HAR §11-45-116
HAR §11-45-116. General requirements
Cite as Haw. Code R. § 11-45-116
(a) The licensee shall
be responsible for directing the operation of the x-ray system(s)
under the licensee's administrative control. The licensee or the
licensee's agent shall assure that the requirements of this
subchapter are met in the operation of the x-ray system(s).
(1)
An x-ray system which does not meet the provisions of
this subchapter shall not be operated for diagnostic
purposes if so directed by the department.
(2)
Individuals who will be operating x-ray systems shall
be adequately instructed on the operating procedures
and be competent in the safe use of the equipment.
(3)
Individuals practicing radiologic technology shall be
licensed under chapter 11-44.
(4)
A chart shall be provided in the vicinity of the
diagnostic x-ray system's control panel which
specifies, for all examinations performed with that
system, the following information:
(A)
Patient's body part and anatomical size, or body
part thickness, or age (for pediatrics), versus
technique factors to be utilized;
(B)
Type and size of the film or screen-film
combination to be used;
(C)
Type and focal distance of the grid to be used,
if any;
(D)
Source-image receptor distance to be used
(except for dental intraoral radiography);
(E)
Type and location of placement of patient
shielding (e.g., gonad, etc.) to be used; and
(F)
For mammography, indication of kilovoltage
peak/target/filter combination.
(5)
The licensee of a facility shall establish and make
available to x-ray operators written safety
procedures, including patient holding and any
restrictions of the operating technique required for
the safe operation of the particular x-ray system.
The operator shall be able to demonstrate familiarity
with these procedures.
§11-45-116
45-88
(6)
Except for human patients who cannot be moved out of
the room, only the staff, ancillary personnel, or
other persons required for the medical procedure or
training shall be in the room during the radiographic
exposure. Other than the patient being examined:
(A)
All individuals shall be positioned such that no
part of the body will be struck by the useful
beam unless protected by not less than 0.5
millimeter lead equivalent material or otherwise
authorized by a qualified medical physicist.
(B)
The x-ray operator, other staff, ancillary
personnel and other persons required for the
medical procedure shall be protected from the
direct scatter radiation by protective aprons or
whole body protective barriers of not less than
0.25 millimeter lead equivalent material, unless
otherwise authorized by a qualified medical
physicist.
(C)
Human patients who cannot be removed from the
room shall be protected from the direct scatter
radiation by whole body protective barriers of
not less than 0.25 millimeter lead equivalent
material, or shall be so positioned that the
nearest portion of the body is at least two
meters from both the tube head and the nearest
edge of the image receptor, unless otherwise
authorized by a qualified medical physicist.
(7)
Gonad shielding of not less than 0.5 millimeter lead
equivalent material shall be used for human patients,
who have not passed the reproductive age, during
radiographic procedures in which the gonads are in the
useful beam, except for cases in which this would
interfere with the diagnostic procedure.
(8)
Individuals shall not be exposed to the useful beam
except for healing arts purposes and unless such
exposure has been authorized by a licensed
practitioner of the healing arts. This provision
specifically prohibits deliberate exposure for the
following purposes:
(A)
Exposure of an individual for training,
demonstration, or other non-healing arts
purposes; and
(B)
Reserved.
(9)
When a patient or film requires auxiliary support
during a radiation exposure:
(A)
Mechanical holding devices shall be used. The
written safety procedures, required by paragraph
(5), shall list individual projections where
holding devices cannot be utilized;
(B)
Written safety procedures, as required by
paragraph (5), shall indicate the requirements
for selecting a holder and the procedure the
holder shall follow;
(C)
The human holder shall be instructed in personal
§11-45-116
45-89
radiation safety and protected as required by
paragraph (6);
(D)
No individual shall be used routinely to hold
film or human patients;
(E)
In those cases where an individual must hold the
film (except during intraoral examinations), any
portion of the body other than the area of
clinical interest struck by the useful beam
shall be protected by not less than 0.5
millimeter lead equivalent material; or
(F)
Each facility shall have leaded aprons and
gloves available in sufficient numbers to
provide protection for all personnel who are
involved with x-ray operations and who are
otherwise not shielded.
(10)
Procedures and auxiliary equipment designed to
minimize patient and personnel exposure commensurate
with the needed diagnostic information shall be
utilized.
(A)
The speed of the screen and film combinations
used shall be the fastest speed consistent with
the diagnostic objective of the examinations.
Film cassettes without intensifying screens
shall not be used for any routine diagnostic
radiological imaging, with the exception of
standard film packets for intraoral use in
dental radiography.
