R 333.5317
R 333.5317 Conditions of operation.
Cite as Mich. Admin. Code R 333.5317
Rule 317. (1) An installation shall be operated in compliance with all limitations
determined necessary and specified in writing by the department.
(2) The output of the x-ray generator shall be calibrated before use for the treatment
of patients for each technique or condition of use. The department shall be informed by
telephone or in writing of completion of initial calibration before patient treatment is
initiated. A written report of this initial calibration shall be submitted within 30 days to
the department. Recalibration shall be required after each tube replacement and after any
changes or replacement in the generating apparatus which could affect a change in the x-
ray output. Check calibrations shall be made on an annual basis and records of all
calibration maintained for not less than 7 years.
(3) X-ray therapy equipment capable of operating above 150 kVp shall not be
operated routinely until the radiation safety of the installation has been established by a
protection survey conducted pursuant to R 333.5063. The department shall be informed
by telephone or in writing of completion of the initial survey before patient treatment is
initiated. A written report of this initial survey shall be submitted within 30 days to the
department. All x-ray therapy equipment shall be operated in conformance with
recommendations of the protection survey.
(4) Both the control panel and the patient shall be observable during exposure.
(5) When a patient must be held in position for radiation therapy, mechanical
supporting or restraining devices should be used. If the patient must be held by an
individual, upon approval by the radiologist in charge followed by written notice to the
department, that individual shall be provided protection equivalent to 7 half-value layers
and shall be positioned so that no part of his or her body can be struck by the useful beam
and is as far as possible from the edge of the useful beam. The exposure of an individual
used for this purpose shall be monitored and a permanent record maintained. The
individual selected for this purpose shall not otherwise be occupationally exposed to
ionizing radiation.
(6) With the exception of subrule (5) of this rule, an individual other than the patient
shall not be permitted in the treatment room when the tube is operated at potentials
exceeding 85 kVp. At potentials of 85 kVp or below, other individuals may be permitted
in the treatment room by the radiologist in charge if they are essential to conduct the
treatment, but only if they are protected as specified in subrule (5) of this rule and their
radiation exposure is monitored and permanently recorded.
(7) Personnel monitoring shall be performed in controlled areas for each individual
occupationally exposed to ionizing radiation from therapeutic x-ray equipment.
Individual monitoring devices, such as film badge dosimeters or thermoluminescent
dosimeters, shall be permanently assigned to each occupationally exposed individual.
This monitoring shall be continuous during employment as a radiation worker.
(8) Personnel exposure records shall be kept on permanent available file at the facility
where the exposure occurs.
(9) Monitoring devices used to estimate whole body exposure shall normally be worn
on the chest or abdomen. Monitoring of other body parts shall comply with R 333.5065.
(10) Monitoring devices worn to estimate personnel occupational exposure shall not
be worn by the individual when he or she is exposed as a patient for a medical or dental
reason.
(11) Lead, lead rubber, lead foil, and similar materials used for limiting the field shall
not transmit more than 5% of the useful beam under the conditions at which the machine
is operated for therapy. This subrule does not apply to treatment blocks used to adjust or
modify the intended radiation dose to the area of treatment.
(12) A therapeutic x-ray system shall not be left unattended without locking the
apparatus, room, or building in some manner which prevents use of the apparatus by
unauthorized persons.
THERAPEUTIC MACHINES OPERATED AT OR BELOW 85 KVP