32 Ill. Adm. Code 335.8080
Dosimetry Equipment
Section 335
Section 335.8080 Dosimetry
Equipment
a) A licensee shall have a calibrated dosimetry system available
for use. To satisfy this requirement, one of the following two conditions shall
be met:
1) The system shall have been calibrated by the National Bureau
of Standards, by the National Institute of Standards and Technology or by a
calibration laboratory accredited by the American Association of Physicists in
Medicine (AAPM). The calibration shall have been performed within the previous
2 years and after any servicing that may have affected system calibration; or
2) The system shall have been calibrated within the previous 4
years; 18 to 30 months after that calibration, the system shall have been
compared with another dosimetry system that was calibrated within the past 24
months by the National Bureau of Standards, by the National Institute of
Standards and Technology or by a calibration laboratory accredited by the AAPM.
The dosimetry system shall be considered calibrated if a comparison is
performed at a meeting sanctioned by a calibration laboratory or radiological
physics center accredited by the AAPM and the results of the comparison
indicate that the calibration factor of the licensee's system has not changed
by more than two percent. The licensee shall not use the comparison result to
change the calibration factor. When comparing dosimetry systems to be used for
calibrating cobalt-60 teletherapy units, the licensee shall use a teletherapy
unit with a cobalt-60 source. When intercomparing dosimetry systems to be used
for calibrating sealed sources for therapeutic units, the licensee shall use a
comparable unit with beam attenuators or collimators, as applicable, and
sources of the same radionuclide as the source used at the licensee's facility.
b) The
licensee shall have available for use a calibrated dosimetry system for
spot-check measurements. To meet this requirement, the system may be compared
with a system that has been calibrated in accordance with subsection (a) of
this Section. This comparison shall have been performed within the previous
year and after each servicing that may have affected calibration of the
calibrated system. The spot-check system may be the same system used to meet
the requirements in subsection (a) of this Section.
c) The licensee shall retain a record of each calibration and
comparison for the duration of the license. For each calibration or comparison,
the record shall include the date, the manufacturer, the model and serial number
of the instruments that were calibrated or compared as required by subsections
(a) and (b) of this Section, the correction factors that were deduced, the
names of the individuals who performed the calibration or comparison and
evidence that the comparison meeting was sanctioned by a calibration laboratory
or radiological physics center accredited by AAPM.