HAR §11-45-237
HAR §11-45-237. Reserved
Cite as Haw. Code R. § 11-45-237
§11-45-238 Therapeutic radiation machines - photon therapy
systems (five hundred kilovolts and above) and electron therapy
§11-45-238
45-179
systems (five hundred kilo electron volts and above). (a) Each
facility location authorized to use a therapeutic radiation
machine in accordance with this section shall possess
appropriately calibrated portable monitoring equipment. As a
minimum, such equipment shall include a portable radiation
measurement survey instrument capable of measuring dose rates over
the range one millirem (ten microsieverts) per hour to one
thousand millirems (ten millisieverts) per hour. The survey
instrument(s) shall be operable and calibrated in accordance with
section 11-45-240.
(b)
Leakage radiation outside the maximum useful beam in
photon and electron modes:
(1)
The absorbed dose rate due to leakage radiation
(excluding neutrons) at any point outside the maximum
sized useful beam, but within a circular plane of
radius two meters which is perpendicular to and
centered on the central axis of the useful beam at the
nominal treatment distance (i.e. patient plane), shall
not exceed a maximum of 0.2 percent and an average of
0.1 percent of the absorbed dose rate on the central
axis of the beam at the nominal treatment distance.
Measurements shall be averaged over an area not
exceeding one hundred square centimeters at a minimum
of sixteen points uniformly distributed in the plane;
(2)
Except for the area defined in paragraph (1), the
absorbed dose rate in water (excluding that from
neutrons) at one meter from the electron path between
the source and the target or electron window shall not
exceed 0.5 percent of the absorbed dose rate in water
on the central axis of the beam at the nominal
treatment distance. Measurements shall be averaged
over an area not exceeding one hundred square
centimeters at a minimum of sixteen points uniformly
distributed in the plane;
(3)
The neutron absorbed dose rate outside the useful beam
shall be kept as low as practicable. Measurements of
the portion of the leakage radiation dose contributed
by neutrons shall be averaged over an area not
exceeding eight hundred square centimeters; and
(4)
For each therapeutic radiation machine, the licensee
shall determine, or obtain from the manufacturer, the
leakage radiation existing at the positions specified
in paragraphs (1) through (3) for the specified
operating conditions. Records on leakage radiation
measurements shall be maintained at the installation
for inspection.
(c)
Leakage radiation through beam limiting devices:
(1)
Photon radiation: All adjustable or interchangeable
beam limiting devices shall attenuate the useful beam
such that at the nominal treatment distance, the
maximum absorbed dose anywhere in the area shielded by
the beam limiting device(s) shall not exceed two
percent of the maximum absorbed dose on the central
axis of the useful beam measured in a ten centimeters
§11-45-238
45-180
by ten centimeters radiation field;
(2)
Electron radiation: All adjustable or interchangeable
electron applicators shall attenuate the radiation,
including but not limited to photon radiation
generated by electrons incident on the beam limiting
device and electron applicator and other parts of the
radiation head, such that the absorbed dose in a plane
perpendicular to the central axis of the useful beam
at the nominal treatment distance shall not exceed:
(A)
A maximum of two percent of the absorbed dose on
the central axis of the useful beam at the
nominal treatment distance. This limit shall
apply beyond a line seven centimeters outside
the periphery of the useful beam; and
(B)
A maximum of ten percent of the absorbed dose on
the central axis of the useful beam at the
nominal treatment distance. This limit shall
apply beyond a line two centimeters outside the
periphery of the useful beam.
(3)
Measurement of leakage radiation:
(A)
Photon radiation: Measurements of leakage
radiation through the beam limiting devices
shall be made with the beam limiting devices
closed and any residual aperture blocked by at
least two tenth value layers of suitable
absorbing material. In the case of overlapping
beam limiting devices, the leakage radiation
through each set shall be measured independently
at the depth of maximum dose. Measurements
shall be made using a radiation detector of area
not exceeding ten square centimeters;
(B)
Electron radiation: Measurements of leakage
radiation through the electron applicators shall
be made with the electron beam directed into the
air and using a radiation detector of area up to
but not exceeding one square centimeter suitably
protected against radiation which has been
scattered from material beyond the radiation
detector. Measurements shall be made using one
centimeter of water equivalent build up
material.
