77 Ill. Adm. Code 890.740
Kidney Dialysis Machines
Section 890
Section 890.740 Kidney
Dialysis Machines
a) The water supply inlet to kidney dialysis equipment shall have
a reduced pressure principle backflow preventer assembly complying with ASSE
1013 or a fixed air gap.
1) A portable dialysis unit or machine shall have a reduced
pressure principle backflow preventer assembly installed on the water supply
inlet on the unit.
2) Stationary dialysis equipment within a facility shall require,
at the filter room or the dialysis machines, a reduced pressure principle
backflow preventer assembly on the water supply or a water supply with a fixed
air gap.
3) Dialysis equipment shall be installed in accordance with this
Part and the manufacturer's specifications. Any conflicts shall be submitted
to the Department for resolution.
b) The water supply to a dialysis reuse room or dialysis machine
repair room shall be isolated from all other deionized (DI) or reverse osmosis
(RO) water lines by an RPZ or an air gap.
c) A sign no smaller than 8 by 10 inches with the wording
"This Water For Dialysis Only" shall be placed above a sink with DI
water or RO water supplied to the faucet.
d) The discharge for each dialysis unit or machine, portable or
stationary, shall be provided with an individual indirect waste connection to
the sanitary drainage system. Each stand pipe shall be individually trapped
and vented.