32 Ill. Adm. Code 360.110
Therapy Systems Operating Below 1 MeV
Section 360
Section 360.110 Therapy
Systems Operating Below 1 MeV
In addition to the provisions of
Sections 360.10 through 360.30 of this Part, the requirements of this Section
apply to x-ray therapy systems and associated facilities operating at energies
less than 1 MeV.
a) Facility Design
1) A therapeutic radiological physicist shall be consulted in the
design of an x-ray therapy installation.
2) Shielding requirements
A) Each x-ray therapy installation shall be provided with such
primary and secondary barriers as are necessary to assure compliance with 32
Ill. Adm. Code 340.
B) For all x-ray therapy systems capable of operating above 150
kVp installed after October 15, 1993, facility design information shall be
submitted to the Agency for review prior to installation of the x-ray therapy
system. Information submitted to the Agency shall include, but need not be
limited to, the following:
i) Name and address of the planned installation.
ii) Name, address and telephone number of the therapeutic
radiological physicist who was consulted in the design of the installation.
iii) A scale drawing that includes the location of the therapy
system, control panel and doors to the room.
iv) The structural composition and thickness of all walls, doors,
partitions, floor and ceiling of the installation.
v) The occupancy of areas adjacent to the installation.
vi) Calculations that demonstrate the adequacy of the amount of
shielding specified for each primary and secondary protective barrier.
vii) Projected weekly dose rates in areas adjacent to the
installation.
3) Interlock. X-ray therapy systems operating at greater than
150 kVp shall have an interlock installed on each door of the therapy room. The
interlock shall be wired into the electrical circuit in such a manner that when
the door is opened, for any reason, the generation of x-rays will automatically
be terminated and irradiation can be resumed only by manually resetting the controls
on the control panel after the door is closed.
4) Doors. The doors to the therapy room shall be designed and
installed to allow opening from the inside at all times and shall be capable of
being opened manually.
5) Warning Lights. X-ray therapy systems operating above 150 kVp,
and all therapy rooms to which access is possible through more than one
entrance shall be provided with warning lights in a readily observable position
near the outside of all access doors. The warning lights shall indicate when
the useful beam is on.
6) Operator and control position
A) X-ray Therapy Systems Operating at 150 kVp and Below. The
control panel and operator shall be located either outside the therapy room or
behind a protective barrier within the room.
B) X-ray Therapy Systems Operating Above 150 kVp. The control
panel and operator shall be located outside the therapy room.
7) Viewing System. Windows, mirrors, closed-circuit television
or an equivalent system shall be provided to permit continuous visual
observation of the patient during irradiation and shall be located so that the
operator can observe the patient from the control panel.
AGENCY NOTE:
When the primary viewing system is electronic, a back-up system should be
available for use in the event of failure of the primary system in order to
ensure compliance with the requirements of subsection (e)(5) of this Section.
8) Communication. The facility design shall permit two-way aural
communications between the patient and the operator at the control panel.
9) Signs required by 32 Ill. Adm. Code 340.920 shall be posted in
the facility.
b) Equipment Requirements
1) Leakage Radiation. When the tube is operated at its maximum
rated continuous current for the maximum rated tube potential, the leakage
radiation shall not exceed the value specified in the table below at the
distance specified in the table for the classification of that x-ray system.
Radiation measurements shall be averaged over an area up to, but not exceeding,
100 square centimeters.
X-Ray
System
Leakage Limit
Measurement Location
Contact Therapy
25.8 microC/kg (0.1 R) per
hour
5 centimeters from the tube
housing
0 - 499 kVp
258 microC/kg (1 R) per hour
1 meter from the source
500 kVp - 999 kVp
0.1 percent of useful beam or
258 microC/kg (1 R) per hour, whichever is greater
1 meter from the source
2) Beam-Limiting Devices
A) Permanent fixed diaphragms or cones used for limiting the
useful beam shall provide the same or a higher degree of protection as required
for the tube housing assembly.
B) Removable beam-limiting devices shall, for the portion of the
useful beam to be blocked by these devices, transmit not more than one percent
of the useful beam at the maximum kilovoltage and maximum treatment filter.
This requirement does not apply to auxiliary blocks or materials placed in the
useful beam to shape the useful beam to the individual patient.
C) Adjustable beam-limiting devices installed after October 15,
1993 shall meet the requirements of subsection (b)(2)(B) of this Section.
