19 CSR 20-10.130
Special Requirements for Medical Fluoroscopic Installations
PURPOSE: This rule establishes special
requirements for medical fluoroscopic installations.
(1) All fluoroscopes shall be so constructed
that the entire cross-section of the useful
beam is attenuated by a primary barrier. Collimators and adjustable diaphragms or shutters shall be provided to restrict the size of
the useful beam to less than the area of the
barrier. For conventional fluoroscopes, this
requirement is met if, when the adjustable
diaphragm is open to its fullest extent, an
unilluminated two centimeter (2 cm) margin
is left on the fluorescent screen with the
screen centered in the beam at a distance of
thirty-five centimeters (35 cm) (fourteen
inches (14")) from the panel or table top. The
margin requirement does not apply to installations where image amplifiers are used but a
protective shield shall be provided in these
installations so that the useful beam does not
produce a radiation hazard.
(2) Accessory shielding devices (such as
cones, curtains, bucky slot covers, shielding
between patient and fluoroscopist) shall be
used as required to reduce the exposure dose
rate to not more than fifty (50) milliroentgens
per hour in the area adjacent to the machine
normally occupied by the fluoroscopist and
his/her assistants during fluoroscopy.
(3) The target to panel distance shall not be
less than eighteen inches (18").
(4) A manually reset cumulative timing
device shall be used which will either indicate elapsed time by an audible signal or turn
off the apparatus when the total exposure
exceeds a predetermined limit given in one
(1) or a series of exposures. The device shall
have a maximum range of five (5) minutes.
(5) The primary beam barrier should have a
lead equivalent of at least 2.0 millimeters
(mm) and shall not be less than 1.5 mm for
one hundred (100) kvp, should be at least 2.4
mm and shall be not less than 1.8 mm for one
hundred twenty-five (125) kvp, and should be
at least 2.7 mm and shall be not less than 2.0
mm for one hundred fifty (150) kvp. For conventional fluoroscopes, this requirement may
be assumed to have been met if the exposuredose rate measured at the viewing surface of
the fluorescent screen does not exceed fifty
(50) milliroentgens per hour with the screen
in the primary beam of the fluoroscope without a patient, under normal operating conditions.
(6) Collimators and adjustable diaphragms or
shutters to restrict the size of the useful beam
shall provide a minimum of 2.0 mm lead
equivalent protection for one hundred (100)
kvp, 2.4 mm for one hundred twenty-five
(125) kvp and 2.7 mm for one hundred fifty
(150) kvp.
(7) For routine fluoroscopy, the dose rate
measured at the panel or table top shall not
exceed six (6) roentgens per minute.
(8) Mobile fluoroscopic equipment shall meet
the requirements of sections (1)–(7) of this
rule except that—
(A) In the absence of panel or table top, a
cone or spacer frame hall limit the target-toskin distance to not less than twelve inches
(12");
(B) It shall be impossible to operate the
machine when the collimating cone or
diaphragm is not in place; and
(C) The maximum permissible exposuredose rate of six (6) roentgens per minute shall
be measured at the minimum target-to-skin
distance.
AUTHORITY: section 192.420, RSMo 1986.*