32 Ill. Adm. Code 335.8040
Safety Procedures and Instructions for Remote Afterloader Units, Teletherapy Units and Gamma Stereotactic Radiosurgery Units
Section 335
Section 335.8040 Safety Procedures
and Instructions for Remote Afterloader Units, Teletherapy Units and Gamma
Stereotactic Radiosurgery Units
a) A licensee using sealed sources in remote afterloader units,
teletherapy units or gamma stereotactic radiosurgery units for therapeutic
medical uses shall:
1) Secure the unit, the console, the console keys and the
treatment room when not in use or unattended, if applicable;
2) Permit only individuals approved by the authorized user,
Radiation Safety Officer or authorized medical physicist to be present in the
treatment room during treatment or emergencies with the sources;
3) Prevent dual operation of more than one radiation producing
device in a treatment room, if applicable; and
4) Develop,
implement and maintain written procedures for responding to an abnormal
situation when the operator is unable to place the sources in the shielded
position, or remove the patient or human research subject from the radiation
field with controls from outside the treatment room. These procedures shall
include:
A) Instructions
for responding to equipment failures and the names of the individuals
responsible for implementing corrective actions;
B) The
process for restricting access to and posting of the treatment area to minimize
the risk of inadvertent exposure; and
C) The
names and telephone numbers of the authorized users, the authorized medical
physicist and the Radiation Safety Officer to be contacted if the unit or
console operates abnormally.
b) A copy of the procedures required by subsection (a)(4) and the
manufacturer's instruction manual shall be physically located at the unit
console.
c) A licensee shall post instructions at the unit console to
inform the operator of:
1) The
procedures located there as required by subsection (b); and
2) The
names and telephone numbers of the authorized users, the authorized medical
physicist and the Radiation Safety Officer to be contacted if the unit or
console operates abnormally.
d) Operational and Safety
Training
1) Prior
to the first use for patient treatment of a new unit or an existing unit with a
manufacturer upgrade that affects the operation and safety of the unit, the
licensee shall ensure that vendor operational and safety training is provided
to all individuals who will operate the unit. The vendor operational and
safety training must be provided by the device manufacturer or by an individual
certified by the device manufacturer to provide the operational and safety
training.
2) Initially
and at least annually, the licensee shall provide operational and safety
instructions to all individuals who operate the unit at the facility, as
appropriate to the individual's assigned duties, in:
A) The
procedures identified in subsection (a)(4); and
B) The
operating procedures for the unit.
e) A
licensee shall ensure that operators, authorized medical physicists and
authorized users participate in drills of the emergency procedures, initially
and at least annually.
f) A
licensee shall retain a record of the instruction required by subsection (d).
The record shall be retained for five years and include a list of the topics
covered, the date of the instruction, the names of the attendees and the names
of the individuals who provided instruction.
g) A
licensee shall retain a copy of the procedures required by subsections (a)(4)
and (d)(2)(B) until the licensee no longer possesses the remote afterloader,
teletherapy unit or gamma stereotactic radiosurgery unit.
h) A
licensee shall maintain a copy of the record documenting results of the drills
of emergency procedures required by subsection (e) for five years.