(4)
When adjustable beam limiting devices are used, the
position and shape of the radiation field shall be
indicated by a light field.
(d)
Filters/wedges:
(1)
Each filter and/or wedge which is removable from the
system shall be clearly marked with an identification
number. For removable wedge filters, the nominal
wedge angle shall appear on the wedge or wedge tray
(if permanently mounted to the tray). If the wedge or
wedge tray is damaged, the wedge transmission factor
shall be redetermined;
(2)
If the absorbed dose rate information required by
subsection (a) relates exclusively to operation with a
§11-45-238
45-181
field flattening or beam scattering filter in place,
such filter shall be removable only by the use of
tools;
(3)
For equipment manufactured after the effective date of
this chapter which utilize a system of wedge filters,
interchangeable field flattening filters, or
interchangeable beam scattering foils:
(A)
Irradiation shall not be possible until a
selection of a filter or a positive selection to
use "no filter" has been made at the treatment
control panel, either manually or automatically;
(B)
An interlock system shall be provided to prevent
irradiation if the filter selected is not in the
correct position;
(C)
A display shall be provided at the treatment
control panel showing the wedge filter(s),
interchangeable field flattening filter(s),
and/or interchangeable beam scattering foil(s)
in use; and
(D)
An interlock shall be provided to prevent
irradiation if any filter and/or beam scattering
foil selection operation carried out in the
treatment room does not agree with the filter
and/or beam scattering foil selection operation
carried out at the treatment control panel.
(e)
For equipment manufactured after the effective date of
this chapter, the licensee shall determine during acceptance
testing, or obtain from the manufacturer, data sufficient to
ensure that x-ray stray radiation in the useful electron beam,
absorbed dose at the surface during x-ray irradiation and stray
neutron radiation in the useful x-ray beam are in compliance with
International Electrotechnical Commission (IEC) Document 601-2-1.
(f)
All therapeutic radiation machines subject to this
section shall be provided with beam monitoring devices. The
sensors for this device shall be fixed in the useful beam during
treatment to indicate the air kerma rate or dose rate.
(1)
Equipment manufactured after the effective date of
this chapter shall be provided with at least two
independently powered integrating dose meters.
Alternatively, a common power supply may be used if
the production of radiation is terminated upon failure
of any common element.
(2)
Equipment manufactured on or before the effective date
of this chapter shall be provided with at least one
radiation detector. This detector shall be
incorporated into a useful beam monitoring system;
(3)
The detector and the system into which that detector
is incorporated shall meet the following requirements:
(A)
Each detector shall be removable only with tools
and, if movable, shall be interlocked to prevent
incorrect positioning;
(B)
Each detector shall form part of a beam
monitoring system from which readings in dose
monitor units the absorbed dose at a reference
§11-45-238
45-182
point in the treatment volume can be calculated;
(C)
Each beam monitoring system shall be capable of
independently monitoring, interrupting, and
terminating irradiation; and
(D)
For equipment manufactured after the effective
date of this chapter, the design of the beam
monitoring systems shall ensure that the:
(i)
Malfunctioning of one system shall not
affect the correct functioning of the
secondary system; and
(ii)
Failure of any element common to both
systems which could affect the correct
function of both systems shall terminate
irradiation or prevent the initiation of
radiation.
(E)
Each beam monitoring system shall have a legible
display at the treatment control panel. For
equipment manufactured after the effective date
of this chapter, each display shall:
(i)
Maintain a reading until intentionally
reset;
(ii)
Have only one scale and no electrical or
mechanical scale multiplying factors;
(iii)
Utilize a design such that increasing dose
is displayed by increasing numbers; and
(iv)
In the event of power failure, the beam
monitoring information required in clause
(iii) displayed at the control panel at
the time of failure shall be retrievable
in at least one system for a twenty minute
period of time.
(g)
Beam symmetry:
(1)
Bent-beam linear accelerators subject to this section
shall be provided with auxiliary device(s) to monitor
beam symmetry;
(2)
The device(s) referenced in paragraph (1) shall be
able to detect field asymmetry greater than ten
percent; and
(3)
The device(s) referenced in paragraph (1) shall be
configured to terminate irradiation if the
specifications in paragraph (2) cannot be maintained.
