HAR §11-45-240
HAR §11-45-240. Calibration and check of survey instruments
Cite as Haw. Code R. § 11-45-240
(a) The licensee shall ensure that the survey instruments used to
show compliance with this subchapter have been calibrated before
first use, at intervals not to exceed twelve months, and following
repair.
(b)
To satisfy the requirements of subsection (a), the
licensee shall:
(1)
Calibrate all required scale readings up to one
thousand millirems (ten millisieverts) per hour with
an appropriate radiation source;
(2)
Calibrate at least two points on each scale to be
§11-45-241
45-191
calibrated. These points should be at approximately
one-third and two-thirds of scale rating.
(c)
To satisfy the requirements of subsection (b), the
licensee shall:
(1)
Consider a point as calibrated if the indicated dose
rate differs from the calculated dose rate by not more
than ten percent; and
(2)
Consider a point as calibrated if the indicated dose
rate differs from the calculated dose rate by not more
than twenty percent if a correction factor or graph is
conspicuously attached to the instrument.
(d)
The licensee shall retain a record of each calibration
required in subsection (a). The record shall include:
(1)
A description of the calibration procedure; and
(2)
A description of the source used and the certified
dose rates from the source, and the rates indicated by
the instrument being calibrated, the correction
factors deduced from the calibration data, the
signature of the individual who performed the
calibration, and the date of calibration.
(e)
The licensee may obtain the services of individuals
licensed to perform calibrations of survey instruments. Records
of calibrations which contain information required by subsection
(d) shall be maintained by the licensee. [Eff 11/12/99] (Auth:
HRS §§321-10, 321-11, 321-71) (Imp: HRS §§321-1, 321-11(21),
321-71)