HAR §11-45-175
HAR §11-45-175. Area requirements
Cite as Haw. Code R. § 11-45-175
(a) The local components of
an analytical x-ray system shall be located and arranged and shall
include sufficient shielding or access control such that no
radiation levels exist in any area surrounding the local component
group which could result in a dose to an individual present
therein in excess of the dose limits given in subchapter 4. For
systems utilizing x-ray tubes, these levels shall be met at any
specified tube rating.
(b)
Radiation surveys, as required by subchapter 4, of all
analytical x-ray systems sufficient to show compliance with
subsection (a) shall be performed:
(1)
Upon installation of the equipment, and at least once
every twelve months thereafter;
(2)
Following any change in the initial arrangement,
number, or type of local components in the system;
(3)
Following any maintenance requiring the disassembly or
removal of a local component in the system;
§11-45-177
45-143
(4)
During the performance of maintenance and alignment
procedures if the procedures require the presence of a
primary x-ray beam when any local component in the
system is disassembled or removed;
(5)
Any time a visual inspection of the local components
in the system reveals an abnormal condition; and
(6)
Whenever personnel monitoring devices show a
significant increase over the previous monitoring
period or the readings are approaching the limits
specified in subchapter 4.
(c)
Radiation survey measurements shall not be required if
a licensee can demonstrate compliance with subchapter 4 to the
satisfaction of the department.
(d)
Each area or room containing analytical x-ray
equipment shall be conspicuously posted with a sign or signs
bearing the radiation symbol and the words "CAUTION - X-RAY
EQUIPMENT" or words having a similar intent in accordance with
subchapter 4. [Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-
71) (Imp: HRS §§321-1, 321-11(21), 321-71)