32 Ill. Adm. Code 335.8160
Full Calibration Measurements on Remote Afterloader Units
Section 335.8160 Full Calibration Measurements on Remote
Afterloader Units
a) A
licensee authorized to use a remote afterloader unit for medical use shall
perform full calibration measurements on each unit:
1) Before the first
medical use of the unit;
2) Before
medical use under the following conditions:
A) Following
replacement of the source or following reinstallation of the unit in a new
location outside the facility;
B) Following
any repair of the unit that includes removal of the source or major repair of
the components associated with the source exposure assembly;
3) At
intervals not exceeding 1 quarter for high dose-rate, medium dose-rate and
pulsed dose-rate remote afterloader units with sources whose half-life exceeds
75 days; and
4) At
intervals not exceeding 1 year for low dose-rate remote afterloader units.
b) To
satisfy the requirement of subsection (a) of this Section, full calibration
measurements shall include, as applicable, determination of:
1) The
output within ± 5 percent;
2) Source
positioning accuracy to within ± 1 millimeter;
3) Source
retraction with backup battery upon power failure;
4) Length
of the source transfer tubes;
5) Timer
accuracy and linearity over the typical range of use;
6) Length
of the applicators; and
7) Function
of the source transfer tubes, applicators and transfer tube-applicator
interfaces.
c) A
licensee shall use the dosimetry system described in subsection 335.8080(a) to
measure the output.
d) A
licensee shall make full calibration measurements required by subsection (a) of
this Section in accordance with published protocols accepted by nationally
recognized bodies.
e) In
addition to the requirements for full calibrations for low dose-rate remote
afterloader units in subsection (b) of this Section, a licensee shall perform
an autoradiograph of the sources to verify inventory and sources arrangement at
intervals not exceeding 1 quarter.
f) For
low dose-rate remote afterloader units, a licensee may use measurements
provided by the source manufacturer that are made in accordance with
subsections (a) through (e) of this Section.
g) A
licensee shall mathematically correct the outputs determined in subsection
(b)(1) of this Section for physical decay at intervals consistent with 1
percent physical decay.
h) Full
calibration measurements required by subsection (a) of this Section and
physical decay corrections required by subsection (g) of this Section shall be
performed by the authorized medical physicist.
i) A
licensee shall maintain a record of the remote afterloader unit full
calibrations required by this Section for 5 years.
j) The records
shall include for each full calibration required by subsection (a) of this
Section:
1) The
date of the calibration;
2) The
manufacturer's name, model and serial number of the remote afterloader unit, together
with the sources and the instruments used to calibrate it;
3) The
results and an assessment of the full calibrations;
4) The
results of the autoradiograph required for low dose-rate remote afterloader
units; and
5) The
signature of the authorized medical physicist who performed the full
calibration.