HAR §11-45-118
HAR §11-45-118. Fluoroscopic x-ray systems
Cite as Haw. Code R. § 11-45-118
(a) All
fluoroscopic x-ray systems shall be image intensified or used in
conjunction with a digital imaging system.
(b)
The useful beam shall be limited as follows:
(1)
Primary barrier:
(A)
The fluoroscopic imaging assembly shall be
provided with a primary protective barrier which
intercepts the entire cross section of the
useful beam at any source-image receptor
distance; and
(B)
The x-ray tube used for fluoroscopy shall not
produce x-rays unless the barrier is in position
to intercept the entire useful beam.
(2)
Fluoroscopic beam limitation:
(A)
For certified fluoroscopic systems with or
without a spot film device, neither the length
nor the width of the x-ray field in the plane of
the image receptor shall exceed that of the
visible area of the image receptor by more than
three percent of the source-image receptor
distance. The sum of the excess length and the
excess width shall be no greater than four
percent of the source-image receptor distance.
(B)
For uncertified fluoroscopic systems with a spot
film device, the x-ray beam with the shutters
fully opened (during fluoroscopy or spot
filming) shall be no larger than the largest
spot film size for which the device is designed.
Measurements shall be made at the minimum
§11-45-118
45-94
source-image receptor distance available but at
no less than twenty centimeters table top to the
film plane distance.
(C)
For uncertified fluoroscopic systems without a
spot film device, the requirements of
subparagraph (A) apply.
(D)
Other requirements for fluoroscopic beam
limitation:
(i)
Means shall be provided to permit further
limitation of the field. Beam-limiting
devices manufactured after May 22, 1979,
and incorporated in equipment with a
variable source-image receptor distance
and/or a visible area of greater than
three hundred square centimeters shall be
provided with means for stepless
adjustment of the x-ray field;
(ii)
All equipment with a fixed source-image
receptor distance and a visible area of
three hundred square centimeters or less
shall be provided with either stepless
adjustment of the x-ray field or with
means to further limit the x-ray field
size at the plane of the image receptor to
one hundred twenty-five square centimeters
or less;
(iii)
If provided, stepless adjustment shall, at
the greatest source-image receptor
distance, provide continuous field sizes
from the maximum attainable to a field
size of five centimeters by five
centimeters or less;
(iv)
For equipment manufactured after February
25, 1978, when the angle between the image
receptor and beam axis is variable, means
shall be provided to indicate when the
axis of the x-ray beam is perpendicular to
the plane of the image receptor; and
(v)
For rectangular x-ray fields used with
circular image reception, the error in
alignment shall be determined along the
length and width dimensions of the x-ray
field which pass through the center of the
visible area of the image receptor.
(E)
Spot-film devices which are certified components
shall meet the following additional
requirements:
(i)
Means shall be provided between the source
and the patient for adjustment of the x-
ray field size in the plane of the film to
the size of that portion of the film which
has been selected on the spot film
selector. Such adjustment shall be
automatically accomplished except when the
§11-45-118
45-95
x-ray field size in the plane of the film
is smaller than that of the selected
portion of the film. For spot film
devices manufactured after June 21, 1979,
if the x-ray field size is less than the
size of the selected portion of the film,
the means for adjustment of the field size
shall be only at the operator's option;
(ii)
It shall be possible to adjust the x-ray
field size in the plane of the film to a
size smaller than the selected portion of
the film. The minimum field size at the
greatest source-image receptor distance
shall be equal to, or less than, five
centimeters by five centimeters;
(iii)
The center of the x-ray field in the plane
of the film shall be aligned with the
center of the selected portion of the film
to within two percent of the source-image
receptor distance; and
(iv)
On spot-film devices manufactured after
February 25, 1978, if the angle between
the plane of the image receptor and beam
axis is variable, means shall be provided
to indicate when the axis of the x-ray
beam is perpendicular to the plane of the
image receptor, and compliance shall be
determined with the beam axis indicated to
be perpendicular to the plane of the image
receptor.
