with
MUG shall be as specified in Paragraph (4) of this Section for
transferring a total coliform-positive culture to EC medium.
Observe fluorescence with an ultraviolet light (366 nm) in the dark
after incubating tube at 44.5 + 0.2 C for 24 + 2 hours; or
(ii)
Nutrient agar supplemented with 100 ug/ml 4-methlymbelliferyl-
beta-D-glucuronide (MUG) (final concentration). Nutrient Agar is
described in the 18th edition of Standard Methods for the
Examination of Water and Wastewater, 1992, p.9-47 to 9-48. This
test is used to determine if a total coliform-positive same, as
determined by the Membrane Filter Technique or any other
method in which a membrane filter is used, contains E. Coli.
Transfer the membrane filter containing a total coliform
colony(ies) to nutrient agar supplemented with 100 ug/ml (final
concentration) of MUG. After incubating the agar plate at 35 C for
4 hours, observe the colony(ies under ultraviolet light (366 nm) in
the dark for fluorescence. If fluorescence is visible, E. Coli are
present.
(iii)
Minimal Medium ONPG-MUG (MMO-MUG) Test, as set forth in
the article “National Field Evaluation of a Defined Substrate
Method for the Simultaneous Detection of Total Coliforms and
Escherichia Coli from Drinking Water: Comparison with
Presence-Absence Techniques” (Edgerg et al.), Applied and
Environmental Microbiology, Volume 55, pp. 1003-1008, April
1989. (Note: The Autoanalysis Colilert System is an MMO-MUG
test). If the MMO-MUG test is total coliform-positive after a 24-
hour incubation, test the medium for fluorescence with a 366-nm
ultraviolet light (preferably with a 6-watt lamp) in the dark. If
fluorescence is observed, the sample is E. Coli-positive. If
fluorescence is questionable (cannot be definitely read) after 24
hours incubation, incubate the culture for an additional four (4)
hours (but not to exceed 28 hours total), and again test the medium
296
for fluorescence. The MMO-MUG Test with hepes buffer in lieu
of phosphate buffer is the only approved formulation for the
detection of E. Coli.
(iv)
The Colisure Test. A description of the Colisure Test may be
obtained from the Millipore Corporation, Technical Services
Department, 80 Ashby Road, Bedford, MA 01730.
6.
As an option to the Minimal Medium ONPG-MUG (MMO-MUG) Test, as
set forth in Paragraph 5(iii) above, a system with a total coliform positive,
MUG-negative, MMO-MUG test may further analyze the culture for the
presence of E. Coli by transferring a 0.1 ml, 28-hour MMO-MUG culture
to EC Medium + MUG with a pipet. The formulation and incubation
conditions of EC Medium + MUG, and observation of results are
described in Paragraph 5(i) above.
C.
Invalidation of Samples
1.
A total coliform sample invalidated under this Paragraph does not count
towards meeting the minimum monitoring requirements of this Section.
2.
The Director will invalidate a total coliform-positive sample and
document same in writing only if:
i)
The laboratory establishes that improper sample analysis caused
the total coliform-positive result;
ii)
The Director determines that the total coliform-positive sample
resulted from a domestic or other non-distribution system
plumbing problem; or
iii)
The Director has substantial grounds to believe that a total
coliform-positive result is due to a circumstance or condition
which does not reflect water quality in the distribution system. (In
this case, the system must still collect all repeat samples required.)
3.
A total coliform-positive sample will not be invalidated solely on the
grounds that all repeat samples are total coliform negative.
4.
A laboratory must invalidate a total coliform sample, unless total
coliforms are detected, if
i)
The sample produces a turbid culture in the absence of gas
production using the method cited in Section 16.4 b) 4) a);
297
ii)
The sample produces a turbid culture in the absence of an acid
reaction; using the method cited in Section 16.4 b) 4) c); or
iii)
It exhibits confluent growth, or produces colonies too numerous to
count, using the method cited in Section 16.4 b) 4) b).
5.
If a laboratory invalidates a sample for the above reasons, the system must
collect another sample from the same location as the original sample
within 24 hours of being notified of the result. The system must continue
to re-sample within 24 hours and have the samples analyzed until it
obtains a valid result. The Director may extend the 24-hour limit on a
case-by-case basis if the system has a logistical problem in collecting the
repeat samples within 24 hours that is beyond its control. In the case of a
extension, the Director will specify how much time the system has to
collect the repeat samples.
Footnotes:
1
Except where noted all methods refer to the 18th edition of Standard Methods for
the Examination of Water and Wastewater, 1992, American Public Health
Association, 1015 Fifteenth Street NW, Washington, DC 20005.
2
The time from sample collection to initiation of analysis may not exceed 8 hours.
3
Lactose broth, as commercially available, may be used in lieu of lauryl tryptose
broth, if the system conducts at least 25 parallel tests between this medium and
lauryl tryptose broth using the water normally tested, and this comparison
demonstrates that the false-positive are for total coliforms, using lactose broth, is
less than 10 percent.
4
Media should cover inverted tubes at least one-half to two-thirds after the sample
is added.
5
No requirement exists to run the completed phase on 10 percent of all total
coliform-positive confirmed tubes.
6
The ONPG-MUG Test is also known as the Autoanalysis Colilert System.
7
A-1 Broth may be held up to three (3) months in a tightly closed screwcap tube at
4 C.
8
The time from sample collection to initiation of analysis may not exceed 30 hours.
9
Six-times formulation strength may be used if the medium is filter-sterilized
rather than autoclaved.
298
10
The Colisure Test must be incubated for 28 hours before examining the results. If
an examination of the results at 28 hours is not convenient, then results may be
examined at any time between 28 hours and 48 hours. A description of the
Colisure Test may be obtained from the Millipore Corporation, Technical
Services Department, 80 Ashby Road, Bedford, MA 01730.
SECTION II
A.
Inorganic Chemistry
References for Sections 5.0, 6.0, 16.1 and 17.2 of the Regulations
1.
Surface Water Treatment Rule Monitoring
a)
PWSs which must conduct analyses to meet the requirements of
Section 5.0 for turbidity, temperature and measure residual
disinfectant concentrations must use the methods contained in the
18th edition of Standard Methods for the Examination of Water
and Wastewater, 1992 with other analytical test procedures are
contained in Technical Notes on Drinking Water Methods, EPA-
600/R-94-173, October 1994, which is available at NTIS PB95-
104766. Residual disinfectant concentrations for free chlorine and
combined chorine also may be measured by using DPD
colorimetric test kits. Free and total chlorine residuals may be
measured continuously by adapting a specified chlorine residual
method for use with a continuous monitoring instrument provided
the chemistry, accuracy and precision remain same. Instruments
used for continuous monitoring must be calibrated with a grab
sample measurement at least every five (5) days, or with a protocol
approved by the Director.
2.
Residual Disinfectant Concentration:
a)
Free Chlorine:
(i)
Amperometric Titrationation Technique.
Citation 4500-Cl D
(ii)
DPD Ferrous Titrimetric.
Citation 4500-Cl F
(iii)
DPD Colorimetric.
Citation 4500-Cl G
(iv)
Syringaldazine (FACTS).
Citation 4500-Cl H
299
b)
Total Chlorine:
(i)
Amperometric Titrationation Technique.
Citation 4500-Cl D
(ii)
Amperometric Titrationation (low level measurement).