(h)
Selection and display of dose monitor units:
(1)
Irradiation shall not be possible until a selection of
a number of dose monitor units has been made at the
treatment control panel;
(2)
The pre-selected number of dose monitor units shall be
displayed at the treatment control panel until reset
manually for the next irradiation;
(3)
After termination of irradiation, it shall be
necessary to reset the dosimeter display before
subsequent treatment can be initiated; and
(4)
For equipment manufactured after the effective date of
this chapter, after termination of irradiation, it
shall be necessary for the operator to reset the pre-
§11-45-238
45-183
selected dose monitor units before irradiation can be
initiated.
(i)
For equipment manufactured after the effective date of
this chapter, a system shall be provided from which readings the
air kerma rate or absorbed dose rate at a reference point in the
treatment volume can be calculated. The radiation detectors
specified in subsection (f) may form part of this system. In
addition:
(1)
The dose monitor unit dose rate shall be displayed at
the treatment control panel;
(2)
If the equipment can deliver under any conditions an
air kerma rate or absorbed dose rate at the nominal
treatment distance more than twice the maximum value
specified by the manufacturer, a device shall be
provided which terminates irradiation when the air
kerma rate or absorbed dose rate exceeds a value twice
the specified maximum. The dose rate at which the
irradiation shall be terminated shall be a record
maintained by the licensee;
(3)
For equipment manufactured after the effective date of
this chapter, if the equipment can deliver under any
conditions an air kerma rate or absorbed dose rate at
the nominal treatment distance more than ten times the
maximum value specified by the manufacturer, a device
shall be provided to prevent the air kerma rate or
absorbed dose rate anywhere in the radiation field
from exceeding twice the specified maximum value and
to terminate irradiation if the excess absorbed dose
at the nominal treatment distance exceeds four hundred
rads (four grays); and
(4)
For each therapeutic radiation machine, the licensee
shall determine, or obtain from the manufacturer, the
maximum value(s) specified in paragraphs (2) and (3)
for the specified operating conditions. Records of
these maximum value(s) shall be maintained at the
installation for inspection.
(j)
Termination of irradiation by the beam monitoring
system or systems during stationary beam radiation therapy:
(1)
Each primary system shall terminate irradiation when
the pre-selected number of dose monitor units has been
detected by the system;
(2)
If the original design of the equipment included a
secondary dose monitoring system, that system shall be
capable of terminating irradiation when not more than
fifteen percent or forty dose monitor units above the
pre-selected number of dose monitor units set at the
control panel has been detected by the secondary dose
monitoring system; and
(3)
For equipment manufactured after the effective date of
this chapter, an indicator on the control panel shall
show which monitoring system has terminated
irradiation.
(k)
It shall be possible to terminate irradiation and
equipment movement or go from an interruption condition to
§11-45-238
45-184
termination condition at any time from the operator's position at
the treatment control panel.
(l)
If a therapeutic radiation machine has an interrupt
mode, it shall be possible to interrupt irradiation and equipment
movements at any time from the treatment control panel. Following
an interruption it shall be possible to restart irradiation by
operator action without any reselection of operating conditions.
If any change is made of a pre-selected value during an
interruption, irradiation and equipment movements shall be
automatically terminated.
(m)
A suitable irradiation control device shall be
provided to terminate the irradiation after a pre-set time
interval.
(1)
A timer shall be provided which has a display at the
treatment control panel. The timer shall have a pre-
set time selector and an elapsed time indicator;
(2)
The timer shall be a cumulative timer which activates
with an indication of "BEAM-ON" and retains its
reading after irradiation is interrupted or
terminated. After irradiation is terminated and
before irradiation can be reinitiated, it shall be
necessary to reset the elapsed time indicator;
(3)
For equipment manufactured after the effective date of
this chapter, after termination of irradiation and
before irradiation can be reinitiated, it shall be
necessary for the operator to reset the pre-set time
selector;
(4)
The timer shall terminate irradiation when a
pre-selected time has elapsed, if the dose monitoring
systems have not previously terminated irradiation.
