WAC 246-240-396. Therapy-related computer systems

Last amended: 2006Year: 2026Length: 108 wordsOfficial source
The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of: (1) The source-specific input parameters required by the dose calculation algorithm; (2) The accuracy of dose, dwell time, and treatment time calculations at representative points; (3) The accuracy of isodose plots and graphic displays; (4) The accuracy of the software used to determine sealed source positions from radiographic images; and (5) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.