19 CSR 20-10.120
General Requirements for Diagnostic X-ray Equipment
PURPOSE: This rule establishes general
requirements for diagnostic X-ray equipment.
(1) The X-ray tube housing shall be of the
protective diagnostic type.
(2) Total filtration shall be as follows:
(A) For fluoroscopic and radiographic
equipment operating at seventy (70) kilovolts
peak (kvp) and below, the total filtration permanently in the useful beam, for routine use,
shall be equivalent to at least 1.5 millimeters
(mm) of aluminum. This condition shall be
considered fulfilled if the half value layer
(HVL) of the useful beam is 1.5 mm of aluminum or greater.
(B) For fluoroscopic and radiographic
equipment capable of operating above seventy (70) kvp, the total filtration permanently in
the useful beam shall be equivalent to at least
2.5 mm of aluminum. This condition shall be
considered fulfilled if the HVL of the useful
beam is 2.5 mm of aluminum or greater.
(3) The exposure switch for routine diagnostic X-rays or routine diagnostic fluoroscopy
shall be of the dead-man type or time limiting switch acceptable to the department.
(4) Diaphragms or cones shall be used for
collimating the useful beam and shall provide
the same degree of protection as the tube
housing.
AUTHORITY: section 192.420, RSMo 1986.*
This rule was previously filed as 13 CSR 5090.120. Original rule file Nov. 9, 1964, effective Dec. 9, 1964.
*Original authority: 192.420, RSMo 1963.