Citation 4500-Cl E
(iii)
DPD Ferrous Titrimetric.
Citation 4500-Cl F
(iv)
DPD Colorimetric.
Citation 4500-Cl G
(v)
Iodometric Electrode.
Citation 4500-Cl I
c)
Chlorine Dioxide:
(i)
Amperometric Titrationation Technique.
Citation 4500-ClO2 C
(ii)
DPD Method.
Citation 4500-ClO2 D
(iii)
Amperometric Titrationation.
Citation 4500-ClO2 E
d)
Ozone:
(i)
Indigo Method.
Citation 4500-O3 B
3.
Turbidity:
a)
Nephelometric Method.
Citation1 2130 B
b)
Nephelometric Method.
Citation8 180.1
c)
Great Lakes Instruments.
Citation9 Method 2
Footnotes:
300
1
Except where noted all methods refer to the 18th edition of Standard Methods for the
Examination of Water and Wastewater, 1992, American Public Health Association, 1015
Fifteenth Street NW, Washington, DC 20005.
8
“Methods for the Determination of Inorganic Substances in Environmental Samples”,
EPA-600/R-93-100, August 1993. Available at NTIS, PB94-121811.
9
GLI Method 2. “Turbidity”, November 2, 1992, Great Lakes Instruments, Inc., 8855
North 56th Street, Milwaukee, Wisconsin 53223.
4.
Regulated Inorganic Chemical Monitoring
a)
Methodology and Detection Limits
PWSs conducting analyses of inorganic chemicals as required in
Sections 6.0, 16.0 and 17.0 of these regulations shall conduct these
analyses in accordance with one of the following analytical
methods or their equivalent as determined by EPA. Criteria for
analyzing arsenic, barium, beryllium, cadmium, chromium, copper,
lead, nickel, selenium, sodium and thallium are contained in
Technical Notes on Drinking Water Methods, EPA-600\R-94-173,
October 1994, which is available at NTIS PB95-104766.
ii)
Effective January 23, 2006, arsenic sampling results will be
reported to the nearest 0.001 mg/L.
Contaminant13
Methodology13
EPA
ASTM3
SM4
Other
Minimum
Detection
Limit
(mg/L)
Antimony
ICP-Mass Spectrometry
2200.8
0.0004
Hydride-Atomic
Absorption
D-3697-92
0.001
Atomic Absorption;
Platform
2200.9
0.0008
Atomic Absorption;
Furnance
3113B
0.003
Arsenic14
Inductively Coupled
Plasma 15
2200.7
3120B
ICP-Mass Spectrometry
2200.8
160.0014
Atomic Absorption;
Platform
2200.9
170.0005
Atomic Absorption;
Furnace
D-2972-
93C
3113B
0.001
Hydride Atomic
Absorption
D-2972-
93B
3114B
0.001
Asbestos
Transmission Electron
Microscopy
9100.1
0.01
MFL
Transmission Electron
Microscopy
10100.2
0.01
MFL
301
Contaminant13
Methodology13
EPA
ASTM3
SM4
Other
Minimum
Detection
Limit
(mg/L)
Barium
Inductively Coupled
Plasma
2200.7
3120B
0.002
ICP-Mass Spectrometry
2200.8
0.002
Atomic Absorption; Direct
3111D
0.1
Atomic Absorption;
Furnace
3113B
0.002
Beryllium
Inductively Coupled
Plasma
2200.7
3120B
0.0003
ICP-Mass Spectrometry
2200.8
0.0003
Atomic Absorption;
Platform
2200.9
0.00002
Atomic Absorption;
Furnace
D-3645-
93B
3113B
0.0002
Cadium
Inductively Coupled
Plasma
2200.7
0.001
ICP-Mass Spectrometry
2200.8
0.001
Atomic Absorption;
Platform
2200.9
Atomic Absorption;
Furnace
3113B
0.0001
Chromium
Inductively Coupled
Plasma
2200.7
3120B
0.007
ICP-Mass Spectrometry
2200.8
0.007
Atomic Absorption;
Platform
2200.9
Atomic Absorption;
Furnace
3113B
0.001
Cyanide
Manual Distillation
followed by
4500-CN-C
Spectrophotometric,
Amenable
D2036-91B
4500CN-G
0.02
Spectrophotometric
Manual
D2036-91A
4500-CN-E
51-3300-
85
0.02
Semi-automated
6335.4
0.005
Selective Electrode
4500CN-F
0.05
Fluoride
Ion Chromatography
6300.0
D4327-91
4110B
Manual Distill.;Color.
SPADNS
4500F-B,D
1.0
Manual Electrode
D1179-93B
4500F-C
1.0
Automated Electrode
11380-
75WE
1.0
Automated Alizarin
4500F-E
11129-
71W
1.0
Mercury
Manual, Cold Vapor
2245.1
D3223-91
3112B
0.0002
Automated, Cold Vapor
1245.2
0.0002
ICP-Mass Spectrometry
2200.8
Nickel
Inductively Coupled
Plasma
2200.7
3120B
0.005
ICP-Mass Spectrometry
2200.8
0.0005
302
Contaminant13
Methodology13
EPA
ASTM3
SM4
Other
Minimum
Detection
Limit
(mg/L)
Atomic Absorption;
Platform
2200.9
0.0006
Atomic Absorption; Direct
3111B
Atomic Absorption;
Furnace
3113B
0.001
Nitrate
Ion Chromatography
6300.0
D4327-91
4110B
8B-1011
0.01
Automated Cadmium
Reduction
6353.2
D3867-90A
4500-NO3-
F
0.05
Ion Selective Electrode
4500-NO3-
D
7601
1
Manual Cadmium
Reduction
D3867-90B
4500-NO3-
E
0.01
Nitrite
Ion Chromatography
6300.0
D4327-91
4110B
8B-1011
0.004
Automated Cadmium
Reduction
6353.2
D3867-90A
4500-NO3-
F
0.05
Manual Cadmium
Reduction
D3867-90B
4500-NO3-
E
0.01
Spectrophotometric
4500-NO2-
B
0.01
Selenium
Hydride-Atomic
Absorption
D3859-93-
A
3114B
0.002
ICP-Mass Spectrometry
2200.8
Atomic Absorption;
Platform
2200.9
Atomic Absorption;
Furnace
D3859-93B
3113B
0.002
Thallium
ICP-Mass Spectrometry
2200.8
0.0003
Atomic absorption,
furnance
279.2
3113B
0.001
Atomic Absorption;
Platform
2200.9
0.0007
Lead
Atomic absorption;
furnace
D3559-90D
3113B
ICP-Mass Spectrometry
2200.8
Atomic absorption;
platform
2200.9
Copper
Atomic absorption;
furnace
D1688-90C
3113B
Atomic absorption; direct
aspiration
D1688-90A
3111B
ICP
2200.7
3120B
ICP-Mass spectrometry
2200.8
Atomic absorption;
platform
2200.9
pH
Electrometric
1150.1
D1293-84
4500-H+-B
1150.2
Conductivity
Conductance
D1125-91A
2510-B
Calcium
EDTA titrimetric
D511-93A
3500-Ca-D
303
Contaminant13
Methodology13
EPA
ASTM3
SM4
Other
Minimum
Detection
Limit
(mg/L)
Atomic absorption; direct
aspiration
D511-93B
3111B
Inductively-coupled
plasma
2200.7
3120B
Alkalinity
Titrimetric
D1067-92B
2320B
Electrometric titration
51-1030-
85
Orthophosphate12
Colorimetric, automated,
ascorbic acid
6365.1
4500-P-F
Colorimetric, ascorbic
acid, single reagent
D515-88A
4500-P-E
Colorimetric,
phosphomolybdate;
51-1601-
85
Automated-segmented
flow;
51-2601-
90
Automated discrete
51-2598-
85
Ion Chromatography
6300.0
D4327-91
4110
Silica
Colorimetric, molybdate
blue;
51-1700-
85
automated-segmented
flow
51-2700-
85
Colormetric
D859-88
Molybdosilicate
4500-Si-D
Heteropoly blue
4500-Si-E
Automated method for
molybdate-reactive silica
4500-Si-F
Inductively-coupled
plasma
2200.7
3120B
Temperature
Thermometric
2550
Sodium
Inductively-coupled
plasma
2200.7
atomic absorption; direct
aspiration
3111B
Footnotes:
1
Methods 150.1, 150.2 and 245.2 are available from US EPA, EMSL, Cincinnati, OH
45268. The identical methods were formerly in “Methods for Chemical Analysis of
Water and Wastes”, EPA-600/4-79-020, March 1983, which is available at NTIS, PB84-
128677.
