Rule 312. (1) The tube housing shall be of the therapeutic type.
(2) Permanent diaphragms or cones used for collimating the useful beam shall afford
the same degree of attenuation as is required of the housing.
(3) Adjustable or removable beam-limiting devices shall transmit not more than 5%
of the useful beam as determined at the maximum tube potential and with maximum
treatment filter.
(4) Filters shall be so mounted as to prevent their movement during the treatment.
(5) The filter slot shall be so constructed that the radiation escaping through it does
not produce an exposure rate exceeding 1 R/h at 1 meter, or if the patient is likely to be
exposed to radiation escaping from the slot, 30 R/h at 5 centimeters (2 inches) from the
external opening.
(6) A removable filter shall be permanently marked with its thickness and material.
(7) A filter indication system shall be used on therapy machines which use
changeable filters. It shall indicate, from the control panel, the presence or absence of a
filter and it shall be designed to permit easy recognition of the filter in place.
(8) The x-ray tube shall be so mounted that it cannot turn or slide with respect to the
housing aperture. A reproducible means of measuring the focal spot to patient distance
shall be provided.
(9) Means to immobilize the tube housing during stationary portal treatment shall be
provided.
(10) An easily discernible indicator which shows whether or not x-rays are being
produced shall be on the control panel.
(11) Beam monitoring devices shall be fixed in the useful beam to indicate an error
due to incorrect filter, tube current, or tube potential, unless the device introduces more
filtration than is clinically acceptable.
(12) A suitable exposure control device, such as an automatic timer, exposure meter,
or dose meter, shall be provided to terminate the exposure after a preset time interval or
preset exposure or dose limit. If a timer is used, it shall permit accurate presetting and
determination of exposure times as short as 1 second. Means for the operator to
terminate the exposure at any time shall be provided.
(13) Mechanical or electrical stops, or both shall be provided to insure that the useful
beam is oriented only toward primary barriers.
(14) Interlocks shall be provided so that, when a door to the treatment room is
opened, the machine turns off automatically or the radiation level within the room is
reduced to an average of not more than 2 mR/h and a maximum of 10 mR/h at a distance
of 1 meter in any direction from the source. After the shut-off or reduction in exposure
rate, it shall be possible to restore the machine to full operation only from the control
panel.
(15) The x-ray control circuit shall be so designed that it is not possible to energize
the x-ray tube to produce x-rays without resetting the x-ray "ON-OFF" switch at the
control panel.
(16) When the relationship between the beam interceptor (when present) and the
useful beam is not permanently fixed, mechanical or electrical stops shall be provided to
ensure that the beam is oriented only toward primary barriers.
(17) X-ray machines with electron beam extraction capability shall be provided with
such additional safety devices as determined necessary and specified in writing by the
department to prevent accidental electron beam exposure.
(18) To reduce the electron contamination of high energy treatment beams, shadow
trays, or other accessories placed in the primary beam shall be placed at a sufficient
distance from the patient that the electron contamination contribution to the skin dose is
minimal.