32 Ill. Adm. Code 335.8170
Periodic Spot-Checks for Remote Afterloader Units
Section 335.8170
Periodic Spot-Checks for Remote Afterloader Units
a) A
licensee authorized to use a remote afterloader unit for medical use shall
perform spot-checks of each remote afterloader facility and on each unit:
1) Before
the first use of a high dose-rate, medium dose-rate or pulsed dose‑rate
remote afterloader unit on a given day;
2) Before
each patient treatment with a low dose-rate remote afterloader unit; and
3) After
each source installation.
b) A
licensee shall perform the measurements required by subsection (a) of this
Section in accordance with written procedures established by the authorized
medical physicist. That individual need not actually perform the spot-check
measurements.
c) A
licensee shall have the authorized medical physicist review the results of each
spot‑check within 15 days. The authorized medical physicist shall notify
the licensee as soon as possible in writing of the results of each spot-check.
d) To
satisfy the requirements of subsection (a) of this Section, spot-checks must,
at a minimum, assure proper operation of:
1) Electrical
interlocks at each remote afterloader unit room entrance;
2) Source
exposure indicator lights on the remote afterloader unit, on the control
console and in the facility;
3) Viewing
and intercom systems in each high dose-rate, medium dose-rate and pulsed
dose-rate remote afterloader facility;
4) Emergency
response equipment;
5) Radiation
monitors used to indicate the source position;
6) Timer
accuracy;
7) Clock
(date and time) in the unit's computer; and
8) Decayed
sources activity in the unit's computer.
e) If
the results of the checks required in subsection (d) of this Section indicate
the malfunction of any system, a licensee shall lock the control console in the
off position and not use the unit except as may be necessary to repair, replace
or check the malfunctioning system.
f) A
licensee shall retain a record of each spot-check for remote afterloader units
required by this Section for 5 years.
g) The record must include,
as applicable:
1) The
date of the spot-check;
2) The
manufacturer's name, model and serial number for the remote afterloader unit
and source;
3) An
assessment of timer accuracy;
4) Notations
indicating the operability of each entrance door electrical interlock,
radiation monitors, source exposure indicator lights, viewing and intercom
systems and clock and decayed source activity in the unit's computer; and
5) The
name of the individual who performed the periodic spot-check and the signature
of the authorized medical physicist who reviewed the record of the spot-check.
h) A
licensee shall retain a copy of the procedures required by subsection (b) of
this Section until the licensee no longer possesses the remote afterloader unit.