(n)
Equipment capable of both x-ray therapy and electron
therapy shall meet the following additional requirements:
(1)
Irradiation shall not be possible until a selection of
radiation type (x-rays or electrons) has been made at
the treatment control panel;
(2)
The radiation type selected shall be displayed at the
treatment control panel before and during irradiation;
(3)
An interlock system shall be provided to ensure that
the equipment can principally emit only the radiation
type which has been selected;
(4)
An interlock system shall be provided to prevent
irradiation with x-rays, except to obtain a
verification film, when electron applicators are
fitted;
(5)
An interlock system shall be provided to prevent
irradiation with electrons when accessories specific
for x-ray therapy are fitted; and
(6)
An interlock system shall be provided to prevent
irradiation if any selected operations carried out in
the treatment room do not agree with the selected
operations carried out at the treatment control panel.
(o)
Equipment capable of generating radiation beams of
different energies shall meet the following requirements:
(1)
Irradiation shall not be possible until a selection of
§11-45-238
45-185
energy has been made at the treatment control panel;
(2)
The measured energy value selected shall be displayed
(megavolts for photons and mega electron volts for
electrons) at the treatment control panel before and
during irradiation; and
(3)
Irradiation shall not be possible until the
appropriate flattening filter or scattering foil for
the selected energy is in its proper location.
(p)
Therapeutic radiation machines capable of both
stationary beam radiation therapy and rotational arc radiation
therapy shall meet the following requirements:
(1)
Irradiation shall not be possible until a selection of
stationary beam radiation therapy or rotational arc
radiation therapy has been made at the treatment
control panel;
(2)
The mode of operation shall be displayed at the
treatment control panel;
(3)
An interlock system shall be provided to ensure that
the equipment can operate only in the mode which has
been selected;
(4)
An interlock system shall be provided to prevent
irradiation if any selected parameter in the treatment
room does not agree with the selected parameter at the
treatment control panel;
(5)
Rotational arc radiation therapy shall be controlled
to obtain the selected relationships between
incremental dose monitor units and incremental angle
of movement for equipment manufactured after the
effective date of this chapter:
(A)
An interlock system shall be provided to
terminate irradiation if the number of dose
monitor units delivered in any fifteen degrees
of arc differs by more than twenty percent from
the selected value;
(B)
Where gantry angle terminates the irradiation in
rotational arc radiation therapy, the dose
monitor units shall differ by less than five
percent from the value calculated from the
absorbed dose per unit angle and total angle
relationship;
(C)
An interlock shall be provided to prevent the
gantry moving more than five degrees beyond the
selected angular limits during rotational arc
radiation therapy;
(D)
An interlock shall be provided to require that a
selection of direction be made at the treatment
control panel in all units which are capable of
both clockwise and counter-clockwise rotational
arc radiation therapy.
(6)
Where the beam monitor system terminates the
irradiation in rotational arc radiation therapy, the
termination of irradiation shall be as required by
subsection (j); and
(7)
For equipment manufactured after the effective date of
§11-45-238
45-186
this chapter, an interlock system shall be provided to
terminate irradiation if movement of the gantry:
(A)
Occurs during stationary beam radiation therapy;
or
(B)
Stops during rotational arc radiation therapy
unless such stoppage is a pre-planned function.
