R 333.5315 Enclosures.
Cite as Mich. Admin. Code R 333.5315
Rule 315. (1) An enclosure shall be a permanent part of the building or equipment.
Portable protective barriers shall not be used for permanent installations.
(2) The degree of protection required for an enclosure shall be determined by the
workload, use and occupancy factors, and the tube potential, tube current, mechanical
movement, and distance. The design shall be subject to approval by the department.
(3) All wall, ceiling, and floor areas that can be irradiated by the useful beam plus an
additional area extending at least 30 centimeters (1 foot) beyond shall be provided with a
primary protective barrier.
(4) For equipment capable of operating above 150 kVp, the control station shall be
outside of the therapy room.
(5) The enclosure shall be so constructed that individuals may at all times be able to
escape from within.
(6) If the radiation exposure rate within the therapy room is so high that an individual
who is accidentally in the treatment room when the machine is turned "ON" may receive
as much as 1250 mR exposure during the time required to reach an access door, special
cut-off, or panic buttons shall be required. When pressed, these buttons, operable by
open hand at appropriate positions about the treatment room, shall cause the irradiation to
be terminated.
(7) Effective means shall be provided to prevent access to the treatment room during
exposure. For equipment capable of operating above 150 kVp, each access door to the
treatment room shall be provided with a fail-safe interlock. The interlock system shall be
so designed that the failure of any 1 component cannot jeopardize the safety of the
system, such as the use of series connected double switch assemblies at access doors, and
dual interlock relays. If an access door is opened when the machine is "ON", the
interlock shall cause termination or reduction of exposure as specified in R 333.5312(14).
(8) Red warning signal lights, energized only when the useful beam in "ON", shall be
located on the control panel and near each entrance to the therapy room. Under
conditions as specified in subrule (6) of this rule a visible signal shall also be located
within the therapy room. Depending upon control panel and door locations, a single
warning signal light may be sufficient to comply with this subrule.