2
“Methods for the Determination of Metals in Environmental Samples–Supplement I”,
EPA-600/R-94-111, May 1994. Available at NTIS, PB 95-125472.
304
3
The procedures shall be done in accordance with the Annual Book of ASTM Standards,
1994 and 1996, Vols. 11.01 and 11.02, American Society for Testing and Materials. The
previous versions of D1688-95A, D1688-95C (copper), D3559-95D (lead), D1293-95
(pH), D1125-91A (conductivity) and D859-94 (silica) are also approved. These previous
versions D1688-90A, C; D3559-90D, D1293- 84, D1125-91A and D859-88, respectively
are located in the Annual Book of ASTM Standards, 1994, Vols. 11.01. Copies may be
obtained from the American Society for Testing and Materials, 100 Barr Harbor Drive,
West Conshohocken, PA 19428.
4
The procedures shall be done in accordance with the 18th and 19th editions of Standard
Methods for the Examination of Water and Wastewater, 1992 and 1995, respectively,
American Public Health Association; either edition may be used. Copies may be obtained
from the American Public Health Association, 1015 Fifteenth Street NW, Washington,
DC 20005.
5
Available from Books and Open-File Reports Section, U.S. Geological Survey, Federal
Center, Box 25425, Denver, CO 80225-0425.
6
“Methods for the Determination of Inorganic Substances in Environmental Samples”,
EPA-600/R-93-100, August 1993. Available at NTIS, PB94-121811.
7
The procedure shall be done in accordance with the Technical Bulletin 601 “Standard
Method of Test for Nitrate in Drinking Water”, July 1994, PN 221890-001, Analytical
Technology, Inc. This incorporation by reference was approved by the Director of the
Federal Register in accordance with 5 U.S.C. 552(a) and 1 CRF Part 51. Copies may be
obtained from ATI Orion, 529 Main Street, Boston, MA 02129. Copies may be inspected
at EPA’s Drinking Water Docket, 401 M Street, SW, Washington, DC 20460; or at the
Office of the Federal Register, 800 North Capital Street, NW, Suite 700, Washington,
DC.
8
Method B-1011, “Waters Test Method for Determination of Nitrite/Nitrate in Water
Using
Single
Column
Ion
Chromatography”,
Millipore
Corporation,
Waters
Chromatography Division, 34 Maple Street, Miliford, MA 01757.
9
Method 100.1, “Analytical Method for Determination of Asbestos Fibers in Water”,
EPA-600/4-83-043, EPA, September 1983. Available at NTIS, PB83-260471.
10
Method 100.2, “Determination of Asbestos Structure Over 10µm in Length in Drinking
Water”, EPA-600/R-94-134, June 1994. Available at NTIS, PB94-201902.
11
The procedures shall be done in accordance with the Industrial Method No. 129-71W,
“Fluoride in Water and Wastewater”, December 1972, and Method No. 380-75WE,
“Fluoride in Water and Wastewater”, February 1976, Technicon Industrial Systems. This
incorporation by reference was approved by the Director of the Federal Register in
accordance with 5 U.S.C. 552(a) and 1 CRF Part 51. Copies may be obtained from the
Technicon Industrial Systems, Tarrytown, NY 10591. Copies may be inspected at EPA’s
305
Drinking Water Docket, 401 M Street, SW, Washington, DC 20460; or at the Office of
the Federal Register, 800 North Capital Street, NW, Suite 700, Washington, DC.
12
Unfiltered, no digestion or hydrolysis.
13
Because MDLs reported in EPA Methods 200.7 and 200.9 were determined using a 2X
preconcentration step during sample digestion, MDLs determined when samples are
analyzed by direct analysis (i.e., no sample digestion) will be higher. For direct analysis
of cadmium and arsenic by Method 200.7, and arsenic by Method 3120 B sample
preconcentration using pneumatic nebulization may be required to achieve lower
detection limits. Preconcentration may also be required for direct analysis of antimony,
lead, and thallium by Method 200.9; antimony and lead by Method 3113 B; and lead by
Method D3559-90D unless multiple in-furnace depositions are made.
14
If ultrasonic nebulization is used in the determination of arsenic by Methods 200.7,
200.8, or SM 3120 B, the arsenic must be in the pentavalent state to provide uniform
signal response. For methods 200.7 and 3120 B, both samples and standards must be
diluted in the same mixed acid matrix concentration of nitric and hydrochloric acid with
the addition of 100 µL of 30% hydrogen peroxide per 100ml of solution. For direct
analysis of arsenic with method 200.8 using ultrasonic nebulization, samples and
standards must contain one (1) mg/L of sodium hypochlorite.
15
After January 23, 2006 analytical methods using the ICP-AES technology, may not be
used because the detection limits for these methods are 0.008 mg/L or higher. This
restriction means that the two (2) ICP-AES methods (EPA Method 200.7 and SM 3120
B) approved for use for the MCL of 0.05 mg/L may not be used for compliance
determinations for the revised MCL of 0.010 mg/L. However, prior to 2006, systems may
have compliance samples analyzed with these less sensitive methods.
16
Using selective ion monitoring, EPA Method 200.8 (ICP-MS) is capable of obtaining a
MDL of 0.0001 mg/L.