(F)
If a means exists to override any of the
automatic x-ray field size adjustments required
in subparagraph (E), that means:
(i)
Shall be designed for use only in the
event of system failure;
(ii)
Shall incorporate a signal visible at the
fluoroscopist's position which shall
indicate whenever the automatic field size
adjustment is overridden; and
(iii)
Shall be clearly and durably labeled as
follows:
FOR X-RAY FIELD
LIMITATION SYSTEM FAILURE
(c)
X-ray production in the fluoroscopic mode shall be
controlled by a device which requires continuous pressure by the
fluoroscopist for the entire time of any exposure. When recording
serial fluoroscopic images, the fluoroscopist shall be able to
terminate the x-ray exposure(s) at any time, but means may be
provided to permit completion of any single exposure of the series
in process.
(d)
Entrance exposure rate allowable limits:
(1)
Fluoroscopic equipment which is provided with
§11-45-118
45-96
automatic exposure rate control shall not be operable
at any combination of tube potential and current which
shall result in an exposure rate in excess of ten
roentgens (2.58 millicoulombs per kilogram) per minute
at the point where the center of the useful beam
enters the patient, except:
(A)
During recording of fluoroscopic images, or
(B)
When an optional high level control is provided.
When so provided, the equipment shall not be
operable at any combination of tube potential
and current which shall result in an exposure
rate in excess of five roentgens (1.29
millicoulombs per kilogram) per minute at the
point where the center of the useful beam enters
the patient unless the high level control is
activated. Special means of activation of high
level controls shall be required. The high
level control shall only be operable when
continuous manual activation is provided by the
operator. A continuous signal audible to the
fluoroscopist shall indicate that high level
control is being employed.
(2)
Fluoroscopic equipment which is not provided with
automatic exposure rate control (manual mode) shall
not be operable at any combination of tube potential
and current which shall result in an exposure rate in
excess of five roentgens (1.29 millicoulombs per
kilogram) per minute at the point where the center of
the useful beam enters the patient, except:
(A)
During recording of fluoroscopic images, or
(B)
When an optional high level control is
activated. Special means of activation of high
level controls shall be required. The high
level control shall only be operable when
continuous manual activation is provided by the
operator. A continuous signal audible to the
fluoroscopist shall indicate that the high level
control is being employed.
(3)
Fluoroscopic equipment which is provided with both
automatic exposure rate control mode and a manual mode
shall not be operable at any combination of tube
potential and current which shall result in an
exposure rate in excess of ten roentgens (2.58
millicoulombs per kilogram) per minute in either mode
at the point where the center of the useful beam
enters the patient, except:
(A)
During recording of fluoroscopic images; or
(B)
When the mode or modes have an optional high
level control, in which case that mode or modes
shall not be operable at any combination of tube
potential and current which shall result in an
exposure rate in excess of five roentgens (1.29
millicoulombs per kilogram) per minute at the
point where the center of the useful beam enters
§11-45-118
45-97
the patient, unless the high level control is
activated. Special means of activation of high
level controls shall be required. The high
level control shall only be operable when
continuous manual activation is provided by the
operator. A continuous signal audible to the
fluoroscopist shall indicate that the high level
control is being employed.
(4)
Any fluoroscopic equipment manufactured after May 19,
1995 which can exceed five roentgens (1.29
millicoulombs per kilogram) per minute shall be
equipped with an automatic exposure rate control. All
entrance exposure rate limits shall be ten roentgens
(2.58 millicoulombs per kilogram) per minute with an
upper limit of twenty roentgens (5.16 millicoulombs
per kilogram) per minute when high level control is
activated.
(5)
Compliance with the requirements of this subsection
shall be determined as follows:
(A)
If the source is below the x-ray table, the
exposure rate shall be measured one centimeter
above the tabletop or cradle.
(B)
If the source is above the x-ray table, the
exposure rate shall be measured at thirty
centimeters above the tabletop with the end of
the beam-limiting device or spacer positioned as
closely as possible to the point of measurement.
(C)
For a C-arm type of fluoroscope, the exposure
rate shall be measured thirty centimeters from
the input surface of the fluoroscopic imaging
assembly, with the source positioned at any
available SID, provided that the end of the
beam-limiting device or spacer is no closer than
thirty centimeters from the input surface of the
fluoroscopic imaging assembly.
(D)
For a lateral type fluoroscope, the exposure
rate shall be measured at a point fifteen
centimeters from the centerline of the x-ray
table and in the direction of the x-ray source
with the end of the beam-limiting device or
spacer positioned as closely as possible to the
point of measurement. If the tabletop is
movable, it shall be positioned as closely as
possible to the lateral x-ray source, with the
end of the beam-limiting device or spacer no
closer than fifteen centimeters to the
centerline of the x-ray table.