(q)
Facility design requirements for therapeutic radiation
machines operating above five hundred kilovolts: In addition to
shielding adequate to meet requirements of section 11-45-241, the
following design requirements are made:
(1)
All protective barriers shall be fixed, except for
access doors to the treatment room or movable beam
interceptors;
(2)
In addition to other requirements specified in this
subchapter, the control panel shall also:
(A)
Be located outside the treatment room;
(B)
Provide an indication of whether electrical
power is available at the control panel and if
activation of the radiation is possible;
(C)
Provide an indication of whether radiation is
being produced; and
(D)
Include an access control (locking) device which
shall prevent unauthorized use of the
therapeutic radiation machine;
(3)
Windows, mirrors, closed-circuit television or an
equivalent viewing system shall be provided to permit
continuous observation of the patient following
positioning and during irradiation and shall be so
located that the operator may observe the patient from
the treatment control panel. The therapeutic
radiation machine shall not be used for patient
irradiation unless at least one viewing system is
operational;
(4)
Provision shall be made for continuous two-way aural
communication between the patient and the operator at
the control panel. The therapeutic radiation machine
shall not be used for irradiation of patients unless
continuous two-way aural communication is possible;
(5)
Treatment room entrances shall be provided with
warning lights in a readily observable position near
the outside of all access doors, which shall indicate
when the useful beam is "ON" and when it is "OFF";
(6)
Interlocks shall be provided such that all entrance
doors shall be closed before treatment can be
initiated or continued. If the radiation beam is
interrupted by any door opening, it shall not be
possible to restore the machine to operation without
closing the door and reinitiating irradiation by
manual action at the control panel;
(7)
If the shielding material in any protective barrier
requires the presence of a beam interceptor to ensure
compliance with subchapter 4, interlocks shall be
provided to prevent the production of radiation,
unless the beam interceptor is in place, whenever the
§11-45-238
45-187
useful beam is directed at the designated barrier(s);
(8)
At least one "scram button" or other emergency power
cutoff switch shall be located in the radiation
therapy room and shall terminate all equipment
electrical power including radiation and mechanical
motion. This switch is in addition to the termination
switch required by subsection (k). All emergency power
cutoff switches shall include a manual reset so that
the therapeutic radiation machine cannot be restarted
from the unit's control console without resetting the
emergency cutoff switch;
(9)
All safety interlocks shall be designed so that any
defect or component failure in the safety interlock
system prevents or terminates operation of the
therapeutic radiation machine; and
(10)
Surveys for residual activity shall be conducted on
all therapeutic radiation machines capable of
generating photon and electron energies above ten
megavolts before machining, removing, or working on
therapeutic radiation machine components which may
have become activated due to photo-neutron production.
(r)
Qualified medical physicist support:
(1)
The services of a qualified medical physicist shall be
utilized in facilities having therapeutic radiation
machines with energies of five hundred kilovolts and
above. The qualified medical physicist shall be
responsible for:
(A)
Full calibration(s) required by subsection (t)
and protection surveys required by section 11-
45-234(a);
(B)
Supervision and review of dosimetry;
(C)
Beam data acquisition and transfer for
computerized dosimetry, and supervision of its
use;
(D)
Quality assurance, including quality assurance
check review required by subsections (u) and
(v);
(E)
Consultation with the authorized user in
treatment planning, as needed; and
(F)
Perform calculations/assessments regarding
misadministrations.
(2)
If the qualified medical physicist is not a full-time
employee of the licensee, the operating procedures
required by subsection (s) shall also specifically
address how the qualified medical physicist is to be
contacted for problems or emergencies, as well as the
specific actions, if any, to be taken until the
qualified medical physicist can be contacted.
(s)
Operating procedures:
(1)
No individual, other than the patient, shall be in the
treatment room during treatment or during any
irradiation for testing or calibration purposes;
(2)
Therapeutic radiation machines shall not be made
available for medical use unless the requirements of
§11-45-238
45-188
section 11-45-234(a), subsections (t) and (u) have
been met;
(3)
Therapeutic radiation machines, when not in operation,
shall be secured to prevent unauthorized use;
(4)
If a patient must be held in position during
treatment, mechanical supporting or restraining
devices shall be used; and
(5)
A copy of the current operating and emergency
procedures shall be maintained at the therapeutic
radiation machine control console.
(t)
Full calibration measurements:
(1)
Full calibration of a therapeutic radiation machine
subject to this section shall be performed by, or
under the direct supervision of, a qualified medical
physicist:
(A)
Before the first medical use following
installation or reinstallation of the
therapeutic radiation machine;
(B)
At intervals not exceeding one year; and
(C)
Before medical use under the following
conditions:
(i)
Whenever quality assurance check
measurements indicate that the radiation
output differs by more than five percent
from the value obtained at the last full
calibration and the difference cannot be
easily discerned; and
(ii)
Following any component replacement, major
repair, or modification of components that
could significantly affect the
characteristics of the radiation beam.
