HAR §11-45-121
HAR §11-45-121. Computed tomography x-ray systems
Cite as Haw. Code R. § 11-45-121
(a)
General requirements:
(1)
Means shall be provided to terminate the x-ray
exposure automatically by either de-energizing the x-
ray source or shuttering the x-ray beam in the event
of equipment failure affecting data collection. Such
termination shall occur within an interval that limits
the total scan time to no more than one hundred ten
percent of its preset value through the use of either
a backup timer or devices which monitor equipment
function.
(2)
A visible signal shall indicate when the x-ray
exposure has been terminated through the means
required by paragraph (1).
(3)
The operator shall be able to terminate the x-ray
exposure at any time during a scan, or series of scans
under computed tomography x-ray system control, of
greater than one-half second duration.
(4)
For any single tomogram system, means shall be
provided to permit visual determination of the
tomographic plane or a reference plane offset from the
tomographic plane.
(5)
For any multiple tomogram system, means shall be
provided to permit visual determination of the
location of a reference plane. This reference plane
can be offset from the location of the tomographic
planes.
(6)
If a device using a light source is used to satisfy
paragraphs (4) and (5), the light source shall provide
illumination levels sufficient to permit visual
determination of the location of the tomographic plane
or reference plane under ambient light conditions of
up to five hundred lux.
(7)
The computed tomography x-ray control and gantry shall
provide visual indication whenever x-rays are produced
and, if applicable, whether the shutter is open or
closed.
(8)
Each emergency button or switch shall be clearly
labeled as to its function.
(9)
The computed tomography x-ray system shall be designed
such that the computed tomography conditions of
§11-45-121
45-111
operation to be used during a scan or a scan sequence
shall be indicated before the initiation of a scan or
a scan sequence. On equipment having all or some of
these conditions of operation at fixed values, this
requirement may be met by permanent markings.
Indication of computed tomography conditions of
operation shall be visible from any position from
which scan initiation is possible.
(10)
When data are not being collected for image
production, the radiation adjacent to the tube port
shall not exceed that permitted by section 11-45-117
(c).
(11)
The angular position where the maximum surface
computed tomography dose index occurs shall be
identified to allow for reproducible positioning of a
computed tomography dosimetry phantom.
(12)
Additional requirements applicable to computed
tomography x-ray systems containing a gantry
manufactured after September 3, 1985:
(A)
The total error in the indicated location of the
tomographic plane or reference plane shall not
exceed five millimeters.
(B)
If the x-ray production period is less than one-
half second, the indication of x-ray production
shall be actuated for at least one-half second.
Indicators at or near the gantry shall be
discernible from any point external to the
patient opening where insertion of any part of
the human body into the primary beam is
possible.
(C)
The deviation of indicated scan increment versus
actual increment shall not exceed plus or minus
one millimeter with any mass from zero to one
hundred kilograms resting on the support device.
The patient support device shall be incremented
from a typical starting position to the maximum
incremented distance or thirty centimeters,
whichever is less, and then returned to the
starting position. Measurement of actual versus
indicated scan increment may be taken anywhere
along this travel.
(D)
Premature termination of the x-ray exposure by
the operator shall necessitate resetting of the
computerized tomography conditions of operation
before the initiation of another scan.
(b)
Provision shall be made for two-way aural
communication between the patient and the operator at the control
panel.
(c)
Viewing systems:
(1)
Windows, mirrors, closed-circuit television, or an
equivalent shall be provided to permit continuous
observation of the patient during irradiation and
shall be so located that the operator can observe the
patient from the control panel.
§11-45-121
45-112
(2)
When the primary viewing system is by electronic
means, an alternate viewing system (which may be
electronic) shall be available for use in the event of
failure of the primary viewing system.
(d)
Surveys:
(1)
All computed tomography x-ray systems installed after
the effective date of these rules and those systems
not previously surveyed shall have a survey made by,
or under the direction of, a qualified medical
physicist before use on human patients unless
otherwise authorized by the director. In addition,
such surveys shall be done after any change in the
facility or equipment which might cause a significant
increase in radiation hazard.
(2)
A written report of the survey shall be available for
inspection by the department for three years.
(e)
Radiation calibrations:
(1)
Calibration of the radiation output of the computed
tomography x-ray system shall be performed by, or
under the direction of, a qualified medical physicist
who is physically present at the facility during such
calibration.