D) Adjustable beam-limiting devices installed on or before October
15, 1993 shall, for the portion of the x-ray beam to be blocked by these
devices, transmit not more than five percent of the useful beam at the maximum
kilovoltage and maximum treatment filter.
3) Filter System. The filter system shall be designed so that:
A) The filters are securely positioned and will not become
dislodged when the machine is positioned at any possible orientation;
B) The radiation dose at one meter from the filter insertion slot
opening does not exceed 258 mC/kg (1 R) per hour when the machine is operated
at its maximum current and maximum tube potential;
C) Each filter is labeled with its composition and thickness (for
wedge filters, the wedge angle and maximum design field size shall appear on
the wedge or wedge tray);
D) If the x-ray therapy system uses changeable filters, there is a
filter indication system which permits recognition of any added filter in place
and indicates from the control panel the presence of a particular filter or
absence of any filter; and
E) For x-ray therapy systems installed after October15, 1993, an
interlock prevents irradiation if the selected filter is not installed.
4) Tube/Aperture Alignment. The x-ray tube shall be mounted so
that it cannot turn or slide with respect to the housing aperture.
5) Tube Housing Stability. The tube housing shall remain stable
during treatment unless tube housing movement is a designed function of the
system.
6) Source-Skin Distance (SSD) Indication
A) Means shall be provided to indicate the SSD.
B) The SSD shall be indicated in centimeters and/or inches and the
measured SSD shall correspond to the indicated value to within 0.5 percent.
7) Timer. A timer, which has a display at the control panel,
shall be provided and shall meet the following requirements:
A) The timer shall be activated with the production of radiation;
B) For systems equipped with a shutter mechanism to control
irradiation, the timer shall be activated when the shutter is opened;
C) The timer shall terminate irradiation when a preselected time
has elapsed;
D) The timer shall permit presetting and determination of exposure
times at least as short as 1 second; and
E) The timer shall not permit an exposure if the operator has not
selected a time for the exposure.
AGENCY NOTE:
The control panel should be equipped with a count-up timer to serve as a
back-up to the control timer.
8) Control Panel Functions. The control panel, in addition to
the displays required in other provisions of this Section, shall have:
A) An indication of whether x-rays are being produced;
B) A means for indicating x-ray tube potential and current; and
C) A means for terminating an exposure at any time.
9) Shutters. Equipment that is provided with shutters shall meet
the following requirements:
A) The shutters shall have a lead equivalency not less than that
of the tube housing assembly;
B) The shutter shall be controlled electrically by the operator at
the control panel; and
C) An indication of shutter position shall appear at the control
panel.
10) Multiple Tubes. Control panels capable of energizing more
than one x-ray tube shall meet the following requirements:
A) It shall be possible to energize only one x-ray tube at any
time;
B) There shall be an indication at the control panel identifying
which x-ray tube is energized; and
C) There shall be an indication at the tube housing assembly
when that tube is energized.
11) Low-Filtration X-Ray Tubes. Each x-ray therapy system equipped
with a beryllium window shall be clearly labeled as such upon the tube housing
assembly and at the control panel.
c) Radiation Protection Survey. A radiation protection survey
shall be performed by a therapeutic radiological physicist on each x-ray therapy
system. The registrant shall maintain at the facility a copy of the most recent
radiation protection survey report for review by the Agency. Radiation
protection surveys shall meet the following additional requirements:
1) X-ray therapy systems installed after October 15, 1993 shall
have a radiation protection survey performed by a physicist before the therapy
system is first used for irradiation of a patient.
2) For all x-ray therapy systems, a radiation protection survey
shall be performed by a physicist after any change in the x-ray therapy system
or facility that might produce a radiation hazard. The survey shall be
performed before the therapy system is used to treat patients.
3) Survey reports shall include, but need not be limited to, the
following:
A) A diagram of the facility that details building structures and
the position of the control panel, x-ray therapy system and associated
equipment;
B) A description of the x-ray therapy system, including the
manufacturer, model number and range of kilovolt potential;
C) A description of the instrumentation used to determine
radiation measurements, including the date and source of the most recent
calibration for each instrument used;
D) Conditions under which radiation measurements were taken; and
E) Survey data including:
i) Projected weekly dose equivalent in areas adjacent to the
therapy room; and
ii) A description of workload, use and occupancy factors employed
in determining the projected weekly dose equivalent.
4) The registrant shall retain a copy of the radiation protection
survey report and a copy of the report shall be provided to the Agency within
30 days after completion of the survey.
5) Any deficiencies detected during the radiation protection
survey that would constitute or result in a violation of 32 Ill. Adm. Code 340
shall be corrected prior to using the machine for treatment of patients.
