32 Ill. Adm. Code 335.8090
Full Calibration Measurements for Teletherapy
Section 335
Section 335.8090 Full
Calibration Measurements for Teletherapy
a) A licensee authorized to use a teletherapy unit for medical
use shall perform full calibration measurements as described in subsection (b) of
this Section, on each teletherapy unit:
1) Before the first medical use of the unit; and
2) Before medical use under the following conditions:
A) Whenever spot-check measurements indicate that the output
differs by more than five percent from the output obtained at the last full
calibration, corrected mathematically for radioactive decay;
B) Following replacement of the source or following reinstallation
of the teletherapy unit in a new location;
C) Following any repair of the teletherapy unit that includes
removal of the source or major repair of the components associated with the
source exposure assembly; and
3) At intervals not exceeding 1 year.
b) To satisfy the requirement of subsection (a) of this Section,
full calibration measurements shall include determination of:
1) The output, within three percent, for the range of field sizes
and for the distance or range of distances used for medical use;
2) The coincidence of the radiation field and the field indicated
by the light beam localizing device;
3) The uniformity of the radiation field and its dependence on
the orientation of the useful beam;
4) Timer constancy and linearity over the range of use;
5) On-off error; and
6) The accuracy of all distance measuring and localization
devices in medical use.
c) A licensee shall use the dosimetry system described in Section
335.8080 of this Part to measure the output for one set of exposure conditions.
The remaining radiation measurements required by subsection (b)(1) of this
Section may then be made using a dosimetry system that indicates relative dose
rates.
d) A licensee shall make full calibration measurements required
by subsection (a) of this Section in accordance with published protocols
accepted by nationally recognized bodies.
e) A licensee shall mathematically correct for physical decay the
outputs determined in subsection (b)(1) of this Section. These corrections
shall be for intervals not exceeding 1 month for cobalt-60 and intervals not
exceeding 6 months for cesium-137 or at intervals consistent with 1 percent
decay for all other nuclides.
f) Full calibration measurements required by subsection (a) of
this Section and physical decay corrections required by subsection (e) of this
Section shall be performed by an authorized medical physicist.
g) A licensee shall retain a record of each calibration for the
duration of the license. The record shall include the date of the calibration,
the manufacturer's name, model and serial numbers for both the teletherapy unit
and the source, the model and serial numbers of the instruments used to
calibrate the teletherapy unit, tables that describe the output of the unit
over the range of field sizes and for the range of distance used in radiation
therapy, a determination of the coincidence of the radiation field and the
field indicated by the light beam localizing device, an assessment of timer
constancy and linearity, the calculated on-off error, the determined accuracy
of each distance measuring or localization device and the signature or initials
of the authorized medical physicist.