32 Ill. Adm. Code 335.8150
Full-Inspection Servicing for Teletherapy and Gamma Stereotactic Radiosurgery Units
Section 335
Section 335.8150 Full-Inspection
Servicing
for Teletherapy and Gamma Stereotactic Radiosurgery Units
a) A licensee shall have each teletherapy unit and gamma
stereotactic radiosurgery unit fully inspected and serviced during each source
replacement to assure proper functioning of the source exposure mechanism and
other safety components. The interval between each full-inspection servicing
shall not exceed 5 years for each teletherapy unit and shall not exceed 7 years
for each gamma stereotactic radiosurgery unit.
b) This inspection and servicing may only be performed by persons
specifically licensed to do so by the Agency, the U.S. Nuclear Regulatory
Commission, or an Agreement State.
c) A licensee shall maintain a record of the inspection and
servicing for teletherapy and gamma stereotactic radiosurgery units required by
this Section for the duration of use of the unit.
d) The record shall
contain:
1) The
inspector's radioactive materials license number;
2) The
date of the inspection;
3) The
manufacturer's name and model number and serial number of both the treatment
unit and source;
4) A
list of components inspected and serviced and the type of service; and
5) The signature
of the inspector.