HAR §11-45-174
HAR §11-45-174. Equipment requirements
Cite as Haw. Code R. § 11-45-174
(a) A safety device
which prevents the entry of any portion of an individual's body
into the primary x-ray beam path or which causes the beam to be
shut off upon entry into its path shall be provided on all open-
beam configurations. A licensee may apply to the department for
an exemption from the requirement of a safety device. Such
application shall include:
(1)
A description of the various safety devices that have
been evaluated;
(2)
The reason each of these devices cannot be used; and
(3)
A description of the alternative methods that shall be
employed to minimize the possibility of an accidental
exposure, including procedures to assure that
operators and others in the area shall be informed of
the absence of safety devices.
(b) Open-beam configurations shall be provided with a
readily discernible indication of:
(1)
X-ray tube "on-off" status located near the radiation
source housing, if the primary beam is controlled in
this manner; and/or
(2)
Shutter "open-closed" status located near each port on
the radiation source housing, if the primary beam is
controlled in this manner.
(c)
An easily visible warning light labeled with the words
"X-RAY ON", or words having a similar intent, shall be located:
(1)
Near any switch that energizes an x-ray tube and shall
be illuminated only when the tube is energized; or
(2)
In the case of a radioactive source, near any switch
that opens a housing shutter and shall be illuminated
only when the shutter is open.
(d)
Warning devices shall be labeled so that their purpose
is easily identified. On equipment installed after the effective
date of these rules, warning devices shall have fail-safe
characteristics.
(e)
Unused ports on radiation source housings shall be
secured in the closed position in a manner which shall prevent
casual opening.
(f)
All analytical x-ray equipment shall be labeled with a
readily discernible sign or signs bearing the radiation symbol and
§11-45-174
45-142
the words:
(1)
"CAUTION - HIGH INTENSITY X-RAY BEAM", or words having
a similar intent, on the x-ray source housing; and
(2)
"CAUTION RADIATION - THIS EQUIPMENT PRODUCES RADIATION
WHEN ENERGIZED", or words having a similar intent,
near any switch that energizes an x-ray tube if the
radiation source is an x-ray tube; or
(3)
"CAUTION - RADIOACTIVE MATERIAL", or words having a
similar intent, on the source housing in accordance
with subchapter 4 if the radiation source is a
radionuclide.
(g)
On open-beam configurations installed after the
effective date of these rules, each port on the radiation source
housing shall be equipped with a shutter that cannot be opened
unless a collimator or a coupling has been connected to the port.
(h)
Each radiation source housing shall be subject to the
following requirements:
(1)
Each x-ray tube housing shall be equipped with an
interlock that shuts off the tube if it is removed
from the radiation source housing or if the housing is
disassembled; and
(2)
Each radioactive source housing or port cover or each
x-ray tube housing shall be so constructed that, with
all shutters closed, the radiation measured at a
distance of five centimeters from its surface is not
capable of producing a dose in excess of 2.5 millirems
(0.025 millisievert) in one hour. For systems
utilizing x-ray tubes, this limit shall be met at any
specified tube rating.
(i)
Each x-ray generator shall be supplied with a
protective cabinet which limits leakage radiation measured at a
distance of five centimeters from its surface such that it is not
capable of producing a dose in excess of 0.25 millirem (2.5
microsieverts) in one hour. [Eff 11/12/99] (Auth: HRS §§321-10,
321-11, 321-71) (Imp: HRS §§321-1, 321-11(21), 321-71)