HAR §11-45-120
HAR §11-45-120. Intraoral dental radiographic systems
Cite as Haw. Code R. § 11-45-120
(a)
X-ray systems designed for use with an intraoral image receptor
shall be provided with means to limit source-to-skin distance to
not less than eighteen centimeters.
(b)
Radiographic systems designed for use with an
intraoral image receptor shall be provided with means to limit the
x-ray field such that the x-ray field at the minimum source-to-
skin distance shall be containable in a circle having a diameter
of no more than seven centimeters.
(c)
Radiation exposure control for certified and non-
certified systems:
(1)
Exposure initiation:
(A)
Means shall be provided to initiate the
radiation exposure by a deliberate action on the
part of the operator, such as the depression of
a switch. Radiation exposure shall not be
initiated without such an action; and
(B)
It shall not be possible to make an exposure
when the timer is set to a "zero" or "off"
position if either position is provided.
(2)
Exposure termination:
(A)
Means shall be provided to terminate the
exposure at a preset time interval, preset
product of current and time, a preset number of
pulses, or a preset radiation exposure to the
image receptor.
(B)
An x-ray exposure control shall be incorporated
into each x-ray system such that an exposure can
be terminated by the operator at any time,
except for exposures of one-half second or less.
(C)
Termination of an exposure shall cause automatic
resetting of the timer to its initial setting or
to "zero".
(3)
Means shall be provided for visual indication
observable at or from the operator's protected
position whenever x-rays are produced. In addition, a
signal audible to the operator shall indicate that the
exposure has terminated.
(4)
With a timer setting of 0.5 seconds or less, the
§11-45-120
45-109
difference between the maximum exposure time (Tmax)
and the minimum exposure time (Tmin) shall be less
than or equal to ten percent of the average exposure
time (T), when four timing tests are performed:
(Tmax - Tmin) < 0.10T
(5)
Exposure control location and operator protection:
(A)
Stationary x-ray systems shall be required to
have the x-ray exposure control permanently
mounted in a protected area, so that the
operator is required to remain in that protected
area during the entire exposure; and
(B)
Mobile and portable x-ray systems which are:
(i)
Used for greater than one week in the same
location, i.e., a room or suite, shall
meet the requirements of subparagraph (A);
(ii)
Used for less than one week in the same
location shall be provided with either a
protective barrier at least 6.5 feet (two
meters) high for operator protection, or
means to allow the operator to be at least
twelve feet (3.7 meters) from the tube
housing assembly while making exposures.
(d)
The coefficient of variation for exposure
reproducibility shall not exceed 0.05 when all technique factors
are held constant. This requirement shall be deemed to have been
met if, when four exposures are made within a period of one hour
at identical technique factors, the difference between the maximum
exposure value (Emax) and the minimum exposure value (Emin) shall
be less than or equal to ten percent of the average exposure:
(Emax - Emin) < 0.10E
(e)
When the equipment allows a choice of x-ray tube
current settings and is operated on a power supply as specified by
the manufacturer for any fixed x-ray tube potential within the
range of forty to one hundred percent of the maximum rating, the
average ratios of exposure to the indicated milliampere seconds
product obtained at any two consecutive tube current settings
shall not differ by more than 0.10 times their sum:
(X1 - X2) < 0.10(X1 + X2)
where X1 and X2 are the average mR/mAs (C/kg/mAs) values obtained
at each of two consecutive tube current settings.
(f)
Deviation of technique factors from indicated values
shall not exceed the limits specified for that system by its
manufacturer. In the absence of manufacturer's specifications the
deviation shall not exceed ten percent of the indicated value.
(g)
Dental x-ray machines with a nominal fixed kilovolts
peak of less than fifty kilovolts peak shall not be used to make
diagnostic dental radiographs of humans.
(h)
Administrative controls:
§11-45-120
45-110
(1)
Patient and film holding devices shall be used when
appropriate.
(2)
The tube housing and the position indicating device
shall not be hand-held during an exposure.
(3)
The x-ray system shall be operated in such a manner
that the useful beam at the patient's skin does not
exceed the requirements of subsection (b).
(4)
Dental fluoroscopy without image intensification shall
not be used. [Eff 11/12/99] (Auth: HRS §§321-10,
321-11, 321-71) (Imp: HRS §§321-1, 321-11(21), 321-
71)