R 333.5321 X-ray equipment.

Last amended: 2016Year: 2026Length: 99 wordsOfficial source

Cite as Mich. Admin. Code R 333.5321

Rule 321. (1) The x-ray equipment shall comply with the requirements of R 333.5312, excluding subrules (11), (14), and (16). (2) Maximum potential shall be mechanically or electronically limited to 85 kVp. (3) A contact therapy machine shall meet the additional requirement that the leakage radiation at 5 centimeters (2 inches) from the surface of the tube housing shall not exceed 0.1 R/h. As used in this subrule, "contact therapy machine" means an x-ray therapy machine designed for source to skin treatment distances of 5 centimeters or less at tube potentials in the range of 20 to 50 kVp.