(6)
Periodic measurement of entrance exposure rate shall
be performed as follows:
(A)
Such measurements shall be made by, or under the
supervision of, a qualified medical physicist at
least annually or after any maintenance of the
system which might affect the exposure rate.
(B)
Results of these measurements shall be posted
§11-45-118
45-98
where any fluoroscopist may have ready access to
such results while using the fluoroscope and in
the record required in section 11-45-116
(a)(13)(C). The measurement results shall be
stated in roentgens (coulombs per kilogram) per
minute and include the technique factors used in
determining such results. The name of the
individual performing the measurements and the
date the measurements were performed shall be
included in the results.
(C)
Conditions of periodic measurement of maximum
entrance exposure rate are as follows:
(i)
The measurement shall be made under the
conditions that satisfy the requirements
of this subsection;
(ii)
The kilovolts peak and milliamperes shall
be typical of clinical use of the x-ray
system;
(iii)
The x-ray system(s) that incorporates
automatic exposure rate control shall have
an appropriate phantom placed in the
useful beam to produce a milliamperage
and/or kilovoltage typical of the use of
the x-ray system; and
(e)
Barrier transmitted radiation rate limits:
(1)
The exposure rate due to transmission through the
primary protective barrier with the attenuation block
in the useful beam, combined with radiation from the
image intensifier, if provided, shall not exceed two
milliroentgens (0.516 microcoulombs per kilogram) per
hour at ten centimeters from any accessible surface of
the fluoroscopic imaging assembly beyond the plane of
the image receptor for each roentgen per minute of
entrance exposure rate.
(2)
Measuring compliance of barrier transmission.
(A)
The exposure rate due to transmission through
the primary protective barrier combined with
radiation from the image intensifier shall be
determined by measurements averaged over an area
of one hundred square centimeters with no linear
dimension greater than twenty centimeters.
(B)
If the source is below the tabletop, the
measurement shall be made with the input surface
of the fluoroscopic imaging assembly positioned
thirty centimeters above the tabletop.
(C)
If the source is above the tabletop and the
source-image receptor distance is variable, the
measurement shall be made with the end of the
beam-limiting device or spacer as close to the
tabletop as it can be placed, provided that it
shall not be closer than thirty centimeters.
(D)
Movable grids and compression devices shall be
removed from the useful beam during the
measurement.
§11-45-118
45-99
(f)
During fluoroscopy and cinefluorography the tube
potential and tube current shall be continuously indicated.
(g)
The source-to-skin distance shall not be less than:
(1)
Thirty-eight centimeters on stationary fluoroscopic
systems manufactured on or after August 1, 1974,;
(2)
A distance of 35.5 centimeters on stationary
fluoroscopic systems manufactured before August 1,
1974;
(3)
Thirty centimeters on all mobile and portable
fluoroscopes; and
(4)
Twenty centimeters for all image-intensified
fluoroscopes used for specific surgical application.
(h)
Fluoroscopic timer:
(1)
Means shall be provided to preset the cumulative on-
time of the fluoroscopic x-ray tube. The maximum
cumulative time of the timing device shall not exceed
five minutes without resetting.
(2)
A signal audible to the fluoroscopist shall indicate
the completion of any preset cumulative on-time. Such
signal shall continue to sound while x-rays are
produced until the timing device is reset.
(i)
Control of scattered radiation:
(1)
Fluoroscopic table designs when combined with
procedures utilized shall be such that no unprotected
part of any staff or ancillary individual's body shall
be exposed to unattenuated scattered radiation which
originates from under the table. The attenuation
required shall be not less than 0.25 millimeter lead
equivalent.
(2)
Equipment configuration when combined with procedures
shall be such that no portion of any staff or
ancillary individual's body, except the extremities,
shall be exposed to the unattenuated scattered
radiation emanating from above the tabletop unless
that individual:
(A)
Is at least one hundred twenty centimeters from
the center of the useful beam; or
(B)
The radiation has passed through not less than
0.25 millimeter lead equivalent material
including, but not limited to, drapes, bucky-
slot cover panel, or self-supporting curtains,
in addition to any lead equivalency provided by
the protective apron.
