HAR §11-114-5
HAR §11-114-5. Instrument approvals
Length: 589 wordsOfficial source
Cite as Haw. Code R. § 11-114-5
(a) Breath
alcohol tests shall be performed using a model of:
(1) Breath alcohol testing instrument;
(2) Breath alcohol testing instrument
accessories; and
(3) Accuracy verification device
which are approved by the DUI coordinator.
(b) The model specifications of NHTSA for
evidential breath alcohol testing devices and for
calibrating units (referred to in this subchapter as
accuracy verification devices) for breath alcohol
testers, as contained in 49 CFR, No. 242, pp.
48854-48865 and 49 CFR, No. 242, pp. 48865-48872,
respectively, are integrated into and made a part of
§11-114-
114-6
this subchapter. Accordingly, those models of
instruments, accessories, and calibrating units
appearing in the "Conforming Products List of
Evidential Breath Measurement Devices" as contained in
57 CFR, No. 46, pp. 8375-8376, and "Conforming Products
List of Calibrating Units for Breath Alcohol Testers"
as contained in 56 CFR, No. 54, pp. 1187-11819, are
approved by the DUI coordinator for purposes of this
subchapter.
(c) The DUI coordinator may approve, in writing,
modified versions of approved instruments, accessories,
and accuracy verification devices. Approval will be
contingent upon the continued performance of the
instrument, accessory, or calibrating within the
specifications set forth in subsection (b).
(d) All breath alcohol testing devices approved
by the director of health as of the effective date of
this chapter shall remain approved unless the approval
is specifically revoked by the director of health in
writing. [EffJDECJ30,J1993JJ] (Auth: HRS §321-161)
(Imp: HRS §§321-161; 286-2; 286-151; 286-152; 286-153;
286-154; 286-155.5; 286-158; 286-160; 286-162; 286-163;
286-171; 286-251; 286-254; 286-255; 286-256; 286-257;
286-258; 286-259; 291-4; 291-4.5; 291-5; 291-6; 291C-
162; 321-1; 321-9; 321-10; 321-18; 321-19; 321-20)
§11-114-6 Procedure approvals and measurement
requirements. (a) Except as provided in subsection
(c), every breath alcohol testing procedure shall be
approved by the DUI coordinator in writing and shall
include, but not be limited to:
(1) Performance of an accuracy verification test
with each breath alcohol test, using an
approved accuracy verification device; and
(2)
Inclusion of an air blank before and after
each breath test.
(b) With every breath alcohol test the following
shall be met:
(1) The person to be tested shall not have
ingested alcoholic beverages, eaten, smoked,
or vomited for at least fifteen minutes
before the breath alcohol test;
(2) The test shall be conducted using an approved
§11-114-
114-7
instrument in conformance with section 11-
114-5 and following an approved procedure as
specified in subsection (a);
(3) A copy of the approved breath alcohol testing
procedure shall be accessible to the operator
or supervisor;
(4) The test shall be conducted by a person who
is licensed as a breath alcohol testing
supervisor or operator pursuant to section
11-114-9 or 11-114-10;
(5) The result of the accuracy verification test
shall be within the range of plus or minus
0.01, or plus or minus ten per cent of the
target value, whichever is greater;
(6) Breath alcohol and accuracy verification test
results shall not be rounded up; and
(7) Each breath alcohol test shall have been
preceded by an accuracy test which meets the
criteria of section 11-114-7, by not more
than thirty-one days.
(c) Any breath alcohol testing procedure approved
by the director of health as of the effective date of
this chapter shall continue to be approved and remain
in effect unless superseded or revoked by the director
of health in writing. [EffJDECJ30,J1993JJ] (Auth: HRS
§321-161)(Imp: HRS §§321-161; 286-2; 286-151; 286-152;
286-153; 286-154; 286-155.5; 286-158; 286-160; 286-162;
286-163; 286-171; 286-251; 286-254; 286-255; 286-256;
286-257; 286-258; 286-259; 291-4; 291-4.5; 291-5; 291-
6; 291C-162; 321-1; 321-9; 321-10; 321-18; 321-19; 321-
20)