(2)
Calibration of a computed tomography x-ray system
shall be performed at intervals specified by a
qualified medical physicist and after any change or
replacement of components which, in the opinion of the
qualified medical physicist, could cause a change in
the radiation output.
(3)
Calibration of the radiation output of a computed
tomography x-ray system shall be performed with a
calibrated dosimetry system. The calibration of such
system shall be traceable to a national standard. The
dosimetry system shall have been calibrated within the
preceding two years.
(4)
Computed tomography dosimetry phantom(s) shall be used
in determining the radiation output of a computed
tomography x-ray system. Such phantom(s) shall meet
the following specifications and conditions of use:
(A)
Computed tomography dosimetry phantom(s) shall
be right circular cylinders of polymethyl
methacrylate of density 1.19 plus or minus 0.01
grams per cubic centimeter. The phantom(s)
shall be at least fourteen centimeters in length
and shall have diameters of 32.0 centimeters for
testing computed tomography x-ray systems
designed to image any section of the body and
16.0 centimeters for systems designed to image
the head or for whole body scanners operated in
the head scanning mode.
(B)
Computed tomography dosimetry phantom(s) shall
provide means for the placement of a
dosimeter(s) along the axis of rotation and
along a line parallel to the axis of rotation
1.0 centimeter from the outer surface and within
§11-45-121
45-113
the phantom. Means for the placement of
dosimeters or alignment devices at other
locations may be provided.
(C)
Any effects on the doses measured due to the
removal of phantom material to accommodate
dosimeters shall be accounted for through
appropriate corrections to the reported data or
included in the statement of maximum deviation
for the values obtained using the phantom.
(D)
All dose measurements shall be performed with
the computed tomography dosimetry phantom placed
on the patient couch or support device without
additional attenuation materials present.
(5)
Calibration shall be required for each type of head,
body, or whole-body scan performed at the facility.
(6)
Calibration shall include the center multiple scan
average dose and the surface multiple scan average
dose.
(7)
Calibration procedures shall be in writing. Records
of calibrations performed shall be maintained for
inspection by the department for three years.
(f)
Spot checks shall be conducted:
(1)
Procedures shall be in writing and shall have been
developed by a qualified medical physicist.
(2)
Procedures shall incorporate the use of a computed
tomography dosimetry phantom which has a capability of
providing an indication of contrast scale, noise,
nominal tomographic section thickness, the resolution
capability of the system for low and high contrast
objects, and measuring the mean computed tomography
number for water or other reference material.
(3)
Spot checks shall be included in the calibration
required by subsection (e) and at time intervals and
under system conditions specified by a qualified
medical physicist.
(4)
Spot checks shall include acquisition of images
obtained with the computed tomography dosimetry
phantom(s) using the same processing mode and computed
tomography conditions of operation as are used to
perform calibrations required by subsection (e). The
images shall be retained, until a new calibration is
performed, in two forms as follows:
(A)
Photographic copies of the images obtained from
the image display device; and
(B)
Images stored in digital form on a storage
medium compatible with the computed tomography
x-ray system.
(5)
Written records of the spot checks shall be
maintained for inspection by the department for three
years.
(g)
Operating procedures:
(1)
The computed tomography x-ray system shall be operated
by an individual who has been specifically trained in
its operation and licensed under chapter 11-44.
§11-45-121
45-114
(2)
Information shall be available at the control panel
regarding the operation and calibration of the system.
Such information shall include the following:
(A)
Dates of the latest calibration and spot checks
and the location within the facility where the
results of those tests may be obtained;
(B)
Instructions on the use of the computed
tomography dosimetry phantom(s) including a
schedule of spot checks appropriate for the
system, allowable variations for the indicated
parameters, and the results of at least the most
recent spot checks conducted on the system;
(C)
The distance in millimeters between the
tomographic plane and the reference plane if a
reference plane is utilized; and
(D)
A current technique chart available at the
control panel which specifies for each routine
examination the computed tomography conditions
of operation and the number of scans per
examination.
(3)
If the calibration or spot check of the computed
tomography x-ray system identifies that a system
operating parameter has exceeded a tolerance
established by the qualified medical physicist, use of
the computed tomography x-ray system on patients shall
be limited to those uses permitted by established
written instructions of the qualified medical
physicist. [Eff 11/12/99] (Auth: HRS §§321-10, 321-
11, 321-71) (Imp: HRS §§321-1, 321-11(21), 321-71)