(B)
The radiation exposure to the patient shall be
the minimum exposure required to produce images
of good diagnostic quality.
(C)
Portable or mobile x-ray equipment shall be used
only for examinations where it is impractical to
transfer the patient(s) to a stationary x-ray
installation.
(D)
If grids are used between the patient and the
image receptor to decrease scatter to the film
and improve contrast the grid shall:
(i)
Be positioned properly, i.e., tube side
facing the right direction, and grid
centered to the central ray; and
(ii)
If of the focused type, be of the proper
focal distance for the SID's being used.
(11)
All individuals who are associated with the operation
of an x-ray system are subject to the requirements of
subchapter 4.
(12)
Reserved.
(13)
The licensee shall maintain the following information
for each x-ray system for inspection by the
department:
(A)
Model and serial numbers of all major
components, and user's manuals for those
components;
(B)
Tube rating charts and cooling curves;
(C)
Records of surveys, calibrations, maintenance,
§11-45-116
45-90
and modifications performed on the x-ray
system(s);
(D)
Copy of all correspondence with this department
regarding that x-ray system.
(14)
Each facility shall maintain either an x-ray log
containing the name of the patient, the type of x-ray
examination, the date of the x-ray examination, the
name of the referring practitioner, and the name of
the individual who performed the x-ray examination
(facilities subject to chapter 11-44), or a patient
record specifying the type of x-ray examination and
the date of the x-ray examination. When the patient
or film must be provided with human auxiliary support,
the name of the human holder shall be recorded.
(b)
Each installation using a radiographic x-ray system
and using analog image receptors (radiographic film) shall have
available suitable equipment for handling and processing
radiographic film in accordance with the following provisions:
(1)
Manually developed film:
(A)
Processing tanks shall be constructed of
mechanically rigid, corrosion resistant
material; and
(B)
The temperature of solutions in the tanks shall
be maintained within the range of sixty degrees
Fahrenheit to eighty degrees Fahrenheit (sixteen
degrees Celsius to twenty-seven degrees
Celsius). Film shall be developed in accordance
with the time-temperature relationships
recommended by the film manufacturer, or, in the
absence of such recommendations, with Appendix A
of this subchapter.
(C)
Devices shall be utilized which will indicate
the actual temperature of the developer and
signal the passage of a preset time appropriate
to the developing time required.
(2)
Automatic processors and other closed processing
systems:
(A)
Films shall be developed in accordance with the
time-temperature relationships recommended by
the film manufacturer; in the absence of such
recommendations, the film shall be developed
according to Appendix B of this subchapter.
(B)
The specified developer temperature and
immersion time shall be posted in the darkroom
or on the automatic processor.
(3)
Processing deviations from the requirements of
paragraph (1) shall be documented by the licensee in
such manner that the requirements are shown to be met
or exceeded (e.g., extended processing, and special
rapid chemistry).
(c)
Other requirements:
(1)
Pass boxes, if provided, shall be so constructed as to
exclude light from the darkroom when cassettes are
placed in or removed from the boxes, and shall
§11-45-117
45-91
incorporate adequate shielding from stray radiation to
prevent exposure of undeveloped film.
(2)
The darkroom shall be light tight and use proper
safelighting such that any film type in use exposed in
a cassette to x-radiation sufficient to produce an
optical density from 1.0 to 2.0 when processed shall
not have a fog density greater than 0.1 (0.05 for
mammography) when exposed in the darkroom for two
minutes with all safelights on. If used, daylight
film handling boxes shall preclude fogging of the
film.
(3)
Darkrooms typically used by more than one individual
shall be provided a method to prevent accidental entry
while undeveloped films are being handled or
processed.
(4)
Film shall be stored in a cool, dry place and shall be
protected from exposure to stray radiation. Film in
open packages shall be stored in a light tight
container.
(5)
Film cassettes and intensifying screens shall be
inspected periodically and shall be cleaned and
replaced as necessary to best assure radiographs of
good diagnostic quality.
(6)
Outdated x-ray film shall not be used for human
diagnostic radiographs, unless the film has been
stored in accordance with the manufacturer's
recommendations and a sample of the film passes a
sensitometric test for normal ranges of base plus fog
and speed.
(7)
Film developing solutions shall be prepared in
accordance with the directions given by the
manufacturer, and shall be maintained in strength by
replenishment or renewal so that full development is
accomplished within the time specified by the
manufacturer.
(d)
The exhibits at the end of chapter 11-45 entitled,
"Appendix A of Subchapter 6, Time-Temperature Chart (2/2/93)" and
"Appendix B of Subchapter 6, Temperature-Immersion Chart
(2/2/93)", are made a part of this section.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)