17
The MDL reported for EPA method 200.9 (Atomic Absorption; Platform-Stabilized
Temperature) was determined using a 2x concentration step during sample digestion. The
MDL determined for samples analyzed using direct analysis (i.e., no sample digestion)
will be higher. Using multiple depositions, EPA 200.9 is capable of obtaining MDL of
0.0001 mg/L.
b)
Sampling Protocol
i)
Sample collection for the inorganic chemicals listed below
shall be conducted using the sample preservation containers
and maximum holding time procedures specified as
follows:
Antimony:
306
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Arsenic:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Asbestos:
Preservative: Cool 4˚C
Container: Plastic or glass
Barium:
Preservative: cool,4˚C
Container: Plastic or glass
Time: ASAP but not over 6 months
Beryllium:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Cadmium:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Chromium:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Cyanide:
Preservative: Cool 4˚C NaOH to pH > 12 (6g
Ascorbic acid if chlorine is present)
Container: Plastic or glass
Time: ASAP, but not over 14 days
307
Fluoride:
Preservative: None
Container: Plastic or glass
Time: ASAP, but not over 1 month
Mercury:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 28 days
Nickel:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Nitrate:
Preservative:
Chlorinated - Cool, 4˚C
Non-Chlorinated - Con H2SO4 to pH <2
Container:
Chlorinated - Plastic or glass
Non-Chlorinated - Plastic or glass
Time:
Chlorinated - ASAP but not over 28 days
Non-Chlorinated - ASAP but not over 14
days
Nitrite:
Preservative: Cool, 4˚C
Container: Plastic or glass
Time: ASAP but not over 48 hours
Selenium:
Preservative: Conc. HNO2 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Thallium:
308
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
Lead:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not over 6 months
NOTE: The technique applicable to total metals
must be used and samples cannot be filtered.
Copper:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP but not more than 6 months
NOTE: The technique applicable to total metals
must be used and samples cannot be filtered.
pH:
Preservative: None
Container: Plastic or glass
Time: Test immediately
Conductivity:
Preservative: Cool, 4˚C
Container: Plastic or glass
Time: ASAP, but not more than 28 days
Calcium:
Preservative: Conc. HNO3 to pH <2
Container: Plastic or glass
Time: ASAP, but not more than 6 months
Alkalinity:
Preservative: Cool, 4˚C
Container: Plastic or glass
Time: ASAP, but not more than 14 days
Orthophosphate:
309
Preservative: Cool, 4˚C
Container: Plastic or glass
Time: ASAP but not more than 48 hours
Silica:
Preservative: Cool, 4˚C
Container: Plastic only
Time: ASAP but not more than 28 days
Sodium:
Container: Plastic or glass
Temperature:
Preservative: None
Container: Plastic or glass
Time: Test immediately
Turbidity:
Preservative: Cool 4˚C
Container: Plastic or glass
Time: ASAP but not over 48 hours
Note:
(1) For approved analytical procedures for metals, the technique applicable to total metals must
be used.
(2) For cyanide determinations samples must be adjusted with sodium hydroxide to pH 12 at the
time off collection.
(3) When chilling is indicated the sample must be shipped and stored at 4 deg. C or less.
(4) Acidification of nitrate or metals samples may be with a concentrated acid or a dilute (50%
by volume) solution of the applicable concentrated acid. Acidification of samples for metals
analysis is encouraged and allowed at the laboratory rather than at the time of sampling provided
the shipping time and other instructions in Section 8.3 of EPA Methods 200.7 or 200.8 or 200.9
are followed.
(5) Plastic or glass may be hard or soft.
(6) Follow additional (if any) information on preservation, containers, or holding times that is
specified in method.
c)
Acceptance Criteria
310
(i)
For a laboratory to receive certification to conduct analyses
for the inorganic chemicals listed in Appendix 1 Section II
(4) (a) of these regulations, the laboratory must:
(A)
Analyze Performance Evaluation samples which
include those substances provided by EPA
Environmental Monitoring Systems Laboratory or
equivalent samples provided by the State.
(B)
Achieve quantitative results on the analyses that are
within the following acceptance limits:
Contaminant
Acceptance limit
Antimony
+30 at >0.006 mg/1
Arsenic1
±30 at >0.003 mg/1
Asbestos
2 standard deviations based on study statistics
Barium
+15% at >0.15 mg/1
Beryllium
+15% at >0.001 mg/1
Cadmium
+20% at >0.002 mg/1
Chromium
+15% at >0.01 mg/1
Cyanide
+25% at >0.1 mg/1
Fluoride
+10% at >1 to 10 mg/1
Mercury
+30% at >0.0005 mg/1
Nickel
+15% at >0.01 mg/1
Nitrate
+10% at >0.4 mg/1
Nitrite
+15% at >0.4 mg/1
Selenium
+20% at >0.01 mg/1
Thallium
+30% at >0.002 mg/1
1 The arsenic acceptance limit criteria becomes effective January 23, 2006
Lead: +30 percent of the actual amount in the
Performance Evaluation sample when the actual
amount is greater than or equal to 0.005 mg/L. The
Practical Quantitation Level, or PQL for lead is
0.005 mg/L.
Copper: +10 percent of the actual amount in the
Performance Evaluation sample when the actual
amount is greater than or equal to 0.050 mg/L The
Practical Quantitation Level, or PQL for copper is
0.03 mg/L.
Achieve method detection limits as follows for lead
and copper:
311
Lead: 0.001 mg/L must be achieved (only if source
water compositing is done under S6.23(a)(4)).
Copper: 0.001 mg/L or 0.020 mg/L when atomic
absorption direct aspiration is used (only if source
water compositing is done under S6.23(a)(4).
(C)
The Director has the authority to allow the use of
previously collected monitoring data for purposes of
monitoring, if the data were collected and analyzed
in accordance with the requirements of this subpart
for lead and copper monitoring.
(D)
All lead and copper levels measured between the
PQL and MDL must be with reported as measured
or they can be reported as one-half the PQL
(0.0025mg/L). All levels below the lead and copper
MDLs must be reported as zero.
(E)
All copper levels measured between the PQL and
MDL must be either reported as measured or they
can be reported as one-half the PQL (0.015 mg/L).
All levels below the copper MDL must be reported
as zero.
4.
Special Inorganic Chemical Monitoring
a.
System monitoring for the unregulated inorganic contaminant
sulfate shall use one (1) of the method(s) identified below:
i)
EPA Method 300.0, and 375.2.2 are in “Methods for the
Determination of Inorganic Substances in Environmental
Samples”, EPA600/R-93-100, August 1993 Available at
NTIS, PB94-121811.
ii)
Method D4327-91 shall be done in accordance with the
Annual book of ASTM Standards, 1994, Vol. 11.01 and
11.02, American Society for Testing and Materials, 1961
Race Street, Philadelphia, PA 19103.
iii)
Method 4110, 4500-SO4-F and 4500-SO4-C,D shall be
followed in accordance with the Standard Methods for the
Examination of Water and Wastewater 18th Edition
Supplement, 1992, American Public Health Association.
Copies may be obtained from the American Public Health
Association, 1015 Fifteenth Street NW, Washington, DC
312
20005. Copies may be inspected at EPA's Drinking Water
Docket, 401 M Street, SW., Washington, DC 20460; or at
the Office of the Federal Register, 800 North Capitol
Street, NW., Site 700, Washington DC.
B.
Volatile Organic Chemistry (VOC’s)
References for Sections 16.2, 16.6, 17.3 and 17.5 of the Regulations
1.
Regulated Volatile Organic Chemicals
a)
Methodology
(i)
PWSs conducting analyses of inorganic chemicals as listed
below and as required in Section 16.0 and 17.0 of these
regulations shall conduct these analyses in accordance with
one (1) of the following analytical methods or their
equivalent as determined by EPA:
aa)
Method 502.2 is in “Methods for the Determination
of Organic Compounds in Drinking Water, EPA-
600/4-88-039, December 1988, Revised, July 1991.
bb)
Method 551 is in Methods for the Determination of
Organic
Compounds
in
Drinking
Water–
Supplement I, EPA-600-4-90-020, July 1990.
cc)
Method 524.2 is in Methods for the Determination
of Organic Compounds in Drinking Water –
Supplement II, EPA-600/R-92-129, August 1992.
