32 Ill. Adm. Code 390.40
Equipment Controls
Section 390
Section 390.40 Equipment
Controls
a) All meters and controls on the accelerator control console
shall be clearly identified and easily discernible. Accelerator control
consoles shall be equipped with a keyswitch or other device that will render
the console inoperative when the key or device is removed. Only one key shall
be available to the operating crew.
b) All entrances into a target room or other high radiation area
shall be provided with a minimum of two personnel interlocks.
c) The interlock system shall be designed to prevent restarting
of the accelerator without manually resetting the accelerator "ON"
switch at the control console after the tripping of a shielding interlock or a
power failure. At the time of such an occurrence, the registrant is required
to resurvey the radiation area prior to reactivation of the accelerator.
Records documenting the circumstances surrounding such occurrences shall be
maintained for review by the Agency.
d) A scram or panic button or other emergency power cut-off
switch shall be located and easily identifiable in all accessible high
radiation areas. A visible and/or audible signal system shall be installed
within the protective enclosure that will be activated for a reasonable length
of time before the power to the accelerator can be activated.
e) Electrical circuit diagrams of the accelerator and the
associated interlock system shall be kept current and on file at each
accelerator facility.
f) All safety and warning devices, including interlocks, shall be
checked and appropriately serviced each month. A log and written records of
these tests shall be kept by the registrant and made available for inspection
by the Agency.