19 CSR 20-10.150
Special Requirements for Dental Radiographic Installations
PURPOSE: This rule establishes special
requirements for dental radiographic installations.
(1) The diameter of the useful beam at the
cone tip should be not more than 2.75 inches
and shall be not more than three inches (3")
for routine intraoral radiography.
(2) A device shall be provided to terminate
the exposure after a preset time or exposure.
(3) The exposure control switch shall be provided with a cord sufficiently long so that the
operator can stand at least six feet (6') from
the tube housing for all exposures and well
away from the useful beam.
AUTHORITY: section 192.420, RSMo 1986.*
This rule was previously filed as 13 CSR 5090.150. Original rule filed Nov. 9, 1964,
effective Dec. 9, 1964.
*Original authority: 192.420, RSMo 1963.