77 Ill. Adm. Code 250.1325
Surgical Smoke Plume Evacuation System Equipment and Policies
Section 250.1325 Surgical Smoke Plume Evacuation System
Equipment and Policies
a)
To
protect patients and health care workers from the hazards of surgical smoke
plume, hospitals shall adopt policies to ensure the elimination of surgical
smoke plume by use of a surgical smoke plume evacuation system for each
procedure that generates surgical smoke plume from the use of energy-based
devices, including, but not limited to, electrosurgery and lasers.
(Section
6.32(b) of the Act)
1) The
facility's surgical department shall perform a risk assessment to identify all
procedures that are performed with energy-based surgical devices (e.g. lasers,
electrosurgical instruments, and ultrasonic devices) that generate a surgical
smoke plume and will require the use of a surgical smoke plume evacuation
system.
2) All
surgical team members shall be trained on the methods for mitigating the
hazards and minimizing exposure to surgical smoke plume, positioning and
operating surgical smoke plume evacuation pursuant to the manufacturer's
instructions, and the requirements in facility policies and procedures for
management of surgical smoke plume.
3) Staff
shall wear appropriate respiratory protection when needed as secondary
protection against residual smoke in accordance with the hospital's respiratory
protection plan.
4) To
protect against potential smoke hazards, the facility’s policy and procedure
shall minimally include:
A) During
utilization of the smoke evacuator, the suction nozzle inlet shall be positioned
as close to the surgical site as possible to maximize capture of airborne
contaminants.
B) The
smoke evacuator shall be “ON” (activated) at all times when airborne particles
are produced during all surgical or other procedures
C) New
tubing shall be used before each procedure and the smoke evacuator filter shall
be replaced as recommended by the manufacturer. Consider all tubing, filters
and absorbers as infectious waste and dispose of appropriately in accordance
with OSHA bloodborne pathogens standards.
D) Inspection
of smoke evacuator systems regularly, including inspection immediately prior to
use, to ensure proper functioning.
b)
The
hospital shall report to the Department that policies
required
under
subsection (a)
have been adopted
. The hospital shall provide the
Department a letter identifying the date of the adoption of the facility's
policy for utilization of smoke evacuation system. (Section 6.32(c) of the Act)