32 Ill. Adm. Code 370.170
Mammography Units Used for Localization or Biopsy Procedures
Section 370
Section 370.170 Mammography
Units Used for Localization or Biopsy Procedures
a) Personnel. The following requirements apply to all personnel
involved in localization or biopsy procedures performed with mammography units:
1) The mammography unit shall be operated by or under the
direction of a physician licensed under the Medical Practice Act of 1987 [225
ILCS 60].
2) Radiologic technologists operating mammography units for
localization or biopsy procedures shall meet the general requirements,
mammography requirements and continuing education and experience requirements
as specified in Section 370.70(b) of this Part.
3) Medical physicists who perform and provide oversight of
quality assurance programs for mammography units used for biopsy procedures
shall meet the requirements of Section 370.70(c) of this Part.
b) Equipment. Mammography units used for localization or biopsy
procedures shall meet the requirements of Section 370.80 of this Part, except
that digital output mammography systems that do not use screen-film image
receptors are exempt from the requirements of Section 370.80 of this Part as
they relate to screen-film image receptors.
c) Quality assurance. Each facility shall establish and maintain
a quality assurance program to ensure the safety, reliability, clarity and
accuracy of mammography localization or biopsy procedures performed at the
facility.
1) Each facility shall have the services of a medical physicist
available to survey mammography equipment and to oversee the equipment-related
quality assurance practices of the facility.
2) The quality assurance program shall be in writing and shall
have been developed by a medical physicist. The program shall include, but
need not be limited to, the following:
A) Specifications of the tests that are to be performed, including
instructions to be employed in the performance of those tests; and
B) Specifications of the frequency at which tests are to be
performed, the acceptable tolerance for each parameter measured and actions to
be taken if tolerances are exceeded.
3) The medical physicist shall conduct a review of the quality
assurance program each year. Such review shall include evaluation of the
results of quality assurance testing.
d) Each facility shall maintain written records of the radiation
dose measurements and quality assurance testing performed, as required in this
Section, for inspection by the Agency for a period of at least one year. Such
records shall include, but need not be limited to, the following:
1) The date of the test and identification of the person
performing the test;
2) Identification of the type of testing that was performed; and
3) Notation of whether the results of the testing were within the
parameters established by the medical physicist.
AGENCY NOTE:
The Agency recommends that facilities performing interventional mammography
seek accreditation through the Stereotactic Breast Biopsy Program of the
American College of Radiology.