WAC 246-226-050
WAC 246-226-050. Dose limits
The CTDI vol for the following CT procedure on phantoms may not exceed the dose limits established in Table 1.
Table 1: Dose limits
CT Procedure Phantom Size Dose Limit: CTDI vol (mGy) Adult head 16 cm 80 Adult abdomen 32 cm 30 Pediatric head (one year old) 16 cm 40 Pediatric abdomen (40 pounds) 16 cm 20
CT Procedure
Phantom Size
Dose Limit:
CTDI vol (mGy)
Adult head
16 cm
80
Adult abdomen
32 cm
30
Pediatric head (one year old)
16 cm
40
Pediatric abdomen
(40 pounds)
16 cm
20