(j)
Fluoroscopic systems equipped with spot film
(radiographic) mode shall meet the exposure reproducibility
requirements of section 11-45-119 when operating in the spot film
mode.
(k)
Radiation therapy simulation systems shall be exempt
from all the requirements of subsections (b), (d), (e), and (h)
provided that:
(1)
Such systems are designed and used in such a manner
that no individual other than the patient is in the x-
ray room during periods of time when the system is
producing x-rays; and
§11-45-118
45-100
(2)
Systems which do not meet the requirements of
subsection (h) are provided with a means of indicating
the cumulative time that an individual patient has
been exposed to x-rays. Procedures shall require in
such cases that the timer be reset between
examinations.
(l)
All newly-assembled, installed, or reassembled
fluoroscopic x-ray systems shall be evaluated by, or under the
supervision of, a qualified medical physicist before use on human
patients unless otherwise authorized by the director.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)
§11-45-119 Radiographic systems other than fluoroscopic,
dental intraoral, or computed tomography systems. (a) The useful
beam shall be limited to the area of clinical interest. This
shall be deemed to have been met if a positive beam limiting
device has been properly used or if evidence of collimation is
shown on at least three sides or three corners of the film, (e.g.,
projections from the shutters of the collimator, cone cutting at
the corners, or borders at the film's edge).
(1)
General purpose stationary, mobile, and portable x-ray
systems, including veterinary systems installed after
the effective date of this chapter:
(A)
Means shall be provided for independent stepless
adjustment of the size of the x-ray field. Each
dimension of the minimum field size at an
source-image receptor distance of one hundred
centimeters shall be equal to or less than five
centimeters.
(B)
A method shall be provided for visually defining
the perimeter of the x-ray field. The total
misalignment of the edges of the visually
defined field with the respective edges of the
x-ray field along either the length or width of
the visually defined field shall not exceed two
percent of the distance from the source to the
center of the visually defined field when the
surface upon which it appears is perpendicular
to the axis of the x-ray beam.
(2)
In addition to the requirements of paragraph (1), all
stationary general purpose x-ray systems shall meet
the following requirements:
(A)
A method shall be provided to indicate when the
axis of the x-ray beam is perpendicular to the
plane of the image receptor, to align the center
of the x-ray field with respect to the center of
the image receptor to within two percent of the
source-image receptor distance, and to indicate
the source-image receptor distance to within two
percent;
(B)
The beam-limiting device shall indicate
numerically the field size in the plane of the
§11-45-119
45-101
image receptor to which it is adjusted; and
(C)
Indication of field size dimensions and source-
image receptor distances shall be specified in
inches and/or centimeters, and shall be such
that aperture adjustments result in x-ray field
dimensions in the plane of the image receptor
which correspond to those indicated by the beam-
limiting device to within two percent of the
source-image receptor distances when the beam
axis is indicated to be perpendicular to the
plane of the image receptor.
(3)
Radiographic equipment designed for only one image
receptor size at a fixed source-image receptor
distances shall be provided with means to limit the
field at the plane of the image receptor to dimensions
no greater than those of the image receptor, and to
align the center of the x-ray field with the center of
the image receptor to within two percent of the
source-image receptor distance, or shall be provided
with means to both size and align the x-ray field such
that the x-ray field at the plane of the image
receptor does not extend beyond any edge of the image
receptor.
(4)
Radiographic systems other than those designated in
paragraphs (1), (2), and (3), and veterinary systems
installed before the effective date of this chapter:
(A)
Means shall be provided to limit the x-ray field
in the plane of the image receptor so that such
field does not exceed each dimension of the
image receptor by more than two percent of the
source-image receptor distance when the axis of
the x-ray beam is perpendicular to the plane of
the image receptor.
(B)
Means shall be provided to align the center of
the x-ray field with the center of the image
receptor to within two percent of the source-
image receptor distance, or means shall be
provided to both size and align the x-ray field
such that the x-ray field at the plane of the
image receptor does not extend beyond any edge
of the image receptor. Compliance shall be
determined with the axis of the x-ray beam
perpendicular to the plane of the image
receptor.
