HAR §11-45-117
HAR §11-45-117. All diagnostic x-ray systems
Cite as Haw. Code R. § 11-45-117
(a) The control
panel containing the main power switch shall bear the warning
statement, legible and accessible to view: "WARNING: This x-ray
unit may be dangerous to patient and operator unless safe exposure
factors and operating instructions are observed."
(b)
On battery-powered x-ray generators, visual means
shall be provided on the control panel to indicate whether the
battery is in a state of charge adequate for proper operation.
(c)
The leakage radiation from the diagnostic source
assembly measured at a distance of one meter in any direction from
the source shall not exceed one hundred milliroentgens (25.8
§11-45-117
45-92
microcoulombs per kilogram) in one hour when the x-ray tube is
operated at its leakage technique factors. Compliance shall be
determined by measurements averaged over an area of one hundred
square centimeters with no linear dimension greater than
twenty centimeters.
(d)
The radiation emitted by a component other than the
diagnostic source assembly shall not exceed two milliroentgens
(0.516 microcoulombs per kilogram) in one hour at five centimeters
from any accessible surface of the component when it is operated
in an assembled x-ray system under any conditions for which it was
designed. Compliance shall be determined by measurements averaged
over an area of one hundred square centimeters with no linear
dimension greater than twenty centimeters.
(e)
Beam quality:
(1)
The half-value layer of the useful beam for a given x-
ray tube potential shall not be less than the
appropriate value specified in Appendix C of this
subchapter under "specified dental systems," for any
dental system designed for use with intraoral image
receptors and manufactured after December 1, 1980; and
under "other x-ray systems," for all other x-ray
systems subject to this section. If it is necessary
to determine such half-value layer at an x-ray tube
potential which is not listed in Appendix C, linear
interpolation or extrapolation may be made. Positive
means shall be provided to insure that at least the
minimum filtration needed to achieve the above beam
quality requirements is in the useful beam during each
exposure.
(2)
For capacitor energy storage equipment, compliance
with the requirements of paragraph (1) shall be
determined with the maximum quantity of charge per
exposure. This shall be deemed to have been met if an
milliamperes-second of ten has been used.
(3)
The required minimal half-value layer of the useful
beam shall include the filtration contributed by all
materials which are permanently between the source and
the patient.
(4)
For x-ray systems which have variable kilovoltage peak
and variable filtration for the useful beam, a device
shall link the kilovolts peak selector with the
filter(s) and shall prevent an exposure unless the
minimum amount of filtration necessary to produce the
half-value layer required by paragraph (1) is in the
useful beam for the given kilovolts peak which has
been selected.
(f)
Where two or more radiographic tubes are controlled by
one exposure switch, the tube or tubes which have been selected
shall be clearly indicated before initiation of the exposure.
This indication shall be both on the x-ray control panel and at or
near the tube housing assembly which has been selected.
(g)
The tube housing assembly supports shall be adjusted
such that the tube housing assembly shall remain stable during an
exposure unless tube housing movement is a designed function of
§11-45-118
45-93
the x-ray system.
(h)
Technique indicators:
(l)
The technique factors to be used during an exposure
shall be indicated before the exposure begins. If
automatic exposure controls are used, the technique
factors which are set before the exposure shall be
indicated.
(2)
The requirement of paragraph (l) may be met by
permanent markings on equipment having fixed technique
factors. Indication of technique factors shall be
visible from the operator's position except in the
case of spot films made by the fluoroscopist.
(i)
Diagnostic x-ray systems and their associated
components used on humans and certified pursuant to the federal x-
ray equipment performance standard, 21 C.F.R. Part 1020, shall be
maintained in compliance with applicable requirements of that
standard and this subchapter.
(j)
All position locking, holding, and centering devices
on x-ray system components and systems shall function as intended.
(k)
The exhibit at the end of chapter 11-45 entitled,
"Appendix C of Subchapter 6, Half-Value Layer (7/25/96)", is made
a part of this section. [Eff 11/12/99] (Auth: HRS §§321-10,
321-11, 321-71) (Imp: HRS §§321-1, 321-11(21), 321-71)