32 Ill. Adm. Code 335.7060
Surveys After Source Implant and Removal
Section 335.7060 Surveys After Source Implant and
Removal
a) Immediately
after implanting sources in a patient or a human research subject, the licensee
shall monitor the area to locate and account for all sources that have not been
implanted.
b) Immediately
after removing the last temporary implant source from a patient or a human
research subject, the licensee shall monitor the patient or the human research
subject with a radiation detection survey instrument to confirm that all sources
have been removed.
c) A
licensee shall maintain a record of the monitoring required by this Section for
5 years. The record shall include the monitoring date, the measured dose rate
at several points in each area, expressed in units, multiples or subunits of Sieverts
or rem per hour, the manufacturer, model and serial number of the survey
instrument used to perform the monitoring, and the identity of the person who
performed the monitoring.