(C)
The requirements of subparagraphs (A) and (B)
may be met with a system that meets the
requirements for a general purpose x-ray system
as specified in paragraph (1) or, when alignment
means are also provided, may be met with either:
(i)
An assortment of removable,
fixed-aperture, beam-limiting devices
sufficient to meet the requirement for
each combination of image receptor size
and source-image receptor distance for
§11-45-119
45-102
which the unit is designed with each such
device having clear and permanent markings
to indicate the image receptor size and
source-image receptor distance for which
it is designed; or
(ii)
A beam-limiting device having multiple
fixed apertures sufficient to meet the
requirement for each combination of image
receptor size and source-image receptor
distance for which the unit is designed.
Permanent, clearly legible markings shall
indicate the image receptor size and
source-image receptor distance for which
each aperture is designed and shall
indicate which aperture is in position for
use.
(b)
Radiation exposure control:
(1)
Means shall be provided to initiate the radiation
exposure by a deliberate action on the part of the
operator, such as the depression of a switch.
Radiation exposure shall not be initiated without such
an action. In addition, it shall not be possible to
initiate an exposure when the timer is set to a "zero"
or "off" position if either position is provided.
(2)
Means shall be provided to terminate the exposure at a
preset time interval, preset product of current and
time, a preset number of pulses, or a preset radiation
exposure to the image receptor. Termination of an
exposure shall cause automatic resetting of the timer
to its initial setting or to "zero."
(A)
When manual exposure control is provided, an x-
ray exposure control shall be incorporated into
each x-ray system such that an exposure can be
terminated by the operator at any time
("deadman" switch) except for:
(i)
Exposure of one-half second or less; and
(ii)
During serial radiography when means shall
be provided to permit completion of any
single exposure of the series in progress.
(B)
When an automatic exposure control is provided:
(i)
Indication shall be made on the control
panel when this mode of operation is
selected;
(ii)
If the x-ray tube potential is equal to or
greater than fifty-one kilovolts peak, the
minimum exposure time for field emission
equipment rated for pulsed operation shall
be equal to or less than a time interval
equivalent to two pulses;
(iii)
The minimum exposure time for all
equipment other than that specified in
clause (ii) shall be equal to or less than
one-sixtieth second or a time interval
required to deliver five milliamperes-
§11-45-119
45-103
seconds, whichever is greater;
(iv)
Either the product of peak x-ray tube
potential, current, and exposure time
shall be limited to not more than sixty
kilowatt-seconds per exposure, or the
product of x-ray tube current and exposure
time shall be limited to not more than six
hundred milliamperes-seconds per exposure,
except that when the x-ray tube potential
is less than fifty-one kilovolts peak, the
product of x-ray tube current and exposure
time shall be limited to not more than two
thousand milliamperes-seconds per
exposure; and
(v)
A visible signal shall indicate when an
exposure has been terminated at the limits
required by clause (iv), and manual
resetting shall be required before further
automatically timed exposures can be made.
(3)
Means shall be provided for visual indication of x-ray
production observable at or from the operator's
protected position whenever x-rays are produced. In
addition, a signal audible to the operator shall
indicate that the exposure has terminated.
(4)
With a timer setting of 0.5 seconds or less, the
difference between the maximum exposure time (Tmax)
and the minimum exposure time (Tmin) shall be less
than or equal to ten percent of the average exposure
time (T), when four timer tests are performed:
(Tmax - Tmin) < 0.1T
(5)
The x-ray exposure control shall be so placed that the
operator can view the patient while making any
exposure.
(6)
Operator protection, except veterinary systems:
(A)
Stationary x-ray systems shall be required to
have the x-ray exposure control permanently
mounted in a protected area so that the operator
is required to remain in that protected area
during the entire exposure.
(B)
Mobile and portable x-ray systems which are:
(i)
Used continuously for greater than one
week in the same location, i.e., a room or
suite shall be considered as stationary
systems under subparagraph (A).
(ii)
Used for less than one week in the same
location shall be provided with either a
protective barrier at least 6.5 feet (two
meters) high for operator protection
during exposures, or means shall be
provided to allow the operator to be at
least twelve feet (3.7 meters) from the
tube housing assembly during the exposure.
§11-45-119
45-104
(7)
Operators of all stationary, mobile or portable x-ray
systems used for veterinary work shall be provided
with a 6.5 foot (two meters) high protective barrier
for operator protection during exposures, or a means
to allow the operator to be at least twelve feet (3.7
meters) from the tube housing assembly during
exposures, or a protective apron and protective
gloves.
