105 CMR 120.645
Shielding
(A) The radiation dose rate in areas that are normally occupied during operation of a panoramic
irradiator may not exceed 0.02 millisievert (two millirems) per hour at any location 30
centimeters or more from the wall of the room when the sources are exposed. The dose rate must
be averaged over an area not to exceed 100 square centimeters having no linear dimension
greater than 20 centimeters. Areas where the radiation dose rate exceeds 0.02 millisievert (two
millirems) per hour must be locked, roped off, or posted.
(B) The radiation dose at 30 centimeters over the edge of the pool of a pool irradiator may not
exceed 0.02 millisievert (two millirems) per hour when the sources are in the fully shielded
position.
(C) The radiation dose rate a one meter from the shield of a dry-source-storage panoramic
irradiator when the source is shielded may not exceed 0.02 millisievert (two millirems) per hour
and at five centimeters from the shield may not exceed 0.2 millisievert (20 millirems) per hour.