Contaminant
Method
Benzene
502.2, 524.2.
Carbon tetrachloride
502.2, 524.2, 551.
Chlorobenzene
502.2, 524.2
1,2-Dichlorobenzene
502.2, 524.2.
1,4-Dichlorobenzene
502.2, 524.2.
1,2-Dichloroethane
502.2, 524.2.
Cis-Dichloroethylene
502.2, 524.2.
Trans-dichloroethylene
502.2, 524.2.
Dichloromethane
502.2, 524.2.
1,2-Dichloropropane
502.2, 524.2.
Ethylbenzene
502.2, 524.2.
Styrene
502.2, 524.2.
Tetrachloroethylene
502.2, 524.2, 551.
1,1,1-Trichloroethane
502.2, 524.2, 551.
Trichloroethylene
502.2, 524.2, 551.
313
Toluene
502.2, 524.2.
1,2,4-Trichlorobenzene
502.2, 524.2.
1,1-Dichloroethylene
502.2, 524.2.
1,1,2-Trichloroethane
502.2, 524.2.
Vinyl chloride
502.2, 524.2.
Xylenes (total)
502.2, 524.2.
b.
Certification Criteria
(i)
To receive certification to conduct analyses for the
contaminants listed in Appendix 1 Section II B (1), above
the laboratory must:
(ii)
Analyze Performance Evaluation samples which include
these
substances
provided
by
EPA
Environmental
Monitoring Systems Laboratory or equivalent samples
provided by the State.
(iii)
Achieve
the
quantitative
acceptance
limits
under
Paragraphs (iv) and (v) of this Section for at least 80
percent of the regulated organic chemicals listed in ref.
(iv)
Achieve quantitative results on the analyses performed
under Paragraph (ii) of this Section that are within +20% of
the actual amount of the substances in the Performance
Evaluation sample when the actual amount is greater than
or equal to 0.010 mg/1.
(v)
Achieve quantitative results on the analyses performed
under Paragraph (ii) of this Section that are within +40
percent of the actual amount of the substances in the
Performance Evaluation sample when the actual mount is
less than 0.010 mg/1.
(vi)
Achieve a method detection limit of 0.0005 mg/1.
(vii) To receive certification for vinyl chloride, the laboratory
must:
aa)
Analyze Performance Evaluation samples provided
by
EPA
Environmental
Monitoring
Systems
Laboratory or equivalent samples provided by the
State.
bb)
Achieve quantitative results on the analyses
performed under Paragraph ref of this Section that
314
are within +40 percent of the actual amount of vinyl
chloride in the Performance Evaluation sample.
cc)
Achieve a method detection limit of 0.0005 mg/1.
dd)
Obtain certification for the contaminants listed in
above.
2.
Total Trihalomethane Chemistry
a.
Methodology
i)
Sampling and analyses made pursuant to Section 16.0 shall
be conducted by the total trihalomethane methods as listed
below and in Technical Notes on Drinking Water Methods,
EPA-600\R-94-173, October 1994, which is available at
NTIS PB95-104766.
aa)
Method 502.2 is in “Methods for the Determination
of Organic Compounds in Drinking Water, EPA-
600/4-88-039, December 1988, Revised, July 1991.
bb)
Method 551 is in Methods for the Determination of
Organic
Compounds
in
Drinking
Water–
Supplement I, EPA-600-4-90-020, July 1990.
cc)
Method 524.2 is in Methods for the Determination
of Organic Compounds in Drinking Water –
Supplement II, EPA-600/R-92-129, August 1992.
3.
Unregulated Contaminants and Special Monitoring
a.
Unregulated Volatile Organic Contaminants Methodology
i)
Analysis for the unregulated contaminants listed in Section
16.6 shall be conducted using EPA Methods 502.2 or
524.2, or their equivalent as determined by EPA, except
analysis
for
bromodichloromethane,
bromoform,
chlorodibromomethane and chloroform may be conducted
by
EPA
Method
551,
and
analysis
for
1,2,3-
trichloropropane also may be conducted by EPA Method
504.1. A source for the EPA methods is listed below:
aa)
Method 502.2 is in “Methods for the Determination
of Organic Compounds in Drinking Water, EPA-
600/4-88-039, December 1988, Revised, July 1991.
315
bb)
Method 524.2 is in Methods for the Determination
of Organic Compounds in Drinking Water –
Supplement II, EPA-600/R-92-129, August 1992.
cc)
Method 551 is in Methods for the Determination of
Organic
Compounds
in
Drinking
Water–
Supplement I, EPA-600-4-90-020, July 1990.
dd)
EPA Method 504.1 is available from US EPA
EMSL, Cincinnati OH 45268.
4.
Compositing of Samples:
All samples must be composited in the laboratory and analyzed within
fourteen (14) days of collection.
a.
The following procedure must be followed for the compositing
samples prior to GC analysis.
i)
Add 5 ml or equal larger amounts of each sample (up to 5
samples are allowed) to a 25 ml glass syringe. Special
precautions must be made to maintain zero headspace in
the syringe.
ii)
The samples must be cooled at 4 C during this step to
minimize volatilization losses.
iii)
Mix well and draw out a 5-ml aliquot for analysis.
iv)
Follow sample introduction, purging and desorption steps
described in the method.
v)
If less than five (5) samples are used for compositing, a
proportionately small syringe may be used.
b.
The following procedure must be followed for the compositing
samples prior to GC/MS analysis.
i)
Inject 5-ml or equal larger amounts of each aqueous sample
(up to 5 samples are allowed) into a 25-ml purging device
using the sample introduction technique described in the
method.
ii)
The total volume of the sample in the purging device must
be 25 ml.
316
iii)
Purge and desorb as described in the method.
C.
Synthetic Organic Chemistry (SOC's)
References for Sections 16.2, 16.7, 17.3 and 17.5 of the Regulations
1.
Regulated Synthetic Organic Chemicals
a.
Methodology
(i)
PWSs conducting analyses of the inorganic chemicals
listed below as required in Section 16.0 and 17.0 of these
regulations shall conduct these analyses in accordance with
one (1) of the following analytical methods or their
equivalent as determined by EPA.
(ii)
Methods 505, 507, 508, 508A, 515.1 and 531.1 are in
“Methods for the Determination of Organic Compounds in
Drinking Water, EPA-600/4-88-039, December 1988,
Revised, July 1991.
(iii)
Methods 506, 547, 550, 550.1 and 551 are in Methods for
the Determination of Organic Compounds in Drinking
Water–Supplement I, EPA-600-4-90-020, July 1990.
(iv)
Methods 515.2, 524.2, 548.1, 549.1, 552.1 and 555 are in
Methods for the Determination of Organic Compounds in
Drinking Water – Supplement II, EPA-600/R-92-129,
August 1992.
(v)
Method 1613 is titled “Tetra-through Octa-Chlorinated
Dioxins and Furans by Isotope-Dilution HRGC/HRMS”,
EPA-821-B-94-005, October 1994.