(c)
All mobile or portable radiographic systems shall be
provided with means to limit the source-to-skin distance to equal
to or greater than thirty centimeters, except for veterinary
systems.
(d)
When all technique factors are held constant,
including control panel selections associated with automatic
exposure control systems, the coefficient of variation of exposure
for both manual and automatic exposure control systems shall not
exceed 0.05. This requirement shall be deemed to have been met
if, when four exposures are made at identical technique factors,
the difference between the maximum exposure (Emax) and the minimum
exposure (Emin) shall be less than or equal to ten percent of the
average exposure (E):
(Emax - Emin) < 0.1E
(e)
Radiation from capacitor energy storage equipment in
standby status: Radiation emitted from the x-ray tube when the
exposure switch or timer is not activated shall not exceed a rate
of two milliroentgens (0.516 microcoulombs per kilogram) per hour
at five centimeters from any accessible surface of the diagnostic
source assembly, with the beam-limiting device fully open.
(f)
Deviation of technique factors from indicated values
shall not exceed the limits specified for that system by its
manufacturer. In the absence of manufacturer's specifications,
the deviation shall not exceed ten percent of the indicated value.
(g)
Linearity, uncertified x-ray systems only: The
following requirements apply when the equipment is operated on a
power supply as specified by the manufacturer for any fixed x-ray
tube potential within the range of forty percent to one hundred
percent of the maximum rated.
(1)
Equipment having independent selection of x-ray tube
current: The average ratios of exposure to the
indicated milliampere seconds product (mR/mAs (or
C/kg/mAs)) obtained at any two tube current settings
shall not differ by more than 0.10 times their sum:
[X1 - X2] < 0.10(X1 + X2)
where X1 and X2 are the average mR/mAs (or C/kg/mAs)
values obtained at any two tube current settings.
(2)
Equipment having a combined x-ray tube current-
exposure time product (mAs) selector, but not a
separate tube current (mA) selector: The average
ratios of exposure to the indicated milliampere
seconds product (mR/mAs (or C/kg/mAs)) obtained at any
§11-45-119
45-105
two mAs selector settings shall not differ by more
than 0.10 times their sum:
[X1 - X2] < 0.10(X1 + X2)
where X1 and X2 are the average mR/mAs (or C/kg/mAs)
values obtained at any two mAs selector settings.
(3)
Determination of compliance shall be based on four
exposures, of no less than 0.05 seconds each, taken
within a time period of one hour, at each of the two
settings. These two settings may include any two
focal spot sizes provided that neither focal spot size
is equal to or less than 0.45 millimeter, in which
case the two settings shall be restricted to the same
focal spot size. For purposes of this requirement,
focal spot size is the nominal focal spot size
specified by the x-ray tube manufacturer.
(h)
Additional requirements applicable to certified
systems only: diagnostic x-ray systems incorporating one or more
certified component(s) shall be required to comply with the
following additional requirement(s) which relate to that certified
component(s).
(1)
Linearity: When the equipment is operated on a power
supply as specified by the manufacturer in accordance
with the requirements of applicable federal standards,
for any fixed x-ray tube potential within the range of
forty to one hundred percent of the maximum rated.
(A)
For equipment having independent selection of x-
ray tube current, the average ratios of exposure
to the indicated milliampere seconds product
obtained at any two consecutive tube current
settings shall not differ by more than 0.10
times their sum:
[X1 - X2] < 0.10(X1 + X2)
where X1 and X2 are the average mR/mAs
(C/kg/mAs) values obtained at each of two
consecutive tube current settings or at two
settings differing by no more than a factor of
two where the tube current selection is
continuous.
(B)
For equipment having selection of x-ray tube
current-exposure time product manufactured after
May 3, 1994, the average ratios of exposure to
the indicated milliampere seconds product
obtained at any two consecutive milliampere
seconds selector settings shall not differ by
more than 0.10 times their sum:
[X1 - X2] < 0.10(X1 + X2)
where X1 and X2 are the average mR/mAs
(C/kg/mAs) values obtained at each of two
§11-45-119
45-106
consecutive milliampere seconds selector
settings or at two settings differing by no more
than a factor of two where the milliampere
seconds selector provides continuous selection.
