WAC 246-225-100
WAC 246-225-100. Radiographic systems—Standby radiation from capacitor energy storage equipment
Radiation emitted from the X-ray tube when the exposure switch or timer is not activated shall not exceed a rate of 2 milliroentgens per hour at 5 centimeters from any accessible surface of the diagnostic source assembly, with the beam-limiting device fully open.