32 Ill. Adm. Code 360.40
General Equipment and Operation Requirements for Diagnostic X-Ray Systems
Section 360
Section 360.40 General
Equipment and Operation Requirements for Diagnostic X-Ray Systems
The requirements of this Section
apply to all diagnostic x-ray systems. Additional requirements for specific
equipment application classes are in Sections 360.41 through 360.100 of this
Part.
a) Half-Value Layer
1) The half-value layer of the useful beam for a given x-ray tube
potential shall not be less than the values shown in Section 360. Table B of
this Part.
2) For capacitor energy storage equipment, compliance with the
requirements of this subsection (a) shall be determined with the system fully
charged and a setting of 10 mAs for each exposure.
b) Beam-On Indicators
1) The control panel shall include a device (usually a
milliammeter or labeled indicator lamp) which will give positive indication of
the production of x-rays whenever the x-ray tube is energized.
2) Where two or more radiographic tubes are controlled by one
exposure switch, the tube or tubes which have been selected shall be clearly
indicated prior to initiation of the exposure. This indication shall be both
on the x-ray control panel and at or near the tube housing assembly which has
been selected.
c) Mechanical Support of Tube Head. The tube housing assembly
supports shall be adjusted such that the tube housing assembly will remain
stable during an exposure unless tube housing movement is a designed function
of the x-ray system. The tube housing assembly supports shall not be hand-held
unless the manufacturer has specifically designed the system to be operated
while hand-held.
d) Diagnostic Source Assembly Leakage Radiation Limits. The
leakage radiation measured at a distance of 1 meter from the source shall not
exceed 25.8 microC/kg (100mR) in 1 hour when the tube is operated at its
leakage technique factors.
e) Radiation From Capacitor Energy Storage X-ray Equipment in
Standby Status. Radiation emitted from the x-ray tube when the exposure switch
or timer is not activated shall not exceed a rate of 0.516 microC/kg (2mR) per
hour at 5 centimeters from any accessible surface of the diagnostic source
assembly, with the beam-limiting device fully open.
f) Technique Indicators
1) The technique factors to be used during an exposure shall be
indicated at the control panel before the exposure begins. If automatic
exposure controls are used, the technique factors which are set prior to the
exposure shall be indicated at the control panel.
2) The requirement of subsection (f)(1) of this Section may be
met by permanent markings on equipment having fixed technique factors.
Indication of technique factors shall be visible from the operator's position
except in the case of spot films.
3) The indicated technique factors of exposure time and kilovolts
peak (kVp) shall correspond to the actual exposure factors within ten percent
of the indicated values.
g) Reproducibility of Exposures
1) For any specific combination of selected technique factors
utilized, the coefficient of variation of radiation exposures shall not exceed
0.05 for any specific combination of selected technique factors.
AGENCY NOTE: It will not be necessary to calculate the
coefficient of variation if for the first four measurements the value of the
average exposure (Eavg) is greater than or equal to ten times the maximum
exposure (Emax) minus the minimum exposure (Emin). This requirement is
mathematically represented by the following:
Eavg = 10 (Emax - Emin)
2) For systems using automatic exposure control (AEC) (i.e.,
systems employing photo-multiplier tubes, or ionization chambers to terminate
the x-ray exposure), compliance measurements shall be performed with the system
operating in the AEC mode. Attenuating material shall be placed in the beam to
provide exposure times in the range of those used clinically.
AGENCY NOTE: The intent of this subsection (g) is to require
testing of the system in a manner that is clinically relevant. Reproducibility
of exposures should be measured at technique factors that are commonly used and
are subject to variation. For AEC systems, commonly used settings in
combination with an appropriate thickness of attenuating material should be
used to provide exposure times in the clinical range.
h) Patient or Film Support
1) When a patient or film must be provided with auxiliary support
during a radiation exposure:
A) No person shall be used routinely to hold film or patients; and
B) Unless the procedure precludes their use, mechanical holding
devices shall be used to restrain patients. For example, mechanical holding
devices could not be used if the devices would preclude clear visualization of
the tissue being examined.