6) The facility shall be operated in compliance with any
limitations indicated by the therapeutic radiological physicist as a result of
the radiation protection survey required by the Agency.
d) Calibrations and Quality Assurance Checks.
1) Each x-ray therapy system installed after October 15, 1993
shall be calibrated by a therapeutic radiological physicist before the therapy
system is first used for irradiation of a patient. The calibration of the x-ray
therapy system shall include, but need not be limited to, determination of the
following:
A) The radiation output, expressed as exposure rate in air or dose
rate in tissue, as a function of distance, field size, x-ray tube potential and
current, filters and treatment applicators used;
B) The half-value layer for each kilovoltage setting and filter
combination used;
C) The degree of congruence between the radiation field and the
field indicated by each beam-limiting device; and
D) An evaluation of the uniformity of the radiation field.
2) Quality assurance checks shall be made by a therapeutic
radiological physicist at intervals not to exceed 1 year. Quality assurance
checks shall include, but need not be limited to, determination of the
following:
A) The radiation output for a set of operating conditions
specified by the therapeutic radiological physicist;
B) The coincidence of the radiation field and the field indicated
by the beam-limiting device, except for systems equipped with fixed diaphragms
or cones; and
C) The therapeutic radiological physicist shall establish criteria
for quality assurance check measurements and shall determine corrective actions
to be implemented if the criteria are exceeded.
AGENCY NOTE:
Quality assurance checks should be performed at a frequency which is
appropriate for the particular therapy system, as determined by the therapeutic
radiological physicist and based on the history of stability of the radiation
output of the machine. A suggested frequency is one that would result in a
quality assurance check being performed at least once during a typical
patient's course of treatment.
3) Whenever service or maintenance is performed on the therapy
system, a therapeutic radiological physicist shall be notified and shall
determine whether a calibration or quality assurance check is necessary to
verify the characteristics of the beam.
4) Measurements of the radiation output of the x-ray therapy
system shall be performed using a dosimetry system that has been calibrated by
a calibration laboratory accredited by the American Association of Physicists
in Medicine (AAPM). Calibration of the dosimetry system shall have been
performed using a radiation beam of comparable half-value layer to the x-ray
system to be calibrated. The dosimetry system shall meet one of the two
conditions below:
A) The calibration of the dosimetry system shall have been
performed within the previous 2 years and after any servicing that may have
affected the calibration of the dosimetry system; or
B) The dosimetry system shall have been calibrated within the
previous 4 years and shall have been subjected to a protocol which provides for
checks of dosimetry constancy and provides for corrective action when results
deviate by more than two percent from the expected values.
5) The registrant shall maintain at the facility records of
machine calibrations, quality assurance checks and instrument calibrations for
inspection by the Agency for a period of 5 years. Records to be maintained by
the registrant shall include, but need not be limited to, the following:
A) Records of machine calibrations and quality assurance checks
shall include identification of the x-ray therapy system, radiation
measurements, the date the measurements were performed and the signature of the
therapeutic radiological physicist who performed the measurements.
B) Instrument calibration records shall include the date of the
last calibration and identity of the calibration laboratory. If a dosimetry
system has been subjected to a protocol as described in subsection (d)(4)(B) of
this Section, records shall be maintained that show the date and results of
each constancy check performed on the system.
e) Operating Procedures
1) No x-ray therapy system shall be left unattended unless the
system is secured against unauthorized use.
2) When a patient must be held in position for radiation therapy,
mechanical supporting or restraining devices shall be used.
3) Other than the patient, no individual shall be in the therapy
room unless such individual is protected by a barrier sufficient to meet the
requirements of 32 Ill. Adm. Code 340.
4) Other than the patient, no individual shall be in the therapy
room during exposures from x-ray therapy systems operating above 150 kVp.
5) The x-ray therapy system shall not be used for treatment of
patients unless the operator can maintain visual observation of the patient and
audible communication with the patient.
6) On contact therapy systems, a shield of at least 0.5
millimeter lead equivalency at 100 kVp shall be positioned over the entire
useful beam exit port during periods when the tube is energized and the beam is
not being used.
7) The tube housing assembly shall not be held by hand during
operating unless the x-ray therapy system is designed to require such holding
and the peak tube potential of the system does not exceed 50 kilovolts. In such
cases, the person holding the tube shall wear protective gloves and apron of
not less than 0.5 millimeter lead equivalency at 100 kVp.