(D)
Notwithstanding the requirements of subparagraph
(C):
(i)
Full calibration of therapeutic radiation
machines with multi-energy and/or
multi-mode capabilities is required only
for those modes and/or energies that are
not within their acceptable range; and
(ii)
If the repair, replacement or modification
does not affect all modes and/or energies,
full calibration shall be performed on the
effected mode/energy that is in most
frequent clinical use at the facility.
The remaining energies/modes may be
validated with quality assurance check
procedures against the criteria in clause
(i).
(2)
To satisfy the requirement of paragraph (1), full
calibration shall include all measurements required
for annual calibration by American Association of
Physicists in Medicine (AAPM) Report 13, "Physical
Aspects of Quality Assurance in Radiation Therapy";
(3)
The licensee shall use the dosimetry system described
in section 11-45-234(c) to measure the radiation
§11-45-238
45-189
output for one set of exposure conditions. The
remaining radiation measurements required in paragraph
(2) may then be made using a dosimetry system that
indicates relative dose rates; and
(4)
The licensee shall maintain a record of each
calibration for the duration of the license. The
record shall include the date of the calibration, the
manufacturer's name, model number, and serial number
for the therapeutic radiation machine, the model
numbers and serial numbers of the instruments used to
calibrate the therapeutic radiation machine, and the
signature of the qualified medical physicist
responsible for performing the calibration.
(u)
Periodic quality assurance checks:
(1)
Periodic quality assurance checks shall be performed
on all therapeutic radiation machines subject to this
section at intervals not to exceed one week;
(2)
To satisfy the requirement of paragraph (1), quality
assurance checks shall include determination of all
parameters for periodic quality assurance checks
contained in American Association of Physicists in
Medicine (AAPM) Report 13, "Physical Aspects of
Quality Assurance in Radiation Therapy";
(3)
The licensee shall use a dosimetry system which has
been inter-compared within the previous six months
with the dosimetry system described in section 11-45-
234(c)(1) to make the periodic quality assurance
checks required in paragraph (2);
(4)
The licensee shall perform periodic quality assurance
checks required by paragraph (1) in accordance with
procedures established by the qualified medical
physicist;
(5)
The licensee shall review the results of each periodic
radiation output check according to the following
procedures:
(A)
The authorized user and qualified medical
physicist shall be immediately notified if any
parameter is not within its acceptable range.
The therapeutic radiation machine shall not be
made available for subsequent medical use until
the qualified medical physicist has determined
that all parameters are within their acceptable
range;
(B)
If all quality assurance check parameters appear
to be within their acceptable range, the quality
assurance check shall be reviewed and signed by
either the authorized user or qualified medical
physicist within three treatment days; and
(C)
The qualified medical physicist shall review and
sign the results of each radiation output
quality assurance check at intervals not to
exceed one month.
(6)
Therapeutic radiation machines subject to this section
shall have safety quality assurance checks performed
§11-45-238
45-190
at intervals not to exceed one week;
(7)
To satisfy the requirement of paragraph (6), safety
quality assurance checks shall ensure proper operation
of:
(A)
Electrical interlocks at each external beam
radiation therapy room entrance;
(B)
Proper operation of the "BEAM-ON", interrupt and
termination switches;
(C)
Beam condition indicator lights on the access
doors, control console, and in the radiation
therapy room;
(D)
Viewing systems;
(E)
Electrically operated treatment room door(s)
from inside and outside the treatment room;
(F)
At least one emergency power cutoff switch. If
more than one emergency power cutoff switch is
installed and not all switches are tested at
once, each switch shall be tested on a rotating
basis. Safety quality assurance checks of the
emergency power cutoff switches may be conducted
at the end of the treatment day in order to
minimize possible stability problems with the
therapeutic radiation machine.
(8)
The licensee shall promptly repair any system
identified in paragraph (7) that is not operating
properly; and
(9)
The licensee shall maintain a record of each quality
assurance check required by paragraphs (1) and (7) for
three years. The record shall include the date of the
quality assurance check, the manufacturer's name,
model number, and serial number for the therapeutic
radiation machine, the manufacturer's name, model
number and serial number of the instrument(s) used to
measure the radiation output of the therapeutic
radiation machine, and the signature of the individual
who performed the periodic quality assurance check.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71)
(Imp: HRS §§321-1, 321-11(21), 321-71)