1370-02-.21
Records, Services And Tests
Cite as Tenn. Comp. R. & Regs. 1370-02-.21
(1)
Records
(a)
Client’s Record. The record for each client shall include a case history, the results of all
required tests and a legend (key) explaining the symbols of results used in the
audiogram. Records of the required test and dispensing procedures shall be
maintained as a part of the client’s file and shall show also the model number or name,
serial number, type of circuit response or any adjustment made on instruments that
have adjustable circuits. Such records shall be maintained and preserved for a
minimum of three (3) years after the last entry in the records.
(b)
Orientation Program. Every hearing aid client shall be offered at the site of the sale a
personalized counseling program provided by the HIS or an individual licensed by the
Council. The counseling program shall be provided with each new fitting and designed
to ensure that proper use and operation of the instrument are understood by the client.
(c)
Earmold. Every air conduction fitting that requires an earmold shall include a properly
fitting custom earmold except in cases where, in the dispenser’s judgment, the client
would be better fitted without the use of a new mold. In special fittings where an
earmold is not required, such shall be noted in the record.
(d)
Examination of the Ear. The dispenser shall make an examination of the client’s
external ears by means of an otoscope, or other instrument designed for that purpose,
before any audiometric test is made and before and after the making of an earmold
impression. Any abnormality detected by this inspection must be reported to the client
and the client advised to seek medical advice and this shall be recorded on the test
record.
(2)
Bill of Sale for Hearing Equipment - To insure that appropriate follow up services are
available to consumers in Tennessee, licensees shall deliver to each person supplied with a
hearing instrument a bill of sale which contains each of the following:
GENERAL RULES GOVERNING HEARING INSTRUMENT SPECIALISTS
CHAPTER 1370-02
(a)
The signature of the licensed hearing instrument specialist, the address of the
specialist’s regular place of business, the specialist’s license number with an expiration
date.
(b)
The make and model of the hearing instrument supplied and the amount charged
including a description, i.e., new, used, rebuilt.
(c)
Clear statement of terms of sale including provisions required by T.C.A. § 63-17-217.
(d)
Name and address of the Council.
(e)
Statement that the Council will receive complaints on any matter relating to the fitting
and dispensing of hearing instruments.
(f)
If the licensee has no location in Tennessee, an outline of follow up services must be
made available to Tennessee residents who purchase the dealer’s services within
Tennessee.
(3)
Tests
(a)
Puretone Audiometry.
1.
Thresholds shall be determined for each ear with air and bone conducted
puretone stimuli. Air conduction stimuli must include 0.5, 1, 2, and 4K Hertz.
Bone conduction stimuli must include .05, 1, 2, and 4K Hertz. On a client whose
bone conduction threshold has been previously charted and that chart is a part of
the client’s permanent record, the licensee, at his discretion, may omit any new
bone conduction testing, if there has been no appreciable change in the air
conduction thresholds.
2.
Appropriate masking must be used with air and bone conducted stimuli whenever
necessary to obtain valid results.
(b)
Discrimination Tests. There shall be a measurement of the client’s speech
discrimination ability at each ear using widely recognized test material prior to fitting.
Appropriate masking must be used with speech stimuli whenever necessary to obtain
valid results.
(c)
Post-Fitting Tests. There shall be a measurement of the client’s hearing ability using
widely recognized test material while the client is wearing the selected hearing aid(s)
adjusted for the selection gain, M.P. output, and frequency response.
(d)
Tests for Tolerance. There shall be a measurement of the client’s tolerance for
amplified auditory stimuli using speech, puretones, or a complex signal representative
of the speech spectrum (0.5, 1 and 2K Hz) or a method deemed acceptable by the
Council.
(e)
Omitted Tests.
1.
The required tests may be omitted only under the most unusual circumstances,
such as the inability of the client to perform the task.
2.
When a fitting is being made upon referral, the specified tests may be omitted if
the referring source has forwarded to the dispenser copies of those test results.
However, these tests must have been performed within ninety (90) days. The
GENERAL RULES GOVERNING HEARING INSTRUMENT SPECIALISTS
CHAPTER 1370-02
licensed dispenser who omits specified tests and accepts tests performed by
referring professional shall orally inform and conspicuously set forth in the sales
contract that the hearing aid dispenser is not responsible for the validity and
accuracy of the tests.