2) When a patient or film must be held by an individual, written
safety procedures, as required by Section 360.30(j) of this Part, shall
indicate the criteria for selecting a holder and the procedure the holder shall
follow.
AGENCY NOTE: The radiation dose received by radiation
workers, patients and the general public can be reduced if mechanical patient
and film support devices are used for radiographic and fluoroscopic
procedures. In the event that an individual must be used in lieu of mechanical
patient or film support devices to hold patients or films, every effort should
be made to limit the individual's radiation dose. This can be accomplished by
not assigning to a single individual the task of supporting patients and films
during radiographic and fluoroscopic examinations. Rather, a number of
individuals may be rotated through the assignment, thereby reducing the
radiation dose to one individual.
i) Personnel Protection
1) Except for patients who cannot be moved out of the room, only
the individuals required for the medical procedure or training shall be in the
room during the radiographic/flouroscopic exposure.
2) Individuals who must be in the room with the patient being
radiographed or fluoroscoped shall be protected by 0.25 millimeter lead
equivalent apparel or device or shall be positioned at a distance such that the
individual does not receive a radiation dose in excess of the limits specified
in 32 Ill. Adm. Code 340.310.
j) Technique Guides
1) In the vicinity of each radiographic x-ray system's control
panel, a technique guide shall be provided which specifies for routine
examinations performed with that system, the following information:
A) Patient's anatomical size versus technique factors to be
utilized;
B) Type of screen-film combination utilized, if more than one; and
C) SID to be used.
2) For automatic exposure control (AEC) systems with selectable
exposure detectors and density settings, the technique guide shall also specify
the appropriate exposure detectors and density setting to be utilized for each
radiographic examination listed.
3) For AEC systems, if operated in a non-automatic mode, the
technique guide shall specify the requirements of subsections (j)(1)(A) through
(C) of this Section to be followed.
AGENCY NOTE: The Agency recognizes that alternate means may
be available at the control panel to indicate technique factors for
computerized imaging systems.
k) Patient Dose Criteria. Procedures and auxiliary equipment
designed to minimize patient and occupational dose commensurate with needed
diagnostic information shall be used.
AGENCY NOTE: It is the intent of this subsection (k) to
provide for the optimum optical density, resolution and contrast on the film
while minimizing patient dose. X-ray films, intensifying screens and other
image recording devices should be as sensitive as is consistent with the
requirements of the examination.
l) X-ray Film Processing Systems. The darkroom safe light
illumination shall be adequate for the film speeds and the darkroom operating
procedures used to prevent fogging of unprocessed film. The following
additional requirements apply to film processing systems:
1) Manual film processing systems shall be monitored by the
registrant to assure:
A) The use of a dedicated darkroom timer with an adjustable preset
function. The timer shall be used to adjust film processing time according to
solution temperature.
B) The use of a dedicated darkroom thermometer. The thermometer
shall be used to adjust the film processing time according to solution
temperature.
C) The use of a film processing guide. The guide shall contain,
at a minimum, information regarding times and temperatures (as recommended by
the film manufacturer) used by the registrant to develop radiographs.
D) The frequency at which film processing chemicals are changed is
appropriate for the conditions of use.
2) Automated film processing shall be monitored by the registrant
to assure:
A) The temperature of film processing chemicals and the film
transport speed is appropriate for the type of films being utilized.
B) The film processing chemicals used and their replenishing rate
(if applicable) are appropriate for the type of films and quantity processed.
m) Gonadal Shielding. Except for cases in which it would
interfere with the diagnostic procedure, gonadal shielding of not less than 0.5
millimeter of lead equivalent shall be used for patients (who have not passed
the reproductive age) during those radiographic procedures in which the gonads
are in the useful beam.
AGENCY NOTE: Protection of the embryo or fetus from
radiation dose during radiological examination or treatment of a woman of
childbearing age (potentially pregnant) should be given special consideration.