WAC 246-226-020

WAC 246-226-020. Equipment requirements

Last amended: 2017Year: 2026Length: 211 wordsOfficial source
The CT X-ray system must: (1) Meet the requirements of 21 C.F.R. Sec. 1020.33 (last amended June 10, 2005) at the time of installation and while the CT X-ray system is registered with the department under chapter 246-224 WAC; and (2) Be equipped: (a) With a visible signal that indicates when the X-ray exposure is occurring; (b) So that the operator can immediately terminate an X-ray exposure of greater than one-half second duration at any time during the X-ray exposure; (c) So that the parameters used during a CT procedure are: (i) Displayed prior to beginning a scan; and (ii) Visible by the operator from any location scanning can be initiated. (d) So that radiation leakage from the tube port does not exceed limits established in WAC 246-225-040 (3) and (4) when data are not being collected for image production; (e) So that the accuracy of the laser or optical positioning system is within five millimeters maximum deviation on the axial position (z-axis); (f) With an X-ray production indicator of at least one-half second at or near the gantry that is visible from any point outside the gantry opening; (g) So that premature termination of the X-ray exposure by the operator requires resetting of the parameters prior to initiating another scan.
WAC 246-226-020: WAC 246-226-020. Equipment requirements | Justis AI