32 Ill. Adm. Code 370.APPENDIX A
A Mammography Dose Measurement Protocol
Section 370
Section 370.APPENDIX A Mammography
Dose Measurement Protocol
The technique factors used for
performing a mammography examination shall not permit the mean glandular
absorbed dose to exceed the limits specified in Section 370.110(e)(7) of this
Part. Radiation measurements shall be performed with an integrating radiation
measuring device that is appropriate to the high beam intensity and
mammographic kilovoltage peak (kVp) used, and sufficiently sensitive to
determine compliance with the criteria specified in Section 370.110(i)(2) of
this Part. The instrument shall have been calibrated as specified in Section
370.110(i)(2) of this Part.
The mammography exam dose limits
are based on an average compressed breast value of 4.2 centimeters having an
average density (i.e., 50 percent adipose and 50 percent glandular).
Perform the following steps to
determine the mean glandular dose to a nominal 4.2 centimeter compressed
breast:
a) Measure and record the x-ray system's useful beam half-value
layer (HVL). (See Section 370.110(e)(5) of this Part.) Any compression device
normally in the useful beam during mammography procedures shall be required to
be placed between the x-ray tube target and measuring device when determining
the HVL. The useful beam shall be collimated to a size encompassing the
detector.
AGENCY NOTE: Filters used for the HVL evaluation should be
placed as close to the target as practical. The HVL for screen-film
mammography should not exceed the minimum acceptable HVL by more than 0.1
millimeter of aluminum equivalent.
b) Determine the glandular dose to entrance exposure factor from
the Mammography Dose Evaluation Table (see Table A of this Part) using the
appropriate HVL, kVp and x-ray tube target-filter material.
AGENCY NOTE: The kVp of screen-film mammography systems with
molybdenum target-filter combinations should be accurately measured to
determine the appropriate glandular dose to entrance exposure factor from Table
A of this Part.
c) If the equipment has the capability for variable source-image
receptor distance, set the craniocaudal source-image receptor distance (SID)
for the image receptor system used.
d) Position in the useful beam any compression apparatus normally
used.
AGENCY NOTE: Some mammography systems have the capability of
providing automatic adjustment of technique factors through feedback from the
position of the compression device. On such systems, the compression device
should be lowered to a position 4.2 centimeters above the breast support
assembly (BSA). The device should then be removed, inverted and replaced to
allow placement of the phantom and measuring device on the BSA below the
compression device. If the compression device cannot be replaced in an
inverted position, the device should be placed in the beam using auxiliary
support.
e) Placement of the Radiation Measuring Device
1) For systems equipped with automatic exposure control (AEC):
A) Place a properly loaded film cassette in the cassette holder.
AGENCY NOTE: The loaded cassette is placed in the cassette
holder to simulate, as much as is possible, the conditions under which actual
patient exposures are made. Following radiation measurements, the film should
be discarded and the cassette reloaded with unexposed film.
B) Place a mammography phantom (see the definition for
"Phantom" in Section 370.20 of this Part) on the breast support
assembly (BSA). Align the phantom so that the edge of the phantom is aligned
with the chest wall side of the BSA and the phantom is over the automatic
exposure control device(s).
C) Place a radiation measuring device in the useful beam so the
center axis of the device is parallel to the breast support assembly (BSA).
The geometric center of the measuring device shall be positioned 4.5
centimeters above the BSA, 2.5 centimeters from the chest wall edge of the BSA
and immediately adjacent to either side of the mammography phantom.
2) For systems not equipped with AEC, place a radiation measuring
device in the useful beam so that the center axis of the device is parallel to
the breast support assembly (BSA). The geometric center of the measuring
device shall be positioned so that it is centered 4.5 centimeters above the
BSA, 2.5 centimeters from the chest wall edge of the BSA and at the center line
of the BSA. No part of the device's detector area shall be outside of the
useful beam.
f) Collimate the x-ray field to the size normally used and assure
that the area covered by the useful beam includes the detector area of the
radiation measuring device and the mammography phantom if the equipment is
equipped with automatic exposure.
g) Set the appropriate technique factors or automatic exposure
controls normally used for a nominal 4.2 centimeter compressed breast.
h) Measure and record the exposure in air with the radiation
measuring device.
i) Calculate the mean glandular dose for a 4.2 centimeter
compressed breast by multiplying the measured exposure in millicoulombs per
kilogram or in roentgens by the glandular dose to entrance exposure factor,
which was determined using the procedure described in subsection (b) of this
Section.
EXAMPLE: A mammography system is provided with a molybdenum
target-filter combination, and the HVL and kVp are determined to be 0.3 and 30,
respectively. Therefore, for a 4.2 centimeter compressed breast, the glandular
dose to entrance exposure factor from the Mammography Dose Evaluation Table
(Table A of this Part) would be 159 mrad. The measured roentgen output
determined in subsection (h) of this Appendix is determined to be 1.8 R.
Therefore, the mean glandular dose would be 1.8 R multiplied by 159 mrad/R.
This results in a mean glandular dose measurement of 286 mrad. As such, the
system would be in compliance with Section 370.110(e)(7) of this Part.