(2)
Beam limitation for stationary and mobile general
purpose x-ray systems:
(A)
There shall be provided a means of stepless
adjustment of the size of the x-ray field. The
minimum field size at an source-image receptor
distance of one hundred centimeters shall be
equal to or less than five centimeters by five
centimeters.
(B)
When a light localizer is used to define the x-
ray field, it shall provide an average
illumination of not less than one hundred sixty
lux or fifteen footcandles at one hundred
centimeters or at the maximum source-image
receptor distance, whichever is less. The
average illumination shall be based upon
measurements made in the approximate center of
each quadrant of the light field. Radiation
therapy simulation systems manufactured on and
after May 27, 1980, are exempt from this
requirement.
(C)
The edge of the light field at one hundred
centimeters or at the maximum source-image
receptor distance, whichever is less, shall have
a contrast ratio, corrected for ambient
lighting, of not less than four in the case of
beam-limiting devices designed for use on
stationary equipment, and a contrast ratio of
not less than three in the case of beam-limiting
devices designed for use on mobile equipment.
The contrast ratio is defined as I1/I2 where I1
is the illumination three millimeters from the
edge of the light field toward the center of the
field; and I2 is the illumination three
millimeters from the edge of the light field
away from the center of the field. Compliance
shall be determined with a measuring instrument
aperture of one millimeter in diameter.
(3)
Beam limitation for portable x-ray systems shall meet
the beam limitation requirements of paragraphs (1) and
(2).
(4)
Stationary, general purpose x-ray systems which
contain a tube housing assembly, an x-ray control,
and, for those systems so equipped, a table, all
certified in accordance with 21 C.F.R. ? 1020.30(c)
shall have manual collimation for field limitation and
alignment.
(5)
If positive beam limitation is installed on
stationary, general purpose x-ray systems after the
effective date of these rules, the following
requirements shall be met:
§11-45-119
45-107
(A)
Positive beam limitation shall prevent the
production of x-rays when:
(i)
Either the length or width of the x-ray
field in the plane of the image receptor
differs, except as permitted by
subparagraph (E), from the corresponding
image receptor dimensions by more than
three percent of the source-image receptor
distance; or
(ii)
The sum of the length and width
differences as stated in clause (i)
without regard to sign exceeds four
percent of the source-image receptor
distance.
(iii)
The beam limiting device is at an source-
image receptor distance for which positive
beam limitation is not designed for
sizing.
(B)
If a means of overriding the positive beam
limitation system exists, that means shall be
designed for use only in the event of positive
beam limitation system failure or if the system
is being serviced.
(C)
If in a position that the operator would
consider it part of the operational controls or
if it is referenced in the operator's manual or
in other materials intended for the operator,
(i)
Shall require that a key be utilized to
defeat the positive beam limitation;
(ii)
Shall require that the key remain in place
during the entire time the positive beam
limitation system is overridden; and
(iii)
Shall require that the key or key switch
be clearly and durably labeled as follows:
FOR X-RAY FIELD LIMITATION SYSTEM FAILURE
(D)
Compliance with subparagraph (A) shall be
determined when the equipment indicates that the
beam axis is perpendicular to the plane of the
image receptor. Compliance shall be determined
no sooner than five seconds after insertion of
the image receptor.
(E)
The positive beam limitation system shall be
capable of operation, at the discretion of the
operator, such that the size of the field may be
made smaller than the size of the image receptor
through stepless adjustment of the field size.
The minimum field size at an source-image
receptor distance of one hundred centimeters
shall be equal to or less than five centimeters
by five centimeters.
(F)
Positive beam limitation shall be designed such
that if a change in image receptor does not
§11-45-119
45-108
cause an automatic return to positive beam
limitation function as described in subparagraph
(B), then any change of image receptor size or
source-image receptor distance shall cause the
automatic return.
(6)
Except for dental panoramic systems, termination of
exposure shall cause automatic resetting of the timer
to its initial setting or to "zero."
(i)
A tube stand or other mechanical support shall be used
for portable x-ray systems so that the x-ray tube housing assembly
need not be hand-held during exposures. [Eff 11/12/99] (Auth:
HRS §§321-10, 321-11, 321-71) (Imp: HRS §§321-1, 321-11(21),
321-71)