32 Ill. Adm. Code 360.100
Veterinary Radiographic Systems
Section 360
Section 360.100 Veterinary
Radiographic Systems
In addition to the provisions of
Sections 360.10, 360.30 and 360.40 (except Section 360.40(a)) of this Part, the
requirements of this Section apply to x-ray equipment and associated facilities
used for radiography with veterinary systems.
a) Beam Limitation. The useful beam shall be limited to the area
of clinical interest. The size of the image receptor used for each
radiographic projection shall be consistent with the objectives of the
examination.
1) Limitation Criteria. Means shall be provided to limit the
x-ray field in the plane of the image receptor so that the field does not
exceed each dimension of the image receptor by more than two percent of the SID
when the axis of the x-ray beam is perpendicular to the plane of the image
receptor.
2) Means shall be provided to align the center of the x-ray field
with the center of the image receptor to within two percent of the SID.
3) The requirements of subsection (a)(1) of this Section may be
met with:
A) An adjustable collimator with a field defining light; or
B) An assortment of removable, fixed-aperture, beam-limiting
devices sufficient to meet the requirement for each combination of image
receptor size and SID for which the unit is used, with each such device having
permanent, clearly legible markings in centimeters and/or inches, to indicate
the image receptor size and SID for which it is designed; or
C) A beam-limiting device having multiple fixed apertures
sufficient to meet the requirement for each combination of image receptor size
and SID for which the unit is used. Permanent, clearly legible markings, in
centimeters and/or inches, shall indicate the image receptor size and SID for
which each aperture is designed and shall indicate which aperture is in
position for use.
4) SID Indication
A) Means shall be provided to indicate the SID.
B) SIDs shall be indicated in centimeters and/or inches and the
measured SID shall correspond to the indicated value to within two percent.
b) Exposure Switch Arrangement. The exposure control switch
shall be arranged so the operator can be at least 1.83 meters (6 feet) from the
animal, the x-ray tube and the useful beam.
c) Radiation Exposure Control Devices
1) Means shall be provided to terminate the exposure at a preset
time interval, preset product of current and time, preset number of pulses or
preset radiation exposure to the image receptor. Also, it shall not be
possible to make an exposure when the timer is set to a zero or off position if
either position is provided.
2) The exposure switch shall be a dead-man switch.
d) Veterinary fluoroscopic, computed tomography and therapy
systems shall meet the requirements specified in Sections 360.50, 360.75,
360.110 and 360.120 of this Part, except that the requirements pertaining to
aural communication specified in Sections 360.75(b)(2), 360.110(a)(8) and
(e)(5) and 360.120(a)(6) and (g)(1)(H) of this Part, need not be satisfied
unless a human is used to hold the animal.
e) Additional Requirements for Veterinary X-Ray Systems
1) All individuals whose presence is required during an x-ray
examination shall be protected from scatter radiation by protective aprons or
gowns of not less than 0.25 millimeter lead equivalent or whole body protective
barriers.
2) All exams and retakes shall be ordered by the veterinarian.
3) Unless required to restrain an animal, the operator shall
stand at least 1.83 meters (6 feet) away from the useful beam and the animal
during radiographic exposures.
4) No individual, other than the operator, shall be in the x-ray
room or area while exposures are being made unless such individual's assistance
is required.
5) When an animal must be held in position during radiography,
mechanical supporting or restraining devices shall be used when technique
permits.
6) When a person is required to hold an animal during a
radiographic procedure, the individual shall be protected with appropriate
shielding devices, such as protective gloves and apron, and the person shall be
so positioned that no part of his/her body except hands and arms will be struck
by the useful beam.
AGENCY NOTE:
Veterinarians should review 32 Ill. Adm. Code 340.520 to determine if
individuals who hold animals will need to use individual monitoring devices.