HAR §11-114-7. Accuracy tests
Length: 232 wordsOfficial source
Cite as Haw. Code R. § 11-114-7
(a) Every accuracy
test procedure shall be approved by the DUI coordinator
in writing and shall include, but not be limited to the
following requirements:
(1) The test shall be conducted by a supervisor;
(2) At least two different reference samples and
an air blank shall be run with each accuracy
test;
(3) Reference samples shall be chosen so that
their target values are not less than 0.04gm
alcohol /210 liters and not greater than
§11-114-
114-8
0.25gm alcohol /210 liters;
(4) Reference sample target values shall differ
from each other by at least 0.04gm alcohol
/210 liters;
(5) Reference sample test results which vary from
the target value by more than plus or minus
0.01gm alcohol /210 liters or plus or minus
ten percent, whichever is greater, shall be
cause for the breath alcohol testing
instrument used to be removed from service
until the fault has been corrected; and
(6) An accuracy test shall be performed on an
operating instrument at intervals not to
exceed thirty-one days.
(b) Operating instruments do not include
instruments in storage or being repaired. Upon return
to operation from storage or repair an accuracy test
shall be performed. [EffJDECJ30,J1993JJ] (Auth: HRS