32 Ill. Adm. Code 335.7070
Calibration Measurements of Brachytherapy Sources
Section 335.7070 Calibration Measurements of
Brachytherapy Sources
a) Before
the first medical use of a brachytherapy source, a licensee shall have:
1) Determined
the source output or activity using a dosimetry system that meets the
requirements of subsection 335.8080(a);
2) Determined
source positioning accuracy within applicators; and
3) Used
published protocols currently accepted by nationally recognized bodies to meet
the requirements of subsections (a)(1) and (a)(2). Copies of these protocols
shall be maintained on file by the licensee for 5 years after the
discontinuation of use of brachytherapy sources.
b) A
licensee may use measurements provided by the source manufacturer or by a
calibration laboratory accredited by the American Association of Physicists in
Medicine or other calibration laboratory approved by the Agency that are made
in accordance with subsection (a).
c) A
licensee shall mathematically correct the outputs or activities determined in
subsection (a) for physical decay at intervals consistent with 1 percent
physical decay.
d) A
licensee shall maintain a record of the calibrations of brachytherapy sources
required by this Section for 5 years after the last use of the source. The
record shall include the:
1) Date of
the calibration;
2) Manufacturer's
name, model number, and serial number for the source, and the instruments used
to calibrate the source;
3) Source
output or activity;
4) Source
positioning accuracy within the applicators; and
5) Name
of the individual, source manufacturer, or the calibration laboratory that
performed the calibration.