The documents referenced in items ii to v above are
available from the National Technical Information Service,
NTIS PB91-231480, PB91-146027, PB92-207703 and
PB95-104774, U.S. Department of Commerce, 5285 Port
Royal Road, Springfield, Virginia 22161.
vi)
EPA Methods 504.1, 508.1 and 525.2 are available from
US EPA EMSL, Cincinnati OH 45268.
vii)
Methods 6651 and 6610 shall be followed in accordance
with the 18th edition of Standard Methods for the
317
Examination of Water and Wastewater, 1992, American
Public Health Association. Copies may be obtained from
the American Public Health Association, 1015 Fifteenth
Street NW., Washington DC 299995. Copies may be
inspected at EPA's Drinking Water Docket, 401 M Street,
SW., Washington, DC 20460; or at the Office of the
Federal Register, 800 North Capitol Street, NW., Suite 700,
Washington, DC.
Other analytical test procedures are contained in Technical
Notes on Drinking Water Methods, EPA-600/R-94-173,
October 1994, NTIS PB95-104766. This document also
contains approved analytical methods which will not be
acceptable after July 1, 1996.
Synthetic Organic Chemicals
Contaminant
Method
2,3,7,8-TCDD (dioxin)
1613.
2,4-D
515.2, 555, 515.1.
2,4,5-TP (Silvex)
515.2, 555, 515.1.
Alachlor
5051, 507, 525.2, 508.1.
Atrazine
5051, 507, 525.2, 508.1.
Benzo(a)pyrene
525.2, 550, 550.1.
Carbofuran
531.1, 6610.
Chlordane
505, 508, 525.2, 508.1.
Dalapon
552.1, 515.1.
Di(2-ethylhexyl) adipate
506, 525.2.
Di(2-ethylhexyl) phthalate
506, 525.2.
Dibromochloropropane (DBCP)
504.1, 551.
Dinoseb
515.2, 555, 515.1.
Diquat
549.1.
Endothall
548.1.
Endrin
505, 508, 525.2, 508.1.
Ethylene dibromide (EDB)
504.1, 551.
Glyphosate
547, 6651.
Heptachlor
505, 508, 525.2, 508.1.
Heptachlor Epoxide
505, 508, 525.2, 508.1.
Hexachlorobenzene
505, 508, 525.2, 508.1.
Hexachlorocyclopentadiene
505, 525.2, 508, 508.1.
Lindane
505, 508, 525.2, 508.1.
Methoxychlor
505, 508, 525.2, 508.1.
Oxamyl
531.1, 6610.
PCBs2(as decachlorobiphenyl)
508A.
(as Aroclors)
505, 508.
Pentachlorophenol
515.2, 525.2, 555, 515.1.
Picloram
515.2, 555, 515.1.
Simazine
5051, 507, 525.2, 508.1.
Toxaphene
505, 508, 525.2.
Total Trihalomethanes
502.2, 524.2, 551.
318
Synthetic Organic Chemicals
Contaminant
Method
1 A nitrogen-phosphorous detector should be substituted for the electron capture detector in Method 505 (or another
approved method should be used) to determine alachlor, atrazine and simazine, if lower detection limits are required.
2 PCBs are qualitatively identified as Aroclors and measured for compliance purposes as decachlorobiphenyl.
(ix)
Polychlorinated biphenyls (PCBs) (as decachlorobiphenyl)
aa)
Analysis for PCBs shall be conducted as follows
using either Method 505 or Method 508.
bb)
If PCBs (as one (1) of seven (7) Aroclors) are
detected (as designated in this Paragraph) in any
sample analyzed using Method 505 or 508, the
system shall reanalyze the sample using Method
508A to quantitate PCBs (as decachlorobiphenyl).
Aroclor
Detection limit (mg/L)
1016
0.00008
1221
0.02
1232
0.0005
1242
0.0003
1248
0.0001
1254
0.0001
1260
0.0002
cc)
Compliance with the PCB MCL shall be determined
based upon the quantitative results of analysis using
Method 508A.
b.
Laboratory Criteria
i.
Analysis under this Section shall only be conducted by
laboratories that have received certification by EPA or the
State and have met the following conditions:
(aa)
To receive certification to conduct analyses for the
contaminants in B, (SOC's) above the laboratory
must:
(i-a)
Analyze Performance Evaluation samples
which include those substances provided by
EPA Environmental Monitoring and Support
Laboratory or equivalent samples provided
by the State.
319
(i-b) Achieve quantitative results on the analyses
that are within the following acceptance
limits:
Contaminant
Acceptance Limits (percent)
Alachlor
+45
Aldicarb
2 standard deviations
Aldicarb sulfoxide
2 standard deviations
Aldicarb sulfone
2 standard deviations
Atrazine
+45
Benzo(a)oyrene
2 standard deviations
Carbofuran
+45
Chlordane
+45
Dalapon
2 standard deviations
Di(2-ethylhexyl)adipate
2 standard deviations
Dibromochloropropane (DBCP)
+40
2,3,7,8-TCDD (Dioxin)
2 standard deviations
2,4-D
+50
2,4,5-TP (Silvex)
+50
Di(2-ethylhexyl)phthalate
2 standard deviations
Dinoseb
2 standard deviations
Diquat
2 standard deviations
Endothall
2 standard deviations
Endrin
+30
Ethylene dibromide (EDB)
+40
Glyphosate
2 standard deviations
Heptachlor
+45
Heptachlor epoxide
+45
Hexachlorobenzene
2 standard deviations
Hexachloro-cyclopentadiene
2 standard deviations
Lindane
+45
Methoxychlor
+45
Oxamyl
2 standard deviations
PCBs (as Decachlorobiphenyl)
0-200
Picloram
2 standard deviations
Pentachlorophenol
+50
Simazine
2 standard deviations
Toxaphene
+45
2,4,5-TP (Silvex)
+50
(bb)
Detection shall be defined as greater than or equal
to
the
following
concentrations
for
each
contaminant:
Contaminant
Detection Limit (mg/L)
Alachlor
0.0002
Aldicarb
0.0005
320
Contaminant
Detection Limit (mg/L)
Aldicarb sulfoxide
0.0005
Aldicarb sulfone
0.0008
Atrazine
0.0001
Benzo(a)oyrene
0.00002
Carbofuran
0.0009
Chlordane
0.0002
Dalapon
0.001
Di(2-ethylhexyl)adipate
0.0006
Dibromochloropropane (DBCP)
0.00002
2,3,7,8-TCDD (Dioxin)
0.000000005
2,4-D
0.0001
2,4,5-TP (Silvex)
0.0002
Di(2-ethylhexyl)phthalate
0.0006
Dinoseb
0.0002
Diquat
0.0004
Endothall
0.009
Endrin
0.00001
Ethylene dibromide (EDB)
0.00001
Glyphosate
0.006
Heptachlor
0.00004
Heptachlor epoxide
0.00002
Hexachlorobenzene
0.0001
Hexachloro-cyclopentadiene
0.0001
Lindane
0.00002
Methoxychlor
0.0001
Oxamyl
0.002
PCBs (as Decachlorobiphenyl)
0.0001
Picloram
0.0001
Pentachlorophenol
0.00004
Simazine
0.00007
Toxaphene
0.001
2.
Special Monitoring
a.
