105 CMR 120.319
Permanent Radiographic Installations
(A) Permanent radiographic installations shall have high radiation area entrance controls of the
type described in 105 CMR 120.227(A)(2) and (3) and (B).
(B) Each entrance that is used for personnel access to the high radiation area shall have both
conspicuous visible and audible warning signals to warn of the presence of radiation. The visible
signal shall be activated by radiation. The audible signal shall be activated when an attempt is
made to enter the installation while the source is exposed.
(C) The control device or alarm system shall be tested for proper operation with a source of
radiation at the beginning of each day of equipment use. The test shall include a check for the
visible and/or audible signals. Entrance control devices that reduce the radiation level upon entry
as described in 105 CMR 120.227(A)(1) shall be tested monthly. If a control device or alarm
system is operating improperly, it shall be immediately labeled as defective and repaired within
seven calendar days. The facility may continue to be used during this seven-day period, provided
the licensee or registrant implements the continuous surveillance requirements of 105 CMR
120.331, ensures that radiographic personnel use an alarming ratemeter, and complies with the
requirements of 105 CMR 120.330(B). Records of these tests shall be maintained for Agency
inspection for five years.