32 Ill. Adm. Code 335.8050
Safety Precautions for Remote Afterloader Units, Teletherapy Units and Gamma Stereotactic Radiosurgery Units
Section 335
Section 335.8050
Safety
Precautions for Remote Afterloader Units, Teletherapy Units and Gamma
Stereotactic Radiosurgery Units
a) A licensee shall control access to the treatment room by a
door at each entrance.
b) A licensee shall equip each entrance to the treatment room
with an electrical interlock system that shall:
1) Prevent the operator from initiating the treatment cycle
unless each treatment room entrance door is closed;
2) Cause the sources to be shielded when an entrance door is
opened; and
3) Prevent the sources from being exposed following an interlock
interruption until all treatment room entrance doors are closed and the sources
on-off control is reset at the console.
c) A licensee shall require any individual entering the treatment
room to assure, through the use of appropriate radiation monitors, that
radiation levels have returned to ambient levels.
d) Except for low-dose remote afterloader units, a licensee shall
construct or equip each treatment room with viewing and intercom systems to
permit continuous observation of the patient or the human research subject from
the treatment console during irradiation.
e) For licensed activities where sources are placed within the
patient's or human research subject's body, a licensee shall only conduct
treatments that allow for expeditious removal of a decoupled or jammed source.
f) In
addition to the requirements specified in subsections (a) through (e) of this
Section, a licensee shall:
1) For medium
dose-rate and pulsed dose-rate remote afterloader units, require:
A) An
authorized medical physicist and either an authorized user or a physician,
under the supervision of an authorized user, who has been trained in the
operation of, and emergency response for, the unit to be physically present
during the initiation of all patient treatments involving the unit; and
B) An
authorized medical physicist and either an authorized user or an individual,
under the supervision of an authorized user, who has been trained to remove the
source applicators in the event of an emergency involving the unit, to be
immediately available during continuation of all patient treatments involving
the unit.
2) For
high dose-rate remote afterloader units, require:
A) An authorized
user and an authorized medical physicist to be physically present during the
initiation of all patient treatments involving the unit; and
B) An
authorized medical physicist and either an authorized user or a physician,
under the supervision of an authorized user, who has been trained in the
operation and emergency response for the unit, to be physically present during
continuation of all patient treatments involving the unit.
3) For
gamma stereotactic radiosurgery units, require an authorized user and an
authorized medical physicist to be physically present throughout all patient
treatments involving the unit.
4) Notify
the Radiation Safety Officer, or his/her designee, and an authorized user as
soon as possible if the patient or human research subject has a medical
emergency or dies.
g) A
licensee shall have applicable emergency response equipment available near each
treatment room to respond to a source:
1) Remaining
in the unshielded position; or
2) Lodged
within the patient following completion of the treatment.