Methodology
i)
Systems shall monitor for the unregulated organic
contaminants listed in 16.7 and referenced in Section 17.5
using the method(s) identified below and using the
analytical test procedures contained in Technical Notes on
Drinking Water Methods, EPA-600/R-94-173, October
1994, which is available at NTIS, PB95-104766.
Contaminants
Method
aldicarb
531.1, 6610.
aldicarb sulfone
531.1, 6610.
aldicarb sulfoxide
531.1, 6610.
321
Contaminants
Method
aldrin
505, 508, 525.2, 508.1
butachlor
507, 525.2
carbaryl
531.1, 6610.
dicamba
515.2, 555, 515.1.
dieldrin
505, 508, 525.2, 508.1
3-hydroxycarbofuran
531.1, 6610.
methomyl
531.1, 6610.
metolachlor
507, 525.2, 508.1.
metribuzin
507, 525.2, 508.1.
propachlor
508, 525.2, 508.1.
ii)
Methods 505, 507, 508, 515.1 and 531.1 are in “Methods
for the Determination of Organic Compounds in Drinking
Water, EPA-600/4-88-039, December 1988, Revised, July
1991.
iii)
Methods 515.2, and 555 are in Methods for the
Determination of Organic Compounds in Drinking Water –
Supplement II, EPA-600/R-92-129, August 1992.
The documents referenced in ii & iii above, are available
from the National Technical Information Service, NTIS
PB91-231480, PB91-146027, PB92-207703 and PB95-
104774, U.S. Department of Commerce, 5285 Port Royal
Road, Springfield, Virginia 22161.
iv)
Method 6610 shall be followed in accordance with the
Standard Methods for the Examination of Water and
Wastewater 18th Edition Supplement, 1994, American
Public Health Association. Copies may be obtained from
the American Public Health Association, 1015 Fifteenth
Street NW, Washington, DC 20005. Copies may be
inspected at EPA's Drinking Water Docket, 401 M Street,
SW., Washington, DC 20460; or at the Office of the
Federal Register, 800 North Capitol Street, NW., Site 700,
Washington DC.
v)
EPA Methods 508.1 and 525.2 are available from US EPA
EMSL, Cincinnati OH 45268.
D.
Radiological Chemistry
Reference for Section 16.5 of the Regulations.
1.
The methods specified in Interim Radiochemical Methodology for
Drinking Water, Environmental Monitoring and Support Laboratory,
322
EPA-600/4-75-008, USEPA, Cincinnati, Ohio 45268, or those listed
below, are to be used to determine compliance with S16.5 (radioactivity)
except in cases where alternative methods have been approved by the
Director.
a)
Gross Alpha and Beta-Method 302 “Gross Alpha and Beta
Radioactivity in Water” Standard Methods for the Examination of
Water and Wastewater, 13th Edition, American Public Health
Association, New York, NY.,1971.
b)
Total Radium–Method 304 “Radium in Water by Precipitation”
Ibid.
c)
Radium-226–Method 305 “radium-226 by Radon in Water” Ibid.
d)
Strontium-89, 90 – Method 303 “Total Strontium and Strontium-
90 in Water” Ibid.
e)
Tritium–Method 306 “Tritium in Water” Ibid.
f)
Cesium-134 – ASTM D-2459 “Gamma Spectrometry in Water, “
1975 Annual Book of ASTM Standards, water and Atmospheric
Analysis, Part 31, American Society for Testing and Materials,
Philadelphia, PA. (1975).
g)
Uranium-ASTM D-2907 “Microquantities of Uranium in Water by
Fluorometry,” Ibid.
2.
When the identification and measurement of radionuclides other than
those listed in Paragraph (1) of this Section is required, the following
references are to be used, except in cases where alternative methods have
been approved by the Director.
a)
Procedures for Radiochemical Analysis of Nuclear Reactor
Aqueous Solutions, H.L. Krieger and S. Gold, EPA-R4-73-014.
USEPA, Cincinnati, Ohio, May 1973.
b)
HASL Procedure Manual, Edited by John H. Harley. HASL 300,
ERDA Health and Safety Laboratory, New York, NY., 1973.
3.
For the purpose of monitoring radioactivity concentrations in drinking
water, the required sensitivity of the radioanalysis is defined in terms of a
detection limit. The detection limit shall be that concentration which can
be counted with a precision of plus or minus 100 percent at the 95 percent
confidence level (1.96 σ where σ is the standard deviation of the net
counting rate of the sample).
323
a)
To determine compliance with 16.5 (b) and 16.5 (c), the detection
limit shall not exceed the concentrations in Table B.
TABLE B–Detection Limits for Gross Alpha Particle Activity, Radium 226, Radium 228, and Uranium
Contaminant
Detection Limit
Gross alpha particle activity
3 pCi/l
Radium-226
1 pCi/l
Radium-228
1 pCi/l
Uranium
1 ug/L
For monitoring data collected prior to December 8, 2003, the
following detection limits apply: gross alpha particle activity 1.5
pCi/L, combined radium 0.5 pCi/L, uranium- none applicable.
b)
To determine compliance with 16.5 (d), Man-made Beta Particle
and Photon Emitters, the detection limits shall not exceed the
concentrations listed in Table C.
TABLE C–Detection Limits for Man-made Beta Particle and Photon Emitters
Radionuclide
Detection Limit
Tritium
1,000 pCi/l
Strontium-89
10 pCi/l
Strontium-90
2 pCi/l
Iodine-131
1 pCi/l
Cesium-134
10 pCi/l
Gross beta
4 pCi/l
Other radionuclides
1/10 of the applicable limit
c)
To judge compliance with the maximum contaminant levels listed
in 16.5 (b), (c), and (d), averages of data shall be used and shall be
rounded to the same number of significant figures as the maximum
contaminant level for the substance in question.
324
APPENDIX 2
RESERVED
325
APPENDIX 3
DWQ PENALTY MATRIX (1)
The Division of Drinking Water Quality has classified its regulations into the following three (3)
categories for use when assessing Administrative Penalties:
Categories*
Category I Penalty Range $1,000 - $5,000/day/violation
These types of violation have a direct impact on public health and will be given a
high priority.
Î
Exceeding any MCL including
Bacteria
Inorganic
Pesticides/Organic
Turbidity
Radiological
Î
Failure to maintain required chlorine residual
Î
Failure to adhere to new source approval requirements/plan requirements
Category II Penalty Range $100 - $1000/day/violation
These types of violations/noncompliance, also have a direct impact on public
health but are mainly noncompliance with technical safeguards.
Î
Failure to monitor as required
Î
Failure to comply with reporting requirements
Î
Failure to make public notice as required
Î
Failure to notify DWQ within 48 hrs after confirmation check samples
reveal MCL violations
Î
Denial of right of entry provisions
Î
Failure to comply with operators certification requirements
Category III Penalty Range $100 - $300/day/violation
These types of violations have an indirect impact on public health and are
generally related to poor record keeping.
Î
Failure to submit monitoring reports (monitoring was done but system did
not send report to DWQ until it was requested)
Î
Late submittal of monitoring reports
Î
Failure to keep required records on file as required
326
* Violation of a Department Order is a separate and additional violation from the violation or
violations which gave rise to the issuance of the order, and is given a Base Number of $1000.00.
No distinction should be made between a unilateral order and a consent order for the purpose of
assessing administrative penalties.
The above classification is subject to change as the Division gets more experience with the
Administrative Penalties regulations. The Legal office will be kept informed of all changes.
327
DWQ PENALTY MATRIX (2)
Sections R46-13-DWQ
Noncompliance
Categories
1.0
Definitions
N/A
2.0
Coverage
I
a)
Approval required
I
c)
Right of entry
II
3.0
New Water Sources
I
4.0
Approval of Treatment Works, Storage and Pumping Facilities
I
5.0
Filtration and Disinfection
5.1
General Requirements
5.2
Criteria for avoiding filtration
II
5.3
Disinfection
I
5.4
Filtration
I
5.5
Analytical and monitoring requirements
II
5.6
Monitoring requirements for systems that do not provide filtration
II
5.7
Monitoring requirements for systems using filtration equipment
II
5.8
Reporting and record keeping requirements
III
6.0
Control of Lead and Copper
6.80
General requirements
6.81
Applicability of corrosion control treatment steps to small, medium-size
I
and large water systems
6.82
Description of corrosion control treatment requirements
I
6.83
Source water treatment requirements
I
6.84
Lead service line replacement requirements
I
6.85
Public education and supplemental monitoring requirements
II
6.86
Monitoring requirements for lead and copper in tap water
II
6.87
Monitoring requirements for water quality parameters
II
6.88
Source monitoring requirements for lead and copper in water
II
6.89
Analytical methods
II
6.90
Reporting requirements
III
6.91
Record keeping requirements
III
7.0
Connections Between Distribution Systems
II
8.0
Contamination of Tanks
I
328
8.1
Tanks Connected to Unsafe Supplies
8.2
Avoidance of Contamination in Tanks
9.0
Assurance of Safety in Public Supply
II
329
Sections R46-13-DWQ
Noncompliance
Categories
10.0 Correction of Unsafe Conditions
I
11.0 Reports as to Public Supplies
II
12.0 Certified Laboratories
II
13.0 Ground Water Microbiology
II
14.0 Consecutive Water System Monitoring
N/A
15.0 Variances and Exemptions
N/A
16.0 Community Water System Requirements
I
Maximum Contaminant levels for 16.1 Inorganic Chemicals
16.2 Organic Chemicals
16.3 Turbidity
16.4 Microbiological
16.5 Radioactivity
Monitoring Requirements, Analytical Techniques, and Monitoring
II
Frequency for 16.1, 16.2, 16.3, 16.4, 16.5, 16.6 and 16.7
16.8 Public Notification
II
16.9 Records
III
17.0 Non-Community Water System Requirements
I
Maximum Contaminant levels for
17.1 Microbiological
17.2 Inorganic Chemicals
17.3 Organic Chemicals
17.4 Turbidity
Monitoring Requirements, Analytical Techniques and
II
Monitoring Frequency for 17.1, 17.2, 17.3, 17.4 and 17.5
17.6 Public Notification
II
17.7 Records
III
330
DWQ PENALTY MATRIX (3)
18.0
Fee Schedule
N/A
19.0
Rules Governing Practices and Procedures
N/A
20.0
Violations, Noncompliance, and Enforcement
21.0
Severability
N/A
Other Areas of Non-Compliance
Violations of approval letter requirements
I
Contamination incidents
I
Non-compliance with orders
I
331
DWQ PENALTY MATRIX (4)
PWSS Civil or Complaint for Penalty Calculation Work Sheet
PWS Name or Owner Name
DATE / /
PWS ID#
LOCATION
Violation Cited
I.
Calculate Statutory Maximum Penalty
(A)
Length of Violation (in days)
(B)
Maximum Penalty
________
Civil Penalty - $5,000/day
Statutory Maximum Penalty
II.
Calculate Economic Benefit Component
1.
Estimate avoided and delayed costs through
reasonable methodology.
This must be documented.
III.
Calculate Gravity Component
2.
BASE NUMBER
________
3.
Impact (+ or -)
________
4.
Extent (+ or -)
________
5.
# of Violations (+ or -)
________
6.
GRAVITY BASE NUMBER
(Total lines 2,3,4 and 5)
________
**(Total must be within class range)
________
7.
NUMBER OF DAYS (If applicable . . .)
(Must be at least one (1))
________
8.
TOTAL GRAVITY BASE NUMBER
(Multiply 6 by 7)
________
9.
PRELIMINARY SETTLEMENT AMOUNT
(Economic Benefit + Gravity Component
________
IV.
Adjustment Factors TO TOTAL GRAVITY BASE NUMBER
10.
History of Violations
(+)
0 to 50% ...
_____%
11.
Lack of Good Faith
(+)
0 to 100% ... _____%
12.
Financial Condition
(+ or -)
0 to 50% ...
_____%
13.
Public Interest
(+ or -)
0 to 50% ...
_____%
14.
Special Circumstances
(+ or -)
0 to 50% ...
_____%
15.
Litigation Considerations
(-)
0 to 90% ...
_____%
TOTAL PERCENTAGE ADJUSTMENTS
16.
(Add lines 10 thru 15)
_____%
17.
MULTIPLY LINE 16 BY LINE 8
_____$
332
18.
Enforcement Costs
_____$
V.
Final Settlement Amount
19.
TOTAL PENALTY (Add lines 1,8,17 and 18)
_____$
COMMENTS (Briefly note reason for any adjustments)
333
APPENDIX 4
List of Potential Sources of Groundwater Contamination
•
Agricultural related activities (pesticide and fertilizer storage and application,
machinery maintenance and fueling
•
Airports-commercial (maintenance and repair, fuel storage)
•
Animal care and holding areas (stables, kennels, pet shops)
•
Asphalt, coal, tar and concrete companies
•
Automotive repair shops
•
Automotive body shops
•
Auto parts stores
•
Beauty salons
•
Boat builders and refinishers
•
Bus and truck terminals
•
Chemical manufacturers
•
Construction sites
•
Dredge disposal sites
•
Dry cleaners
•
Food processors (meat packers, dairies, bakeries)
•
Fuel oil distributors (product storage, equipment maintenance and storage)
•
Funeral homes and cemeteries
•
Furniture strippers, refinishers
•
Golf courses
•
Hotels and motels
•
Industrial manufacturers
•
Junkyard and salvage yards
•
Land application of sewage sludge
•
Landfills and dumps
•
Laundromats
•
Machine shops
•
Medical facilities (hospitals, clinics, laboratories)
•
Metal and drum cleaning/reconditioning
•
Military facilities (past and present)
•
Nurseries
•
Nursing homes
•
Paint shops
•
Photographic processors
•
Pipelines (oil and sewer)
•
Printers and blueprint shops
•
Prisons
•
Railroad yards
•
Repair shops (engines, appliances, etc.)
•
Research laboratories
•
Residential development (lawn care, septic systems)
•
Restaurants and taverns
334
•
Retail shopping centers, malls
•
Road salt storage
•
Rust proofers
•
Sand and gravel mining operations
•
Sawmills
•
Schools, colleges and trade centers
•
Service stations (gas stations)
•
Storm water management facilities (leaching systems)
List of Potential Sources of Groundwater Contamination
•
Transmission line rights of way
•
Transportation corridors (road deicing, materials transport)
•
Utility substations/transformers
•
Waste storage, treatment and recycling (hazardous and non-hazardous)
•
Water transfer stations
•
Wastewater treatment plants (past or present sludge disposal)
•
Wood preservers